WatchBP Office Central

EN 3

Cardiovascular Screening Solution for:

-Central blood pressure measurement

-Ankle Brachial Index assessment

-Atrial fibrillation detection

-Simultaneous assessment of the Inter-Arm

Blood Pressure Difference

Instruction Manual

WatchBP Office Central

Page 2: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

Preface

The WatchBP Office Central has been validated against simultaneous recorded intra-arterial blood pressure measurement, which is the

gold standard for validation of central blood pressure measurement, and showed extreme high accuracy in measuring central systolic,

central diastolic and central pulse pressure 1. The WatchBP central has passed the validation protocol of the BHS for accuracy in blood

pressure measurement (A/A) and is equipped with other innovative screening tools such as:

• An implemented atrial fibrillation (Afib) detection system that has proven excellent accuracy 2, 3 and is officially recommendation by

the National Institute of Clinical health and Excellence (NICE) in the UK 4.

• Two arm cuffs that allow simultaneous blood pressure measurement of both arms for accurate determination of inter-arm BP

differences 5, 6. This mode is highly supported by the latest ESH guidelines (2013) stating that Inter-arm difference in blood pressure is

only meaningful when obtained from simultaneous arm blood pressure measurement 7.

• Ankle-brachial index (ABI) performance that is validated against Doppler for accuracy with positive results and additionally showed

that the assessment led to a significant time reduction as compared to Doppler 8. Finally, the automated calculation of the ABI value

guarantees uniformity and prevents errors due to registration and calculation failures 9.1. Cheng, H.M., et al., Measurement Accuracy of a Stand-Alone Oscillometric Central Blood Pressure Monitor: A Validation Report for Microlife WatchBP Office Central. Am J Hypertens.

2013. 26: 42-50.

2. Stergiou, G.S., et al., Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation. J Hum Hypertens, 2009. 23: 654-8.

3. Wiesel, J., et al., Detection of atrial fibrillation using a modified microlife blood pressure monitor. Am J Hypertens, 2009. 22: 848-52.

4. NICE, WatchBP Home A for opportunistically detecting atrial fibrillation during diagnosis and monitoring of hypertension. http://guidance.nice.org.uk/MTG13, 2013.

5. Lohmann, F.W., et al., Interarm differences in blood pressure should be determined by measuring both arms simultaneously with an automatic oscillometric device. Blood Press Monit, 2011. 16: 37-42.

6. Stergiou, G.S., et al., Automated device that complies with current guidelines for office blood pressure measurement: design and pilot application study of the Microlife WatchBP Office device. Blood Press Monit, 2008. 13: 231-5.

7. Mancia, G., et al., 2013 ESH/ESC Guidelines for the management of arterial hypertension: The Task Force for the management of arterial hypertension of the European Society of Hypertension (ESH) and of the European Society of Cardiology (ESC). J of Hypertens. 2013, 31:1281–1357.

8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood pressure monitor: validation vs. Doppler measurement and cardiovascular risk factor profile. Hypertens Res, 2011. 34: 825-30.

9. Verberk, W.J., et al., Automated oscillometric determination of the ankle-brachial index: a systematic review and meta-analysis. Hypertens Res, 2012. 35: 883-91.

Page 3: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

EN

WatchBP Office Central

Product descriptionName of parts and display ..................................6-7 Components and accessories ................................. 8

Before using WatchBP Office CentralSelecting and fitting the correct cuff ...................... 9

Selective Cuffs for upper arm and ankle ..............10

Application of the ankle cuff ................................ 11

Three operation modes«ABI» Mode .......................................................... 12 «CENTRAL» Mode .................................................13 «SCREEN» Mode ............................................. 14–15 Inter-arm difference (IAD) ................................... 15

Taking measurements using WatchBP Office Central«SCREEN» Mode ............................................. 16–17 Taking less than three measurements ................ 17

«CENTRAL» Mode ........................................... 18–19 Skipping the countdown time ..............................20 Setting measurement intervals ...........................20 «ABI» Mode .................................................... 21–22

Special FunctionsCentral blood pressure ........................................23Central pulse pressure ..........................................23Mean arterial pressure and pulse pressure ......... 24Appearance of the atrial fibrillation indicator for early detection ............................................................. 25About Atrial Fibrillation .......................................26 ABI (Ankle Brachial Index) .................................. 27Hiding measurement values ................................ 28Viewing and transferring Measurements«SCREEN» Mode ................................................... 29 «CENTRAL» Mode ................................................30 «ABI» Mode ..........................................................31

Transferring measurements ................................32

AppendixRechargeable battery and power adaptor ..........33 Troubleshooting ...................................................34 Error messages ............................................... 35–36 Safety, care, accuracy test and disposal ......... 38–39 Technical specifications .......................................40

Table of Contents

Page 4: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

6

L

R

AUSCULT.ROUTINE

SCREEN

ABI CENTRAL SCREEN

MAPIAD

I/O switch

Main Adaptor Socket

Battery Compartment

Cuff Socket (L)

Cuff Socket (R)

Recharge Indicator

Display

M Button (Memory)

Mode Switch

Start/Stop Button

DC 7.5V 2.0ADC 7.5V 2.0A

Product description

Name of parts

USB Port

Page 5: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

7

WatchBP Office Central

EN

Mean Arterial Pressure (MAP)

Pulse Pressure (PP)

Mode Indication

External Power

Atrial Fibrillation Detection

Central Systolic/ Pulse Pressure

L/R Cuff

Supine Position Reminder

Hide Display

Memory Mode

ABI indicator and Pulse Rate

Battery Symbol

Value for Left Arm

Systolic Value

Diastolic Value

Inter-arm Difference

Value for Right Arm

Systolic Value

Diastolic Value

Display

Page 6: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

8

Components and accessories

WatchBP Office Central

Blood Pressure Device

AC Adaptor UES18LCP-075200SPA X1

(Input: 100-240V~50/60Hz 0.5A

Output: +7.5V 2A)

Cuff for upper arm

M-L size (22cm~42cm) X 2

Instruction Manual Quick Start Guide

AUSCULT.ROUTINE

SCREEN

ABI CENTRAL SCREEN

MAPIAD

AUSCULT.ROUTINE

SCREEN

ABI ROUTINE SCREEN

MAPIAD

L

R

ROUTINE AUSCULT. IAD

I/O switch

Main Adaptor Socket

Battery Compartment

Cuff Socket (L)

Cuff Socket (R)

DC 7.5V 2.0A

Product description

Quick start guide

Flip over the quick start guide to see ‘How to take blood pressure measurement in three different modes’.

Printed 2009 August

Artery Mark

Cuff Connector

Recharge Indicator

Start/Stop Button

Mode Switch

M Button (Memory)

AUSCULT.ROUTINE

SCREEN

ABI ROUTINE SCREEN

MAPIAD

INDE

X LI

NE

22-32 cm

8.7-12.6 inches

M size

WatchBP

2-3cm 2-3cm

LEFTRIGHT

OK ARTERY MARK

Artery Mark

Instruction Manual

Bedienungsanleitung

Mode d’emploi

Manuale di istruzioni

EN 3 DE 43 FR 83 IT 123

Cuff for ankle

M size (22cm~32cm) X 1

INDEX LINE 8 LE

F

USB Cable

Page 7: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

EN9

WatchBP Office Central

M-L (Medium - Large size), arm cuff22 - 42 cm (8.7 - 16.5 inches) With air tube 130 cm

M (Medium size), ankle cuff22-32 cm (12.6 - 16.5 inches) With air tube 130 cm

Before using WatchBP Office Central

Selecting the correct cuff Two upper arm cuffs and an ankle cuff are provided within the standard package of the WatchBP Office Central.

Fitting the cuff properly1) Place the cuff over the left (or right) upper

arm so that the air tube and artery mark arrow point toward the lower arm.

2) Wrap the cuff around the arm. Make sure that the lower edge of the cuff lies approximately 2cm ( ¾ inch) above the elbow.

* Remove all clothing covering or constricting the measurement arm before applying the cuff.

Connect the cuff to the device by inserting the cuff connector into the cuff connector socket.

R

R

INDEX LINE

8 LEF

Page 8: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

10

Selective Cuffs for upper arm and ankle

M - L (Medium - Large size, standard delivered)22 - 42 cm (8.7 - 16.5 inches) With air tube 130 cm

* Please contact Microlife or its distributor to purchase L-XL size or other size cuffs.

INDE

X LI

NE

22-32 cm

8.7-12.6 inches

M size

WatchBP

2-3cm 2-3cm

LEFTRIGHT

OK ARTERY MARK

INDE

X LI

NE

WatchBP

2-3cm 2-3cm

LEFTRIGHT

OK ARTERY MARK

32 - 42 cm

12.6 - 16.5 inches

L size

M (Medium size, standard delivered) 22 - 32 cm (8.7 - 12.6 inches) With air tube 200 cm

L (Large size, to be ordered) 32 - 42 cm (12.6 - 16.5 inches) With air tube 130 cm

M (Medium size, to be ordered) 22 - 32 cm (8.7 - 12.6 inches) With air tube 130 cm

L-XL (Large to extra-large size, to be ordered)32 - 52 cm (12.6-20.5 inches) With air tube 130 cm

For upper arm

For ankle

INDE

X LI

NE

22-32 cm

8.7-12.6 inches

M size

WatchBP

2-3cm 2-3cm

LEFTRIGHT

OK ARTERY MARK

INDE

X LI

NE

WatchBP

2-3cm 2-3cm

LEFTRIGHT

OK ARTERY MARK

32 - 42 cm

12.6 - 16.5 inches

L size

L (Large size, to be ordered) 32 - 42 cm (12.6 - 16.5 inches) With air tube 200 cm

Rigid cuff Soft cuffs

Page 9: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

EN11

WatchBP Office Central

Application of the ankle cuff (in ABI mode)1) The patient has to lie down in supine position.

2) Place the ankle cuff on the leg. Make sure the edge of the ankle cuff lies approximately 2 to 3cm ( ¾ to 1 inch) above the ankle and notice that the artery mark is on the posterior tibial artery.

INDEX LINE

8

LEF

INDEX LINE

8

LEF

INDEX LINE

8

LEF

INDEX LINE

8

LEF

posterior tibial artery.

2-3 cm (3/4 to 1 inch)

3) Wrap and tighten the cuff around the leg.

4) Leave a little free space between the leg of the patient and the cuff. Two fingers should fit between the leg and the cuff.

INDEX LINE

8

LEF

INDEX LINE

8

LEF

Page 10: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

12

Three operation modes

AUSCULT.ROUTINE

SCREEN

ABI CENTRAL SCREEN

MAPIAD

«ABI» ModeSelect «ABI» Mode for Ankle-Brachial measurement. Select the arm with the higher blood pressure value as determined with the «SCREEN» Mode

Select an operation modeThe WatchBP Office Central device includes three operation modes: «ABI», «CENTRAL», and «SCREEN» (Inter-arm difference). Use the Mode Switch to select the desired mode.

ABI CENTRAL SCREEN ABI CENTRAL SCREEN ABI CENTRAL SCREEN ABI CENTRAL SCREEN ABI CENTRAL SCREEN ABI CENTRAL SCREEN

Page 11: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

EN13

WatchBP Office Central

ABI CENTRAL SCREEN ABI CENTRAL SCREEN ABI CENTRAL SCREEN

«CENTRAL» ModeSelect «CENTRAL» Mode to perform central blood pressure measurements on the preferred arm for prompt and accurate office measurements.

Automated double measurements

In «CENTRAL» Mode, the WatchBP Office Central device automatically takes two consecutive

* The user can manually set measurement intervals of 15, 30, 45 or 60 seconds in CENTRAL Mode. (Please refer to special functions section page 19 “Setting measurement intervals”)

1 15 seconds

2

measurements at 15 second intervals on default. At each measurement brachial blood pressure is measured first, thereafter the cuff pressure will be hold around 60 mmHg for 30 seconds for assessing the central blood pressure value. the readings of the two brachial blood pressure values are averaged and displayed and the readings of the two central blood pressure values are averaged and displayed after the measurement process is finished. For verifying each measurement separately press the M button to see the 2nd measurement, press the M-button again to see the 1st measurement.

Page 12: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

14

Three operation modes (cont.)

Simultaneous dual-arm measurements

In «SCREEN» Mode, the WatchBP Office Central device measures the patient’s blood pressure on both arms simultaneously, for determining the arm with the highest blood pressure value and revealing the possible presence of peripheral arterial disease.

«SCREEN» ModeSelect «SCREEN» Mode to complete fully-automated triple measurements on both arms according to recommended ESH/AHA blood pressure measurement protocols for a patient’s first office visit.

ABI CENTRAL SCREEN ABI CENTRAL SCREEN ABI CENTRAL SCREEN

Page 13: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

EN15

WatchBP Office Central

Automated triple measurements

In «SCREEN» Mode, the WatchBP Office Central device automatically takes three consecutive measurements on both arms at fixed one minute* intervals. The results of these three measurements are then averaged to conclude the blood pressure measurement.

* The measuring intervals are fixed at one minute in SCREEN mode. The interval time can be shortened by pressing the on/off button during the countdown period.

* The arm with the higher blood pressure value should be taken for ABI measurement or future blood pressure measurements.

Determine the proper arm and inter-arm difference

If the difference of blood pressure readings between two arms is more than 20mmHg for Systolic or 10mmHg for Diastolic at all three measurements, the device indicates the arm with the higher blood pressure value by displaying L or R and the “IAD” icon. The “IAD” icon and readings of the higher arm flash to indicate that more attention is needed as this patient might have peripheral arterial disease. .

IAD SCREEN

IAD

L

60 seconds

60 seconds

1 2 3

Page 14: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

16

«SCREEN» Mode1) Turn on the power – Turn on the device by

switching the I/O switch at the back of the device to the ON position.

3) Switch to «SCREEN» Mode – Slide the Mode Switch to «SCREEN» Mode.

* Additional visual instruction can be found on the cuff

4) Inflation by fuzzy logic – The device will adjust inflation pressure by fuzzy logic technology in order to determine the optimal inflation pressure.

5) Start measurements – Press the Start/Stop Button to start the measurement sequence.

2) Apply the cuff – Properly fit one cuff to each of the patient’s arms.

DC 7.5V 2.0ADC 7.5V 2.0A

ABI CENTRAL SCREEN ABI CENTRAL SCREEN ABI CENTRAL SCREEN

Taking measurements using WatchBP Office Central

60-second countdown...

Page 15: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

EN17

WatchBP Office Central

6) Three consecutive measurements – The device will take three consecutive measurements with one minute intervals between measurements.

* A 60-second countdown before each measurement.

60 seconds

60 seconds

1 2 3

7) The average value of the measurements – The average value will be displayed after the measurements.

* If one of the three measurements fail, a 4th

measurement will be conducted.

* If the 3rd measurements is cancelled by pressing the Start/Stop after the 2nd measurement, the averages are calculated and saved to the memory as well. Press the M button displays the average, you can continue to press M-button to see the reading of the 2nd and the 1st measurement.

60 seconds

Taking less than three measurementsThe measurement sequence can be stopped at anytime by pressing the Start/Stop Button in «SCREEN» Mode. The device enters Standby Mode and the remaining measurements are cancelled. Average of the measurements is saved and can be viewed by pressing the M Button.

Cancel remaining measurements at anytime during the measurement sequence.

SCREEN SCREEN

Page 16: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

18

Taking measurements using WatchBP Office Central (cont.)

«CENTRAL» Mode1) Turn on the power – Turn on the device by

switching the I/O switch at the back of the device to the ON position.

3) Switch to «CENTRAL» Mode – Slide the Mode Switch to «CENTRAL» Mode.

4) Select the inflation cuff – Set the device to the preferred measurement side (left or right). Press and hold the M Button for 3 seconds.

2) Apply the cuff – Properly fit one cuff to the preferred measurement arm.

Press and hold for 3 seconds...

* Additional visual instruction can be found on the cuff

DC 7.5V 2.0ADC 7.5V 2.0A

ABI CENTRAL SCREEN ABI CENTRAL SCREEN ABI CENTRAL SCREEN

Page 17: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

EN19

WatchBP Office Central

5) Press the M Button to switch between left(L) and right(R). Press the Start/Stop Button to save the setting.

6) Start measurements – Press the Start/Stop Button to start the measurements.

15-second countdown...

7) Automatic two measurements – The device will take two consecutive measurements with 15 second intervals in between on default.

15 seconds

1 2

* The measurement intervals can be set to 15, 30, 45 or 60 seconds in CENTRAL Mode. (Please refer to special functions section on page 19 “Setting up measurement intervals”)

8) Inflation by fuzzy logic – The device will adjust inflation pressure by fuzzy logic technology.

9) The average value of the measurements– The average values of both the brachial and central blood pressure will be displayed and stored automatically.

switch

Page 18: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

20

Taking measurements using WatchBP Office Central (cont.)

Skipping the countdown timeThe 60 seconds countdown before measurement in «SCREEN» Mode and 15 seconds in «CENTRAL» Mode can be skipped by pressing the Start/Stop Button. While the Start/Stop Button is pressed, the device will immediately begin the next measurement.

* The device can be set into Standby Mode by pressing the Start/Stop Button after the completion of measurements. The device will automatically switch to Standby Mode if left unattended for five minutes.

Skip the countdown time and begin measurement.

Standby Mode

Setting measuring intervals in «CENTRAL» Mode

1) Pressing and holding the Start/Stop Button for 3 seconds.

Press and hold for 3 seconds...

2) Press the M Button to adjust the measurement interval, then press the Start/Stop Button to confirm, the device will go back to Standby Mode.

15 306045

* The default measuring interval is 15 seconds. The interval can be set as 15, 30, 45 or 60 seconds.

Page 19: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

EN21

WatchBP Office Central

«ABI» Mode1) Turn on the power – Turn on the device by

switching the I/O switch at the back of the device into the ON position.

2) Switch to «ABI» Mode – Slide the Mode Switch to «ABI» Mode. The supine position reminder will display.

3) The patient has to lie down in supine position.

4) Apply the cuff* – Properly fit the arm cuff to the upper arm and the ankle cuff to the leg of the preferred measurement side of the body. The preferred arm is the arm with the higher blood pressure value, which can be determined in «SCREEN» Mode.

5) Make sure the upper arm cuff is connected to the left cuff socket and the ankle cuff is connected to the right cuff socket.

DC 7.5V 2.0ADC 7.5V 2.0A

* Additional instruction can be found on the cuff.ABI CENTRAL SCREEN ABI CENTRAL SCREEN ABI CENTRAL SCREEN

INDEX LINE

8

LEF

INDEX LINE

8

LEF

ABI

L

R

Page 20: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

22

6) Start measurements – Press the Start/Stop Button to start the measurement.

7) Inflation by fuzzy logic – The device will adjust inflation pressure automatically by fuzzy logic technology.

8) Display of the measurement values– The measurement values will be displayed and the Ankle-brachial index is automatically calculated when the measurement is completed. The Ankle-brachial index and the blood pressure value display first, the brachial index value is replaced by the pulse value for 3 seconds. All the values are saved automatically.

ABI

INDEX

ABI

ABI

Taking measurements using WatchBP Office Central (cont.)

Page 21: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

EN23

WatchBP Office CentralSpecial Functions

Central blood pressure

Central blood pressure is the pressure in the ascending aorta, the largest artery that originates from the left ventricular of the heart and from where oxygen is distributed to all parts of the body through the systemic circulation.

How is central blood pressure measured?

The WatchBP Office Central uses brachial pulse volume plethysmography (PVP) waveforms 1 to determine central blood pressure. The WatchBP Office Central is designed to perform PVP at a cuff pressure of 60 mmHg. Based on the analysis of PVP waveforms the central systolic blood pressure value and the central pulse pressure value are then determined 1.

Accuracy of the WatchBP Office Central

The accuracy of central blood pressure measurement performed with an automated oscillometric device can only reliably be determined against intra-arterial blood pressure measurement. The WatchBP Office Central has been validated against simultaneous recorded intra-arterial blood pressure measurement in 85 subjects and showed high accuracy 2.

CPP (Central Pulse Pressure)

Central Pulse Pressure by this monitor is estimated directly through pulse volume plethysmography (PVP) waveform analysis.

1. Sung, S.H., et al., Measurement of central systolic blood pressure by

pulse volume plethysmography with a noninvasive blood pressure

monitor. Am J Hypertens, 2012. 25: 542-8.

2. Cheng, H.M., et al., Measurement accuracy of a stand-alone

oscillometric central blood pressure monitor: a validation report for

Microlife WatchBP Office Central. Am J Hypertens, 2013. 26: 42-50.

Page 22: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

24

MAP (Mean Arterial Pressure)The WatchBP Office Central device measures the true mean arterial pressure (MAP). Each measurement includes a single MAP value. The average measurement will display the average MAP value. When viewing in Memory mode, the MAP value will be displayed with the systolic and diastolic pressure once every 5 seconds.

PP (Pulse Pressure)This measurement device provides the pulse pressure (PP) of the patient: pulse pressure = systolic - diastolic pressure. Each measurement includes the calculation of a single PP value. The average measurement will display the average PP value. When viewing in Memory mode, the PP value will be displayed with the systolic/diastolic pressure once every 5 seconds.

* The Mean Arterial Pressure (MAP) in this device is determined from the maximum peak of the oscillometric envelope curve.

ROUTINE ROUTINE

Special Function (cont.)

Page 23: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

EN25

WatchBP Office Central

Appearance of the atrial fibrillation indicator for early detection The WatchBP Office Central is designed to screen for asymptomatic atrial fibrillation during blood pressure measurements in the «SCREEN» Mode. If two out of three measurements in the «SCREEN» Mode detect atrial fibrillation, the Afib icon is displayed. This device is able to detect atrial fibrillation with high accuracy: a sensitivity of 97% and a specificity of 89%.

1,2*

1. Joseph Wiesel, et al. Detection of Atrial Fibrillation Using a

Modified Microlife Blood Pressure Monitor. Am J Hypertens 2009;

22, 848–852.

2. G S Stergiou, et al. Diagnostic accuracy of a home blood pressure

monitor to detect atrial fibrillation. J Hum Hypertens 2009, 23,

654-658.

* This device detects atrial fibrillation, a major cause of stroke. Not all risk factors for stroke, including atrial flutter, may be detected by this device.

* This device may not detect atrial fibrillation in people with pacemakers or defibrillators.

SCREEN

Page 24: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

26

About Atrial Fibrillation (Afib) Afib is a common heart rhythm problem. It affects more than 2 million people in North America. It is more common in older age and it is found in 18% individuals aged 85 years and older. It is a common cause of major strokes. About 15% of all strokes are caused by Afib.

The elderly, or those with high blood pressure, diabetes or heart disease are more likely to get a stroke if they have Afib.

Afib can last from a few minutes, to days or weeks and even years. Afib can cause blood clots in the upper chambers of the heart (the atria). These clots can break off and flow to the brain causing a stroke.

The use of blood thinners, such as warfarin, can lower the risk of a stroke in patients with Afib.

A doctor can confirm the presence of Afib by using a 12-lead ECG. Sometimes Afib is present incidentally. Therefore, a doctor may not see it on regularly scheduled visits.

One method of detecting Afib is by mean of palpations. This method is not very reliable. Failure of detecting Afib may lead to the occurrence of a stroke in the end; whereas early detection followed by adequate treatment can reduce the chance of stroke by two-thirds.

Special Function (cont.)

Page 25: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

EN27

WatchBP Office Central

ABI (Ankle Brachial Index)The systolic blood pressure values of a person’s arm and leg need to be measured In order to determine ABI (Ankle Brachial Index).

The ankle-brachial index (ABI) is then calculated using the quotient of the systolic pressure from the leg measurement and the systolic pressure from the arm measurement.

A low ankle-brachial index (ABI) is indicative of systemic vascular disease, and places a patient in the high-risk category.

ABI

INDEX

Page 26: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

28

Hiding measurement valuesThe WatchBP Office Central features a Hide function in order to prevent unnecessary elevated blood pressure in patients due to nervousness triggered by visible blood pressure values.

1) Activate the «Hide» Function – Press M button; L or R flashes; keep pressing and holding the M Button for 7 more seconds until On or Off flashes.

2) Select ON or OFF – Press the M Button again to turn the “Hide” function ON or OFF.

3) Confirm – Press Start/Stop to confirm the setting.

Press and hold for 7 seconds...

* The default setting of “Hide” is set at “ON” in both «SCREEN» Mode and «CENTRAL».

* This feature is available in both SCREEN mode and CENTRAL mode.

* When the «Hide» function is ON, only the average values are displayed after the consecutive measurements and the Hide icon is displayed.

L or R flashes...

Special Function (cont.)

Page 27: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

EN29

WatchBP Office CentralViewing and transferring measurement readings

2) Viewing MAP and PP values – Ten seconds after the display of the average blood pressure, the device displays values for Pulse Pressure (PP) and Mean Arterial Pressure (MAP) .

3) Viewing individual measurements – All individual readings can be viewed by repeatedly pressing the M Button. The display will flash a number before the reading indicating the sequence of the measurement.

10 seconds...

4) Values of the individual measurement (including PP and MAP values) will be displayed in the order listed in step 2.

123

«SCREEN» Mode 1) Viewing the average of all measurements

When the last measurement is performed in «SCREEN» mode, press the M button displays the average of the 3 measurements of the «SCREEN» mode.

Page 28: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

30

2) Viewing MAP, PP, central systolic blood pressure (CBP) and central pulse pressure (CPP) values – The device switches the display for the average blood pressure, Pulse Pressure (PP), Mean Arterial Pressure (MAP), central systolic blood pressure (CBP) and central pulse pressure (CPP) automatically.

3) Viewing individual measurements – All individual readings can be viewed by repeatedly pressing the M Button. The display will flash a number before the reading indicating the sequence of the measurement.

4) The device switches the display for values of the individual measurement (including PP, MAP, CBP and CPP values).

12

«CENTRAL» Mode 1) Viewing the average of all measurements

When the last measurement is performed in «CENTRAL» mode, press the M button displays the average of the 2 measurements of the «CENTRAL» mode.

Viewing and transferring Measurements (cont.)

Page 29: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

EN31

WatchBP Office Central

3) Viewing MAP and PP values – Ten seconds after display of the average blood pressure value, the values for Pulse Pressure (PP) and Mean Arterial Pressure (MAP) are displayed.

* There is only one set of memory capacity in the device. Only the last measurements will be saved in the memory.

* The memory will be deleted when the power is turned off.

«ABI» Mode 1) When the last measurement is performed in

«ABI» mode, press the M button displays the result of the ABI measurement.

2) Viewing the Ankle-brachial index (ABI) At first the INDEX icon, the ABI and the blood pressure values will be displayed

ABI

INDEX

ABI

Page 30: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

32

R

Transferring data to the computer

1) Start the software program and connect the device to the computer using the cable supplied.

2) Refer to the software user manual for detailed information and instructions.

When measuring blood pressure can not use computer to transfer data, the connected computer needs to meet the related safety standards.

Transferring measurements

Installation of the software program

1) Put the CD in the CD-ROM drive of your computer or click on «setup.exe» in the CD’s directory alternatively.

2) Follow the instructions provided in the installation window on the computer screen.

3) When the installation is finished, be sure to restart the computer before you start working with the program.

R

* System Requirements: 550MHz CPU. 256MB Memory, 1024x768 pixel resolution, 256 color, CD-ROM drive, 1 free USB port, 40MB free hard disk space, Microsoft Windows XP / Vista / Win7.

System Requirements: 550MHz CPU. 256MB Memory, 1024x768

pixel resolution, 256 color, CD-ROM drive, 1 free USB port, 40MB

free hard disk space, Microsoft Windo XP / Vista / Win7

Install CD

WatchBP Analyzer Office CentralVersion 2.1.0.8

© 2

012

Mic

rolif

e Co

rpor

atio

n. A

ll Ri

ght

Rese

rved

.

CENTRAL

Viewing and transferring Measurements (cont.)

Page 31: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

EN33

WatchBP Office Central

DC 7.5V 2.0ADC 7.5V 2.0A

Rechargeable BatteryThe device is equipped with a built-in, rechargeable Ni-MH battery pack which delivers up to 400~500 measurement cycles. The batteries can be recharged while the power adaptor is plugged in. It does not affect the measurement.

Appendix

Using a power adaptorPlease only use the Microlife adaptor supplied with the device to recharge the device.

1) Plug the adapter cable into the Power Plug of the WatchBP Office Central device.

2) Plug the adaptor plug into the wall socket. When the power adaptor is connected, no battery power is consumed.

The device program version is EB68h-20130131.* First time use , charge the batteries until the recharge indicator turns to green

* An orange recharge indicator means recharge in progress.

* A green recharge indicator means recharge is completed.

AUSCULT.ROUTINE

SCREEN

ABI CENTRAL SCREEN

MAPIAD

DC 7.5V 2.0ADC 7.5V 2.0A

Page 32: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

34

Troubleshooting

Problem Possible cause How to make correction

No power(No LCD display) Power supply is not properly plugged in

Plug power supply into wall socket.

Battery is fully discharged Recharge the rechargeable battery by plugging in the power supply.

Cuff does not inflate properly

Loose connection of the tube

Make sure the tube of the cuff is securely connected to the device.

Leakage of the tube or bladder

Check for cracks on the tube or the bladder. Please contact Microlife customer service for this issue.

No result displayed between measurements

The Hide function is activated

Disable “hide” function

Page 33: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

EN35

WatchBP Office Central

If an error occurs during measurement, the measurement is interrupted and an error message «Err» is displayed.

• Please contact your local Microlife service center if the error persists.

• If you think the results are unusual, please read through the information in this instruction manual carefully.

Error Description Potential cause and remedy

Err 1 Signal too weak

The pulse signals on the cuff are too weak. Re-position the cuff and repeat the measurement.

Err 2 Error signal During the measurement, error signals were detected by the cuff, caused for instance by movement or muscle tension. Repeat the measurement, keeping your arm still.

Error messages

Page 34: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

36

Err 11 Signal too weak during central blood pressure measurement

The pulse signals on the cuff are too weak. Re-position the cuff and repeat the measurement.

Err 12 Error signal during Central blood pressure measurement

During the measurement, error signals were detected by the cuff, caused for instance by movement or muscle tension. Repeat the measurement, keeping your arm still

Err 3 No pressure in the cuff

An adequate pressure cannot be generated in the cuff. A leak may have occurred. Replace the batteries if necessary. Repeat the measurement.

Err 5 Abnormal result

The measuring signals are inaccurate and no result can therefore be displayed. Read through the checklist for performing reliable measurements and then repeat the measurement.

Error messages

Page 35: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

EN37

WatchBP Office Central

Err 13 Cuff pressure errors during Central blood pressure measurement

An adequate pressure cannot be generated in the cuff. A leak may have occurred. Check that the cuff is correctly connected and is not too loose. Replace the batteries if necessary. Repeat the measurement.

Err 15 Abnormal result of Central blood pressure reading

The measuring signals are inaccurate and no result can therefore be displayed. Read through the checklist for performing reliable measurements and then repeat the measurement.

HI Pulse or cuff pressure too high

The pressure in the cuff is too high (over 300 mmHg) OR the pulse is too high (over 200 beats per minute). Relax for 5 minutes and repeat the measurement.

LO Pulse too low The pulse is too low (less than 40 beats per minute). Repeat the measurement.

Page 36: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

38

Protect the device from water and moisture

Protect the device from extreme heat and cold

Protect device from impact and drops

Protect the device from direct sunlight

Avoid proximity to electromagnetic fields, such as those produced by mobile phones

Never open device

Safety, care, accuracy test and disposal

Safety and protectionThis device may be used only for the purpose described in this booklet. The device comprises of sensitive components and must be treated with caution. The manufacturer cannot be held liable for damage caused by incorrect application.

Observe the storage and operating conditions described in the “Technical specifications” section of this manual.

• Ensure that children do not use the device unsupervised; some parts are small enough to be swallowed.

• Only activate the pump when cuff is installed.

• Do not use the device if you think it is damaged or if anything appears unusual.

• Read the further safety instructions in the individual sections of the instruction manual.

• Do not connect the device to a computer until prompted to do so by the computer software.

Read the instruction manual carefully before using this device, especially the safety instructions, and keep the instruction manual for future use. Do not use the instrument close to strong electromagnetic fields such as mobile telephones or radio installations

Page 37: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

EN39

WatchBP Office Central

ABI CENTRARR L SCREEN

CENTRAL

CENTRAL

Do not wash and scrub cuff

Do not iron the cuff!

Accuracy testWe recommend the WatchBP Office Central device be tested for accuracy every 2 years or after mechanical impact (e.g. being dropped). Please contact Microlife to arrange for an accuracy test.

Cleaning the cuff Do not wash and scrub cuff . DO NOT iron the cuff cover.

Device care Clean the device with a soft, dry cloth.Cleaning the adapter

Clean the adapter with a dry cloth.

DisposalBatteries and electronic instruments must be disposed of in accordance with the locally applicable regulations, and not as domestic waste.

Page 38: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

40

Technical specifications Operation temperature/humidity: • 10 to 40 °C (50 to 104 °F), 15 - 90 %

relative maximum humidity at an atmospheric pressure range of 860hPa to 1060 hPa.

Transportation/ Storage temperature/humidity:

• -20 to 55 °C (-4 to 131 °F)

• 15 - 90 % relative maximum humidity

Weight: • 1100 g (including rechargeable battery pack)

Dimensions: • 200 x 125 x 90 mm

Measuring method: • Oscillometric, corresponding to Korotkoff

Measurement range: • 30 - 280 mmHg – blood pressure

• 40 - 200 beats per minute – pulse

Cuff pressure display:

Voltage source:

• Range: 0 - 299 mmHg

• Resolution: 1 mmHg

• Static accuracy: pressure within ± 3 mmHg

• Pulse accuracy: ±5 % of the readout value

• Rechargeable battery pack; 4.8V C3500 mAh

• Mains adapter DC 7.5V, 2 A

Reference to Standards: • Device corresponds to the requirements of the standard for non-invasive blood pressure monitor.EN 1060-1EN 1060-3EN 1060-4IEC 60601-1IEC 60601-1-2

Electromagnetic compatibility:

• Device fulfills the stipulations of the standard IEC 60601-1-2.

The stipulations of the EU Directive 93/42/EEC for Medical Devices Class IIa have been fulfilled.

Type BF applied part

Manufacturer Reference number

Class II equipment Caution

For indoor use

Microlife reserves the right to alter technical specifications without prior written notice.

0044

Page 39: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

EN41

WatchBP Office Central

This device is covered by a two-year

guarantee and accessories are cover by

a one-year guarantee from the date of

purchase. This guarantee is valid only

on presentation of the guarantee card

completed by the owner confirming date of

purchase or purchase receipt.

Product model: WatchBP Office Central

REF: TWIN200 CBP

Name:

Address:

Date:

Telephone:

E-mail:

Guarantee card

Page 40: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

42

Page 41: WatchBP Office Central - quickmedical...J of Hypertens. 2013, 31:1281–1357. 8. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood

EN43

WatchBP Office Central

IB WatchBP Office Central EN 4716

Europe / Middle-East /

Africa

Microlife WatchBP AG

Espenstrasse 139

9443 Widnau, Switzerland

Tel +41 71 727 7000

Fax +41 71 727 7011

Email: [email protected]

www.watchbp.com

Asia

Microlife Corporation

9F, 431, RuiGang Road, NeiHu

Taipei, 114, Taiwan, R.O.C.

Tel +886 2 8797 1288

Fax+886 2 8797 1283

Email:

[email protected]

www.watchbp.com

United States, Canada

Microlife Medical Home

Solutions, Inc.

2801 Youngfield St., Suite 241

Golden, CO 80401, USA

Tel. +1 303 274 2277

Fax +1 303 274 2244

Email: [email protected]

www.watchbp.com

Central / South

America

Microlife USA, Inc.

1617 Gulf To Bay Blvd., 2nd Floor

Clearwater, FL 33755, USA

Tel +1 727 442 5353

Fax +1 727 442 5377

Email: [email protected]

www.watchbp.com