ESI school 16-20 May 2011 X-ray area detectors for SR 1 X-ray area detectors for synchrotron experiments Characteristics and Technologies Cyril Ponchut ESRF Instrumentation Services and Development Division

Upload: hoangdat

Post on 01-Jul-2018

231 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

ESI school 16-20 May 2011 X-ray area detectors for SR 1

X-ray area detectors for synchrotron experimentsCharacteristics and Technologies

Cyril PonchutESRF Instrumentation Services and Development Division

Page 2: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 2

Outline

Introduction

2D detector parameters

2D detector principles

2D detector technologies

Page 3: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 3

Introduction

The beamline user problem

X-ray area detector on a SR beamline

Page 4: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 4

?

Introduction

Image : “The ATLAS experiment at CERN”

The beamline user problemWhat is the optimum detector for a given (range of ) experiment(s) ?

Page 5: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 5

X-ray area detection on a SR beamline

sample

Φo(t,E)

X-ray beam

x

y

detector

Si (x,y,t,E)

Introduction

Page 6: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 6

Detector parameters and characterization

2D detector parameters2D detector modelGainNoiseDynamic range

Page 7: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 7

2D detectors parameters

MTFLSFspatial distortions

Frame rateReadout dead timeDecay time

GainNoiseDynamic rangeLinearity

Energy resolutionEnergy threshold

QEDQE

X-ray area detector = device that measures the intensity of an

incident X-ray flux with a certain efficiency, as a function of

position, time , energy.

Detector parameters and characterization

Page 8: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 8

Gain

inputoutputG =

G = image level (ADU) per incident X-ray

integrated pixel signal in the exposed region

X-ray counts through pinhole measured with a counter

Measuring G :

G includes X-ray interaction probability => depends on energy

G =

Detector parameters and characterization

ADU : analog to digital units

Page 9: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 9

Noise

)),(( jiIVarNo = ROIji ∈),(

1

10

100

1000

1 10 100 1000 10000X-ray photons/pixel

r.m.s

. im

age

nois

e quantumreadoutproportionalDSNUtotal

quantum readout proportional

Noise components :

22222DPRQ NNNNN +++=

DSNU

stochastic spatial

Image noise : ADU/pixel r.m.s.

Detector parameters and characterization

incoming photons

(Poisson)

electronics

inhomogeneities(fixed pattern

noise)

dark pattern

Page 10: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 10

Dynamic range

min0

max0

SS

DR =

maxXIDR =

DRDRbits 2log=

0

max

NII

DR dX −=

( S0min ≈ No = readout noise )

DR = 10000 = 80 dB = 13.2 bits

Not to be confused with ADC range = log2Imax

IXmax

Id

input X-ray flux

imag

e le

vel

S0min

S0max

energy integratingIXmax

1imag

e le

vel

S0max

0

photon counting

S0mininput X-ray flux

Detector parameters and characterization

Page 11: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 11

Linearity

Integral non-linearity

><><−

=G

GSGINL i |))(max(| <G>

SiSi maxQuick non-linearity test

Detector parameters and characterization

Page 12: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 12

Quantum efficiency

QE = X-ray interaction probability = characteristic of the X-ray conversion medium

QE can be deduced from knowledge of detection material, but not measurable directly

⇒ Need for a general and measurable quantity for detection efficiency :

DQE

(Detective Quantum Efficiency)

QE does not take into account signal degradation across the system

Detector parameters and characterization

Page 13: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 13

Detective Quantum Efficiency (DQE)

20

0

NSG

DQE⋅

=

DQE = SNRout2/SNRin

2 = (So2/ No

2 ) / (Si2/ Ni

2 ) (Gruner, 1978)

So/ Si = GNi

2 = Si (Poisson statistics)

Measuring the DQE :

Detector parameters and characterization

Page 14: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 14

DQE approximated expression

i

e

QSgnQDQE

2

2

1 −+

≈Si

X-rays incident

QSiQ

X-rays detected

So = SiQg

output signal

Si SiQ SiQg2 + ne-2

signal

variance(Poisson) (Poisson)

Input-equivalent noise Neq = ne-/gQ :

i

eq

SQNQDQE 2

1+

“perfect” detector (ne-= 0) => DQE = Q

g

Detector parameters and characterization

Page 15: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 15

Line-spread function (LSF)

Slit aperture << pixel size

IEEE TNS 52 (2005)

Tilted slit (Fujita, 1992) : oversampling 1/sinθ=> Provides the presampling LSF

Measuring the LSF :

Detector parameters and characterization

For non-isotropic spatial response : PSF (point-spread function)

Page 16: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 16

Contrast Transfer Function (CTF)

Imax

Imin

(square modulation)

contrastmodulation

)()()()(

)(minmax

minmax

νννν

νIIII

CTF+−

=

Measuring the CTF :

Detector parameters and characterization

Page 17: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 17

Modulation Transfer Function (MTF)

minmax

minmax

IIII

MTF+−

= (sine modulation) Not measurable directly

∫ −= dxxLSFMTF xjπνν 2exp)()(

Indirect measurement from LSF :

Detector parameters and characterization

Page 18: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 18

DQE in the Fourier domain

0,0 0,5 1,00,0

0,5

nps

(Xph

/pixe

l-1)

frequency (normalized)

nps2||)),((||),( jiIFFTlkNPS =

0,0 0,5 1,00,0

0,5

1,0

DQE

frequency (normalized)

DQE

)(.)()(

2

ννν

o

o

NPSNMTFSGDQE ⋅⋅

=

Noise power spectrum

angularaveraging

Frequency-dependent DQE

Detector parameters and characterization

∫=π

θν2

0,) )dθNPS(νNPS(

Page 19: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 19

Time resolutionDetector parameters and characterization

Tjitter

frame rate F = 1/TcycleTdecayTcycle

exposure

readout

Treadout

Texp

triggered acquisition

experiment

Tcycle

exposure

Texp

Page 20: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 20

Summary : basic area detector specification list

Detection field, pixel size, LSF Gain, noise, dynamic range Linearity DQE at a given energy Frame rate, readout dead time, minimum exposure time Energy range

Detector parameters and characterization

Page 21: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 21

2D X-ray detection principles

Indirect conversion

X-ray convertersOptical coupling

Direct conversion

Page 22: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 22

Indirect conversion

X-ray to light converter

Impinging X-ray photon

optical coupling

image sensor

X-ray 2D detection principles

Page 23: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 23

X-ray to light converters

phosphor powder+ binder

reflective layer

spatial resolution ≈ e

e=5…100 µm

Gd2O2S:Tb (P43), YAG:Ce (P46),…

2D X-ray detection principles

columnar CsI

spatial resolution < e

e = 100…500 µm

growth substrate

scintillator layertransparentsubstrate

microscopeobjective

LuAG, YAG:Ce, LAG:Ce,…

spatial resolution = 0.5…5 µm

α

e=1…25 µm

NA=n.sin α

Martin, Koch, JSR (2006)

Page 24: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 24

Absorption efficienciesX-ray 2D detection principles

1.E-03

1.E-02

1.E-01

1.E+00

0 50 100 150

keV

abso

rptio

nCsI 100 µmGOS 40 µmYSO 10 µmYAG 10 µm

High-resolution single crystal film scintillators

Phosphor powder screens

Absorption efficiency / spatial resolution trade–off

Page 25: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 25

Decay times2D X-ray detection principles

10-6

10-5

0.0001

0.001

0.01

0.1

1

1 10 100 1000 104

Rel

ativ

e am

plitu

de

Time (ms)

Afterglow depend on exposure timeGGG: Eu 14 bit dynamic @ 80ms16-17 bit are resolved

100ms

1s 10s

12 b

it14

bit

16 b

it

Bit depth

speed / dynamic range trade–off

Data from Thierry Martin (ESRF)

Page 26: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 26

Decay times

speed / sensitivity trade–off

Phosphors with peak emission in the 530-550 nm range :

P20

P43

P46

0

20

40

60

1.E-05 1.E-04 1.E-03 1.E-02 1.E-01

decay time at 1% (s)

effic

ienc

y (p

h/ke

V)

Page 27: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 27

Optical coupling

Lens

Fiberoptic taper

Direct coupling

2D X-ray detection principles

Page 28: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 28

Optical coupling efficiency

222

)/11(41 GNTNATopt ++

=⋅=ηSingle lens

p p’G = p’/p

G = 1:1 and N = F/2 : ηopt = 0.015

Demagnification (G < 1 ) => coupling efficiency ∼1/N2G2

Magnification (G > 1 ) => coupling efficiency ∼NA2 ~1/N2

DfN =

2D X-ray detection principles

Field width / sensitivity trade-off

Improvable using tandem lenses :

f f’

G = f’/f

Page 29: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 29

Signal propagation in indirect detection

opticalsystem

imagesensor

chargesignal

X-rayconverter

10 keV 500 hν 5 hν 3 e-

~50 hν/keV

~0.01 hν/hν(demag.lens)

QE ~ 65%

image

1 ADU

signal processing

2D X-ray detection principles

SX-ray ~ ne- : single-photon discrimination (generally) not possible

Page 30: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 30

Signal propagation in direct detection

chargesignal

depletedsemiconductorpixel sensor

10 keVX-ray

2760 e-

εh+/e-= 3.62 eV

imagesignal processing“perfect” material

(e.g. Si) :

Imperfect material ( ≠ Si )

incomplete depletion → reduced DQE

SX-ray >> ne- : single photon detection possible

incomplete charge collection → reduced sensitivity

polarization → image afterglow

GaAs

CdTeCdZnTea-Se

HgI2

PbI2

2D X-ray detection principles

Page 31: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 31

LSF : indirect vs direct detection

IEEE TNS 52 (2005)

2 different detectors with identical pixel sizes and X-ray absorptions :

2D X-ray detection principles

Lens-coupled CCDDirect detection in Si

-3 0 3100

101

102

103

104

105

106

107

(b)

Inte

nsity

(ADU

)

Distance to slit centre (mm)

MPX CCD

Page 32: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 32

Summary : indirect vs. direct detection

wide range of spatial resolutions

versatile

2D X-ray detection principles

indirect direct

fixed spatial resolutionlow gain at large input fields

dynamic range limitations

sharp LSF

high gain (=> photon counting)

Page 33: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 33

2D X-ray detector technologies

Image sensors : CCD, CMOS Optically-coupled CCD detectors Photon-counting ASICS and detectors Flatpanels Image plates Gas filled multiwire proportional chambers

Page 34: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 34

CCD image sensors

Image credit : DALSA

+V1

p Si bulk

SiO2

poly-Si electrodes

+V2

+V3

pixel wells

MOS gate structure, 3-phase CCD

horizontal shift

verticalshift

output video signal

output node

V=q/C

Full frame transfer exposure during readout => SMEARING

2D X-ray detector technologies

Page 35: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 35

Electron-multiplication CCD

High gain

Self-amplification in horizontal shift register

Dark noise amplification

Excess noise at high gain

2D X-ray detector technologies

Page 36: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 36

CMOS image sensor2D X-ray detector technologies

row

add

ress

ing

column ADCmemory

MUX

active pixel matrix

Example layout Example characteristics (MICRON):

Active Array 1,280H x 1,024VImaging Area 15.36mm(H) x 12.29mm(V)Pixel Size 12.0µmDynamic Range 59dBResponsivity 1600 LSB/lux-secFrame Rate 0-500+ fpsShutter type TrueSNAPData Rate 660 Mp/sMaster Clock 66 MHzData Format 10-bit digital

Page 37: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 37

CCD vs. CMOS in brief

High frame rate

Fixed pattern noise

Readout noise

Short exposures

readout time

smearing (full frame)

Low readout noise

High dynamic range

CCD CMOS

2D X-ray detector technologies

Page 38: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 38

Application : high-energy imaging (ESRF ID15)

manual zoom setting

folding mirror

screen holder with motorized focusing

camera

Variable field CCD detector

Field size 7-19 mmPixel size 3.8-9.4 µmConverter YAG 280 µmGain 2.4 ADU/keVLSF FWHM 13-25 µmFrame rate ~10 Hz

Design : ESRF

2D X-ray detector technologies

Adjustable field size with zoom lens

Page 39: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 39

CCD detector for EDXAS

Image : ESRF ID24http://www.esrf.fr/UsersAndScience/Experiments/XASMS/ID24/

CCD sensor

Lens

Lens

MCP (fast shutter)

Phosphor screen

2D X-ray detector technologies

Field size 50x3 mm2

Pixel size 25 µmNoise 4 ADU/pixel rmsDynamic range 14 bitsFrame rate 1 k Hz (kinetics mode)

Based on ESRF FReLoN CCD cameraRSI 78 (2007)

Page 40: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 40

High-resolution CCD detector

Substrate 170µm

Sensitive layer 1-25µm

Substrate 170µm

Sensitive layer 1-25µm

Submicron spatial resolution

Design : ESRF (Thierry Martin)

2D X-ray detector technologies

Page 41: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 41

Large mosaic CCD

Field size 312 x 312 mm2

Pixel size 52 µmADC range 16 bits

Noise 4 ADU/pixel r.m.sGain 2.1 ADU/10 keV X-rayDQE 0.6 @ 10 keV

LSF FWHM 124 µmFrame rate 0.26 Hz (1 s exposure)

2D X-ray detector technologies

Page 42: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 42

Large mosaic CCD : image corrections

Correction- Spatial distortion - flatfield

raw image

2D X-ray detector technologies

Page 43: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 43

Hybrid photon-counting pixel detectors

CERN, the Medipix Collaborationhttp://medipix.web.cern.ch/MEDIPIX/

Llopart et al. IEEE TNS 49 ( 2002 ) 2279

Each pixel is a photon counter

2D X-ray detector technologies

Example : Medipix2

SC sensor

pixel ASIC

chipboard PCBwirebonds

bump-bonding contacts

sensor bias

Page 44: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 44

2D photon-counting : charge sharing 2D X-ray detector technologies

5 10 15 20 25 300.0

0.5

1.0

1.5

rela

tive

coun

tsEnergy (keV)

pixel 170 µm pixel 55 µm

q

Page 45: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 45

MAXIPIX

ESRF developmentBased on MEDIPIX2 chiphttp://medipix.web.cern.ch/MEDIPIX/

2D X-ray detector technologies

Detection areas256x256512x512

1280x256pixel2

Pixel size 55 µm2

X-ray converter Si 500 µmCounter depth 13.5 bits

Frame rate 280-1400 HzReadout dead time 0.29 ms

Fast readout photon-counting detector

Page 46: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 46

Other photon-counting pixel detectors

Detection areas487x195to

2463x2527pixel2

Pixel size 172 µm2

X-ray converter Si 300 µmCounter depth 20 bits

Frame rate 12-200 HzReadout dead time 2.7-3.6 ms

PILATUS (DECTRIS)

2D X-ray detector technologies

XPAD2 (CNRS CPPM) Detection areas 120x75 mm2

Pixel size 330 µm2

X-ray converter Si 500 µmCounter depth 15 bits

Readout dead time 2 ms

Delpierre et al. NIMA 572 (2007) 250

Page 47: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 47

Example application : SAXS

Colloidal PMMA, 40% concentration, 8.33 keV

Reduced exposure time Arbitrarily high dynamic rangeNo profile distortion at low signal

MPX2 : 1 s CCD : 100 x 1 spoint detector scan : 5s/point

data : F. Zontone (ESRF)

IEEE TNS 52(5) (2005) 1760

2D X-ray detector technologies

Photon-counting detector (Medipix2)

Page 48: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 48

Fluorescence rejection

5 10 15 20 25 300

5x103

1x104

elasticscatter

fluorescencebackground

coun

ts/k

eV

energy threshold (keV)

ROI 1 (backgd.+ring) ROI 2 (backgd.)

Cu foil sample, 20 keV beam

threshold settings

NIMA 510 (2003) 29

2D X-ray detector technologies

Photon-counting detector (Medipix1)

Page 49: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 49

Flatpanel + scintillator

Field size 407 x 296 mm2

Pixel size 154 µmNoise 0.2-2.6 ADU/pixel r.m.sGain 0.2-1 ADU/X-ray @ 37 keVDQE 0.85 @ 37 keVLSF FWHM 210 µmFrame rate 7.5-30 HzEnergy range 30 - 100 keV

Main application : high energy diffraction

2D X-ray detector technologies

PIXIUM (Thales)

Images : Trixell

Page 50: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 50

Flatpanel + semiconductor

Main application : diffraction

Field size 358 x 430 mm2

Pixel size 139 µm2

X-ray converter a-SeADC range 16 (20) bitsNoise 2.8 ADU/pixel r.m.sGain 0.6 ADU/17.4 keVDQE 0.6 @ 17.4 keVLSF FWHM ~130 µmFrame rate 0.3 Hz (1 s exposure)

2D X-ray detector technologies

Page 51: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 51

Flatpanel + semiconductor

CCD fiberopticpixel size 100 µm

a-Se flatpanelpixel size 139 µm

Pixel-limited spatial resolution Large uniform detection area Image afterglow

2D X-ray detector technologies

Page 52: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 52

Image plate scanner2D X-ray detector technologies

Photostimulable luminescent screen

Von Seggern et al., J. Appl. Phys. 64(3) (1988) 1405

Image : NONIUS

Field size 345 mm diametrePixel size 100 µmX-ray converter BaFBr:Eu2+

ADC range 17 bitsNoise 1 ADU/pixel r.m.sGain 1 ADU/8 keV X-rayReadout time 108 s

Page 53: X-ray area detectors for synchrotron experiments · X-ray area detectors for synchrotron experiments ... Introduction 2D detector ... Xmax. image level. 1. S 0max. 0. photon counting

Eiroforum school on Instrumentation - 16-20 May 2011 53

Conclusion

An obviously incomplete overview of 2D X-ray detectors for synchrotrons experiments

Detector parameters and characterization

This lecture only pretends to be :

A guide to help asking oneself the right questions when having to choose or design a 2D X-ray detector

A incentive to learn more about modern 2D X-ray detection technologies

Thank you for your attention