x-ray emission spectrumkarary.edu.sd/site/assets/files/22114/04_x-ray_emission_spectrum-1.pdf• a...
TRANSCRIPT
-
X-ray Emission Spectrum
Dr. Ahmed Alsharef Farah
Dr. Ahmed Alsharef Farah 1
-
• X-ray photons produced by an X-ray tubeare heterogeneous in energy.
• The energy spectrum shows a continuousdistribution of energies for thebremsstrahlung photons superimposed bycharacteristic radiation of discreteenergies.
Dr. Ahmed Alsharef Farah 2
-
There are two types of X-ray spectrum:
1. Bremsstrahlung or continuous spectrum.2. Characteristic spectrum.
Dr. Ahmed Alsharef Farah 3
-
• A bremsstrahlung spectrum consists of X-ray photons of all energies up to maximumin a continuous fashion, which is alsoknown as white radiation, because of itssimilarity to white light.
• A characteristic spectrum consists of X-rayphotons of few energy, which is also calledas line spectrum.
• The position of the characteristic radiationdepends upon the atomic number of thetarget.
Dr. Ahmed Alsharef Farah 4
-
Dr. Ahmed Alsharef Farah 5
-
Characteristic X-ray Spectrum:
• The discrete energies of characteristic x-rays are characteristic of the differencesbetween electron binding energies in aparticular element.
• A characteristic x-ray from tungsten, forexample, can have 1 of 15 different energiesand no others.
Dr. Ahmed Alsharef Farah 6
-
Characteristic X-rays of Tungsten and Their EffectiveEnergies (keV).
Dr. Ahmed Alsharef Farah 7
-
Characteristic x-ray emission spectrum for tungstencontains 15 different x-ray energies.
Dr. Ahmed Alsharef Farah 8
-
• Such a plot is called the characteristic x-rayemission spectrum.
• Five vertical lines representing K x-rays andfour vertical lines representing L x-rays areincluded.
• The lower energy lines representcharacteristic emissions from the outerelectron shells.
Dr. Ahmed Alsharef Farah 9
-
• The relative intensity of the K x-rays isgreater than that of the lower energycharacteristic x-rays because of the natureof the interaction process.
• K x-rays are the only characteristic x-raysof tungsten with sufficient energy to be ofvalue in diagnostic radiology.
• Although there are five K x-rays, it iscustomary to represent them as one with asingle vertical line, at 69 keV.
Dr. Ahmed Alsharef Farah 10
-
The characteristic x-ray emission spectrum isrepresented by a line at 69 keV.
Dr. Ahmed Alsharef Farah 11
-
Bremsstrahlung X-ray Spectrum:
• If it were possible to measure the energycontained in each bremsstrahlung x-rayemitted from an x-ray tube, one would findthat these energies range from the peakelectron energy all the way down to zero.
• In other words, when an x-ray tube isoperated at 90 kVp, bremsstrahlung x-rayswith energies up to 90 keV are emitted.
Dr. Ahmed Alsharef Farah 12
-
• A typical bremsstrahlung x-ray emissionspectrum is shown in Figure below.
Dr. Ahmed Alsharef Farah 13
-
Question:• At what kVp was the x-ray imaging system
presented in Figure above operated?Answer:• Because the bremsstrahlung spectrum intersects the
energy axis at approximately 90 keV, the imagingsystem must have been operated at approximately 90kVp.
Dr. Ahmed Alsharef Farah 14
-
• The general shape of the bremsstrahlung x-ray spectrum is the same for all x-rayimaging systems.
• The maximum energy (in keV) of abremsstrahlung x-ray is numerically equalto the kVp of operation.
Dr. Ahmed Alsharef Farah 15
-
• The greatest number of x-rays is emittedwith energy approximately one third of themaximum energy.
• The number of x-rays emitted decreasesrapidly at very low energies.
Dr. Ahmed Alsharef Farah 16
-
• The energy of an x-ray is equal to theproduct of its frequency (f) and Planck’sconstant (h). X-ray energy is inverselyproportional to its wavelength. As x-raywavelength increases, x-ray energydecreases.
• The minimum wavelength of x-rayemission corresponds to the maximum x-ray energy, and the maximum x-rayenergy is numerically equal to the kVp.
Dr. Ahmed Alsharef Farah 17
-
Question
• What would be the expected emissionspectrum for an x-ray imaging system witha pure molybdenum (Mo) target (effectiveenergy of K x-ray = 19 keV) operated at 90kVp?
Dr. Ahmed Alsharef Farah 18
-
Answer
• The spectrum should look something likeFigure below.
Dr. Ahmed Alsharef Farah 19
-
• The curve intersects the energy axis at 0 and90 keV and has the general shape shown inFigure below.
Dr. Ahmed Alsharef Farah 20
-
• The bremsstrahlung spectrum is muchlower because the atomic number of Mo islow (Z = 42), and x-ray production is muchless efficient.
• A line extends above the curve at 19 keV torepresent the K-characteristic x-rays ofmolybdenum.
Dr. Ahmed Alsharef Farah 21
-
• The total number of x-rays emitted froman x-ray tube could be determined byadding together the number of x-raysemitted at each energy over the entirespectrum, a process called integration.
Factors affecting the x-ray emission spectrum
Dr. Ahmed Alsharef Farah 22
-
• Graphically, the total number of x-raysemitted is equivalent to the area under thecurve of the x-ray emission spectrum.
• The general shape of an emissionspectrum is always the same, but its relativeposition along the energy axis can change.
• The farther to the right a spectrum is, thehigher the effective energy or quality ofthe x-ray beam.
Dr. Ahmed Alsharef Farah 23
-
• The larger the area under the curve, thehigher is the x-ray intensity or quantity.
• A number of factors under the control ofradiographers influence the size and shapeof the x-ray emission spectrum andtherefore the quality and quantity of the x-ray beam.
Dr. Ahmed Alsharef Farah 24
-
Factors that affect the size and relative position of x-ray emission spectrum.
Dr. Ahmed Alsharef Farah 25
-
1. Effect of mA and mAs:
• If one changes the current from 200 to 400mA while all other conditions remainconstant, twice as many electrons will flowfrom the cathode to the anode, and the mAswill be doubled.
• This operating change will produce twice asmany x-rays at every energy.
Factors affecting the x-ray emissionspectrum:
Dr. Ahmed Alsharef Farah 26
-
• In other words the x-ray emissionspectrum will be changed in amplitude butnot in shape.
• Increasing the mA does not affect theshape of the spectrum but increases theoutput of both bremsstrahlung andcharacteristic radiation proportionately.
• The area under the x-ray emissionspectrum varies in proportion to changes inmA or mAs, as does the x-ray quantity.
Dr. Ahmed Alsharef Farah 27
-
Change in mA or mAs results in a proportionate changein the amplitude of the x-ray emission spectrum at allenergies.
Dr. Ahmed Alsharef Farah 28
-
2. Effect of kVp:
• As the kVp is raised, the area under thecurve increases to an area approximatingthe square of the factor by which kVp wasincreased.
• Accordingly, the x-ray quantity increaseswith the square of this factor.
Dr. Ahmed Alsharef Farah 29
-
• When kVp is increased, the relativedistribution of emitted x-ray energy shiftsto the right to a higher average x-ray energy.
• The maximum energy of x-ray emissionalways remains numerically equal to thekVp.
Dr. Ahmed Alsharef Farah 30
-
• Increasing the kV shifts the spectrumupwards and to the right.
• It increases the maximum and effectiveenergies and the total number of x-rayphotons.
• Below a certain kV (70 kV for a tungstentarget), the characteristic K-radiation isnot produced.
Dr. Ahmed Alsharef Farah 31
-
Effect of tube kilovoltage (kV) on X-ray spectra forthree tube potentials: A, 40 kV; B, 80 kV; and C, 120kV.
Dr. Ahmed Alsharef Farah 32
-
• A 15% increase in kVp does not double thex-ray intensity but is equivalent to doublingthe mAs to the image receptor.
• To double the output intensity by increasingkVp, one would have to raise the kVp by asmuch as 40%.
Dr. Ahmed Alsharef Farah 33
-
• Radiographically, only a 15% increase inkVp is necessary because with increasedkVp, the penetrability of the x-ray beam isincreased.
• Therefore, less radiation is absorbed by thepatient, leaving a proportionately greaternumber of x-rays to expose the imagereceptor.
Dr. Ahmed Alsharef Farah 34
-
3. Effect of Added Filtration:
• Adding filtration to the useful x-ray beamreduces x-ray beam intensity whileincreasing the average energy.
• Added filtration more effectively absorbslow-energy x-rays than high-energy x-rays;therefore, the bremsstrahlung x-rayemission spectrum is reduced further on theleft than on the right.
Dr. Ahmed Alsharef Farah 35
-
Adding filtration to an x-ray tube results in reduced x-ray intensity but increased effective energy.
Dr. Ahmed Alsharef Farah 36
-
The emission spectra represented here resulted from operationat the same mA and kVp but with different filtration.
Dr. Ahmed Alsharef Farah 37
-
• Adding filtration is sometimes calledhardening the x-ray beam because of therelative increase in average energy.
• The characteristic spectrum is notaffected, nor is the maximum energy of x-ray emission.
Dr. Ahmed Alsharef Farah 38
-
4. Effect of Target Material:
• The atomic number of the target affectsboth the number (quantity) and theeffective energy (quality) of x-rays.
• As the atomic number of the target materialincreases, the efficiency of the production ofbremsstrahlung radiation increases, andhigh-energy x-rays increase in number to agreater extent than low-energy x-rays.
Dr. Ahmed Alsharef Farah 39
-
• The change in the bremsstrahlung x-rayspectrum is not nearly as pronounced as thechange in the characteristic spectrum.
• After an increase in the atomic number ofthe target material, the characteristicspectrum is shifted to the right, representingthe higher energy characteristic radiation.This phenomenon is a direct result of thehigher electron binding energies associatedwith increasing atomic number.
Dr. Ahmed Alsharef Farah 40
-
• Changing the target to one of loweratomic number reduces the output ofbremsstrahlung but does not otherwiseaffect its spectrum, unless the filtration isalso changed.
• The photon energy of the characteristiclines will also be less.
• Elements of low atomic number alsoproduce low-energy characteristic x-rays.
Dr. Ahmed Alsharef Farah 41
-
Discrete emission spectrum shifts to the right with anincrease in the atomic number of the target material.
Dr. Ahmed Alsharef Farah 42
-
• Tungsten is the primary component of x-raytube targets, but some specialty x-ray tubesuse gold as target material.
• The atomic number for tungsten is 74 andgold is 79.
Dr. Ahmed Alsharef Farah 43
-
• Molybdenum (Z = 42) and rhodium (Z =45) are target elements used formammography.
• In many dedicated mammography imagingsystems, these elements are incorporatedseparately into the target.
• The x-ray quantity from suchmammography target material is lowowing to the inefficiency of x-rayproduction.
Dr. Ahmed Alsharef Farah 44
-
5. Effect of Voltage Wave form:
• There are five voltage waveforms:I. Half-wave–rectified.II. Full-wave–rectified.III. Three-phase/six-pulse.IV. Three-phase/ 12-pulse.V. High-frequency waveforms.
Dr. Ahmed Alsharef Farah 45
-
Waveforms of high-voltage generators:(a) singlephase, half wave–rectified; (b) single-phase,full wave–rectified; (c) three-phase, six-pulse; and (d)high-frequency generator.
Dr. Ahmed Alsharef Farah 46
-
• Whatever the kV waveform, the maximumand minimum photon energies areunchanged.
Dr. Ahmed Alsharef Farah 47
-
Three-phase and high-frequency operations areconsiderably more efficient than single-phase operation.
Dr. Ahmed Alsharef Farah 48
-
• The number of x-rays emitted is low at lowervoltages and increases at higher voltages.
• The quantity of x-rays is much greater atpeak voltages than at lower voltages.
• Consequently, voltage waveforms of three-phase or high-frequency operation result inconsiderably more intense x-ray emissionthan those of single-phase operation.
Dr. Ahmed Alsharef Farah 49
-
• The relationship between x-ray quantityand type of high-voltage generator providesthe basis for another rule of thumb used byradiologic technologists.
• If a radiographic technique calls for 72kVp on single-phase equipment, then onthree-phase equipment, approximately 64kVp — a 12% reduction — will producesimilar results.
Dr. Ahmed Alsharef Farah 50
-
Dr. Ahmed Alsharef Farah 51