The Toshiba FDX4343R X-ray Flat Panel Detector (FPD) for radiographic applications delivers high sensitivity, high resolution, low noise and superior image quality. It has an active area of 43 x 43cm and features excellent diagnostic images, while delivering the benefits of lower radiation dose to the patient. The FDX4343R achieves a new level of functionality and reliability in developing functional FPDs for system manufacturers. FDX4343R X-Ray FLAT PANEL DETECTOR Digital Radiography Excellent Sensitivity & Resolution Excellent Image Quality Excellent Reliability Product Information PI No. PE-FDX4343R 2010-06-25

Upload: others

Post on 25-Jan-2022

4 views

Category:

Documents


0 download

TRANSCRIPT

The Toshiba FDX4343R X-ray Flat Panel Detector (FPD)

for radiographic applications delivers high sensitivity,

high resolution, low noise and superior image quality.

It has an active area of 43 x 43cm and features excellent

diagnostic images, while delivering the benefits of

lower radiation dose to the patient.

The FDX4343R achieves a new level of

functionality and reliability in

developing functional FPDs for

system manufacturers.

FDX4343RX-Ray FLAT PANEL DETECTOR

Digital Radiography

Excellent Sensitivity & Resolution

Excellent Image Quality

Excellent Reliability

Product Information PI No. PE-FDX4343R 2010-06-25

Sales & Marketing Department1385 Shimoishigami, Otawara-shi, Tochigi, 324-8550 JapanPHONE: +81-287-26-6531 FAX: +81-287-26-6059

http://www.toshiba-tetd.co.jp/

2010-08-1000 FXSS

Toshiba Electron Tubes & Devices Co., Ltd.meets the Environmental Management SystemStandard, ISO14001.

Toshiba Electron Tubes & Devices Co., Ltd.meets internationally recognized Standards forQuality Management System ISO9001,ISO13485.

Characteristics

Dimension

512

50143

495

1

0.9

0.8

0.7

0.6

0.5

0.4

0.3

0.2

0.1

0

MTF

Frequency (Ip/mm)

0 0.5 1 1.5 2 2.5 3 3.5

MTF of FDX4343R

High MTF!

Application General radiography

Scintillator CsI : Tl (direct vapor deposition) + reflective coating

Pixel Pitch 143×143μm

Pixels 3008 (H)×3072 (V) (approx. 9.2 million pixels)

Image Size 16.9 (H)×17.3 inchs (V) (43×43.9 cm)

A /D Conversion 14 bit

Output data 16 bit grayscales tiff

Preview image access time Approx. 4 seconds after X-ray exposure

MTF (2 lp /mm) 36 % typ.

DQE (0) 70 % typ.

Power (Voltage /Frequency) AC100~240V 50/60Hz

Operating environment Sensor Unit : 10-35℃ , 30-80% RH (non-condensing)

Dimensions Sensor Unit : 512×495×43mm

Weight Sensor Unit : 16kg

Standard components Sensor unit, Power unit, Cables

Certification •IEC60601-1,IEC60601-1-2 •MDD93/42/EEC(CEmarking)

FDX4343R Specification

Excellent Sensitivity & Resolution◦ Toshiba’s advanced and proven fine structured CsI/Tl and direct vapor

deposition technology deliver higher sensitivity and better resolution.◦ Reflection coating on CsI/Tl screen enables excellent Detective

Quantum Efficiency (DQE) and low noise.◦ A high Modulation Transfer Function (MTF) FPD is the result of these

outstanding technologies.

Excellent Image Quality◦ Achieves a raw image with a low-noise through the use of Toshiba’s

own circuit technology.◦ Images have a shorter lag time with an improved refresh function,

suitable for radiographic double exposure.◦ Driver software (attached to the product) provides excellent, clear

raw images, adopting real-time offset, gain and defect corrections.

Excellent Reliability◦ Excellent durability by using CsI/TI screen direct vapor deposition

method.◦ The structure is highly reliable and protected from degradation due to

the use of a unique moisture-proof sealing method for the CsI/TI screen.

* The information contained herein is presented only as a guide for the applications of our products.   No Responsibility is assumed by TOSHIBA ELECTRON TUBES & DEVICES CO., LTD. (TETD) for any

infringements of patents or other rights of the third parties which may result from its use. No license is granted by implication or otherwise under any patent or patent rights of TETD or others.

* The information contained herein may be changed without prior notice. It is therefore advisable to contact TETD before proceeding with the design of equipment incorporating this product.

Excellent X-ray image CsI screen technology

(mm)