XN-Series Technology
“What’s Inside the Box?”
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
OBJECTIVES
Describe the technologies available on the XN-Series and the reason(s) for selection
Utilize example case studies to illustrate the new capabilities of PLT-F technology used on the XN-Series
Describe the advanced reagent technology of the RU-20
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
2
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
BEYOND TECHNOLOGY BEYOND EXPECTATIONS
Quality
Dependability
Accuracy
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
greater accuracy, specificity, productivity
NEXT GENERATION
DIAGNOSTICSsuperior insight and control, enhancing speed to treatment
ADVANCED TOOLS & TECHNOLOGIES
automatically balances work-flow and routine tasks
PROCESS OPTIMIZATION
driving improvement,supporting your staff,
helping you plan for the future
HARMONIZED SUPPORT
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
greater accuracy, specificity, productivity
NEXT GENERATION
DIAGNOSTICSsuperior insight and control, enhancing speed to treatment
ADVANCED TOOLS & TECHNOLOGIES
automatically balances work-flow and routine tasks
PROCESS OPTIMIZATION
3
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
THE MOST ADVANCED HEMATOLOGY DIAGNOSTICS ON THE MARKET
Fluorescent Platelet
Rerun/Reflex Testing
Body Fluid Differentials
Advanced Clinical
Parameters
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
XN-SERIES
Same Analytical Module
XN-1000 XN-2000XN-3000 XN-9000
Up to 100 /hr /module
Common Reagent System
Common QC System
Common Software System
XN-1000
SA-01
4
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
SINGLE MODULE – FULLY LOADED
Every XN Configuration can include:
All the parameters
All the features
All the capabilities
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
P R O C E S S O P T I M I Z AT I O N
EXCEPTIONS TO THE RULE: DIFFERENT XN-1000 SAMPLERS
Simple SamplerSA-01
Reflex SamplerSA-10
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
P R O C E S S O P T I M I Z AT I O N
DIFFERENT XN-1000 SAMPLERS
Simple SamplerSA‐01
Reflex SamplerSA‐10
Smart Start
Reflex/Rerun capabilities
Rack bar code read capability
MAP tubes in rack mode
MAP tubes in manual mode
5
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
P R O C E S S O P T I M I Z AT I O N
XN WITH SIMPLE SAMPLER OPTIONS
© 2015 Sysmex America, Inc. All rights reserved.
XN-Series Technology
Analytical Technologies
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
SYSMEX XE VS. XN TECHNOLOGY
RBC, PLT, Hgb
WBC / BASO
NRBC DIFFRET / PLT‐O
RBC, PLT, Hgb
WNR WDF RET PLT‐F
NEW NEW NEW
N E X T G E N E R AT I O N D I A G N O S T I C S
6
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
SYSMEX FOUNDATIONAL TECHNOLOGY
RBC & PlateletDC Detection / Hydrodynamic Focusing
HemoglobinAbsorbance / Sulfolyser
N E X T G E N E R AT I O N D I A G N O S T I C S
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
FLUORESCENT FLOW CYTOMETRY
WNR channel WDF channel PLT-F channel RET channel
N E X T G E N E R AT I O N D I A G N O S T I C S
Side Fluorescent LightDNA/RNA information
Side Scattered LightCell inside structure information
Forward Scattered LightCell size information
Dichroic Mirror
Laser Beam wavelength=633nm
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
WNR CHANNEL
BASO LYMPH MONO Granulocyte
NRBC RBC
Polymethine Dye
7
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
WNR CHANNEL
NRBC
WBC
BASO
Debris
SFL
FS
C NRBCs with EVERY CBC
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
ADVANTAGE OF FLUOROCELL WNR
SFL
FS
C
SFL
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
NRBC CORRELATION DATA
Correlation Coefficient (R) = 0.9179 N = 171
Correlation Coefficient (R) = 0.9219 N = 171
Data from the University of Iowa / Dec 2011
8
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
MEGAKARYOCYTES
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
MEGAKARYOCYTES
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
WNR CHANNEL
Maximized efficiencyNRBC with every CBC
No additional steps or reagents
Accurate WBC counts in the presence of NRBCs
Virtually eliminates interferences from
Lyse resistant RBCs
Lipids
Megakaryocytes
Improves turn around times
BASO
LYMPH
MONO
Granulocyte
NRBC
RBCDebris
FS
C
SFL
9
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
WDF CHANNEL
LYMPH MONO NEUT EO
Atypical LYMPH Immature WBC
Polymethine Dye
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
WDF CHANNEL
LYMPH NEUT+BASO
EO
MONO
Debris
IG
SF
L
SSC
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
IG – XN DIFF ABSTRACT
10
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
ENHANCED FLAGGING ALGORITHMS
(Sysmex Adaptive Flagging Algorithm
based on Shape-recognition)
LDA (Linear Discriminant Analysis)
Detects abnormal cells - (with high sensitivity)
SAFLAS method
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
Normal Abnormal (Blasts?) Abnormal (Abn Lympho?)
WDF SAFLAS METHOD
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
LOW WBC ANALYSIS MODE
WB (Whole Blood) mode LW (Low WBC) mode
Dilution Rate :Analysis Volume :Whole Counts *2 :
Dilution Rate :Analysis Volume :Whole Counts *2 :
x 61.158.2µL9,530
x 61.1174.6µL28,600
Analysis Time extended to 3 times the Whole Blood WBC count time
11
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
LOW WBC ANALYSIS MODE
Analysis time extended to 3 times WB WBC count time
Better accuracy and precision on counts < 0.50 x 103
No Vote Outs – Differential results on all counts
Increased reportable differentials
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
WDF CHANNEL
Improved efficiency More reportable automated differentials
6-part differential with reportable IG%
Rapid TAT
Decreased manual interventions
Enhanced Flagging
Auto-correction of lymphswhen NRBC are presentLYMPH
MONO
NEUT
EO
Atypical LYMPH
Immature WBC
SF
L
SSC
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
BODY FLUID ANALYSIS
Cerebrospinal fluid (CSF)
Body fluid mode (target species)
Pleural fluid
Peritoneal fluid
Synovial fluid
HF-BF
MN
PMN
MO-BF
LY-BF
EO-BFNE-BF
WDF Scattergram
12
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
BODY FLUID ANALYSIS
No pre-analysis preparation
No additional reagents
Automated background counts
Rapid analysis
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
PLT-F CHANNEL
RBC PLT IPF
Oxazine Dye
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
WHY A SECOND PLATELET CHANNEL?
Impedance platelet analysis (size) has limitations
Identification and discrimination of platelets from interfering particles with the same size
13
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
False Increases
False Decreases
BacteriaImmune complexes
WBC fragments
RBC fragments
Microcytes
Giant platelets
Platelet clumps
INTERFERENCES INIMPEDANCE PLATELET COUNTING
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
PLT-F CHANNEL
RBC
PLT-F
IPF
FS
C
SFL
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
GIANT PLATELETS & PLT-F
14
© 2015 Sysmex America, Inc. All rights reserved.
DxH 800 Platelet 114Manual plt estimate 200
© 2015 Sysmex America, Inc. All rights reserved.
DxH 800 Platelet 114Manual plt estimate 200
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
PLT-F CHANNEL
Second method of platelet
Fluorescent Dye specific for platelet organelleExtended count time (6 times) for accurate platelet enumeration, especially in low platelet counts
Good comparison with CD41/CD61
Minimizes interference from RBC fragments, microcytic RBC’s and WBC fragments
Automated reflex with on board rules“BPR systems” only
15
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
RET CHANNEL
RET RBCWBC PLT
Polymethine Dye
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
RET CHANNEL
PLT
RET
RBC
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
RETICULOCYTE PARAMETERS
• Reticulocytes – # and % of immature RBC’s
• Immature Reticulocyte Fraction– Newly released from the
marrow, a direct cellular measurement of erythropoiesis
• Reticulocyte Hemoglobin– Direct cellular measure of iron
availability
16
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
RETICULOCYTE CHANNEL
Clinical UtilityMonitor RBC development at the cellular level
IRF for Retic production
RET-He for effectiveness of hematopoiesis
Clinically relevant information for the management of anemia in conjunction with other available clinical information
Operational EfficiencyReduced Interference From:
WBCs
Howell Jolley Bodies
Parasites
Sickle cells
Quick and automatic
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
XN TECHNOLOGY SUMMARY
WNR ChannelMaximized efficiency
NRBC the first time – all the time
Accurate WBC Counts in the presence of NRBCs
WDF ChannelEnhanced Flagging
6-part differential (including IG)
Improved Sensitivity and Specificity
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
N E X T G E N E R AT I O N D I A G N O S T I C S
XN TECHNOLOGY SUMMARY
Low WBCBetter accuracy and precision on counts < 0.50 x 103
No Vote Outs – Differential results on all counts
Increased Reportable differentials
PLT-FFluorescent Dye specific for platelet organelle
Extended count time (6 times) for accurate low platelet enumeration
Good comparison with CD41/CD61
Automated reflex with on board rules
17
© 2015 Sysmex America, Inc. All rights reserved.
XN Technology
Reagents – RU-20
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
A D VA N C E D T O O L S & T E C H N O L O G I E S
SMART REAGENT MANAGEMENT SYSTEM
RU-20 Reagent Management System25x concentrate
Reduces reagent changes and instrument idle time
Increases free space in the lab.
CELLPACK DCL compared with RU-20 and CELLPACK DST
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
A D VA N C E D T O O L S & T E C H N O L O G I E S
WHY IT MATTERS……
1 cube of diluent weighs ~48lbsAt 300 samples/day you move about 156 cubes/year
RU-20 = 1 cube/50 days~7.2 cubes/year
Model that over 5 years With an average of 3 physical moves of each cube
18
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
A D VA N C E D T O O L S & T E C H N O L O G I E S
Why it matters……
© 2015 Sysmex America, Inc. All rights reserved.
XN Technology
Automation & Process Optimization
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
P R O C E S S O P T I M I Z AT I O N
© 2015 Sysmex America, Inc. All rights reserved.
THE SYSMEX® XN-SERIES brings greater accuracy and efficiency throughout the entire testing chain
S Y S M E X X N - S E R I E S
X N - 1 0 0 0 ™X N - 2 0 0 0 ™
X N - 3 0 0 0 ™
19
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
P R O C E S S O P T I M I Z AT I O N
XN-SERIES ON-BOARD RULES
Enhanced Walk-Away CapabilitiesAutomated Reflex or Rerun testing
No Need To Search For Tube
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
P R O C E S S O P T I M I Z AT I O N
CO-PRIMARY SYSTEM
Fully Automated Workload Balancing
Automatic re-run/reflex testing
XN-2000 and XN-3000
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
P R O C E S S O P T I M I Z AT I O N
AUTOMATED SLIDE MAKING
4th generation system
Fully integrated
Fully reflexive slide preparation
20
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
P R O C E S S O P T I M I Z AT I O N
SLIDEMAKER / STAINER
Improves smear TATRules to define specific slide types
Non-batch staining
Consistent smear quality
Re-cycling reagent system
Provides positive patient ID:3 slide print options available
Low sample volume requirements:
Micro-sample port
Blood aspirated and slide made when indicated
© 2015 Sysmex America, Inc. All rights reserved.© 2015 Sysmex America, Inc. All rights reserved.
P R O C E S S O P T I M I Z AT I O N
© 2015 Sysmex America, Inc. All rights reserved.
THE SYSMEX® XN-SERIES brings greater accuracy and efficiency throughout the entire testing chain
S Y S M E X X N - S E R I E S
X N - 9 0 0 0 ™
© 2015 Sysmex America, Inc. All rights reserved.
Now You Know What’s Inside the Box
Thank You