05-4d ct lung planning

Upload: chanmama

Post on 07-Apr-2018

219 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/4/2019 05-4d Ct Lung Planning

    1/104

    4D CT For Lung Planning4D CT For Lung Planning

    25 June 2009

    Dr Ho Gwo FuangClinical Oncologist

    University of Malaya Medical Centre

  • 8/4/2019 05-4d Ct Lung Planning

    2/104

    LungLung tumourtumouris a MOVING targetis a MOVING target

  • 8/4/2019 05-4d Ct Lung Planning

    3/104

    Problems of Treating a MovingProblems of Treating a Moving

    TargetTarget

    Interfraction movement

    Intrafraction movement

  • 8/4/2019 05-4d Ct Lung Planning

    4/104

    Target Miss or Organ Hit

    Miss Target Hit organ

  • 8/4/2019 05-4d Ct Lung Planning

    5/104

    Respiratory MotionRespiratory Motion

    Traditionally, treatment volume is defined on static CTimage

    Static CT imaging does not precisely define tumour inmotion due to respiration.

    Target may move in and out of treatment field

  • 8/4/2019 05-4d Ct Lung Planning

    6/104

    ICRU 52 and 60ICRU 52 and 60

    CTV

    PTV2 cm

    2 cm

    1 cm1 cm

    GTV is expanded to CTV(microscopic spread) and PTV (set-

    up and other errors)

    To account for respiratory

    movement, margins are added to

    the clinical target volume

    Treated volume

    Increases the normal tissue dose

    and limits the target dose

  • 8/4/2019 05-4d Ct Lung Planning

    7/104

    UnUn--gated Treatmentgated Treatment

    PTV increases to include the target in motion

    for treatment planning and dose delivery Large volume of normal tissue is exposed un-necessarily for high radiation dose

  • 8/4/2019 05-4d Ct Lung Planning

    8/104

    Target MovementTarget Movement -- 3D Solution3D Solution

    Expand the PTV to cover the maximum ranges of

    target motions along all three directions

    Problems: Difficult to generate isodose distributions conforming to

    the moving target such as lung tumor with 3DCRT

    Unable to minimize the doses to the surrounding normal

    tissues

    Therefore limiting total dose and dose fraction sizes

  • 8/4/2019 05-4d Ct Lung Planning

    9/104

    ChallengeChallenge

    Goal How do you accurately deliver appropriate

    dose distributions conforming to a moving

    lung target and meantime effectively

    minimize doses to surrounding normal

    lung tissues?

  • 8/4/2019 05-4d Ct Lung Planning

    10/104

    Solutions for organ or target motionSolutions for organ or target motion

    managementmanagement

    1) Breath-hold technique Radiation is delivered with breath-hold

    2) Tracking technique Radiation is delivered by tracking the motion of thetarget. Dynamic Tracking

    Real time Tracking

    3) Gating technique

    Gated radiation delivery is based on the selected phaseof breathing cycle

  • 8/4/2019 05-4d Ct Lung Planning

    11/104

    Solutions for organ or target motionSolutions for organ or target motion

    managementmanagement

    1) Breath-hold technique Radiation is delivered with breath-hold

    2) Tracking technique Radiation is delivered by tracking the motion of thetarget. Dynamic Tracking

    Real time Tracking

    3) Gating technique

    Gated radiation delivery is based on the selected phaseof breathing cycle

  • 8/4/2019 05-4d Ct Lung Planning

    12/104

    Active Breathing CoordinatorActive Breathing Coordinator

    Valve-controlled cessation of

    inhalation and exhalation duringpredetermined comfortable level ofmoderately deep inspiration

    A

    B

    C

    D

  • 8/4/2019 05-4d Ct Lung Planning

    13/104

    ABC TreatmentABC Treatment

    ABC Immobilise the tumour motion results inreduced margins

    Lower doses to normal tissue

    Prescribed dose can be higher

    Free breathing Breath-hold

  • 8/4/2019 05-4d Ct Lung Planning

    14/104

    Additional benefit of inspirationAdditional benefit of inspiration

    techniques with ABCtechniques with ABC

    Healthy tissue density lung tissue is less dense at deep

    inspiration, therefore less healthy tissue

    in path of irradiation

    Expiration / normal inspiration Deep Inspiration

  • 8/4/2019 05-4d Ct Lung Planning

    15/104

  • 8/4/2019 05-4d Ct Lung Planning

    16/104

    ABC ApparatusABC Apparatus

    Digitalspirometer

    Balloon

    valve

  • 8/4/2019 05-4d Ct Lung Planning

    17/104

    Visual display of breathingVisual display of breathing

    motionmotion

  • 8/4/2019 05-4d Ct Lung Planning

    18/104

    Breath Holding & its ChallengesBreath Holding & its Challenges

    Initial patient set up

    Detecting patient movement during

    treatment

    Patient training

    Patient compliance

    Repeatability of breath holding

    Margins added to compensate for all theabove

    Requires a breath-hold that is:

    reproducible

    consistent

    Accurate

  • 8/4/2019 05-4d Ct Lung Planning

    19/104

    Solutions for organ or target motionSolutions for organ or target motion

    managementmanagement

    1) Breath-hold technique Radiation is delivered with breath-hold

    2)Tracking technique Radiation is delivered by tracking the motion of the

    target. Dynamic Tracking

    Real time Tracking

    3) Gating technique

    Gated radiation delivery is based on the selected phaseof breathing cycle

  • 8/4/2019 05-4d Ct Lung Planning

    20/104

    TrackingTracking

    Dynamic tracking

    e.g. Cyberknife, RTRT system

    Real time tracking

    e.g. Calypso System

  • 8/4/2019 05-4d Ct Lung Planning

    21/104

    ConceptConcept

    External position sensor

    Internal fiducial

  • 8/4/2019 05-4d Ct Lung Planning

    22/104

    Internal movement

    Location of the tumour known using fiducial tracking

    External movement

    Tracking Marker system monitors external movements

    Correspondence model

    Relationship between internal and external movements Continuously follow the internal movement via externalmovement

    Model update continuously throughout the treatment

    ConceptConcept

  • 8/4/2019 05-4d Ct Lung Planning

    23/104

    CyerknifeCyerknife

    Diagnostic

    X-Ray Sources

    ImageDetector

    TreatmentCouch

    LinearAccelerator

    Robotic Arm

  • 8/4/2019 05-4d Ct Lung Planning

    24/104

    Prediction AlgorithmPrediction Algorithm 200ms delay allows robot to smooth jerky offset

    calculations

    Based on pattern searching

    1) Look at the record of model results just before

    2) Compare this pattern with the record of model results over a

    longer period of elapsed time

    3) Find the position at which they match. Sample the model

    position 200ms later this is the prediction

  • 8/4/2019 05-4d Ct Lung Planning

    25/104

    KUKA RobotKUKA Robot

    Made by KUKAof Germany

  • 8/4/2019 05-4d Ct Lung Planning

    26/104

    Solutions for organ or target motionSolutions for organ or target motion

    managementmanagement

    1) Breath-hold technique Radiation is delivered with breath-hold

    2) Tracking technique Radiation is delivered by tracking the motion of the

    target. Dynamic Tracking

    Real time Tracking

    3) Gating technique

    Gated radiation delivery is based on the selected phaseof breathing cycle

  • 8/4/2019 05-4d Ct Lung Planning

    27/104

    Beam Off

    Beam OffBeam On

    Beam On

    Treatment Field

    1.1. 2.2.

    3.3. 4.4.

    Gating -

    Treatment beam is turned on and off as tumor enters and exits

    a static treatment field

    GatingGating

  • 8/4/2019 05-4d Ct Lung Planning

    28/104

    GatingGating

    Respiration-gated radiotherapy offers asignificant potential for improvement in the

    irradiation of tumour sites affected by

    respiratory motion such as lung, breast andliver tumours.

    An increased conformality of irradiationfields leading to decreased complications

    rates of organs at risk

  • 8/4/2019 05-4d Ct Lung Planning

    29/104

    Respiratory motion solutionsRespiratory motion solutions Breath-hold techniques (e.g. ABC)

    Uncomfortable for patients, limited applicability (MSKCC: 7/13patients)

    Increases treatment time (MSKCC: 17 to 33 minutes for conventionalRT)

    Respiratory gating Residual motion within

    gating window

    Increases treatment time Baseline shift

    4D Radiotherapy

    Hardware/Softwarecomplexity

    -0.6

    -0.4

    -0.2

    0

    0.2

    0.4

    0.6

    0 5 10 15 20

    Time (s)

    Displacement(cm)

    Exhale gate

    Inhale Gate

  • 8/4/2019 05-4d Ct Lung Planning

    30/104

    4D Radiotherapy4D Radiotherapy

    The explicit inclusion of the temporal changesin anatomy during the imaging, planning and

    delivery of radiotherapy

  • 8/4/2019 05-4d Ct Lung Planning

    31/104

    4D Solution for Organ Motion4D Solution for Organ Motion

    4D CT provides insight into organ motionduring respiration, with volumetric anatomic

    data set

    Treatment planning explicitly accounts for

    the internal target motion

    This can be implemented at various levels ofcomplexity

  • 8/4/2019 05-4d Ct Lung Planning

    32/104

    Utilization of 4D CT in RadiotherapyUtilization of 4D CT in Radiotherapy

    Treatment Planning

    Image Acquisition & Registration Acquisition of a sequence of CT

    image sets over consecutive phases of abreathing cycle

    Designing treatment plans on CT

    image sets obtained for each phase

    of the breathing cycle

    Continuous delivery of the 4D

    treatment plans throughout the breathingcycle

    Treatment Delivery

  • 8/4/2019 05-4d Ct Lung Planning

    33/104

    CT Motion ArtifactsCT Motion Artifacts

    CT data acquisition is serial

    Data at adjacent couch positions are acquired serially

    Collection of projection data one slice afteranother in combination with motion of the

    scanned object leads to significant interplay

    Depending on the relative motion of the advancing

    scan plane and the tumour, different artifacts can be

    imaged

  • 8/4/2019 05-4d Ct Lung Planning

    34/104

    3D CT - Distorted images, incorrect anatomicalDistorted images, incorrect anatomical

    positions, volumes or shapespositions, volumes or shapes

  • 8/4/2019 05-4d Ct Lung Planning

    35/104

    Gated CT

    Keall et alAust Phys Eng Sci Med 2002

  • 8/4/2019 05-4d Ct Lung Planning

    36/104

    Standard helical CT scan acquiredunder light breathing

    One respiratory phase of a 4D CTscan

  • 8/4/2019 05-4d Ct Lung Planning

    37/104

    4D

    Images

  • 8/4/2019 05-4d Ct Lung Planning

    38/104

    4D CT Scan4D CT Scan

    Assumptions Organ motions such as lungs are related to the motionsof an external marker

    Concepts If patients can breathe periodically and regularly, the

    CT image acquisition is fast enough to generate many

    images at all phases in a series of respiratory cycles

    When the scan is done, all the images of the selectedphase are retrospectively organized to form 4D video

    images

  • 8/4/2019 05-4d Ct Lung Planning

    39/104

    4D CT: Data Acquisition (1)4D CT: Data Acquisition (1)

    4D CT technique images multiple respiratory

    states within one data acquisition

    Temporally oversampling data acquisition at each

    couch position CT tube rotates continuously for the duration of the

    respiratory cycle and acquire projection data thorugh all

    respiratory states (typically 10-20)

    Typically 5-10 revolutions during the respiratory cycle

    is achieved

  • 8/4/2019 05-4d Ct Lung Planning

    40/104

    4D CT: Data Acquisition (2)4D CT: Data Acquisition (2)

    A 4D CT scan consisting of a series of 3D CT

    image sets acquired at different respiratory phases

    Typically 10-20 images per slice are reconstructed,

    representing 10-20 respiratory phase states This results in a total number of images between

    1000 and 2000 per 4D CT study

    The acquisition time decreases linearly with

    number of rings of detectors

  • 8/4/2019 05-4d Ct Lung Planning

    41/104

    External Sorting SignalExternal Sorting Signal

    After 4D CT data acquisition and image

    reconstruction

    in order to sort these images into specific temporallycoherent volumes, additional information is required

    One such sorting signal is the rise and fall of theabdominal surface, as a surrogate for respiratory

    motion

  • 8/4/2019 05-4d Ct Lung Planning

    42/104

    4D CT imaging4D CT imaging

  • 8/4/2019 05-4d Ct Lung Planning

    43/104

    Retrospectively Reconstructed CT SlicesRetrospectively Reconstructed CT Slices After data acquisition, a software (e.g. GE

    Advantage4D software) is used to retrospectivelysort the images into multiple temporally coherentvolumes

    The software loads the 4D CT images as well asthe respiratory trace (recorded by the RPM system)

    Based on the data acquisition time stamps in theimage Dicom headers and the correlation signal inthe RPM trace, a specific respiratory phase can be

    assigned to each image

  • 8/4/2019 05-4d Ct Lung Planning

    44/104

    Respiration Waveform from

    RPM Respiratory Gating System

    X-ray on

    Exhalation

    Inhalation

    First couch position Second couch position Third couch position

    Image acquired

    signal to RPM

    system

    Retrospective 4D-CT imaging

    TinsuTinsu PanPan

    Respiratory Bin

    R i 4D I A i i i

  • 8/4/2019 05-4d Ct Lung Planning

    45/104

    0:00

    00.0 cm

    Retrospective 4D Image Acquisition

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    R i 4D I A i i i

  • 8/4/2019 05-4d Ct Lung Planning

    46/104

    0:01

    00.0 cm

    Retrospective 4D Image Acquisition

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    R i 4D I A i i i

  • 8/4/2019 05-4d Ct Lung Planning

    47/104

    0:02

    00.0 cm

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    Retrospective 4D Image Acquisition

    R t ti 4D I A i iti

  • 8/4/2019 05-4d Ct Lung Planning

    48/104

    0:03

    00.0 cm

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    Retrospective 4D Image Acquisition

    R t ti 4D I A i iti

  • 8/4/2019 05-4d Ct Lung Planning

    49/104

    0:04

    00.0 cm

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    Retrospective 4D Image Acquisition

    R t ti 4D I A i iti

  • 8/4/2019 05-4d Ct Lung Planning

    50/104

    0:05

    00.0 cm

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    Retrospective 4D Image Acquisition

    R t ti 4D I A i iti

  • 8/4/2019 05-4d Ct Lung Planning

    51/104

    0:06

    00.0 cm

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    Retrospective 4D Image Acquisition

    R t ti 4D I A i iti

  • 8/4/2019 05-4d Ct Lung Planning

    52/104

    0:00

    01.0 cm

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    Retrospective 4D Image Acquisition

    R t ti 4D I A i iti

  • 8/4/2019 05-4d Ct Lung Planning

    53/104

    0:01

    01.0 cm

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    Retrospective 4D Image Acquisition

    R t ti 4D I A i iti

  • 8/4/2019 05-4d Ct Lung Planning

    54/104

    0:02

    01.0 cm

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    Retrospective 4D Image Acquisition

    Retrospecti e 4D Image Acq isition

  • 8/4/2019 05-4d Ct Lung Planning

    55/104

    0:03

    01.0 cm

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    Retrospective 4D Image Acquisition

    Retrospective 4D Image Acquisition

  • 8/4/2019 05-4d Ct Lung Planning

    56/104

    0:04

    01.0 cm

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    Retrospective 4D Image Acquisition

    Retrospective 4D Image Acquisition

  • 8/4/2019 05-4d Ct Lung Planning

    57/104

    0:05

    01.0 cm

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    Retrospective 4D Image Acquisition

    Retrospective 4D Image Acquisition

  • 8/4/2019 05-4d Ct Lung Planning

    58/104

    0:06

    01.0 cm

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    Retrospective 4D Image Acquisition

    Retrospective 4D Image Acquisition

  • 8/4/2019 05-4d Ct Lung Planning

    59/104

    0:00

    02.0 cm

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    Retrospective 4D Image Acquisition

    Retrospective 4D Image Acquisition

  • 8/4/2019 05-4d Ct Lung Planning

    60/104

    0:01

    02.0 cm

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    Retrospective 4D Image Acquisition

    Retrospective 4D Image Acquisition

  • 8/4/2019 05-4d Ct Lung Planning

    61/104

    0:02

    02.0 cm

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    Retrospective 4D Image Acquisition

    Retrospective 4D Image Acquisition

  • 8/4/2019 05-4d Ct Lung Planning

    62/104

    0:03

    02.0 cm

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    Retrospective 4D Image Acquisition

    Retrospective 4D Image Acquisition

  • 8/4/2019 05-4d Ct Lung Planning

    63/104

    0:04

    02.0 cm

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    Retrospective 4D Image Acquisition

    Retrospective 4D Image Acquisition

  • 8/4/2019 05-4d Ct Lung Planning

    64/104

    0:05

    02.0 cm

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    Retrospective 4D Image Acquisition

    Retrospective 4D Image Acquisition

  • 8/4/2019 05-4d Ct Lung Planning

    65/104

    0:06

    02.0 cm

    4 x 2.5 mm Multi Slice

    (10 mm total coverage)

    X-ray Tube

    Table location

    Acq. Time

    Retrospective 4D Image Acquisition

  • 8/4/2019 05-4d Ct Lung Planning

    66/104

    4D CT scan Acquired

  • 8/4/2019 05-4d Ct Lung Planning

    67/104

    Identify the motion

    Decide the treatment window with min.displacement and max. beam-on time

  • 8/4/2019 05-4d Ct Lung Planning

    68/104

    Contour CTV and critical structures on one or

    several phases of CT images in the window

    Determine the max. displacement of CTV of allphases in the window

  • 8/4/2019 05-4d Ct Lung Planning

    69/104

    Treatment planning with IGTV/ ICTV

    Beam placements and calculations for treatment plan Plan review, approval, and validation

  • 8/4/2019 05-4d Ct Lung Planning

    70/104

    4D CT planning4D CT planning

    The explicit inclusion of temporal effects in

    radiotherapy treatment planning is referred to as

    4D treatment planning

    Intrafrational motion can be included in treatment

    planning at different levels

  • 8/4/2019 05-4d Ct Lung Planning

    71/104

    4D CT planning4D CT planning Simplest solution:

    To generate a composite target volume thatencompass the CTV throughout organ motion

    during the respiratory cycle

    Advanced:

    Multiple dose calculations & deformable

    registration

  • 8/4/2019 05-4d Ct Lung Planning

    72/104

    Targets DelineationTargets Delineation

    With the 4D CT dataset, we can design the

    internal gross target volume (IGTV), that is the

    volume containing the GTV throughout its motion

    during respiration

    One method of combining the data from the

    multiple CT datasets is to create a maximalintensity projection, which can be used as an aid in

    contouring the IGTV

  • 8/4/2019 05-4d Ct Lung Planning

    73/104

    MIP and Average CTMIP and Average CT

    The MIP (or MIV) is a 3D CT dataset created by

    assigning each voxel the value of the highest value

    voxel at that location across the breathing phases

    The average CT is a 3D CT dataset created by

    performing a voxel-by-voxel numerical averaging

    over all the breathing phases

  • 8/4/2019 05-4d Ct Lung Planning

    74/104

    4D CT simulation demosntrated tumour motion during breathing cycle. (A) End of Inspiration, (B) End

    of Expiration, (C)Average CT, (D) MIP. In this patient, MIV image was used to design IGTV

  • 8/4/2019 05-4d Ct Lung Planning

    75/104

    Targets DelineationTargets Delineation

    Another approach is to contour the GTV with the

    end of inspiration and expiration breath-holding

    and then combine these two volumes to form the

    IGTV this approach can be used with regularspiral CT without 4D

    All CT databases are transferred to the treatment-planning system for reference.

  • 8/4/2019 05-4d Ct Lung Planning

    76/104

    4D CT inhale /exhale fusion4D CT inhale /exhale fusion

    coronalcoronal saggitalsaggital

    Maria Hawkins, RMH

  • 8/4/2019 05-4d Ct Lung Planning

    77/104

    Treatment PlanningTreatment Planning

    All 10 respiratory-phase datasets, the MIP, and the

    average CT along with extended range free-

    breathing CT acquired during the same imaging

    session are transferred to the treatment planningsystem

    The information is crucial for target delineatingusing the internal taget volume (ITV) approach to

    take tumour motion into consideration

  • 8/4/2019 05-4d Ct Lung Planning

    78/104

    4D4D vsvs 3D Target Volumes3D Target Volumes

    4D target volumes differ from those derived

    from conventional helical scanning

    Differences: the shapes of volumes of

    interest and their centroids change more

    accurate from 4D CT

  • 8/4/2019 05-4d Ct Lung Planning

    79/104

    3D BEV 4D BEV

    BEVsBEVs

  • 8/4/2019 05-4d Ct Lung Planning

    80/104

    MLC leaf motionMLC leaf motion

    3D 4DKeall et alPMB 2001 46:1-10

  • 8/4/2019 05-4d Ct Lung Planning

    81/104

    Image RegistrationImage Registration Another approach is to use a deformable

    registration technique in which the tumour volumeoutlined on the expiratory phase of the 4D imagesis registered on other phases of the images to

    create a union of target contours, enclosing allpossible positions of the target

    Mathematically, this is to find the transformation matrix, that maps

    an arbitrary point from the fixed image to the corresponding pointon the floating image (or vice versa)

    e.g. using a freeware tool vtkCISG (Hartkens et al 2002)

    The resulting IGTV contour is then evaluatedacross all phases

  • 8/4/2019 05-4d Ct Lung Planning

    82/104

    Treatment Planning with ImageTreatment Planning with Image

    RegistrationRegistration

    A treatment plan was created on the end-exhale

    CT image set and then automatically created on

    each of the 3D CT image sets corresponding with

    subsequent respiration phases, based on the beam

    arrangement and dose prescription in the end-

    exhale plan.

    Dose calculation using e.g. Monte Carlo, issimultaneously performed on each of the 3D

    image sets

  • 8/4/2019 05-4d Ct Lung Planning

    83/104

    Treatment PlanningTreatment Planning

    The dose distribution from each respiratory phase

    CT image set was mapped back to the end-exhale

    CT image set for analysis

    4D dataset therefore provides the ability to study theimpact of respiratory motion on dose distribution

    The use of deformable image registration to merge all

    the statistically noisy dose distributions back onto one

    CT image set effectively yields a 4D Monte Carlo

    calculation with a statistical uncertainty equivalent to a

    3D calculation

  • 8/4/2019 05-4d Ct Lung Planning

    84/104

    4D Imaging and Treatment Planning; Eike Rietzel and George T.Y. Chen

    3D ( lid) 4D (d h d) DVH3D ( lid) 4D (d h d) DVH

  • 8/4/2019 05-4d Ct Lung Planning

    85/104

    3D (solid) vs 4D (dashed) DVHs3D (solid) vs 4D (dashed) DVHs

    4D Planning Flow Chart

  • 8/4/2019 05-4d Ct Lung Planning

    86/104

    Acquire 4D CT

    Define anatomy

    Create/adjust treatment plan

    Evaluate dose distribution

    1

    4

    3

    2

    Proceed to treatment6

    Plan acceptable?No

    Yes

    Deformableregistration

    Automatedplanning

    Defor

    mable

    registra

    tion

    5

  • 8/4/2019 05-4d Ct Lung Planning

    87/104

    Treatment deliveryTreatment delivery

    Treatment delivered to the planned volume

    Treatment can be delivered with a gated beam

    control

  • 8/4/2019 05-4d Ct Lung Planning

    88/104

    Beam ONBeam ONBeam OFFBeam OFF

    Respiratory Gated Radiation Therapy

  • 8/4/2019 05-4d Ct Lung Planning

    89/104

  • 8/4/2019 05-4d Ct Lung Planning

    90/104

    4D Solutions4D Solutions

  • 8/4/2019 05-4d Ct Lung Planning

    91/104

    4D Solutions4D SolutionsUnUn--gated Acquisitiongated AcquisitionGated TreatmentGated Treatment

    4D (x,y,z,t) CT (GE Adv4D)

    allows to acquire all images in free breathing.

    divides the respiratory cycle in various phases.

    Gated window (Varian RPM) is set for minimum tumour

    position uncertainty and maximum beam-on time interval

  • 8/4/2019 05-4d Ct Lung Planning

    92/104

    4D radiotherapy delivery4D radiotherapy delivery

    Linac Controller MLC Workstation

    MLC Controller4DC

    Tracking Signal

    Treatment

    parameters

    Linac Controller MLC Workstation

    MLC Controller4DC

    Tracking Signal

    Treatment

    parameters

    GE Avd4D CT Scanner withGE Avd4D CT Scanner with

  • 8/4/2019 05-4d Ct Lung Planning

    93/104

    GE Avd4D CT Scanner with

    VarianVarians RPM Respiratory Gating Systems RPM Respiratory Gating System

  • 8/4/2019 05-4d Ct Lung Planning

    94/104

    Gated IMRT for LungGated IMRT for Lung

    Gating for Liver IrradiationGating for Liver Irradiation

  • 8/4/2019 05-4d Ct Lung Planning

    95/104

    Maria Hawkins, Royal Marsden Hospital

    Gating for Liver IrradiationGating for Liver Irradiation

    Liver motion, 1-5 cm, increases volume of normal

    tissue irradiated

    Respiratory gating reduces volume of liverirradiated

    GTVPTV

    Volume to

    be irradiated

    Free breathing Breath hold RT

  • 8/4/2019 05-4d Ct Lung Planning

    96/104

    What is The Next Exciting NewDimension in IGRT ?

    4D PET/CT4D PET/CT

    f /C h l i

  • 8/4/2019 05-4d Ct Lung Planning

    97/104

    Impact of PET/CT on Therapy PlanningImpact of PET/CT on Therapy Planning

    PET/CT helps to find unsuspected involvementin the mediastinum

    Can PET/CT help to re-define the treatment volume of a

    primary tumour ???

    Images courtesy of Community Cancer Center, FL

    CT PET PET/CT Fusion

    Gated PETGated PET

  • 8/4/2019 05-4d Ct Lung Planning

    98/104

    time

    7

    3

    4 5

    6

    8

    3

    45

    6

    7

    Bin 8

    8

    2

    Trigger

    1

    Bin 1

    2

    1

    Trigger

    Prospective fixed forward time binning (DLS&DST)

    Ability to reject cycles that dont match

    Single 15 cm FOV Gated PET

    Helical CT attenuation correction

    Gated PETGated PETC i

  • 8/4/2019 05-4d Ct Lung Planning

    99/104

    Images courtesy of Holy Name Hospital

    PET MIP

    4D PET MIP

    Primary tumor

    ?

    CT PET Fusion

    Trans.

    Coronal

    4D PET Coronal

    GatedGated acqacq. statistics. statistics Single FOVSingle FOV

    Helical CTACHelical CTAC

    10 minute scan duration10 minute scan duration

    8 respiratory gated bins8 respiratory gated bins

    ImpactImpact

    Max Intensity project (MIP)Max Intensity project (MIP)

    Increased quantitativeIncreased quantitative

    accuracyaccuracy

    Motion assessmentMotion assessment

  • 8/4/2019 05-4d Ct Lung Planning

    100/104

    4D Radiotherapy: Caveats4D Radiotherapy: Caveats Other variables exist:

    Cardiac motion

    Interfractional motion set up errors, variations inphysiologic state (e.g. stomach size)

    The data acquired in 4D CT is synthesized frommultiple breaths during an acquisition time of afew minutes

    Reproducitiliby of this pattern during each treatmentfraction is implicitly assumed when analyzing theseresulting 4D dose distributions

    Possible variations may be monitored by examining therespiratory trace on a daily basis

  • 8/4/2019 05-4d Ct Lung Planning

    101/104

    4D Radiotherapy: Looking Ahead4D Radiotherapy: Looking Ahead Techniques to deliver 4D treatment is still being developed

    and refined to take full advantage of the knowledgeprovided by 4D CT

    Image guided therapy in the treatment room could wellinclude 4D cone beam CT, if appropriate

    The aim is to mitigate the dose-perturbing effects of motion,and possibly lead to safe decrease of geometric margins andincreased therapeutic gain

    Incorporation of functional imaging informations

  • 8/4/2019 05-4d Ct Lung Planning

    102/104

    Respiratory motion causes problems during the

    imaging, planning and treatment stages ofradiotherapy

    Several methods have been proposed to address

    respiratory motion:1) Target positioning

    2) Robotic tracking

    3) Real-time monitoring

    4) 4D planning & gating

    4D radiotherapy has some advantages over existingmethods

    There are still many unanswered questions

    ConclusionsConclusions

  • 8/4/2019 05-4d Ct Lung Planning

    103/104

    AcknowlegementAcknowlegement I am indebted to Professor Andrew Wu, PhD, of

    Department of Radiologic Sciences,Thomas JeffersonUniversity, Philadelphia, Pennsylvania, Varian, Elekta and

    Siemens for lending me slides & videos for this talk.

  • 8/4/2019 05-4d Ct Lung Planning

    104/104

    Thank You !Thank You !