1 staff progress report david urner university of oxford

35
1 StaFF Progress Report David Urner University of Oxford

Upload: roberta-charles

Post on 16-Jan-2016

218 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: 1 StaFF Progress Report David Urner University of Oxford

1

StaFF Progress Report

David Urner

University of Oxford

Page 2: 1 StaFF Progress Report David Urner University of Oxford

2

Cast

Dr. David Urner Dr. Paul Cole Dr. Armin Reichold

Stephanie Yang(mechanical engineering)

Tony Handford(workshop)

Roy Wastie(electrical engineering)

Page 3: 1 StaFF Progress Report David Urner University of Oxford

3

Measuring Motion

• At the ILC beam delivery system many magnets have to be stable with respect to each other to achieve high luminosities. – Final focus doublet– Critical magnets in BDS– Position monitoring of BPM’s in energy chicane.

• Often no direct line of sight:– Correlate position information of magnet to stable

platform (e.g. anchored in ground) interferometrically.– Can be coupled with very large accelerometer

performing better at small frequencies.– Correlate the stable platforms interferometrically.

Page 4: 1 StaFF Progress Report David Urner University of Oxford

4

Generic Tools

• Straightness monitor

• Distance meter (build first)

Page 5: 1 StaFF Progress Report David Urner University of Oxford

5

Distance Meter: Method of Measurement

• Distance meter 2 modes:– Michelson mode:

• fast, • relative distances • resolution nm

– FSI mode:• slow• Absolute distances • Precision 1m.

DistanceMeters

Laser Reference Interferometer

Wavelength: 1550nm DAQ

Pump(optical amplifier)

• Keep costs down for distance meters so that overall cost scale favourably.

Page 6: 1 StaFF Progress Report David Urner University of Oxford

6

Laser tune

I

The Distance Meter: FSI Mode

• Blue (long) arm reflected at Retro-reflector returning light at same angle– Slightly defocus lens → returning light is spread to ~1mm circle at launch plane.

• Red (short arm) reflected at far end of lens– Both arms cross same amount of material (1. order) – Close end of lens has to be anti-reflection coated.– Chose lens: short arm is reflected into small region.

• Red (short arm) and blue (long arm) interfere.• FSI: Needs only one return line.• Tune laser from 1530-1560 nm.

– #wavelength-#wavelength changes – For constant tuning speed: constant FSI frequency.

Page 7: 1 StaFF Progress Report David Urner University of Oxford

7

The Distance Meter: FSI Mode

• Measure fringe frequency using fast Fourier transformation.– #fringes * frequency → total phase advance.– Known effective length of reference → effective length of

distance meter.– Fourier spectrum measures all frequencies → all interferences at

all distances!

I

0.3 0.4 0.5

0

5

10

15

Am

plitu

de /

Arb

uni

ts

Length / m

Page 8: 1 StaFF Progress Report David Urner University of Oxford

8

Laser Reference Interferometer

DistanceMeters

Wavelength: 1550nmTuning: 1530-1560

DAQ

FSI: Reference Interferometer

• Laser: Constant tuning speed?– Unfortunately no.

• Reference interferometer developed for LiCAS.• Use Reference interferometer to unfold phase advance. • Then correct phase information from all distance meters.

time

Phas

e

Page 9: 1 StaFF Progress Report David Urner University of Oxford

9

The Distance Meter: Michelson mode

• Length motion leads to change in interference pattern– Measure intensity I.

IFixed Laser Frequency

Page 10: 1 StaFF Progress Report David Urner University of Oxford

10

The Distance Meter: Michelson mode

• Length motion leads to change in interference pattern– Measure intensity I.

• Add more lines– Each line will have another path length difference.– 4 lines enough to calculate exact motion.

Fixed Laser Frequency

Page 11: 1 StaFF Progress Report David Urner University of Oxford

11

Pay Attention to Systematic Effects

• Understanding the behaviour of laser is key!

• FSI mode: – Can we get better handle on tuning speed?

Reference Interferometer

DistanceMeters

Wavelength: 1550nmTuning: 1530-1560nm

DAQ

Laser

Page 12: 1 StaFF Progress Report David Urner University of Oxford

12

Piezzo Driven Monitor

Detector

Detector

PZT

MovingMirror

Mirror

Mirror

Mirror

Splitter

Splitter

Splitter

Tuning laser

SinglePointmeasurement

980 1000 1020 1040 1060 1080

-4

-3

-2

-1

0

1

2

3

4

Phase stepping cycle

Laser is tuning continuously

Pha

se

/rad

ian

s

(Derive phases modulo 2from intensity data)

Moving mirror: Measure Intensity pattern

Page 13: 1 StaFF Progress Report David Urner University of Oxford

13

3–D mechanical model, detector side removed for clarity

Page 14: 1 StaFF Progress Report David Urner University of Oxford

14

Systematic Effects: Laser

• Understanding the behaviour of laser is key!

• FSI mode: – Can we get better handle on tuning speed?

Reference Interferometer

DistanceMeters

Wavelength: 1550nmTuning: 1530-1560nm

DAQ

Laser

Piezzo driven monitor

Page 15: 1 StaFF Progress Report David Urner University of Oxford

15

Systematic Effects: Laser

• Understanding the behaviour of laser is key!• Michelson mode:

– Stable frequency needed (unbalanced arm length) at level ~30kHz!

• Lock Laser to absorption line (very hard at 1550nm).• Equip reference with Michelson Mode readout. → Change of

reference interferometer information measures frequency change (assuming length is stable).

Reference Interferometer

DistanceMeters

Wavelength: 1550nmTuning: 1530-1560nm

DAQ

Laser

Piezzo driven monitor

Lock to absorption line

Page 16: 1 StaFF Progress Report David Urner University of Oxford

16

Other Systematic Effects

• FSI mode: – Length change of distance meter during tuning (1nm → ~40nm error).

• Scan rapidly → Piezzo driven monitor will not work!• Use Michelson information of distance meter to track length change.• Use both methods.

• Vacuum enclosure of distance meters needed.• Temperature effects.

Reference Interferometer

DistanceMeters

Wavelength: 1550nmTuning: 1530-1560nm

DAQ

Laser

Piezzo driven monitor

Lock to absorption line

Page 17: 1 StaFF Progress Report David Urner University of Oxford

170 10 20 30 40

-5

0

5

10

15

0 10 20 30 40-2

0

2

Te

mp

era

ture

ch

an

ge

/ m

K

Time / s

Re

f V

/ m

V

Time / s

Reference channels Thermometer channels

First Temperature Measurements

Page 18: 1 StaFF Progress Report David Urner University of Oxford

18

Status of Distance Meter Development

• Double peak found in two different prototypes.

• Ruled out possibility of analysis artefact.

• No obvious reflective surfaces

• Software in place now to analyse data within minutes after data taking should enable us to trace the problem

0.3 0.4 0.5

0

5

10

15

Am

plitu

de /

Arb

uni

ts

Length / m

Raw data of 2 channels recorded simultaniously

Fourier spectrum

Page 19: 1 StaFF Progress Report David Urner University of Oxford

19

Distance meter simulation

• Simulation done with Zemax

• Use non-sequential mode– Take into account

polarisation → correct interference pattern

– Allows stray light analysis

• Allow analysis of chromatic aberrations

Page 20: 1 StaFF Progress Report David Urner University of Oxford

20

Distance meter simulation

• Simulation done with Zemax

• Use non-sequential mode– Take into account

polarisation → correct interference pattern

– Allows stray light analysis

• Allow analysis of chromatic aberrations

Page 21: 1 StaFF Progress Report David Urner University of Oxford

21

Simulated Interference Patters

Page 22: 1 StaFF Progress Report David Urner University of Oxford

22

A Straightness Monitor Made from Distance Meters

Setup planned at KEK

• Red lines: Distance meter. • Multilateration measure 6D coord. of A with respect to B.

A

B

Page 23: 1 StaFF Progress Report David Urner University of Oxford

23

A Straightness Monitor Made from Distance Meters

• Information related via central triangle

Floor node

A

B

Ceiling node 1Ceiling node 1

Page 24: 1 StaFF Progress Report David Urner University of Oxford

24

A Straightness Monitor Made from Distance Meters

• 3 nodes on each object, with 3 distance meters to each triangle node

Floor node

A

B

Ceiling node 1Ceiling node 1

Page 25: 1 StaFF Progress Report David Urner University of Oxford

25

A Straightness Monitor Made from Distance Meters

• 3 nodes on each object, with 3 distance meters to each triangle node

Floor node

A

B

Ceiling node 1Ceiling node 1

Page 26: 1 StaFF Progress Report David Urner University of Oxford

26

ATF at KEK

Page 27: 1 StaFF Progress Report David Urner University of Oxford

27

Implement system at ATF/KEK relating positions of nano-BPM’s

• Advantage: – Nano-BPM have 5-100 nm resolution: cross check of results– Test of distance meter in accelerator environment

Nano-BPM Built by SLAC group Nano-BPM

Built byKEK group

Page 28: 1 StaFF Progress Report David Urner University of Oxford

28

Spider web Design with Opto-Geometrical Simulation: Simulgeo

Page 29: 1 StaFF Progress Report David Urner University of Oxford

29

Spider web Design with Opto-Geometrical Simulation: Simulgeo

• Allows objects to be placed (6D) in hierarchal structure– Reference placements.– Fixed placements (with error).– Variable placements (the

objects to measure).

• Objects can be points, mirrors, distance meters…– Distance meter assume

measurement between points with error.

• Big matrix inversion takes into account all errors and constrains 6D position of all points.

Page 30: 1 StaFF Progress Report David Urner University of Oxford

30

Spider web Design with Opto-Geometrical Simulation: Simulgeo

• Resolution of distancemeter: 1nm

• Mount precision of distancemeter: 1nm

• Angle precision of distancemeter holder: 10 rad.

SLAC BPM: referenceKEK BPM variable (6D):

Position: x:32 y:19 z:2 nmAngle: x:0.01 y:0.01 z:0.1 rad

~1m absolute distance resolution needed to determine constants required to solve geometry.

Page 31: 1 StaFF Progress Report David Urner University of Oxford

31

Triangle Nodes

• Distance meter heads located in triangle nodes.

• Floor node– Overall resolution improves if

firmly anchored.– Dome anchored separately

from interferometers.

• Ceiling nodes: position stability unimportant.

Page 32: 1 StaFF Progress Report David Urner University of Oxford

32

First Concept on how to Align Distance Meters in Network

Page 33: 1 StaFF Progress Report David Urner University of Oxford

33

BPM Nodes

• One wide angle retro- reflector (cateye) for each node

• Challenges: – Relative position between

retro-reflector needs to be known to 1nm

• Requires measurement between 3 nodes on each nano-BPM.(blue lines).

– Attachment of vacuum lines to BPM’s

• Requires zero-force design.

Page 34: 1 StaFF Progress Report David Urner University of Oxford

34

Force Free Mount

Here attach vacuum tube for interferometerAttached to BPM.

Holds retro reflector.

Firmconnection

Strain Gauge• Needs bellow to allow motion of BPM– Vacuum causes a force

order of 100N!

• Develop small force vacuum mount using double bellow system.

• Allows small motion (~1 mm) of BPM-system

• Test stand to measure remaining (perpendicular) force on BPM frame. -1mm-3mm 1mm 3mm

1N

2N

-1N

0N

-2N

-3N

-3N

Force exerted on carbon frame (BPM)±1mm: < 0.5N/mm±3mm: < 0.8N/mm

Force exerted by perpendicular motion

Page 35: 1 StaFF Progress Report David Urner University of Oxford

35

Concluding Remarks

• Developing– Software to understand distance meter.– Hardware to characterize laser.– Temperature sensing system.

• First optical simulation in place.• Force Free mount system seems to work.• Starting work on Mount/Alignment system for

distance meter setup at KEK• Still much to do

– but things start to fall into place