131210415 plusoptix folder a09 gb:210108403 … generation documentation pediatric autorefractor...
TRANSCRIPT
¸ Designed for infants, children and uncooperative patients
¸ From one meter away in less than 1 second
Measur ing pr inc ip le
MEASUREMENT
Spherical range* -7.0/+5.0 dpt in 0.25 dpt steps
Cylindrical range* -7.0/+5.0 dpt in 0.25 dpt steps
Axis 1-180° in 1° steps
Pupil size 4.0 – 8.0 mm in 0.1 mm steps
Acquisition time dynamic, in average 0.8 seconds
Measuring distance 1 Meter (+/- 5 cm)
Fixation target Warble sound
SYSTEM SPECIFICATIONS
Monitor resolution 1024 x 768 pixel
Interfaces 4 x USB, 1 x VGA and 1 x RJ -45
Voltage 100 – 120 V / 220 – 240 VAC
Frequency 50 – 60 Hz
Standards EN 60601
Certified CE, FDA and CMDCAS
*) Binocular and monocular Spherical equivalent
This information has been handed over by:
[ ]
Pediatric AutorefractorplusoptiX A09
Binocular refraction measurement
plusoptiX A09 EmilyplusoptiX A09 Emily
The measuring principle is based on eccentric photo-retinos-copy. Infrared light is projected through the pupils onto theretina. Depending on the refractive error, the reflected lightforms a specific brightness pattern within the pupil. The sphe-rical refraction is calculated based on this crescent pattern.To determine cylinder and axis the same measurement isrepeated in three meridians. The measurement with infraredlight is completely innocuous. Infrared light is also contained indaylight and not visible.
Made in Germany
January 2011
Plusoptix GmbHNeumeyerstr. 4890411 NuernbergGermanyTel.: +49-911-598 399 - 10Fax: +49-911-583 399 - 90 www.plusoptix.eu
ContactFor further information please contact our sales repre-sentative in your country:
www.plusoptix.eu/salespartner
Connect ion to :Á Monitor
Á Mouse and keyboard
Á Computer network
Á Printer
No computer required.
131210415_Plusoptix_Folder_A09_GB:210108403_PlusOptix_Flyer_4000 15.12.10 11:45 Seite 1
[New Generat ion
Documentat ionPediatric Autorefractor plusoptiX A09 offers electronic aswell as paper documentation options.
If you use an EMR-System you can import patient data andexport readings automatically.
If a label printer is connected, you can print self adhesivelabels for your paper records.
If a printer is connected, you can print screenshots and anoptional letter size measurement report for parents shownbelow.
Without NotebookplusoptiX A09 Emily
App l i cat ionPediatric Autorefractor plusoptiX A09 is a binocular hand-heldautorefractometer appropriate for infants and children. It provi-des reliable measurements of refraction, pupil size, interpupilla-ry distance and eye alignment. It can also be used to assessthe accuracy of an eyeglass prescription.
Used in un-dilated pupils, measurements of hyperopia, myopia,astigmastism and anisometropia provide valuable data in a ped-iatric vision examination. Results can additionally be used as astarting point for retinoscopy or to confirm retinoscopic results.
Un ique featuresPediatric Autorefractor plusoptiX A09 provides three uniquefeatures:
Á measurement distance is 1 meter, i.e. 3.3 feet
Á both eyes are measured simultaneously
Á acquisition time is 0.8 seconds
These features make it a very effective tool for gathering visi-on data on non-verbal, uncooperative patients and childrenwith nystagmus.
[Measurement is fully automated and can be performed by paramedical staff easily.
[ [INSERTyour commentsfor the parents
HERE(free editable text modules)