2.0. introduction - inflibnetshodhganga.inflibnet.ac.in/bitstream/10603/36019/10/10_chapter...
TRANSCRIPT
2.0. Introduction
This chapter presents the sampling procedure, preservation techniques and
procedure for analysis of waste water samples. Purification of solvents chemicals
and reagents used are given. Methods of treatment and an introductory treatment
on correlation analysis are also presented in this chapter.
2.1 Waste water samples
2.1.1 Sampling procedure
The industrial waste water samples for this study were collected from four
different sampling sites(industrial discharge points) A, B, C, and D, viz., A -Match
industry, B - Fire works unit, C - Printing industry and D - Plate washing unit of a
printing industry. From each and every sampling site, samples were collected
bimonthly over a period of one year. On the day of sampling, the samples were
collected once in 4 hours (for 24 hours) and mixed in equal proportion to get
uniform, homogeneous average samples. The average analytical data of the
replicate samples from each site have been grouped into the sample groups A, B, C
and D. Sampling units and sampling sites were selected by random selection
procedure. The samples were collected bimonthly over a period of 12 months
during the period, March 1999 to April 2000.
The waste water samples were collected from the discharge points (outlet),
where the industrial effluents were discharged by the four different industries. A
polythene can(2 litre capacity) was used to collect the effluent samples from the
sampling point for analysis. For the present study, spot sampling technique was
adopted[237], A series of samples collected in this technique over a period of one
year will reflect the seasonal variation of WOPs over that period of time[237].
2.1.2 Sample preservation
The polythene bottles (of 100/500 ml capacity) used for sample preservation
were thoroughly cleaned by rinsing with 8M HN03 solution, followed by repeated
washing with distilled water and finally with double distilled (DD) water. The bottles
were also rinsed thrice with the waste water sample before collection. Water quality
parameters (WQPs) such as temperature, pH and electrical conductivity were
determined in the field itself (within 30 min. or quickly after sampling).
The other WQPs except BOD were determined within 72 h, from the time of
collection of sample. During the period of analysis the water samples were
preserved as per the preservation technique recommended by APHA[238]. This is
essential for retarding biological action, hydrolysis of chemical compounds and
complexes and reduction of volatility of constituents. The details of preservation
techniques employed in the present study are summarised in table 2.1.
The waste water samples collected were always kept in a suitable
polythene/glass container in a refrigerator (at temperature: 15 - 20°C), after adding
the necessary preservatives. The waste water samples were taken out from the
refrigerator only at the time of analysis.
2.2 Purification of solvents
2.2.1 Acetone
Acetone (E. Merck, India) was refluxed for two hours with potassium
permanganate and distilled. After distillation it was dried over anhydrous potassium
carbonate, filtered and fractionated through a Vigruex fractionating column (b. pt :
56 - 57°C). This method is essentially similar to that described by Sachs[239].
2.2.2 Glycerol
Pure glycerol was prepared by the usual standard method of purification.
Commercial glycerol(E. Merck, India) was dehydrated using anhydrous potassium
carbonate[240] and distilled under reduced pressure (b. pt. :80 - 81°C/10 mm).
2.2.3 Chloroform
Chloroform (E. Merck, India) was dried over anhydrous potassium
carbonate, filtered and fractionated through a Vigruex fractionating column (b. pt. 61
-62°C).
2.3 Chemicals and reagents
All the chemicals and reagents used in the present investigation were of
either analytical grade or laboratory reagent grade, procured from E. Merck/ Glaxo/
s.d. fine chemicals/Ranbaxy/Nice, India and they were used as such, without any
41
further purification. Doble distilled (DD) water was used during the entire course of
this work to prepare all the solutions and reagents.
2.3.1 Double distilled water
Deionised water was first distilled in an all glass apparatus. Distilled water
was once again distilled over alkaline potassium permanganate[241] in an all glass
apparatus, which was protected from carbon dioxide using soda lime guard tube. All
the reagents/solutions for analysis of WQPs were prepared using this double
distilled water (DD water) and stored in brown bottles.
2.4 Calibration
The burettes, graduated pipettes and standard measuring flasks used in the
present investigation were calibrated at room temperature (30 ± 0.1 °C). Sampling
pipettes were calibrated at room temperature (30 ± 0.1°C)and also at the sampling
temperatures (Temperature range: 25-35°C, error ± 0.1 °C). The calibrations were
carried out by the method recommended by Vogel[242], Pure carbon dioxide free
DD water was used for all the calibrations.
2.5 Measurement of water quality parameters
The industrial waste water samples were collected, preserved and the
following WQPs were measured within a period of 72 h, except BOD. The samples
for analysis were collected and preserved as per the standard methods
recommended by APHA[238 - 243]. The methods of chemical analysis[244 - 247]
and the measurement procedure for all the WQPs were similar to the standards
recommended[248] by Bureau of Indian Standards (BIS). Samples were filtered
through Whatman No. 42 filter paper in all the cases, before the chemical analysis
42
and data measurement, except for the determination of total suspended solids, pH
and temperature. Further the waste water samples were acidified with few ml of
con. HN03, digested and filtered through Whatman No. 42 filter paper for the
analysis of metal ions.
WQPs such as temperature(T), pH(pH), electrical conductivity(EC), total
suspended solids(TSS), total dissolved solids(TDS), alkalinity(ALK), total hardness
(THA), temporary hardness(HAT), permanent hardness(HAP), chloride(CL),
sulphate (SUL), phosphate(P04), sodium(NA), potassium(K), calcium(CA),
magnesium(MG), iron(FE), chromium(CR), biochemical oxygen demand(BOD) and
chemical oxygen demand(COD) were determined[238 - 248] and water quality
index (WQI) was calculated[9].
The notations provided in the parenthesis are the ones used in the
software (CORREL) for correlation analysis.
2.5.1 Temperature (T)
The temperature of the waste water samples were noted at the time of
sampling using a precision thermometer (accuracy = 0.1 °C). Standard error= ±
0.1 °C; Unit: °C.
2.5.2 pH (PH)
The pH values of the effluent samples were also measured at the time of
sampling under field conditions. The pH values of the samples were measured
using a pen type digital pH meter (Hanna Instruments, Portugal) with accuracy of
0.1. The instrument was set ready by using the standard buffer solutions of pH 4.0
and 9.2. The pH meter was washed thoroughly with distilled water and then, with
DD water. Finally it was rinsed with the waste water sample for which pH value was
43
to be measured and then it was carefully wiped with a tissue paper. The pH meter
was then dipped into the waste water sample taken in a 100 ml beaker and after
some time (1-2 min.) the pH reading was noted from the display[249]. Standard
error = ± 0.1.
2.5.3 Electrical conductivity (EC)
The instrument used for the measurement of electrical conductivity (EC)
was an Elico conductivity bridge (Model CM 185, India). The conductivity cell was
platinised by the usual standard method, washed well with distilled water and then
with DD water. The cell was finally rinsed with the waste water sample, for which
EC value was to be determined. Then the conductivity cell was dipped into the
waste water sample, taken in a 100 ml beaker and the conductance was noted in
micro(u) mhos. The electrical conductivity of the waste water sample was
calculated by using the formula[250]:
Electrical conductivity = Specific conductance x 1.03 ... (2.1)
The cell constant of the conductivity cell (= 1.03/cm) employed for the
determination of electrical conductivity (EC). Unit = urn ho/cm; Error = ± 0.5
f.im ho/cm.
2.5.4 Total suspended solids (TSS)
Gooch crucible was fixed with vacuum assembly and washed with three
successive 20 ml portions of DD water and it was removed from the filter assembly
and dried in an air oven (Toshniwal, India) at 103 -105°C for 1 h. After drying it was
cooled in a desiccator and weighed. The cycle of drying, cooling, desiccating and
weighing was repeated until a constant weight(B in mg) was obtained. Gooch
crucible was stored in a desiccator until needed.
44
difference between the values of total hardness and permanent hardness. These
values were also found to be in good agreement with the estimated values[255].
The residue obtained after filtration of the boiled water, in the experiment carried
out for the determination of the values of permanent hardness, was dissolved in
known volume of water and employed in complexometric titration with EDTA
solution. The procedure followed in the titration was similar to that employed for the
estimation of total hardness of waste water. Unit = mg/l of calcium carbonate.
Standard error = ± 0.1 mg/l.
2.5.8 Permanent hardness (HAP)
The permanent hardness or the carbonate hardness (in terms of mg/l of
calcium carbonate) of waste water was determined by the complexometric method
employing EDTA solution for titration. Exactly 250 ml of industrial effluent sample
was taken in a 500 ml beaker and it was boiled, till its volume reduced to about 50
ml. Then, it was cooled and filtered through Whatman No. 42 filter paper. The
precipitate was washed well with DD water and the washings were also added to
the filtrate. The solution was made up to 250 ml in a standard measuring flask.
Exactly 50 ml of the sample was titrated against N/100 EDTA solution, as
described in the experiment for the determination of the values of total hardness.
From the titre value, the permanent hardness was calculated in terms of mg/l of
calcium carbonate[256J by using the formula:
A standard calibration curve was drawn by plotting turbidity vs concentration of
sulphate ions (Fig. 2.1).
The industrial effluent (containing sulphate ion) was suitably diluted and the
turbidity measurement was made. The amount of sulphate ions present in the
waste water samples were determined for the diluted effluent samples, by the
interpolation technique using the calibration curve[260]. Finally the amount of
sulphate ions present in the effluent was calculated. Unit = mg/l; Standard error
= ±0.1 mg/l.
2.5.11 Sodium (NA)
The estimation of sodium and potassium was carried out flame
photometrical!y[261, 262] by employing flame photometric technique using a
Systronics Mediflame (Model No. 127, India).
Exactly 20 ml of waste water sample was taken in a silica crucible and
evaporated to dryness in a steam bath. It was then heated in a muffle furnace
(AUSCO, India) at 550 - 600°C. The ash obtained was dissolved in a minimum
quantity! 1 ml] of con. HN03 and 10-15 ml of warm DD water. This was then filtered
(Whatman No. 41) and diluted to known volume (preferably 100 ml) so that the final
nitric acid concentration was about 1 % (v/v). This sample was used for estimation
of sodium ions.
Standard solutions containing 0.1 mg/l to 40 mg/l of sodium ions were
prepared by diluting the standard stock solution of sodium chloride (100 mg/l). By
using these solutions, the flame photometer was calibrated. The flame photometer
scale reading for 40 mg/l solution was adjusted to be 100 and the scale reading for
the blank (pure DD water) was set to be zero. A standard calibration curve was
drawn by plotting fiarrie photometer reading (showing the concentration of sodium
50
ions) vs concentration of standard sodium ions(Fig. 2.2). The acid digested and
suitably diluted industrial effluent sample was introduced into the burner unit of the
instrument and the reading was noted. From the calibration curve, the amount of
sodium ions present in the diluted sample was determined by interpolation method.
Finally the amount of Na+ ions present in the effluent sample was calculated.
Unit = mg/l; Standard error = ± 0.1 mg/l.
2.5.12 Potassium (K)
The standard series of solutions containing 0.1 mg/l to 4 mg/l of potassium
ions were prepared from a stock solution of potassium chloride (100 mg/l) and the
flame photometer (Systronics Mediflame, model 127, India) was calibrated
(Settings: 4 mg/l solution to 100 unit and DD water to zero unit in the photometer
scale readings). A standard calibration curve was drawn by plotting photometer
scale reading vs concentration of potassium ions. The industrial effluent sample
(acid digested and suitably diluted ) was introduced into the burner unit of the flame
photometer. The flame photometer scale reading was noted. From the calibration
curve, the amount of potassium ions present in the waste water sample was
determined by employing the interpolation method[263]. The amount of K+ ions
present in the effluent sample was calculated Unit = mg/l; Standard error = ±
0.1 mg/l.
2.5.13 Calcium (CA)
Exactly 100 ml of industrial effluent (waste water sample) was filtered
through Whatman No. 42 filter paper and to the clear filtrate, sufficient amount of
20% (w/v) potassium hydroxide solution was added to adjust the pH of the solution
to pH 12. About 0.5 g of Patton and Reeder's indicator (2-hydroxy-1~(2-hydroxy-4-
51
B = mg of CaC03 equivalent to 1.00 ml of EDTA titrant
Unit: mg/l; Standard error = ± 0.1 mg/l.
2.5.15 Phosphate (P04)
Exactly 100 ml of industrial effluent was taken in a 250 ml beaker. To this
waste water sample, 1 ml con. sulphuric acid and 5 ml con. nitric acid were added.
The resulting solution was digested. This method of acid digestion was repeated
for 3 - 4 times, the residue was then dissolved with minimum amount of acid
mixture (20 ml) and transferred into a 100 ml standard measuring flask. Ammonium
molybdate solution was prepared as follows: Ammonium molybdate (25 g) was
dissolved in 75 ml of distilled water. About 280 ml. of con. sulphuric acid was added
to 400 ml. of DD water. Ammonium molybdate solution was added to diluted
sulphuric acid solution and made upto 1 litre. Stannous chloride solution was
prepared by dissolving freshly prepared stannous chloride (2.5 g) in 100 ml of
glycerol by heating in a water bath. Ammonium molybdate reagent (4 ml) and
stannous chloride solution in glycerol (0.5 ml) were added to the acid digested
water sample taken in standard measuring flask and the solution was made upto
the mark with DD water. After allowing the solution to stand for 10 min., the
absorbance (O.D.) of the reaction product[266] was measured at the wavelength of
690 nm by using Spectronic 20 D+ (Milton Roy, U.S.A), spectrophotometer. The
amount of phosphate present in the waste water sample was determined from the
standard calibration curve by employing the interpolation method. The standard
calibration curve was drawn by plotting the values of absorbance vs concentration
of the standard phosphate solutions (Fig 2.3). Unit = mg/l; Standard error =
±0.1 mg/l.
53
2.5.16 Iron(FE)
The total iron present in the waste water sample was estimated as Fe2+
ions. Exactly 50 ml of the industrial effluent was taken in a 250 ml beaker and the
sample was evaporated to dryness. After cooling the residue, 5 ml of dilute HCI
solution (1:1 v/v) was added. The resulting solution was digested. The method of
acid digestion was repeated by 3 - 4 times by adding 20 ml of DD water every time.
Then 2 ml of con. HCI and 1 ml of hydroxylamine hydrochloride reagent (10% w/v)
solution, were added and the mixture was heated to boiling. After cooling, the
mixture was transferred into a 100 ml standard measuring flask. Then, 10 ml of
ammonium acetate buffer solution (250 g of ammonium acetate was dissolved in
the 150 ml distilled water. To this, 700 ml of glacial acetic acid was added) and 4 ml
of 1, 10-phenanthroline solution (100 mg of 1,10-phenanthroline dissolved in 100
ml of water by adding 2 drops of con. HCI) were added and made upto the mark
with DD water. After allowing the solution to stand for 10 min. the absorbance
(O.D.) of the resulting reaction mixture[267] was measured at the wavelength of
501 nm using Spectronic 20 D+ (Milton Roy, USA) spectrophotometer.
The amount of total iron present in the waste water sample was determined
from the standard calibration curve by interpolation method. A standard calibration
curve (Fig. 2.4) was drawn by plotting the absorbance vs concentration of standard
iron solutions, prepared from ferrous ammonium sulphate (100 mg/l stock solution).
Unit: mg/l; Standard error = ± 0.1 mg/l.
2.5.17 Chromium (CR)
The total chromium present in the waste water sample was estimated as
Cr6+ ion. Exactly 50 ml of the industrial effluent was taken in a 250 ml beaker. To
this waste water sample, 5 ml of con. sulphuric acid and 2 ml of con. nitric acid
55
were added. The resulting acidic solution was digested. The method of acid
digestion was repeated by 3 - 4 times by adding 10 ml of DD water, each time.
Then, the residue was dissolved in 50 ml of DD water, cooled and transferred into a
separating funnel (250 ml). To this, 5 ml of ice cold cupferron solution (5 % w/v)
was added and shaken well. Then 5 ml of chloroform was added to the separating
funnel and shaken well. The chloroform layer was separated and discarded. This
process of extraction with chloroform (by adding 5 ml each time) was repeated by 3
- 4 times. Then the aqueous layer was transferred to a clean dry 250 ml conical
flask. The separating funnel was washed with DD water and the washings were
also collected in the conical flask.
The solution was boiled for 5 min. (to remove chloronvif any) and cooled.
Nitric acid (5 ml) and sulphuric acid (3 ml) were added and boiled. After cooling, a
few drops of (1:1) liquor ammonia was added, so that the solution became just
basic to methyl orange. Then (1:1 v/v) sulphuric acid solution was added dropwise
until it became acidic and two drops were added in excess. Potassium
permanganate (4% w/v) solution was added to this solution dropwise until the
solution became light pink colour and boiled. Then sodium azide solution (0.5 %
w/v) was added dropwise and boiled until the solution became colourless. Finally
0.25 ml of ortho phosphoric acid was added to this.
The pH of the solution was adjusted to pH 1 by adding 4N solution of
sulphuric acid and the resulting solution was transferred into a 100 ml standard
measuring flask. To this, 2 ml of diphenyl carbazide solution (250 mg of 1,5-
diphenyl carbazide was dissolved in 50 ml acetone) was added and the solution
was made up to the mark with DD water. After allowing the solution to stand for 10
min, the absorbance (O.D.) of the resulting coloured solution of the compiex[266]
56
dipotasium hydrogen phosphate, 33.4 g disodium hydrogen phosphate and 1.7 g of
ammonium chloride dissolved in 1 litre of DD water) per litre of aerated DD
water (i.e., D.D water saturated with air using aerator).
2.1.19 Dissolved Oxygen (DO)
To the effluent sample(25 ml) diluted in a BOD bottle(300 ml), 1 ml of
Winkler's solution (364 g manganous sulphate monohydrate in 1 litre) was added,
followed by 1 ml of alkali - azide solution (700 g of potassium hydroxide and 150 g
of potassium iodide were dissolved in 1 litre. To this, 40 ml of solution containing 10
g of sodium azide was added). The solution was shaken well by inverting the bottle
several times. To this one ml of con. sulphuric acid was added and mixed well by
inverting the bottle few times, A volume corresponding to 200 ml original sample
after making the correction for sample loss by displacement with added reagents
{i.e., 201 ml) was titrated against 0.025 M sodium thiosulphate solution using starch
( 1 % w/v) indicator. The end point is the disappearance of blue colour. Dissolved
oxygen (DO) was calculated[266] by using the relationship:
1 ml of 0.025 M sodium thiosulphate solution = 1 mg of dissolved
oxygen per litre.
Unit = mg/l; Standard error = ± 0.1 mg/l
2.5.20 Chemical oxygen demand (COD)
Exactly 50 ml of industrial effluent sample was taken in a 500 ml round
bottomed (RB) flask. The following reagents were added under the ice cold
condition: a known volume (excess) of 0.25 N potassium dichromate, con. sulphuric
acid containing 1 g of silver sulphate (equal to the volume of sample and potassium
dichromate added) and 1 g of mercuric sulphate. The reaction mixture was refluxed
for 3 h in a heating mantle (Toshniwal, India). After cooling, the excess unreacted
58
2.7. Removal studies
2.7.1 Materials
In removal studies, commercial activated carbon (AC) and cationic resin -
CR (Amberlite IR 120 strong acid, Loba) were used in batch and column modes.
AC and cationic resin were used in column studies. AC was washed with water and
dried. AC was further activated by digesting it with 4N sulphuric acid solution at
80°C for 60 min. Then, the sample was thoroughly washed with distilled water and
finally with DD water, until the washings were free, from acid (tested with barium
chloride solution; till no precipitate was formed). Activated carbon after activation
was powdered mechanically, to obtain the desired particle size (90 micron) before
using it for batch and column experiments. CR was used after treating it with 4N
HCI solution. Alum was procured commercially and used as such.
2.7.2 Primary treatment
A known volume of effluent sample was neutralised by adding required
volume of dil. H2S04 (pH « 7) treated with 10, 20, 30, 40 and 50 mg/l of alum and
stirred in a jar test apparatus. After allowing the solution to stand for 10 - 15 min.,
62
Control experiments with AC or CR and DD water were carried out as
blanks. The experiments were replicated and the average value was reported (error
= ±1%).
2.7.4.2 Column studies
A known weight (20 g) of cationic resin (CR) sample was treated with 4 N
HCI solution, washed well and dried. A column was prepared using a graduated
burette glass wool plug and the acid treated CR sample, which was taken in the
form of slurry and DD water.
Waste water(influent) sample (1 litre) obtained after secondary treatment
was added slowly to the column in several portions. The flow of effluent was
adjusted to 1 ml/ min. The CR sample exchanged its H+ ion with the metal ions
present in the influent. The effluent(out-coming) sample was collected and analysed
for various WQPs such as hardness, metal ions, EC and TDS. The percentage
removal was determined. The column was regenerated by washing it with HCI (40
ml 4.0 N) and subsequently with DD water till the washings showed a negative
indication for chloride ions to AgN03test. The experiments were repeated with other
waste water samples also using AC instead of CR.
The column capacity was determined with a standard solution of (40 ml of
0.05 N) sodium sulphate instead of effluent. The quantity of H+ ions liberated to the
effluent was estimated by titrating it with NaOH solution(0.05 M), using phenol-
phthalein as indicator. From the titre value, the column capacity or cation exchange
capacity of the resin sample, with respect to exchange with Na+ ions, was
calculated as detailed below:
64