PHYSICAL PRINCIPLES OF COMPUTED TOMOGRAPHY

Page 2: 7 Physical principles of CT

RADIOGRAPHY LIMITATIONS

SUPERIMPOSITION DIFFICULTY IN DISTINGUISHING

BETWEEN HOMOGENOUS OBJECTS OF NON-UNIFORM THICKNESS.

Page 3: 7 Physical principles of CT

SUPERIMPOSITION

Page 4: 7 Physical principles of CT

TISUE DIFFERENCE SENSITIVITY 5%-10%

Page 5: 7 Physical principles of CT

TOMOGRAPHY ( CONVENTIONAL)

ELIMINATES TISSUE SUPERIMPOSITION INCREASES CONTRAST OF LOW

SUBJECT CONTRAST TISSUES

Page 6: 7 Physical principles of CT
Page 7: 7 Physical principles of CT

TOMOGRAPHY

Page 8: 7 Physical principles of CT

TOMOGRAPHY

Page 9: 7 Physical principles of CT

TOMOGRAPHY

Page 10: 7 Physical principles of CT
Page 11: 7 Physical principles of CT

CT ADVANTAGES

Page 12: 7 Physical principles of CT

LIMITATIONS OF CT

UNABLE TO DIFFERENTIATE BETWEEN TISSUES WITH SLIGHT CONTRAST DIFFERENCES < 1%.

Page 13: 7 Physical principles of CT
Page 14: 7 Physical principles of CT

GOALS OF CT

MINIMAL SUPERIMPOSITION IMAGE CONTRAST IMPROVEMENT SMALL TISSUE DIFFERENCE

RECORDING

Page 15: 7 Physical principles of CT

CT DATA AQUISITION

Page 16: 7 Physical principles of CT
Page 17: 7 Physical principles of CT

SLIP RINGS

Page 18: 7 Physical principles of CT
Page 19: 7 Physical principles of CT

SEQUENTIAL-SLICE BY SLICE SCANNING

Page 20: 7 Physical principles of CT

SCANNING

Page 21: 7 Physical principles of CT

TRANSMISSION

RELATIVE TRANSMISSION=Io/I

Page 22: 7 Physical principles of CT
Page 23: 7 Physical principles of CT
Page 24: 7 Physical principles of CT

Total # of trans. measurements=

# of views X # of rays in each view

Page 25: 7 Physical principles of CT

ATTENUATION

Page 26: 7 Physical principles of CT
Page 27: 7 Physical principles of CT

DATA AQUSITION GEOMETRIES

CONTINUOUS STATIONARY

Page 28: 7 Physical principles of CT

CONTINUOUS

Page 29: 7 Physical principles of CT

STATIONARY

Page 30: 7 Physical principles of CT

CT 120-140 KVP

REDUCED DEPENDENCY ON ATTENUATION COEFFICIENT

REDUCED CONTRAST INCREASED PHOTON FLUX

Page 31: 7 Physical principles of CT

ORIGINAL CLINICAL CT SCANS COMPOSED OF

80 X 80 MATRIX

PIXELS

6400

Page 32: 7 Physical principles of CT

X

Page 33: 7 Physical principles of CT

Y

Page 34: 7 Physical principles of CT

Z

Page 35: 7 Physical principles of CT

ISOCENTER

Page 36: 7 Physical principles of CT

SCAN FOV

SFOV

DETECTORS

Page 37: 7 Physical principles of CT

SCAN FOV-SMALL

SFOV

DETECTORS

Page 38: 7 Physical principles of CT

SFOV - HEAD

Page 39: 7 Physical principles of CT
Page 40: 7 Physical principles of CT

TOO SMALL OF SFOV – OUT OF FIELD ARTIFACT

Page 41: 7 Physical principles of CT

SCAN FOV-RESOLUTION

SFOV

Page 42: 7 Physical principles of CT

RECONSTRUCTION

Ц CT#RECONSTRUCTION

Page 43: 7 Physical principles of CT

SCAN FOV-RESOLUTION

SFOV

Page 44: 7 Physical principles of CT

DISPLAY FOV vs SCANNING FOV

DFOV CAN BE EQUAL OR LESS OF SFOV SFOV – AREA OF MEASUREMENT

DURING SCAN DFOV - DISPLAYED IMAGE

Page 45: 7 Physical principles of CT

PIXEL SIZE

PIXEL SIZE= FOV (mm)/ MATRIX SIZE

Page 46: 7 Physical principles of CT

MOST SCANNERS PIXEL SIZE

1 TO 10mm

Page 47: 7 Physical principles of CT

EXAMPLE:

FOV= 40 CM= 40 X 10 MM=400 mm MATRIX= 512 X 512 = 5122

400/512 = 0.78 mm0.8 mm

Page 48: 7 Physical principles of CT

EACH PIXEL IN CT HAS RANGE OF GRAY SHADES

2 8 = 256 SHADES 2 12 = 4096 SHADES = -100 TO 3095

SHADES OF GRAY

Page 49: 7 Physical principles of CT

PIXEL vs VOXEL

PIXELVOXEL

Page 50: 7 Physical principles of CT

PIXEL SIZE DEPENDS ON:

MATRIX SIZE FOV

Page 51: 7 Physical principles of CT

VOXEL SIZE DEPENDS

FOV MATRIX SIZE SLICE THICKNESS

Page 52: 7 Physical principles of CT

IMAGE DISPLAY

Page 53: 7 Physical principles of CT

GRAY SCALE DISPLAY MONITOR RESOLUTION IS RELATED TO THE SIZE OF THE PIXEL MATRIX

64 X 64 128 X 128 256 X 256 512 X 512 1024 X 1024 2048 X 2048 (HIGH PERFORMANCE MONITORS)

Page 54: 7 Physical principles of CT

MATRIX

Page 55: 7 Physical principles of CT

PIXEL MATRIX

Page 56: 7 Physical principles of CT

IN CT DIGITAL RECONSTRUCTED IMAGE IS CONVERTED IMAGE IS CONVERTED INTO A GRAY SCALE IMAGE.

Page 57: 7 Physical principles of CT

DAC

ANALOG

DIGITAL

Page 58: 7 Physical principles of CT

The high performance video display of the microcomputer is connected to the system via an interface board. The video display is a form of cathode ray tube sometimes referred to as a raster display. The term raster describes the technique of producing the picture or text which is formed by a beam of electrons that repeatedly scans across the screen to form a uniform pattern of closely spaced, horizontal lines (the raster), covering the entire screen. The screen consists of a phosphor that converts the energy of the electron beam into visible light. A picture is formed by "turning on and off" the electron beam at appropriate points in the scanning of the screen surface.

Page 59: 7 Physical principles of CT

CRT OPERATION

Page 60: 7 Physical principles of CT

IMAGE FORMATION ON THE MONITOR

Page 61: 7 Physical principles of CT

IMAGE FORMATION ON THE MONITOR

Page 62: 7 Physical principles of CT
Page 63: 7 Physical principles of CT

CT NUMBER

Page 64: 7 Physical principles of CT
Page 65: 7 Physical principles of CT

LINEAR ATTENUATION COEFFICIENT ( cm-1)

BONE 0.528 BLOOD 0.208 G. MATTER 0.212 W. MATTER 0.213 CSF 0.207 WATER 0.206 FAT 0.185 AIR 0.0004

Page 66: 7 Physical principles of CT

CT # vs BRIGHTNESS LEVEL

+ 1000

-1000

Page 67: 7 Physical principles of CT

CT # 1000

Page 68: 7 Physical principles of CT

CT # - 500

Page 69: 7 Physical principles of CT

CT # OF CYST5

Page 70: 7 Physical principles of CT

CT # OF LIPOMA ( FATTY TUMOR)

                                                                        

                                                

-100

Page 71: 7 Physical principles of CT
Page 72: 7 Physical principles of CT
Page 73: 7 Physical principles of CT
Page 74: 7 Physical principles of CT

W 120L 40

Page 75: 7 Physical principles of CT

W 80L 40

Page 76: 7 Physical principles of CT

DATA FLOW IN CTREFERENCE DETECTOR

REFERENCE DETECTOR

ADCPREPROCESSOR

COMPUTER

RAW DATA

CONVOLVED DATABACK

PROJECTORRECONSTRUCTED DATA

PROCESSORS

DISK TAPE DAC CRT DISPLAY

Page 77: 7 Physical principles of CT

CT ADVANTAGES AND DISADVANTAGES

EXCELLENT LOW CONTRAST RESOLUTION WINDOWING- IMAGE MANIPULATION TAILORED TO

OBSERVER NEEDS SPIRAL CT-SINGLE BREATH HOLD STUDIES ( CTA, MPR, VIRTUAL REALITY CT, CT ENDOSCOPY) CT ASISST IN RADIATION THERAPY BONE SCAN PACKAGE XENON CT PERFUCION CT DIGITAL PROCESSING ABILITY