Page 1: Agilent Bio-Monolith Protein A and Protein G Affinity ... · Bio-Monolith Protein G, 4.95 x 5.2 mm 5190-6900 Sophisticated Agilent Solutions for Every Separation Challenge Infinitely

Agilent Bio-Monolith Protein A and Protein G Affinity Columns

WIDER SELECTIVITY FOR mAb QUANTITATION AND ANALYSIS

Page 2: Agilent Bio-Monolith Protein A and Protein G Affinity ... · Bio-Monolith Protein G, 4.95 x 5.2 mm 5190-6900 Sophisticated Agilent Solutions for Every Separation Challenge Infinitely

2

AGILENT BIO-MONOLITH PROTEIN A AND PROTEIN G AFFINITY COLUMNS

• Fast separations decrease method development time and costs and provide a larger window of opportunity for harvest.

• High binding capacity allows titer determination across relevant concentration ranges and enables purification for further characterization.

Wide concentration range

Figure 1. Linearity of IgG1 quantitation achieved using the Agilent Bio-Monolith Protein A column over the typical mAb titer range encountered with CHO cells.

DETERMINE MAB TITERS QUICKLY AND ACCURATELY

The ability to measure monoclonal antibody (mAb) titers and to purify small amounts of product candidate for further characterization is essential for the development of an economic production process.

Agilent Bio-Monolith Protein A and Protein G columns deliver consistently excellent performance under the conditions encountered during method development, optimization, and production. These columns are ideal for titer determination during process development, cell clone selection, and optimization of the fermentation conditions, and in production to determine the optimum time to harvest – helping to ensure correct cell line choice and identification of harvest times to increase process yield and economics.

• Now with the new Protein G Monolith column there is a wider range of selectivity for IgG subclasses.

• Long column life and minimal clogging, even with complex samples, means less frequent column cleaning and replacement.

An unusual shape? Perhaps. But Agilent Bio-Monolith Protein A and Protein G columns deliver the high capacity and performance you need for reproducible affinity chromatography.

R2 = 0.9999

0

1000

2000

3000

4000

5000

6000

7000

8000

9000

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2

Area

min0.5 1 1.5 2 2.5 3 3.5 4

mAU

0

500

1000

1500

2000

2500

2 mg/mL1mg/mL

0.5 mg/mL0.2 mg/mL

0.1 mg/mL0.05 mg/mL

0.02 mg/mL

min1.2 1.4 1.6 1.8 2 2.2 2.4

mAU

0

50

100

150

200

250

300

0.02 mg/mL0.05 mg/mL

0.1 mg/mL

0.2 mg/mL

0.5 mg/mL

Concentration (mg/mL)

Page 3: Agilent Bio-Monolith Protein A and Protein G Affinity ... · Bio-Monolith Protein G, 4.95 x 5.2 mm 5190-6900 Sophisticated Agilent Solutions for Every Separation Challenge Infinitely

3

AGILENT BIO-MONOLITH PROTEIN A AND PROTEIN G AFFINITY COLUMNS

Proven specificity

Figure 2. Rapid and specific capture of IgG1 from cell culture supernatant using the Bio-Monolith Protein A column. (A) Escherichia coli cell lysate supernatant (2 ug of IgG1 and 38 ug of cell lysate proteins of ug total) was loaded onto the column.

Long column life3

lysate proteins

min 0 0.25 0.5 0.75 1 1.25 1.5 1.75 2 2.25

mAU

0

50

100

150

200

250

0.11

3 min 0 0.25 0.5 0.75 1 1.25 1.5 1.75 2 2.25

mAU

0 200

600

1000

1400

lysate proteins contains IgG1

wash elute Re-equilibrate

flow through (8 µm)

Abso

rban

ce (U

V 28

0 m

m)

Retention time (min)

min 0.25 0.5 0.75 1 1.25 1.5 1.75 2 2.25

mAU

0 20 40 60 80

A

B

baseline expansion

IgG1

Reliable accuracy with large capacity

Key:Lane 1: Whole serum prior to separationLane 2: IgG standardLane 3: Peak 1 (flow-through fraction)Lane 4: Peak 2 (protein A-bound fraction; i.e. IgG1 and IgG2)

0

200

400

600

800

1000

1200

0 0.5 1 1.5 2 2.5 30

20

40

60

80

100

120

Relat

ive a

bsor

banc

e [mA

U]

% G

radie

nt

Time [min]

Figure 4. Rapid human polyclonal IgG quantitation from human plasma using the Bio-Monolith Protein A column. IgG was bound and a 100% buffer B step gradient was applied; IgG eluted at 1.5 minutes.

100

300

500

700

900

1100

1300

1500

1700

1900

0 100 200 300 400 500 600 700 800 900 1000

Peak

are

a

Injection numberFigure 3. Lifetime of the Agilent Bio-Monolith Protein G column is demonstrated over 1000 cycles of IgG3 binding and elution. Reproducibility of peak area enables accurate quantitation of the IgG3 titer.

Peak 1

Peak 2

Page 4: Agilent Bio-Monolith Protein A and Protein G Affinity ... · Bio-Monolith Protein G, 4.95 x 5.2 mm 5190-6900 Sophisticated Agilent Solutions for Every Separation Challenge Infinitely

Learn more http://www.agilent.com/chem/ AdvanceBio

Order your Agilent Bio-Monolith Protein A and Protein G columns now at www.agilent.com/chem/store

To find your local Agilent Representative or Agilent Authorized Distributor, www.agilent.com/chem/contactus

USA and Canada 1-800-227-9770 [email protected]

Europe [email protected]

Asia Pacific [email protected]

India [email protected]

For Research Use Only. Not for use in diagnostic procedures. This information is subject to change without notice.

© Agilent Technologies, Inc., 2015Published in the USA, May 23, 2015 5991-5961EN

Ordering informationAgilent Bio-Monolith Protein A and Protein G ColumnsDescription Part NumberBio-Monolith Protein A, 4.95 x 5.2 mm 5069-3639Bio-Monolith Protein G, 4.95 x 5.2 mm 5190-6900

Sophisticated Agilent Solutions for Every Separation ChallengeInfinitely better technology – The Agilent 1200 Infinity SeriesWhether you need a workhorse LC system for routine analysis or a sophisticated, high-resolution LC/MS system, the Agilent 1200 Infinity Series delivers ultimate resolution and sensitivity. Boost your separation power per time, and ensure easy method transferability between systems – without redevelopment or revalidation. Learn more at agilent.com/chem/infinity.

Faster, more consistent biomolecule characterization – Agilent AdvanceBio columnsAgilent AdvanceBio columns deliver consistent, exceptional performance for separating and characterizing peptides and proteins. These state-of-the-art columns increase accuracy and productivity and eliminate interference that can impede analysis; they are rigorously tested and backed by Agilent’s 60-day full satisfaction warranty.

Learn more at agilent.com/chem/AdvanceBio.