application areas: 7039test method astm d7039 and iso 20884 dimensions 37 cm (w) x 50 cm (d) x 34 cm...
TRANSCRIPT
better analysis counts
Application Areas: • Totalsulfuranalysisfromultralowsulfur
fuelstocrudes.
• Foruseinrefinerylabs,pipelineterminals,additiveplants,testingvansandinspectionlaboratories.
• ComplieswithASTMD7039andISO20884.
Features and Benefits:• LOD:0.4ppmat300s.
• DynamicRange: Standard: 0.4ppmto3000ppm XRpackage: 0.4ppmto10%
• Fitsonanybenchandcompatibleforuseinmobilelabsandvans:
37cm(w)x50cm(d)x34cm(h)
• Plug-it-inandmeasure:noadditionalutilitiesrequired.
• TouchScreenuserinterface.
• Userprogrammablemeasurementtime:30-900s.
• Onecalibrationcurvewillrunbothdieselandgasolineupto3000ppm
• Noconversiongasses,heatingelements,quartztubesorcolumns.
• 75Wair-cooledexcitationtube.
• Robustpolyamidewindowforeasycleaning.
Options: • ExtendedRangepackage:from0.4ppmup
to10%.
• LIMSdataoutputsoftwarecapability.
7039
Sulfur Analysis in Petroleum and Bio Fuels
Fromultralowsulfurdieselandgasolinetoheavyfueloilandcrudes,theSindie
7039analyzerdeliversunprecedentedprecisionandaccuracy.Theanalyzeristhe
ultimateeasy-to-useandrobustsulfuranalyzerforpetroleumandbiofuels.The
idealanalyticalsolutionfortheenergyindustrywheredetection,performanceand
reliabilityarecritical.SINDIE7039complieswithASTMD7039andISO20884
methods.Plugitinandmeasure.Resultswithonetouch.Unrivaledprecision.
PrecisionTypicalrepeatability(r)andreproducibility(R)values
indieselfuel,at95%confidence.300smeasurement
time.
Sulfur Concentration r R (ppm)
2 0.4 1.0
8 0.7 1.2
15 0.9 1.7
100 3 6
500 6 12
better analysis counts15 Tech Valley Drive • East Greenbush, New York 12061, USA • 518.880.1500 • Fax: 518.880.1510 e-mail: [email protected] • website: www.xos.com
© XOS (2012) MAR2012
Optic 1 Optic 2PolychromaticIncident X-rays
from tube
Sulfur
MWD XRF Monochromatic Wavelength Dispersive X-Ray Fluorescence (MWD XRF) utilizesstate-of-the-
artfocusingandmonochromatingopticstoincreaseexcitationintensityanddramaticallyimprove
signal-to-backgroundoverhighpowertraditionalWDXRFinstruments.Thisenablessignificantly
improveddetectionlimitsandprecisionandareducedsensitivitytomatrixeffects.Amonochromatic
andfocusedprimarybeamexcitesthesampleandsecondarycharacteristicfluorescencex-raysare
emittedfromthesample.Asecondmonochromatingopticselectsthesulfurcharacteristicx-rays
anddirectsthesex-raystothedetector.MWDXRFisadirectmeasurementtechniqueanddoesnot
requireconsumablegassesorsampleconversion.
SINDIE-XRCalibrationCurve
Low RangeCalibration
Test Method ASTM D7039 and ISO 20884
Dimensions 37 cm (w) x 50 cm (d) x 34 cm (h)
Power 100-120 VAC, 47-63 HZ at 6.0 Amps/200-240 VAC, 47-63 HZ at 6.0 Amps
Sample Cup Volume 10 ml
I/O Ports Ethernet 10/100 base T, RS232
Optional Computer Interface Pentium, 100MHz, 32MB RAM/Windows 98 or newer operating system
Ambient Temperature Requirements 5-40o C (40-104o C)
Dynamic Range Standard: 0.4 – 3000 ppm, XR Package: 0.4 ppm – 10%
Measurement User selectable: 30-900 s
Calibration 8 calibration curves. Automatic and Manual Calibration functionality.
Product Specifications
better analysis counts
Application Areas: • Totalsulfuranalysisfromultralowsulfur
fuelstocrudes.
• Foruseinrefinerylabs,pipelineterminals,additiveplants,testingvansandinspectionlaboratories.
• ComplieswithASTMD7039andISO20884.
Features and Benefits:• LOD:0.15ppmat300s.
• DynamicRange: Standard: 0.15ppmto3000ppm XRpackage: 0.15ppmto10%
• Availablewith8cellauto-sampler.
• Fitsonanybenchandcompatibleforuseinmobilelabs/vans.
• Plug-it-inandmeasure:powerisonlyutility.
• Touchscreenuserinterface.
• UtilizesACCU-CELLpre-assembledandpre-ventedsamplecupsforenhancedpreci-sion,extremeease-of-useandenhancedproductivity.
• Userprogrammablemeasurementtime:30-900s.
• Onecalibrationcurvewillrunbothdieselandgasolineupto3000ppm.
• Noconversiongasses,heatingelements,quartztubesorcolumns.
• 75Wair-cooledexcitationtube.
• Robustpolyamidewindowforeasycleaning.
Options: • ExtendedRangepackage:from0.15ppm upto10%.
• 8cellAutosampler.
• LIMSdataoutputsoftwarecapability.
7039G3
Sulfur Analysis with Unprecedented Precision
Nocompromisesindetection,performanceandreliability-theSINDIE7039G3
analyzeristheidealsulfuranalyticalsolutionfortheenergyindustry.Fromultra
lowsulfurdieselandgasolinetoheavyfueloilandcrudes,theanalyzerdelivers
unprecedentedprecisionandaccuracy.TheSindie7039G3bench-topanalyzeris
easy-to-use,robustandcomplieswithbothASTMD7039andISO20884methods.
Plugitinandmeasure.Resultswithonetouch.Unrivaledprecision.
PrecisionTypicalrepeatability(r)andreproducibility(R)valuesin
dieselfuel,at95%confidence.300smeasurementtime.
Sulfur Concentration r R (ppm)
2 0.3 0.7
5 0.5 0.8
8 0.6 1.0
15 0.8 1.4
100 2 4
500 5 10
better analysis counts15 Tech Valley Drive • East Greenbush, New York 12061, USA • 518.880.1500 • Fax: 518.880.1510 e-mail: [email protected] • website: www.xos.com
© XOS (2012) MAR2012
Optic 1 Optic 2PolychromaticIncident X-rays
from tube
Sulfur
MWD XRF Monochromatic Wavelength Dispersive X-Ray Fluorescence (MWD XRF) utilizesstate-of-the-artfocusingandmonochromatingopticstoincreaseexcitationintensityanddramaticallyimprovesignal-to-backgroundoverhighpowertraditionalWDXRFinstruments.Thisenablessignificantlyimproveddetectionlimitsandprecisionandareducedsensitivitytomatrixeffects.Amonochromaticandfocusedprimarybeamexcitesthesampleandsecondarycharacteristicfluorescencex-raysareemittedfromthesample.Asecondmonochromatingopticselectsthesulfurcharacteristicx-raysanddirectsthesex-raystothedetector.MWDXRFisadirectmeasurementtechniqueanddoesnotrequireconsumablegassesor
sampleconversion.
SINDIE Autosampler•8samplecellcapacity
•Increasesproductivity
•UtilitizesXOSAccu-Cellcups
ACCU-CELL Sample Cups•Noassemblyofseparatefilm&cupcomponents
•Pre-ventedsamplecups
•Eliminatessample&cupcontamination
•Onedischargeof1mlpipettewillfillthecup
SINDIE-XRCalibrationCurve
Low RangeCalibration
Test Method ASTM D7039 and ISO 20884
Dimensions 37 cm (w) x 50 cm (d) x 34 cm (h)
Power 100-120 VAC, 47-63 HZ at 6.0 Amps/200-240 VAC, 47-63 HZ at 6.0 Amps
Sample Cup Volume 1 ml
I/O Ports Ethernet 10/100 base T, RS232
Optional Computer Interface Pentium, 100 MHz, 32 MB RAM/Windows 98 or newer operating system
Ambient Temperature Requirements 5-40o C (40-104o C)
Dynamic Range Standard: 0.15 – 3000 ppm, XR Package: 0.15 ppm – 10%
Measurement User selectable: 30-900 s
Calibration 8 calibration curves. Automatic and Manual Calibration functionality
Product Specifications
better analysis counts
Application Areas: • Total sulfur analysis from ultra low sulfur fuels to crudes.
• For use in refinery labs, pipeline terminals, additive plants and inspection laboratories.
• Complies with ASTM D2622, D7039 and ISO 20884.
Features and Benefits:• LOD: 0.4 ppm at 300 s.
• Dynamic Range: 0.4 ppm to 10%
• Fits on any bench: 37 cm (w) x 50 cm (d) x 34 cm (h).
• Plug-it-in and measure: power is only utility.
• Touch Screen user interface.
• User programmable measurement time: 30-900 s.
• Two calibrations covers both gasoline and diesel matrices over full dynamic range:
- 0.4 – 3000 ppm wt - 0.3 – 10% wt
• No conversion gasses, heating elements, quartz tubes or columns.
• 75 W air-cooled excitation tube.
• Robust polyamide window for easy cleaning.
Options:• LIMS data output software capability.
Sulfur Analysis with Compliance Flexibility
The Sindie 2622 bench-top analyzer complies with ASTM D2622, D7039 and ISO
20884 methods, enabling complete flexibility in sulfur analysis. No compromises in
detection, performance and reliability - the SINDIE 2622 analyzer is the ideal sulfur
analytical solution from ultra low sulfur diesel and gasoline to heavy fuel oil and
crudes. One analyzer – three compliance solutions.
PrecisionTypical repeatability (r) and reproducibility (R) values
in diesel fuel, at 95% confidence. 300s measurement
time.
Sulfur Concentration r R (ppm)
4 0.4 1.0
8 0.7 1.2
15 0.9 1.7
100 3 6
500 6 12
better analysis counts15 Tech Valley Drive • East Greenbush, New York 12061, USA • 518.880.1500 • Fax: 518.880.1510 e-mail: [email protected] • website: www.xos.com
© XOS (2012) MAR2012
Optic 1 Optic 2PolychromaticIncident X-rays
from tube
Sulfur
MWD XRF Monochromatic Wavelength Dispersive X-Ray Fluorescence (MWD XRF) utilizes state-of-the-
art focusing and monochromating optics to increase excitation intensity and dramatically improve
signal-to-background over high power traditional WD XRF instruments. This enables significantly
improved detection limits and precision and a reduced sensitivity to matrix effects. A monochromatic
and focused primary beam excites the sample and secondary characteristic fluorescence x-rays are
emitted from the sample. A second monochromating optic selects the sulfur characteristic x-rays
and directs these x-rays to the detector. MWD XRF is a direct measurement technique and does not
require consumable gasses or sample conversion.
SINDIE-XRCalibrationCurve
Low RangeCalibration
Test Method ASTM D7039, D2622 and ISO 20884
Dimensions 37 cm (w) x 50 cm (d) x 34 cm (h)
Power 100-120 VAC, 47-63 HZ at 6.0 Amps/200-240 VAC, 47-63 HZ at 6.0 Amps
Sample Cup Volume 10 ml
I/O Ports Ethernet 10/100 base T, RS232
Ambient Temperature Requirements 5-40o C (40-104o F)
Dynamic Range 0.4 ppm – 10%
Measurement User selectable: 30-900 s
Calibration 8 calibration curves. Automatic and Manual Calibration functionality.
Product Specifications