EDGE®
Automated Extraction System
Method Note Compendium
Updated September 25, 2019
![Page 2: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/2.jpg)
Agriculture
Extraction of Pesticides From Avocados 2
Extraction of Polyphenols From Beans 3
Extraction of Polyphenols From Cacao 4
Extraction of Cannabinoids From Cannabis Plant 5
Extraction of Mycotoxins From Cannabis Plant 6
Extraction of Pesticides From Cannabis Plant 7
Extraction of Total Fat From Feeds/Foods 8
Extraction of PCBs From Feeds/Foods (Wet) 9
Extraction of Dioxins From Feeds/Foods (Wet) 10
Extraction of PCBs From Fish Meal (Dry) 11
Extraction of PAHs From Fish Tissue (Dry) 12
Extraction of PAHs From Fish Tissue (Wet) 13
Extraction of Pesticides From Foods (Wet) 14
Extraction of Pesticides From Hops 15
Extraction of Fat From Nuts (Brazil) 16
Extraction of Fat From Nuts (Hazelnuts) 17
Extraction of Fat From Pumpkin Seeds 18
Extraction of Brodifacoum From Rat Poison Powder 19
Extraction of Pesticides From Rice 20
Extraction of PCBs From Soy Meal (Dry) 21
Extraction of Pesticides From Spices (Dry) 22
Extraction of Pesticides From Strawberries 23
Cannabis
Extraction of THC/THCA From Cannabis Flower 25
Extraction of Cannabinoids From Cannabis Plant 26
Extraction of Mycotoxins From Cannabis Plant 27
Extraction of Pesticides From Cannabis Plant 28
ContentsEnvironmental & Regulatory
Extraction of SVOCs Environmental Samples 30
Extraction of SVOCs EPA 3545A 31
Extraction of Dioxins From Calcined Kiln Dust 32
Extraction of SVOCs, Pesticides, TPH, PCB From Cement Mix 33
Extraction of SVOCs From Environmental Samples 34
Extraction of PAHs From Fish Tissue (Wet) 35
Extraction of Organics From Lake Sediment 36
Extraction of PCBs From PUF Filters (Large) 37
Extraction of PCBs From PUF Filters (Small) 38
Extraction of PAHs From Quartz Air Filter 39
Extraction of Dioxins From Soil 40
Extraction of Diesel and Oils From Soil 41
Extraction of PAHs From Soil, Loam or Clay 42
Extraction of PCBs From Soil, Loam or Clay 43
Extraction of Dioxins From Soil, Loam or Clay 44
Cleaning of PUF Filters 45
Food Testing
Extraction of Pesticides From Avocados 47
Extraction of Polyphenols From Beans 48
Extraction of Fat From Bone Meal 49
Extraction of Fat From Bread 50
Extraction of Polyphenols From Cacao 51
Extraction of Cannabinoids From Cannabis Plant 52
Extraction of Mycotoxins From Cannabis Plant 53
Extraction of Pesticides From Cannabis Plant 54
Extraction of Lipids From Chicken (Baked) 55
Extraction of Lipids From Chicken (Fried) 56
Hydrolysis & Extraction of Total Fat From Chocolate 57
Hydrolysis & Extraction of Total Fat From Cookies 58
![Page 3: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/3.jpg)
Extraction of Fat From DATEM Powder 59
Extraction of Flavonoids from Echinacea 60
Extraction of Fat From Egg Powder 61
Extraction of Total Fat From Feeds/Foods 62
Extraction of PCBs From Feeds/Foods (Wet) 63
Extraction of Dioxins From Feeds/Foods (Wet) 64
Extraction of PCBs From Fish Meal (Dry) 65
Extraction of PAHs From Fish Tissue (Dry) 66
Extraction of PAHs From Fish Tissue (Wet) 67
Extraction of Pesticides From Foods (Wet) 68
Extraction From Goldenseal Root 69
Extraction of Pesticides From Hops 70
Extraction of Total Fat From Hot Dogs 71
Extraction of PAHs From Infant Formula 72
Extraction of Total Fat From Meats (High-Fat) 73
Extraction of Total Fat From Meats (Potted) 74
Extraction of Fat From Nuts (Brazil) 75
Extraction of Fat From Nuts (Cashews) 76
Extraction of Fat From Nuts (Hazelnuts) 77
Extraction of Total Fat From Nuts (Peanuts) 78
Extraction of Total Fat From Pepperoni 79
Extraction of Lipids From Potatoes (Baked) 80
Extraction of Lipids From Potatoes (Fried) 81
Extraction of Nutrients From Protein Powder 82
Extraction of Total Fat From Protein Powder 83
Extraction of Fat From Pumpkin Seeds 84
Extraction of Pesticides From Rice 85
Extraction of PCBs From Soy Meal (Dry) 86
Extraction of Pesticides From Spices (Dry) 87
Extraction From Spirulina Algae 88
Extraction of Pesticides From Strawberries 89
Extraction of PAHs From Yerba Mate 90
Geoscience & Mining
Extraction of Dioxins From Calcined Kiln Dust 92
Extraction of SVOCs, Pesticides, TPH, PCB From Cement Mix 93
Extraction of Molybdenum 94
Industrial Hygiene
Extraction of PCBs From PUF Filters (Large) 96
Extraction of PCBs From PUF Filters (Small) 97
Extraction of Brodifacoum From Rat Poison Powder 98
Cleaning of PUF Filters 99
Materials Science
Extraction of Film Powder 101
Extraction of Pigments From Plastics 102
Extraction of Additives From Polypropylene Powder 103
Extraction of Phthalates From Polyvinyl Chloride 104
Extraction of PCBs From PUF Filters (Large) 105
Extraction of PCBs From PUF Filters (Small) 106
Extraction of Brodifacoum From Rat Poison Powder 107
Extraction of Antioxidants From Resin 108
Extraction of PAHs From XAD Resin 109
Cleaning of PUF Filters 110
Metals & Alloys
Extraction of Molybdenum 112
Nutraceuticals
Extraction of Flavonoids From Echinacea 114
Extraction of PAHs From Fish Tissue (Dry) 115
Extraction of PAHs From Fish Tissue (Wet) 116
Extraction From Goldenseal Root 117
![Page 4: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/4.jpg)
Extraction of PAHs From Infant Formula 118
Extraction of Nutrients From Protein Powder 119
Extraction of Total Fat From Protein Powder 120
Extraction From Spirulina Algae 121
Extraction of PAHs From Yerba Mate 122
Personal Care Products
Extraction of PAHs From Infant Formula 124
Extraction of Mouthwash 125
Extraction of Toothpaste 126
Plastics, Polymers, & Oils
Extraction of Film Powder 128
Extraction of Additives From PET 129
Extraction of Additives From Pigment 130
Extraction of Pigments From Plastics 131
Extraction of Additives From Polyethylene Powder 132
Extraction of Chemical Modifier From Polyethylene Resin 133
Extraction of Additives From Polypropylene Powder 134
Extraction of Phthalates From Polyvinyl Chloride 135
Extraction of Diesel and Oils From Soil 136
![Page 5: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/5.jpg)
Agriculture
1
![Page 6: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/6.jpg)
Extraction of PesticidesFrom AvocadosFrom Avocados
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscM2 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc. 2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is required.
Solvents1% Acetic Acid in Acetonitrile
SorbentsQ-Matrix Hydraᵀᴹ (2.5 g)C18 (2 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 4020 1:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 1% Acetic Acid in Acetonitrile 15 0:1540
2 1% Acetic Acid in Acetonitrile 15 0:0040
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label2
![Page 7: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/7.jpg)
Extraction of PolyphenolsFrom BeansFrom Beans
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight0.1 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
Solvents(1:1) Methanol/Water
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 8020 10:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (1:1) Methanol/Water 15 0:1580
2 (1:1) Methanol/Water 15 0:0080
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label3
![Page 8: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/8.jpg)
Extraction of PolyphenolsFrom CacaoFrom Cacao
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight0.1 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
Solvents(1:1) Methanol/Water
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 8020 10:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (1:1) Methanol/Water 15 0:1580
2 (1:1) Methanol/Water 15 0:0080
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label4
![Page 9: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/9.jpg)
Extraction of CannabinoidsFrom Cannabis PlantFrom Cannabis Plant
Agriculture, Cannabis, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesPossessing, using, distributing, or selling marijuana or marijuana-based products constitute federal crimes in the United States, even where a state law decriminalizes or legalizes such activities. CEM Corporation produces instruments that are intended for use in testing laboratories and applications only where such use is permitted under applicable state/country law.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between two C9 discs. 2. Weigh the sample into the Q-Cup.
SolventsOptions:MethanolIsopropanol
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 9010 0:3055
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Methanol 30 0:3090
2 Isopropanol 20 0:3090
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label5
![Page 10: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/10.jpg)
Extraction of MycotoxinsFrom Cannabis PlantFrom Cannabis Plant
Agriculture, Cannabis, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1.5 g
Q-DiscM2 + C9C4
EquipmentEDGE40 mL collection vials - clear
NotesPossessing, using, distributing, or selling marijuana or marijuana-based products constitute federal crimes in the United States, even where a state law decriminalizes or legalizes such activities. CEM Corporation produces instruments that are intended for use in testing laboratories and applications only where such use is permitted under applicable state/country law.Either C4 or M2/C9 Q-Disc can be used.
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc. Layr the M2 disc (textured sided up) on top of the C9
disc, if M2 used. 2. Weigh the sample into the Q-Cup.
SolventsOptions:Methanol May also use a percentage of Water/Methanol as needed
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 5015 2:0055
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Methanol 15 0:3050
2 Methanol 15 0:3050
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label6
![Page 11: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/11.jpg)
Extraction of PesticidesFrom Cannabis PlantFrom Cannabis Plant
Agriculture, Cannabis, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1.5 g
Q-DiscM2 + C9
EquipmentEDGE40 mL collection vials - clear
NotesPossessing, using, distributing, or selling marijuana or marijuana-based products constitute federal crimes in the United States, even where a state law decriminalizes or legalizes such activities. CEM Corporation produces instruments that are intended for use in testing laboratories and applications only where such use is permitted under applicable state/country law.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the sorbent. No mixing is required.
Solvents1% Acetic Acid in Acetonitrile
SorbentsContents of Agilent QuEChERS SPE 15 mL Tube
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 4020 2:0055
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 1% Acetic Acid in Acetonitrile 15 0:3040
2 1% Acetic Acid in Acetonitrile 15 0:3040
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label7
![Page 12: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/12.jpg)
Extraction of Total FatFrom Feeds/FoodsFrom Feeds/Foods
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight≤ 5 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.4. Dry Q-Cups with sample and sand in the oven for 1 hour at 100° C while tilted at an angle in ~2" aluminum pans or equivalent. This provides stability and enables easier visual identification that no sample has been lost.
SolventsPetroleum Ether
SorbentsWet Samples Only: Sand (use equal amount to sample size)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14010 1:00010
2 14010 1:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Petroleum Ether 15 0:15150
2 Petroleum Ether 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label8
![Page 13: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/13.jpg)
Extraction of PCBsFrom Feeds/Foods (Wet)From Feeds/Foods (Wet)
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between the two C9 discs.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is needed.
SolventsDichloromethane
SorbentsC18 (2.5 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 13520 2:00510
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Dichloromethane 15 0:15150
2 Dichloromethane 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label9
![Page 14: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/14.jpg)
Extraction of DioxinsFrom Feeds/Foods (Wet)From Feeds/Foods (Wet)
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC4
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is needed.
SolventsDichloromethane
SorbentsC18 (2 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 15020 2:00510
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Dichloromethane 15 0:15150
2 Dichloromethane 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label10
![Page 15: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/15.jpg)
Extraction of PCBsFrom Fish Meal (Dry)From Fish Meal (Dry)
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC1M1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesLess than 5 grams of sample can be used with a ratio of 1:2 C18: Sample.Either C1 or M1/C9 Q-Disc can be used.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. If using the M1 disc, layer a C9 disc on top of M1.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the sorbent. No mixing is required.
SolventsOptions:Dichloromethane(1:1) Hexane/Acetone
SorbentsC18 (2.5 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12020 3:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Dichloromethane 15 0:15150
2 Extraction Solvent 10 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label11
![Page 16: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/16.jpg)
Extraction of PAHsFrom Fish Tissue (Dry)From Fish Tissue (Dry)
Agriculture, Food Testing, NutraceuticalEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®.2. Weigh the sorbent in the Q-Cup. Tap lightly to spread evenly.3. Weigh the sample into the Q-Cup layering the sample on top of the sorbent. No mixing is required.
SolventsDichloromethane
SorbentsC18 (1.7 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12020 3:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Dichloromethane 15 0:30150
2 Dichloromethane 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label12
![Page 17: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/17.jpg)
Extraction of PAHsFrom Fish Tissue (Wet)From Fish Tissue (Wet)
Agriculture, Environmental and RegulatEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesIf there is moisture in collection vial after the extraction:Using a glass funnel with glass wool and about 20 g of Sodium Sulfate. Pour over into new collection vial and rinse thoroughly with fresh solvent.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between the two C9 discs.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is needed.
SolventsOptions:Dichloromethane(1:1) Hexane/Acetone
SorbentsC18 (0.8 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12020 3:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Extraction Solvent 15 0:15150
2 Extraction Solvent 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label13
![Page 18: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/18.jpg)
Extraction of PesticidesFrom Foods (Wet)From Foods (Wet)
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between two C9 discs.2. Weigh the sorbents into the Q-Cup. Spread evenly across the sample by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the sorbents. No mixing is required.
Solvents1% Acetic Acid in Acetonitrile
SorbentsQ-Matrix Hydraᵀᴹ (2.5 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 4020 3:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 1% Acetic Acid in Acetonitrile 15 0:1540
2 1% Acetic Acid in Acetonitrile 15 0:0040
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label14
![Page 19: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/19.jpg)
Extraction of PesticidesFrom HopsFrom Hops
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight0.5 g
Q-DiscM2 + C9
EquipmentEDGE40 mL collection vials - clear
NotesCanadian Pesticide Mix 4 in LCMS grade obtained from SPEX CertiPrep.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc.2. Add sorbents to bottom of Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping. No mixing is required.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is required.4. Spike with 100 ug/mL Canadian Pesticide Mix 4 and leave in hood for approximately 30 minutes to evaporate solvent.
Solvents1% Acetic Acid in Acetonitrile
SorbentsQ-Matrix Hydraᵀᴹ (1.25 g)Supelᵀᴹ QuE Citrate (1.25 g)Primary secondary amine (1.25 g)Siliabond C18 (1.25 g)*Different sorbents can be used
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 4015 3:00510
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 1% Acetic Acid in Acetonitrile 15 0:0040
2 1% Acetic Acid in Acetonitrile 15 0:0040
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label15
![Page 20: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/20.jpg)
Extraction of FatFrom Nuts (Brazil)From Nuts (Brazil)
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight12 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and the sand in the Q-Cup.
SolventsPetroleum Ether
SorbentsSand (12 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14020 1:0000
2 14015 1:0005
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Petroleum Ether 15 0:15150
2 Petroleum Ether 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label16
![Page 21: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/21.jpg)
Extraction of FatFrom Nuts (Hazelnuts)From Nuts (Hazelnuts)
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight12 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.
SolventsPetroleum Ether
SorbentsSand (12 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14020 1:0000
2 14015 1:0005
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Petroleum Ether 15 0:15150
2 Petroleum Ether 0 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label17
![Page 22: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/22.jpg)
Extraction of FatFrom Pumpkin SeedsFrom Pumpkin Seeds
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight12 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.
SolventsPetroleum Ether
SorbentsSand (12 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14020 1:0000
2 14015 1:0005
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Petroleum Ether 15 0:15150
2 Petroleum Ether 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label18
![Page 23: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/23.jpg)
Extraction of BrodifacoumFrom Rat Poison PowderFrom Rat Poison Powder
Agriculture, Industrial Hygiene, MaterialEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscC4
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
SolventsMethanol
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 13015 2:001010
2 13015 2:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Methanol 15 0:15150
2 Methanol 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label19
![Page 24: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/24.jpg)
Extraction of PesticidesFrom RiceFrom Rice
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscM2 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc. 2. Weigh the sorbents into the Q-Cup. Spread evenly across the sample by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the sorbent. No mixing is required.
Solvents1% Acetic Acid in Acetonitile
SorbentsQ-Matrix Hydraᵀᴹ (2.5 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 4020 3:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 1% Acetic Acid in Acetonitrile 15 0:1540
2 1% Acetic Acid in Acetonitrile 15 0:0040
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label20
![Page 25: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/25.jpg)
Extraction of PCBsFrom Soy Meal (Dry)From Soy Meal (Dry)
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC1M1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesLess than 5 grams of sample can be used with a ratio of 1:2 C18/Sample.Either C1 or M1/C9 Q-Disc can be used.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. If using M1 disc, layer on top of C9 disc.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the sorbent. No mixing is required.
SolventsOptions:Dichloromethane(1:1) Hexane/Acetone
SorbentsC18 (2.5 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12020 3:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Dichloromethane 15 0:15150
2 Extraction Solvent 10 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label21
![Page 26: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/26.jpg)
Extraction of PesticidesFrom Spices (Dry)From Spices (Dry)
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesMatrix matched calibration curve required without the use of sorbents.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between the two C9 discs.2. Weigh the sample into the Q-Cup.
Solvents1% Acetic Acid in Acetonitrile
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 4015 3:00510
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 1% Acetic Acid in Acetonitrile 15 0:1040
2 1% Acetic Acid in Acetonitrile 15 0:0040
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label22
![Page 27: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/27.jpg)
Extraction of PesticidesFrom StrawberriesFrom Strawberries
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between 2 C9 Q-Disc. 2. Weigh the sorbents into the Q-Cup. Spread evenly across the sample by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the sorbent. No mixing is required.
Solvents1% Acetic Acid in Acetonitrile
SorbentsQ-Matrix Hydraᵀᴹ (2.5 g)Primary Second Amine (2 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 4020 3:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 1% Acetic Acid in Acetonitrile 15 0:1040
2 1% Acetic Acid in Acetonitrile 15 0:0040
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label23
![Page 28: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/28.jpg)
Cannabis
24
![Page 29: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/29.jpg)
Extraction of THC/THCAFrom Cannabis PlantFrom Cannabis Plant
CannabisEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesPossessing, using, distributing, or selling marijuana or marijuana-based products constitute federal crimes in the United States, even where a state law decriminalizes or legalizes such activities. CEM Corporation produces instruments that are intended for use in testing laboratories and applications only where such use is permitted under applicable state/country law.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup®. Sandwich the G1 disc between the two C9 discs.2. Weigh the sample into the Q-Cup.
SolventsOptions:MethanolIsopropanol
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 9010 0:3055
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Methanol 30 0:3090
2 Isopropanol 20 0:3090
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label25
![Page 30: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/30.jpg)
Extraction of CannabinoidsFrom Cannabis PlantFrom Cannabis Plant
Agriculture, Cannabis, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesPossessing, using, distributing, or selling marijuana or marijuana-based products constitute federal crimes in the United States, even where a state law decriminalizes or legalizes such activities. CEM Corporation produces instruments that are intended for use in testing laboratories and applications only where such use is permitted under applicable state/country law.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between two C9 discs. 2. Weigh the sample into the Q-Cup.
SolventsOptions:MethanolIsopropanol
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 9010 0:3055
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Methanol 30 0:3090
2 Isopropanol 20 0:3090
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label26
![Page 31: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/31.jpg)
Extraction of MycotoxinsFrom Cannabis PlantFrom Cannabis Plant
Agriculture, Cannabis, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1.5 g
Q-DiscM2 + C9C4
EquipmentEDGE40 mL collection vials - clear
NotesPossessing, using, distributing, or selling marijuana or marijuana-based products constitute federal crimes in the United States, even where a state law decriminalizes or legalizes such activities. CEM Corporation produces instruments that are intended for use in testing laboratories and applications only where such use is permitted under applicable state/country law.Either C4 or M2/C9 Q-Disc can be used.
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc. Layr the M2 disc (textured sided up) on top of the C9
disc, if M2 used. 2. Weigh the sample into the Q-Cup.
SolventsOptions:Methanol May also use a percentage of Water/Methanol as needed
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 5015 2:0055
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Methanol 15 0:3050
2 Methanol 15 0:3050
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label27
![Page 32: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/32.jpg)
Extraction of PesticidesFrom Cannabis PlantFrom Cannabis Plant
Agriculture, Cannabis, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1.5 g
Q-DiscM2 + C9
EquipmentEDGE40 mL collection vials - clear
NotesPossessing, using, distributing, or selling marijuana or marijuana-based products constitute federal crimes in the United States, even where a state law decriminalizes or legalizes such activities. CEM Corporation produces instruments that are intended for use in testing laboratories and applications only where such use is permitted under applicable state/country law.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the sorbent. No mixing is required.
Solvents1% Acetic Acid in Acetonitrile
SorbentsContents of Agilent QuEChERS SPE 15 mL Tube
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 4020 2:0055
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 1% Acetic Acid in Acetonitrile 15 0:3040
2 1% Acetic Acid in Acetonitrile 15 0:3040
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label28
![Page 33: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/33.jpg)
Environmental & Regulatory
29
![Page 34: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/34.jpg)
Extraction of SVOCsEnvironmental SamplesEnvironmental Samples
Environmental and RegulatoryEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight30 g
Q-DiscC4
EquipmentEDGE40 mL collection vials - clear
NotesThis is a general method for the miscellaneous extraction of environmental samples.
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
SolventsOptions:AcetoneDichloromethaneHexaneToluene
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12015 2:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Dichloromethane 20 0:15150
2 Extraction Solvent 10 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label30
![Page 35: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/35.jpg)
Extraction of SVOCsEPA 3545AEPA 3545A
Environmental and RegulatoryEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight30 g
Q-DiscC4
EquipmentEDGE40 mL collection vials - clear
NotesSee the CEM Application note: Extraction of Semi-Volatile Organic Compounds from Soil in Accordance with EPA 3545A if needed.
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
Solvents(1:1) Acetone/Hexane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12015 2:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (1:1) Acetone/Hexane 30 0:15150
2 (1:1) Acetone/Hexane 30 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label31
![Page 36: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/36.jpg)
Extraction of DioxinsFrom Calcined Kiln DustFrom Calcined Kiln Dust
Environmental and Regulatory, GeoscieEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight10 g
Q-DiscC9 + G1 + C9
EquipmentEDGE60 mL collection vials - amber
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup®. Sandwich the G1 Q-Disc® between the two C9 Q-Discs. 2. Weigh the sample into the Q-Cup.
SolventsOptions:DichloromethaneMethyl tert-butyl ether
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 15020 2:0000
2 15020 2:0000
3 15020 2:0000
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Dichloromethane 20 0:30150
2 Methyl tert-butyl ether 20 0:30150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label32
![Page 37: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/37.jpg)
Extraction of SVOCs, Pesticides, TPH, PCBFrom Cement MixFrom Cement Mix
Environmental and Regulatory, GeoscieEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight15 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between two C9 discs. 2. Weigh the sample into the Q-Cup.
Solvents(1:1) Acetone/Dichloromethane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 10020 3:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (1:1) Acetone/Dichloromethane 15 0:30150
2 (1:1) Acetone/Dichloromethane 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label33
![Page 38: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/38.jpg)
Extraction of SVOCsFrom Environmental SamplesFrom Environmental Samples
Environmental and RegulatoryEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight30 g
Q-DiscC4
EquipmentEDGE40 mL collection vials - clear
NotesThis is a general method for the miscellaneous extraction of environmental samples.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®.2. Weigh the sample into the Q-Cup.
SolventsOptions:Acetone DichloromethaneHexaneToluene
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12015 2:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Dichloromethane 20 0:15150
2 Extraction Solvent 10 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label34
![Page 39: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/39.jpg)
Extraction of PAHsFrom Fish Tissue (Wet)From Fish Tissue (Wet)
Agriculture, Environmental and RegulatEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesIf there is moisture in collection vial after the extraction:Using a glass funnel with glass wool and about 20 g of Sodium Sulfate. Pour over into new collection vial and rinse thoroughly with fresh solvent.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between the two C9 discs.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is needed.
SolventsOptions:Dichloromethane(1:1) Hexane/Acetone
SorbentsC18 (0.8 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12020 3:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Extraction Solvent 15 0:15150
2 Extraction Solvent 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label35
![Page 40: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/40.jpg)
Extraction of OrganicsFrom Lake SedimentFrom Lake Sediment
Environmental and Regulatory, Organic EDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, grind, or cryomill sample. Assemble the Q-Cup® with the Q-Disc®.2. Weigh the sample into the Q-Cup.
Solvents(1:1) Methanol/Dichloromethane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 10020 2:0000
2 10020 2:0000
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (1:1) Methanol/Dichloromethane 15 0:20150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label36
![Page 41: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/41.jpg)
Extraction of PCBsFrom PUF Filters (Large)From PUF Filters (Large)
Environmental and Regulatory, IndustriaEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample WeightEntire Filter
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Assemble the Q-Cup® with the Q-Disc®. 2. Compress and push the filter into the base of the Q-Cup.
SolventsHexane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 11520 3:00155
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Hexane 10 0:15150
2 Hexane 10 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label37
![Page 42: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/42.jpg)
Extraction of PCBsFrom PUF Filters (Small)From PUF Filters (Small)
Environmental and Regulatory, IndustriaEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample WeightEntire Filter
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Assemble the Q-Cup® with the Q-Disc®. 2. Compress and push the filter into the base of the Q-Cup.
SolventsHexane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 10015 0:10100
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Hexane 10 0:15150
2 Hexane 10 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label38
![Page 43: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/43.jpg)
Extraction of PAHsFrom Quartz Air FilterFrom Quartz Air Filter
Environmental and RegulatoryEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight3 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sorbent in the Q-Cup. Tap lightly to spread evenly.3. Add the sample on top of the sorbent. No mixing is required.
Solvents(1:1) Hexane/Dichloromethane
SorbentsSand (3 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12020 2:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (1:1) Hexane/Dichloromethane 15 0:15100
2 (1:1) Hexane/Dichloromethane 15 0:00100
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label39
![Page 44: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/44.jpg)
Extraction of DioxinsFrom SoilFrom Soil
Environmental and RegulatoryEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight10 g
Q-DiscC1
EquipmentEDGE60 mL collection vials - clear
NotesNone
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
SolventsOptions:DichloromethaneMethy tert-butyl etherToluene
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 15020 1:0000
2 15020 1:0000
3 15020 1:0000
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Extraction Solvent 30 0:30150
2 Extraction Solvent 30 0:30150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label40
![Page 45: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/45.jpg)
Extraction of Diesel and OilsFrom SoilFrom Soil
Environmental and Regulatory, Plastics,EDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between the two C9 discs. 2. Weigh the sorbent into the Q-Cup.3. Weigh the sample into the Q-Cup on top of the sorbent. No mixing is required.
SolventsDichloromethane
SorbentsDiatomaceous Earth (1 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 10020 2:0000
2 10010 2:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Dichloromethane 15 0:30150
2 Dichloromethane 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label41
![Page 46: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/46.jpg)
Extraction of PAHsFrom Soil, Loam or ClayFrom Soil, Loam or Clay
Environmental and RegulatoryEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight30 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
Solvents(1:1) Acetone/Hexane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12015 1:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (1:1) Acetone/Hexane 15 0:30150
2 (1:1) Acetone/Hexane 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label42
![Page 47: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/47.jpg)
Extraction of PCBsFrom Soil, Loam or ClayFrom Soil, Loam or Clay
Environmental and RegulatoryEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight30 g
Q-DiscC4
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
Solvents(1:1) Acetone/Hexane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12015 1:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (1:1) Acetone/Hexane 15 0:30150
2 (1:1) Acetone/Hexane 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.d) Bring to volume or evaporate as necessary for analysis.
label43
![Page 48: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/48.jpg)
Extraction of DioxinsFrom Soil, Loam or ClayFrom Soil, Loam or Clay
Environmental and RegulatoryEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscC1
EquipmentEDGE60 mL collection vials - clear
NotesNone
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
SolventsToluene
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 15020 1:0000
2 15020 1:0000
3 15015 1:0005
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Toluene 30 0:30150
2 Toluene 30 0:30150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label44
![Page 49: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/49.jpg)
Cleaning ofPUF FiltersPUF Filters
Environmental and Regulatory, IndustriaEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample WeightEntire Filter
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Assemble the Q-Cup® with the Q-Disc®. 2. Compress and push the filter into the base of the Q-Cup.
SolventsOptions:HexaneAcetone
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 11510 3:00100
2 11510 3:00100
3 11530 3:00100
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Hexane 15 0:15150
2 Acetone 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label45
![Page 50: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/50.jpg)
Food Testing
46
![Page 51: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/51.jpg)
Extraction of PesticidesFrom AvocadosFrom Avocados
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscM2 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc. 2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is required.
Solvents1% Acetic Acid in Acetonitrile
SorbentsQ-Matrix Hydraᵀᴹ (2.5 g)C18 (2 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 4020 1:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 1% Acetic Acid in Acetonitrile 15 0:1540
2 1% Acetic Acid in Acetonitrile 15 0:0040
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label47
![Page 52: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/52.jpg)
Extraction of PolyphenolsFrom BeansFrom Beans
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight0.1 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
Solvents(1:1) Methanol/Water
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 8020 10:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (1:1) Methanol/Water 15 0:1580
2 (1:1) Methanol/Water 15 0:0080
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtainrequired results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position andload method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label48
![Page 53: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/53.jpg)
Extraction of FatFrom Bone MealFrom Bone Meal
Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight10 g
Q-DiscC4
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Assemble the Q-Cup® with the Q-Disc®.2. Weigh the sorbent into the Q-Cup.3. Weigh the sample into the Q-Cup. Gently mix the sample and the sorbent in the Q-Cup.
SolventsPetroleum Ether
SorbentsDiatomaceous Earth (5 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14020 1:0000
2 14010 1:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Petroluem Ether 15 0:15150
2 Petroluem Ether 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label49
![Page 54: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/54.jpg)
Extraction of FatFrom BreadFrom Bread
Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight4 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Assemble the Q-Cup® with the Q-Disc®.2. Dry the sample and mill, or blend, to a fine powder.3. Weigh the sample into the Q-Cup.
SolventsPetroleum Ether
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14010 1:00010
2 14010 1:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Petroleum Ether 15 0:15150
2 Petroleum Ether 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label50
![Page 55: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/55.jpg)
Extraction of PolyphenolsFrom CacaoFrom Cacao
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight0.1 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
Solvents(1:1) Methanol/Water
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 8020 10:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (1:1) Methanol/Water 15 0:1580
2 (1:1) Methanol/Water 15 0:0080
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label51
![Page 56: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/56.jpg)
Extraction of CannabinoidsFrom Cannabis PlantFrom Cannabis Plant
Agriculture, Cannabis, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesPossessing, using, distributing, or selling marijuana or marijuana-based products constitute federal crimes in the United States, even where a state law decriminalizes or legalizes such activities. CEM Corporation produces instruments that are intended for use in testing laboratories and applications only where such use is permitted under applicable state/country law.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between two C9 discs. 2. Weigh the sample into the Q-Cup.
SolventsOptions:MethanolIsopropanol
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 9010 0:3055
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Methanol 30 0:3090
2 Isopropanol 20 0:3090
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label52
![Page 57: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/57.jpg)
Extraction of MycotoxinsFrom Cannabis PlantFrom Cannabis Plant
Agriculture, Cannabis, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1.5 g
Q-DiscM2 + C9C4
EquipmentEDGE40 mL collection vials - clear
NotesPossessing, using, distributing, or selling marijuana or marijuana-based products constitute federal crimes in the United States, even where a state law decriminalizes or legalizes such activities. CEM Corporation produces instruments that are intended for use in testing laboratories and applications only where such use is permitted under applicable state/country law.Either C4 or M2/C9 Q-Disc can be used.
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc. Layr the M2 disc (textured sided up) on top of the C9
disc, if M2 used. 2. Weigh the sample into the Q-Cup.
SolventsOptions:Methanol May also use a percentage of Water/Methanol as needed
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 5015 2:0055
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Methanol 15 0:3050
2 Methanol 15 0:3050
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label53
![Page 58: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/58.jpg)
Extraction of PesticidesFrom Cannabis PlantFrom Cannabis Plant
Agriculture, Cannabis, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1.5 g
Q-DiscM2 + C9
EquipmentEDGE40 mL collection vials - clear
NotesPossessing, using, distributing, or selling marijuana or marijuana-based products constitute federal crimes in the United States, even where a state law decriminalizes or legalizes such activities. CEM Corporation produces instruments that are intended for use in testing laboratories and applications only where such use is permitted under applicable state/country law.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the sorbent. No mixing is required.
Solvents1% Acetic Acid in Acetonitrile
SorbentsContents of Agilent QuEChERS SPE 15 mL Tube
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 4020 2:0055
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 1% Acetic Acid in Acetonitrile 15 0:3040
2 1% Acetic Acid in Acetonitrile 15 0:3040
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label54
![Page 59: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/59.jpg)
Extraction of LipidsFrom Chicken (Baked)From Chicken (Baked)
Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight9 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®.2. Weigh the sample into the Q-Cup.
Solvents(2:1) Chloroform/Methanol
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 3030 4:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (2:1) Chloroform/Methanol 30 0:2030
2 (2:1) Chloroform/Methanol 30 0:0030
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label55
![Page 60: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/60.jpg)
Extraction of LipidsFrom Chicken (Fried)From Chicken (Fried)
Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight9 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®.2. Weigh the sample into the Q-Cup.
Solvents(2:1) Chloroform/Methanol
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 3030 4:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (2:1) Chloroform/Methanol 15 0:2030
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label56
![Page 61: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/61.jpg)
Hydrolysis & Extraction of Total FatFrom ChocolateFrom Chocolate
Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesStep 2- EDGE Method1. Assemble the Q-Cup with the Q-Disc®. Sandwich the G1 disc between the two C9 discs. 2. Place the Filter Paper into the Q-Cup.
ProcedureStep 1- Hydrolysis Method 1. Weigh out 1 g of milled or ground sample. 2. Add 1 mL Ethanol and 5 mL (25:11) Concentrated Hydrochloric Acid:Water. Heat for 1 hr at 100° C in a covered dish. 3. Add 5 mL of Ethanol. Filter the sample using Whatman 3 Filter Paper with a filter funnel, a vacuum pump, and filter flask. Wash any residual material from the beaker in the filter funnel using Ethanol. 4. Dry the filter paper in an oven at 100° C for 30 minutes. After the drying step, place Filter Paper into the Q-Cup® and extract.
SolventsExtraction:Petroleum Ether(25:11) Concentrated Hydrochloric Acid/WaterWash:Water
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14010 1:00010
2 14010 1:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Water 15 0:00100
2 Petroleum Ether 15 0:15100
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label57
![Page 62: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/62.jpg)
Hydrolysis & Extraction of Total FatFrom CookiesFrom Cookies
Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesStep 2- EDGE Method1. Assemble the Q-Cup with the Q-Disc®. Sandwich the G1 disc between the two C9 discs. 2. Place the Filter Paper into the Q-Cup.
ProcedureStep 1- Hydrolysis Method 1. Weigh out 1 g of milled or ground sample. 2. Add 1 mL Ethanol and 5 mL (25:11) Concentrated Hydrochloric Acid:Water. Heat for 1 hr at 100° C in a covered dish. 3. Add 5 mL of Ethanol. Filter the sample using Whatman 3 Filter Paper with a filter funnel, a vacuum pump, and filter flask. Wash any residual material from the beaker in the filter funnel using Ethanol. 4. Dry the filter paper in an oven at 100° C for 30 minutes. After the drying step, place Filter Paper into the Q-Cup® and extract.
SolventsExtraction:Petroleum Ether(25:11) Concentrated Hyrdochloric Acid/WaterWash:Water
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14010 1:00010
2 14010 1:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Water 15 0:00150
2 Petroleum Ether 15 0:15150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label58
![Page 63: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/63.jpg)
Extraction of FatFrom DATEM PowderFrom DATEM Powder
Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscC4
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Assemble the Q-Cup® with the Q-Disc®.2. Weigh the sample into the Q-Cup.3. Weigh the sand directly on top of the sample to compact down.
SolventsPetroleum Ether
SorbentsSand (1 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 10010 1:0000
2 10010 1:0000
3 10010 1:0000
4 1005 1:0005
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Petroluem Ether 15 0:15150
2 Petroluem Ether 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
59
![Page 64: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/64.jpg)
Extraction of FlavonoidsFrFromom Echinacea
Food TestingEDGE
label
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscM2 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Layer the M2 Q-Disc (textured side up) on top of theC9 Q-Disc.2. Weigh the sample into the Q-Cup.
Solvents(70:30) Methanol/Water
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 6020 1:00020
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (70:30) Methanol/Water 15 0:1060
2 (70:30) Methanol/Water 10 0:0060
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtainrequired results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position andload method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label 60
![Page 65: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/65.jpg)
Extraction of FatFrom Egg PowderFrom Egg Powder
Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC4
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
SolventsToluene
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14020 1:0000
2 14015 1:0005
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Toluene 15 0:15150
2 Toluene 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label61
![Page 66: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/66.jpg)
Extraction of Total FatFrom Feeds/FoodsFrom Feeds/Foods
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight≤ 5 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.4. Dry Q-Cups with sample and sand in the oven for 1 hour at 100° C while tilted at an angle in ~2" aluminum pans or equivalent. This provides stability and enables easier visual identification that no sample has been lost.
SolventsPetroleum Ether
SorbentsWet Samples Only: Sand (use equal amount to sample size)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14010 1:00010
2 14010 1:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Petroleum Ether 15 0:15150
2 Petroleum Ether 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label62
![Page 67: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/67.jpg)
Extraction of PCBsFrom Feeds/Foods (Wet)From Feeds/Foods (Wet)
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between the two C9 discs.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is needed.
SolventsDichloromethane
SorbentsC18 (2.5 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 13520 2:00510
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Dichloromethane 15 0:15150
2 Dichloromethane 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label63
![Page 68: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/68.jpg)
Extraction of DioxinsFrom Feeds/Foods (Wet)From Feeds/Foods (Wet)
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC4
EquipmentEDGE60 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is needed.
SolventsOptions:DichloromethaneMethyl tert-butyl etherToluene
SorbentsC18 (2 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 15020 1:0000
2 15020 1:0000
3 15020 1:0000
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Extraction Solvent 30 0:30150
2 Extraction Solvent 30 0:30150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label64
![Page 69: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/69.jpg)
Extraction of PCBsFrom Fish Meal (Dry)From Fish Meal (Dry)
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC1M1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesLess than 5 grams of sample can be used with a ratio of 1:2 C18: Sample.Either C1 or M1/C9 Q-Disc can be used.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. If using the M1 disc, layer a C9 disc on top of M1.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the sorbent. No mixing is required.
SolventsOptions:Dichloromethane(1:1) Hexane/Acetone
SorbentsC18 (2.5 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12020 3:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Dichloromethane 15 0:15150
2 Extraction Solvent 10 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label65
![Page 70: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/70.jpg)
Extraction of PAHsFrom Fish Tissue (Dry)From Fish Tissue (Dry)
Agriculture, Food Testing, NutraceuticalEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®.2. Weigh the sorbent in the Q-Cup. Tap lightly to spread evenly.3. Weigh the sample into the Q-Cup layering the sample on top of the sorbent. No mixing is required.
SolventsDichloromethane
SorbentsC18 (1.7 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12020 3:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Dichloromethane 15 0:30150
2 Dichloromethane 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label66
![Page 71: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/71.jpg)
Extraction of PAHsFrom Fish Tissue (Wet)From Fish Tissue (Wet)
Agriculture, Environmental and RegulatEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesIf there is moisture in collection vial after the extraction:Using a glass funnel with glass wool and about 20 g of Sodium Sulfate. Pour over into new collection vial and rinse thoroughly with fresh solvent.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between the two C9 discs.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is needed.
SolventsOptions:Dichloromethane(1:1) Hexane/Acetone
SorbentsC18 (0.8 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12020 3:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Extraction Solvent 15 0:15150
2 Extraction Solvent 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label67
![Page 72: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/72.jpg)
Extraction of PesticidesFrom Foods (Wet)From Foods (Wet)
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between two C9 discs.2. Weigh the sorbents into the Q-Cup. Spread evenly across the sample by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the sorbents. No mixing is required.
Solvents1% Acetic Acid in Acetonitrile
SorbentsQ-Matrix Hydraᵀᴹ (2.5 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 4020 3:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 1% Acetic Acid in Acetonitrile 15 0:1540
2 1% Acetic Acid in Acetonitrile 15 0:0040
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label68
![Page 73: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/73.jpg)
ExtractionFrom Goldenseal RootFrom Goldenseal Root
Food Testing, NutraceuticalsEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Assemble the Q-Cup® with the Q-Disc®.2. Weigh the sample into the Q-Cup.
SolventsAcetonitrile
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 10015 1:00105
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Acetonitrile 15 0:15150
2 Acetonitrile 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label69
![Page 74: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/74.jpg)
Extraction of PesticidesFrom HopsFrom Hops
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight0.5 g
Q-DiscM2 + C9
EquipmentEDGE40 mL collection vials - clear
NotesCanadian Pesticide Mix 4 in LCMS grade obtained from SPEX CertiPrep.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc.2. Add sorbents to bottom of Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping. No mixing is required.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is required.4. Spike with 100 ug/mL Canadian Pesticide Mix 4 and leave in hood for approximately 30 minutes to evaporate solvent.
Solvents1% Acetic Acid in Acetonitrile
SorbentsQ-Matrix Hydraᵀᴹ (1.25 g)Supelᵀᴹ QuE Citrate (1.25 g)Primary secondary amine (1.25 g)Siliabond C18 (1.25 g)*Different sorbents can be used
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 4015 3:00510
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 1% Acetic Acid in Acetonitrile 15 0:0040
2 1% Acetic Acid in Acetonitrile 15 0:0040
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label70
![Page 75: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/75.jpg)
Extraction of Total FatFrom Hot DogsFrom Hot Dogs
Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and the sand in the Q-Cup.4. Dry Q-Cups with the sample and the sand in the oven for 1 hour at 100° C while tilted at an angle in ~2" aluminum pans or equivalent. This provides stability and enables easier visual identification that no sample has been lost.
SolventsExtraction:Petroleum EtherWash:Water
SorbentsSand (2 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14010 1:00010
2 14010 1:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Water 10 0:00150
2 Petroleum Ether 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label71
![Page 76: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/76.jpg)
Extraction of PAHsFrom Infant FormulaFrom Infant Formula
Food Testing, Nutraceuticals, Personal EDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight4 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between two C9 discs.2. Weigh the sample into the Q-Cup.
SolventsCyclohexane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12020 1:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Cyclohexane 15 0:30150
2 Cyclohexane 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label72
![Page 77: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/77.jpg)
Extraction of Total FatFrom Meats (High-Fat)From Meats (High-Fat)
Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.4. Dry Q-Cups with the sample and the sand in the oven for 1 hour at 100° C while tilted at an angle in ~2" aluminum pans or equivalent. This provides stability and enables easier visual identification that no sample has been lost.
SolventsPetroleum Ether
SorbentsSand (2 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14010 1:00010
2 14010 1:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Petroleum Ether 15 0:15150
2 Petroleum Ether 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label73
![Page 78: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/78.jpg)
Extraction of Total FatFrom Meats (Potted)From Meats (Potted)
Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and the sand in the Q-Cup.4. Dry Q-Cups with the sample and the sand in the oven for 1 hour at 100° C while tilted at an angle in ~2" aluminum pans or equivalent. This provides stability and enables easier visual identification that no sample has been lost.
SolventsExtraction:Petroleum EtherWash:Water
SorbentsSand (2 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14010 1:00010
2 14010 1:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Water 10 0:00150
2 Petroleum Ether 15 0:15150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label74
![Page 79: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/79.jpg)
Extraction of FatFrom Nuts (Brazil)From Nuts (Brazil)
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight12 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and the sand in the Q-Cup.
SolventsPetroleum Ether
SorbentsSand (12 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14020 1:0000
2 14015 1:0005
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Petroleum Ether 15 0:15150
2 Petroleum Ether 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label75
![Page 80: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/80.jpg)
Extraction of FatFrom Nuts (Cashews)From Nuts (Cashews)
Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight12 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®.2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.
SolventsPetroleum Ether
SorbentsSand (12 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14020 1:0000
2 14015 1:0005
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Petroleum Ether 15 0:15150
2 Petroleum Ether 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label76
![Page 81: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/81.jpg)
Extraction of FatFrom Nuts (Hazelnuts)From Nuts (Hazelnuts)
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight12 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.
SolventsPetroleum Ether
SorbentsSand (12 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14020 1:0000
2 14015 1:0005
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Petroleum Ether 15 0:15150
2 Petroleum Ether 0 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label77
![Page 82: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/82.jpg)
Extraction of Total FatFrom Nuts (Peanuts)From Nuts (Peanuts)
Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight12 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and the sand in the Q-Cup.4. Dry Q-Cups with the sample and the sand in the oven for 1 hour at 100° C while tilted at an angle in ~2" aluminum pans or equivalent. This provides stability and enables easier visual identification that no sample has been lost.
SolventsPetroleum Ether
SorbentsSand (12 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14020 1:0000
2 14015 1:0005
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Petroleum Ether 15 0:15150
2 Petroleum Ether 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label78
![Page 83: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/83.jpg)
Extraction of Total FatFrom PepperoniFrom Pepperoni
Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and the sand in the Q-Cup.4. Dry Q-Cups with the sample and the sand in the oven for 1 hour at 100° C while tilted at an angle in ~2" aluminum pans or equivalent. This provides stability and enables easier visual identification that no sample has been lost.
SolventsPetroleum Ether
SorbentsSand (2 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14010 1:00010
2 14010 1:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Water 10 0:00150
2 Petroleum Ether 15 0:15150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label79
![Page 84: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/84.jpg)
Extraction of LipidsFrom Potatoes (Baked)From Potatoes (Baked)
Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight9 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®.2. Weigh the sample into the Q-Cup.
Solvents(3:2) Hexane/Isopropanol
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 3030 4:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (3:2) Hexane/Isopropanol 30 0:2030
2 (3:2) Hexane/Isopropanol 30 0:0030
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label80
![Page 85: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/85.jpg)
Extraction of LipidsFrom Potatoes (Fried)From Potatoes (Fried)
Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight9 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®.2. Weigh the sample into the Q-Cup.
Solvents(3:2) Hexane/Isopropanol
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 3030 4:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (3:2) Hexane/Isopropanol 15 0:2030
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label81
![Page 86: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/86.jpg)
Extraction of NutrientsFrom Protein PowderFrom Protein Powder
Food Testing, NutraceuticalsEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscC4
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and the sand in the Q-Cup.
Solvents(80:20) Methanol/Water
SorbentsSand (3 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 15015 0:101010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (80:20) Methanol/Water 15 0:15150
2 (80:20) Methanol/Water 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label82
![Page 87: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/87.jpg)
Extraction of Total FatFrom Protein PowderFrom Protein Powder
Food Testing, NutraceuticalsEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and the sand in the Q-Cup.
SolventsPetroleum Ether
SorbentsSand (6 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12015 2:00105
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Petroleum Ether 15 0:15150
2 Petroleum Ether 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label83
![Page 88: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/88.jpg)
Extraction of FatFrom Pumpkin SeedsFrom Pumpkin Seeds
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight12 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.
SolventsPetroleum Ether
SorbentsSand (12 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14020 1:0000
2 14015 1:0005
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Petroleum Ether 15 0:15150
2 Petroleum Ether 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label84
![Page 89: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/89.jpg)
Extraction of PesticidesFrom RiceFrom Rice
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscM2 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc. 2. Weigh the sorbents into the Q-Cup. Spread evenly across the sample by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the sorbent. No mixing is required.
Solvents1% Acetic Acid in Acetonitile
SorbentsQ-Matrix Hydraᵀᴹ (2.5 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 4020 3:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 1% Acetic Acid in Acetonitrile 15 0:1540
2 1% Acetic Acid in Acetonitrile 15 0:0040
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label85
![Page 90: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/90.jpg)
Extraction of PCBsFrom Soy Meal (Dry)From Soy Meal (Dry)
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC1M1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesLess than 5 grams of sample can be used with a ratio of 1:2 C18/Sample.Either C1 or M1/C9 Q-Disc can be used.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. If using M1 disc, layer on top of C9 disc.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the sorbent. No mixing is required.
SolventsOptions:Dichloromethane(1:1) Hexane/Acetone
SorbentsC18 (2.5 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12020 3:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Dichloromethane 15 0:15150
2 Extraction Solvent 10 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label86
![Page 91: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/91.jpg)
Extraction of PesticidesFrom Spices (Dry)From Spices (Dry)
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesMatrix matched calibration curve required without the use of sorbents.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between the two C9 discs.2. Weigh the sample into the Q-Cup.
Solvents1% Acetic Acid in Acetonitrile
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 4015 3:00510
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 1% Acetic Acid in Acetonitrile 15 0:1040
2 1% Acetic Acid in Acetonitrile 15 0:0040
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label87
![Page 92: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/92.jpg)
ExtractionFrom Spirulina AlgaeFrom Spirulina Algae
Food Testing, NutraceuticalsEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Assemble the Q-Cup® with the Q-Disc®.
2. Weigh the sample into the Q-Cup.
SolventsAcetonitrile
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 10015 1:00105
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Acetonitrile 15 0:00150
2 Acetonitrile 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label88
![Page 93: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/93.jpg)
Extraction of PesticidesFrom StrawberriesFrom Strawberries
Agriculture, Food TestingEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between 2 C9 Q-Disc. 2. Weigh the sorbents into the Q-Cup. Spread evenly across the sample by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the sorbent. No mixing is required.
Solvents1% Acetic Acid in Acetonitrile
SorbentsQ-Matrix Hydraᵀᴹ (2.5 g)Primary Second Amine (2 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 4020 3:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 1% Acetic Acid in Acetonitrile 15 0:1040
2 1% Acetic Acid in Acetonitrile 15 0:0040
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label89
![Page 94: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/94.jpg)
Extraction of PAHsFrom Yerba MateFrom Yerba Mate
Food Testing, NutraceuticalsEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight0.5 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
SolventsOptions:(1:1) Hexane/AcetoneDichloromethaneAcetone
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 120-15015 1:00105
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Extraction Solvent 15 0:30150
2 Extraction Solvent 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label90
![Page 95: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/95.jpg)
Geoscience & Mining
91
![Page 96: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/96.jpg)
Extraction of DioxinsFrom Calcined Kiln DustFrom Calcined Kiln Dust
Environmental and Regulatory, GeoscieEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight10 g
Q-DiscC9 + G1 + C9
EquipmentEDGE60 mL collection vials - amber
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup®. Sandwich the G1 Q-Disc® between the two C9 Q-Discs. 2. Weigh the sample into the Q-Cup.
SolventsOptions:DichloromethaneMethyl tert-butyl ether
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 15020 2:0000
2 15020 2:0000
3 15020 2:0000
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Dichloromethane 20 0:30150
2 Methyl tert-butyl ether 20 0:30150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label92
![Page 97: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/97.jpg)
Extraction of SVOCs, Pesticides, TPH, PCBFrom Cement MixFrom Cement Mix
Environmental and Regulatory, GeoscieEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight15 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between two C9 discs. 2. Weigh the sample into the Q-Cup.
Solvents(1:1) Acetone/Dichloromethane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 10020 3:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (1:1) Acetone/Dichloromethane 15 0:30150
2 (1:1) Acetone/Dichloromethane 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label93
![Page 98: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/98.jpg)
Extraction ofMolybdenumMolybdenum
Geoscience and Mining, Metals and AllEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
SolventsAcetone
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14015 1:00105
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Acetone 15 0:15140
2 Acetone 15 0:00140
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label94
![Page 99: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/99.jpg)
Industrial Hygiene
95
![Page 100: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/100.jpg)
Extraction of PCBsFrom PUF Filters (Large)From PUF Filters (Large)
Environmental and Regulatory, IndustriaEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample WeightEntire Filter
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Assemble the Q-Cup® with the Q-Disc®. 2. Compress and push the filter into the base of the Q-Cup.
SolventsHexane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 11520 3:00155
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Hexane 10 0:15150
2 Hexane 10 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label96
![Page 101: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/101.jpg)
Extraction of PCBsFrom PUF Filters (Small)From PUF Filters (Small)
Environmental and Regulatory, IndustriaEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample WeightEntire Filter
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Assemble the Q-Cup® with the Q-Disc®. 2. Compress and push the filter into the base of the Q-Cup.
SolventsHexane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 10015 0:10100
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Hexane 10 0:15150
2 Hexane 10 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label97
![Page 102: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/102.jpg)
Extraction of BrodifacoumFrom Rat Poison PowderFrom Rat Poison Powder
Agriculture, Industrial Hygiene, MaterialEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscC4
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
SolventsMethanol
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 13015 2:001010
2 13015 2:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Methanol 15 0:15150
2 Methanol 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label98
![Page 103: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/103.jpg)
Cleaning ofPUF FiltersPUF Filters
Environmental and Regulatory, IndustriaEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample WeightEntire Filter
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Assemble the Q-Cup® with the Q-Disc®. 2. Compress and push the filter into the base of the Q-Cup.
SolventsOptions:HexaneAcetone
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 11510 3:00100
2 11510 3:00100
3 11530 3:00100
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Hexane 15 0:15150
2 Acetone 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label99
![Page 104: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/104.jpg)
Materials Science
100
![Page 105: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/105.jpg)
Extraction ofFilm PowderFilm Powder
Materials Science, Plastics, Polymers, aEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscM1 + C9
EquipmentEDGE40 mL collection vials - clear
Notes*Plastics and polymers should only be extracted at temperatures below melting point range.Expansion of the sample during extraction is common during polymer based extraction.
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc. Layer the M1 disc on top of the C9 disc. 2. Weigh the sample into the Q-Cup.
SolventsOptions:CyclohexaneDichloromethane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 *40-7015 10:0005
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Cyclohexane 15 0:0070
2 Dichloromethane 15 0:1570
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label101
![Page 106: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/106.jpg)
Extraction of PigmentsFrom PlasticsFrom Plastics
Materials Science, Plastics, Polymers, aEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscM1 + C9
EquipmentEDGE40 mL collection vials - clear
Notes*Plastics and polymers should only be extracted at temperatures below melting point range.Expansion of the sample during extraction is common during polymer based extraction.
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc. Layer the M1 disc on top of the C9 disc. 2. Weigh the sample into the Q-Cup.
SolventsOptions:CyclohexaneDichloromethane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 *40-8015 10:0005
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Cyclohexane 10 0:1580
2 Dichloromethane 10 0:0080
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label102
![Page 107: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/107.jpg)
Extraction of AdditivesFrom Polypropylene PowderFrom Polypropylene Powder
Materials Science, Plastics, Polymers, aEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
Notes*Plastics and polymers should only be extracted at temperatures below melting point range.Expansion of the sample during extraction is common during polymer based extraction.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between the two C9 discs. 2. Weigh the sample into the Q-Cup.
SolventsIsopropanol
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12030 15:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Isopropanol 15 0:15150
2 Isopropanol 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label103
![Page 108: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/108.jpg)
Extraction of PhthalatesFrom Polyvinyl ChlorideFrom Polyvinyl Chloride
Materials Science, Plastics, Polymers, aEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
Notes*Plastics and polymers should only be extracted at temperatures below melting point range.Expansion of the sample during extraction is common during polymer based extraction.
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc. Sandwich the G disc between 2 C9 discs. 2. Weigh the sample into the Q-Cup
Solvents(1:1) Isopropanol/Cyclohexane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 *8020 10:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (1:1) Isopropanol/Cyclohexane 10 0:1580
2 (1:1) Isopropanol/Cyclohexane 10 0:0080
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label104
![Page 109: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/109.jpg)
Extraction of PCBsFrom PUF Filters (Large)From PUF Filters (Large)
Environmental and Regulatory, IndustriaEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample WeightEntire Filter
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Assemble the Q-Cup® with the Q-Disc®. 2. Compress and push the filter into the base of the Q-Cup.
SolventsHexane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 11520 3:00155
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Hexane 10 0:15150
2 Hexane 10 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label105
![Page 110: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/110.jpg)
Extraction of PCBsFrom PUF Filters (Small)From PUF Filters (Small)
Environmental and Regulatory, IndustriaEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample WeightEntire Filter
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Assemble the Q-Cup® with the Q-Disc®. 2. Compress and push the filter into the base of the Q-Cup.
SolventsHexane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 10015 0:10100
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Hexane 10 0:15150
2 Hexane 10 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label106
![Page 111: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/111.jpg)
Extraction of BrodifacoumFrom Rat Poison PowderFrom Rat Poison Powder
Agriculture, Industrial Hygiene, MaterialEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscC4
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
SolventsMethanol
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 13015 2:001010
2 13015 2:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Methanol 15 0:15150
2 Methanol 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label107
![Page 112: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/112.jpg)
Extraction of AntioxidantsFrom ResinFrom Resin
Materials ScienceEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
SolventsMethanol
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 10015 2:00510
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Methanol 15 0:20150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label108
![Page 113: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/113.jpg)
Extraction of PAHsFrom XAD ResinFrom XAD Resin
Materials ScienceEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight15 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesBe sure that sample or sand is present in the Q-Cup when running this method. Sample size does matter, as you decrease the sample size you will build up more pressure within the cell.
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
SolventsDichloromethane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 15020 2:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Dichloromethane 15 0:30150
2 Dichloromethane 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label109
![Page 114: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/114.jpg)
Cleaning ofPUF FiltersPUF Filters
Environmental and Regulatory, IndustriaEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample WeightEntire Filter
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Assemble the Q-Cup® with the Q-Disc®. 2. Compress and push the filter into the base of the Q-Cup.
SolventsOptions:HexaneAcetone
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 11510 3:00100
2 11510 3:00100
3 11530 3:00100
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Hexane 15 0:15150
2 Acetone 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label110
![Page 115: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/115.jpg)
Metals & Alloys
111
![Page 116: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/116.jpg)
Extraction ofMolybdenumMolybdenum
Geoscience and Mining, Metals and AllEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
SolventsAcetone
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 14015 1:00105
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Acetone 15 0:15140
2 Acetone 15 0:00140
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label112
![Page 117: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/117.jpg)
Nutraceuticals
113
![Page 118: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/118.jpg)
Extraction of FlavonoidsFrom EchinaceaFrom Echinacea
Food Testing, NutraceuticalsEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscM2 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc.2. Weigh the sample into the Q-Cup.
Solvents(70:30) Methanol/Water
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 6020 1:00020
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (70:30) Methanol/Water 15 0:1060
2 (70:30) Methanol/Water 10 0:0060
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label114
![Page 119: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/119.jpg)
Extraction of PAHsFrom Fish Tissue (Dry)From Fish Tissue (Dry)
Agriculture, Food Testing, NutraceuticalEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®.2. Weigh the sorbent in the Q-Cup. Tap lightly to spread evenly.3. Weigh the sample into the Q-Cup layering the sample on top of the sorbent. No mixing is required.
SolventsDichloromethane
SorbentsC18 (1.7 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12020 3:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Dichloromethane 15 0:30150
2 Dichloromethane 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label115
![Page 120: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/120.jpg)
Extraction of PAHsFrom Fish Tissue (Wet)From Fish Tissue (Wet)
Agriculture, Environmental and RegulatEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesIf there is moisture in collection vial after the extraction:Using a glass funnel with glass wool and about 20 g of Sodium Sulfate. Pour over into new collection vial and rinse thoroughly with fresh solvent.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between the two C9 discs.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is needed.
SolventsOptions:Dichloromethane(1:1) Hexane/Acetone
SorbentsC18 (0.8 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12020 3:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Extraction Solvent 15 0:15150
2 Extraction Solvent 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label116
![Page 121: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/121.jpg)
ExtractionFrom Goldenseal RootFrom Goldenseal Root
Food Testing, NutraceuticalsEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Assemble the Q-Cup® with the Q-Disc®.2. Weigh the sample into the Q-Cup.
SolventsAcetonitrile
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 10015 1:00105
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Acetonitrile 15 0:15150
2 Acetonitrile 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label117
![Page 122: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/122.jpg)
Extraction of PAHsFrom Infant FormulaFrom Infant Formula
Food Testing, Nutraceuticals, Personal EDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight4 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between two C9 discs.2. Weigh the sample into the Q-Cup.
SolventsCyclohexane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12020 1:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Cyclohexane 15 0:30150
2 Cyclohexane 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label118
![Page 123: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/123.jpg)
Extraction of NutrientsFrom Protein PowderFrom Protein Powder
Food Testing, NutraceuticalsEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscC4
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and the sand in the Q-Cup.
Solvents(80:20) Methanol/Water
SorbentsSand (3 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 15015 0:101010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (80:20) Methanol/Water 15 0:15150
2 (80:20) Methanol/Water 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label119
![Page 124: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/124.jpg)
Extraction of Total FatFrom Protein PowderFrom Protein Powder
Food Testing, NutraceuticalsEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and the sand in the Q-Cup.
SolventsPetroleum Ether
SorbentsSand (6 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12015 2:00105
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Petroleum Ether 15 0:15150
2 Petroleum Ether 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label120
![Page 125: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/125.jpg)
ExtractionFrom Spirulina AlgaeFrom Spirulina Algae
Food Testing, NutraceuticalsEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight2 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Assemble the Q-Cup® with the Q-Disc®.
2. Weigh the sample into the Q-Cup.
SolventsAcetonitrile
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 10015 1:00105
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Acetonitrile 15 0:00150
2 Acetonitrile 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label121
![Page 126: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/126.jpg)
Extraction of PAHsFrom Yerba MateFrom Yerba Mate
Food Testing, NutraceuticalsEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight0.5 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
SolventsOptions:(1:1) Hexane/AcetoneDichloromethaneAcetone
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 120-15015 1:00105
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Extraction Solvent 15 0:30150
2 Extraction Solvent 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label122
![Page 127: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/127.jpg)
Personal Care Products
123
![Page 128: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/128.jpg)
Extraction of PAHsFrom Infant FormulaFrom Infant Formula
Food Testing, Nutraceuticals, Personal EDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight4 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between two C9 discs.2. Weigh the sample into the Q-Cup.
SolventsCyclohexane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12020 1:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Cyclohexane 15 0:30150
2 Cyclohexane 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label124
![Page 129: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/129.jpg)
Extraction ofMouthwashMouthwash
Personal Care (Health and Beauty)EDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between the two C9 discs.
2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping. 3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is needed.
SolventsHexane
SorbentsOptions:Sodium Polyacrylate (5 g)Diatomaceous Earth (5 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 4020 3:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Hexane 15 0:15100
2 Hexane 15 0:00100
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label125
![Page 130: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/130.jpg)
Extraction ofToothpasteToothpaste
Personal Care (Health and Beauty)EDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
Notes*for sand- mix into sample*for Diatamaceous Earth- layer under sample, do not mix.
Procedure1. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between the two C9 discs. 2. Weigh the sorbent/sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup*.
SolventsHexane
SorbentsOptions:Sand (15 g)Diatamaceous Earth (5 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 4020 3:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Hexane 15 0:1540
2 Hexane 15 0:0040
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label126
![Page 131: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/131.jpg)
Plastics, Polymers, & Oils
127
![Page 132: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/132.jpg)
Extraction ofFilm PowderFilm Powder
Materials Science, Plastics, Polymers, aEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscM1 + C9
EquipmentEDGE40 mL collection vials - clear
Notes*Plastics and polymers should only be extracted at temperatures below melting point range.Expansion of the sample during extraction is common during polymer based extraction.
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc. Layer the M1 disc on top of the C9 disc. 2. Weigh the sample into the Q-Cup.
SolventsOptions:CyclohexaneDichloromethane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 *40-7015 10:0005
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Cyclohexane 15 0:0070
2 Dichloromethane 15 0:1570
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label128
![Page 133: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/133.jpg)
Extraction of AdditivesFrom PETFrom PET
Plastics, Polymers, and OilsEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
Notes*Plastics and polymers should only be extracted at temperatures below melting point range.Expansion of the sample during extraction is common during polymer based extraction.
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
Solvents(40:60) Acetone/Hexane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 115*25 5:00510
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (40:60) Acetone/Hexane 15 0:15150
2 (40:60) Acetone/Hexane 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label129
![Page 134: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/134.jpg)
Extraction of AdditivesFrom PigmentFrom Pigment
Plastics, Polymers, and OilsEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
SolventsDimethylformamide
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 10020 1:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Dimethylformamide 10 0:15125
2 Dimethylformamide 10 0:00125
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label130
![Page 135: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/135.jpg)
Extraction of PigmentsFrom PlasticsFrom Plastics
Materials Science, Plastics, Polymers, aEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscM1 + C9
EquipmentEDGE40 mL collection vials - clear
Notes*Plastics and polymers should only be extracted at temperatures below melting point range.Expansion of the sample during extraction is common during polymer based extraction.
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc. Layer the M1 disc on top of the C9 disc. 2. Weigh the sample into the Q-Cup.
SolventsOptions:CyclohexaneDichloromethane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 *40-8015 10:0005
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Cyclohexane 10 0:1580
2 Dichloromethane 10 0:0080
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label131
![Page 136: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/136.jpg)
Extraction of AdditivesFrom Polyethylene PowderFrom Polyethylene Powder
Plastics, Polymers, and OilsEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscM1 + C9
EquipmentEDGE40 mL collection vials - clear
Notes*Plastics and polymers should only be extracted at temperatures below melting point range.Expansion of the sample during extraction is common during polymer based extraction.
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Layer the C9 Q-Disc on top of the M1 Q-Disc. 2. Weigh the sample into the Q-Cup.
SolventsIsopropanol
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 *9030 15:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Isopropanol 15 0:1590
2 Isopropanol 15 0:0090
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label132
![Page 137: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/137.jpg)
Extraction of Chemical ModifierFrom Polyethylene ResinFrom Polyethylene Resin
Plastics, Polymers, and OilsEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight0.1 g
Q-DiscC1
EquipmentEDGE40 mL collection vials - clear
Notes*Plastics and polymers should only be extracted at temperatures below melting point range.Expansion of the sample during extraction is common during polymer based extraction.
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. 2. Weigh the sample into the Q-Cup.
Solvents(40:60) Acetone/Hexane
SorbentsNone
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 *6020 15:001010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (40:60) Acetone/Hexane 10 0:1560
2 (40:60) Acetone/Hexane 10 0:0060
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label133
![Page 138: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/138.jpg)
Extraction of AdditivesFrom Polypropylene PowderFrom Polypropylene Powder
Materials Science, Plastics, Polymers, aEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
Notes*Plastics and polymers should only be extracted at temperatures below melting point range.Expansion of the sample during extraction is common during polymer based extraction.
Procedure1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between the two C9 discs. 2. Weigh the sample into the Q-Cup.
SolventsIsopropanol
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 12030 15:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Isopropanol 15 0:15150
2 Isopropanol 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label134
![Page 139: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/139.jpg)
Extraction of PhthalatesFrom Polyvinyl ChlorideFrom Polyvinyl Chloride
Materials Science, Plastics, Polymers, aEDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight1 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
Notes*Plastics and polymers should only be extracted at temperatures below melting point range.Expansion of the sample during extraction is common during polymer based extraction.
Procedure 1. Homogenize, mill, or cryomill the sample. Assemble the Q-Cup® with the Q-Disc. Sandwich the G disc between 2 C9 discs. 2. Weigh the sample into the Q-Cup
Solvents(1:1) Isopropanol/Cyclohexane
SorbentsN/A
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 *8020 10:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 (1:1) Isopropanol/Cyclohexane 10 0:1580
2 (1:1) Isopropanol/Cyclohexane 10 0:0080
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.
label135
![Page 140: Automated Extraction System Method Note Compendiumcem.com/.../literature/MetNote_EDGE_Compendium_1.pdf · Automated Extraction System Method Note Compendium Updated September 25,](https://reader035.vdocument.in/reader035/viewer/2022062917/5ed6a77f39f16f294d5736cf/html5/thumbnails/140.jpg)
Extraction of Diesel and OilsFrom SoilFrom Soil
Environmental and Regulatory, Plastics,EDGE
Note: Temperature and hold times may vary depending on the sample and analytes of interest.
Heating Program
Sample Weight5 g
Q-DiscC9 + G1 + C9
EquipmentEDGE40 mL collection vials - clear
NotesNone
Procedure1. Assemble the Q-Cup® with the Q-Disc®. Sandwich the G1 disc between the two C9 discs. 2. Weigh the sorbent into the Q-Cup.3. Weigh the sample into the Q-Cup on top of the sorbent. No mixing is required.
SolventsDichloromethane
SorbentsDiatomaceous Earth (1 g)
Wash Program
Cycle Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)
1 10020 2:0000
2 10010 2:00010
Cycle Solvent Volume (mL) Hold (mm:ss)Temp (°C)
1 Dichloromethane 15 0:30150
2 Dichloromethane 15 0:00150
General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.