Calorimeter Calibration and Jet Energy Scale
Jean-Francois Arguin
November 28th, 2005
Physics 252B, UC Davis
![Page 2: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/2.jpg)
Outline
Quick remainder of calorimetry Calibration before the experiment starts: test beam Calibration when the experiment is running:
Hardware calibration Collider data
Measuring jets at high-energy colliders Example of a physics measurement: top quark
mass
![Page 3: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/3.jpg)
Basics of Calorimetry
Incident particle creates a shower inside material Shower can be either
electromagnetic or hadronic
Energy is deposited in material through ionization/excitation
![Page 4: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/4.jpg)
Basics of Calorimetry II
Basic principle of calorimetry: deposited energy is proportional to incident energy
Calorimeter calibration translate detector response to incident energy
Great feature of showers for detector use: length is proportional to logE
![Page 5: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/5.jpg)
Electromagnetic showers
Created by incident photon and electron electrons emit
bremstrahlung photons undergo pair
production
Length of shower expressed in term of X
0
X0 depends on material
95% containment requires typically about 20X
0
![Page 6: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/6.jpg)
Created by incident charged pion, kaon, proton, etc Typical composition:
50% EM (e.g. ) 25% Visible non-EM energy 25% invisible energy (nuclear break-ups)
Requires longer containment (expressed in λ)
Hadronic showers
0
![Page 7: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/7.jpg)
Calorimeter detectors
Detector hardware must: Favor shower development Collect deposited energy
Can do both at the same time (e.g. BaBar/Belle crystal calorimeters)
Or have calorimeters with alternating passive and sensitive material
Example of electron shower with lead absorber:
![Page 8: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/8.jpg)
Sampling calorimetry (Ex.: CDF)
Scintillators (sensitive material) emit lights with passage of ionizing particles
Collect light deposited in sensitive material using wavelength shifter (WLS)
WLS → photomultipliers that convert light into electric signal
![Page 9: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/9.jpg)
CDF Calorimeters Segmentation
Calorimeter is segmented into towers that are read-out independently
Lead (iron) interspersed with scintillators for EM (HAD) calorimeters
Each central tower covers
Each tower has an EM calorimeter followed by an HAD calorimeter
151.0
![Page 10: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/10.jpg)
CDF Calorimeters
Three regions: central, wall and plug
Use “projective” geometry Designed to measure
electrons, photons, quarks, gluons, hadrons, neutrinos
Note: design of calorimeter performed with a simulation of the most important processes you plan to measure (ex.: Higgs at LHC)
![Page 11: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/11.jpg)
Construction
Go on and built the thing after it is designed! Many institutions in the world participate
![Page 12: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/12.jpg)
First calibration: test beam
Take one calorimeter “wedge”, send beam of particles with known energy
Obtain correspondence detector response → energy in GeV
A few towers only submitted to test beam Set absolute scale for all
towers Relative scale for other
towers obtained later
Wedge getting ready to receive beam:
![Page 13: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/13.jpg)
How does the test beam works (Ex.: plug calorimeter)
Performed at Fermilab meson beam facilities
Beams characteristics: Various types for EM and HAD
showers: electrons, pions, muons Various energy: 5-120 GeV (electrons),
5-220 GeV (pions)
Beams can be contaminated → bias the calibration constants E.g. use Cherenkov detector in front of
calorimeter to identify proton contamination in pion beam
WhyMuons?
![Page 14: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/14.jpg)
Calorimeter response linearity
Extract calibration constant for many energy point
Can test linearity of calorimeter
Can add “artificial” material in front of calorimeter to simulate tracker+magnet material
Send pions and electrons to hadronic calorimeter Why sending
electrons in hadron
calorimeter?
![Page 15: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/15.jpg)
Performance determined from test beam
• From RMS of tower response to same beam energy → measure calorimeter resolution
• Can test tower transverse uniformity (influences resolution)
• Stochastic term resolution:– EM:
– HAD: EEE /%80/
EEE /%14/
![Page 16: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/16.jpg)
Final detector assembly: getting ready for physics!
![Page 17: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/17.jpg)
The Tevatron
Proton-antiproton collisions at Most energetic collider in the
world Collisions every 0.4 μs Circumference of 6.3 km
TeVs 96.1
![Page 18: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/18.jpg)
The CDF Detector
CDF II: general purpose CDF II: general purpose solenoidal detectorsolenoidal detector
7 layers of silicon tracking – Vertexing, B-tagging
COT: drift chamber– coverage – Resolution:
Muon chambers– Proportional chamber
interspersed with absorber– Provide muon ID up-to
Calorimeters Central, wall, plug calorimeter
1|| %1.0/ 2 Tp p
T
5.1||
![Page 19: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/19.jpg)
Calibration when the detector is installed
Only a few towers saw test beam, how to calibrate the whole thing?? Test beam sets the absolute scale as a function energy Two solutions:
Hardware calibrations Physics calibration (using collider data)
These calibrations need to: Cross-check absolute scale (e.g. test beam not 100% realistic) Track detector response through time
Expected degradation of scintillator and PMT PMT sensitive to temperature
Uniform response through all towers
![Page 20: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/20.jpg)
Hardware calibration
Can use radioactive sources that have very well defined decay energy Cobalt 60 (2.8 MeV) Cesium 137 (1.2 MeV)
Source calibration can be performed between colliders run
Sources are movable and can expose one tower at a time
Check uniformity over all towers and over time
Sources are sensitive to both scintillator and PMT responses
![Page 21: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/21.jpg)
Laser calibrations
The lasers are connected directly to PMTs Skip scintillator/WLS steps
Used to uniformize PMTs response over towers and time
![Page 22: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/22.jpg)
Physics calibrations
Use real collider data For calibration, you have to have some “known” and
some “unknown” (the calorimeter response) Examples of “known” information:
Mass of a well-known particles Ex.: Z→ee (Z mass measured at LEP)
Energy deposited by muons over a given length Muon sample
Energy measured in tracker (assuming tracker in calibrated) Redundant to energy measured in calorimeter for electrons
![Page 23: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/23.jpg)
Example: Z boson mass
Z mass measured with great accuracy at LEP using beam energy
Background is very small for Z→ee
Sample is relatively small, but good enough
Z mass peak:
![Page 24: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/24.jpg)
Example: E/p of electrons
Used for relative scale over towers
Cannot be used in forward region (no tracker)
In plug: rely on sourcing and lasers
![Page 25: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/25.jpg)
Example: muons for HAD calorimeters
Muon calibrate detector response to ionizing energy
Use muon from J/ψ for identification (mass not used like Z boson)
Again, not used for PHA (rely on sourcing, laser)
![Page 26: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/26.jpg)
Physics with photons/electrons
Calorimeter calibration not the only issue
Electron/photon physics also rely on tracking
Removal of background E.g. remove pion
background by studying shower shape
Search for new physics: Z' candidate:
Precision measurement: W mass:
![Page 27: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/27.jpg)
What are jets?
Jets are a collimated group of particles that result from the fragmentation of quarks and gluons
They are measured as clusters in the calorimeter
momentum of cluster of towers is correlated with the momentum of the original quark and lepton
Why not using tracker
(has better resolution)?
![Page 28: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/28.jpg)
Phenomenology of jets
Quark/gluon produced from ppbar interaction
Fragmentation into hadrons Jets clustering algorithm:
Adds towers inside cone
Fraction of energy is out-of-cone
Underlying event contributes
![Page 29: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/29.jpg)
Jet versus calorimeter energy scale
Jets are complicated processes Previous calorimeter calibrations are not sufficient to get
calibrated jet energy More work needs to be done!!
Jet energy scale is crucial for many important measurements: Top quark mass (used to constrain Higgs boson) Jet cross-sections (comparison to QCD predictions)
Measurements often performed by comparing real data with simulations Need to get both physics and detector simulation right
![Page 30: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/30.jpg)
Relative energy scale
Jet energy measurement depend on location in detector
True even after all previous calibrations!
How come? Jets are wide Some regions of CDF
calorimeter are not instrumented
Relative energy scale: Use QCD dijet events Should have equal
transverse momentum
![Page 31: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/31.jpg)
Absolute energy scale
Response to single pion non-linear (in test beam)
However, jets are identified as one single objects
For a 50 GeV jet: calibration is not the same whether: One 50 GeV pion 10 times 5 GeV pions
Solution: Get the average energy scale
Simulate an “average” particles configuration inside jet
Use test beam information to get calibration factor for single particles
![Page 32: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/32.jpg)
Out-of-cone energy
Cone of fixed radius used to identify jets
Need to correct for fraction of energy out-of-cone (typically 15%)
This is mostly physics related How well is the physics
generator representing fragmentation?
![Page 33: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/33.jpg)
Underlying event energy
Proton/antiproton remnants splash energy in calorimeter
Spoils jet energy measurement
Depends on the number of ppbar interaction per event
Extracted from “minimum bias” events
Small effect: ~0.4 GeV per jet
![Page 34: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/34.jpg)
Final jet energy scale uncertainty
Estimate of jet energy scale uncertainty is important to estimate systematic uncertainties of measurements
Dominated by out-of-cone (low-pT) and absolute energy scale (high-pT)
Ranges from 10% to 3% energy uncertainties
![Page 35: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/35.jpg)
Example physics measurement: top quark mass
Top produced in pairs at Tevatron
Top decays to W boson and b-quark 100% of time in SM
Typical event selections: Well-identified electron(s) or
muon(s) Large missing ET Several reconstructed jets
identified in calorimeters
Note: 4 jets in final state!
![Page 36: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/36.jpg)
Identification of b-quark jets
• Complicated final state:
• Which jets come from which parton?
• Can identify b-quark jets using one characteristic: – Long b-quark lifetime
• Note: lots of semileptonc B-hadrons decay (involving neutrino)– Require special b-jets
calibration
jjjjltt
![Page 37: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/37.jpg)
Top mass reconstruction
Event-by-event kinematic fitter (assumes event is ttbar)
Attempts all jet-parton assignments Assign b-tag jets to b-
quarks
The one most consistent with ttbar hypothesis is kept
More correct combinations with b-tags!
![Page 38: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/38.jpg)
The strategy
Construct reconstructed top mass distributions for many true top mass So-called “templates”
Compare distribution reconstructed in data with templates Using likelihood fit
Account for background contamination Dominated by W+jets
production
![Page 39: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/39.jpg)
The measurement (spring 2005)
Using 138 candidate ttbar events, fit yields:
Mtop= 173.2 +2.9/-2.8 (stat.) +/- 3.4 (syst.) GeV/c2
By shifting by JES uncertainty defined before: Mtop changes by 3.1 GeV/c !
JES uncertainty limiting factor for Mtop measurement
![Page 40: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/40.jpg)
Improvement: W→jj calibration
Inside ttbar events, invariant mass of two jets from W boson decay should equal MW
Can use W→jj decays to further constraint JES
Use same data for measurement and calibration… cheating?? No: Mjj (almost) independent
of Mtop
Remaining correlations are accounted for
![Page 41: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/41.jpg)
The measurement(adding W→jj information)
Using same dataset as previously:
Mtop= 173.5 +2.7/-2.6 (stat.) +/- 2.8 (syst.) GeV/c2
Total Mtop uncertainty improved by 10%
JES uncertainty decreased by 20%
Good prospect for future
![Page 42: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/42.jpg)
Impact of Mtop measurement
Mtop, MW connected to Higgs boson mass through radiative corrections
MH< 186 GeV/c2 @ 95%C.L.
Can constrain mass of supersymmetric particles
![Page 43: Calorimeter Calibration and Jet Energy Scale Jean-Francois Arguin November 28 th, 2005 Physics 252B, UC Davis](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d625503460f94a43c38/html5/thumbnails/43.jpg)
Conclusion
Detector calibration needed to translate detector response in energy
Various techniques used for calorimetry: Test beam Radioactive sources Lasers Collider data
Calorimeter can be used to measure: Electrons, photons, jets, missing ET
Good calorimeter and jet calibration needed for measurements like top quark mass