DASGIP® PHPO Sensor ModulesUser Manual
User Manual pHpO
Copyright © 2003 DASGIP AG, Juelich, Germany 1
Publisher DASGIP Information and Process Technology GmbH
Rudolf-Schulten-Straße 5
52428 Juelich
Telephone: +49 (0)2461 / 980-0
Fax: +49 (0)2461 / 980-100
Version
http://www.dasgip.de email:
Decai^an 2014, 1...1
Copyright Copyright © 2003 DASGIP GmbH, Juelich
Re-printing and copying – even in parts – only allowed with prior permission of DASGIP GmbH.
Trademarks fedbatch-pro®/cellferm-pro® are registered trademarks of the DASGIP GmbH, Juelich.
MS-Excel®, MS-Windows® are registered trademarks of the Microsoft Corporation.
Flow-NG is a trademark of the DASGIP GmbH, Juelich.
*Figures and descriptions in this document can differ fromthe delivered product.
User Manual pHpO Table of Contents
Copyright © 2003 DASGIP AG, Jülich, Germany 2
Table of Contents
Preamble ............................................................................................... 4
1 Notes .............................................................................................. 5
1.1 Safety Notes............................................................................. 6
1.2 General Notes .......................................................................... 8
2 Introduction.................................................................................... 9
2.1 General..................................................................................... 9
2.2 Operation with DASGIP EasyAccess ....................................... 10
3 Connection Elements .................................................................... 11
3.1 The Front of the Device.......................................................... 11
3.2 The Back of the Device........................................................... 13
3.2.1 Communication Connections......................................................14
4 Commissioning ............................................................................. 16
4.1 Transport ................................................................................ 16
4.2 What’s Included...................................................................... 16
4.3 Set-up ..................................................................................... 17
4.4 Connections............................................................................ 18
4.4.1 Electrical Power Supply ..............................................................18
4.4.2 Sensors......................................................................................19
4.4.3 Communication .........................................................................19
4.5 Operation ............................................................................... 20
4.5.1 Calibrations ...............................................................................20
4.5.2 pO2 Sensors ...............................................................................20
5 Maintenance and Trouble-shooting ............................................. 23
5.1 Maintenance and Cleaning..................................................... 23
5.2 Fault Repair and FAQ’s........................................................... 24
6 Annex............................................................................................ 26
6.1 Technical Data........................................................................ 26
6.2 Connector Pin Assignments.................................................... 26
6.2.1 General .....................................................................................26
6.2.2 pH Socket ..................................................................................29
6.2.3 pO2 Socket.................................................................................29
6.2.4 Pt Socket ...................................................................................29
6.2.5 AUX Socket ...............................................................................30
User Manual pHpO Table of Contents
Copyright © 2003 DASGIP AG, Jülich, Germany 3
6.3 Accessories............................................................................. 31
Index ................................................................................................... 32
User Manual pHpO Preamble
Copyright © 2003 DASGIP AG, Jülich, Germany 4
DASGIP AG, located in Jülich, Germany, founded in 1991 develops and distrib-utes parallel Systems for microbiology and cell culture.
Our team consists of specialists for hardware, software, engineering, mechanical engineering, measurement and control technique, microbiology and cell culture. We are continually working at new hardware and software solutions to actively support our customers with the developing and optimization or their biotechno-logical processes.
Both our developing experience and our customers’ tips and suggestions have contributed to this manual. Our goal was to make this manual as practical and user friendly as possible to make it a true guide and reference book that can help you in every situation.
Our mission is to help you in every possible way with cellferm-pro® and the product option MetabolicActivity to allow you to achieve the best possible results working with DASGIP hardware and software.
Preamble
User Manual pHpO 1 Notes
Copyright © 2003 DASGIP AG, Jülich, Germany 5
1 Notes To avoid improper use and to keep your warranty, please read these instructions carefully prior to starting to use the sensor module. Please heed to the following instructions and safety notes to avoid damages or injuries to people, effects or the equipment.
Please do not hesitate to contact us, if you have any questions or queries.
In this manual the following symbols are used:
This symbol warns the personnel concerned of possible dangers. Please read the corresponding note thoroughly to avoid injuries to people.
Heed to the notes labeled with this symbol to avoid accidents due to electric shock or damage to appliances due to overvoltage.
Paragraphs labeled with this symbol contain important information to avoid pos-sible damages to the sensor module.
This symbol highlights valuable tips and suggestions to ease the work with the sensor module and points out possible sources of error.
Appropriation and use of the system without heeding the instruc-tions and safety notes as well as improper use render warranty null and void for the system failure and damage or injuries to persons or effects.
+49 (0) 2461 980 0 [email protected]
Symbols Used
User Manual pHpO 1 Notes
Copyright © 2003 DASGIP AG, Jülich, Germany 6
The following safety notes will help to avoid possible dangers when using the pHpO Module and give advice regarding the correct behavior to avoid dangerous situations. Special dangers when working with the pHpO Module are commented on in particular in the corresponding chapters of this manual.
Operation only by qualified personnel! The pHpO Module has been developed for the use in a laboratory. We assume, that the user is familiar with the necessary safety instructions as well as the op-erating of electrical equipment and will act accordingly.
Not accredited for the use in medical diagnostics or therapy! The pHpO Module was not developed for the use in medical diagnostics or ther-apy and is not to be used in such applications. It is not a medical device.
Do not open device! Danger of electric shock! The housing of the device is not to be opened by the user. In the inside of the housing there are no parts that have to be exchanged or adjusted by the user.
Do not use the device with damaged power cable or power plug!Danger of electrical shock! Replace damaged parts only with original parts of the manufacturer.
Service and Repair Work Service and repair work, beyond the maintenance work described in the user manual are only to be performed by DASGIP’s service technicians or technicians authorized by DASGIP.
Heed to limiting connection values! Please heed to the connection values listed in this manual regarding the connec-tion of the sensors.
1.1 Safety Notes
User Manual pHpO 1 Notes
Copyright © 2003 DASGIP AG, Jülich, Germany 7
Do not pull plug during operation! Do not pull the power plug while the device in running as this could result in the pHpO Module being damaged.
User Manual pHpO 1 Notes
Copyright © 2003 DASGIP AG, Jülich, Germany 8
Take note of user manuals and documentations of other manufactur-ers! Please heed to the instructions and safety notes of all devices included in the deliverable, especially the sensor manufacturer’s instructions.
Ambient Conditions During set-up and operation of the pHpO Module the necessary ambience condi-tions have always to be met (compare Chapter 4.3). If the device is to be operated at high ambience temperature and/or high humid-ity, the suitability of the device has to be discussed with DASGIP first.
The device is not protected against spray water!
Electromagnetic Radiation Strong electromagnetic radiation, e.g. from cell phones, might affect the meas-urement signals.
Reconsignment for Repair or Return You have to notify us of a reconsignment. If you fail to do so we will refuse acceptance. Furthermore, you have to declare, that the consignment is free from any decontaminations. Please read and heed to the information regarding the Decontamination Declaration at the end of this manual and complete the Decontamination Declaration carefully. Please insure, that the Decontamination Declaration reaches DASGIP 10 days prior to the delivery of the consignment.
For the reconsignment the original packaging should be used. Please heed to the instruction sheet How to Pack DASGIP-Modules at the end of the user manual!
If the original packaging is not available, the client is responsible for proper and suitable packaging. Please do not hesitate to inquire at DASGIP to have original packaging material sent to you.
1.2 General Notes
User Manual pHpO 2 Introduction
Copyright © 2003 DASGIP AG, Jülich, Germany 9
2 Introduction
This manual is for all the product variants of the parallel sensor module pHpO. The variants only differ in the number and combination of the measurement channels (e.g. pH4pO4, pH8pO8, etc.). The manual describes the installation and appropriation of the module as a stand-alone device or as part of a DASGIP par-allel cultivation system cellferm-pro® or fedbatch-pro®. Furthermore, it contains important information and Tipps regarding the work with the sensor module.
The pHpO Module support standard pH and pO2 sensors. The temperature meas-urement is done using the NTC senors integrated in the pO2 sensors. Separate inlets for Pt100 handheld sensors facilitate an automatic temperature compensa-tion during measurement and calibration. The individual measurement channels of the device are separated galvanicly from each other; they cannot affect or damage each other.
With the sensor module pHpO you can monitor online e.g. the pH and pO2 val-ues in several parallel reactors or culture vessels. Different combinations of pH ,pO2 and temperature measurement offer various possibilities of the parallel monitoring of process values.
All sensor module variants can be operated as stand-alone devices or connected with a PC. For the control and monitoring of DASGIP devices the software DAS-GIP EasyAccess is availabe. In addition, the devices can be integrated into a DASGIP parallel cultivation system (e.g. cellferm-pro® or fedbatch-pro®).
2.1 General
The User Manual
The Device
Applications
Integration into a DASGIP System
User Manual pHpO 2 Introduction
Copyright © 2003 DASGIP AG, Jülich, Germany 10
Using the Software DASGIP EasyAccess on the CD included in the delivery, the DASGIP pHpO Sensor Modules can be controlled as stand-alone devices via a PC.
DASGIP EasyAccess is user interface based on MS-Internet Explorer, facilitating the control and monitoring of all important functions of the DASGIP devices. Via a simple web-page the set point values and process values can be accessed. DASGIP EasyAccess also offers effective support regarding process data logging.
Figure 1: Example of a pO2 calibration using DASGIP EasyAccess
Operating the web-pages is very easy. Extended training is not needed. DAS-GIP EasyAccess offers effective support in everyday laboratory tasks.
Detailed instructions regarding the work with DASGIP EasyAccess can be found in the DASGIP EasyAccess User Manual. The integration of the module into other applications using the DASGIP EasyAccess Driver Software is also described and examples are given.
2.2 Operation with DASGIP EasyAccess
User Manual DAS-GIP EasyAccess
User Manual pHpO 3 Connection Elements
Copyright © 2003 DASGIP AG, Jülich, Germany 11
3 Connection Elements
On the front of the device, an operation indicator lamp and the connection sock-ets for the different sensor types.
Figure 2: Front of Device (Example pH4pO4)
The following table lists the information regarding all elements on the front of the device:
No. Name Function
1 Status indicator light (Power Supply)
Status indicator light for the power supply of the device:
Gray = OFF Green = ON
2 pH Sockets Input range: -472mV - 472mV Measurement range: pH0 - pH14 ( for a sensor temperature of max. 67°C )
3 pO2 Sockets pO2 measurement:
Input range: 0 - 400nA Measurement range: 0 - 250% DO
3.1 The Front of the Device
User Manual pHpO 3 Connection Elements
Copyright © 2003 DASGIP AG, Jülich, Germany 12
No. Name Function
Temperature measurement:
Input range: infinite - 1684 Ohm Measurement range: -273°C - 104°C
(0K - 377K)
4 Pt Sockets Input range: 90.3Ohm – 144.7Ohm Measurement Range: -25°C - 116°C
5 AUX Sockets Input range: 0 - 10,605V and 0 – 40.95mA
Measurement Range: 0 - 10V and 0 - 20mA
6 Stowage Com-partment (hinged)
To lay cables and tubes for the operation with DAS-GIP fedbatch-pro®/ cellferm-pro® Systems (DASGIP Parallel Cultivation System).
Please insure that no cables or tubes are caught in the hinge of the stowage flap!
Please do not use sensors which deliver signals outside the ranges given in the table.
The connector assignment of all sensor sockets are given in the annex.
User Manual pHpO 3 Connection Elements
Copyright © 2003 DASGIP AG, Jülich, Germany 13
At the back of the device, the power supply connector unit, communication con-nections, serial number and various venting slots can be found. Further informa-tion regarding the individual elements are listed in the Table further down.
Figure 3: Back of a pHpO Module
No. Name Function
1 Power Switch Power switch of the device. On and off positions are labeled according to standard: I = ON O = OFF
2 Fuse Holder Fuse holder for two micro-fuses 20mm x 5mm. Value and characteristics of the fuse are given on the back panel of the device.
3 Power Connection IEC connector to connect a power cable. Power sup-ply specifications are given on the back panel of the device.
4 Serial Number The serial number of the device should be stated in every enquiry with the manufacturer.
5 Communication Connections
RS232 and RS485 communication connections are explained in Chapter 3.2.1.
6 Venting Slots The venting slots are needed for heat dissipation and are not to be covered (Chapter 4.3).
3.2 The Back of the Device
User Manual pHpO 3 Connection Elements
Copyright © 2003 DASGIP AG, Jülich, Germany 14
At the back of the device are several communication connections (compare Figure 4).
Figure 4: Communication connections
In the following table the individual elements of the communication connections are listed and explained:
No. Name Function
1 "DTP in" DTP input
SUB D9F/RS232 port for the DTP communication with the process computer. The DTP input will be connected directly to the process computer or the „DTP out“ of another DTP capable de-vice in a DTP network (compare Figure 5).
2 "DTP out" DTP output
SUB D9M/RS232 port to connect additional DTP capable device to a DTP network (compare Figure 5).
3 "AUX" SUB D9M/RS232 port to connect additional devices and the floating alarm contacts.
4 "State" DTP state indicator lamp:
Gray = Connections DTP-in/out activated Green = Connections RS485 activated
3.2.1 Communication Connections
User Manual pHpO 3 Connection Elements
Copyright © 2003 DASGIP AG, Jülich, Germany 15
No. Name Function
5 "Rx/Tx" DTP communication indication lamp
Gray = no signal Green = receiving Orange = sending Red = communication error
6 "RS485" RJ45 ports for a RS485 network; floating.
Please do not confuse the similarly built "RS485" connections with the Ethernet connections.
Additional information regarding the allocation of the individual connections can be found in the corresponding Chapter in the Annex.
User Manual pHpO 4 Commissioning
Copyright © 2003 DASGIP AG, Jülich, Germany 16
4 Commissioning
Unpacking
The sensor Module pHpO device is delivered completely wrapped and should be taken out of the box following the Packaging Instructions for DASGIP Modules at the end of this manual. Whenever possible, the device should be unpacked at the location of installation.
Onword Transport
For the transport to another location or shipping for servicing, the original pack-aging should be used. This is the only way to insure the device being sufficiently protected. Check the Re-Consignment Information and the Packaging Instruc-tions for DASGIP Modules for additional advice (at the end of this manual).
Included in the deliverable of the sensor module are – depending on the device variant – the following parts:
Name Qty. Comment
Sensor Module pHpO 1 Several possible variants
Power connection cable 2.5m 1
User Manual pHpO
1
DTP communication cable 2m 1 optional; for the operation as part of a DASGIP Parallel Cultivation System
pH sensors 1-16 optional, depending on the device variant
4.1 Transport
4.2 What’s Included
User Manual pHpO 4 Commissioning
Copyright © 2003 DASGIP AG, Jülich, Germany 17
Name Qty. Comment
pH sensor cables 1-16 optional, depending on the device variant
pO2 sensors 1-8 optional, depending on the device variant
pO2 sensor cables 1-8 optional, depending on the device variant
Pt100 handheld sensor 1-2 optional, depending on the device variant
To insure undisrupted operation of the pHpO Module you should respect the limits regarding the ambient conditions listed in the Table further down and the following notes regarding the set-up location.
The set-up location should be level and appropriate for the weight of the device (compare Chapter 6.1).
DASGIP Modules can be stacked. A maximum number of three DASGIP Modules are to be stacked, otherwise the stack becomes unstable. Please consider the resulting total weight.
The ambience of the pHpO Module has to be clean, dry and largely dust free. The limits for ambience conditions have to be complied with during transport, storage and operation.
When setting up the device it has to be insured, that sufficient convection can develop both on the front and back of the device, otherwise the developing heat cannot escape.
Do not cover the venting slots!
4.3 Set-up
Set-Up Location
Ambience
Ventilation
User Manual pHpO 4 Commissioning
Copyright © 2003 DASGIP AG, Jülich, Germany 18
Ambient Conditions
Property Value
Temperature range during opera-tion
5-40°C
Maximum relative humidity dur-ing operation
80%rH, non condensing
Maximum height above sea level during operation
2000m above sea level (6000ft ASL)
Temperature range during stor-age
0-70°C
Maximum relative humidity dur-ing storage
90%rH, non condensing
When setting up a complete cultivation system please note also the setting up instructions in the cellferm-pro® or fedbatch-pro® User Manual and the set-up conditions for DASGIP Systems.
Make sure that the power switch is in position “0” (=OFF). Connect the power socket with the power supply socket using the power cable included in the deliv-ery.
Do not use the device with damaged power cable or power plug!Danger of electrical shock!
Cultivation System
4.4 Connections
4.4.1 Electrical Power Supply
User Manual pHpO 4 Commissioning
Copyright © 2003 DASGIP AG, Jülich, Germany 19
The sensors have to be connected to the corresponding sockets via the sensor cables and screwed tight.
Please heed to the input connection values regarding the connection of the sen-sors (Chapter 3.1).
The pO2 sensors have to be polarized for 2 – 6h prior to operation.
In the case of the device being operated in a DTP network, connect the port “DTP in” either with a serial interface of the process control computer (usually COM1) or with the port “DTP out” of another communication participant.
Figure 5: Example of a typical DTP network
If two devices of the same type are to be operated in a DTP network, the ad-dresses have to be adapted e.g. according to the following table.
Changing the DTP address can be accomplished either using the configuration menu of the device or the configuration software DASGIP DTP Insight.
The corresponding device addresses for the operation of the sensor module as a part of a fedbatch-pro® or cellferm-pro® system can be taken from the following Table:
4.4.2 Sensors
4.4.3 Communication
DTP Network (optional)
DTP Address (optional)
User Manual pHpO 4 Commissioning
Copyright © 2003 DASGIP AG, Jülich, Germany 20
Channel MX4/4 pHpO MP8 Line1
MP8 Line2
MP8 Line3
MP8 Line4
TC SC
1-4 1 5 9 11 13 15 17 21
5-8 2 18
9-12 3 10 12 14 16 19 22
13-16 4 20
Please insure, that the calibrated sensors are connected to their corresponding sockets. If the sockets are interchanged, the calibration has to be run again for the changed slot.
For autoclaving the sensors can be disconnected from the sensor module. Please insure the correct allocation of the sensors to their slots.
Polarisation
Prior to operation, the pO2 sensors need a polarization time of 2 – 6h.
Connect the sensor to the corresponding socket and switch on the sensor mod-ule. Polarized sensors may be disconnected briefly from the sensor module. After a disconnection time an equally long polarization time should be kept.
Please heed to the sensor manufacturer’s notes.
4.5 Operation
4.5.1 Calibrations
4.5.2 pO2 Sensors
User Manual pHpO 4 Commissioning
Copyright © 2003 DASGIP AG, Jülich, Germany 21
Inspection of the pO2 Sensors
Prior to every calibration and using the pO2 sensors, you have to thoroughly in-spect the sensors:
1. Inspect the sensor membrane for visible signs of damages or soiling.
2. Clean the soiled membranes by wiping them with a soft, damp tissue.
3. Optional:Renew the electrolyte in regular intervals (compare instructions below).
4. Additionally, check, whether all connections are free of crystallizations, andcheck cables for bends, embrittlement and damages.
Please heed also to the instructions given in the pO2 sensor’s operating manual.
Electrolyte Change
Depending on mode and duration of use, the electrolyte has to be exchanged in regular intervals. At least every 6 months. In the following paragraph the electro-lyte exchange is described for standard Mettler Toledo sensors.
Attention! The 02 electrolyte is strongly alkaline. Please wear appropriate protective clothing – especially gloves. The liquid will certainly escape the membrane body and come into contact with your hands when opening the membrane body. Avoid contact with skin and mucosas, especially the eyes. Rinse contaminated areas thoroughly with water.
1. Unscrew cap sleeve from sensor body.
2. Carefully pull sleeve off sensor.
3. Remove membrane body from cap sleeve.
4. If body is tight fitted, eject by pushing it with flat finger tip.
5. Rinse interior body with DI water and carefully dap it dry with a paper tis-sue.
6. Examine O-rings and trapezoidal seal visually for mechanical defects.
7. Half-fill membrane body with O2 electrolyte.
8. Remove air bubbles by carefully tapping membrane body.
9. Slip membrane body over interior body while holding sensor in vertical posi-tion tip down.
10. Remove excess electrolyte with a paper tissue.
11. Carefully slip the clean and dry cap sleeve over fitted membrane body andscrew tight.
User Manual pHpO 4 Commissioning
Copyright © 2003 DASGIP AG, Jülich, Germany 22
After each exchange of electrolyte or membrane body, sensor has to be re-polarized and recalibrated due to change of setpoint and slope.
The sensor has to completely polarize (6-8h). Afterwards, carry out a calibration!
Please heed also to the manufacturer’s notes in the corresponding documenta-tion for the pO2 sensors.
User Manual pHpO 5 Maintenance and Trouble-shooting
Copyright © 2003 DASGIP AG, Jülich, Germany 23
5 Maintenance and Trouble-shooting
Maintenance Intervals
The devices are tested by the manufacturer and are for the most part mainte-nance-free. To insure a fault-free operation we recommend to send the Module to us for a calibration service every one to two years.
In regular intervals, the internal buffer battery of the device should be renewed. Depending on the number of times the module was used this should be done every one to three years.
Please do not hesitate to contact the DASGIP service if you experience any errors or faults you cannot repair using the information given in the following Chapter or the user manual.
Cleaning
Superficial soiling can be removed using a damp cloth. Depending on the type of soiling, a mild detergent can be used. Remove residual acid or base remnants with clear water.
Please do not use scoring agents or corrosive detergents!
The device is not protected against spray!
5.1 Maintenance and Cleaning
[email protected] +49(0)2461 980-0
User Manual pHpO 5 Maintenance and Trouble-shooting
Copyright © 2003 DASGIP AG, Jülich, Germany 24
1 All pH sensors show pH7.
Cause
No voltage is measured.
Fix
Is the sensor module switched on?
2 One sensor shows pH7.
Cause
The sensor is probably short circuited or defect.
Fix
Check all connections and – if needed – connect another sen-sor.
Check connections for dampness or crystallized media compo-nents.
6 pH sensors show implausible values.
Cause
Sensors are not correctly connected.
The diaphragm is soiled.
The sensor is defect.
Fix
Check connections and cables.
Clean sensor.
Replace sensor.
7 pH sensors show slow response time.
Cause
Sensor is electrostatically charged.
Diaphragm is soiled.
Fix
Wait, avoid rubbing the sensor shaft.
Clean sensor.
5.2 Fault Repair and FAQ’s
User Manual pHpO 5 Maintenance and Trouble-shooting
Copyright © 2003 DASGIP AG, Jülich, Germany 25
8 How can the device be integrated into other systems?
Answer
Via the digital communication interface, all parameters, proc-ess values and set point values can be accessed. All DASGIP devices use the communication protocol DTP. For the easy integration of the device into Windows-PC appli-cations, a library is provided by DASGIP EasyAccess. The cor-responding driver can be found on the CD that was part of the delivery. An integration into other control or operating systems can be implementing using the protocol. Please inquire with DASGIP for the protocol specifications. An integration into other control or system software can be done by implementing the protocol. Please contact DASGIP for the protocol specifications. Due to the multitude of parame-ters, set point and process values in the device, no analogue interfaces are provided. The AUX port provides floating relay contacts for the indication of cumulate alarms.
User Manual pHpO 6 Annex
Copyright © 2003 DASGIP AG, Jülich, Germany 26
6 Annex
Property Value
Power Supply 110-230V, 50/60 Hz
Power Consumption max. 60VA; Type 35 VA@230V
Protection Class I
Overvoltage Category 2
Weight approx. 8 kg
Dimensions (WxDxH) 300 x 320 x 190 mm
External Digital Interface RS232 / RS485
For a better understanding of the Tables further down, the following Figure shows the communication ports at the back of the device.
6.1 Technical Data
6.2 Connector Pin Assignments
6.2.1 General
User Manual pHpO 6 Annex
Copyright © 2003 DASGIP AG, Jülich, Germany 27
Figure 6: Communication ports
DTP in/out, AUX, Alarm Contacts
PIN DTP in DTP out AUX
1 Alarm
2 TXD TXD RXD
3 RXD RXD TXD
4
5 GND GND GND
6 Alarm(normally closed)
7 CTS CTS
8 RTS RTS
9 Alarm(normally open)
The lines RXD, TXD,GND, RTS, CTS of the ports DTP-in and DTP-out are directly interconnected.
The maximal permissible load of the relay contacts is 30 VDC/1A.
In undisturbed state, the pins 1 and 6 are interconnected (break contact/ nor-mally closed). In case of an error, the pins 1 and 9 are connecting (working con-tact/ normally open).
User Manual pHpO 6 Annex
Copyright © 2003 DASGIP AG, Jülich, Germany 28
RS 485
PIN RS485
1
2
3
4 B/D1
5 A/D0
6
7 +5V emergency power
8 Common
The RS 485 communication ports are isolated galvanically from the remaining device electronics.
The connector pin assignment is according to the specifications "Modbus Over Serial Line Specification and Implementation Guide Version 1.0", Chapter 3.5.1. The lines B/D1, A/DO, Common of the two RJ45 port are directly interconnected.
The ends of the RS485 network have to be terminated using an appropriate re-sistance / condenser serial connection. Typical values are 1nF/16V and 120 Ohm.
The +5V emergency power at Pin 7 can be used for the polarization of a RS485 network. For polarization, B/D1 is connected to +5V and A/D0 to Common using in each case a resistance of 560 Ohm. The inner resistance of the voltage source +5V is 100 Ohm.
User Manual pHpO 6 Annex
Copyright © 2003 DASGIP AG, Jülich, Germany 29
PIN assignment
1 pH+
2
3 GND
Figure 7: Connector pin assignments of the pH socket
PIN assignment
1 NTC
2 UVisiFerm
3 GND
4
5 pO2
6 Upol
Figure 8: Connector pin assignments of the pO2 socket
PIN assignment
1 Pt100
2
3 GND2
4 GND1
Figure 9: Connector pin assignments of the Pt socket
6.2.2 pH Socket
6.2.3 pO2 Socket
6.2.4 Pt Socket
1 3
2
1
2 4
5
6
1
3 2
4
3
User Manual pHpO 6 Annex
Copyright © 2003 DASGIP AG, Jülich, Germany 30
PIN assignment
1 In 0-10V
2 Out +24V,max. 25mA
3 GND
4 In 0-20mA
5 GND
Figure 10: Connector pin assignments of the AUX socket
6.2.5 AUX Socket
1
2 4
5
3
User Manual pHpO 6 Annex
Copyright © 2003 DASGIP AG, Jülich, Germany 31
For the pHpO Module the following accessories are available from DASGIP:
Item Order No.
L120mm 78108031
L225mm 78108032 pO2 Sensor VisiFerm D12mm
autoclavable optical sensor L325mm 78108033
DASGIP pO2 cable with plug Type VP8 for VisiFerm sensors L3m 78522042
L120mm 78108018
L225mm 78108026 pO2 Sensor MTI D12mm, autoclavable
L325mm 78108022
pO2 Membrane Kit MTI 4 membrane bodies, O-rings, 25mL electrolyte 78108003
Electrolyte for pO2 Sensor MTI 25mL 78108015
L120mm 78108012
L150mm 78108019
L220mm 78108020 pO2 Sensor BJC D12mm, autoclavable
320mm 78108021
1 membrane bodies, O-rings, 25mL electrolyte 78108024
4 membrane bodies, O- rings, 25mL electrolyte 78108017 pO2 Membrane Kit BJC
25 membrane bodies 78108029
Electrolyte for pO2 Sensor BJC 250mL 78108027
DASGIP pO2 cable with plug Type 82 L3m 78522040
L120mm 78103207
L225mm 78103220 pH Sensor MTI D12mm, autoclavable
L325mm 78103209
DASGIP pH/Redox cable with AK9 plug L3m 78522020
6.3 Accessories
User Manual pHpO Index
Copyright © 2003 DASGIP AG, Jülich, Germany 32
Index
A
Accessories................................. 31
Ambience ................................... 17
Ambient Conditions .................... 18
AUX ........................................... 14
AUX Sockets............................... 12
B
Back of the Device ...................... 13
C
Calibrations................................ 20
Cleaning..................................... 23
Commissioning........................... 16
Communication
Connections............................ 19
Communication Connections 13, 14
Connection Elements .................. 11
Back of the Device .................. 13
Front of the Device.................. 11
Connections
electric.................................... 18
Connections ............................... 11
Connections
Communication....................... 19
Connector Pin Assignments......... 26
D
Dangers........................................ 6
DASGIP EasyAccess .................... 10
Decontamination Declaration........ 8
Device AddressesCompare DTP Addresses
DTP Address ............................... 19
DTP in ........................................ 14
DTP Network .............................. 19
DTP out ......................................14
E
EasyAccess .................................10
Electrical Power Supply ...............18
Electrolyte Change ......................21
F
Fault Repair ................................24
Front of the Device......................11
Fuse Holder ................................13
M
Maintenance...............................23
Measurement Range...................11
N
Notes ...........................................5
General .....................................8
Safety........................................6
P
pH Sockets..................................11
pO2 Sensors
Electrolyte Change...................21
Inspection ...............................21
pO2 Sockets ................................11
Polarization ................................20
Power Connection.......................13
Power Switch..............................13
Pt-Sockets ..................................12
R
Reconsignment .............................8
Rx/Tx ..........................................15
S
Safety Notes .................................6
Sensors
User Manual pHpO Index
Copyright © 2003 DASGIP AG, Jülich, Germany 33
Connect .................................. 19
Serial Number............................. 13
Set-up ........................................ 17
State .......................................... 14
Symbols........................................ 5
T
Technical Data............................ 26
Transport....................................16
Troubleshooting..........................23
V
Ventilation..................................17
W
What’s Included..........................16
���������AB��C�A�DE�F��AE��
���������DE�F��AE��������E����������EE�C��F���������������AB��CDE�CF��D������������������AB�������DF��E�� � �
�D��������� !"��#�#������$�%�������� !"��#�!##������&&&'��()�*'�D� �
�
�+D�F�A�(C��DF,�
-E��F�DF�C��*F�.��D�����(C��E��D���A�DEC�B�E���E)�������F�FDC�FED��)���(���F�FD*��F"���ECDE�EAD�+/�-0��(1(�������F�(��D�BD�*��E���E��F��C��E'��������E���������������A������BA�����DE�������A����������A�F���������� � DE�� ��������A����E��� !�A� ��"#E$%����&� AD���������C&� ����AB�� A�DE�F��AE��
��E�����������EF������������E��BEA�����A�AE����������$��BD������&�E)���F�(�A�E�2D�����D����C��E��(�)ED��D�DACF�E�A������F�����A�E�*F�EC�CBD���E�������CBD����C�"�(�)E�CBD����E�����'�'A��E���������DE�F�&��������A�F����(A��E�����������$��
o ���*F�CDAC���F�D�*���DD(,��((�A��CD(��E��CBD�DE.�F�E�DEC�����FDC�FED��D3��*�DEC���(C�2D��FDD����(�2(C�EAD(�CB�C��FD�2���B�4�F���(��F��E��(�2(C�EAD(�CB�C�����FD3��FD�(*DA����B�E���E)'��BDFD��FD,�*�D�(D��CCDE��C��CBD�5�D�E�E)��E��+DA�EC���E�C��E�$�F���E��(DE��CBD�A��*�DCD����F��.�����%��F�D������(�&D����(��CC�AB�CBD��F�)�E���C��CBD���C(��D����CBD�*�A1�)D'��
6�����E��CBD�5�D�E�E)��E��+DA�EC���E�C��E�$�F���E�CBD������&�E)�*�)D('�
o $�F�D�(�DF�B�E���E)�&D�&�������1D�C���(1�����C��*F�.��D��(�&�CB�(��D��E��F��C��E��2��C�CBD��D.�AD��E��CBD�FD3�D(CD���AC��E('��BDFD��FD,�*�D�(D��������C�CBD�DF.�AD��E���D*��F�$�F�'��BDE�*FD�(DE���C�.�����%��F�D������E���CC�AB��C�C��CBD�*�A1�)D���1D�CBD�5�D�E�E)��E��+DA�EC���E�C��E�$�F�'��
�BD�DF.�AD��E���D*��F�$�F���(���(�����D���E�CBD������&�E)�*�)D('�
o ���*F�CDAC�)���(��F����E������)D���F�E)�CF�E(*�FC�C��E�*�D�(D��(D��E���D3��CD�*�A1�)�E)���F������CD�(�(B�**D��C��+/�-0'�0�D�(D��CCDE��C��CBD���AC,�CB�C�(��D�)���(��FD�.DF��(DE(�C�.D,�D')'�(DE(�F(��F�)��((&�FD,��E��EDD��(*DA����*F�CDAC��E���1D�*���D��*�A1�)�E)'�/�(��(��D�*�FC(,�D')'���1D�+/�-0������D(,��FD�.DF��BD�.�'��B�(,�CBD��FD3��FD���*���D��2�C���(��.DF��(CF�E)�*�A1�)�E)�CB�C���%D(�CBD��D.�AD�.DF��&D��'����DE(�FD�(�.D�CF�E(*�FC�+/�-0�*F�.��D(�(*DA����*�A1�)�E)���CDF������F�+/�-0������D(��E��.D((D�(��E��CBD�+/�-0�7��2��A1'�-��CBDFD��(�E����D3��CD�*�A1�)�E)���CDF�����.����2�D��C����F�(�CD,�*�D�(D�A�EC�AC�+/�-0�(DF.�AD'��
0�D�(D���E���E�CBD������&�E)��E(CF�AC��E(��2��C�CBD��(�)D����+/�-0�(*DA����*�A1�)�E)'�
o 0�D�(D�(B�*�*�FC(�C���
8�F�*D,�/�F�A���E��/(���� /�DF�A���
+/�-0�-E��F��C��E��E��0F�AD((� Eppendorf North America��DABE���)����2:�0��DF.�AD� DF.�AD�-��9�E)DE2F��AB�;� !�!�<D��F����+F�.D,�=E�C�7�����������AB� BFD&(2�F�,�</�#!�����DF��E�� =E�CD��C�CD(����/�DF�A��
o $�F��E��3�D(C��E��2��C�CBD�FDC�FE����)���(�DF.�AD��D����(�B�**��C��)�.D������E��(�**�FC'�
����>##�;�>��>���� �!��##��;!��� ��(DF.�AD?��()�*'�D� (DF.�AD'��DF�A�?��()�*'A���
������������A�BC�D�E���D�C��FC�E�
A���������C�E�D�C�������CB������B��C�C����E���������������AB��CDE�CF��D������������������AB�������DF��E�� � �
�D��������� !"��#�#������$�%�������� !"��#�!##������&&&'��()�*'�D� �
+�D�(D��
����
��
�CBD���F��,�����%��F�D�
����C��-.�
/+*@kk`i_jma Ijmoc <h`md^\��E���CC�AB�CBD��F�)�E���C
�D�
���CBD�*�A4�)D5�
�
$F���3��*�E������ �
.��FD((������ � �
� � ���� � �
+B�ED�� ������� ����������������������������������������������������$�%����� �
6*DF�C�F����� � �
�A�D� ����!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!��
"�C����D��#�$���%&�'( ����7777777777777777777��
��B)�D�*C+ ���� !!!!!!!!!!!!!!!!!!!!!!!!!!!!��
�
�������*�E����,���C���C�D�E���D�C� ��
-������E.����$�$�����D��%����D��A�D������D(�
A�BC�D�E���D�C��/�D�C��-��0�����.�1D-��.�*�-��
#CD�*C+�A�D�.������D����
� � � �
� � � �
� � � �
� � � �
�
/�ADFC����CB�C�&D�C��4�E�C�AD����CBD��**DE��%����CB�(���F��8+�)D���9:1��E��CB�C������E��F��C��E�*FD(DECD���;�,D� �(� A��*�DCD� �E�� �AA�F�CD'� .��� EDAD((�F�� �AC��E(� ��F� A�D�E�E)� �E�� �DA�EC���E�C��E� B�,D� ;DDE�C�4DE1� �E�� CBD� (�AAD((� A�E��F�D�� �E�� &�CED((D�'� �BD� *�FC(1� A��*�EDEC(1� �E�"�F� �����D(� �E� CB�(�(B�*�DEC��FD�(��D���F�(B�**�E)��E��B�E���E)'��<��D�� � �
+�(�C��E�� � �
-�CD�� � �
��)E�C�FD����777777777777777777777777777777��
�
FC��"�C����D����$���2�'��������C$������A�����������B������� �
A�D�.������D������!!!!!!!!!!!!!!!!7777�
Cleaning and Decontamination Form
DASGIP Information and Process Technology GmbH ���� Rudolf-Schulten-Straße 5 ���� 52428 Jülich ���� Germany
Tel: +49 2461/980-0 ���� Fax: +49 2461/980-100 ���� www.dasgip.de
In order to provide maximum protection for our employees, associates, and the environment, please pay attention to the consecutively described arrangements for sending equipment to the DASGIP as well as for the on-site service:
DASGIP requires complete disclosure of biohazard and/or toxicity information on materials contacting or processed in our equipment. In the case of rental equipment, DASGIP will require this information before the rental equipment is shipped to the customer site.
Equipment returned to DASGIP must be free of substances that are bio-hazardous or any substances that are not bio-hazardous but may require special handling precautions. The cleaning and decontamination procedure must be completely documented (Page 1 of this form).
For equipment returned to DASGIP for service or any other reason please fill out the form (Page 1) completely and pre-send it via fax or email to DASGIP and attach the original to the outside of the package. Equipment which was sent to DASGIP without the completed form will not be handled. If the form is missing or contaminations are visible and no declaration is subsequently handed in, the goods will be returned at the sender’s costs.
This form must also be completed before any on-site service, whether covered under warranty or not!
Based on the information provided, each material should be assigned to one of the four Biosafety Levels as defined in the European Directive 2000/54/EG. The use of rental equipment is restricted to materials assigned to Biosafety Level 1. The use of materials assigned to a higher Biosafety Level needs prior written agreement by DASGIP.
Substances that are not bio-hazardous but may require special handling precautions must also be listed. Information should include name of chemical, chemical formula and hazard rating. Examples of these substances include, but are not limited to, acids, bases, radioactives and flammables.
Material Safety Data Sheets (MSDS) for any of the above items must be included with this form.
Cleaning and Decontamination
The cleaning and decontamination protocol described below is understood as a minimal procedure for cleaning and decontamination of equipment in contact with material appending to Biosafety Level 1.
Any substances that are bio-hazardous or any substances that are not bio-hazardous but may require special handling precautions must be removed, neutralized or inactivated in an appropriate manner. Check your in-house standard operating procedures concerning cleaning and decontamination to minimize any risk.
Please be aware, the undersigned is solely responsible and liable that all send parts, components and modules are safe for shipping and handling.
In the case that you are using alternative cleaning agents, please check chemical compatibility charts for all surface materials, in particular, but are not limited to, with components made out of acrylic glass. If in doubt, please check with DASGIP for advise.
Cleaning and Decontamination Form
DASGIP Information and Process Technology GmbH ���� Rudolf-Schulten-Straße 5 ���� 52428 Jülich ���� Germany
Tel: +49 2461/980-0 ���� Fax: +49 2461/980-100 ���� www.dasgip.de
Protocol (Applicable for Biosafety Level 1 only):
A. Control System Components, Modules, Cables, Gas Tubing, Incubators, Racks and Waterbath:
• Control system components include, but are not limited to, process computer, monitor, mouse, keyboard, uninterruptiblepower supply (UPS), balances, magnetic stirrer control box or immersion thermostats.
• Modules include, but are not limited to, sensor module, multi pump module, gas mixture module, one or multi positionmagnetic stir plates.
• Cables include, but are not limited to, power cords, extension cords, sensor cables (pH and pO2), DTP control cables,sensor module control cables or multi position power strips.
• Incubators include, but are not limited to, dry incubator, UNITRON-pro incubator shaker or shaking tables.
• Gas tubing include, but are not limited to, gas tubing, push-in connectors or stainless steel inline filter.
• Racks and waterbath include, but are not limited to, acrylic racks incl. shelves, acrylic stands or acrylic or stainless steelwaterbath.
Wipe down all surfaces, if applicable inside and outside, with 80% EtOH.
B. Glass Ware, Stainless Steel Parts, PTFE Parts, Caps and Fittings, Pump and Feed Tubing:
• Glass ware includes, but is not limited to, glass body of culture vessels, glass body of reservoir bottles, glass body ofharvesting bottles, glass ball agitators or spargers with porous glass frits.
• Stainless steel and PTFE parts include, but are not limited to, stainless steel flanges, stainless steel pipes, stainless steelspargers, heating and cooling coils of immersion thermostat, PTFE fittings or PTFE head gears, paddle type agitator.
• Caps and fittings include, but are not limited to, screw caps, screw caps with port, stainless steel caps, stainless steelcaps with port, tube fittings, Luer and Luer-lock fittings or compression fittings.
• Pump and feed tubing include, but are not limited to, PTFE feed lines, Pharmed tubing or Maprene tubing.
Soak in 0.2M NaOH for 1 hour followed by a 3x deionized water rinse (if applicable inside and outside). Autoclave at 121 °C for 30 min. in sterilization pouches indicating the exposure to sterilization conditions and ship sealed pouches back.
C. Sensors:
• Sensors include, but are not limited to, pH sensors, pO2 sensors or temperature sensors.
Soak sensor tip in 0.2M NaOH for 1 hour followed by a 3x deionized water rinse (if applicable inside and outside). Autoclave sensor, if applicable (e.g. an exception would be a hand held temperature sensor with permanent attached cable), at 121 °C for 30 min. in sterilization pouches indicating the exposure to sterilization conditions and ship back sealed pouches.
D. Not Returnable Parts: Parts not to be returned unless unused or otherwise advised by DASGIP include, but are not limited to, venting filters, gas in- and outlet filter (with the exception of the stainless steel inline gas filters), silicone tubing, silicone washers, silicone septa, silicone gaskets, rubber seals, sampling valves, or one way valves or cannulas.
���������A�B�C����DE�F�
���������E�F���E��A���E���������E�E����F���������������AB��CDE�CF��D������������������AB�������DF��E�� � �
�D��������� !"��#�#������$�%�������� !"��#�!##������&&&'��()�*'�D� �
+�D�(D�C�����A�CBD���F��,�����%��F�D�
����C��-.�
/+*@kk`i_jma Ijmoc <h`md^\��E���CC�AB�CBD��F�)�E���E ���A��CBD�*�A4�)D5�
�
$F���3��*�E������ �
.��FD((������� � �
� � ����� � �
+B�ED�� ������� ����������������������������������������������������$�%����� �
6*DF�C�F����� � �
�����!����"""""""""""""""""""""""""""""""""��
#�E�������$�����%&�'(!����7777777777777777777��
���)���*E+!���� """"""""""""""""""""""""""""��
�BD���(CD��D8��*�DEC�FD8��FD(�(DF,�AD��E�"�F�FD*��F��F���-.�/+��
�
���������� � B�C����������������A���,���F���E�E��-E�������E���.E�)��������EFF���+�
�
�D�(�E���F�(DF,�AD��F�FD*��F!������
/�BDFD9��ADFC����CB�C�����EDAD((�F���AC��E(���F�A�D�E�E)��E���DA�EC���E�C��E�B�,D�C�4DE�*��AD1�A�FFD(*�E��E)���A��DEC�C��E����
CB�(� �(� �CC�ABD�� :(�CD� !� ��� CBD� 3�D�E�E)� �E��-DA�EC���E�C��E� $�F�;� �E�� CB�C� CBD� *�FC(1� A��*�EDEC(� �E�"�F������D(� �E� CB�(�
(B�*�DEC��FD�(��D���F�(B�**�E)��E��B�E���E)'��
�
�
�)E�C�FD�����777777777777777777777777777777�
������C��E� �����* F���� �������* F����$E����E����� ��EF�������/A�C���F���
� � � �
� � � �
� � � �
� � � �
�����������ABCDEF�����������������������������������
ABCDEF�E���������������F���������������� �D�!��������������AB��CDE�CF��D������������������AB�������DF��E�� � �
�D��������� !"��#�#������$�%�������� !"��#�!##������&&&'��()�*'�D� �
�
��
� � �!'�+,�-.������D'� �'�.�D�(D��*DE�CBD����*������E��/� 0'�/�E(DFC�CBD�&B�CD�(*�ADF��(�
(B�&E'��
� � ��'�1��(D�CBD����*���*��(���F��(��*�((�2�D'�
�'�3(D�CBD�FD���E�E)�*�FC(���F�*�A4�E)��(�(B�&E'''��
'�'''�E�'''�
� � �5'�'''*�C�CBD������D��E�CBD�2�%���4D�CB�('�
�'�$�E�(BD�/� �'�'''�E��(��D'�
� � �� � �� � �
�B�E4�������F����F�A��*DF�C��E'��
6D(C�FD)�F�(�� �
+,�-.�-E��F��C��E��E��.F�AD((��DABE���)����27�
� �
�����������ABCDEF�"��!������������������������������
ABCDEF�E���������������F���������������� �D�!��������������AB��CDE�CF��D������������������AB�������DF��E�� � �
�D��������� !"��#�#������$�%�������� !"��#�!##������&&&'��()�*'�D� �
� � �!'�-��CBD�6��2��A4��(�D8��**D��&�CB��19+:�A�2�D�CF����E�"�F����(4�B���DF�/�
�'�/�FD��;D�CBD�A�2�D�CF��:�CBD����(4�B���DF��E��CBD�(���A�E�F�E)('�
0'����FD��;D�CBD�19+��*DE�CBD�19+��E�CBD�2�A4(��D�<��BD%�(AFD&(=��E��FD��;D�CBD�C&��(AFD&(��C�CBD�2�CC��'��
� � ��'��D��;D�CBD�(*��(B�&�CDF�*F�CDAC��E�2���D�<CBFDD�BD%�(AFD&(��E�CBD�2�A4(��D='��
�'�6��2��A4�FD������F�*�A4�E)'� '�.�C�CBD�6��2��A4��EC��CBD�C&�������*�DAD('�
� �
�
5'�.�C�CBD�6��2��A4��E�CBD�2�%���4D�CB�('�
�'�$�E�(BD���E��(��D'� �
� � �� � ��B�E4�������F����F�A��*DF�C��E'��6D(C�FD)�F�(�
� �
+,�-.�-E��F��C��E��E��.F�AD((��DABE���)����27�
� �
�