University of Colorado, Boulder CU Scholar Chemistry & Biochemistry Graduate eses & Dissertations Chemistry & Biochemistry Spring 1-1-2011 Design and Development of a Droplet Mass Spectrometer Luis Alberto Cuadra-Rodriguez University of Colorado at Boulder, [email protected] Follow this and additional works at: hp://scholar.colorado.edu/chem_gradetds Part of the Physical Chemistry Commons is Dissertation is brought to you for free and open access by Chemistry & Biochemistry at CU Scholar. It has been accepted for inclusion in Chemistry & Biochemistry Graduate eses & Dissertations by an authorized administrator of CU Scholar. For more information, please contact [email protected]. Recommended Citation Cuadra-Rodriguez, Luis Alberto, "Design and Development of a Droplet Mass Spectrometer" (2011). Chemistry & Biochemistry Graduate eses & Dissertations. Paper 29.

Upload: others

Post on 30-Apr-2020

5 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

University of Colorado, BoulderCU ScholarChemistry & Biochemistry Graduate Theses &Dissertations Chemistry & Biochemistry

Spring 1-1-2011

Design and Development of a Droplet MassSpectrometerLuis Alberto Cuadra-RodriguezUniversity of Colorado at Boulder, [email protected]

Follow this and additional works at: http://scholar.colorado.edu/chem_gradetds

Part of the Physical Chemistry Commons

This Dissertation is brought to you for free and open access by Chemistry & Biochemistry at CU Scholar. It has been accepted for inclusion inChemistry & Biochemistry Graduate Theses & Dissertations by an authorized administrator of CU Scholar. For more information, please [email protected].

Recommended CitationCuadra-Rodriguez, Luis Alberto, "Design and Development of a Droplet Mass Spectrometer" (2011). Chemistry & BiochemistryGraduate Theses & Dissertations. Paper 29.

Page 2: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

Design and Development of a Droplet Mass Spectrometer

by

Luis A. Cuadra-Rodríguez

B.S., University of Puerto Rico-Rio Piedras, 2004

A thesis submitted to the

Faculty of the Graduate School of the

University of Colorado in partial fulfillment

of the requirements for the degree of

Doctor of Philosophy

Department of Chemistry and Biochemistry

2011

Page 3: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

This thesis entitled:

Design and Development of a Droplet Mass Spectrometer

Written by Luis A. Cuadra-Rodríguez

has been approved for the Department of Chemistry and Biochemistry program by

___________________________

Dr. G. Barney Ellison

___________________________

Dr. Veronica Bierbaum

Date: April 11, 2011

The final copy of this thesis has been examined by both the signatories, and we find that both the content and the form meet acceptable presentation standards of scholarly work in the above

mentioned discipline.

Page 4: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

iii

Cuadra-Rodríguez, Luis A. (Ph.D. Physical Chemistry)

Design and Development of a Droplet Mass Spectrometer

Thesis directed by Dr. G. Barney Ellison

A droplet mass spectrometer has been built to study the oxidation of salt-water droplets

coated by organic surfactants. These studies are to model the complex interfacial chemistry of

nascent organic marine aerosols. The model predicts the aerosols to have a structure of an

“inverted micelle” where a core salt-water droplet is coated by surfactants. Reactions with

atmospheric oxidants are predicted to transform the hydrophobic surface to hydrophilic, making

the aerosols better cloud condensation nuclei.

A small review on single particle mass spectrometry details the consideration for the

design and development criteria for the construction of the instrument. The design and

development of this laser based instrument is discussed in detail, describing each component of

the spectrometer. The essence of the experiment is to oxidize the surfactant coating, and then to

vaporize with a laser pulse the water droplet to release the ions from solution (similar to

MALDI), for subsequent analysis with a quadrupole mass analyzer.

Quantitative experiments were carried out to understand water evaporation in an

aerodynamic lens inlet used in single particle mass spectrometers. Also, the process of

generating surfactant coated water droplets in the laboratory is described. Finally, experiments

to study the microphysical characterization of common hygroscopic particles coated by

surfactants are discussed.

Page 5: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

iv

Acknowledgments

Coming from far away in Puerto Rico, my transition to grad school and to Boulder has

been smooth and joyful. This is in part because of the beautiful mountains around, but most

importantly for the people that make this place amazing, scientifically and socially.

First, I would like to thank my advisor, Prof. G. Barney Ellison. Barney as an advisor

and educator is the ultimate motivator. Just by talking to him two times I decided to join his

research group and an exciting new experiment. In his mind impossible experiments do not

exist. As he likes to point out; “impossible is just a few years from now”. Barney’s exciting

approach to science gets everyone going and that is something I will never forget. Like the time

Beth, Barney and I were having breakfast talking about experiments to try. After a while the

lady next to us came over to our table and told Barney that he should talk to K-12 kids to get

them into science. Imagine the extent of Barney’s enthusiasm. Also, I will not forget all the

help and harsh remarks about my presentation and thesis preparation, it proved to be very

valuable advice.

I would like to thank my parents (Olga Iris Rodriguez and José L. Cuadra) for their

invaluable support. I never take for granted the sacrifices they have to go through and the love

they have given to raise me and my sisters. Mom’s strength, support and wise advice were

always there, whether I wanted it or not and that is important for me. Dad’s kindness and good

sense of humor always made things go smoothly. Because of them I feel very fortunate about

my life and the privilege to choose to do what I like in life. And that includes science as a

career. Just as mom likes to say I’m the crazy scientist. My sisters (Ailleen and Suhail) for

being the best siblings, and most importantly friends I could ever ask for. I certainly love talking

Page 6: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

v

to them and answering questions about experiments and lasers. Thanks to my brothers-in-law

(Hector Acevedo and Miguel Claudio) for their unconditional support.

Next I would like to thank all the people that make the project I was involved with move

forward. Dr. Alla Zelenyuk (PNNL) taught me probably everything I know about aerosols. Also

she gave free rein in her lab to use all the instruments and run experiments without any

supervision even though I was a rookie. The guys at the machine shop were the ones that

constructed the instrument we designed, especially Donald “the Don” David, Craig Joy, Ken

Smith and Jim Kastengren. They are helpful and get our lab experiments in working order.

Thanks to Sid Gustafson for the shop class and for his suggestions when working on a machining

project. I appreciate talks that we have with many different scientists that help us in our

experiments. These are: Dr. Dan Murphy (NOAA) for visiting to see our instrument, Dr. Kim

Prather (UCSD), Dr. Stephen Barlow (PNNL), Dr. Jim Cowin (PNNL), Dr. Mike Alexander

(PNNL), Dr. Alex Laskin for the use of the SEM machine and many others that in some way or

another helped us.

I can’t say enough about the interaction and opportunity that our group meeting provides.

Thanks to the great mentors Dr. Carl Lineberger, Dr. Veronica Bierbaum and Dr. Mathias Weber

and their research groups. The interaction with them and their group members made me a better

scientist. Especial thanks to Dr. Nicole Eyet, Dr. Stephanie Villano and Dr. Oscar Martinez for

the help with the electron impact ionization source. I also give thanks to all the chemistry staff

and people from the stockroom (Mark Stewart, Lynn Sarkis, Becky Thirlaway, Cori Fagan-

Edminster, Cheryl Graham, Maia and Hillary).

I would like to thank the DOE Global Change Educational Program Graduate Research

Environmental Fellowship for financial support throughout my graduate school career. Ricky

Page 7: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

vi

Petty, Milton Constantin, Jeff Gaffney, Nancy Marley and many others made sure the people in

the program had successful graduate careers. They never hesitated to show their support. The

same can be said about Barbara Kraus, Kristin Lopez, Amahl Scheppach and Dr. Mark

Hernandez, Ruth Lopez from the Colorado Diversity Initiative office. It was always a pleasure

to work with them and undergrad summer interns in their quest to do research.

I’m fortunate to have excellent lab mates that I can call friends outside the lab. Dr. Xu

Zhang and Dr. Mark Nimlos were here when I joined the lab and were people I looked up to.

Thanks Mark for having those ski trips in Monarch and letting us stay at your cabin. Dr. Al

Maccarone and Dr. Tim D’Andrea made my entry to lab fun and filled with knowledge and

motivation. Al gave me really good advice on how to deal with lab work and with Barney.

From Tim I learned how to reach and teach people since he is a great instructor. Also, I will not

forget the time we went to ski at Eldora in the morning and then came to lab in the afternoon to

work. Thanks to Dr. Brad Flowers for working with me in the instrument assembly and initial

tests. Also, for talking to me about possible post doctoral opportunities.

Most recent lab mates include AnGayle “AJ” Vasiliou, Dr. Beth Reed, Jong Kim, Adam

Scheer, Kim Jacsh, Dr. Krzysztof “the Pole” Piech and Tom Ormond. I really appreciate the fact

that we have lunch together and have very entertaining talks. The dynamic we have in lab makes

the learning experience much more valuable. This is the same as I used to do with my friends

back in PR and it made me feel at home. I think it is fair to say that we talked about SS, science

and sports. Even though we have fun in the lab, we can always talk about experimental

problems and help each other think through these challenges, which is how grad students learn in

lab. AJ has been a true friend that cares about my well being and for that I’m thankful. I will

always root for any sports team that plays a Boston team, just as friends do! I appreciate the

Page 8: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

vii

support in good and bad situations and for managing the softball team Disrespect. I thank Beth

for sharing the challenges and problems in lab work and developing our instrument. I’ll never

forget the good times at happy hour talking about it. Also thanks for caring about me, the chats,

sharing time and dinners outside the lab. I will miss Jong’s ridiculous stories that turn out to be

true in some ways. Like the one about the killer frog. It is always fun to talk to Adam about

anything, including cinco de mustache. Keep the Spanish learning going and in a few years

we’ll be talking in my native language. Kim, thanks for being an honest person and friend. Take

care of those moose faces and your weak teeth. There is lot to say about “the Pole” who is a true

character. Never forget all the good stuff he taught us in Polish, especially how to say shut up.

Even though I don’t know Tom for a long period of time, it is good to have him in lab being part

of the group, in lab and outside lab. Keep the hard work.

Last but not least I would like to thank all the friends I made along the way here in

Colorado. Thanks to Suinya Perez for her support, all the years we spent together and the things

we shared. I wish you the best in the future. Nestor Lopez and Mabel Ramirez are a couple of

the best friends I’ve ever met. Nestor, I will not forget going to the gym at 6 am every morning,

shooting games, “el vinito” during happy hour and all the other things that you share with true

friends and brothers. I cannot forget the support and good times that I spent with many of these

friends that hopefully don’t get left behind. Katy Plath, Chris Arellano, Ruth Lopez, Joan

Marcano, Eric Antillon, Danielle Griego, Niel Rodriguez, Idalis Villanueva, Fernando Rivera,

Marlene Rivera, Fernando “Fer” Rivera, Gianina Caballero, Javier Torres, Myrna Melendez,

Carlos Figueroa, Melba Figueroa, Carlos Rosado and many others. I cannot forget the friends

that I’ve had before coming to CO in Miguel Fernandez, Gina Batista and my godchildren

Gimarie and Miguelito. Miguel has been my best friend since I was 11 years old and I

Page 9: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

viii

appreciate all his support. Included on the list are Anette Casiano, Brenda Valle, Rolando

Santos, Rafael Gonzalez, Viviana Lopez, Danny Cruz, Vanessa Marrero and Maritere Padilla.

I’m sorry if I forgot anyone but the pages are short and so is my brain. I truly appreciate

everyone’s support!

Page 10: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

ix

Table of Contents

Chapter I Introduction.................................................................................................................... 1

1.1 Background................................................................................................................... 1

1.2 Marine Aerosol Model.................................................................................................. 3

1.3 Atmospheric Processing................................................................................................ 5

1.4 Structure of Thesis ........................................................................................................ 7

References for Chapter I ........................................................................................................... 10

Chapter II Single Particle Mass Spectrometry............................................................................. 13

2.1 Introduction................................................................................................................. 13

2.2 Introduction to Single Particle Mass Spectrometry .................................................... 14

2.2.1 SPMS Inlets ................................................................................................................ 15

2.2.2 Particle Detection and Sizing...................................................................................... 18

2.2.3 Particle Vaporization and Ionization .......................................................................... 21

2.2.4 Mass Analysis ............................................................................................................. 24

2.3 Photoionization ........................................................................................................... 25

2.4 Droplet Spectrometer Objective ................................................................................. 26

2.5 Previous Work ............................................................................................................ 27

References for Chapter II .......................................................................................................... 34

Chapter III Droplet Mass Spectrometer Design........................................................................... 37

3.1 Introduction................................................................................................................. 37

3.2 Aerosol Generation ..................................................................................................... 41

3.2.1 Inorganic Salts ............................................................................................................ 41

Page 11: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

x

3.2.2 Aerosol Chemical Components in the Laboratory...................................................... 45

3.2.3 Bubbler........................................................................................................................ 47

3.2.4 Vibrating Orifice Aerosol Generator .......................................................................... 49

3.3 Droplet Spectrometer: Design..................................................................................... 51

3.3.1 Capillary Inlet ............................................................................................................. 53

3.3.2 Drift Tube.................................................................................................................... 55

3.3.2.1 Water Cluster Dissociation ......................................................................................... 56

3.3.3 Droplet Detection........................................................................................................ 58

3.3.4 Droplet Vaporization .................................................................................................. 58

3.3.4.1 Laser system................................................................................................................ 58

3.3.4.2 Energy Requirements.................................................................................................. 59

3.3.4.3 Laser Triggering.......................................................................................................... 62

3.3.5 Signal Estimates.......................................................................................................... 65

3.3.6 Ion Detection............................................................................................................... 66

3.3.7 Expected Results......................................................................................................... 68

3.4 Droplet Spectrometer: Development .......................................................................... 71

3.4.1 EI Source..................................................................................................................... 71

3.4.2 Droplet Detection........................................................................................................ 75

3.4.3 Droplet Vaporization .................................................................................................. 77

References for Chapter III......................................................................................................... 82

Chapter IV Evaporation of Water from Particles in the Aerodynamic Lens Inlet: An Experimental Study....................................................................................................................... 85

4.1 Introduction................................................................................................................. 85

Page 12: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

xi

4.1.1 Aerodynamic Lens Inlet.............................................................................................. 87

4.1.2 SPLAT ........................................................................................................................ 88

4.2 Experimental ............................................................................................................... 90

4.3 Results and Discussion ............................................................................................... 93

4.3.1 (NH4)2SO4 Particles .................................................................................................... 93

4.3.2 NaC Particles............................................................................................................. 97

4.3.3 Other Particles........................................................................................................... 100

4.4 Conclusion ................................................................................................................ 102

References for Chapter IV....................................................................................................... 104

Chapter V Measurements and Interpretation of the Effect of a Soluble Surfactant on the Density, Shape and Water Uptake of Hygroscopic Particles .................................................................... 107

5.1 Introduction............................................................................................................... 107

5.2 Experimental ............................................................................................................. 110

5.3 Results and Discussion ............................................................................................. 116

5.3.1 NaNO3 Particles Coated with SDS ........................................................................... 116

5.3.1.1 NaNO3/SDS Particles Density Measurements.......................................................... 116

5.3.1.2 NaNO3/SDS Growth Factor Measurements.............................................................. 118

5.3.1.3 Vacuum Aerodynamic Size Distributions of NaNO3/SDS Particles ........................ 120

5.3.2 NaC Particles Coated with SDS.............................................................................. 124

5.3.2.1 NaC/SDS Density and Shape Measurements.......................................................... 124

5.3.2.2 NaC/SDS Growth Factor Measurements ................................................................ 130

5.4 Conclusions............................................................................................................... 132

References for Chapter V........................................................................................................ 136

Page 13: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

xii

Bibliography ............................................................................................................................... 139

Page 14: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

xiii

List of Figures

Figure 1.1 Chemical structure of a nascent organic marine aerosol. A salt/water core is encapsulated by a monolayer ( 2.5 nm thick) of organic surfactant. Circles represent the polar head (-CO2

-) while the squiggly lines represent the alkyl chain (CH3(CH2)10-). ........................................................................................................ 4

Figure 1.2 Atmospheric processing of an organic marine aerosol. .......................................... 5

Figure 2.1 Schematic of main components of a Single Particle Mass Spectrometer............. 15

Figure 2.2 Type of inlets for Single Particle Mass Spectrometry: (a) conical nozzle, (b) orifice, (c) nozzle with short capillary and (d) nozzle with long capillary. Particles are sampled from atmospheric pressure on the inlets’ open ended orifice going into vacuum (10-3 Torr – 10-6 Torr) through the 500 µm apertures. Adapted from reference 19.................................................................................................. 17

Figure 2.3 Aerodynamic lens inlet schematic. A 100 µm orifice is placed at the aerosol inlet entrance to constrict the sampling flow to about 100 cm3 min-1. The pressure after the orifice is the aerodynamic lens inlet working pressure, usually 2 Torr. Adapted from reference 17. .................................................................................. 18

Figure 2.4 Mass spectra of individual succinic acid particles. Spectra in (a) were generated by laser ablation with excimer laser UV radiation at 193 nm and in (b) by IR laser (10.6 µm) vaporization followed by ionization with UV radiation at 193 nm. Adapted from reference 16. .................................................................................. 24

Figure 2.5 Droplet vaporization by an IR laser pulse. R-CO2- represents surfactant coating

where R is a long alkyl chain................................................................................ 27

Figure 2.6 LILBID mass spectrometer schematic. Adapted from reference 9....................... 28

Figure 2.7 Spectrum of Li+ ions desorbed using LILBID technique from an aqueous liquid jet. Adapted from reference 9. .............................................................................. 30

Figure 2.8 SPMS from reference 32. The spectrometer consists of an aerodynamic lens inlet, two differential pumping chambers, two diode lasers and two PMTs for light scattering detection and CO2 and UV lasers for vaporization and ionization, respectively. .......................................................................................................... 31

Figure 2.9 Integrated Rb+ ion signal as a function of RbC concentration as measured with the TOFMS after IR laser vaporization. Na+ signal comes from impurities in solution. Adapted from reference 5. .................................................................... 33

Figure 3.1 Experiment schematic diagram............................................................................. 39

Figure 3.2 Oxidation of organic surfactant layer on marine organic aerosols. ...................... 40

Page 15: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

xiv

Figure 3.3 Phase diagram transformation, growth () and evaporation () for a (NH4)2SO4 particle as a function of RH. Red arrow indicates efflorescence point of 37-40% RH and blue arrow indicates deliquescence point of 80% RH. Adapted from reference 3............................................................................................................. 43

Figure 3.4 NaNO3 growth factor as a function of relative humidity (RH)............................. 46

Figure 3.5 Aerosol generation schematic. (A) shows the bubbler method and (B) the VOAG approach................................................................................................................ 48

Figure 3.6 VOAG aerosol generation numbers. A 25 µm droplet is dried and then grown by H2O uptake at 86% RH to produce a 1 µm droplet coated by SDS surfactant (see text for description)............................................................................................... 51

Figure 3.7 Droplet mass spectrometer overview (see text for description)............................ 52

Figure 3.8 Droplet spectrometer’s inlet.................................................................................. 54

Figure 3.9 Drift tube diagram. Particles enter the drift region, are detected by light scattering from continuous green laser beams, and trigger the 10 Hz YAG/OPO IR laser beam for droplet vaporization to release ions into the gas phase. ........................ 56

Figure 3.10 Absorption spectrum for bulk water in the infrared. Red arrow points at the peak absorption of water at 3400 cm-1 or 2.94 µm. Adapted from reference 22. ........ 60

Figure 3.11 Triggering schematic. PMTs signal from aerosol light scattering to trigger the YAG flashlamp and Q-switch. tT is the particle transit time between green laser beams, tF is the time for the flashlamp relative to PMT2 signal and tQ is the time to fire the Q-switch to fire the laser. ..................................................................... 64

Figure 3.12 QMS chamber. ...................................................................................................... 67

Figure 3.13 Expected results from the oxidation of SDS coated NaNO3/water droplets......... 70

Figure 3.14 Schematic set up for EI source ionization and ion extraction into the QMS region................................................................................................................................ 73

Figure 3.15 EI mass spectra for N2+ and Ar+ extracted from the drift tube into the QMS....... 74

Figure 3.16 Particle light scattering signal from PMTs to measure transit time, tT. (A) shows tT = 133 µs with drift tube pressure at 40 mTorr and (B) tT = 265 µs for a drift tube pressure of about 1 Torr........................................................................................ 76

Figure 3.17 Droplet spectrometer calibration with PSL particles with 495 nm, 900 nm, 1000 nm and 1600 nm diameters. Sodium nitrate (SN) poly-disperse particles from a 0.1 M solution were also measured for comparison. ............................................ 77

Page 16: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

xv

Figure 3.18 Plates setup schematic for droplet vaporization optimization. The top plate repels ions to the bottom plate. The op-amp circuit is used to measure incident current on the grounded bottom plate. .............................................................................. 78

Figure 3.19 Circuit logic sequence of acquired signals to trigger the YAG/OPO laser and detect incident current on the stainless steel plates after droplet vaporization (see text for description)............................................................................................... 79

Figure 4.1 Calibration curve to measure particle vacuum aerodynamic diameter (dva) as a function of particle transit time (tT). PSL particles were used as size standards. The data are fitted with a power function. ............................................................ 89

Figure 4.2 Experimental schematic for three types of measurements. In all measurements droplets are size selected with a DMA conditioned at high RH. In (1) the wet mono-disperse aerosol flow is sent to SPLAT for dva measurements. In setup (2) the particles are dried through a diffusion dryer and then delivered to the spectrometer. In (3) particles are dried and then sent to the DMA conditioned at low RH to measured dry mobility diameter (dm) distributions. ............................ 92

Figure 4.3 Two vacuum aerodynamic size distributions of ammonium sulfate size selected droplets with dm = 206 nm at 84% RH. The dashed and solid lines indicate dry and wet runs, respectively..................................................................................... 94

Figure 4.4 Calculated dva as a function of water content for an (NH4)2SO4 droplet with dm,wet = 330 nm at 79 ± 1% RH. The efflorescence point and the measured vacuum aerodynamic diameters for wet and dry runs are indicated by ▲ and ∆, respectively. .......................................................................................................... 96

Figure 4.5 Two vacuum aerodynamic size distributions of NaC particles size selected at dm = 558 nm with the DMA at 84% RH. The dashed and solid lines indicate dry and wet runs, respectively. The circled numbers label three observed NaC “modes”................................................................................................................................ 98

Figure 4.6 Plots of the ratios of the measured vacuum aerodynamic and mobility diameters as a function of the particle mobility diameters selected by the DMA at 84% RH. Open symbols represent dried particles, filled symbols indicate measurements of wet droplets at high RH. The three NaC particle “modes” are indicated in the numbered circles. ................................................................................................ 100

Figure 5.1 Measurement to characterize inorganic salt/SDS particles’ hygroscopic properties. (1) shows the scheme to measure particle density. (2) and (3) are the schemes to measure GF by increasing the RH and decreasing RH, respectively. ................ 112

Figure 5.2 Measured and fit densities of internally mixed NaNO3/SDS particles with mobility diameters of 296 nm and 500 nm as a function of SDS concentration. The dashed and dash-dot lines are calculated assuming constant SDS densities of

Page 17: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

xvi

0.37 g cm-3 and 1.13 g cm-3, respectively. The solid line shows an empirical fit to the data. ............................................................................................................... 117

Figure 5.3 Measured growth factors of NaNO3/SDS particles as a function of SDS WF. The dashed and dash-dot lines show fits assuming constant SDS densities of 0.37 g cm-3 and 1.13 g cm-3, respectively. The solid line uses the SDS density that was calculated on the basis of the data shown in Figure 5.2...................................... 119

Figure 5.4 Measured dva distributions of 500 nm mobility diameter NaNO3/SDS particles for a number of SDS weight fractions indicated in the upper left corners. .............. 121

Figure 5.5 The effect of heating NaNO3/SDS particles with a mobility diameter of 500 nm on the particle dva distributions and hence densities, for a number of SDS WF values indicated in the upper left corners. .......................................................... 123

Figure 5.6 Measured dva distributions of NaC/SDS particles with a mobility diameter of 500 nm for a number of SDS WF values indicated in the upper right hand corners. Note that both line shape and line positions are sensitive to particle SDS concentration....................................................................................................... 125

Figure 5.7 Measured and fit densities of internally mixed NaC/SDS particles with mobility diameters of 296 nm and 500 nm as a function of SDS WF. The dashed and dash-dot lines are calculated assuming particle sphericity and constant SDS densities of 0.37 g cm-3 and 1.13 g cm-3, respectively. The solid line shows an empirical fit to the observed data................................................................................................. 127

Figure 5.8 Microscopic images of internally mixed NaC/SDS particles with SDS WF of 0.0098, 0.090 and 0.37 shown in the upper, middle and lower frames, respectively. ........................................................................................................ 128

Figure 5.9 Particle dynamic shape factor (DSF), particle calculated effective density and the SDS density as a function of the SDS WF.......................................................... 130

Figure 5.10 DSF corrected fit and measured GF for the NaC/SDS particles. The fit uses the DSF and densities shown in Figure 5.9. ............................................................. 132

List of Tables

Table 4.1 Summary of dva Measurements of (NH4)2SO4 Droplets Selected by the DMA at

three RH values..................................................................................................... 95

Table 4.2 Summary of Measurements on Six Different Particle Types ............................. 101

Page 18: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

________________

Chapter I

Introduction

________________

1.1 Background

Aerosols are defined as relatively stable suspensions of solid and/or liquid particles in air.

Aerosol particles may be emitted directly into the atmosphere, then known as primary particles,

or formed by chemical reactions or physical transformations, and referred to as secondary

particles. Ambient particles are usually classified into coarse mode particles, with diameters

greater than 2.5 μm, and fine mode particles, with diameters smaller than 2.5 μm.1,2 In general,

fine and coarse particles originate separately, are transformed differently, removed from the

atmosphere by different mechanisms and have different chemical composition. Coarse particles

are mainly generated from mechanical processes, such as windblown dust, ocean waves,

crushing or grinding operations and dust kicked up on unpaved roads by cars. Fine mode

particles are mainly formed by gas-to-particle conversion in a process known as nucleation.1-3

Fine mode particles continue to grow by the coagulation of smaller particles and further gas

condensation.

Particle size range in the atmosphere is between 10 nm -10 µm. Particles bigger than 10

µm are more easily scavenged from the atmosphere by sedimentation or precipitation processes.

Therefore particles smaller than 10 µm have longer lifetimes in the atmosphere. Atmospheric

particle lifetimes range from hours to weeks.3 For this reason aerosol particles can influence the

atmosphere’s chemistry and subsequently change climate directly or indirectly. Aerosol particles

can absorb and scatter terrestrial and solar radiation affecting directly the transport of light

Page 19: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

2

through earth’s atmosphere.4-6 They also are involved in the formation of clouds by acting as

cloud condensation nuclei (CCN) and ice nuclei which have a decisive influence on earth’s

temperature.7

Atmospheric particles originate from a variety of anthropogenic and natural sources.

Therefore particle composition and concentration depend strongly on their origin. Concentration

in urban areas is approximately 104 particles cm-3 while in rural environments is on the order of

102 particles cm-3. For the last two decades organic matter in aerosol particles has been one of

the central topics in atmospheric research.8-10 Since the earth’s surface is approximately 70%

ocean, organic marine aerosols have gained more attention in atmospheric research.

The first Aerosol Characterization Experiment (ACE1) field campaign demonstrated that

organic matter content in marine aerosols is approximately 10% of dry particle mass in pristine

marine environments9,11 and it could be as high as 50%. The aerosol composition in ACE1 was

measured using particle analysis by mass spectrometry12 and showed that the organic mass

content depends on sampling altitude and location. At the marine boundary layer pure inorganic

salts are not prevalent and a large fraction of the particle composition was found to be organic

matter.9,11 This organic material on particles suggests that the particles’ chemistry and cloud

formation properties are different than pure sea salt. Motivated by these findings Ellison et al.13

proposed a model for organic marine aerosol chemistry. This model was previously proposed by

Gill et al.14 who predicted it will have no effects in the atmosphere. The revision by Ellison et

al. describes the model’s implications in the atmosphere’s composition, cloud formation and

climate change. This model is the main focus of this thesis.

Page 20: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

3

1.2 Marine Aerosol Model

Wave action from the ocean releases sea spray aerosols into the atmosphere with

concentrations1 of approximately 300 particles cm-3. These aerosol droplets contain water,

inorganic sea salts and organic material from the ocean. The organic molecules in the aerosol

droplets come from marine life and are long chain hydrocarbons with polar head groups such as

-CO2- or -NH3

+. Figure 1.1 shows a model for the structure of the organic marine aerosols. The

circles represent polar head groups and the squiggly lines the hydrocarbon chains. These

amphiphilic hydrocarbons are of biological origin and reside at the surface of the ocean with the

hydrophobic tail exposed to the atmosphere and the polar head at the water surface. The organic

surfactants are likely to partition to the marine aerosol’s surface. Water evaporation and droplet

coagulation processes result in a coated droplet with an organic surfactant monolayer.

Nucleation of partially coated droplets also leads to the formation of surfactant coated marine

droplets. This results in an “inverted micelle” structure for marine aerosols where an inert

hydrophobic surface is exposed to the atmosphere with a salt/water core. In Figure 1.1 a 1 µm

marine droplet has about 1010 water molecules in the brine core and approximately 107 surfactant

molecules at the surface. The organic layer film thickness is around 25 Å or 2.5 nm.

Page 21: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

4

Figure 1.1 Chemical structure of a nascent organic marine aerosol. A salt/water core is encapsulated by a monolayer ( 2.5 nm thick) of organic surfactant. Circles represent the polar head (-CO2

-) while the squiggly lines represent the alkyl chain (CH3(CH2)10-).

The “inverted micelle” model for organic marine aerosols accounts for the organic mass

loading found in field studies. It also provides an aerosol chemical structure and predicts them to

be poor cloud condensation nuclei.13 The hydrophobic organic layer is predicted to inhibit water

condensation and therefore particle growth into cloud droplets. Furthermore water exchange

from the bulk and atmospheric water vapor is predicted to be suppressed.

Donaldson and Vaida reviewed the current state of organic surfactants’ effects on aerosol

particles’ properties in the atmosphere.15 Transport of gases at the surface, CCN activity, uptake

of hydrophobic compounds or pollutants, reaction rates and mechanisms were all found to be

influenced by the presence of an organic film at the particles’ surface.16-21 This has implications

for aqueous phase heterogeneous chemistry. An emphasis on differentiating between soluble

and insoluble surfactants was also made. Surfactants reside mostly at the droplets’ surface but

soluble surfactants could partition to the bulk. This solubility creates less compact (more

Page 22: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

5

porous) films. It also makes the interaction between soluble surfactant, the atmosphere and the

bulk more dynamic. On the other hand insoluble surfactants exist only at the droplets’ surface

which makes more compact hydrocarbon films.15

1.3 Atmospheric Processing

Oxidation of the organic surfactants by atmospheric oxidants (i.e. OH, Cl, O2, NO3) will

transform the hydrophobic marine aerosol surface to a more hydrophilic one. This atmospheric

processing leads to the formation of more polar groups such as aldehydes, ketones, alcohols and

carboxylic acids at the particles’ surface. This is depicted in Figure 1.2. The oxidation

processes13 change the particles’ surface properties, reactivity, and enhances the organic marine

aerosols’ ability to condense water vapor making them better CCN.

Figure 1.2 Atmospheric processing of an organic marine aerosol.

The primary atmospheric oxidant during the daytime is the hydroxyl (OH) radical. Even

though OH radical is very reactive, its concentration1,2,22,23 in the atmosphere is relatively high,

(OH) = 1 x 106 cm-3. The OH radical homogeneous reactions with hydrocarbons have been

studied and are very fast.24,25 Hydrogen abstraction and addition to double bonds are possible

Page 23: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

6

OH radical reactions with organic molecules in the atmosphere. Because of the large bond

energy of water,26 OH radicals will abstract hydrogen atoms when they strike a film of saturated

organics. The bond energies of most alkanes27,28 are less than 100 kcal mol-1 making the

hydrogen abstraction exothermic by almost 1 eV ( 20 kcal mol-1). The hydrogen abstraction24,25

of ethane (CH3CH3) and propane (CH3CH2CH3) have rate constants of 3 x 10-13 cm3 s-1 and 1 x

10-12 cm3 s-1, respectively. It is predicted that the hydroxyl radical oxidation of the organic layer

of marine aerosols will have a rate constant k ≥ 1 x 10-12 cm3 s-1. Based on the OH radical

concentration of 1 x 106 cm-3, it was conjectured that organic marine aerosol processing will take

up to 10 hours for a surfactant coated droplet with a diameter of 0.2 µm.

The air-particle interfacial chemistry is a heterogeneous process. These processes have

been studied to a lesser degree when compared to gas phase homogeneous reactions. Bertram et

al.29 measured OH loss via chemical ionization mass spectrometry in a flow tube reactor coated

with organics. Heterogeneous loss of OH was found to be very efficient. The reaction

probability of the OH radical with the organic film was estimated to be roughly 50%. D’Andrea

et al.30 studied OH radical reaction with saturated and unsaturated hydrocarbon self assembled

monolayers (SAM) or thin films in high vacuum. Reactions were monitored using infrared

spectroscopy. For the saturated hydrocarbon film it was found that complete methyl (-CH3)

oxidation by hydrogen abstraction occurred in 10 minutes. Further exposure to OH does not

change the hydrocarbon film, suggesting that OH does not penetrate deeply into the organic film.

The alkene-terminated SAM only required 6% of an OH monolayer to deplete all double bonds

in the film. This was attributed to radical induced polymerization between hydrocarbon chains.

The hydrogen abstraction from saturated hydrocarbons at the organic marine aerosols

surface is the first step in the atmospheric processing. This reaction will form an alkyl radical

Page 24: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

7

(R•) at the particle’s surface. In the atmosphere the alkyl radical (R•) reacts with molecular

oxygen (O2) to form an alkyl peroxyl radical (ROO•). Subsequent reactions in the atmosphere

can transform the ROO• radicals at the particle’s surface into stable polar functional groups.

This atmospheric processing transforms the hydrophobic surface into a hydrophilic surface that

can bind water more readily, enhancing the CNN activity. Moreover, due to chemical

transformations particles’ optical properties could change affecting radiative transfer on earth.

The fundamental molecular reactions driving the chemical transformations in Figure 1.2

are poorly understood. The main objective of this thesis is the study of the initial steps of

atmospheric processing of organic marine aerosol as in the “inverted micelle” model. This

interfacial chemistry is proposed to be done in the laboratory setting at atmospheric pressure (1

atm) and temperature (298 K). Reactions will be carried out in a flow tube and analyzed with a

unique droplet mass spectrometer. The droplet spectrometer is specifically designed to test the

marine aerosol model and to retain the intrinsic liquid water content of marine aerosols.

1.4 Structure of Thesis

Chapter II describes single particle mass spectrometers used for the analysis of

atmospheric particles. These spectrometers provide chemical and physical information at a

single particle level. The main components of the spectrometers are described in some detail

along with considerations for the design of a droplet mass spectrometer. The study of water

droplets is one of the main considerations since the organic marine aerosol is comprised mostly

of water.

Chapter III focuses on the description of the droplet spectrometer design and all its

components. Particles that resemble the “inverted micelle” are generated using sodium nitrate,

NaNO3, as the inorganic salt in the water droplet’s core and sodium dodecyl sulfate (SDS) as the

Page 25: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

8

organic surfactant. Aerosol generation is described to produce a 1 µm salt/water droplet coated

with organic surfactant. The droplet spectrometer is a laser based instrument in which no

ionization step (photoionization or electron impact) is required for analysis therefore minimal

surfactant fragmentation is expected. A YAG/OPO infrared laser beam tuned at the peak

absorption of bulk water (2.94 µm or 3400 cm-1) is utilized to vaporize the water droplets. The

surfactants are ions in solution and at the droplet’s interface. Consequently the laser

vaporization releases the ions into the gas phase for mass analysis with a quadrupole mass

spectrometer.

Chapter III also discusses the steps in the development of the spectrometer. These steps

involve optimization of each of the droplet spectrometer components. Ion generation and

detection by electron impact ionization is explained in order to optimize ion detection with the

quadrupole mass spectrometer. Particle detection along with laser triggering is discussed in the

process for particle analysis.

Chapters IV and V describes the work done in collaboration with Dr. Alla Zelenyuk at

Pacific Northwest National Laboratory. Chapter IV focuses on quantitative experiments to

describe water evaporation from particles in the aerodynamic lens inlet. The particle transport

from atmospheric pressure into the high vacuum environment is accompanied by a large drop in

relative humidity. This change in RH may result in significant particle water loss and phase

change. Vacuum aerodynamic distribution measurements of DMA (differential mobility

analyzer) size selected particles were done at controlled RH conditions for different particle

types. It was found that almost all particles studied, except NaC, lose significant amounts of

water in the inlet. Thus it was concluded that to avoid data misinterpretation particles must be

dried before sampling with an aerodynamic lens inlet. The results from this study were taken into

Page 26: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

9

consideration for the droplet spectrometer inlet design. Since evaporation was observed to be

severe in the aerodynamic lens inlet, a capillary inlet was chosen for the droplet spectrometer.

Chapter V discusses the effects of sodium dodecyl sulfate (SDS) organic surfactant on

NaC and NaNO3 particles. These experiments were done in order to characterize physical

properties of model particles for organic marine aerosols. Particles’ density, vacuum

aerodynamic size distributions, shape and growth factors were measured in a wide SDS

concentration range, from pure inorganic salt to pure SDS particles. Independent measurements

were used to characterize the particles’ density and growth factor behavior quantitatively. It was

found that the SDS density is not constant as a function of SDS concentration for the mixed

salt/SDS particles. The SDS changing density was used to calculate the mixed particles’ growth

factors to reach an excellent agreement with the measured growth factors.

Page 27: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

10

References for Chapter I

(1) Finlayson-Pitts, B. J.; Pitts Jr., J. N. Chemistry of the Upper and Lower Atmosphere; Academic Press: San Diego, 1999.

(2) Seinfeld, J. H.; Pandis, S. N. Atmospheric Chemistry and Physics - From Air Pollution to Climate Change (2nd Edition); John Wiley & Sons, 2006.

(3) Poschl, U., Atmospheric aerosols: Composition, transformation, climate and health effects, Angewandte Chemie-International Edition 2005, 44, 7520-7540.

(4) Cruz, C. N.; Pandis, S. N., A study of the ability of pure secondary organic aerosol to act as cloud condensation nuclei, Atmospheric Environment 1997, 31, 2205-2214.

(5) Vaida, V.; Tuck, A. F.; Ellison, G. B., Optical and chemical properties of atmospheric organic aerosols, Physics and Chemistry of the Earth Part C-Solar-Terrestrial and Planetary Science 2000, 25, 195-198.

(6) Penner, J. E.; Hegg, D.; Leaitch, R., Unraveling the role of aerosols in climate change, Environmental Science & Technology 2001, 35, 332a-340a.

(7) Pruppacher, H. R.; Klett, J. D. Microphysics of Clouds and Precipitation; 2nd ed.; Kluwer Academic Publishing: Dordrecht: Holland, 1997.

(8) Lee, S. H.; Murphy, D. M.; Thomson, D. S.; Middlebrook, A. M., Chemical components of single particles measured with Particle Analysis by Laser Mass Spectrometry (PALMS) during the Atlanta SuperSite Project: Focus on organic/sulfate, lead, soot, and mineral particles, Journal of Geophysical Research-Atmospheres 2002, 107, 1-(1-13).

(9) Middlebrook, A. M.; Murphy, D. M.; Thomson, D. S., Observations of organic material in individual marine particles at Cape Grim during the First Aerosol Characterization Experiment (ACE 1), Journal of Geophysical Research-Atmospheres 1998, 103, 16475-16483.

(10) Murphy, D. M.; Thomson, D. S.; Middlebrook, A. M.; Schein, M. E., In situ single-particle characterization at Cape Grim, Journal of Geophysical Research-Atmospheres 1998, 103, 16485-16491.

(11) Murphy, D. M.; Anderson, J. R.; Quinn, P. K.; McInnes, L. M.; Brechtel, F. J.; Kreidenweis, S. M.; Middlebrook, A. M.; Posfai, M.; Thomson, D. S.; Buseck, P. R., Influence of sea-salt on aerosol radiative properties in the Southern Ocean marine boundary layer, Nature 1998, 392, 62-65.

Page 28: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

11

(12) Thomson, D. S.; Murphy, D. M., Analyzing Single Aerosol-Particles in Real-Time - Composition Analysis of Aerosols Is Critical to an Understanding of Their Origins and Fate - This Is Done by Combining Time-of-Flight Mass-Spectroscopy with Optical-Particle Detections and Sizing, Chemtech 1994, 24, 30-35.

(13) Ellison, G. B.; Tuck, A. F.; Vaida, V., Atmospheric processing of organic aerosols, Journal of Geophysical Research-Atmospheres 1999, 104, 11633-11641.

(14) Gill, P. S.; Graedel, T. E.; Weschler, C. J., Organic Films on Atmospheric Aerosol-Particles, Fog Droplets, Cloud Droplets, Raindrops, and Snowflakes, Reviews of Geophysics 1983, 21, 903-920.

(15) Donaldson, D. J.; Vaida, V., The influence of organic films at the air-aqueous boundary on atmospheric processes, Chemical Reviews 2006, 106, 1445-1461.

(16) Svenningsson, B.; Rissler, J.; Swietlicki, E.; Mircea, M.; Bilde, M.; Facchini, M. C.; Decesari, S.; Fuzzi, S.; Zhou, J.; Monster, J.; Rosenorn, T., Hygroscopic growth and critical supersaturations for mixed aerosol particles of inorganic and organic compounds of atmospheric relevance, Atmospheric Chemistry and Physics 2006, 6, 1937-1952.

(17) McNeill, V. F.; Wolfe, G. M.; Thornton, J. A., The oxidation of oleate in submicron aqueous salt aerosols: Evidence of a surface process, Journal of Physical Chemistry A 2007, 111, 1073-1083.

(18) Hearn, J. D.; Smith, G. A., Ozonolysis of mixed oleic acid/n-docosane particles: The roles of phase, morphology, and metastable states, Journal of Physical Chemistry A 2007, 111, 11059-11065.

(19) Hearn, J. D.; Renbaum, L. H.; Wang, X.; Smith, G. D., Kinetics and products from reaction of Cl radicals with dioctyl sebacate (DOS) particles in O2: a model for radical-initiated oxidation of organic aerosols, Physical Chemistry Chemical Physics 2007, 9, 4803-4813.

(20) Decesari, S.; Facchini, M. C.; Mircea, M.; Cavalli, F.; Fuzzi, S., Solubility properties of surfactants in atmospheric aerosol and cloud/fog water samples, Journal of Geophysical Research-Atmospheres 2003, 108, 6-(1-9).

(21) Covert, D. S.; Gras, J. L.; Wiedensohler, A.; Stratmann, F., Comparison of directly measured CCN with CCN modeled from the number-size distribution in the marine boundary layer during ACE 1 at Cape Grim, Tasmania, Journal of Geophysical Research-Atmospheres 1998, 103, 16597-16608.

Page 29: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

12

(22) Bell, N.; Heard, D. E.; Pilling, M. J.; Tomlin, A. S., Atmospheric lifetime as a probe of radical chemistry in the boundary layer, Atmospheric Environment 2003, 37, 2193-2205.

(23) Atkinson, R.; Aschmann, S. M., OH Radical Production from the Gas-Phase Reductions of O3 with a Series of Alkenes Under Atmospheric Conditions, Environ. Sci. Technol. 1993, 27, 1357-1363.

(24) Atkinson, R., Gas-Phase Tropospheric Chemistry of Organic-Compounds, Journal of Physical and Chemical Reference Data 1994, R1.

(25) Donahue, N. M.; Anderson, J. G.; Demerjian, K. L., New rate constants for ten OH alkane reactions from 300 to 400 K: An assessment of accuracy, J. Phys. Chem. A 1998, 102, 3121-3126.

(26) Ruscic, B.; Feller, D.; Dixon, D. A.; Peterson, K. A.; Harding, L. B.; Asher, R. L.; Wagner, A. F., Evidence for a lower enthalpy of formation of hydroxyl radical and a lower gas-phase bond dissociation energy of water, Journal of Physical Chemistry A 2001, 105, 1-4.

(27) Blanksby, S. J.; Ellison, G. B., Bond dissociation energies of organic molecules, Accounts of Chemical Research 2003, 36, 255-263.

(28) Berkowitz, J.; Ellison, G. B.; Gutman, D., Three Methods to Measure RH Bond-Energies, Journal of Physical Chemistry 1994, 98, 2744-2765.

(29) Bertram, A. K.; Ivanov, A. V.; Hunter, M.; Molina, L. T.; Molina, M. J., The Reaction Probablity of OH on Organic Surfaces of Tropospheric Interest, J. Phys. Chem. A 2001, 105, 9415-9421.

(30) D'Andrea, T. M.; Zhang, X.; Jochnowitz, E. B.; Lindeman, T. G.; Simpson, C. J. S. M.; David, D. E.; Curtiss, T. J.; Morris, J. R.; Ellison, G. B., Oxidation of organic films by beams of hydroxyl radicals, Journal of Physical Chemistry B 2008, 112, 535-544.

Page 30: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

________________

Chapter II

Single Particle Mass Spectrometry

________________

2.1 Introduction

Marine aerosols are examples of organic containing particles in the marine boundary

layer. The study of initial interfacial oxidation of organic marine aerosols by radicals is the main

objective in this thesis (see chapter I). The analysis of salt/water droplets coated with organic

surfactants is proposed using a unique droplet mass spectrometer. When studying aerosol

particles using single particle mass spectrometers, particles are usually dried and then analyzed.

Since our droplet spectrometer is designed to study water droplets without drying, this novel

instrument has specific requirements and design considerations.

The goal in this chapter is to describe the main components of single particle mass

spectrometry (SPMS). These spectrometers are used to study particles’ chemical composition

and transformation in the field. Also they are used in the laboratory setting to study fundamental

chemical and physical properties. Moreover, atmospheric processes and composition could be

studied at an individual particle level making this technique a very powerful tool. Brief

descriptions of the principal SPMS components are given in sections. More extensive reviews

on each SPMS component and historical review has been given previously by Johnston and

Wexler, Noble and Prather, Nash et al. and Murphy.1-4 At the end of the chapter the focus shifts

more towards two specific SPMS techniques5-10 that were considered for the design of our

droplet mass spectrometer. The spectrometer is a laser based instrument that implements droplet

vaporization by an infrared laser.

Page 31: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

14

2.2 Introduction to Single Particle Mass Spectrometry

Single Particle Mass Spectrometry has been the technique of choice to study and measure

atmospheric aerosol particles for the last two decades.1-4 It has replaced traditional collection

methods of aerosol particles on filters or substrates. The collection methods often involve

analysis by common analytical techniques such as: gas chromatography mass spectrometry,

liquid chromatography mass spectrometry, scanning electron microscopy, laser microprobe mass

spectrometry, atomic absorption spectroscopy, among others.2 SPMS is a big improvement to

traditional methods. The most important advantage is that it provides single particle chemical

and physical information in real time.11-16 This in-situ characterization has proved to be very

useful by avoiding ensemble averages from other methods and probing the constantly changing

atmospheric aerosol particles. It also does not require sampling on filters thus eliminating

interference from the substrate, solvent extraction and from mass loss/gain from volatile or

condensable components.

The main characteristic of SPMS is the continuous sampling of particles from the

atmosphere into the high vacuum (usually 10-6 Torr) environment of the mass spectrometer.

Figure 2.1 depicts the main components for most SPMS. The aerosol inlet is the interface

between atmospheric pressure and the SPMS high vacuum system. After particles are sampled -

nominally of µm and sub µm size - through the inlet, they are subsequently detected in the

particle detection and sizing region. The particle detection provides the triggering for

vaporization/ionization and data collection in the mass analysis region. Particles are then

vaporized into monomers and the monomers ionized in the vaporization/ionization region. The

produced ions are analyzed with mass spectrometers. Common mass spectrometers used for this

Page 32: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

15

technique are quadrupole (QMS) and Time-of-Flight mass spectrometers (TOFMS). This allows

for the determination of individual chemical components of single particles.

Figure 2.1 Schematic of main components of a Single Particle Mass Spectrometer

2.2.1 SPMS Inlets

Aerosol mass spectrometers sample particles though inlets from atmospheric pressure

into vacuum. The goal of SPMS inlets is to transmit particles efficiently into high vacuum (10-6

Torr) without changing their size and composition. Another desirable characteristic is that the

particle beam has a low divergence in vacuum. Inlets can be classified in three categories, (1)

orifices, (2) nozzles/capillaries (short and long) and (3) aerodynamic lenses.17-20 These inlets are

often followed by one or more differential pumping stages to separate the expanding gas from

the particle beam.

Figure 2.2 shows orifice and capillary designs.19 These types of inlets were originally

incorporated in SPMS instruments. Orifices are the simplest sampling inlets. Particles sampled

through a choked orifice into the vacuum chamber are focused into the ionization region of the

spectrometer. Its focal length is dependent on particle size, upstream pressure, orifice diameter,

Aerosol Inlet

Particle Detection and

Sizing

Vaporization Ionization

Mass Analysis

Vacuum

Atmosphere

Page 33: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

16

and density.4,20 This makes the orifice very ineffective in transmitting a broad particle size

range. Only a certain particle size is transmitted efficiently into the mass spectrometer for any

given set of instrument conditions (pressure, orifice diameter, etc.). However, orifices have two

advantages; (1) short particle residence time in the inlet which avoids evaporation of water or

volatile compounds and (2) the ability to focus different particle sizes by changing upstream

pressure, especially for particles smaller than 100 nm which are hard to detect by light scattering.

In contrast to orifices, capillaries sample and collimate a wider particle size range (sub-

micrometer to a few micrometers). Particle separation occurs at the entrance of the capillary

inlet. Therefore capillaries do not have particle size dependent focal points, increasing

substantially transmission efficiency of several size particles. Since particles travel parallel to

the capillary axis they spread radially, depending on size, due to free jet expansion into

vacuum.19,20 Capillaries produce a particle beam with smaller divergence angle at long distance

when compared to orifices making a reasonably collimated particle beam ( 1 mm). Capillaries

and orifices share the disadvantage of clogging. In capillaries this is accentuated by particle

growth by condensation and particle deposition on the inlet’s wall. The chances of clogging

increase when sampling in high water vapor environments. This results in changes in particle

size due to water condensation and increases instrument down time due to capillary inlet

clogging.

Page 34: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

17

Figure 2.2 Type of inlets for Single Particle Mass Spectrometry: (a) conical nozzle, (b) orifice, (c) nozzle with short capillary and (d) nozzle with long capillary. Particles are sampled from atmospheric pressure on the inlets’ open ended orifice going into vacuum (10-3 Torr – 10-6 Torr) through the 500 µm apertures. Adapted from reference 19.

Most modern SPMS adopt aerodynamic lens designs17,18,21 as inlets due to their high

transmission efficiency in the sub-micrometer particle size range and the tightly focused (less

than 1 mm in diameter) particle beam produced. See Figure 2.3 for an aerodynamic focusing

lens schematic. Working stagnation pressure for the aerodynamic lens inlet varies from 0.5-5

Torr with the most commonly used pressure of approximately 2 Torr. A choked orifice ( 100

µm in diameter) is placed at the aerodynamic inlet entrance. The working pressure of the

aerodynamic lens inlet is measured after the choked orifice. This orifice sets the working

pressure of the inlet by constricting the sampling flow to approximately 100 cm3 min-1. Inside an

Page 35: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

18

aerodynamic lens inlet a series of orifices with decreasing diameter act as constrictions. The

pressure drop and slight focus due to the orifices contract the particles into the centerline of the

inlet. Once in the centerline, particles accelerate into vacuum through a nozzle, orifice or small

capillary. The degree of acceleration depends on particle size, imparting particles with

differential velocities. The velocity partitioning allows the gathering of particle size information

as the vacuum aerodynamic diameter. Once in the vacuum, particle velocities range from 80-150

m/s.

Figure 2.3 Aerodynamic lens inlet schematic. A 100 µm orifice is placed at the aerosol inlet entrance to constrict the sampling flow to about 100 cm3 min-1. The pressure after the orifice is the aerodynamic lens inlet working pressure, usually 2 Torr. Adapted from reference 17.

2.2.2 Particle Detection and Sizing

Sensitive particle detection becomes important when conducting single particle

measurement and analysis. Due to atmospheric particle concentrations and inlet transmission

efficiency, detection and sizing of particles improves substantially the SPMS duty cycle.

Essentially it allows the synchronization of particle ionization and mass analysis. Light

scattering detection of particles is the most commonly used technique. The regime for particle

light scattering depends on the dimensionless22 size parameter, α

Page 36: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

19

λ

d π α p (2.1)

where dp is the particle diameter and λ is the scattering light wavelength. For α <<< 1 the

particle is subjected to an almost uniform electromagnetic field. The particle oscillates like a

dipole with polarization proportional to the electric field which is the Rayleigh approximation.22

The light scattering intensity (I) of this particle is proportional to the polarizability, p, by;

4

6p2

λ

d I p (2.2)

Light scattering from particles with diameters of the same order as the incident light (α

1) is described by Lorenz-Mie theory23. This applies for homogeneous sphere particles with

diameters from 0.1 µm to 1 µm. Rayleigh scattering does not apply because the local electric

field is no longer uniform at any instant. In the Lorenz-Mie regime there is strong interaction

between the particle and the incident light. There is no simple relation for light scattering

intensity and particle diameter. However, light scattering intensity is proportional to (diameter)6

(I d6) for a given laser intensity and α 1. This makes slight improvements in the particle

diameter detection limit. Intensity falls off rapidly when particle size decreases. The use of

smaller wavelengths will improve detection of particles with diameters less than or equal to 100

nm since they are closer to the λ-4 Rayleigh scattering limit.4 A more detailed description may be

found in chapters 15-16 of Baron and Willeke (2001).

In practice, signal-to-noise constraints will limit light scattering detection and collection.

The design limitations are laser intensity, laser focus, background light detected by the photo-

detector coming from laser windows, and chamber reflections. Painting the chamber black and

using neutral density filters helps with the background scattered light improving signal-to-noise.

The type of ionization for the SPMS will determine the particle detection and sizing parameters

Page 37: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

20

to be utilized. For example, laser systems used for ionization are triggered prior to particle

arrival at the interaction region. This facilitates hitting the particle in space and time with the

laser beam. Therefore complete knowledge of particle velocity is critical for laser ionization

techniques.

Scattered light from particles is detected by photo-detectors such as avalanche

photodiodes and more commonly used photo-multiplier tube (PMT) detectors. These detectors

have been used to collect scattered light using fiber optics with the detector at 90º angle to the

laser beam. Particles of a nominal µm size scatter light mostly in the backward and forward

direction. Collection and detection of scattered light at 30° has been implemented in some

SPMS designs.12,24-26 In this setup fiber optics and optic lens systems have been used to improve

the light collection and detection. Furthermore the use of elliptical mirrors at the particle-laser

interaction has been shown to improve particle size limit detection and efficiency by collecting

scattered light at all solid angles and focusing it at the detector.16

Continuous wave (CW) visible lasers are used to scatter light off particles for detection.

Helium Neon (HeNe) lasers at 632 nm, doubled Neodymium-doped Yttrium Aluminum Garnet

(Nd:YAG) output, or diode lasers at 532 nm are all used. One adaptation is to intersect particles

with one CW laser beam and let the scattered light trigger the ionization laser.13,14 This scheme

only works with lasers that could be triggered very fast (in less than 1 microsecond). Excimer

lasers are good examples since they can be triggered in approximately 500 ns. If particles are

travelling as fast as 250 m/s, they only move approximately (500 x 10-9 s)(250 m/s) or 125 µm

during the excimer triggering sequence, which is well within the ionization laser beam area.

More commonly the particles pass through two CW laser beams12,16 located at a known

distance apart. Scattered light generated by a particle is collected in each laser stage by a PMT

Page 38: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

21

providing the particle’s transit time. This transit time depends on particle size and is used to

trigger vaporization/ionization laser(s) and data acquisition. If the sampling inlet differentially

accelerates particles, size information could be obtained upon instrument calibration with known

particle sizes. The two laser configuration is necessary when using any harmonic of a Nd:YAG

laser for vaporization/ionization. This is because the YAG laser needs to be triggered

approximately 200 µs prior to particle arrival at the vaporization/ionization region of the

spectrometer.

The commercial Aerodyne Aerosol Mass Spectrometer incorporates a chopper instead of

lasers.15 The modulated particle beam travels up to a hot plate/ionization region, with different

sizes arriving at different times. The chopper frequency and ion detection are synchronized for

data collection. The ion detection at the mass spectrometer provides the signal for particle

arrival and finally particle transit time in vacuum. Since the Aerodyne Aerosol Mass

Spectrometer does not incorporate lasers, the detection of particles smaller than 100 nm is

possible.

2.2.3 Particle Vaporization and Ionization

Particle chemical characterization depends on efficiently vaporizing and ionizing

particles for mass analysis. This in turn depends on the particle detection and sizing technique to

be used as described in the previous section. Also the selection of the ionization method is

influenced by the type of aerosol particles to be studied, particle composition and laser

wavelength. Ultra-Violet (UV) laser vaporization/ionization is the most efficient for the

detection of salts and metals. For the detection of organic molecules, UV ionization, electron

impact (EI) ionization and chemical ionization (CI) methods have been used.3

Page 39: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

22

Vaporization/ionization of particles into monomers is accomplished by either striking

particles on a hot metal surface or by vaporization/ionization with a laser. The hot plate

configuration combined with EI ionization for subsequent detection with a QMS is implemented

in the Aerodyne Aerosol Mass Spectrometer.15 EI has the attribute of using 70 eV electrons

which allows comparison to standard libraries for compound identification. However,

atmospheric particles vary significantly in composition and therefore EI presents a challenging

task in identifying species, especially organic molecules. Despite this challenge use of the

Aerodyne spectrometer has provided useful information on atmospheric particles like total

organic mass and degree of oxidation.

There are two common configurations for laser based instruments. The simplest is to use

a UV laser (i.e. 193 nm (6.4 eV) from an excimer laser or 266 nm (4.7 eV) from 4th harmonic of

Nd:YAG) to simultaneously vaporize and ionize particles with the same laser pulse. This

process is often called ablation.12,13,16 UV ionization laser energy is on the order of 10 mJ/pulse.

A 10 nanosecond pulse focused to 1 mm spot size will have an area of 7.3 x 10-3 cm2. This

results in a laser fluence of (10 x 10-3 J/(7.3 x 10-3 cm2)(1 x 10-8 s)) or 3.8 x 108 W/cm2 where

1J/s = 1 W. The vaporization/ionization at 193 nm or 266 nm is a multi-photon process because

ionization energy of the majority of organic molecules is approximately 9-10 eV. Generally

laser ablation produces a large flux of ions and when combined with TOFMS it becomes a very

sensitive technique. The generation of ion signal is dependent on the laser power or fluence.

Laser fluence is in the range of 108-1010 W/cm2. Lower laser fluence allows for a bigger laser

spot size facilitating particle and laser beam alignment. Also it avoids the bias from particle

shape and minimizes molecular fragmentation.

Page 40: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

23

Laser induced ion formation as a function of laser fluence has been studied27 for three

different UV laser lines; 248 nm, 193 nm and 157 nm. As the laser fluence was scanned, the

formation of ions from particles was monitored with a TOFMS. Laser fluence above ion

formation threshold produces more ion signal from particles. It showed that laser fluence

required to produce ions decreases as the laser energy increases from 248 nm to 157 nm. The

increase in wavelength energy for ion formation threshold, also narrows the gap to detect

different particle types.

A more sophisticated configuration is to use two lasers to separate the vaporization and

ionization steps. A powerful (about 500 mJ/pulse) CO2 (λ = 10.6 µm) or a Nd:YAG pumped

Optical Parametric Oscillator (OPO) (λ around 3 µm, 10 mJ/pulse) laser is used to evaporate the

particle into a plume and then ionize the resulting gas components with the UV laser16,24,28,29 (i.e.

193 nm). Figure 2.4 shows how useful is this method when using Single Particle Laser Ablation

Time-of-Flight Mass Spectrometry.16 This study was done with pure succinic acid

(HOOCCH2CH2COOH, m/z = 118) particles generated in the laboratory. Spectra in (a) are from

ablation of different particle sizes with a 193 nm excimer laser. Signal-to-noise is very good

showing the sensitivity of the technique but spectra from one particle size to another is not

reproducible. On the other hand, when the CO2 IR laser is used prior to ionization with the UV

laser (e.g. spectra in (b)), a much better reproducibility is achieved between different particle

sizes. Note that succinic acid parent peak at m/z 118 is not detected in the mass spectra.

Identifications of the species are very difficult if they arre done by the fragmentation pattern.

Page 41: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

24

Figure 2.4 Mass spectra of individual succinic acid particles. Spectra in (a) were generated by laser ablation with excimer laser UV radiation at 193 nm and in (b) by IR laser (10.6 µm) vaporization followed by ionization with UV radiation at 193 nm. Adapted from reference 16.

2.2.4 Mass Analysis

Mass analysis in SPMS is most commonly accomplished by the use of TOFMS,

quadrupole and ion trap mass spectrometers. The Aerodyne spectrometer uses an EI source

coupled with a QMS. EI mass spectra are directly proportional to the particle’s gas phase

components that are volatile at the hot plate temperatures. TOF and ion traps are multiplexed

instruments. Therefore acquisition of a complete mass spectrum from an individual particle is

possible. Moreover, the negative and positive ion spectra could be collected simultaneously for a

Page 42: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

25

single particle when using dual TOF as is the case for the commercially available TSI Inc.

Aerosol TOFMS. Ion traps have this same advantage as the ATOFMS plus collision induced

dissociation capabilities since ions could be trapped for long periods of time with subsequent

detection of all ions. Size information with chemical composition could be coupled allowing for

a more complete characterization of aerosol particles.

2.3 Photoionization

Development of SMPS instruments in recent years has focused more on the use of

aerodynamic lens inlets and lasers for detection and vaporization/ionization in combination with

TOFMS. Examples are Single Particle Laser Ablation Time-of-Flight Mass Spectrometer,16

Aerosol Time-of-Flight Mass Spectrometer,12 Particle Analysis by Laser Mass Spectrometry13

and others.3,4,12

The use of UV lasers makes SPMS very useful for laboratory and field studies due to the

chemical characterization potential. But chemical speciation, especially for particle’s organic

material content, has been impossible. This is primarily due to extensive fragmentation due to

the high laser power required for ionization. The UV laser pulses used in SPMS are 4.7 eV (266

nm), 5 eV (248 nm), 6.4 eV (193 nm) and 7.9 eV (157 nm). Thus, when using UV laser pulses

particles absorb multiple photons at very high power or laser fluence making the spectra

complicated to interpret due to fragmentation. Figure 2.4 shows the extensive fragmentation of

pure succinic acid particles generated in the laboratory. Even though the two step laser IR

vaporization/UV ionization minimizes fragmentation and gives better reproducibility,

fragmentation is inevitable. Spectra are very convoluted so identification of individual organic

species becomes problematic in the laboratory and even more in field measurements.

Page 43: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

26

2.4 Droplet Spectrometer Objective

The main purpose of our design is to be able to study the gas-particle interfacial oxidation

focusing on the organic coating on salt/water droplets. This approach is to attempt to understand

the first steps in the oxidation of organic marine aerosols. Therefore in order to study this

chemistry we have built a droplet mass spectrometer in which no ionization step is required, only

the vaporization step.

The object is not to design a spectrometer to study all type of particles. The target

particles we want to study are µm sized water droplets coated by organic surfactants. A laser

based instrument was designed for this purpose. An infrared (IR) laser is used to vaporize the

aqueous particle or droplet. This main idea is shown in Figure 2.5. Surfactants (RCO2-) coating

the particle are ions at the droplet’s surface and in solution. Vaporization of the droplet releases

the ionic surfactants into the gas phase. Subsequent analysis of ions will be done with a mass

spectrometer. Essentially water in the droplet is used as a Matrix Assisted Laser Desorption

Ionization (MALDI) matrix to absorb all the laser energy. A major goal for this droplet

spectrometer is to detect individual molecular organic species from the droplet surface. This will

test the model30 for nascent organic marine aerosols (see Chapter I).

Page 44: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

27

Figure 2.5 Droplet vaporization by an IR laser pulse. R-CO2- represents surfactant coating

where R is a long alkyl chain.

2.5 Previous Work

The vaporization technique to desorb ions from solution has been used previously.

Brutschy’s group has been using the technique to study bio-molecules and non covalent

complexes. The technique is called Laser Induced Liquid Beam (or Bead)

Ionization/Desorption6,7,10 (LILBID) mass spectrometry and is depicted in Figure 2.6. In

summary, a continuous liquid jet containing the species of interest in solution is sent into a high

vacuum chamber (10-6 Torr) through a 10 µm aperture. The solvents used in the liquid jet are

water and small alcohols such as methanol, ethanol or propanol. The continuous liquid jet

strikes a liquid nitrogen cold trap which helps maintain the pressure at 10-6 Torr. An IR laser

tuned at the absorption frequency of the solvent intersects the liquid beam and the analytes are

desorbed into the gas phase. The desorption is done by using a custom tunable Nd:YAG pumped

OPO with energies of 10 mJ/pulse or a CO2 laser at 10.6 µm. The laser beam is focused mildly

to a 1 mm spot size yielding laser fluences of 106-108 W/cm2. Ions are then accelerated into a

reflecting TOF for mass analysis.

Page 45: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

28

Figure 2.6 LILBID mass spectrometer schematic. Adapted from reference 9.

The technique has also been implemented with micro-droplet instead of liquid jets.

Nominally the micro-droplets size with diameter of 10 µm.8 By using micro-droplets with an on

demand generator the duty cycle improves to 1. The LILBID technique has shown6-8,10 that big

molecules (i.e. proteins) can be desorbed with minimal or no fragmentation from water, buffer or

alcohol solutions. This is done in the presence of small quantities of species, on the order of 10-5

M. The 10 µm droplet has a volume of 5.2 x 10-13 L, therefore it contains (10-5 M)(5.2 x 10-13

L)(6.02 x 1023 molecules/mol) or 3.2 x 106 molecules to be analyzed.

Page 46: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

29

Clustering of ions with the solvent presents a problem when desorbing ionic species from

solution in vacuum. Binding energy30 of carboxylates (RCO2-) to H2O clusters is approximately

20 kcal mol-1. The ion-water clusters will make identification of individual molecular species

more difficult. This can be demonstrated by previous LILBID studies of metals in solution.9

The spectra show extensive water clustering as seen in Figure 2.7. Li+ ions are detected

following desorption from a water liquid jet. An envelop of peaks separated by m/z = 18 shows

Li+(H2O)n water clusters with not much signal from the bare Li+ ion. There are two possible

explanations for the water clustering in this study. The first one is that metal ions are solvated in

the liquid and desorbed with a solvent shell into the gas phase. The other one is that water binds

to the bare metal ions once they are desorbed into the gas phase. Water clustering to ions may

present a problem when using the LILBID technique for atmospheric aerosol particles, especially

organic marine aerosols. The chemistry at the particle interface and interpretation gets more

complicated with water clustering. Differentiation between water cluster peaks from surfactants

or surfactant-radical reaction products could be masked. This is something to consider when

designing an SPMS as the droplet spectrometer proposed here, where the aerosol model has

more water molecules than organic surfactants.

Page 47: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

30

Figure 2.7 Spectrum of Li+ ions desorbed using LILBID technique from an aqueous liquid jet. Adapted from reference 9.

Baer and co-workers developed an aerosol time-of-flight mass spectrometer to study

atmospherically relevant particles in a laboratory setting.24,26,31 Like most SPMS described in

previous sections (see Figure 2.8), their spectrometer uses an aerodynamic lens inlet to sample

particles into high vacuum. Once in vacuum particles intersect two continuous green (532 nm)

diode laser beams separated by 10 cm. The scattered light is collected by PMTs. The PMTs

signal is used to measure particle size and trigger the vaporization/ionization lasers. The two

laser vaporization/ionization configuration was adopted. The IR vaporization laser used is a CO2

laser at 10.6 µm that was kept at high laser power (500 mJ/pulse). This ensured rapid and

complete vaporization of particles. The ionization UV laser (193 nm) is fired following the IR

laser to ionize the resulting plume from vaporization. The IR CO2 and UV excimer lasers were

Page 48: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

31

focused to approximately 1 mm and 2 mm, respectively. Ions are accelerated either into a 15 cm

or 1 m long drift region of TOFMS.

Figure 2.8 SPMS from reference 32. The spectrometer consists of an aerodynamic lens inlet, two differential pumping chambers, two diode lasers and two PMTs for light scattering detection and CO2 and UV lasers for vaporization and ionization, respectively.

For desorption experiments the UV laser was not used. The purpose of the experiment

was to detect the ions present in solution only by using IR desorption from liquid particles.

Ethylene glycol droplets containing RbC were used for the vaporization experiments.5 The

droplet size was kept constant at a diameter of about 4.1 µm and only the RbC concentration

was varied. In order to measure and establish total ion signal, total ion current was collected on

the TOF repeller plate shown in Figure 2.8. An operational amplifier (op-amp) is used measure

the incident current at the repeller plate. The op-amp works in a current to voltage configuration

and has a gain of 390,000 V/A. The field between the repeller and acceleration plates is 200

Page 49: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

32

V/cm. This field and the repeller size (7.5 cm) ensured that no ions were lost due to the droplets’

initial kinetic energy. Positive ion mass spectra and total negative ion output voltage were

collected simultaneously. Integrated output voltage was then converted to the number of ions

detected and compared to the expected number of ions based on the RbC concentration.

Figure 2.9 shows the positive ion signal as a function of RbC concentration.5 Integrated

signal increases as the RbC concentration increases. At a concentration of approximately 10-4

M the Rb+ signal levels off while the appearance of clusters like Rb2C+ becomes relevant.

Moreover the total ion signal measured as negative ions at the repeller, decreases from 100%

detected at 3 x 10-5 M to 0.03% at the higher concentration of 3 x 10-1 M. The evident signal lost

in TOF spectra and absolute ion count is attributed to ion-ion recombination. This quantitative

experiment showed how important ion-ion recombination is for laser desorption experiments.

This is more evident at higher salt concentrations and becomes a limitation on ion signal in the

spectrometer. This constrains the droplet spectrometer in terms of how much ion signal could be

detected from a single organic marine aerosol.

Page 50: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

33

Figure 2.9 Integrated Rb+ ion signal as a function of RbC concentration as measured with the TOFMS after IR laser vaporization. Na+ signal comes from impurities in solution. Adapted from reference 5.

Page 51: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

34

References for Chapter II

(1) Johnston, M. V.; Wexler, A. S., On the Cover- MS of Individual Aerosol-Particles, Analytical Chemistry 1995, 67, A721-A726.

(2) Noble, C. A.; Prather, K. A., Real-time single particle mass spectrometry: A historical review of a quarter century of the chemical analysis of aerosols, Mass Spectrometry Reviews 2000, 19, 248-274.

(3) Nash, D. G.; Baer, T.; Johnston, M. V., Aerosol mass spectrometry: An introductory review, International Journal of Mass Spectrometry 2006, 258, 2-12.

(4) Murphy, D. M., The design of single particle laser mass spectrometers, Mass Spectrometry Reviews 2007, 26, 150-165.

(5) Dessiaterik, Y.; Nguyen, T.; Baer, T.; Miller, R. E., IR vaporization mass spectrometry of aerosol particles with ionic solutions: The problem of ion-ion recombination, Journal of Physical Chemistry A 2003, 107, 11245-11252.

(6) Kleinekofort, W.; Avdiev, J.; Brutschy, B., A new method of laser desorption mass spectrometry for the study of biological macromolecules, Int. J. Mass Spectrom. Ion Process. 1996, 152, 135-142.

(7) Kleinekofort, W.; Pfenninger, A.; Plomer, T.; Griesinger, C.; Brutschy, B., Observation of noncovalent complexes using laser-induced liquid beam ionization desorption, Int. J. Mass Spectrom. Ion Process. 1996, 156, 195-202.

(8) Morgner, N.; Barth, H. D.; Brutschy, B., A new way to detect noncovalently bonded complexes of biomolecules from liquid micro-droplets by laser mass spectrometry, Aust. J. Chem. 2006, 59, 109-114.

(9) Sobott, F.; Wattenberg, A.; Barth, H. D.; Brutschy, B., Ionic clathrates from aqueous solutions detected with laser induced liquid beam ionization/desorption mass spectrometry, Int. J. Mass Spectrom. 1999, 187, 271-279.

(10) Wattenberg, A.; Sobott, F.; Brutschy, B., Detection of intact hemoglobin from aqueous solution with laser desorption mass spectrometry, Rapid Commun. Mass Spectrom. 2000, 14, 859-861.

(11) Mckeown, P. J.; Johnston, M. V.; Murphy, D. M., Online Single-Particle Analysis by Laser Desorption Mass-Spectrometry, Analytical Chemistry 1991, 63, 2069-2073.

Page 52: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

35

(12) Noble, C. A.; Nordmeyer, T.; Salt, K.; Morrical, B.; Prather, K. A., Aerosol Characterization Using Mass-Spectrometry, Trac-Trends in Analytical Chemistry 1994, 13, 218-222.

(13) Thomson, D. S.; Murphy, D. M., Analyzing Single Aerosol-Particles in Real-Time - Composition Analysis of Aerosols Is Critical to an Understanding of Their Origins and Fate - This Is Done by Combining Time-of-Flight Mass-Spectroscopy with Optical-Particle Detections and Sizing, Chemtech 1994, 24, 30-35.

(14) Murphy, D. M.; Thomson, D. S., Laser Ionization Mass-Spectroscopy of Single Aerosol-Particles, Aerosol Science and Technology 1995, 22, 237-249.

(15) Jimenez, J. L.; Jayne, J. T.; Shi, Q.; Kolb, C. E.; Worsnop, D. R.; Yourshaw, I.; Seinfeld, J. H.; Flagan, R. C.; Zhang, X.; Smith, K. A.; Morris, J. W.; Davidovits, P., Ambient aerosol sampling using the Aerodyne Aerosol Mass Spectrometer, J. Geophys. Res. 2003, 108, 8425.

(16) Zelenyuk, A.; Imre, D., Single particle laser ablation time-of-flight mass spectrometer: An introduction to SPLAT, Aerosol Science and Technology 2005, 39, 554-568.

(17) Liu, P.; Ziemann, P. J.; Kittelson, D. B.; McMurry, P. H., Generating particle beams of controlled dimensions and divergence .1. Theory of particle motion in aerodynamic lenses and nozzle expansions., Aerosol Science and Technology 1995, 22, 293-313.

(18) Liu, P.; Ziemann, P. J.; Kittelson, D. B.; McMurry, P. H., Generating particle beams of controlled dimensions and divergence .2. Experimental evaluation of particle motion in aerodynamic lenses and nozzle expansions., Aerosol Science and Technology 1995, 22, 314-324.

(19) Mallina, R. V.; Wexler, A. S.; Johnston, M. V., Particle growth in high-speed particle beam inlets, Journal of Aerosol Science 1997, 28, 223-238.

(20) Mallina, R. V.; Wexler, A. S.; Johnston, M. V., High-speed particle beam generation: Simple focusing mechanisms, Journal of Aerosol Science 1999, 30, 719-738.

(21) Jayne, J. T.; Leard, D. C.; Zhang, X. F.; Davidovits, P.; Smith, K. A.; Kolb, C. E.; Worsnop, D. R., Development of an aerosol mass spectrometer for size and composition analysis of submicron particles, Aerosol Science and Technology 2000, 33, 49-70.

(22) Gebhart, J. In Aerosol Measurement: Principles, Techniques, and Applications; 2nd ed.; Willeke, K., Baron, P. A., Eds.; John Wiley & Sons, Inc.: New York City, 2001.

Page 53: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

36

(23) Rader, D. J.; J., O. H. T. In Aerosol Measurement: Principles, Techniques, and Applications; 2nd ed.; Willeke, K., Baron, P. A., Eds.; John Wiley & Sons, Inc.: New York City, 2001.

(24) Cabalo, J.; Zelenyuk, A.; Baer, T.; Miller, R. E., Two-color laser induced evaporation dynamics of liquid aerosols probed by time-of-flight mass spectrometry, Aerosol Science and Technology 2000, 33, 3-19.

(25) Nordmeyer, T.; Prather, K. A., Real-Time Measurement Capabilities Using Aerosol Time-of-Flight Mass-Spectrometry, Analytical Chemistry 1994, 66, 3540-3542.

(26) Zelenyuk, A.; Cabalo, J.; Baer, T.; Miller, R. E., Mass spectrometry of liquid aniline aerosol particles by IR/UV laser irradiation, Analytical Chemistry 1999, 71, 1802-1808.

(27) Thomson, D. S.; Middlebrook, A. M.; Murphy, D. M., Thresholds for laser-induced ion formation from aerosols in a vacuum using ultraviolet and vacuum-ultraviolet laser wavelengths, Aerosol Science and Technology 1997, 26, 544-559.

(28) Sykes, D. C.; Woods, E.; Smith, G. D.; Baer, T.; Miller, R. E., Thermal vaporization-vacuum ultraviolet laser ionization time-of-flight mass spectrometry of single aerosol particles, Analytical Chemistry 2002, 74, 2048-2052.

(29) Dessiaterik, Y.; Baer, T.; Miller, R. E., Laser ablation of imidazolium based ionic liquids, Journal of Physical Chemistry A 2006, 110, 1500-1505.

(30) Ellison, G. B.; Tuck, A. F.; Vaida, V., Atmospheric processing of organic aerosols, Journal of Geophysical Research-Atmospheres 1999, 104, 11633-11641.

(31) Woods, E.; Miller, R. E.; Baer, T., The internal energy of neutral ethylene glycol molecules created in the laser vaporization of aerosol particles, Journal of Physical Chemistry A 2003, 107, 2119-2125.

Page 54: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

________________

Chapter III

Droplet Mass Spectrometer Design

________________

3.1 Introduction

We have built a droplet mass spectrometer to study the air-particle interfacial chemistry

of marine organic aerosols. The spectrometer is designed to study aqueous particles as a model

for marine aerosols. The model can be studied by following the qualitative experimental

overview in Figure 3.1. The marine organic aerosols are predicted to be salt/water droplets

encapsulated by organic surfactants.1 A 1 µm droplet contains approximately 1010 water

molecules and 107 surfactant molecules. Water droplets are generated from a dilute solution

containing an inorganic salt and organic surfactant salt. The inorganic salt of choice is NaNO3

and surfactant of choice is sodium dodecyl sulfate (SDS) because of their well known physical

properties.2-5 After generation, aerosols then enter a flow tube at atmospheric pressure and

temperature, and containing N2, O2 at atmospheric concentrations, to react with radicals such as

hydroxyl radical (OH) or chlorine radical (C). These radicals abstract a hydrogen atom from

the organic surfactant to form an alkyl radical (R•). Since the flow tube is at atmospheric

pressure, the alkyl radicals are likely to combine with molecular oxygen to form an alkyl peroxyl

radical (ROO•).

The main objective is to study the initial radical chemistry on organic aerosols at the air-

particle interface. This is to test the “inverted micelle” marine aerosol model1 proposed by

Ellison et al. in 1999 (see Figure 3.2). The model suggests that organic surfactants at the

droplet’s interface will inhibit particle growth by uptake of H2O and therefore suppress cloud

Page 55: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

38

formation. However, surfactant reactions with atmospheric oxidants will produce polar groups

(aldehydes, ketones, carboxylic acids, etc.) at the droplet’s surface. This atmospheric processing

makes the surface more hydrophilic and is conjectured to make organic marine aerosols better

cloud condensation nuclei.

To study this atmospheric processing, aerosols are sampled into the droplet mass

spectrometer through a capillary inlet following reaction with hydroxyl (OH) or chlorine (C)

radical in atmospheric conditions. The droplet spectrometer is a laser based spectrometer

coupled to a quadrupole mass spectrometer (QMS). In the droplet spectrometer aerosol particles

intersect an IR laser beam tuned at the peak absorption for bulk water, 2.94 µm or 3400 cm-1.

Organic lipids or surfactants from marine life exist predominantly as ions at the ocean surface.

Consequently the “inverted micelle” model predicts that the surfactants are naturally ions at the

droplet’s surface. Vaporization of the aqueous core with an IR laser will release the surfactant

ions into the gas phase. Ionization steps such as photoionization or electron impact are not

required in the droplet spectrometer instrument. Subsequent extraction and detection with a

commercial QMS is used for mass analysis.

Page 56: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

39

Fig

ure

3.1

E

xper

imen

t sch

emat

ic d

iagr

am.

Page 57: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

40

Fig

ure

3.2

O

xida

tion

of

orga

nic

surf

acta

nt la

yer

on m

arin

e or

gani

c ae

roso

ls.

Page 58: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

41

3.2 Aerosol Generation

The target droplets are predicted to be an “inverted micelle” as shown in Figure 3.2.

Aerosols are produced from a suitable source in the laboratory to serve as a proxy for organic

marine aerosol. To test the “inverted micelle” model droplets of known composition and size are

generated. Organic surfactants cover the salt/water droplets where the hydrophobic part is

exposed to the atmosphere and the polar head is at the droplets’ surface-bulk interface. NaNO3

and sodium dodecyl sulfate (SDS) are chosen for the inorganic salt inside the droplet and organic

surfactant components, respectively. Their physical properties are well known and

characterized.2-5 Descriptions of these physical properties along with aerosol generation follows

in the next sections. The droplets are generated from a dilute NaNO3/SDS solution. The

droplets are then dried and sent to a humidity chamber to grow by water uptake. Since the

physical properties of water uptake in NaNO3/SDS aqueous particles are known, final droplet

size can be predicted precisely and controlled in the laboratory.

3.2.1 Inorganic Salts

Understanding inorganic salt aerosol physical properties is crucial not only to better

describe them in atmospheric models, but also to generate and characterize aerosols in a

laboratory setting. Inorganic salt aerosols undergo phase changes depending on the

environment’s water vapor content or humidity.6 Some common inorganic salt particles grow

spontaneously taking up water when exposed to high humidity environments in a process known

as deliquescence. The particle becomes a saline solution. Once the particle is deliquesced it

continues to grow if the relative humidity (RH) increases. If the humidity decreases, the same

liquid droplet will lose water rapidly after passing its deliquescence RH. This will make the

Page 59: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

42

liquid droplet a super saturated solution until it reaches its efflorescence point. At this point the

particle loses all the water spontaneously and crystallizes to become a solid particle.

The phase diagram with deliquescence and efflorescence points for different sulfates,

sodium nitrate (NaNO3), sodium chloride (NaC) and sea salts has been measured previously.2-4,7

The experiments were done using an electrostatic balance. A salt particle was levitated in the

center of the electrostatic balance and the water vapor was increased in the chamber. The

particle mass change was measured and recorded as a function of RH. The mass change was

measured as the ratio (m/mo) of mass of the droplet (m) to the mass of pure salt particle (mo).

This mass ratio is a measurement of the salt weight fraction (WF). Weight fraction is defined as

the ratio of mass of the salt to mass of the droplet (msalt/mdroplet) or (m/mo)-1 from Figure 3.3.

Figure 3.3 shows the phase change diagram for ammonium sulfate ((NH4)2SO4) particle. As the

RH increases the particle mass does not change until it reaches the deliquescence point at 80%

RH and the particle becomes a droplet. As the water content keeps increasing the particle gains

more mass. Then this droplet is exposed to lower RH by evacuating the chamber while still

measuring the particle mass change. (NH4)2SO4 droplet reaches the efflorescence point between

37-40% RH.

Page 60: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

43

Figure 3.3 Phase diagram transformation, growth () and evaporation () for a (NH4)2SO4 particle as a function of RH. Red arrow indicates efflorescence point of 37-40% RH and blue arrow indicates deliquescence point of 80% RH. Adapted from reference 3.

The water activity (i.e. RH) as a function of salt weight fraction was derived from the

measurements.3 Empirical formulas for the RH as a function of weight fraction (WF) were

obtained by fitting all data points for a particular inorganic salt. Empirical formulas are known

for the droplet density as a function of solute weight fraction for a variety of inorganic salts.3

The empirical formulas for RH and droplet density (drop) are as follows;

100 ) wfC (1.0 RH ii (3.1)

iidrop wfA 0.9971 ρ (3.2)

Page 61: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

44

where Ci and Ai are the empirically derived polynomial coefficients, RH is the relative humidity,

drop is the drop density and WF is the solute weight fraction. With these empirical formulas the

droplet’s solute weight fraction and droplet density can be calculated as a function of RH.*

Ultimately growth factors (GF) can be calculated for a given RH and inorganic salt composition.

GF is defined as the ratio (ddrop/dsalt) of the diameter of the droplet (ddrop) to the diameter of the

dry salt particle (dsalt) after the droplet has been dried completely of all water. Knowing the

droplet density and salt weight fraction as a function of RH the growth factors can be calculated.

The dimensional analysis to obtain the growth factor is as follows:

salt

drop3drop3

saltsalt ρ

WFρ d

6

π d

6

π V (3.3)

where Vdry is the volume of the dry salt particle, ddry is the dry salt particle diameter, ddrop is the

droplet diameter, drop is the calculated droplet density, WF is the salt weight fraction and salt is

the pure salt density.

Rearranging equation 3.3 to obtain growth factors gives:

3

1

drop

salt

salt

drop

wfρ

ρ

d

d GF

(3.4)

Using these equations inorganic salt particle growth factors can be calculated as a

function of RH. By knowing the RH in a controlled environment the pure inorganic particles’

composition as weight fraction could be obtained. Also the RH in a controlled humid

environment can be measured by measuring the growth factor of a pure inorganic salt particle.

*NaNO3 C1-C4 coefficients for eq. 3.1 are -5.52 x 10-3, 1.286 x 10-4, -3.496 x 10-6 and 1.843 x 10-8, respectively. Coefficients A1-A3 for eq. 3.2 are 6.512 x 10-3, 3.025 x 10-5 and 1.437 x 10-7, respectively. A droplet of NaNO3 with WF of 0.33 exposed to an RH of 86% will have a density of 1.25 g cm-3.

Page 62: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

45

3.2.2 Aerosol Chemical Components in the Laboratory

Physical properties of NaNO3 aqueous particles are well known.3 NaNO3 particles

deliquesce at 75% RH and effloresce at a range of 0.05-30% RH. It is known that a NaNO3

particle does not crystallize easily and depends on other impurities in the particle for it to

undergo the complete phase transition to solid. This water retention by NaNO3 means that

particles remain spherical even at low humidity. For that reason NaNO3 is chosen as the

inorganic salt to generate the droplets that represent marine aerosols. By knowing all physical

properties of this inorganic salt, composition and size for the laboratory generated aerosols can

be controlled.

Figure 3.4 shows a plot of equation 3.4 as a function of RH for NaNO3 particles. The

growth factor increases as the RH increases, as expected. The GF can be obtained from this kind

of plot and is used in the laboratory to generate droplets of a known size containing NaNO3 at a

given RH. Notice that only RH points larger than the efflorescence point are plotted due to the

fact that particles are solid below that point. Dry particles will grow only at RH higher than the

deliquescence point of 75 % RH for NaNO3.

Page 63: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

46

Figure 3.4 NaNO3 growth factor as a function of relative humidity (RH).

Sodium dodecyl sulfate (NaO3SO(CH2)11CH3) is the surfactant of choice to generate

coated aqueous NaNO3 droplets. SDS has been used before to model organic coatings in marine

aerosols. The kinetics and chemical properties of SDS particles have been studied.5,8-10

However the physical properties of SDS/H2O particles are not well known such as phase

diagrams with efflorescence and deliquescence points. It has been observed that particles

containing SDS and other surfactants grow when exposed to high relative humidity. The growth

rate is affected by the surfactant coating. Zelenyuk et al. measured growth factors at 86% RH

and density in detail on SDS salt particles (i.e., NaNO3, NaC) as a function of the surfactant

concentration.5 This study provided a physical characterization of surfactant coated salt/water

Page 64: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

47

particles. It demonstrated that these particles retain water and that the physical properties of the

particles are a complex function of surfactant concentration.

Using the NaNO3 and SDS physical properties such as growth factors we have devised

two ways to produce 1 µm surfactant coated aqueous droplets. These are shown in Figure 3.5.

The use of a bubbler and a vibrating orifice aerosol generator is described.

3.2.3 Bubbler

The basic idea to generate aerosols of known composition and size is shown in Figure

3.5. In Figure 3.5 (A), droplets are generated by bubbling air through a dilute salt solution of

pure NaNO3, or NaNO3/SDS mixture. The poly-dispersed aerosol flow produced by this method

is then dried of all water by passing the flow through a diffusion dryer. Aerosols are size

selected (by diameter) with a Differential Mobility Analyzer (DMA). Particles become charged

in the process of aerosol formation. The DMA is capable of selecting different particles sizes by

applying a high electrostatic potential to the particle flow.11,12 Charged particles of different

sizes have different mobilities in an applied electric field. The response in particle size is

monotonic with the electric field. Consequently by setting the field voltage to an appropriate

value, only one particle size is selected. The DMA works at relatively high voltage (up to 10

kV), and it tends to discharge when working with high relative humidity particle flow. The dry

mono-disperse aerosol flow is then sent to a humid chamber where the RH is controlled by

mixing the dry aerosol flow with humid flow. In the humid chamber the particles take up water

and grow based on their GF. For example, a 0.30 L min-1 dry (0% RH) particle flow combined

with a 1.7 L min-1 humid (100% RH) flow will roughly produce a 2 L min-1, 86% RH total flow.

Page 65: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

48

Fig

ure

3.5

A

eros

ol g

ener

atio

n sc

hem

atic

. (A

) sh

ows

the

bubb

ler

met

hod

and

(B)

the

VO

AG

app

roac

h.

Page 66: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

49

For NaNO3 particles the growth factor is 1.73 at 86% RH as obtained3 from the graph in

Figure 3.4. To generate a 1 µm droplet containing NaNO3, a dry particle with diameter of 617

nm is selected with the DMA and then sent to the 86% RH chamber. The result is a mono-

disperse aerosol flow of NaNO3 aqueous droplets.

The GF for SDS coated salt droplets has been measured as a function of the surfactant

solute weight fraction (WF = SDSmass/(SDSmass + NaNO3, mass)). Low SDS concentration is

desired to avoid foam formation by bubbling the dilute solution.5 A particle with SDS weight

fraction of 0.014 in the dry SDS/NaNO3 particle has a GF of 1.68 at 86% RH. Therefore a 630

nm dry particle is selected with the DMA and grown in the 86% RH humid chamber to produce a

1 µm surfactant coated water droplet.

3.2.4 Vibrating Orifice Aerosol Generator

An alternative way of producing a mono-disperse aerosol flow is to use a vibrating orifice

aerosol generator13 (VOAG) which is shown in Figure 3.5 (B). A liquid jet is generated by

injecting a dilute solution through a small orifice (usually 10 or 20 µm in diameter) at high

pressures. The orifice vibrates and uniformly breaks the liquid jet producing a mono-disperse

aerosol flow. The orifice vibration is controlled by adjusting its working frequency which is

between 40-80 kHz. The constant vibrating frequency and steady liquid flow-rate determines the

droplets size precisely and consistently. Droplets generated with this method are on the order of

25 µm or bigger depending on the orifice sized used.

As described in the previous section, to generate a 1 µm size NaNO3 aqueous droplet, a

610 nm dried NaNO3 particle is required. To create a particle of similar size from the VOAG, a

25 µm droplet with a concentration 3.6 x 10-4 M in NaNO3 is needed. Once this droplet is

Page 67: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

50

completely dried of water, the result is a 610 nm particle. Then the dry particle flow is re-

humidified in a controlled 86% RH chamber to grow to a 1 µm droplet.

The same analysis is made with NaNO3/SDS mixture to generate a 1 µm droplet by

VOAG. Figure 3.6 summarizes the drying of the VOAG produced 25 µm droplet. The dry

particle will grow based on its growth factor (1.68) by H2O uptake at 86% RH. The only

difference is that the concentration of SDS in the solution depends on the initial NaNO3

concentration. A solution with concentration of 3.6 x 10-4 M in NaNO3 and 1.5 x 10-6 M in SDS

has an SDS weight fraction of 0.014. Once the solvent evaporates from the 25 µm droplet, a 635

nm NaNO3/SDS dried particle is produced. The final volume, and therefore diameter, in the

dried particle is calculated from each component’s density and mass. The 635 nm dried particle

will grow to nominally 1 µm at 86% RH. The resulting 1 µm droplet will have a surfactant film

thickness of approximately 2 nm. It will contain 2 x 1010 H2O molecules in the core, 7.5 x 106

surfactants at the surface and 3.1 x 105 surfactants in the bulk. Even though the starting solution

has low SDS concentration (1.5 x 10-6 M), it contains enough surfactants to cover the surface of

the 1 µm droplet. The remaining surfactant molecules partition to the droplet’s bulk. If we

assume a SDS critical micelle concentration of 0.0081 M, all the 3.1 x 105 surfactants in the bulk

exist as monomers and not as micelles. Avoiding micelle formation in the bulk is done by

selecting an appropriately low SDS concentration in the starting solution.

Page 68: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

51

Figure 3.6 VOAG aerosol generation numbers. A 25 µm droplet is dried and then grown by H2O uptake at 86% RH to produce a 1 µm droplet coated by SDS surfactant (see text for description).

3.3 Droplet Spectrometer: Design

Water droplets of known size and composition are sent to the droplet spectrometer for

detection and analysis. Figure 3.7 shows the schematic overview of the droplet spectrometer.

The spectrometer consists mainly of three chambers or regions; the inlet, drift tube and

quadrupole mass spectrometer. Particles are sampled from the atmospheric pressure flow tube

through the inlet into vacuum. Once in vacuum, particles travel to the drift tube where they are

detected and vaporized. Vaporization of the bulk water releases previously solvated ions into the

gas phase. The ions are then extracted via electric field into the QMS chamber for mass analysis.

Page 69: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

52

Fig

ure

3.7

D

ropl

et m

ass

spec

trom

eter

ove

rvie

w (

see

text

for

des

crip

tion

).

Page 70: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

53

Figure 3.7 Droplet mass spectrometer overview (see text for description).

3.3.1 Capillary Inlet

Droplets are sampled through a 0.16 cm (O.D.) x 10 cm long capillary with a sampling

orifice I.D. of approximately 0.5 mm. Figure 3.8 shows the schematic of this inlet. Calculated

conductance14 for air is;

L/s 0.085 (L/s)cm10

)Torr (760)cm 05.0(179.6 (L/s)

Pd179.6(Air)C

44

pipe

(3.5)

where Cpipe is the conductance in L/s, d is the pipe’s I.D. in cm, is pipe’s length in cm and P is

the upstream pressure of the pipe in Torr. The calculated capillary conductance is 85 cm3 s-1

while the measured sampling rate is in the range of 150-200 cm3/min. If the droplet

concentration in the flow tube is 102 particles cm-3 then aerosol flux through the capillary is (85

cm3 s-1)(102 droplets cm-3) or 9 x 103 droplets s-1. A roots blower with pumping speed of 475 L/s

maintains the pressure after the capillary at approximately 30 mTorr. A 0.5 mm molecular

skimmer is used to sample the center of the aerosol beam. The skimmer has a smooth hyperbolic

design for better gas separation from the particle beam. The region after the skimmer and before

the drift cell is pumped by a 77 L/s turbo pump. The pressure in this chamber is 0.1-1 mTorr.

Page 71: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

54

Figure 3.8 Droplet spectrometer’s inlet.

The capillary mounts in a threaded Swagelok nut that screw into a xy translation stage.

The Swagelok nut can be screwed in/out for capillary adjustments in the z direction (or

instrument axis). These xyz adjustments allow for rough alignment of the particle beam in the

instrument. The capillary is set at a particular z position, usually between 1-4 mm from the

skimmer as seen from the side window. Then the xy adjustments are made to maximize the

pressure in the drift chamber. This will produce the highest throughput of particles past the

skimmer. Final small adjustments are made with the particle detection, explained later in the

chapter.

The water droplets will “freeze” or cool down dramatically by going from atmospheric

pressure into vacuum. This evaporative cooling is a coupled mass and energy transfer problem.

The droplets experience the low pressure and start evaporating H2O, taking energy from the bulk

Page 72: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

55

in the process. Hydrogen bonding for a water molecule is equivalent to 10 kcal/mol. Water loss

will cool the rest of the water bulk, finally stopping the evaporation and “freezing” the droplet.

However, a water droplet will evaporate depending on the background gas and the pressure. For

example, a 1 µm droplet with accommodation coefficient, α, of 0.1 will completely evaporate in

approximately 100 ms in 1 Torr of N2 background gas.15

3.3.2 Drift Tube

Once sampled and skimmed, aerosols pass into the 25 cm long drift tube shown in Figure

3.9. The drift tube consists of fifteen 60 mm (I.D.) stainless steel concentric rings (or cells)

isolated from each other by plastic PEEK spacers and connected in series by a resistor daisy

chain. Three of the cells are relatively wide to accommodate laser windows. The rings have

ridges with outer diameter flush to the spacers’ inner diameter. This design makes the assembly

straightforward. The cells with the laser windows are pinned to each other for better laser

alignment consistency. The cells are interchangeable allowing changing of the distances

between laser windows and therefore laser beams.

Page 73: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

56

Figure 3.9 Drift tube diagram. Particles enter the drift region, are detected by light scattering from continuous green laser beams, and trigger the 10 Hz YAG/OPO IR laser beam for droplet vaporization to release ions into the gas phase.

3.3.2.1 Water Cluster Dissociation

In the drift tube, the droplets are detected by light scattering and subsequently vaporized

with a YAG/OPO infrared laser beam. Vaporization releases ions into the gas phase for analysis.

A 1 µm surfactant coated water droplet contains about 1010 H2O molecules and around 107

surfactant ion molecules. It is likely that the vaporized surfactant ions will cluster to water as

demonstrated in Chapter II section 2.5 for Li+ ions.16 The ion-water clustering is expected to

complicate the mass spectra analysis. For that reason a drift tube that works at higher pressures

(0.3 – 1 Torr) and a constant electric field is implemented in the droplet spectrometer. The ion-

Page 74: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

57

water clusters are accelerated in the constant field dissociating the clusters with low energy

collisions with the background N2 gas.

Aerosols pass through a Venturi inlet with a 2 mm diameter orifice into the drift tube.

The design17 for the Venturi is modeled after Van Doren et al. Just downstream of the 2 mm

orifice, two concentric annuli are used to inject gas into the tube. The buffer gas passes through

the annuli at high pressures creating a lower pressure right at the orifice. This dynamic process

is called a Venturi effect and often these inlets are called aspirator inlets.18 In Figure 3.9, the

cooled water droplets feel the drag that the Venturi inlet produces and pass from a lower pressure

(1 mTorr) into the drift tube higher pressure (0.3 - 1 Torr). This type of inlet has been well

characterized and is used to study kinetics of ion-neutral reactions and ion mobility because of

the consistent flow rate and laminar flow produced in the drift tube.17-20 When no gas is injected

through the Venturi inlet, the static gas leaking from the capillary inlet through the skimmer sets

the pressure in the drift tube. The distance between the capillary and the skimmer is adjusted

resulting in a drift tube pressure range of 50-180 mTorr.

To dissociate ion-water clusters, a combination of collisions and electric field are

required. The electric field per pressure (E/N) is the characteristic parameter19 used in ion

mobility studies where 1 x 10-17 E/N = 1 Townsend or 1 Td. The drift tube cells are connected in

series by resistors. A voltage is applied across the drift tube to extract ions into the QMS region.

The voltage works together with the relatively higher pressure (0.3-1 Torr) in the drift tube to

dissociate any ion-H2O clusters formed. A voltage of 320 V across the 25 cm long drift tube

imparts a field (E) of 12.8 V cm-1. In a background pressure of 0.4 Torr (N = 1.3 x 1016 cm-3) the

electric field produces an E/N of 1 x 10-15 V cm2 or 100 Td. This produces low energy collisions

to dissociate the ion-water clusters and to extract the gas phase ions into the QMS.

Page 75: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

58

3.3.3 Droplet Detection

The particle beam in the drift cell perpendicularly intersects two continuous (CW) green

laser beams ( 50 mW) of wavelength 532 nm. The CW laser beams are approximately 4 cm

apart and are focused by 20 cm focal length (F.L.) lenses to the centerline of the drift tube. The

calculated focal spot is approximately 500 µm or 0.5 mm in diameter. Forward light scattering

from particles is collected at 30º angle by a 30 mm F.L. lens. The light collected by the lens is

detected by two analog photomultiplier tubes (PMTs). Light scattering signals from the PMTs

are used to find the particle transit time between the two detection lasers. This transit time is

used to trigger the YAG/OPO laser system to vaporize the detected droplets.

Fine adjustment to particle and laser beam overlap alignment is performed using the light

scattering signal. The PMTs signals are sent to an oscilloscope and the laser beams are moved so

that the signal intensity and number of particles detected are maximized. Then the second PMT

signal is monitored so that particle velocity can be measured with the oscilloscope.

3.3.4 Droplet Vaporization

3.3.4.1 Laser system

The vaporization laser system is an Optical Parametric Oscillator/Optical Parametric

Amplifier (OPO/A) pumped by a Continuum Surelite II Nd:YAG. This YAG laser is actively Q-

switched, runs at 10 Hz and has a 10 ns pulse duration. The delay time between flash lamp and

Q-switch is 194 µs with 4.8-5.0 W of power at the fundamental wavelength ( = 1064 nm). The

OPO/A system21 is designed to convert the single fixed frequency from the YAG laser into

tunable radiation in the mid-infrared. The YAG beam is aligned into the OPO/A system

Page 76: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

59

entrance. The beam is split with a beam splitter into two beam lines. One beam is frequency

doubled (to 532 nm) in a KTP (potassium titanyl phosphate) crystal to pump the OPO cavity and

the other beam is sent to a delay line before recombination with the OPO output. In the OPO, a

532 nm KTP crystal is used as the medium for the parametric conversion. This produces a mid-

infrared wavelength at 1.67 µm which then passes to the amplifier. Before the amplifier the mid-

infrared is combined with the delayed fundamental beam ( = 1064 nm). The combined beams

are then directed to the optical parametric amplifier (OPA). The amplifier consists of four 1064

nm KTP crystals where the mid-infrared radiation is produced via difference-frequency mixing.

Two wavelengths result from this process: 2.94 µm and 781 nm. These wavelengths are

opposite in polarization; the 2.94 µm is horizontally polarized and the 781 nm is vertically

polarized. A polarizer is placed at the exit to choose between the wavelengths. Power at 2.94

µm is up to 30 mJ/pulse and up to 60 mJ/pulse at 781 nm. For this experiment only the 2.94 µm

is used because it is the peak absorption for bulk water.22 The laser system is tunable from 2.75

µm to 3.0 µm and from 710 nm to 885 nm. All crystal positions are motor controlled with a

Microsoft Windows based program in a dedicated computer to easily change wavelength.

3.3.4.2 Energy Requirements

How much energy is required to vaporize a 1 µm diameter surfactant coated aqueous

particle? Vaporization energy for water23 is 40.7 kJ/mol (9.72 kcal/mol) or 2.27 kJ/g at 293 K.

A 1 µm water droplet has a volume of 5.2 x 10-13 cm3 and consequently a mass of 5.2 x 10-13 g

assuming a density for water of 1.0 g/cm3. The energy required to vaporize the 1 µm droplet is

(5.2 x 10-13 g )(2.27 kJ/g) or 1.2 x 10-9 J (1.2 nJ).

The mid infrared laser beam from the YAG/OPO system can generate up to 30 mJ/pulse

at 2.94 µm or 3400 cm-1. This is the peak frequency absorption for water as shown in the

Page 77: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

60

spectrum for bulk water in Figure 3.10. Water in the gas phase24 has three absorptions;

asymmetric stretch at 3756 cm-1 (3) , symmetric stretch at 3657 cm-1 (1) and the bending mode

at 1595 cm-1 (2, not shown in spectrum).

Figure 3.10 Absorption spectrum for bulk water in the infrared. Red arrow points at the peak absorption of water at 3400 cm-1 or 2.94 µm. Adapted from reference 22.

The hydrogen bonding network of bulk water washes out the features of the gas phase

spectrum and red-shifts the peak absorption of water. Absorption for the droplet can be

calculated using Beer’s law as a first approximation;

) (α -o e I I (3.6)

where α is the absorption coefficient for water and is the droplet’s diameter. The absorption

coefficient is calculated from the imaginary part25,26 of the index of refraction, ik, for water at a

particular wavelength (λ). The expression is;

λ

k π4 α (3.7)

Page 78: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

61

For the wavelength of 2.94 µm, k is 0.205 and as a result α is 0.88 µm-1. If the laser

beam with 10 mJ of energy is focused to a 1 mm (1000 µm) diameter spot and the particle is 1

µm in diameter, the amount of energy absorbed by the droplet is:

nJ 6 )e -(1μm 1000

μm 1 mJ 10 )e -(1I I)-(I (0.88)(1) -

2

) (α -oo

(3.8)

The energy absorbed by the 1 µm droplet is 6 nJ which is approximately 6 times more

than the required vaporization energy of 1.2 nJ. If the laser beam power is increased to 30

mJ/pulse, the amount of energy absorbed increases to 18 nJ. The droplet temperature rises based

on the energy absorbed (Q), water mass (mwater) and water heat capacity at constant pressure

(Cp).

K 2670 )g 10 x (5.2 g))J/(K (4.184

J 10 x 6 (K)

m C

Q T

13-

-9

waterp

(3.9)

The Cp (4.184 J K-1 g-1) for water27 is assumed to be constant at all temperatures. The 1 µm

water droplet will increase its temperature by 2670 K for a 10 mJ/pulse. A 30 mJ/pulse takes the

temperature to 8000 K.

This is a large amount of energy to be absorbed by the droplet. The wavelength for the

vaporization to be at the peak absorption of water was chosen for two reasons. The first is that

water will absorb readily at this wavelength to vaporize the droplets and release ions from

solution. The fast energy delivery from the laser pulse facilitates the explosive boiling of the

water. The second reason is that the organic surfactant covering the surface of the droplet will be

transparent to this energy. Most organic molecule C-H stretches are below the 3000 cm-1 region

of the infrared spectrum. Since we are using a laser at 3400 cm-1 with a bandwidth of

approximately 3 cm-1, fragmentation of the organic surfactant ions is expected to be minimized

Page 79: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

62

or completely suppressed by this vaporization method. Minimizing the fragmentation of the

organic surfactants is critical to the detection of individual molecular species resulting from the

initial radical induced oxidation of organic marine aerosols.

3.3.4.3 Laser Triggering

The IR vaporization laser beam intersects the particle beam at 3.8 cm and 7.6 cm from

the second and first green laser beams, respectively. These distances depend on the IR cell

position in the drift tube, which can be changed. For the purpose of this description the distance

is set at 7.6 cm from the second detection laser to the vaporization laser. Droplet transit time

between the lasers is less than 1 ms. Droplets must be located in space and time in order to be

vaporized.

A timing circuit has been developed to trigger the YAG/OPO system. Since the particles

travelling at appropriate velocities occur randomly with respect to time, triggering the

YAG/OPO is complex. The YAG laser has to run at an average of 10 Hz to maintain appropriate

thermal load. Moreover the YAG laser needs to be triggered 200 µs in advance of the firing

time. The timing circuit meets the challenge of triggering a fixed frequency laser to hit randomly

occurring particle events.

We have adopted software based logic for the timing circuit in contrast to others that use

hardware.28,29 A schematic of the PMTs signal along with the generated pulses for the YAG

laser flashlamp and Q-switch are shown in Figure 3.11. A micro-processor (40 MHz) gathers the

light scattering signals from the PMTs and is programmed to generate triggering pulses for the

YAG laser. Since the YAG must fire every 100 ms, an internal 10 Hz clock keeps track of this

cycle. An arbitrary 20 ms (from 90-110 ms in the clock) active viewing window is set to look

Page 80: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

63

for a valid aerosol detection event. The arbitrary viewing window keeps the laser firing at an

average repetition rate of 10 Hz with a variation of ±1 Hz.

The valid aerosol detection event is defined as light scattering detection in PMT1 and

PMT2 within the active window, intensity level, and circuit time constraints. All of these

variables are controlled and set with the software through the circuit front panel. The active

window is fixed and cannot be changed. The PMT signal intensity level is usually 30 mV in the

circuit. Analog signal from the PMTs is amplified by a factor of 20 to improve signal-to-noise.

As for constraints, the microprocessor cannot process a particle transit time between PMTs of

less than 120 µs and still successfully trigger the YAG laser in time to intersect the particle.

Therefore particle events faster than 120 µs are ignored. If an acceptable particle detection event

is gathered by the circuit, all other scattering pulses from subsequent particles are ignored until

the whole triggering sequence takes place. This ensures that the YAG laser is not triggered

faster than its working frequency of 10 Hz.

Page 81: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

64

F

igu

re 3

.11

Tri

gger

ing

sche

mat

ic.

PM

Ts

sign

al f

rom

aer

osol

lig

ht s

catt

erin

g to

tri

gger

the

YA

G f

lash

lam

p an

d Q

-sw

itch

. tT is

the

part

icle

tran

sit t

ime

betw

een

gree

n la

ser

beam

s, t F

is th

e ti

me

for

the

flas

hlam

p re

lati

ve

to P

MT

2 si

gnal

and

t Q is

the

tim

e to

fir

e th

e Q

-sw

itch

to f

ire

the

lase

r.

Page 82: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

65

The time between the two PMT signals is the transit time, tT, which is used to trigger the

YAG. The particle should arrive at the laser interaction region at time 2 tT. The logic in the

circuit generates two pulses to trigger the YAG laser, one to fire the flashlamp and the other to

trigger the Q-switch to fire the laser. The Q-switch pulse is generated at tQ = 2tT, since the IR

laser beam/aerosol interaction region is twice as far from the second PMT as the distance

between the two PMTs. The flashlamp pulse is generated at tF = (tQ - 194 µs). The 194 µs is the

factory set optimum delay time between flashlamp and Q-switch for the YAG laser. At the end

of this YAG firing cycle the circuit resets and starts again. If no aerosol is detected within the

active window a pulse is sent to only fire the flashlamp to maintain the heat load on the laser

cavity.

Furthermore, the circuit programming allows changing the acceptable particle transit time

window (i.e., to select for different particle velocities), scattering signal level, and more

importantly tQ. All this helps optimize for appropriate particle velocities, particle counts,

detection efficiencies and time overlap of the YAG/OPO beam with the droplet.

3.3.5 Signal Estimates

Sections 3.2.3 and 3.2.4 described the process of generating 1 µm SDS coated

NaNO3/water droplets. Composition of this 1 µm droplet is depicted in Figure 3.6 along with

estimates. If we follow the concentration estimates, a coated droplet contains 1.6 x 1010 water

molecules in the core, 7.2 x 106 SDS molecules at the surface and 3 x 105 SDS molecules in the

bulk. The surfactant number at the surface is calculated using the estimated10 SDS footprint of

50 Å2/molecule. The bulk SDS molecules are calculated by subtracting the amount at the surface

from the total. Droplet vaporization will release 7.5 x 106 SDS ions per 1 μm droplet into the gas

phase (CH3(CH2)11OSO3-) for detection. Counter ions are not considered in this analysis but for

Page 83: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

66

every surfactant ion there is a Na+ ion. Ion-ion recombination30 represents a problem for the

detection signal as seen in chapter II. Laser repetition rate is 10 Hz and therefore is the highest

rate at which particles could be vaporized and detected. If 10 particles are vaporized per second

then the result is (10 particles/second)(7.5 x 106 SDS ions/particle) or 7.5 x 107 SDS ions s-1.

This corresponds to approximately (7.5 x 107 ions)(1.6 x 10-19 C/unit charge) or 1.2 x 10-11 C/s

(12 pA).

3.3.6 Ion Detection

The ions released from the droplets are accelerated from the drift tube into the next

differentially pumped section between the drift tube region and quadrupole mass spectrometer

chamber. Ions exiting the drift tube are guided with an ion lens pointed in Figure 3.8 into a

sampling skimmer or nose cone. The ion lens consists of four elements, two in the x direction

and two in the y direction. The ion lens can be biased with positive or negative voltage. All four

elements can be biased together at the same voltage or independently biased. This flexibility in

the ion lens allows for optimization of ions going into the quadrupole chamber.

The nose cone attached to the quadrupole chamber (see Figure 3.12) is also

independently biased and in combination with the ion lens better ion throughput into the

quadrupole is achieved. Once ions get through the nose cone they are delivered to the QMS

chamber. In this chamber a cryogenic pump with pumping speed of 1000 L/s keeps the pressure

at around 10-6 Torr. The cryo-pump was chosen to pump away efficiently as much water as

possible. The droplets contain approximately 3 orders of magnitude (1010 H2O molecules) more

water molecules than surfactant ions.

Page 84: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

67

Fig

ure

3.1

2 Q

MS

cha

mbe

r.

Page 85: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

68

The QMS is a commercial instrument from Extrel Company, Model

MAX5001.2APP3/4N6. It has a mass range of 1-500 amu, 2-4 mA/Torr sensitivity and about

106 abundance sensitivity. It consists of an einzel lens, quadrupole bending lens or 90º deflector,

the quadrupole mass filter, and an electron multiplier (not shown in Figure 3.12) for ion

detection. The einzel lens guides and focuses the ions to the center of the quadrupole mass filter.

The 90º deflector is optimized to bend the ions into the QMS direction by the use of fixed

voltages. As with the quadruple mass filter, the 90º deflector rods are biased such that two

opposing rods are positively biased while the other two are negatively biased. The difference

from the mass filter is that they are not necessarily biased at the same voltage and no radio

frequency is applied. Also the voltage depends on the type of ions (positive or negative) to be

analyzed. If the rods are optimized to transmit positive ions, reversing the polarity will transmit

negative ions. The quadrupole mass filter scans to separate mass-to-charge (m/z) ratios for ion

separation. Ions then strike an electron multiplier that is biased at ± 1900 V (depending on ion

polarity) for detection. Extrel Merlin automation software is used to gather the mass spectra

data.

3.3.7 Expected Results

The radical oxidation of the SDS surfactant coated NaNO3/water droplets in an

atmospheric pressure flow tube is predicted to form the alkyl peroxyl radical

(•OOCH2(CH2)11OSO3-) at the droplet surface as depicted in Figure 3.13. The main objective is

to detect this peroxyl radical after oxidation, particle detection and vaporization. The expected

mass spectrum before reaction is of the SDS surfactant (CH3(CH2)11OSO3-) without

fragmentation at m/z = 288 (see Figure 3.13). No or minimal fragmentation is predicted because

the vaporization laser wavelength (2.94 µm or 3400 cm-1) is not absorbed by the organic

Page 86: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

69

surfactants. Also a clean spectrum is expected with no ion-water clusters present. Sulfate ions

are very stable and have similar electron affinities31 as carboxylates on the order of

approximately 3.5 eV. The ion-water clusters that may form from droplet vaporization will be

dissociated with low energy collisions with the background gas in the drift tube (see section

3.3.2.1). Following oxidation of the surfactants a mass spectrum with the alkyl peroxyl radical at

higher mass (m/z = 319) is predicted. This comes from the hydrogen abstraction followed by O2

addition to the alkyl radical (m/z = 288 -1 + 32 = 319). The peroxyl radical is detected in the

presence of the unreacted surfactant in the mass spectrum. The droplet spectrometer will be used

to study the initial steps of the organic marine aerosols oxidation processes at room temperature

and atmospheric pressure. Individual molecular species are expected to be identified from the

droplets surface after radical oxidation.

Page 87: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

70

Fig

ure

3.1

3 E

xpec

ted

resu

lts

from

the

oxid

atio

n of

SD

S c

oate

d N

aNO

3/w

ater

dro

plet

s.

Page 88: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

71

3.4 Droplet Spectrometer: Development

Optimization of the droplet spectrometer components is needed in order to develop the

spectrometer into a functional instrument. To begin, ion detection had to be optimized to deliver

ions from the drift tube into the QMS chamber for analysis. For this purpose an electron impact

(EI) ionization source is used to generate ions in the drift tube for detection with the QMS. Once

ions detection is optimized, droplet detection follows. Droplet detection involves light scattering

collection, transit time measurements and circuit optimization for YAG/OPO laser triggering.

The triggering allows intersecting the particles in space and time with the IR laser beam. Ion

detection from droplet vaporization is performed by hitting droplets with the IR laser beam

between two stainless steel plates. Incident current is then measured at the plates using an

operational amplifier (op-amp). Finally, droplet vaporization and ion detection need to be

combined for the spectrometer to be fully developed and functional.

3.4.1 EI Source

The droplet mass spectrometer is designed to detect ions from the laser-droplet

interaction region from the drift tube into the QMS. Ions are extracted from the drift tube,

sampled through the nose cone into the quadrupole chamber and bent 90º into the QMS for

detection. Optimization of the drift tube cells voltages, the ion lens and nose cone voltages, and

the QMS settings is required to maximize ion detection.

An electron impact source was built to produce ions in the drift tube for extraction and

detection in the quadrupole chamber. The EI source is mounted on a 30 mm diameter, 3 mm

thick copper plate that fits in the PMT window drift tube cell and seals to vacuum. The EI

source consists of three stainless steel leads. A grid is attached to one lead and a Rhenium

(99.9%) wire is attached to the other two stainless steel leads. The leads are long enough for the

Page 89: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

72

filament and grid to be at the center of the drift tube. The assembly also has a 0.25 stainless

steel tube to leak gas into the chamber for ionization.

The EI source is placed in the second drift tube cell (see Figure 3.14). The filament is

floated at around 200 V and the grid at approximately 270 V. The electron energy could be

adjusted but it was set at the standard 70 eV. The voltage of the drift cell is the same as the grid.

The current passing through the filament is about 5-6 A at optimum settings. The filament

slowly ages and the current must be readjusted. The filament must be replaced every three days

of use as the filament breaks or does not generate enough electrons for ionization. The emission

current generated is about 0.1 – 0.6 mA measured at the drift tube cell with a picoammeter

(Keithley 485 Autoranging Picoammeter). The grid is connected in series to the rest of the drift

tube cells stack by 40 k resistors (see Figure 3.14). The last cell is connected to ground by a

250 k resistor in order to have the cell still floating at a voltage. With this setup the last cell

has a voltage of approximately 90 V.

Page 90: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

73

Figure 3.14 Schematic set up for EI source ionization and ion extraction into the QMS region.

The main idea for ion detection optimization is to generate ions and measure ion current

along the way in the drift tube and quadrupole chamber (see Figure 3.14). Once ions are

delivered into the quadrupole chamber the QMS settings are optimized for ion detection. Gas is

leaked through the EI source for ionization and detection to maintain a pressure in the drift tube

of approximately 100 mTorr. Ion current is detected at the ion lens and nose cone after the drift

tube with the picoammeter. The measured current at the ion lens (all four elements combined) is

around 10 nA and about 2 nA at the nose cone. After measuring ions after the drift tube,

voltages were applied to the ion lens and nose cone with its dedicated power supply. All four

elements in the ion lens have a voltage of 43 V and 10 V is applied to the nose cone. These

values were set by measuring ion current in the quadrupole einzel lens and 90º deflector poles.

The ion current at the einzel lens (measured at pin G and E of the QMS) is approximately 30-45

Page 91: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

74

pA. At the 90º deflector poles the ion current measured (at pin A and B) is 10-22 pA. This ion

current corresponds to (22 pA/1.6 x 10-19C/ion) or approximately 1.4 x 1011 ions s-1.

Figure 3.15 shows representative mass spectra for ionization of N2 and Ar gas using this

set up. The spectra show very clean peaks at m/z = 28 and m/z = 40 corresponding to N2+ and

Ar+, respectively. Good signal-to-noise is achieved. Spectra were acquired using Extrel Merlin

automation software. To obtain these spectra the einzel lens and 90º deflector pole voltages were

optimized. Mass spectra were monitored in real time with the spectrometer working in radio

frequency (RF) mode only. This mode allows for all ions to pass through the quadrupole mass

filter and be detected in the electron multiplier with no mass separation. Once maximum signal

is obtained the DC voltage is applied again. Then small voltages adjustments were done to

obtain a better mass spectrum. All voltages are controlled with the software tune files through a

personal computer.

Figure 3.15 EI mass spectra for N2+ and Ar+ extracted from the drift tube into the QMS.

Page 92: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

75

3.4.2 Droplet Detection

Droplet detection was optimized using known size standards polystyrene (PSL) spheres

and particles generated by bubbling a NaNO3 solution. The PMTs light scattering signals are

sent to the circuit box and from the circuit box signal is split to monitor with an oscilloscope.

Signal-to-noise for light scattering is 10 or better. Figure 3.16 (A) shows particle light scattering

signals from PMTs for a 1.6 µm polystyrene (PSL) sphere at a drift tube pressure of 40 mTorr.

The transit time between green laser beams for the 1.6 µm PSL particle is 133 µs. If the pressure

in the drift tube is increased to about 1 Torr the transit time is 265 µs as shown in Figure 3.16

(B). The YAG laser requires at least 194 µs to be triggered. If the pressure in the drift tube is

kept low (40 mTorr) the IR laser beam must be at twice the distance ( 8 cm) than the green

lasers distance ( 4 cm). With higher pressures (1 Torr) the IR laser beam can be at the same

distance from the second detection stage as the distance between the detection lasers. The IR

laser beam is set at 2x the distance of the green laser beams in the current setup. This ensures

PMTs signal could be processed by the circuit logic to generate the pulses to trigger the YAG

laser.

Page 93: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

76

Figure 3.16 Particle light scattering signal from PMTs to measure transit time, tT. (A) shows tT = 133 µs with drift tube pressure at 40 mTorr and (B) tT = 265 µs for a drift tube pressure of about 1 Torr.

The PSL particles used to calibrate the spectrometer had the following diameters; 495

nm, 900 nm, 1000 nm and 1600 nm. Also a poly-disperse aerosol flow generated from a 0.1 M

NaNO3 solution was used for comparison. Figure 3.17 shows the transit time distributions for all

the particle sizes and the NaNO3 poly-disperse flow. The calibration was done at a drift tube

pressure of approximately 550 Torr. All particle sizes have overlapping distributions with

approximately the same average transit time of 188 µs. The NaNO3 distribution is broader but

still shows the same average transit time as the PSL particles. Basically particles are not

differentially accelerated into vacuum based on their size. Therefore no size information could

be obtained. A possible explanation for this is that the capillary inlet has a length to inner

diameter (L/dI.D.) that is too large. This causes the particles to spend more time in the capillary

making the particle equilibrate to the expanding gas final velocity. Measured particle velocities

Page 94: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

77

are in the 275 m/s - 300 m/s range, which is comparable to the expanding background gas

velocity.

Figure 3.17 Droplet spectrometer calibration with PSL particles with 495 nm, 900 nm, 1000 nm and 1600 nm diameters. Sodium nitrate (SN) poly-disperse particles from a 0.1 M solution were also measured for comparison.

3.4.3 Droplet Vaporization

The droplet vaporization optimization is done using two stainless steel plates located in

the IR laser-droplet interaction region as shown in Figure 3.18. The interaction region is located

between the plates at about 8 cm from the second green laser beam. The green laser beams are

about 4 cm apart, and the plate dimension is 1 x 2. They are electrically isolated by ceramic

pieces that hold them together in the corners. The top plate can be positively or negatively

Page 95: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

78

biased for detecting positive ions or negative ions, respectively. When the droplet is vaporized,

the ions released from solution are repelled from the top plate to the grounded bottom plate

where incident current is detected with an op-amp. The setup is similar to that of Dessiaterik et

al.30 in which the circuit works in a current to voltage configuration. The gain in this setup is

higher at 2 x 106 V/A. For positive ion detection the voltage can be increased up to 370 V while

for negative ion detection the voltage can be increased up to -175 V.

Figure 3.18 Plates setup schematic for droplet vaporization optimization. The top plate repels ions to the bottom plate. The op-amp circuit is used to measure incident current on the grounded bottom plate.

NaNO3 droplets were generated by bubbling a 0.1 M solution. The poly-disperse aerosol

flow is sent to the instrument inlet for sampling. This ensures that many droplets are sampled to

reach the detection region for optimization. The number of particles detected in each individual

PMT was observed to be 150-300 particles s-1. With this flux of particles the circuit detected

particles very efficiently. For example, all particles detected in PMT1 were also detected in

PMT2 with no less than five out of five (5:5) per second on average. Sometimes the efficiency

Page 96: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

79

decreased between 70-80%. The RH before the inlet was higher than 90% measured with a

humidity meter (Vaisala HM70 humidity and temperature meter). Since the RH is still high at

the inlet’s entrance all NaNO3 aerosols remain aqueous particles. Figure 3.19 shows the

detection of a droplet by light scattering followed by the circuit logic generated pulses to trigger

the YAG/OPO laser. In order to see all signals the oscilloscope is triggered on the Q-switch

pulse since this pulse is generated only when a real particle event occurs (see section 3.3.4.3).

The droplet transit time, tT, is 137 µs therefore the Q-switch is fired at 274 µs after droplet

detection. The flash lamp pulse is at tF = 217 µs or 80 µs after particle detection at PMT2.

Figure 3.19 Circuit logic sequence of acquired signals to trigger the YAG/OPO laser and detect incident current on the stainless steel plates after droplet vaporization (see text for description).

Page 97: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

80

The incident current detection after droplet vaporization was observed approximately 2-5

µs after the Q-switch pulse as shown in the green circle in Figure 3.19. The plates’ setup for

positive ion detection generates signal of around 100-200 mV about 2 µs after the Q-switch

pulse. This signal is erratic and not reproducible, as a result more difficult to be optimized.

Finding the particles in space/time with the IR laser beam seems to be the main reason for the

erratic signal. The Q-switch generated pulses have a jitter of a few microseconds and this is not

always reproducible for a given particle velocity. Droplets are travelling at about 280 m/s on

average, which means that in 1 µs it would have moved approximately 280 µm. This could be

outside the laser vaporization volume. To amend this different focal length (F.L.) lenses were

used. With a 10 cm F.L. lens the laser spot size is less than 1 mm while with a 25 cm F.L. lens

the laser spot size is 1-2 mm. The laser spot sizes are only estimated by burning paper (Kodak

linotype 1895) since the YAG/OPO infrared beam is elongated in the vertical direction. Bigger

laser spot size seemed to help but signal was still erratic and irreproducible.

In summary, ion detection optimization has been accomplished by using an EI source.

Also particle detection and transit time have been measured in order to optimize circuit logic for

laser triggering. Consequently, ion detection from droplet vaporization was attained using a set

of plates and the incident current was measured but there was no success in reproducibility. If

the ion detection, particle detection and laser triggering would function in a more reproducible

way, interfacial oxidation of organic marine aerosols could be studied. No fragmentation is

expected since the organic surfactants are not likely to absorb the IR laser energy. Extraction of

ions from the drift tube dissociates the ion-water clusters that may form. A clean mass spectrum

is expected as depicted in Figure 3.13. Finally, if the spectrometer is completely functional

Page 98: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

81

individual molecular species could be detected from organics at the droplet’s surface in an

attempt to elucidate the initial chemistry of organic marine aerosols.

Page 99: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

82

References for Chapter III

(1) Ellison, G. B.; Tuck, A. F.; Vaida, V., Atmospheric processing of organic aerosols, Journal of Geophysical Research-Atmospheres 1999, 104, 11633-11641.

(2) Tang, I. N., Thermodynamic and optical properties of mixed-salt aerosols of atmospheric importance, Journal of Geophysical Research-Atmospheres 1997, 102, 1883-1893.

(3) Tang, I. N.; Munkelwitz, H. R., Water Activities, Densities, and Refractive-Indexes of Aqueous Sulfates and Sodium-Nitrate Droplets of Atmospheric Importance, Journal of Geophysical Research-Atmospheres 1994, 99, 18801-18808.

(4) Tang, I. N.; Munkelwitz, H. R., Aerosol Phase-Transformation and Growth in the Atmosphere, Journal of Applied Meteorology 1994, 33, 791-796.

(5) Zelenyuk, A.; Imre, D.; Cuadra-Rodriguez, L. A.; Ellison, B., Measurements and interpretation of the effect of a soluble organic surfactant on the density, shape and water uptake of hygroscopic particles, Journal of Aerosol Science 2007, 38, 903-923.

(6) Seinfeld, J. H.; Pandis, S. N. In Atmospheric Chemistry and Physics - From Air Pollution to Climate Change (2nd Edition); John Wiley & Sons, 2006.

(7) Tang, I. N.; Tridico, A. C.; Fung, K. H., Thermodynamic and optical properties of sea salt aerosols, Journal of Geophysical Research-Atmospheres 1997, 102, 23269-23275.

(8) Harper, K. L.; Allen, H. C., Competition between DPPC and SDS at the air-aqueous interface, Langmuir 2007, 23, 8925-8931.

(9) Li, Z. D.; Williams, A. L.; Rood, M. J., Influence of soluble surfactant properties on the activation of aerosol particles containing inorganic solute, Journal of the Atmospheric Sciences 1998, 55, 1859-1866.

(10) McNeill, V. F.; Patterson, J.; Wolfe, G. M.; Thornton, J. A., The effect of varying levels of surfactant on the reactive uptake of N2O5 to aqueous aerosol, Atmospheric Chemistry and Physics 2006, 6, 1635-1644.

(11) Flagan, R. C. In Aerosol Measurement: Principles, Techniques, and Applications; 2nd ed.; Willeke, K., Baron, P. A., Eds.; John Wiley & Sons, Inc.: New York City, 2001.

(12) TSI, I. Series 3080 Electrostatic Classifier Manual: Shoreview, 2006; Vol. Revision G.

Page 100: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

83

(13) TSI, I. Model 3450 Vibrating Orifice Aerosol Generator Instruction Manual: Shoreview, MN, 2002; Vol. Revision L.

(14) O'Hanlon, J. F. A User's Guide to Vacuum Technology; Third ed.; Wiley-Interscience, 2003.

(15) Murphy, D. M.; Personal communication ed.; Ellison, G. B., Cuadra-Rodriguez, L. A., Eds.: Boulder, CO, 2006.

(16) Sobott, F.; Wattenberg, A.; Barth, H. D.; Brutschy, B., Ionic clathrates from aqueous solutions detected with laser induced liquid beam ionization/desorption mass spectrometry, International Journal of Mass Spectrometry 1999, 187, 271-279.

(17) Van Doren, J. M.; Barlow, S. E.; DePuy, C. H.; Bierbaum, V. M., Tandem Flowing Afterglow-Selected Ion Flow Tube and Its Application to the Thermal-Energy Reactions of 18O-, Journal of the American Chemical Society 1987, 109, 4412-4414.

(18) Dupeyrat, G.; Rowe, B. R.; Fahey, D. W.; Albritton, D. L., Diagnostic studies of venturi inlets for flow reactors, International Journal of Mass Spectrometry and Ion Physics 1982, 44, 1-18.

(19) de Gouw, J. A.; Krishnamurthy, M.; Bierbaum, V. M.; Leone, S. R., Measured and calculated mobilities of cluster ions drifting in helium and in nitrogen, International Journal of Mass Spectrometry 1997, 167, 281-289.

(20) King, G. K.; Maricq, M. M.; Bierbaum, V. M.; DePuy, C. H., Gas-Phase Reactions of Negative-Ions with Alkyl Nitrites, Journal of the American Chemical Society 1981, 103, 7133-7140.

(21) Guyer, D. Optical Parametric Oscillator User Manual: Bellevue, 2006.

(22) Robertson, W. H.; Johnson, M. A., Molecular aspects of halide ion hydration: The cluster approach, Annual Review of Physical Chemistry 2003, 54, 173-213.

(23) Lide, D. R. CRC Handbook of Chemistry and Physics; CRC Press: Boca Raton, FL, 1996.

(24) Shimanouchi, T. Tables of Molecular Vibrational Frequencies Consolidated Volume I; National Bureau of Standards, 1972.

(25) Toon, O. B.; Tolbert, M. A.; Koehler, B. G.; Middlebrook, A. M.; Jordan, J., Infrared optical constants of H2O ice, amorphous nitric acid solutions, and nitric acid hydrates, J. Geophys. Res. 1994, 99, 25631-25654.

Page 101: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

84

(26) Warren, S. G., Optical-Constants of Ice from the Ultraviolet to the Microwave, Applied Optics 1984, 23, 1206-1225.

(27) Silberberg, M. S. Principles of General Chemistry; First ed.; McGraw-Hill Science/Engineering/Math: New York, 2006.

(28) Noble, C. A.; Nordmeyer, T.; Salt, K.; Morrical, B.; Prather, K. A., Aerosol Characterization Using Mass-Spectrometry, Trac-Trends in Analytical Chemistry 1994, 13, 218-222.

(29) Mysak, E. R.; Dessiaterik, Y. N.; McKinney, C. J.; Miller, R. E.; Baer, T., Design of a timing circuit for random laser triggering on aerosol particles, Review of Scientific Instruments 2006, 77, 013301-(1-6).

(30) Dessiaterik, Y.; Nguyen, T.; Baer, T.; Miller, R. E., IR vaporization mass spectrometry of aerosol particles with ionic solutions: The problem of ion-ion recombination, Journal of Physical Chemistry A 2003, 107, 11245-11252.

(31) Rienstra-Kiracofe, J. C.; Tschumper, G. S.; Schaefer, H. F.; Nandi, S.; Ellison, G. B., Atomic and molecular electron affinities: Photoelectron experiments and theoretical computations, Chemical Reviews 2002, 102, 231-282.

Page 102: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

________________

Chapter IV

Evaporation of Water from Particles in the Aerodynamic Lens Inlet: An Experimental Study

________________

4.1 Introduction

Single particle mass spectrometers are used for real time characterization of aerosol

particles in field and laboratory studies. Aerosol particles are sampled from high pressure

environment to vacuum (10-6 Torr) for chemical and physical characterization at a single particle

level. The interface between the atmospheric pressure and high vacuum mass spectrometer

environment has evolved throughout the years. As seen in chapter II orifices and capillary inlets

were originally incorporated as sampling inlets.1,2 In recent designs these inlets have been

replaced by aerodynamic lens inlets.3,4 The aerodynamic lens inlet is better at collimating

particles into a beam and exhibits higher transmission efficiency in the sub-micron particle size

range.5-8 Customized versions of the Liu-type aerodynamic9,10 lens inlet have been adapted for

several single particle mass spectrometers including the one used for this study, single particle

laser ablation time-of-flight mass spectrometer (SPLAT).

It is desirable for sampling inlets to have high transmission efficiency of aerosol particles

without changing the particle composition in the process. The aerosols experience a large

pressure drop when delivered into the high vacuum mass spectrometer environment. As a result

the relative humidity (RH) decreases and water evaporates from the aerosols. This evaporation

process changes the particles’ chemical and physical properties. Water evaporation and

condensation in the inlet present a problem for single particle analysis.11 The water content of

the aerosol depends on the RH it has been exposed to. For example, an ammonium sulfate

Page 103: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

86

((NH4)2SO4) particle that has been sampled at a RH below its deliquescence point ( 80% RH)

could be a solid or liquid particle. This is evident from the (NH4)2SO4 phase diagram shown in

Figure 3.3. If the particle is exposed to an environment with higher than 80% RH and then

sampled at 50% RH the particle will be a saturated liquid droplet containing 31% H2O by mass.

On the other hand if the particle is exposed to a lower RH and sampled at 50% RH then it will be

a dry crystalline (NH4)2SO4 particle.12 This water environment dependent behavior affects

directly physical properties of aerosols and single particle analysis.

Discussion of water evaporation and condensation has been addressed previously in

supersonic inlets.11,13-15 But little attention has been paid to particles transported in aerodynamic

lens inlets, even though they are widely used. DeCarlo et al.13 first mentioned that water

evaporation in the low-pressure region in the aerodynamic lens inlet can cause significant

particle water loss. Moreover a complex behavior of the measured vacuum aerodynamic

diameter has been observed for size selected particles.8 It was concluded that the complex

changes in vacuum aerodynamic diameters were related to the RH the particles were exposed to

prior to sampling and water evaporation in the aerodynamic lens inlet. Also simulations

conducted by McGraw,16 found that the water evaporation induces rapid particle cooling

followed by particle re-warming by collisions with the gas at the 2 Torr pressure of the

aerodynamic lens. Because of re-warming, substantial water losses could occur during the

particle’s transit in the lens. Furthermore, it has been noticed8,13 that water evaporation was

significant in this type of inlet. To minimize ambiguity in data interpretation researchers have

chosen to dry particles prior to sampling.

A recent review14 of the designs of single particle laser mass spectrometers presents an

approximate calculation of water evaporation dynamics in three aerosol inlet types. It was

Page 104: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

87

concluded that evaporation in the Liu-type aerodynamic lens inlet is the most severe.

Evaporation could change the particles’ vacuum aerodynamic diameter distribution

measurements and their interpretation. There are no published experimental studies that provide

quantitative characterization of the water evaporation in an aerodynamic lens inlet. Available

calculations may not be sufficient to efficiently describe and quantify water evaporation in the

inlet.

This study presents the results of a number of measurements to quantify water

evaporation from particles composed of various hygroscopic compounds during their transport in

SPLAT’s aerodynamic lens inlet. Measurements of vacuum aerodynamic size (dva) distribution

were done on size selected droplets (i.e., (NH4)2SO4/H2O particles). The measurements were

done at high RH (74-85%) for wet runs and at low RH (<7%) for dry runs. Comparison of the

distributions at high RH (droplets) and low RH (dry particles) of size selected droplets provided

experimental evidence for water evaporation in the aerodynamic lens inlet. The observations

suggest that in most cases, the droplets’ dva distributions are nearly equal to those of the same

particles after drying. But the data also suggest that for one particle type (NaC) the measured

aerodynamic size distributions can be too complex to be easily interpreted.

4.1.1 Aerodynamic Lens Inlet

SPLAT uses a customized Liu-type aerodynamic lens inlet. A detailed description is

given by Jayne et al.6 The important parameters for SPLAT’s aerodynamic lens inlet include a

volume of 28 cm3, a pressure in the lens of 1.98 Torr, and the flow rate is approximately 100 cm3

min-1. The residence time for the particles in the lens is 37 msec.

In general, when a liquid particle is transported into the near-zero RH environment of the

aerodynamic inlet lens it is expected to lose its water in an attempt to establish a new

Page 105: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

88

equilibrium. As the water evaporates the particle temperature decreases and water evaporation

slows. This evaporative cooling can serve as a self limiting process. However, the pressure in

the aerodynamic lens is 2 Torr and the time the particles spend in the lens is about 37 msec. The

particles undergo a large number of collisions with the gas in the inlet. Therefore particles will

likely re-warm, after their initial evaporative cooling and evaporation can resume.16 The amount

of water lost in such a process depends on the particle composition, the RH, the inlet’s pressure

and the time that the particle spends in the lens.

4.1.2 SPLAT

SPLAT is similar to the single particle mass spectrometer described17 in chapter II

(section 2.5) and has been described in detail elsewhere.8 It uses an aerodynamic lens inlet to

transport particles from the ambient atmosphere into the high vacuum environment. In vacuum,

particles are detected by light scattering using two optical detection stages. In the first detection

stage particles intersect a CW 532 nm (2nd harmonic Nd:YAG) with power up to 300 mW. The

scattered light is collected by an ellipsoidal reflector and detected perpendicular to the particle

beam by a photomultiplier tube (PMT) positioned at the focal spot of the reflector. When a

particle is detected in the first optical stage it is recorded and a clock started. The same particle

is detected in the second optical detection stage located 4.125 ( 10.5 cm) downstream. The

particle transit time between the laser beams is recorded and used to trigger the vaporization and

ionization lasers. A CO2 laser (10.6 µm) and an excimer laser (193 nm) are used for

vaporization and ionization, respectively. The CO2 laser pulse is synchronized to hit the particle

for vaporization at the TOF ionization region. The resulting plume is ionized by an excimer

laser UV light pulse and the particle mass spectrum is recorded. Another possibility is to use the

UV laser pulse for a one step vaporization/ionization process.

Page 106: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

89

For this study only the particle transit times were recorded for the measurements. The

vacuum aerodynamic diameter of each particle is obtained by measuring particle’s transit time

between the two detection stages spaced by 4.125 ( 10.5 cm) and comparing it to a calibration

curve established with polystyrene latex (PSL) spheres used as NIST certified size standards.

The sizing accuracy for this set of measurements was ±1%. Figure 4.1 shows a representative

calibration curve. The data are fitted with a power function. The vacuum aerodynamic diameter

(dva) is obtained as a function of particle transit time and therefore comparison is straightforward

when doing vacuum aerodynamic diameter measurements.

Figure 4.1 Calibration curve to measure particle vacuum aerodynamic diameter (dva) as a function of particle transit time (tT). PSL particles were used as size standards. The data are fitted with a power function.

Page 107: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

90

4.2 Experimental

Ammonium sulfate ((NH4)2SO4) and sodium chloride (NaC) particles were selected in

this study because all of the relevant microphysical and thermodynamic properties are well

known.12 This makes it possible to calculate some of the relevant parameters. In addition we

provide results for other inorganic and organic water soluble compounds.

Particles are generated by aerosolizing water solutions using a TSI atomizer (TSI Inc.,

Model 3076). The wet aerosol flow rate is set to 0.35 L min-1 and used as the sample flow for

the Scanning Mobility Particle Sizer (SMPS, TSI Inc., Model 3936) that has a sheath flow rate

set to 3.5 L min-1. The SMPS uses the differential mobility analyzer (DMA) technique where

particles are size selected based on their charge and mobility at an applied electric field in a

sheath flow. The mobility diameter (dm) of particles depends on its size and charge.18 Therefore

a mobility diameter in the DMA results in the selection of different particle diameters.

The SMPS (or DMA) is designed to use the sample flow as the source of air for the

sheath flow. This feature is intended to equalize the sheath flow temperature and RH with that of

the sample flow. It was observed that it take many hours for the sheath flow to reach the RH of

the sample flow, especially for high RH. This is in agreement with discussions with the

manufacturer representative. Therefore all the measurements were conducted following a

conditioning period of at least 24 hours. During this period the SMPS is exposed to humid air

from the atomizer. The final RH of each of the flows was measured with a RH meter (Vaisala

Inc., Model HMT 331). The RH meter also requires a number of hours to reach equilibrium to

provide a consistent reading. Measurements showed that during high RH measurements, the RH

at the entrance to SPLAT varied between 74 and 85% from run to run, but remained within ±1%

Page 108: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

91

within each run. During each experiment the RH of the aerosol flow was continuously measured

0.5 inches ( 1.3 cm) from the inlet to SPLAT.

Three types of measurement were carried out as depicted in the Figure 4.2 schematic. All

of the measurements were done on DMA size selected particles. Also notice that in all schemes

the wet poly-disperse aerosol flow is sent to a DMA conditioned at high RH. Scheme (1) is the

wet cycle in which DMA size selected droplets (dm,wet) are sampled directly at high RH with

SPLAT for dva measurements. In scheme (2) the size selected droplets were dried by passing the

aerosol flow through a diffusion dryer (TSI Inc., Model 3062). Then the resulting dry mono-

disperse aerosol flow is sampled with SPLAT. For all dry runs the RH at the entrance to SPLAT

was measured to be below 7%. The only difference between these two schemes is that particles

are sample wet or dry by SPLAT. Therefore, the dva measurements for wet and dry runs provide

direct observations of the particle’s water evaporation in the aerodynamic lens inlet.

Lastly, in a number of dry runs as in scheme (3), a second DMA operated at low RH was

used to measure the mobility diameter distribution of the dried aerosol particles. The measured

dry mobility diameters (dm,dry) were checked against the measured vacuum aerodynamic

diameters for ammonium sulfate and sodium chloride. Similar experiments have been conducted

for (NH4)2SO4 and NaC particles when measuring the particle’s dynamic shape factors.19

Page 109: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

92

F

igu

re 4

.2

Exp

erim

enta

l sc

hem

atic

for

thr

ee t

ypes

of

mea

sure

men

ts.

In

all

mea

sure

men

ts d

ropl

ets

are

size

se

lect

ed w

ith

a D

MA

con

diti

oned

at h

igh

RH

. In

(1)

the

wet

mon

o-di

sper

se a

eros

ol f

low

is s

ent t

o S

PL

AT

for

dva

mea

sure

men

ts.

In

setu

p (2

) th

e pa

rtic

les

are

drie

d th

roug

h a

diff

usio

n dr

yer

and

then

del

iver

ed to

the

spec

trom

eter

. In

(3)

par

ticl

es a

re d

ried

and

then

sen

t to

the

DM

A c

ondi

tion

ed

at lo

w R

H to

mea

sure

d dr

y m

obil

ity

diam

eter

(d m

) di

stri

buti

ons.

Page 110: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

93

4.3 Results and Discussion

4.3.1 (NH4)2SO4 Particles

(NH4)2SO4 droplets were size selected with the high RH DMA for a mobility diameter of

206 nm and sent to SPLAT for the vacuum aerodynamic diameter (dva) distribution

measurements. Figure 4.3 shows two measured dva distributions; one for a wet run (as in Figure

4.2 scheme 1) and the other one for a dry run (as in Figure 4.2 scheme 2). The mobility diameter

depends on particle size and charge. Therefore, the DMA set for one mobility diameter results in

different particle diameters that are selected. For example, a DMA dm = 206 nm also results in

doubly charged 330 nm, triple charged 446 nm and quadruple charged 558 nm particles. Note in

Figure 4.3 that the different multiply charged particles can be distinguished in the dva

distributions for a mobility diameter of 206 nm selected by the DMA. The original particle

diameters (dm,wet) are indicated in the dva distributions in Figure 4.3. Also note that the singly

charged 206 nm particles were not detected in SPLAT.

The RH in the DMA is 84 ± 1%. The RH in the DMA was calculated using the wet

mobility diameter (dm,wet) selected in the high RH DMA and the measured dva for the dry runs (as

in Figure 4.2 scheme 2). The size selected droplets are dried before being sampled by SPLAT

corresponding to a dry particle mobility diameter (dm,dry). The ratio of the dm,wet to dm,dry for the

dry runs is the growth factor for the particle. Knowing the growth factor allows for the

determination12,20 of RH precisely. This process is described in Chapter III for the growth factor

analysis of another particle type (NaNO3 particles) in Figure 3.4.

The vacuum aerodynamic diameter is related to shape factors and the dry mobility

diameter by19:

Page 111: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

94

0.096 - ~

1

1.094 2

m,dryd

vad (4.1)

In this equation o and p are standard (1.0 g cm-3) and particle density, respectively, dm,dry is the

dry particle mobility diameter and ~ is the average shape factor for the dry particle. The dry

mobility diameter, dm,dry, can be calculated using equation 4.1. All the parameters in equation

4.1 have been published for a number of inorganic salt particles.19 This approach yields a

calculated RH of 84 ± 1% RH for the run in Figure 4.3 which is in very good agreement with the

measured RH of 83.5% with the humidity meter.

Figure 4.3 Two vacuum aerodynamic size distributions of ammonium sulfate size selected droplets with dm = 206 nm at 84% RH. The dashed and solid lines indicate dry and wet runs, respectively.

Page 112: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

95

The dashed line in Figure 4.3 is the measured vacuum aerodynamic size distribution of

dry (NH4)2SO4 particles. The solid line is the measured dva distribution for (NH4)2SO4 droplets

and sampled directly into SPLAT. The two distributions are nearly the same for dry particles

and droplets before sampling with SPLAT.

The measurements were repeated a number of times at different RH and the results are

presented in Table 4.1. A range of mobility diameters (dm,wet = 330–580 nm) were selected for

the measurements at 74, 79 and 84% RH. The first column lists the calculated RH for each of

the runs. The next two columns provide the ratios of the observed vacuum aerodynamic

diameters dva, and the mobility diameters dm,wet for the dry and wet runs. The last column is a

ratio of the dry and wet vacuum aerodynamic diameters. Note that going from 84% to 74% RH

results in an increase in dva of approximately 12% for a given mobility diameter. Also the

difference between dry and wet runs is between 2-3% for a given RH and mobility diameter.

Table 4.1 Summary of dva Measurements of (NH4)2SO4 Droplets Selected by the DMA at three RH values.

RH% Dry dva/dm

Wet dva/dm

Dry/Wet

84 ± 1 1.08 ± 0.01 1.06 ± 0.01 1.02 ± 0.01

79 ± 1 1.15 ± 0.01 1.13 ± 0.01 1.02 ± 0.01

74 ± 1 1.21 ± 0.01 1.18 ± 0.01 1.03 ± 0.01

Vacuum aerodynamic diameter depends on particle size, density and shape as seen by

equation 4.1. Figure 4.4 shows the calculated vacuum aerodynamic diameter (line) as a function

of water loss of an ammonium sulfate droplet. The droplet was size selected at 330 nm by the

DMA at 79 ± 1% RH. Loss of water decreases particle size. This in turn simultaneously

Page 113: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

96

increases the (NH4)2SO4 weight fraction and hence droplet density. Decrease in size and increase

in density yield a gradually decreasing vacuum aerodynamic diameter as indicated in Figure 4.4.

The calculations stop when 80% of the droplet water is evaporated since at this point the droplet

is expected to effloresce and form a dry crystalline particle. For reference, the measured vacuum

aerodynamic diameter of 372 ± 4 nm for wet runs and 382 ± 4 nm for dry runs are indicated by

▲ and ∆, respectively.

Figure 4.4 Calculated dva as a function of water content for an (NH4)2SO4 droplet with dm,wet = 330 nm at 79 ± 1% RH. The efflorescence point and the measured vacuum aerodynamic diameters for wet and dry runs are indicated by ▲ and ∆, respectively.

Page 114: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

97

The measured dva in the wet run (▲) could be interpreted in two different ways. It could

represent a saturated (NH4)2SO4 droplet that has lost approximately 80% of its water content and

is near its crystallization point as observed in Figure 4.4. On the other hand, it could be a dry

crystalline (NH4)2SO4 particle with a dynamic shape factor of about 1.07 which is higher than the

previously measured19 value of 1.04 ± 1. Rapid droplet drying could yield higher dynamic shape

factors, consistent with previous observations19 and this process could be happening inside the

aerodynamic lens inlet.

4.3.2 NaC Particles

Sodium chloride measurements showed a more complex behavior than ammonium

sulfate. Figure 4.5 shows the measured vacuum aerodynamic size distributions for dry (dashed

line) and wet (solid line) runs. The runs are for a DMA size selected NaC droplet of 558 nm.

Use of equation 4.1 and the known parameters12,19-21 for NaC yields a calculated RH of 84 ± 2%

RH which is consistent with the measured RH of 83%. The dry run measurement shows one

peak for the dva distribution. The wet run shows a more complicated dva distribution in which

three main peaks are observed and numbered in Figure 4.5.

Page 115: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

98

Figure 4.5 Two vacuum aerodynamic size distributions of NaC particles size selected at dm = 558 nm with the DMA at 84% RH. The dashed and solid lines indicate dry and wet runs, respectively. The circled numbers label three observed NaC “modes”.

More experiments were done over a range of particle sizes at the same relative humidity

conditions. The ratios of dva to dm were plotted for the dry run and the three main peaks observed

in the wet run as a function of droplet mobility diameter and the results are shown in Figure 4.6.

The ratio for the dried particles remains constant as a function of particle mobility diameter

(dm,wet) as seen in the open squares. This is expected since only one aerodynamic diameter is

observed in the dva distribution. The filled symbols represent the behavior of the three main

peaks from Figure 4.5 for the wet runs as a function of droplet size (dm,wet). The three “modes”

suggest that the NaC droplets larger or equal to 450 nm lose water in the aerodynamic lens inlet

in three different ways. This physical transformation and mechanism can be very complex and is

Page 116: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

99

not the purpose of this study to model it. These measurements show that the vacuum

aerodynamic size distributions of NaC droplets do not offer sufficient information that they can

be used to measure original particle size, density or water content. Consequently, the different

peaks or modes in the wet run could mislead the interpretation of the dva distributions. Also,

ambiguity about the NaC three “modes” being multiply charged particles measured by SPLAT

was eliminated by selecting mobility diameters larger than 500 nm. Multiply charged particles

with dm = 500 nm are larger than 700 nm in diameter. SPLAT aerodynamic lens inlet

transmission efficiency is the highest for 300 nm particles. Then it drops significantly for

particles larger than 700 nm. Therefore, the larger multiply charged particles with dm = 500 nm

are not sampled and detected by SPLAT.

Page 117: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

100

Figure 4.6 Plots of the ratios of the measured vacuum aerodynamic and mobility diameters as a function of the particle mobility diameters selected by the DMA at 84% RH. Open symbols represent dried particles, filled symbols indicate measurements of wet droplets at high RH. The three NaC particle “modes” are indicated in the numbered circles.

4.3.3 Other Particles

The same measurements were done on other inorganic and organic salt droplets. These

are: sodium nitrate (NaNO3), potassium chloride (KC), rubidium chloride (RbC), tetramethyl

ammonium chloride ((CH3)4NC) and dodecyltrimethylammonium chloride

(CH3(CH2)11N(CH3)3C). Table 4.2 presents a summary of dva to dm ratios for dry and wet runs

and the dry to wet results ratios. Also it shows the particle size range and dry particle density.

Ammonium sulfate is presented for comparison.

Page 118: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

101

Table 4.2 Summary of Measurements on Six Different Particle Types

Compound Density

g cm-3

Particle sizes (nm)

dry dva/dm

wet dva/dm

dry/wet

NaNO3 2.10 260-580 1.52 ± 0.01 1.49 ± 0.02 1.02 ± 0.02

(CH3)4NC 1.17 330-780 0.76 ± 0.01 0.79 ± 0.03 0.98 ± 0.03

CH3(CH2)11N(CH3)3C 0.89 210-710 0.87 ± 0.02 0.82 ± 0.04 1.03 ± 0.04

(NH4)2SO4 1.75 330-580 1.08 ± 0.006 1.06 ± 0.006 1.02 ± 0.01

KC 1.99 550-650 0.73 0.70 1.04

RbC 2.76 322-550 1.10 1.13 1.02

NaNO3 aqueous particles have been shown that do not effloresce to form dry crystalline

particles even at very low RH.22,23 The wet dva shows a larger spread in the distributions that can

be attributed to a size dependent behavior. Regardless of the spread the difference between dry

and wet measurements is small at approximately 2% which is consistent with previous

measurements.23 Also, KC, RbC and (CH3)4NC show only minor differences in the wet and

dry run distributions. These particles were found to lose significant amounts of water whether

they were dried or not dried before being sampled with the particle spectrometer.

CH3(CH2)11N(CH3)3C is a twelve carbon chain water soluble organic salt with a crystal

density of 0.89 g cm-3. This organic salt was chosen because the presence of hydrophobic

compounds is expected to slow water evaporation as predicted by the “inverted micelle” model

for marine aerosols.24 Even though these particles show a larger spread between dry and wet

measurements than the other salt particles, the difference is very small. Measurement of

mobility diameters for a droplet at high RH and after drying were done and showed only minor

differences as shown in the numbers by the last column in Table 4.2.

Page 119: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

102

It is important to know that drying particles prior to sampling does not ensure that

particles have no water content. For example, when drying Ca(NO3)2 aqueous particles an

amorphous phase is formed containing a small amount of water that is difficult to quantify. All

particle types, except NaC, lose significant amount of water in the aerodynamic lens inlet in a

wide range of component’s density (0.89-2.76 g cm-3). The differences between wet and dry

runs for almost all particle types were small and the ratios between dry and wet runs are close to

1. The only particle type that showed major differences was NaC particles as shown in Figures

4.5 and 4.6.

4.4 Conclusion

The measurements carried out by sampling different types of aqueous particles with

SPLAT provided a more quantitative approach for water evaporation in an aerodynamic lens

inlet. The spectrometer sampling of droplets at high RH to measure vacuum aerodynamic

diameters demonstrated that in most cases particles lose a significant amount of water in the

inlet. Measured vacuum aerodynamic diameters were essentially the same for wet and dry runs.

The only exception was NaC particles which showed complex aerodynamic size distributions in

the wet runs that suggest that the particles lose water in at least three different ways.

Consequently, the aerodynamic size distribution could be misinterpreted based on the three

“modes” observed for NaC droplets.

Many single particle mass spectrometers use the aerodynamic lens inlet to sample

ambient particles. Water in the particles potentially could affect particle transmission efficiency

in the inlet, relative intensity of ions in the mass spectra and the laser power threshold to generate

ions from particles.25,26 Drying particles before sampling is recommended since almost all

Page 120: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

103

particles in this study lose their water content in the inlet. Drying is not an insurance that

particles will not retain water. On the other hand, it will provide more reproducible and

comparable measurements reducing ambiguities in vacuum aerodynamic measurements and data

analysis.

Page 121: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

104

References for Chapter IV

(1) Gard, E.; Mayer, J. E.; Morrical, B. D.; Dienes, T.; Fergenson, D. P.; Prather, K. A., Real-time analysis of individual atmospheric aerosol particles: Design and performance of a portable ATOFMS, Analytical Chemistry 1997, 69, 4083-4091.

(2) Thomson, D. S.; Schein, M. E.; Murphy, D. M., Particle analysis by laser mass spectrometry WB-57F instrument overview, Aerosol Science and Technology 2000, 33, 153-169.

(3) Cziczo, D. J.; DeMott, P. J.; Brock, C.; Hudson, P. K.; Jesse, B.; Kreidenweis, S. M.; Prenni, A. J.; Schreiner, J.; Thomson, D. S.; Murphy, D. M., A method for single particle mass spectrometry of ice nuclei, Aerosol Science and Technology 2003, 37, 460-470.

(4) Su, Y.; Sipin, M. F.; Furutani, H.; Prather, K. A., Development and Characterization of an Aerosol Time-of-Flight Mass Spectrometer with Increased Detection Efficiency, Analytical Chemistry 2003, 76, 712-719.

(5) Huffman, J. A.; Jayne, J. T.; Drewnick, F.; Aiken, A. C.; Onasch, T.; Worsnop, D. R.; Jimenez, J. L., Design, modeling, optimization, and experimental tests of a particle beam width probe for the Aerodyne aerosol mass spectrometer, Aerosol Science and Technology 2005, 39, 1143-1163.

(6) Jayne, J. T.; Leard, D. C.; Zhang, X. F.; Davidovits, P.; Smith, K. A.; Kolb, C. E.; Worsnop, D. R., Development of an aerosol mass spectrometer for size and composition analysis of submicron particles, Aerosol Science and Technology 2000, 33, 49-70.

(7) Schreiner, J.; Schild, U.; Voigt, C.; Mauersberger, K., Focusing of aerosols into a particle beam at pressures from 10 to 150 Torr, Aerosol Science and Technology 1999, 31, 373-382.

(8) Zelenyuk, A.; Imre, D., Single particle laser ablation time-of-flight mass spectrometer: An introduction to SPLAT, Aerosol Science and Technology 2005, 39, 554-568.

(9) Liu, P.; Ziemann, P. J.; Kittelson, D. B.; McMurry, P. H., Generating particle beams of controlled dimensions and divergence .1. Theory of particle motion in aerodynamic lenses and nozzle expansions., Aerosol Science and Technology 1995, 22, 293-313.

(10) Liu, P.; Ziemann, P. J.; Kittelson, D. B.; McMurry, P. H., Generating particle beams of controlled dimensions and divergence .2. Experimental evaluation of particle motion in aerodynamic lenses and nozzle expansions., Aerosol Science and Technology 1995, 22, 314-324.

Page 122: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

105

(11) Mallina, R. V.; Wexler, A. S.; Johnston, M. V., Particle growth in high-speed particle beam inlets, Journal of Aerosol Science 1997, 28, 223-238.

(12) Tang, I. N.; Munkelwitz, H. R., Water Activities, Densities, and Refractive-Indexes of Aqueous Sulfates and Sodium-Nitrate Droplets of Atmospheric Importance, Journal of Geophysical Research-Atmospheres 1994, 99, 18801-18808.

(13) DeCarlo, P. F.; Slowik, J. G.; Worsnop, D. R.; Davidovits, P.; Jimenez, J. L., Particle morphology and density characterization by combined mobility and aerodynamic diameter measurements. Part 1: Theory, Aerosol Science and Technology 2004, 38, 1185-1205.

(14) Murphy, D. M., The design of single particle laser mass spectrometers, Mass Spectrometry Reviews 2007, 26, 150-165.

(15) Dahneke, B.; Padliya, D. In Progress in Astronautics and Aeronautics; Potter, J., Ed.; American Institute of Aeronautics and Astronautics: New York, 1977; Vol. 51.

(16) McGraw, R.; Private communication ed.; Zelenyuk, A., Imre, D., Eds.: Richland, WA, 2002.

(17) Cabalo, J.; Zelenyuk, A.; Baer, T.; Miller, R. E., Two-color laser induced evaporation dynamics of liquid aerosols probed by time-of-flight mass spectrometry, Aerosol Science and Technology 2000, 33, 3-19.

(18) TSI, I. Series 3080 Electrostatic Classifier Manual: Shoreview, MN, 2006; Vol. Revision G.

(19) Zelenyuk, A.; Cai, Y.; Imre, D., From agglomerates of spheres to irregularly shaped particles: Determination of dynamic shape factors from measurements of mobility and vacuum aerodynamic diameters, Aerosol Science and Technology 2006, 40, 197-217.

(20) Clegg, S. L.; Brimblecombe, P.; Wexler, A. S., Thermodynamic model of the system H+-NH4

+-Na+-SO42--NO3--Cl--H2O at 298.15 K, Journal of Physical Chemistry A 1998, 102,

2155-2171.

(21) Tang, I. N.; Tridico, A. C.; Fung, K. H., Thermodynamic and optical properties of sea salt aerosols, Journal of Geophysical Research-Atmospheres 1997, 102, 23269-23275.

(22) Hoffman, R. C.; Laskin, A.; Finlayson-Pitts, B. J., Sodium nitrate particles: physical and chemical properties during hydration and dehydration, and implications for aged sea salt aerosols, Journal of Aerosol Science 2004, 35, 869-887.

Page 123: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

106

(23) Zelenyuk, A.; Cai, Y.; Chieffo, L.; Imre, D., High precision density measurements of single particles: The density of metastable phases, Aerosol Science and Technology 2005, 39, 972-986.

(24) Ellison, G. B.; Tuck, A. F.; Vaida, V., Atmospheric processing of organic aerosols, Journal of Geophysical Research-Atmospheres 1999, 104, 11633-11641.

(25) Neubauer, K. R.; Johnston, M. V.; Wexler, A. S., On-line analysis of aqueous aerosols by laser desorption ionization, Int. J. Mass Spectrom. Ion Process. 1997, 163, 29-37.

(26) Neubauer, K. R.; Johnston, M. V.; Wexler, A. S., Humidity effects on the mass spectra of single aerosol particles, Atmospheric Environment 1998, 32, 2521-2529.

Page 124: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

________________

Chapter V

Measurements and Interpretation of the Effect of a Soluble Surfactant on the Density, Shape and Water Uptake of Hygroscopic Particles

________________

5.1 Introduction

Atmospheric particles have been found to contain a significant amount of organic

material.1,2 The chemical and physical properties of all atmospheric particles are strongly

influenced by their size, composition and the environmental conditions, especially relative

humidity (RH). Aerosol particles are responsible for cloud formation and growth by acting as

cloud condensation nuclei (CCN). Ellison et al. proposed3 a model for the chemical structure for

organic marine aerosols as described in Chapter I. In this model organic marine aerosols are

droplets with a salt-water core encapsulated by organic surfactants such as fatty acids (R-CO2-).

The source of the organic surfactants (i.e., lipids, fatty acids) is the marine life biological activity

that is enriched at the ocean’s surface. The polar end of the surfactant is pointing towards the

droplet’s core while the long hydrocarbon chain is at the surface resulting in an “inverted”

micelle structure. The “inverted” micelle model predicts that cloud growth is inhibited by the

organic hydrophobic layer at the aerosol surface making marine aerosols poor cloud

condensation nuclei (CCN).3,4 Atmospheric processing of the inert surface by radical oxidation

will transform the surface from hydrophobic to hydrophilic and therefore making the particles

better CCN. These types of particles have been found in the marine boundary layer.

Since the organic surfactants at the marine aerosol surface could influence cloud

formation, a better characterization of the particles’ water uptake is needed. A common

experiment is to measure growth factors of salt-water particles coated by surfactants to describe

Page 125: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

108

how much water is gained based on surfactant concentration. Here, GF and particle density

measurements are combined to describe water uptake of this complex particle mixture as a

function of surfactant concentration. This will test directly the hypothesis made by Ellison et

al.3, in which the surfactant-coated particles are predicted to inhibit particle growth by water

uptake, making them poor CCN.

The “inverted micelle” model chemical structure, chemical transformation and potential

atmospheric impacts stimulated a number of studies to quantify the effect of surfactants on

particles properties such as the interaction with H2O vapor.5-8 It has been observed that organic

molecules such as octanoic acid, lauric acid, sodium dodecyl sulfate and sodium oleate decrease

the growth factors of sodium chloride7 (NaC) and calcium chloride5,6 (CaC2) particles. It was

found that the maximum observed growth factors for pure organic particles was only 7.5% at

85% RH. The growth factors reach values from 56-83% for internally mixed CaC2/organic

mixed aerosols with an organic weight fraction of 0.25. Moreover, it was shown6 that the

addition of glutaric and pyruvic acids lowered the deliquescence point of NaC. The lowest

deliquescence point observed was 71% RH for an organic weight fraction of 0.60. The addition

of organic molecules does not prevent water uptake or loss on hygroscopic particles. Instead it

changes the amount of water gain or lost with the general trend of decreasing growth factors as

the organic concentration is increased.

The effect of surfactants on the CNN activity of some inorganic salt particles has been

addressed.9 It was observed that organic molecules have effects on solubility, particle weight or

density and surface tension. All of these effects tend to counteract each other producing smaller

than expected changes in CNN activity. The effect of surfactants on the CNN activity of

ammonium sulfate ((NH4)2SO4) particles has been studied.10 It was observed that the surfactants

Page 126: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

109

altered particles’ CNN activation but did not inhibit it with the exception of stearic acid.10 It was

suggested that H2O vapor may not have enough time to diffuse through the thick layer of this

surfactant.

A recent review11 provided a summary of the effects of surfactants on atmospheric

particles. They noted that solubility of the organic surfactant could affect the particles

properties. Soluble and insoluble surfactant effects are expected present as organic coatings in

particles.11 Soluble surfactants could partition to the bulk in liquid particles and therefore form a

porous layer at the droplet’s surface. The porous layer is expected to decrease the potential

inhibitory effects on CNN activity. On the other hand, insoluble surfactants form a tighter

coating on the droplet’s surface since they cannot partition to the bulk. This layer would be more

difficult to penetrate affecting strongly CNN activity and water uptake. Nevertheless, stearic

acid has been the only organic surfactant observed to completely inhibit CNN activity of

common hygroscopic salt particles.

The current studies strongly suggest7,10,12,13 that hygroscopic particles coated by

surfactants gain and lose water. The presence of organic surfactants on particles is expected to

alter the particles’ physical properties as predicted by the “inverted micelle” model for marine

aerosols. Therefore it is important to characterize the effect of organic surfactants on particle

growth as a function of RH and organic fraction concentration. The main goal of this study is to

physically characterize of common hygroscopic salt particles coated with surfactants by

measuring growth factors, particle density and shape as a function of organic surfactant

concentration.

Page 127: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

110

5.2 Experimental

Droplets were generated by forming aerosols from water solutions of pure inorganic salts,

SDS or their mixtures using a TSI atomizer (TSI Inc., Model 3076). The solutions were

prepared by dissolving the appropriate amounts of sodium chloride (NaC), sodium nitrate

(NaNO3), or sodium dodecyl sulfate (NaSO4(CH2)11CH3) in ultrapure, deionized water. It is

assumed that surfactants partition mostly to the droplet’s surface producing an organic coating

on the droplet. The remaining surfactant molecules reside in the bulk of the droplet.

Three types of measurements were carried out and are numbered in Figure 5.1. All

measurements were done throughout the full concentration range as a function of organic weight

fraction: from pure inorganic salt to pure organic surfactant particles. Figure 5.1 (1) shows the

scheme to measure particle effective density. The poly-disperse aerosol stream from the

atomizer is dried by passing the flow through two diffusion dryers (TSI Inc. Model 3062)

connected in series. The dried particles are size-selected by their mobility diameter (dm) using a

differential mobility analyzer (DMA, TSI Inc. 3081) operated at low RH (< 5%). Sequentially

particles are delivered to the single particle laser ablation time-of-flight mass spectrometer

(SPLAT) for vacuum aerodynamic diameter (dva) measurement.14 The simultaneous

measurements of vacuum aerodynamic and mobility diameters are used to calculate particle

density14,15 by:

m

vap d

d0 (5.1)

or

)(

)(1 0

20 mc

pvacp

m

vaeff dC

dC

d

d

(5.2)

Page 128: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

111

where 0 , p , vad and md are standard (1g cm-3) and particle densities, vacuum aerodynamic

diameter and mobility diameter, respectively. Equation 5.1 is used for spherical particles. For

non-spherical particles the ratio of the two measured diameters is defined as the particle effective

density ( mvaeff dd ) as in equation 5.2, is approximately an average of the dynamic shape

factor (DSF) of the particle in the transition regime of the DMA and in the free molecular

regime15 of SPLAT and )(dCc is the Cunningham correction factor. For small shape factors it

was empirically shown15,16 that it is possible to approximate the DSF with Equation 5.3:

096.0094.11

2

p

eff

(5.3)

Page 129: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

112

Fig

ure

5.1

M

easu

rem

ent

to c

hara

cter

ize

inor

gani

c sa

lts/

SDS

par

ticl

es h

ygro

scop

ic p

rope

rtie

s.

(1)

show

s th

e sc

hem

e to

mea

sure

par

ticl

e de

nsit

y.

(2)

and

(3)

are

the

sche

mes

to

mea

sure

GF

by

incr

easi

ng th

e R

H a

nd d

ecre

asin

g R

H, r

espe

ctiv

ely.

Page 130: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

113

Equation 5.1 is used with the two measured diameters (dm and dva) of NaNO3/SDS

internally mixed particles to calculate the density of the SDS fraction as a function of its

concentration while assuming that the NaNO3 density remains constant. These calculated

densities are then used to interpret the measured growth factors (GF). Dynamic shape factors

become important for NaC/SDS since pure sodium chloride dries to a cubic crystalline form.

Therefore equations 5.2 and 5.3 are used to describe the NaC/SDS particles with the same

approach as NaNO3/SDS particles.

A detailed description of the use of SPLAT to measure particle densities is given in

Zelenyuk et al.14,15 The aerodynamic diameter of each particle is obtained by measuring the

particle’s transit time between the two optical detection stages spaced by 4.125 (10.5 cm) and

comparing it to a calibration curve established with polystyrene latex spheres (see Chapter IV)

used as NIST certified size standards. The sizing accuracy for this set of measurements is ±1%.

To characterize the inorganic salt particles coated with the SDS surfactant, they were

collected on a substrate for analysis using Scanning Electron Microscopy (SEM). SEM images

were taken at different surfactant concentrations and were used as a guide for shape factor

determination and particle structure.

Growth factor measurements were done as depicted in Figure 5.1, schemes (2) and (3).

In (2) the GF measurements were done by increasing the RH. For these sets of runs the poly-

disperse aerosol was first dried with two diffusion driers in series. The dried particles were size

selected with a DMA operated at low (<5%) RH. Then the dried mono-disperse particles were

grown by water uptake in a humidity conditioning chamber in which the residence time was

approximately 90 seconds. The resulting humidified particles were sampled by a Scanning

Mobility Particle Sizer (SMPS, TSI Inc., Model 3936) at a rate of 0.3 L min-1. The RH in the 2nd

Page 131: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

114

DMA was brought to equilibrium with the RH in the humidified conditioning chamber for

several hours before measurements, as described in previous studies and Chapter IV.17

In scheme (3) the GF measurements were done by decreasing the RH. In this set of

experiments, the poly-disperse aerosol flow was first sent through the conditioning chamber in

order to establish a constant high RH. The wet particles were size selected with a DMA that was

operated at the same RH. The resulting mono-disperse humidified aerosol flow was dried by

passing through the two diffusion dryers and sampled by the DMA that was operated at low RH

(< 5%).

The scanning on the 2nd DMA to get mobility diameter distributions is done by coupling

the SMPS to a condensation particle counter (CPC) for both GF measurement schemes. To

avoid any ambiguities due to multiply charged particles selected in the DMA, mobility diameters

to match those of multiply charged particles were selected. For example, when the DMA is set

to select 296 nm particles it also selects doubly charged 500 nm particles with it. Therefore

additional experiments were done by selecting particles with a mobility diameter of 500 nm with

the DMA.

The RH of the wet and dry aerosol flows and DMA sheath flows were monitored with

RH meters (Vaisala Inc., Model HUMICAP HMT 75).

Measurements of GF are done by measuring the mobility diameter of DMA size-selected

particles in which one DMA is operated at high RH and the other at low RH. Therefore, GF is

defined as the ratio of the two measured mobility diameters by:

drym

wetmm d

dGF

,

, (5.4)

Page 132: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

115

where dm,wet and dm,dry are measured wet and dry mobility diameters. Moreover, mobility

diameters depend on particle size and shape and hence the measured18-20 growth factor, GFm.

Dynamic shape factors have been used to correct measured GFm for NaC particles to achieve

better agreement between measurements and calculations.

In order to model growth factor measurements a simplified Zdanovskii-Stokes-Robinson

(ZSR) approach is used for the internally mixed salt/SDS particles.21-24 In this model it is

assumed that the organic and inorganic components take up water independently described by

each component growth factor. Also volume additivity and particle sphericity are assumed

which results in the following GF expression:

1/333 ININOO GFGFGF (5.5)

where O and IN are the volume fractions of the organic and inorganic components in the dry

particle, respectively, and OGF and INGF are the growth factors of the organic and inorganic

fractions, respectively.25,26 The volume fractions are calculated from each component’s mass

ratio and density in the pure form. In this simplified approach the inorganic and organic GF and

densities are constant and independent of particle composition.

The ZSR model may have limitations describing internally mixed salt/SDS particles.

Surfactant properties such as solubility, critical micelle concentration (cmc), micelle aggregation

number, size and shape are influenced by the addition of inorganic salts and their concentration.

The crystalline density of SDS is 0.37 g cm-3, but it is known that in the presence of water it can

form different mesophases that are difficult to completely dry.27-30 This SDS surfactant

characteristic combined with the addition of an inorganic salt (in this study NaNO3 and NaC)

makes these particles more complex to describe.

Page 133: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

116

5.3 Results and Discussion

5.3.1 NaNO3 Particles Coated with SDS

5.3.1.1 NaNO3/SDS Particles Density Measurements

Sodium nitrate was selected because its particles do not crystallize even at very low RH

or extreme conditions and always retain some water.16,31-33 The data also suggest that a small

amount of water is retained in “dry” SDS particles, thus it is assumed that the “dry” NaNO3/SDS

particles remain spherical. For that reason, particle density could be calculated with equation 5.1

using the measured vacuum aerodynamic and mobility diameter distributions.

Figure 5.2 shows the measured average dry particle density as a function of SDS weight

fraction (WF) for two DMA selected sizes: 296 nm and 500 nm. The particle density decreases

as the SDS weight fraction increases, with the lowest density at a WF of approximately 0.3. As

the SDS WF continues to increase, the particle densities increase to 1.13 g cm-3 for the pure dry

SDS particles. This density suggests that dry SDS particles retained a small amount of water

since the SDS dry crystalline density is 0.37 g cm-3.

Three fits to the data are shown in Figure 5.2. The particle density was calculated and

modeled using the known NaNO3 particle density32,33 of 2.13 g cm-3, the SDS crystalline density

of 0.37 g cm-3 and the weight fraction of each component, as shown by the dashed line. This

approach only fits the data at one point, the pure NaNO3 particles, and fails to describe any other

density measurement. The same approach is done for the dash-dot fit but using the measured dry

SDS particle density of 1.13 g cm-3 instead of its crystalline density. In that case the calculated

density only fits at two points (see Figure 5.2): the pure NaNO3 and the pure SDS.

Page 134: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

117

.

Figure 5.2 Measured and fit densities of internally mixed NaNO3/SDS particles with mobility diameters of 296 nm and 500 nm as a function of SDS concentration. The dashed and dash-dot lines are calculated assuming constant SDS densities of 0.37 g cm-3 and 1.13 g cm-3, respectively. The solid line shows an empirical fit to the data.

Since using the SDS crystalline and measured densities does a poor job in explaining the

measured particle densities, an empirical approach was used to fit the data. The solid line in

Figure 5.2 shows this empirical fit. For each data point a SDS density is calculated to fit the

measured particle density. In this approach it is assumed that the NaNO3 density remains

constant (2.13 g cm-3) and the volume ratios are proportional to each component’s dry mass

ratio. The calculated SDS density, SDS, is then fitted resulting in:

Page 135: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

118

432-3 F 2.14F 5.19-F 3.94 10 3.61-0.24 WWWWFSDS (5.6)

The surfactant properties such as cmc and structure are dependent on concentration and the

addition of inorganic salts.27-30 Equation 5.6 describes the changing density of the SDS fraction

as a function of its concentration in the internally mixed particle assuming that the NaNO3

density remains constant.

5.3.1.2 NaNO3/SDS Growth Factor Measurements

Two different schemes were used to measure the GF of NaNO3/SDS particles. One

scheme is for the GF measurements by increasing the RH. In this scheme dry particles are size-

selected (by diameter) and re-humidified to measure the droplets’ size distribution at high RH.

The second scheme involves decreasing the RH. In this method, droplets are conditioned and

size-selected at high RH. The droplet stream is then dried to measure the size distribution at low

RH. A detailed description of the experiments can be found in section 5.2. The GF

measurements were found to be independent of the procedure.

Figure 5.3 shows the observed GF measurements of NaNO3/SDS particles at 86 ± 1%

RH. DMA size-selected particles were in the 200–700 nm diameter range. For this size range

the mobility size distributions were found to be independent of particle size and composition.

The GF factor decreases as the organic weight fraction is increased, as observed previously.5-8

Figure 5.3 shows the fits using the ZSR approach (dashed and dash-dot lines) to calculate

GF with equation 5.5. This approach assumes a SDS crystalline density of 0.37 g cm-3 (dashed

line) and pure SDS particle density of 1.13 g cm-3 (dash-dot line). These densities were used to

calculate the SDS volume weight fractions as a function of SDS concentration. The growth

factors used in these calculations were 1.73 and 1.09 for pure NaNO3 and SDS particles,

respectively. Note that the SDS densities used for the fits are incorrect and only fit at the two

Page 136: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

119

concentration limits by construction. On the other hand, calculating GF using the empirical SDS

density (equation 5.6) obtained from the density measurements, achieves a much better

agreement as shown by the solid line. Using the changing SDS density reproduced the measured

growth factors quantitatively. Varying SDS concentration changes its volume fraction but also

changes the SDS density which is different than the constant SDS density assumed in the

simplified ZSR approach.

Figure 5.3 Measured growth factors of NaNO3/SDS particles as a function of SDS WF. The dashed and dash-dot lines show fits assuming constant SDS densities of 0.37 g cm-3 and 1.13 g cm-3, respectively. The solid line uses the SDS density that was calculated on the basis of the data shown in Figure 5.2.

Page 137: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

120

5.3.1.3 Vacuum Aerodynamic Size Distributions of NaNO3/SDS Particles

The observed densities in Figure 5.2 do not show any error bars, instead this figure shows

the average density that is calculated based on the average dva obtained from vacuum

aerodynamic diameter distributions. Figure 5.4 shows the dva distributions for dry NaNO3/SDS

particles with a mobility diameter of 500 nm for a number of SDS weight fractions, from pure

NaNO3 to pure SDS. The data show that the line-width in the distribution changes significantly

as the SDS WF is increased. For low SDS WF the line-widths are narrow. But as the SDS

concentration exceeds a WF of 1.3 x 10-2 the line-widths increase in the dva distributions and the

shape become more complex. The dva distribution line-width for pure NaNO3 particles is 10%,

defined as full-width at half-maximum (FWHM). As the SDS concentration increases the line-

width reaches approximately 65% at a SDS WF of 6.4 x 10-2. If the SDS concentration is

increased further the line-width slowly returns to the original 10% FWHM.

The NaNO3/SDS particles dva distributions showed a long tail in the smaller aerodynamic

diameters or lower density side. The line shape for particles containing small amounts of SDS

was found not to be reproducible. The irreproducibility might be due to the possible mesophases

that the surfactant can form as the concentration is changed. Water retention by these particles

could be influenced by the drying conditions which have been observed to have impact on

density measurements.15,16,32,33

Page 138: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

121

Figure 5.4 Measured dva distributions of 500 nm mobility diameter NaNO3/SDS particles for a number of SDS weight fractions indicated in the upper left corners.

Page 139: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

122

During these studies we hypothesized that the spread in the dva distributions (i.e.,

densities) is due to water retention based on the surfactant mesophases formed. Therefore

changing the drying conditions before sampling by SPLAT should be indicative of the amount of

water retained by the particles.

The particles were passed through a heated tube maintained at approximately 75 Cº. The

residence time in the heated tube is estimated to be about 6.5 s. Heated and unheated particle dva

distributions were measured for a number of SDS weight fractions for dry particles with a

mobility diameter of 500 nm as shown in Figure 5.5. This figure shows two sets of distributions;

the dashed lines represent room temperature (RT) drying with diffusion dryers and the shaded

distributions indicate the addition of the heated tube drying. The particle compositions that have

the broadest distributions at room temperature exhibit the greatest changes upon heating. This

result suggests that there are a number of different types of particles that retain different amounts

of water in the dry aerosol flow making the room temperature distributions broader. Heating

also causes the particles’ density to increase in all cases with the exception of pure SDS particles

which shows a decrease in density upon heating.

Page 140: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

123

Figure 5.5 The effect of heating NaNO3/SDS particles with a mobility diameter of 500 nm on the particle dva distributions and hence densities, for a number of SDS WF values indicated in the upper left corners.

Page 141: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

124

5.3.2 NaC Particles Coated with SDS

5.3.2.1 NaC/SDS Density and Shape Measurements

Sodium chloride particles crystallize to form cubic dried particles. The dynamic shape

factors (DSF) for these particles have been characterized in previous studies.15 The DSF is

required to characterize NaC particles, especially in this study where an organic surfactant is

mixed with the NaC particles. Here the measured dva distributions, GFI, and SEM images are

used to describe NaC/SDS particles.

Figure 5.6 shows the dva distributions of NaC particles coated with SDS with dm,dry =

500 nm for a number of SDS concentrations. The distributions look very similar to the

NaNO3/SDS particles shown in Figure 5.4. The line-width increases as the SDS WF is increased

up to intermediate concentrations. As the SDS concentration continues to increase to pure SDS,

the line-widths decrease. For the pure substances (NaC and SDS) the distributions are narrow,

consistent with previous measurements. The complex line-widths indicate that different particle

types are present and it depends on the amount of water retained by the particles. The pure NaC

particles’ line shape, density and shape factor are also consistent with previously reported

values.15

Page 142: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

125

Figure 5.6 Measured dva distributions of NaC/SDS particles with a mobility diameter of 500 nm for a number of SDS WF values indicated in the upper right hand corners. Note that both line shape and line positions are sensitive to particle SDS concentration.

Page 143: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

126

Figure 5.7 shows the average effective density (equation 5.2) as a function of SDS weight

fraction based on the dva distributions from Figure 5.6. As in the case for NaNO3/SDS, particle

mobility diameters of 296 nm and 500 nm are reported and no error bars are shown. Effective

densities for 296 nm particles are slightly higher than for 500 nm particles but in general the

behavior is the same. The density decreases in the same manner as for the NaNO3/SDS particles.

The difference is that the decrease in the density for NaC/SDS particles is more gradual,

reaching its lowest value at a SDS WF of approximately 0.4. The density at this point is lower

compared to the NaNO3/SDS particles’ lowest density value.

Calculated densities as a function of SDS concentration are also shown in Figure 5.7.

Volume additivity and particle sphericity is assumed. For this calculation a constant density for

NaC (2.165 g cm-3) is used. SDS densities of 0.37 g cm-3 and 1.13 g cm-3 are used, as marked

by dashed and dash-dot lines, respectively. The density calculations do not fit the measured

effective densities. This result was observed for NaNO3/SDS particles as shown in Figure 5.3.

The SDS density is expected to change as a function of concentration. Knowledge of the

particles’ DSF is required since dried NaC/SDS particles are not spherical.

Page 144: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

127

Figure 5.7 Measured and fit densities of internally mixed NaC/SDS particles with mobility diameters of 296 nm and 500 nm as a function of SDS WF. The dashed and dash-dot lines are calculated assuming particle sphericity and constant SDS densities of 0.37 g cm-3 and 1.13 g cm-3, respectively. The solid line shows an empirical fit to the observed data.

Pure NaC particles with dm,dry = 500 nm and dva distribution (as in Figure 5.6) result in a

DSF of 1.10 by using equation 5.3, which is consistent with previous measurements.15,16 The

DSF for pure SDS particles is 1.0. However the shape factors for intermediate concentrations is

more complex and dependent on the SDS concentration. Figure 5.8 shows SEM images of

NaC/SDS particles collected on a substrate for a number of SDS concentrations. The upper

Page 145: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

128

image shows the lowest SDS concentration (WF = 0.0098) in which the mixed particles dried to

a cubic crystalline solid similar to pure NaC particles. For a SDS WF of 0.09 the particles show

less uniformity in their shape with some of them being nearly spherical. For the highest SDS WF

value, the SDS fraction separates from the NaC fraction making the surfactant fraction more

visible in the SEM image. The images also suggest that the particles were deposited wet on the

substrate even though they were dried before sampling. The surfactant fraction is not completely

dry but is a liquid gel-like phase that retains a small amount of water.

Figure 5.8 Microscopic images of internally mixed NaC/SDS particles with SDS WF of 0.0098, 0.090 and 0.37 shown in the upper, middle and lower frames, respectively.

Page 146: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

129

The DSF depends on particle composition as seen in Figure 5.8 and it affects the particle

density calculation, the SDS density and GF calculation. An iterative process is used to derive a

DSF function to calculate the particle density, the SDS density and the GF. The calculated

results were compared to the measured GF and effective densities. The function is then changed

and the process is repeated until a reasonable fit to the entire data set is reached.

Figure 5.9 shows the calculated DSF, the SDS density and calculated particle density

(corrected for DSF) as a function of SDS weight fraction. The DSF decreases slowly from 1.1 to

1.0 as the SDS WF goes from 0 to 0.8. The SDS density increases from 0.39 g cm-3 to 1.13 g

cm-3 with increasing SDS concentration. The corrected calculated particle density describes

more quantitatively the measured effective densities by fitting the data using DSF and SDS

density. The SDS density and DSF functional forms presented in Figure 5.9 are:

5432 0.0554-2.47-5.011.95-0.226.3810 WFWFWFWFWFSDS (5.7)

0.181-0.6871.09-0.9440.467-1.10 5432 WFWFWFWFWFDSF (5.8)

Page 147: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

130

Figure 5.9 Particle dynamic shape factor (DSF), particle calculated effective density and the SDS density as a function of the SDS WF.

5.3.2.2 NaC/SDS Growth Factor Measurements

The growth factors for NaC particles coated with SDS were measured for the complete

composition range by increasing the RH and decreasing the RH. Figure 5.10 shows the observed

growth factors at 82 ± 1% RH. The GF shows an S shaped curve similar to GF for NaNO3/SDS

particles shown in Figure 5.3. The ZSR approach using the crystalline SDS density of 0.37 g

cm-3 or the measured SDS particle density of 1.13 g cm-3 did not describe properly the measured

Page 148: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

131

GF. The calculated GF values using the ZSR approach are not shown in Figure 5.10 but are

similar to what was observed for NaNO3/SDS particles.

To better represent the measured GF, these values must be corrected for DSF and SDS

density as a function of concentration. The calculated GF (solid line in Figure 5.10) uses

equations 5.7 and 5.8 to calculate volume fractions and correct for shape factors as a function of

SDS concentration. These calculations use the GF of 1.09 and 2.27 for pure SDS and NaC,

respectively. This approach resulted in a calculated GF that reproduced quantitatively the

measured growth factors. The SDS density function obtained from the NaNO3/SDS particle

measurements (equation 5.6) was used to calculate the NaC/SDS growth factors. However,

different functional forms for the SDS density are required to properly describe the two different

types of particles (NaC/SDS and NaNO3/SDS).

Page 149: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

132

Figure 5.10 DSF corrected fit and measured GF for the NaC/SDS particles. The fit uses the DSF and densities shown in Figure 5.9.

5.4 Conclusions

The interaction of water vapor with hygroscopic salt particles (NaC and NaNO3) coated

by a soluble surfactant (SDS) were studied. Sodium nitrate was used to represent spherical

particles while sodium chloride represented non-spherical particles. A combination of

independent measurements (growth factor and particle density) was used to characterize these

mixed particles. The GF factor measurements indicated that the surfactant coated particles

Page 150: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

133

gained water when exposed to high RH, which is in slight disagreement to the “inverted” micelle

model. The model predicted that the surfactant coated salt-water droplets will completely inhibit

growth by water uptake. The results point out that the amount of water gained is strongly

dependent on organic surfactant concentration and density in a complex manner. For very low

surfactant concentrations the growth factors are relatively high and comparable to those of pure

inorganic salt particles. When the concentration reaches the point at which a monolayer is

formed at the particle surface, the growth factors decrease significantly. The general trend is that

increasing organic surfactant concentration decreases the GF or water uptake. The changes can

be described quantitatively by coupling the particle density measurements into the GF analysis.

The “inverted” micelle CCN predictions could be incorrect due to the particles’ physical

structure. The NaC/SDS “dry” particle SEM images clearly show the surfactant fractions (for

WF greater than 0.090). The images demonstrate that the particles are deposited wet on the

substrate and there are possibly two phases; the inorganic and organic fractions. The surfactant

in a droplet is assumed to partition to the surface but when the particles are “dried”, the two

components separate into two phases making the characterization and the “inverted” micelle

model test more challenging. Based on these studies, the surfactant coated particles gain and

lose water but the amount changes with organic surfactant concentration.

A better understanding of how SDS fraction affects salt particles is needed to properly

describe these mixed particles. It was observed that the density of the surfactant changes as its

concentration increases. This in turn affects directly the mixed particles’ density assuming that

the NaNO3 density remains constant. The dva distributions used to calculate particle density

suggest that the NaNO3/SDS particles retain water. The distributions are broader as the SDS

fraction is increased suggesting the existence of several different SDS mesophases in the mixed

Page 151: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

134

particles. Heating the particles prior to dva distribution measurements showed significant shifts

in the distributions. This is evidence that particles retain water and that different types of

particles exist, especially at intermediate SDS concentrations. A changing SDS density as a

function of its concentration was derived from the particle density measurements. This result

was used to calculate growth factors by using the SDS fraction density to calculate volume

fraction and assuming that each component takes up water based on its growth factor as a pure

substance. The calculated GF quantitatively described the measured growth factors as opposed

to the simplified ZSR approach.

Sodium chloride coated with SDS surfactant represents particles for which the shape

factor must be included in the effective density and growth factor measurements and analysis.

The effective density for the NaC/SDS particles follows the same trend as for the NaNO3/SDS

particles. The density decreases from the pure NaC particle until it reaches its lowest value.

Further increase in concentration takes the density back up to the pure SDS particle density of

1.13 g cm-3. From the density measurements, functional forms for the dynamic shape factors and

SDS density were derived as a function of SDS concentration. These functions were used to

calculate GF as a function of SDS concentration. The calculated GF fit well the measured GF

and the S shaped curved.

The hygroscopic particles coated by surfactants appear to retain water even after drying.

But the amount of water retained is small compared to the water uptake by these particles at

approximately 80% RH. The simplified ZSR approach falls short to describe the measured

growth factors. Based on the density measurements it was concluded that the configuration of

surfactant molecules strongly depends on their concentrations. Therefore the SDS density

Page 152: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

135

functional forms are needed to understand the behavior of mixed inorganic/organic particles.

This factor is important for modeling atmospheric particles with high organic mass content.

Page 153: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

136

References for Chapter V

(1) Middlebrook, A. M.; Murphy, D. M.; Lee, S. H.; Thomson, D. S.; Prather, K. A.; Wenzel, R. J.; Liu, D. Y.; Phares, D. J.; Rhoads, K. P.; Wexler, A. S.; Johnston, M. V.; Jimenez, J. L.; Jayne, J. T.; Worsnop, D. R.; Yourshaw, I.; Seinfeld, J. H.; Flagan, R. C., A comparison of particle mass spectrometers during the 1999 Atlanta Supersite Project, Journal of Geophysical Research-Atmospheres 2003, 108.

(2) Middlebrook, A. M.; Murphy, D. M.; Thomson, D. S., Observations of organic material in individual marine particles at Cape Grim during the First Aerosol Characterization Experiment (ACE 1), Journal of Geophysical Research-Atmospheres 1998, 103, 16475-16483.

(3) Ellison, G. B.; Tuck, A. F.; Vaida, V., Atmospheric processing of organic aerosols, Journal of Geophysical Research-Atmospheres 1999, 104, 11633-11641.

(4) Gill, P. S.; Graedel, T. E.; Weschler, C. J., Organic Films on Atmospheric Aerosol-Particles, Fog Droplets, Cloud Droplets, Raindrops, and Snowflakes, Reviews of Geophysics 1983, 21, 903-920.

(5) Chen, Y. Y.; Lee, W. M. G., Hygroscopic properties of inorganic-salt aerosol with surface-active organic compounds, Chemosphere 1999, 38, 2431-2448.

(6) Chen, Y. Y.; Lee, W. M. G., The effect of surfactants on the deliquescence of sodium chloride, Journal of Environmental Science and Health Part a-Toxic/Hazardous Substances & Environmental Engineering 2001, 36, 229-242.

(7) Hansson, H. C.; Rood, M. J.; Koloutsou-Vakakis, S.; Hameri, K.; Orsini, D.; Wiedensohler, A., NaCl aerosol particle hygroscopicity dependence on mixing with organic compounds, Journal of Atmospheric Chemistry 1998, 31, 321-346.

(8) McNeill, V. F.; Patterson, J.; Wolfe, G. M.; Thornton, J. A., The effect of varying levels of surfactant on the reactive uptake of N2O5 to aqueous aerosol, Atmospheric Chemistry and Physics 2006, 6, 1635-1644.

(9) Ervens, B.; Feingold, G.; Kreidenweis, S. M., Influence of water-soluble organic carbon on cloud drop number concentration, Journal of Geophysical Research-Atmospheres 2005, 110.

(10) Abbatt, J. P. D.; Broekhuizen, K.; Kumal, P. P., Cloud condensation nucleus activity of internally mixed ammonium sulfate/organic acid aerosol particles, Atmospheric Environment 2005, 39, 4767-4778.

Page 154: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

137

(11) Donaldson, D. J.; Vaida, V., The influence of organic films at the air-aqueous boundary on atmospheric processes, Chemical Reviews 2006, 106, 1445-1461.

(12) Andrews, E.; Larson, S. M., Effect of Surfactant Layers on the Size Changes of Aerosol-Particles as a Function of Relative-Humidity, Environmental Science & Technology 1993, 27, 857-865.

(13) Garland, R. M.; Wise, M. E.; Beaver, M. R.; DeWitt, H. L.; Aiken, A. C.; Jimenez, J. L.; Tolbert, M. A., Impact of palmitic acid coating on the water uptake and loss of ammonium sulfate particles, Atmospheric Chemistry and Physics 2005, 5, 1951-1961.

(14) Zelenyuk, A.; Imre, D., Single particle laser ablation time-of-flight mass spectrometer: An introduction to SPLAT, Aerosol Science and Technology 2005, 39, 554-568.

(15) Zelenyuk, A.; Cai, Y.; Imre, D., From agglomerates of spheres to irregularly shaped particles: Determination of dynamic shape factors from measurements of mobility and vacuum aerodynamic diameters, Aerosol Science and Technology 2006, 40, 197-217.

(16) Cai, Y.; Zelenyuk, A.; Imre, D., A high resolution study of the effect of morphology on the mass spectra of single PSL particles with Na-containing layers and nodules, Aerosol Science and Technology 2006, 40, 1111-1122.

(17) Zelenyuk, A.; Imre, D.; Cuadra-Rodriguez, L. A., Evaporation of water from particles in the aerodynamic lens inlet: An experimental study, Analytical Chemistry 2006, 78, 6942-6947.

(18) DeCarlo, P. F.; Slowik, J. G.; Worsnop, D. R.; Davidovits, P.; Jimenez, J. L., Particle morphology and density characterization by combined mobility and aerodynamic diameter measurements. Part 1: Theory, Aerosol Science and Technology 2004, 38, 1185-1205.

(19) Hameri, K.; Laaksonen, A.; Vakeva, M.; Suni, T., Hygroscopic growth of ultrafine sodium chloride particles, Journal of Geophysical Research-Atmospheres 2001, 106, 20749-20757.

(20) Mikhailov, E.; Vlasenko, S.; Niessner, R.; Poschl, U., Interaction of aerosol particles composed of protein and salts with water vapor: hygroscopic growth and microstructural rearrangement, Atmospheric Chemistry and Physics 2004, 4, 323-350.

(21) Clegg, S. L.; Seinfeld, J. H., Improvement of the Zdanovskii-Stokes-Robinson model for mixtures containing solutes of different charge types, Journal of Physical Chemistry A 2004, 108, 1008-1017.

Page 155: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

138

(22) Clegg, S. L.; Seinfeld, J. H., Thermodynamic models of aqueous solutions containing inorganic electrolytes and dicarboxylic acids at 298.15 K. 2. Systems including dissociation equilibria, Journal of Physical Chemistry A 2006, 110, 5718-5734.

(23) Clegg, S. L.; Seinfeld, J. H.; Brimblecombe, P., Thermodynamic modelling of aqueous aerosols containing electrolytes and dissolved organic compounds, Journal of Aerosol Science 2001, 32, 713-738.

(24) Seinfeld, J. H.; Pandis, S. N. Atmospheric Chemistry and Physics - From Air Pollution to Climate Change (2nd Edition); John Wiley & Sons, 2006.

(25) Cruz, C. N.; Pandis, S. N., Deliquescence and hygroscopic growth of mixed inorganic-organic atmospheric aerosol, Environmental Science & Technology 2000, 34, 4313-4319.

(26) Prenni, A. J.; De Mott, P. J.; Kreidenweis, S. M., Water uptake of internally mixed particles containing ammonium sulfate and dicarboxylic acids, Atmospheric Environment 2003, 37, 4243-4251.

(27) Mazer, N. A.; Benedek, G. B.; Carey, M. C., Investigation of Micellar Phase of Sodium Dodecyl-Sulfate in Aqueous Sodium-Chloride Solutions Using Quasi-Elastic Light-Scattering Spectroscopy, Journal of Physical Chemistry 1976, 80, 1075-1085.

(28) Tabazadeh, A., Organic aggregate formation in aerosols and its impact on the physicochemical properties of atmospheric particles., Atmospheric Environment 2005, 39, 5472-5480.

(29) Myers, D. Surfaces, surfactant science and technology; 3rd ed.; Wiley: New York, 2006.

(30) Rharbi, Y.; Chen, L. S.; Winnik, M. A., Exchange mechanisms for sodium dodecyl sulfate micelles: High salt concentration, Journal of the American Chemical Society 2004, 126, 6025-6034.

(31) Hoffman, R. C.; Laskin, A.; Finlayson-Pitts, B. J., Sodium nitrate particles: physical and chemical properties during hydration and dehydration, and implications for aged sea salt aerosols, Journal of Aerosol Science 2004, 35, 869-887.

(32) Tang, I. N.; Munkelwitz, H. R., Water Activities, Densities, and Refractive-Indexes of Aqueous Sulfates and Sodium-Nitrate Droplets of Atmospheric Importance, Journal of Geophysical Research-Atmospheres 1994, 99, 18801-18808.

(33) Zelenyuk, A.; Cai, Y.; Chieffo, L.; Imre, D., High precision density measurements of single particles: The density of metastable phases, Aerosol Science and Technology 2005, 39, 972-986.

Page 156: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

________________

Bibliography

________________

Abbatt, J. P. D.; Broekhuizen, K.; Kumal, P. P., Cloud condensation nucleus activity of internally mixed ammonium sulfate/organic acid aerosol particles, Atmospheric Environment 2005, 39, 4767-4778.

Andrews, E.; Larson, S. M., Effect of Surfactant Layers on the Size Changes of Aerosol-Particles as a Function of Relative-Humidity, Environmental Science & Technology 1993, 27, 857-865.

Atkinson, R., Gas-Phase Tropospheric Chemistry of Organic-Compounds, Journal of Physical and Chemical Reference Data 1994, R1.

Atkinson, R.; Aschmann, S. M., OH Radical Production from the Gas-Phase Reductions of O3 with a Series of Alkenes Under Atmospheric Conditions, Environ. Sci. Technol. 1993, 27, 1357-1363.

Bell, N.; Heard, D. E.; Pilling, M. J.; Tomlin, A. S., Atmospheric lifetime as a probe of radical chemistry in the boundary layer, Atmospheric Environment 2003, 37, 2193-2205.

Berkowitz, J.; Ellison, G. B.; Gutman, D., Three Methods to Measure RH Bond-Energies, Journal of Physical Chemistry 1994, 98, 2744-2765.

Bertram, A. K.; Ivanov, A. V.; Hunter, M.; Molina, L. T.; Molina, M. J., The Reaction Probablity of OH on Organic Surfaces of Tropospheric Interest, J. Phys. Chem. A 2001, 105, 9415-9421.

Blanksby, S. J.; Ellison, G. B., Bond dissociation energies of organic molecules, Accounts of Chemical Research 2003, 36, 255-263.

Cabalo, J.; Zelenyuk, A.; Baer, T.; Miller, R. E., Two-color laser induced evaporation dynamics of liquid aerosols probed by time-of-flight mass spectrometry, Aerosol Science and Technology 2000, 33, 3-19.

Cai, Y.; Zelenyuk, A.; Imre, D., A high resolution study of the effect of morphology on the mass spectra of single PSL particles with Na-containing layers and nodules, Aerosol Science and Technology 2006, 40, 1111-1122.

Chen, Y. Y.; Lee, W. M. G., Hygroscopic properties of inorganic-salt aerosol with surface-active organic compounds, Chemosphere 1999, 38, 2431-2448.

Chen, Y. Y.; Lee, W. M. G., The effect of surfactants on the deliquescence of sodium chloride, Journal of Environmental Science and Health Part a-Toxic/Hazardous Substances & Environmental Engineering 2001, 36, 229-242.

Page 157: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

140

Clegg, S. L.; Brimblecombe, P.; Wexler, A. S., Thermodynamic model of the system H+-NH4+-

Na+-SO42--NO3--Cl--H2O at 298.15 K, Journal of Physical Chemistry A 1998, 102, 2155-

2171.

Clegg, S. L.; Seinfeld, J. H., Improvement of the Zdanovskii-Stokes-Robinson model for mixtures containing solutes of different charge types, Journal of Physical Chemistry A 2004, 108, 1008-1017.

Clegg, S. L.; Seinfeld, J. H., Thermodynamic models of aqueous solutions containing inorganic electrolytes and dicarboxylic acids at 298.15 K. 2. Systems including dissociation equilibria, Journal of Physical Chemistry A 2006, 110, 5718-5734.

Clegg, S. L.; Seinfeld, J. H.; Brimblecombe, P., Thermodynamic modelling of aqueous aerosols containing electrolytes and dissolved organic compounds, Journal of Aerosol Science 2001, 32, 713-738.

Covert, D. S.; Gras, J. L.; Wiedensohler, A.; Stratmann, F., Comparison of directly measured CCN with CCN modeled from the number-size distribution in the marine boundary layer during ACE 1 at Cape Grim, Tasmania, Journal of Geophysical Research-Atmospheres 1998, 103, 16597-16608.

Cruz, C. N.; Pandis, S. N., A study of the ability of pure secondary organic aerosol to act as cloud condensation nuclei, Atmospheric Environment 1997, 31, 2205-2214.

Cruz, C. N.; Pandis, S. N., Deliquescence and hygroscopic growth of mixed inorganic-organic atmospheric aerosol, Environmental Science & Technology 2000, 34, 4313-4319.

Cziczo, D. J.; DeMott, P. J.; Brock, C.; Hudson, P. K.; Jesse, B.; Kreidenweis, S. M.; Prenni, A. J.; Schreiner, J.; Thomson, D. S.; Murphy, D. M., A method for single particle mass spectrometry of ice nuclei, Aerosol Science and Technology 2003, 37, 460-470.

Dahneke, B.; Padliya, D. In Progress in Astronautics and Aeronautics; Potter, J., Ed.; American Institute of Aeronautics and Astronautics: New York, 1977; Vol. 51.

D'Andrea, T. M.; Zhang, X.; Jochnowitz, E. B.; Lindeman, T. G.; Simpson, C. J. S. M.; David, D. E.; Curtiss, T. J.; Morris, J. R.; Ellison, G. B., Oxidation of organic films by beams of hydroxyl radicals, Journal of Physical Chemistry B 2008, 112, 535-544.

de Gouw, J. A.; Krishnamurthy, M.; Bierbaum, V. M.; Leone, S. R., Measured and calculated mobilities of cluster ions drifting in helium and in nitrogen, International Journal of Mass Spectrometry 1997, 167, 281-289.

DeCarlo, P. F.; Slowik, J. G.; Worsnop, D. R.; Davidovits, P.; Jimenez, J. L., Particle morphology and density characterization by combined mobility and aerodynamic diameter measurements. Part 1: Theory, Aerosol Science and Technology 2004, 38, 1185-1205.

Page 158: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

141

Decesari, S.; Facchini, M. C.; Mircea, M.; Cavalli, F.; Fuzzi, S., Solubility properties of surfactants in atmospheric aerosol and cloud/fog water samples, Journal of Geophysical Research-Atmospheres 2003, 108, 6-(1-9).

Dessiaterik, Y.; Baer, T.; Miller, R. E., Laser ablation of imidazolium based ionic liquids, Journal of Physical Chemistry A 2006, 110, 1500-1505.

Dessiaterik, Y.; Nguyen, T.; Baer, T.; Miller, R. E., IR vaporization mass spectrometry of aerosol particles with ionic solutions: The problem of ion-ion recombination, Journal of Physical Chemistry A 2003, 107, 11245-11252.

Donahue, N. M.; Anderson, J. G.; Demerjian, K. L., New rate constants for ten OH alkane reactions from 300 to 400 K: An assessment of accuracy, J. Phys. Chem. A 1998, 102, 3121-3126.

Donaldson, D. J.; Vaida, V., The influence of organic films at the air-aqueous boundary on atmospheric processes, Chemical Reviews 2006, 106, 1445-1461.

Dupeyrat, G.; Rowe, B. R.; Fahey, D. W.; Albritton, D. L., Diagnostic studies of venturi inlets for flow reactors, International Journal of Mass Spectrometry and Ion Physics 1982, 44, 1-18.

Ellison, G. B.; Tuck, A. F.; Vaida, V., Atmospheric processing of organic aerosols, Journal of Geophysical Research-Atmospheres 1999, 104, 11633-11641.

Ervens, B.; Feingold, G.; Kreidenweis, S. M., Influence of water-soluble organic carbon on cloud drop number concentration, Journal of Geophysical Research-Atmospheres 2005, 110.

Finlayson-Pitts, B. J.; Pitts Jr., J. N. Chemistry of the Upper and Lower Atmosphere; Academic Press: San Diego, 1999.

Flagan, R. C. In Aerosol Measurement: Principles, Techniques, and Applications; 2nd ed.; Willeke, K., Baron, P. A., Eds.; John Wiley & Sons, Inc.: New York City, 2001.

Gard, E.; Mayer, J. E.; Morrical, B. D.; Dienes, T.; Fergenson, D. P.; Prather, K. A., Real-time analysis of individual atmospheric aerosol particles: Design and performance of a portable ATOFMS, Analytical Chemistry 1997, 69, 4083-4091.

Garland, R. M.; Wise, M. E.; Beaver, M. R.; DeWitt, H. L.; Aiken, A. C.; Jimenez, J. L.; Tolbert, M. A., Impact of palmitic acid coating on the water uptake and loss of ammonium sulfate particles, Atmospheric Chemistry and Physics 2005, 5, 1951-1961.

Gebhart, J. In Aerosol Measurement: Principles, Techniques, and Applications; 2nd ed.; Willeke, K., Baron, P. A., Eds.; John Wiley & Sons, Inc.: New York City, 2001.

Gill, P. S.; Graedel, T. E.; Weschler, C. J., Organic Films on Atmospheric Aerosol-Particles, Fog Droplets, Cloud Droplets, Raindrops, and Snowflakes, Reviews of Geophysics 1983, 21, 903-920.

Page 159: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

142

Guyer, D. Optical Parametric Oscillator User Manual: Bellevue, 2006.

Hameri, K.; Laaksonen, A.; Vakeva, M.; Suni, T., Hygroscopic growth of ultrafine sodium chloride particles, Journal of Geophysical Research-Atmospheres 2001, 106, 20749-20757.

Hansson, H. C.; Rood, M. J.; Koloutsou-Vakakis, S.; Hameri, K.; Orsini, D.; Wiedensohler, A., NaCl aerosol particle hygroscopicity dependence on mixing with organic compounds, Journal of Atmospheric Chemistry 1998, 31, 321-346.

Harper, K. L.; Allen, H. C., Competition between DPPC and SDS at the air-aqueous interface, Langmuir 2007, 23, 8925-8931.

Hearn, J. D.; Renbaum, L. H.; Wang, X.; Smith, G. D., Kinetics and products from reaction of Cl radicals with dioctyl sebacate (DOS) particles in O2: a model for radical-initiated oxidation of organic aerosols, Physical Chemistry Chemical Physics 2007, 9, 4803-4813.

Hearn, J. D.; Smith, G. A., Ozonolysis of mixed oleic acid/n-docosane particles: The roles of phase, morphology, and metastable states, Journal of Physical Chemistry A 2007, 111, 11059-11065.

Hoffman, R. C.; Laskin, A.; Finlayson-Pitts, B. J., Sodium nitrate particles: physical and chemical properties during hydration and dehydration, and implications for aged sea salt aerosols, Journal of Aerosol Science 2004, 35, 869-887.

Huffman, J. A.; Jayne, J. T.; Drewnick, F.; Aiken, A. C.; Onasch, T.; Worsnop, D. R.; Jimenez, J. L., Design, modeling, optimization, and experimental tests of a particle beam width probe for the Aerodyne aerosol mass spectrometer, Aerosol Science and Technology 2005, 39, 1143-1163.

Jayne, J. T.; Leard, D. C.; Zhang, X. F.; Davidovits, P.; Smith, K. A.; Kolb, C. E.; Worsnop, D. R., Development of an aerosol mass spectrometer for size and composition analysis of submicron particles, Aerosol Science and Technology 2000, 33, 49-70.

Jimenez, J. L.; Jayne, J. T.; Shi, Q.; Kolb, C. E.; Worsnop, D. R.; Yourshaw, I.; Seinfeld, J. H.; Flagan, R. C.; Zhang, X.; Smith, K. A.; Morris, J. W.; Davidovits, P., Ambient aerosol sampling using the Aerodyne Aerosol Mass Spectrometer, J. Geophys. Res. 2003, 108, 8425.

Johnston, M. V.; Wexler, A. S., On the Cover- MS of Individual Aerosol-Particles, Analytical Chemistry 1995, 67, A721-A726.

King, G. K.; Maricq, M. M.; Bierbaum, V. M.; DePuy, C. H., Gas-Phase Reactions of Negative-Ions with Alkyl Nitrites, Journal of the American Chemical Society 1981, 103, 7133-7140.

Kleinekofort, W.; Avdiev, J.; Brutschy, B., A new method of laser desorption mass spectrometry for the study of biological macromolecules, Int. J. Mass Spectrom. Ion Process. 1996, 152, 135-142.

Page 160: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

143

Kleinekofort, W.; Pfenninger, A.; Plomer, T.; Griesinger, C.; Brutschy, B., Observation of noncovalent complexes using laser-induced liquid beam ionization desorption, Int. J. Mass Spectrom. Ion Process. 1996, 156, 195-202.

Lee, S. H.; Murphy, D. M.; Thomson, D. S.; Middlebrook, A. M., Chemical components of single particles measured with Particle Analysis by Laser Mass Spectrometry (PALMS) during the Atlanta SuperSite Project: Focus on organic/sulfate, lead, soot, and mineral particles, Journal of Geophysical Research-Atmospheres 2002, 107, 1-(1-13).

Li, Z. D.; Williams, A. L.; Rood, M. J., Influence of soluble surfactant properties on the activation of aerosol particles containing inorganic solute, Journal of the Atmospheric Sciences 1998, 55, 1859-1866.

Lide, D. R. CRC Handbook of Chemistry and Physics; CRC Press: Boca Raton, FL, 1996.

Liu, P.; Ziemann, P. J.; Kittelson, D. B.; McMurry, P. H., Generating particle beams of controlled dimensions and divergence .1. Theory of particle motion in aerodynamic lenses and nozzle expansions., Aerosol Science and Technology 1995, 22, 293-313.

Liu, P.; Ziemann, P. J.; Kittelson, D. B.; McMurry, P. H., Generating particle beams of controlled dimensions and divergence .2. Experimental evaluation of particle motion in aerodynamic lenses and nozzle expansions., Aerosol Science and Technology 1995, 22, 314-324.

Mallina, R. V.; Wexler, A. S.; Johnston, M. V., Particle growth in high-speed particle beam inlets, Journal of Aerosol Science 1997, 28, 223-238.

Mallina, R. V.; Wexler, A. S.; Johnston, M. V., High-speed particle beam generation: Simple focusing mechanisms, Journal of Aerosol Science 1999, 30, 719-738.

Mazer, N. A.; Benedek, G. B.; Carey, M. C., Investigation of Micellar Phase of Sodium Dodecyl-Sulfate in Aqueous Sodium-Chloride Solutions Using Quasi-Elastic Light-Scattering Spectroscopy, Journal of Physical Chemistry 1976, 80, 1075-1085.

McGraw, R.; Private communication ed.; Zelenyuk, A., Imre, D., Eds.: Richland, WA, 2002.

Mckeown, P. J.; Johnston, M. V.; Murphy, D. M., Online Single-Particle Analysis by Laser Desorption Mass-Spectrometry, Analytical Chemistry 1991, 63, 2069-2073.

McNeill, V. F.; Patterson, J.; Wolfe, G. M.; Thornton, J. A., The effect of varying levels of surfactant on the reactive uptake of N2O5 to aqueous aerosol, Atmospheric Chemistry and Physics 2006, 6, 1635-1644.

McNeill, V. F.; Wolfe, G. M.; Thornton, J. A., The oxidation of oleate in submicron aqueous salt aerosols: Evidence of a surface process, Journal of Physical Chemistry A 2007, 111, 1073-1083.

Page 161: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

144

Middlebrook, A. M.; Murphy, D. M.; Lee, S. H.; Thomson, D. S.; Prather, K. A.; Wenzel, R. J.; Liu, D. Y.; Phares, D. J.; Rhoads, K. P.; Wexler, A. S.; Johnston, M. V.; Jimenez, J. L.; Jayne, J. T.; Worsnop, D. R.; Yourshaw, I.; Seinfeld, J. H.; Flagan, R. C., A comparison of particle mass spectrometers during the 1999 Atlanta Supersite Project, Journal of Geophysical Research-Atmospheres 2003, 108.

Middlebrook, A. M.; Murphy, D. M.; Thomson, D. S., Observations of organic material in individual marine particles at Cape Grim during the First Aerosol Characterization Experiment (ACE 1), Journal of Geophysical Research-Atmospheres 1998, 103, 16475-16483.

Mikhailov, E.; Vlasenko, S.; Niessner, R.; Poschl, U., Interaction of aerosol particles composed of protein and salts with water vapor: hygroscopic growth and microstructural rearrangement, Atmospheric Chemistry and Physics 2004, 4, 323-350.

Morgner, N.; Barth, H. D.; Brutschy, B., A new way to detect noncovalently bonded complexes of biomolecules from liquid micro-droplets by laser mass spectrometry, Aust. J. Chem. 2006, 59, 109-114.

Murphy, D. M.; Personal communication ed.; Ellison, G. B., Cuadra-Rodriguez, L. A., Eds.: Boulder, CO, 2006.

Murphy, D. M., The design of single particle laser mass spectrometers, Mass Spectrometry Reviews 2007, 26, 150-165.

Murphy, D. M.; Anderson, J. R.; Quinn, P. K.; McInnes, L. M.; Brechtel, F. J.; Kreidenweis, S. M.; Middlebrook, A. M.; Posfai, M.; Thomson, D. S.; Buseck, P. R., Influence of sea-salt on aerosol radiative properties in the Southern Ocean marine boundary layer, Nature 1998, 392, 62-65.

Murphy, D. M.; Thomson, D. S., Laser Ionization Mass-Spectroscopy of Single Aerosol-Particles, Aerosol Science and Technology 1995, 22, 237-249.

Murphy, D. M.; Thomson, D. S.; Middlebrook, A. M.; Schein, M. E., In situ single-particle characterization at Cape Grim, Journal of Geophysical Research-Atmospheres 1998, 103, 16485-16491.

Myers, D. Surfaces, surfactant science and technology; 3rd ed.; Wiley: New York, 2006.

Mysak, E. R.; Dessiaterik, Y. N.; McKinney, C. J.; Miller, R. E.; Baer, T., Design of a timing circuit for random laser triggering on aerosol particles, Review of Scientific Instruments 2006, 77, 013301-(1-6).

Nash, D. G.; Baer, T.; Johnston, M. V., Aerosol mass spectrometry: An introductory review, International Journal of Mass Spectrometry 2006, 258, 2-12.

Neubauer, K. R.; Johnston, M. V.; Wexler, A. S., On-line analysis of aqueous aerosols by laser desorption ionization, Int. J. Mass Spectrom. Ion Process. 1997, 163, 29-37.

Page 162: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

145

Neubauer, K. R.; Johnston, M. V.; Wexler, A. S., Humidity effects on the mass spectra of single aerosol particles, Atmospheric Environment 1998, 32, 2521-2529.

Noble, C. A.; Nordmeyer, T.; Salt, K.; Morrical, B.; Prather, K. A., Aerosol Characterization Using Mass-Spectrometry, Trac-Trends in Analytical Chemistry 1994, 13, 218-222.

Noble, C. A.; Prather, K. A., Real-time single particle mass spectrometry: A historical review of a quarter century of the chemical analysis of aerosols, Mass Spectrometry Reviews 2000, 19, 248-274.

Nordmeyer, T.; Prather, K. A., Real-Time Measurement Capabilities Using Aerosol Time-of-Flight Mass-Spectrometry, Analytical Chemistry 1994, 66, 3540-3542.

O'Hanlon, J. F. A User's Guide to Vacuum Technology; Third ed.; Wiley-Interscience, 2003.

Penner, J. E.; Hegg, D.; Leaitch, R., Unraveling the role of aerosols in climate change, Environmental Science & Technology 2001, 35, 332a-340a.

Poschl, U., Atmospheric aerosols: Composition, transformation, climate and health effects, Angewandte Chemie-International Edition 2005, 44, 7520-7540.

Prenni, A. J.; De Mott, P. J.; Kreidenweis, S. M., Water uptake of internally mixed particles containing ammonium sulfate and dicarboxylic acids, Atmospheric Environment 2003, 37, 4243-4251.

Pruppacher, H. R.; Klett, J. D. Microphysics of Clouds and Precipitation; 2nd ed.; Kluwer Academic Publishing: Dordrecht: Holland, 1997.

Rader, D. J.; J., O. H. T. In Aerosol Measurement: Principles, Techniques, and Applications; 2nd ed.; Willeke, K., Baron, P. A., Eds.; John Wiley & Sons, Inc.: New York City, 2001.

Rharbi, Y.; Chen, L. S.; Winnik, M. A., Exchange mechanisms for sodium dodecyl sulfate micelles: High salt concentration, Journal of the American Chemical Society 2004, 126, 6025-6034.

Rienstra-Kiracofe, J. C.; Tschumper, G. S.; Schaefer, H. F.; Nandi, S.; Ellison, G. B., Atomic and molecular electron affinities: Photoelectron experiments and theoretical computations, Chemical Reviews 2002, 102, 231-282.

Robertson, W. H.; Johnson, M. A., Molecular aspects of halide ion hydration: The cluster approach, Annual Review of Physical Chemistry 2003, 54, 173-213.

Ruscic, B.; Feller, D.; Dixon, D. A.; Peterson, K. A.; Harding, L. B.; Asher, R. L.; Wagner, A. F., Evidence for a lower enthalpy of formation of hydroxyl radical and a lower gas-phase bond dissociation energy of water, Journal of Physical Chemistry A 2001, 105, 1-4.

Schreiner, J.; Schild, U.; Voigt, C.; Mauersberger, K., Focusing of aerosols into a particle beam at pressures from 10 to 150 Torr, Aerosol Science and Technology 1999, 31, 373-382.

Page 163: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

146

Seinfeld, J. H.; Pandis, S. N. Atmospheric Chemistry and Physics - From Air Pollution to Climate Change (2nd Edition); John Wiley & Sons, 2006.

Seinfeld, J. H.; Pandis, S. N. In Atmospheric Chemistry and Physics - From Air Pollution to Climate Change (2nd Edition); John Wiley & Sons, 2006.

Shimanouchi, T. Tables of Molecular Vibrational Frequencies Consolidated Volume I; National Bureau of Standards, 1972.

Silberberg, M. S. Principles of General Chemistry; First ed.; McGraw-Hill Science/Engineering/Math: New York, 2006.

Sobott, F.; Wattenberg, A.; Barth, H. D.; Brutschy, B., Ionic clathrates from aqueous solutions detected with laser induced liquid beam ionization/desorption mass spectrometry, International Journal of Mass Spectrometry 1999, 187, 271-279.

Su, Y.; Sipin, M. F.; Furutani, H.; Prather, K. A., Development and Characterization of an Aerosol Time-of-Flight Mass Spectrometer with Increased Detection Efficiency, Analytical Chemistry 2003, 76, 712-719.

Svenningsson, B.; Rissler, J.; Swietlicki, E.; Mircea, M.; Bilde, M.; Facchini, M. C.; Decesari, S.; Fuzzi, S.; Zhou, J.; Monster, J.; Rosenorn, T., Hygroscopic growth and critical supersaturations for mixed aerosol particles of inorganic and organic compounds of atmospheric relevance, Atmospheric Chemistry and Physics 2006, 6, 1937-1952.

Sykes, D. C.; Woods, E.; Smith, G. D.; Baer, T.; Miller, R. E., Thermal vaporization-vacuum ultraviolet laser ionization time-of-flight mass spectrometry of single aerosol particles, Analytical Chemistry 2002, 74, 2048-2052.

Tabazadeh, A., Organic aggregate formation in aerosols and its impact on the physicochemical properties of atmospheric particles., Atmospheric Environment 2005, 39, 5472-5480.

Tang, I. N., Thermodynamic and optical properties of mixed-salt aerosols of atmospheric importance, Journal of Geophysical Research-Atmospheres 1997, 102, 1883-1893.

Tang, I. N.; Munkelwitz, H. R., Water Activities, Densities, and Refractive-Indexes of Aqueous Sulfates and Sodium-Nitrate Droplets of Atmospheric Importance, Journal of Geophysical Research-Atmospheres 1994, 99, 18801-18808.

Tang, I. N.; Munkelwitz, H. R., Aerosol Phase-Transformation and Growth in the Atmosphere, Journal of Applied Meteorology 1994, 33, 791-796.

Tang, I. N.; Tridico, A. C.; Fung, K. H., Thermodynamic and optical properties of sea salt aerosols, Journal of Geophysical Research-Atmospheres 1997, 102, 23269-23275.

Thomson, D. S.; Middlebrook, A. M.; Murphy, D. M., Thresholds for laser-induced ion formation from aerosols in a vacuum using ultraviolet and vacuum-ultraviolet laser wavelengths, Aerosol Science and Technology 1997, 26, 544-559.

Page 164: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

147

Thomson, D. S.; Murphy, D. M., Analyzing Single Aerosol-Particles in Real-Time - Composition Analysis of Aerosols Is Critical to an Understanding of Their Origins and Fate - This Is Done by Combining Time-of-Flight Mass-Spectroscopy with Optical-Particle Detections and Sizing, Chemtech 1994, 24, 30-35.

Thomson, D. S.; Schein, M. E.; Murphy, D. M., Particle analysis by laser mass spectrometry WB-57F instrument overview, Aerosol Science and Technology 2000, 33, 153-169.

Toon, O. B.; Tolbert, M. A.; Koehler, B. G.; Middlebrook, A. M.; Jordan, J., Infrared optical constants of H2O ice, amorphous nitric acid solutions, and nitric acid hydrates, J. Geophys. Res. 1994, 99, 25631-25654.

TSI, I. Model 3450 Vibrating Orifice Aerosol Generator Instruction Manual: Shoreview, MN, 2002; Vol. Revision L.

TSI, I. Series 3080 Electrostatic Classifier Manual: Shoreview, MN, 2006; Vol. Revision G.

Vaida, V.; Tuck, A. F.; Ellison, G. B., Optical and chemical properties of atmospheric organic aerosols, Physics and Chemistry of the Earth Part C-Solar-Terrestrial and Planetary Science 2000, 25, 195-198.

Van Doren, J. M.; Barlow, S. E.; DePuy, C. H.; Bierbaum, V. M., Tandem Flowing Afterglow-Selected Ion Flow Tube and Its Application to the Thermal-Energy Reactions of 18O-, Journal of the American Chemical Society 1987, 109, 4412-4414.

Warren, S. G., Optical-Constants of Ice from the Ultraviolet to the Microwave, Applied Optics 1984, 23, 1206-1225.

Wattenberg, A.; Sobott, F.; Brutschy, B., Detection of intact hemoglobin from aqueous solution with laser desorption mass spectrometry, Rapid Commun. Mass Spectrom. 2000, 14, 859-861.

Woods, E.; Miller, R. E.; Baer, T., The internal energy of neutral ethylene glycol molecules created in the laser vaporization of aerosol particles, Journal of Physical Chemistry A 2003, 107, 2119-2125.

Woods, E.; Smith, G. D.; Dessiaterik, Y.; Baer, T.; Miller, R. E., Quantitative detection of aromatic compounds in single aerosol particle mass spectrometry, Analytical Chemistry 2001, 73, 2317-2322.

Zelenyuk, A.; Cabalo, J.; Baer, T.; Miller, R. E., Mass spectrometry of liquid aniline aerosol particles by IR/UV laser irradiation, Analytical Chemistry 1999, 71, 1802-1808.

Zelenyuk, A.; Cai, Y.; Chieffo, L.; Imre, D., High precision density measurements of single particles: The density of metastable phases, Aerosol Science and Technology 2005, 39, 972-986.

Page 165: Design and Development of a Droplet Mass Spectrometer · Design and Development of a Droplet Mass Spectrometer Thesis directed by Dr. G. Barney Ellison A droplet mass spectrometer

148

Zelenyuk, A.; Cai, Y.; Imre, D., From agglomerates of spheres to irregularly shaped particles: Determination of dynamic shape factors from measurements of mobility and vacuum aerodynamic diameters, Aerosol Science and Technology 2006, 40, 197-217.

Zelenyuk, A.; Imre, D., Single particle laser ablation time-of-flight mass spectrometer: An introduction to SPLAT, Aerosol Science and Technology 2005, 39, 554-568.

Zelenyuk, A.; Imre, D.; Cuadra-Rodriguez, L. A., Evaporation of water from particles in the aerodynamic lens inlet: An experimental study, Analytical Chemistry 2006, 78, 6942-6947.

Zelenyuk, A.; Imre, D.; Cuadra-Rodriguez, L. A.; Ellison, B., Measurements and interpretation of the effect of a soluble organic surfactant on the density, shape and water uptake of hygroscopic particles, Journal of Aerosol Science 2007, 38, 903-923.