dissolution media preparator dissolution media preparator Ÿ enables filtration, pre-heating,...

2
Dissolution Media Preparator Ÿ Enables filtration, pre-heating, deaeration, and precise dosing of media in the vessel within the dissolution bath Ÿ 21 ltr media tank capacity allows filling of dissolution vessels for up to 3 dissolution tests (n = 6) Ÿ Batch degassing ensures inter-vessel degassing consistency Ÿ Volumetric dispensing with an accuracy better than + 1% as per USP, FDA and Ph. Eur. requirements Ÿ Special nozzle design allows the handling of surfactant and frothing media like Sodium Lauryl Sulphate (SLS) Ÿ Programmable deaeration time, temperature and vacuum for effective deaeration of different media Ÿ Maintains media under vacuum at desired temperature while in idle condition Ÿ Virtually zero dead volume minimizes wastage of media Ÿ Programmable dispensing volumes from 200 ml to 4 ltrs Ÿ Fast filling of dissolution vessels at a dispensing rate of 900 ml/min Ÿ Programmable rinse and cleaning cycle allows the instrument to run through a prevalidated self cleaning procedure Ÿ On-board memory stores 10 batch information of degassed media QbD Quality by Design Ÿ Unparalleled versatility due to 1 ml volume change increments o Ÿ 360 turning radius allows it to be maneuvered in tight places Ÿ Multifunction single hose for dispensing, aspirating, cleaning, rinsing, and draining ensures no lifting of heavy media tanks Versatility Ÿ Ergonomically designed nozzle with embedded dispense switch for ease of operation Ÿ User friendly validation and calibration of media volumes User friendly Ÿ Print out of validation, calibration, dispensed volume and temperature reports Reports 360 turning radius ° CIP Clean-In-Place No lifting of media tank Volume change increment 1 ml Media volume 21 Ltr. Frothy media handled Dead volume 0 Virtually Programmable vacuum EMP-21

Upload: others

Post on 31-Jan-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

  • Dissolution Media PreparatorŸ Enables filtration, pre-heating, deaeration,

    and precise dosing of media in the vessel within the dissolution bath

    Ÿ 21 ltr media tank capacity allows filling of dissolution vessels for up to 3 dissolution tests (n = 6)

    Ÿ Batch degassing ensures inter-vessel degassing consistency

    Ÿ Volumetric dispensing with an accuracy better than + 1% as per USP, FDA and Ph. Eur. requirements

    Ÿ Special nozzle design allows the handling of surfactant and frothing media like Sodium Lauryl Sulphate (SLS)

    Ÿ Programmable deaeration time, temperature and vacuum for effective deaeration of different media

    Ÿ Maintains media under vacuum at desired temperature while in idle condition

    Ÿ Virtually zero dead volume minimizes wastage of media

    Ÿ Programmable dispensing volumes from 200 ml to 4 ltrs

    Ÿ Fast filling of dissolution vessels at a dispensing rate of 900 ml/min

    Ÿ Programmable rinse and cleaning cycle allows the instrument to run through a prevalidated self cleaning procedure

    Ÿ On-board memory stores 10 batch information of degassed media

    QbDQuality by Design

    Ÿ Unparalleled versatility due to 1 ml volume change increments

    oŸ 360 turning radius allows it to be

    maneuvered in tight placesŸ Multifunction single hose for dispensing,

    aspirating, cleaning, rinsing, and draining ensures no lifting of heavy media tanks

    Versatility

    Ÿ Ergonomically designed nozzle with embedded dispense switch for ease of operation

    Ÿ User friendly validation and calibration of media volumes

    User friendly

    Ÿ Print out of validation, calibration, dispensed volume and temperature reports

    Reports

    360 turningradius°

    CIPClean-In-Place No lifting of

    media tankVolume change

    increment

    1ml

    Media volume

    21 Ltr.

    Frothy mediahandled

    Dead volume

    0Virtually

    Programmablevacuum

    EMP-21

  • The information contained in this document is believed to be correct, But E LECTROLAB accepts no liability for any errors and reserves the right to alter specifications without notice

    • Complete range of Dissolution Testers • Disintegration Testers • Friability Testers• Tablet Hardness Testers • Tap Density Testers • Electromagnetic Sieve Shaker • Powder Flow Tester

    Specifications

    Our Products

    Certified ISO 9001 : 2008

    < 6 ppm is recommended dissolved oxygen level for USP Prednisone Tablets RS (Ref : USP Dissolution Toolkit version 2.0)

    USP 35 states “Dissolved gases can cause bubbles to form, which may change the results of the test.”

    Importance of Degassing

    2 to 21 litres

    ± 1%

    900 ml/min

    0.2 to 4 litres

    ± 1%

    900 ml/min

    o o 20 C to 50 Co0.5 C

    Accuracy

    Aspiration :

    Media volume

    Flow rate (during run cycle)

    Dispense :

    Dispense volume

    Temperature :

    Temperature range

    Accuracy

    Flow rate (during run cycle)

    Temperature accuracy

    100 kg

    Stainless Steel

    100 to 500 mm/Hg

    High performance piston pump

    230 VAC/50 Hz 10%110 VAC/50 Hz ± 10%

    ±

    55 x 92 cm cm

    ± 1%Piston pump accuracy

    Degassing level (measured by DO meter)2

    Media tank

    Vacuum

    Piston pump

    Electrical power

    Dimensions (Ø x H)

    Weight

    Model EMP-21

    Other specifications :

    Inter & Intra Laboratory

    VariabilityReliability

    Ÿ 52.3% of mean percent dissolved Ÿ 11.8% of standard deviation

    Effective deaeration reduces variability in results by:

    Ref : Eaton J et al. Perturbation Study of Dissolution Apparatus Variables -A Design of Experiment Approach. Dissolution Technologies;14(2):20-26.

    * Dependent on media type and conditions

    CE

    Optional Accessory : Optical DO meter2

    < 5 ppm of *DO2

    ELECTROLAB

    &Service

    Worldwide

    SalesHead Office:

    401, Tirupati Udyog, I. B. Patel road, Off. Western Express Highway, Goregaon (East), Mumbai- 400063, India.

    Tel: +91-22-40413131 Fax: +91-22-40413199, Website : www.electrolabindia.com, Email : [email protected]

    Page 1Page 2