e very l ife h as a p urpose… ultima ® xl/xt gas monitors

10
EVERY LIFE HAS A PURPOSE… Ultima ® XL/XT Gas Monitors

Upload: phillip-walsh

Post on 28-Dec-2015

217 views

Category:

Documents


3 download

TRANSCRIPT

EVERY LIFE HAS A PURPOSE…

Ultima® XL/XT Gas Monitors

Ultima XL / XT Gas Monitors

Presentation Overview HART:

General Information Network Topology

Ultima XL/XT Features Calibration Options Approvals Sales Support

Ultima XL / XT Gas Monitors

General Information HART (Highway Addressable

Remote Transducer) MSA device is a slave in a polled

network HART is a software protocol

running over 4-20mA Hardware Point to point 1200 baud FSK

Ultima XL / XT Topology

Ultima XL / XT Features

Non-display versions of XE & XA Single sensor gas monitor Enclosure options

Stainless steel - XL Polycarbonate G.P. - XT

Condulet has openings for: Sensor entry Intrinsically Safe HART Port (optional) Signal and Power Wiring

Internal Push-button & LEDs for Calibration

Available in Remote Condulet Versions

Ultima XL / XT Features

Main board configuration

Similar to the Ultima X main board

Single removable design with standoffs

No tools required for insertion into the condulet

Ultima XL / XT Calibration Options

Calibration utilizing: Internal push-button

and LEDs (intrusive) Portable handheld

communicator, host DCS or laptop

Disconnect under power (in-shop calibration)

Sensor X-change program

Ultima XL / XT Global Approvals

FM APPROVAL US/Canada

Class I Div 1&2 Groups A,B,C,D Class I Zone 1&2 Groups IIC, IIB,

IIA, T6 Type 4X, IP 66

Europe ATEX II 2G 3G EEx d nA [ib] IIC

T6 IP 66

International IEC Ex d nA [ib] IIC T6 IP 66

Ultima XL / XT Sales Support Customer requirements

HART handheld controller Must be compatible with the Emerson 375 Emerson 275 will NOT work HART Rev 6 or Rev 7

Latest DD’s downloaded to handheld controller A Device Description (DD) is an electronic data file

prepared in accordance with Device Description Language specifications that describes specific features and functions of a device including details of menus and graphic display features to be used by host applications (including handheld devices) to access all parameters and data in the corresponding device.

Download from Emerson website (or other handheld controller that subscribes to the HART foundation updates)

Through HART membership www.hartcomm2.org Available from MSA

Ultima XL / XT Menu Tree

Calibration

Alarm set points

Detailed device information

4-20 output calibration