errors in sampling/monitoring in an internal corrosion program

22
Lynsay Bensman 06 October 2009 Errors in Sampling/Monitoring in an Internal Corrosion Program NACE Eastern Area Conference

Upload: yvette-salas

Post on 01-Jan-2016

56 views

Category:

Documents


2 download

DESCRIPTION

Errors in Sampling/Monitoring in an Internal Corrosion Program. NACE Eastern Area Conference. Goals of Sampling/Monitoring. Identify potentially corrosive species that may be present in a pipeline system Determine corrosion rate Determine effectiveness of mitigation. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Errors in Sampling/Monitoring in an Internal Corrosion Program

Lynsay Bensman 06 October 2009

Errors in Sampling/Monitoring in an Internal Corrosion Program

NACE Eastern Area Conference

Page 2: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

2

Goals of Sampling/Monitoring

Identify potentially corrosive species that may be present in a pipeline system

Determine corrosion rate

Determine effectiveness of mitigation

Page 3: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

3

Errors in Sampling/Monitoring

Improper tests or equipment

Choosing the incorrect location

Poor timing

Improper sampling techniques

Improper sample preservation

Misinterpreting results

Missing sampling opportunities

SAFETY

Page 4: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

4

Improper Tests or Equipment

Using the wrong equipment for what you are actually trying to test- Monitoring devices

Using a kit that is not the proper range

Using a kit that is expired or not properly working- Check prior to going out in the field

Page 5: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

5

Monitoring Devices

What are you trying to achieve?

Do you need real time monitoring?

Are you looking for pitting?

Is erosion playing a role in the corrosion process?

Page 6: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

6

Incorrect Sampling/Monitoring Locations

Representative or more severe than the corrosion in the system

Not necessarily- The easiest place- The closest place

Dilution effect

How many locations is enough?

Monitoring in conjunction with mitigation

Page 7: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

7

Monitoring Locations

Coupons/probes on stick off the top of the pipeline are not exposed to liquids laying alone the bottom of the line

Coupons/probes off the bottom are not in the gas flow but allow liquids to collect if present

Coupons in flow should be as close to the bottom of the pipe as possible

Page 8: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

8

Poor Timing

Liquid sample collection is not likely to yield optimum results- Right after drip blowing or removal from separators

At intervals that makes it difficult to interpret results- Consideration for injection withdrawal cycles- Consideration for chemical treatment

Page 9: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

9

Improper sampling techniques

Waiting to perform field testing

Performing analyses on samples that do not contain water

Not following instructions

Air

Sample contamination

Page 10: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

10

Waiting to Perform Field Testing

Reasons- Weather- Time- Training

Ramifications- Some data is unreliable if not collected immediately

- Dissolved gases- pH- Bacteria culture testing

Page 11: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

11

Air

Sample containers should be as full as possible to prevent reaction with air

Solid samples can react with oxygen in air leading to misidentification

4FeS + 3O2 2Fe2O3 + 4S + heat

Coupons that have been exposed to air should never be re-inserted

Page 12: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

12

Sample Contamination

Liquid Samples- Should not re-use sample bottles- Grabbing a bottle that is just ‘lying around’ - Not rinsing out test kit containers- Sampling from a common tank- Re-using bacteria inoculation syringes

Pipe Samples- External environment

Coupons- Should never install a coupon if it has been dropped on the ground- Need to make sure wearing gloves when handling the coupon- Coupons should not be re-inserted if they have been removed from the system- Coupons should not be analyzed if they have been removed from the system and not

properly stored prior to analysis

Page 13: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

13

Bacteria Testing

Perform immediately after sample collection or pipe exposure- Prevent contamination

Serial dilution media should match the salinity of the water present in the pipeline

Should not expel any of the inoculation media

Swabs need to be taken over consistent area size

Proper labeling

Page 14: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

14

Gas Testing

What are you actually testing?

Need to purge sample container or testing container

Some measurements are better performed on-site

Storage containers for gas that may contain H2S

What are on-line instruments monitoring

Do you need to check for O2

Page 15: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

15

Improper Storage of Samples

Samples need to be properly labeled (if further analysis is to be performed)

Liquid samples should be stored in a cool environment until analyzed

Bacteria inoculation vials should be stored at similar temperature to pipeline operating temperature

Ok

Page 16: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

16

Misinterpreting Results

Garbage In = Garbage Out

Page 17: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

17

Misinterpreting Results

How do operating conditions affect the results?- Well work- Change in flow direction- Change in flow rate

Iron counts- Iron to Manganese ratios- Correlation to corrosion- Trending

Bacteria

Page 18: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

18

Misinterpreting Results - Bacteria

Planktonic- Free floating- Liquid sampling

Sessile- Attached to pipe wall- Solids or swab sampling

Correlation to MIC- No known correlation between planktonic bacteria levels and the presence of MIC!!!

Page 19: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

19

Misinterpreting Results

Do I have enough data to make a decision?- Trending

Do the data make sense?- Is further investigation needed? (e.g., no liquids are recovered from drips but leaks are still

occurring in the system)

Do the data support current monitoring/sampling and mitigation?

Page 20: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

20

Missing Sampling Opportunities

After pigging operations- Liquids- Solids- Swabs

As pipe is being removed from system- Liquids- Solids- Swabs

Page 21: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

21

SAFETY

Safety is of primary importance!!!

Sampling is always concern when collecting a sample from a live pipeline system

Questions???- Is there potential for projectiles?- Are isolation valves shut?- Are there concerns for being sprayed with liquids?- Are there any pyrophoric materials present?

Page 22: Errors in Sampling/Monitoring in an Internal Corrosion Program

06 October 2009

22

Safeguarding life, property and the environment

www.dnv.com