fmri methods lecture2 – mri physics

55
fMRI Methods Lecture2 – MRI Physics

Upload: neci

Post on 16-Jan-2016

45 views

Category:

Documents


2 download

DESCRIPTION

fMRI Methods Lecture2 – MRI Physics. Magnetic fields. magnetized materials and moving electric charges. Electric induction. Similarly a moving magnetic field can be used to create electric current (moving charge). Electric induction. Or you could use an electric current to move a magnet…. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: fMRI Methods Lecture2 – MRI Physics

fMRI Methods

Lecture2 – MRI Physics

Page 2: fMRI Methods Lecture2 – MRI Physics

magnetized materials and moving electric charges.

Magnetic fields

Page 3: fMRI Methods Lecture2 – MRI Physics

Similarly a moving magnetic field can be used to create electric current (moving charge).

Electric induction

Page 4: fMRI Methods Lecture2 – MRI Physics

Or you could use an electric current to move a magnet…

Electric induction

Page 5: fMRI Methods Lecture2 – MRI Physics

Force and field directions

Right hand rule

Page 6: fMRI Methods Lecture2 – MRI Physics

Protons are positively charged atomic particles that spin about themselves because of thermal energy.

Nuclear spins

Page 7: fMRI Methods Lecture2 – MRI Physics

μ (magnetic moment) = the torque (turning force) felt by a moving electrical charge as it is put in a magnet field.

Magnetic moment

The size of a magnetic moment depends on how much electrical charge is moving and the strength of the magnetic field it is in.

A Hydrogen proton has a constant electrical charge.

Page 8: fMRI Methods Lecture2 – MRI Physics

Earth’s magnetic field is relatively small (0.00005 Tesla), so the spins happen in different directions and cancel out.

Spin alignment

Page 9: fMRI Methods Lecture2 – MRI Physics

But when in a strong external magnetic field (e.g. 1.5 Tesla).

Spin alignment

Page 10: fMRI Methods Lecture2 – MRI Physics

Sum of magnetic moments in a sample with a particular volume at a given time.

Net magnetization (M)

Page 11: fMRI Methods Lecture2 – MRI Physics

Hydrogen protons not only spin. They also precess around the axis of the magnetic field.

PrecessionM

agne

tic f

ield

dire

ctio

n

True for all atoms with an odd number of protons

Page 12: fMRI Methods Lecture2 – MRI Physics

Two factors govern the speed of precession (Larmor frequency): magnetic field strength & gyromagnetic ratio

Larmor frequency = Bo * /2π

Precession speed

Page 13: fMRI Methods Lecture2 – MRI Physics

Gyromagnetic ratio ( )

Magnetic moment / Angular momentum

Combination of electromagnetic and mechanical forces.

Angular momentum is dependant on the mass of the atom.

Gyromagnetic ratio

Page 14: fMRI Methods Lecture2 – MRI Physics

Different atoms have different gyromagnetic ratios:

Gyromagnetic ratio

Nucleus Gyromagnetic ratio (γ)1H 267.5137Li 103.96213C 67.26219F 251.662

23Na 70.76131P 108.291

Page 15: fMRI Methods Lecture2 – MRI Physics

Different atoms placed in the same magnetic field have different Larmor frequencies:

“Tune in” to the Hydrogen frequency.

Larmor frequency

Nucleus Larmor Frequency at 1 Tesla1H 42.576 MGHz7Li 16.546 MGHz13C 10.705 MGHz19F 40.053 MGHz

23Na 11.262 MGHz31P 17.235 MGHz

Page 16: fMRI Methods Lecture2 – MRI Physics

The hydrogen atoms are precessing around z (direction of B0)

Longitudinal & transverse directions

Page 17: fMRI Methods Lecture2 – MRI Physics

Net magnetization is all pointing in the z direction

Steady state

Page 18: fMRI Methods Lecture2 – MRI Physics

Applying a perpendicular magnetic field “flips” the protons

Excitation pulses

Page 19: fMRI Methods Lecture2 – MRI Physics

Excite the sample in a perpendicular direction and let it relax.

Net magnetization of the sample changes as it relaxes, inducing current to move in a near by coil.

Excitation & Relaxation

Larmor frequency

Page 20: fMRI Methods Lecture2 – MRI Physics

Defined by the strength of B1 pulse and how long it lasts (T)

θ = *B1*T

This is one of the parameters we set during a scan

It defines how far we “flip” the protons…

Flip angle

xy

z

xy

z

900 pulse

xy

z

xy

z

1800 pulse

xy

z

xy

z

<900 pulse

xy

z

xy

z

>900 pulse

Page 21: fMRI Methods Lecture2 – MRI Physics

T1: relaxation in the longitudinal directionT2*: relaxation in the transverse plane

T1 and T2/T2*

Changes in the direction of the sample’s net magnetization

Page 22: fMRI Methods Lecture2 – MRI Physics

Realignment of net magnetization with main magnetic field direction

T1

Before excitation At excitation Relaxation

Net magnetization along the longitudinal direction

Page 23: fMRI Methods Lecture2 – MRI Physics

T1

T1 = 63% recovery of original magnetization value M0

Page 24: fMRI Methods Lecture2 – MRI Physics

What influences T1?

Has something to do with the surroundings of the excited atom. The excited hydrogen needs to “pass on” its energy to its surroundings (the lattice) in order to relax.

Different tissues offer different surroundings and have different T1 relaxation times…

We can also introduce external molecules to a particular tissue and change its relaxation time. These are called “contrast agents”…

Page 25: fMRI Methods Lecture2 – MRI Physics

Loss of net magnetization phase in the transverse plain

T2*/T2

Before excitation At excitation Relaxation

Net magnetization in the transverse plain

Page 26: fMRI Methods Lecture2 – MRI Physics

T2/T2*

T2 = 63% decay of magnetization in transverse plain

Page 27: fMRI Methods Lecture2 – MRI Physics

Two main factors effect transverse relaxation:

1. Intrinsic (T2): spin-spin interactions. Mechanical and electromagnetic interactions.

2. Extrinsic (T2’): Magnetic field inhomogeneity. Local fluctuations in the strength of the magnetic field

experienced by different spins.

T2* = T2 - T2’

Page 28: fMRI Methods Lecture2 – MRI Physics

T2’Magnetic field inhomogeneities

Examples of causes:Transition to air filled cavities (sinusoids)Paramagnetic materials like cavity fillingsMost importantly – Deoxygenated hemoglobin

Page 29: fMRI Methods Lecture2 – MRI Physics

What influences T2?

Again, has to do with the molecular neighborhood affecting the amount and quality of spin-spin interactions.

Different tissues will have different T2 relaxation times.

The stronger the static magnetic field, the more interactions there are, quicker T2 decay.

Page 30: fMRI Methods Lecture2 – MRI Physics

MR signalWe only have one measurement:

Measurement of the net magnetization in the transverse plain as the sample relaxes.

Once T2* relaxation is completeProtons precess out of phase in the transverse plain

Net magnetization in transverse plain = 0

Page 31: fMRI Methods Lecture2 – MRI Physics

Two important scanning parameters:

TR – repetition time between excitation pulses.

TE – time between excitation pulse and data acquisition (“read out”).

Creating scanning protocols with different TR and TE lengths will allow us to derive T1 and T2/T2* relaxation times.

TR and TE

Page 32: fMRI Methods Lecture2 – MRI Physics

Short TR = weaker MR signal on consecutive pulses.

TR length & MR signal strength

With short TRs relaxation in the longitudinal direction will not be complete. So there will be fewer relaxed protons to excite.

Page 33: fMRI Methods Lecture2 – MRI Physics

TE: when to measure MR signalWe can measure the amplitude of net magnetization

immediately after excitation or we can wait a bit.

Longer TEs will allow more transverse relaxation to happen and the MR signal will be weaker.

Page 34: fMRI Methods Lecture2 – MRI Physics

We can scan the brain using different pulse sequences by choosing particular TR and TE values to create images

with different contrasts.

Different image contrasts

TR length will determine how much time the sample has had to relax in the longitudinal direction.

TE will determine how much time the sample has had to relax (loose phase) in the transverse plain.

Page 35: fMRI Methods Lecture2 – MRI Physics

Measuring the amount of hydrogen in the voxels regardless of their T1 or T2 relaxation constants.

Proton density contrast

This is done using a very long TR and very short TE

Page 36: fMRI Methods Lecture2 – MRI Physics

Higher intensity in voxels containing more hydrogen protons

Proton density

Page 37: fMRI Methods Lecture2 – MRI Physics

Measuring how T1 relaxation differs between voxels.This is done using a medium TR and very short TE

T1 contrast

You need to know when largest difference between the tissues will take place…

Page 38: fMRI Methods Lecture2 – MRI Physics

Images have high intensity in voxels with shorter T1 constants (faster relaxation/recovery = release of more energy)

T1 contrast

CSF: 1800 msGray matter: 650 ms White matter: 500 ms Muscle: 400 msFat: 200 ms

Page 39: fMRI Methods Lecture2 – MRI Physics

Measuring how T2 relaxation differs between voxels.This is done using a long TR and medium TE

T2/T2* contrast

We can combine a T2 acquisition with proton density…

Page 40: fMRI Methods Lecture2 – MRI Physics

Images have high intensity in voxels with longer T2 constants (slower relaxation = more detectable energy)

T2 contrast

CSF: 200 msGray Matter: 80 ms White Matter: 60 ms Muscle: 50 msFat: 50 ms

Page 41: fMRI Methods Lecture2 – MRI Physics

Same as T2 only smaller numbers (faster relaxation)

T2* contrast

CSF: 100 msGray Matter: 40 ms White Matter: 30 ms Fat: 25 ms

Page 42: fMRI Methods Lecture2 – MRI Physics

T2* = T2 +T2’

T2: Spin-spin interactionsT2’: field inhomogeneities

Exposed iron (heme) molecules create local magnetic inhomogeneities

T2* and BOLD fMRI

BOLD – blood oxygen level dependant

Assuming everything else stays constant during a scan one can measure BOLD changes across time…

Page 43: fMRI Methods Lecture2 – MRI Physics

More deoxygenated blood = more inhomogeneity

more inhomogeneity = faster relaxation (shorter T2*)

Shorter T2* = weaker energy/signal (image intensity)

So what would increased neural activity cause?

T2* and BOLD

Page 44: fMRI Methods Lecture2 – MRI Physics

So what happened in particular time points of this scan?

T2* and BOLD

Page 45: fMRI Methods Lecture2 – MRI Physics

Bloch equation

Page 46: fMRI Methods Lecture2 – MRI Physics

So far we’ve talked about a bunch of forces and energies changing in a sample across time…

How can we differentiate locations in space and create an image?

MR images

Paul Lauterbur Peter Mansfield

2004 Nobel prize in Medicine

Page 47: fMRI Methods Lecture2 – MRI Physics

Create magnetic fields in each direction (x,y,z) that move from stronger to weaker (hence gradient).

Spatial gradients

Page 48: fMRI Methods Lecture2 – MRI Physics

Different magnetic fields at different points in space.

Hydrogen will precess at a different speed in each spatial location.

By “tunning in” on the specific precession speed we can separate different spatial locations.

Similarly to how we “tunned in” on hydrogen atoms…

Spatial gradients

Page 49: fMRI Methods Lecture2 – MRI Physics

Spatial gradients

64 MHz

65 MHz

66 MHz

63 MHz

62 MHz

G

(-)

(+)

Page 50: fMRI Methods Lecture2 – MRI Physics

Lot’s of Fourier transforms.

Work in k-space (a vectorial space that keeps track of the spin phase & frequency variation across magnet space).

It’s possible to turn gradients on and off very quickly (ms).

Image reconstruction

Pulse sequences

Spatial gradients

Page 51: fMRI Methods Lecture2 – MRI Physics

The magnet

Page 52: fMRI Methods Lecture2 – MRI Physics

Main static field

Extremely large electric charge spinning on a helium cooled (-271o c) super conducting coil.

Earth’s magnetic field 30-60 microtesla.

MRI magnets suitable for scanning humans 1.5-7 T.

Page 53: fMRI Methods Lecture2 – MRI Physics

Main coils

The bulk of the structure contains the coils generating the static magnetic field and the gradient magnetic fields.

Page 54: fMRI Methods Lecture2 – MRI Physics

RF coil

Transmit and receive RF coils located close to the sample do the actual excitation and “read out”.

Page 55: fMRI Methods Lecture2 – MRI Physics

Read Chapters 3-5 of Huettel et. al.

Explain how a spin-echo pulse does the magic of separating T2 relaxation from T2* relaxation. You can include figures/drawings if you like.

Homework!