impedance spectroscopy for manufacturing control … · impedance spectroscopy for manufacturing...

110
Impedance Spectroscopy for Manufacturing Control of Material Physical Properties Xiaobei Li A thesis submitted in partial fulfillment of the requirements for the degree of Master of Science in Electrical Engineering University of Washington 2003 Program Authorized to Offer Degree: Department of Electrical Engineering

Upload: tranxuyen

Post on 21-Aug-2018

225 views

Category:

Documents


0 download

TRANSCRIPT

Impedance Spectroscopy for Manufacturing Control of

Material Physical Properties

Xiaobei Li

A thesis submitted in partial fulfillment of the requirements for the degree of

Master of Science in Electrical Engineering

University of Washington 2003

Program Authorized to Offer Degree: Department of Electrical Engineering

University of Washington

Graduate School

This is to certify that I have examined this copy of a master’s thesis by

Xiaobei Li

and have found that it is complete and satisfactory in all respects,

and that any and all revisions required by the final

examining committee have been made.

Committee Members:

_________________________________________

Alexander Mamishev

_________________________________________

Lloyd Burgness

_________________________________________

Karl Böhringer

Date: _________________

In presenting this thesis in partial fulfillment of the requirements for a Master’s degree at

the University of Washington, I agree that the Library shall make its copies freely

available for inspection. I further agree that extensive copying of this thesis is allowable

only for scholarly purposes, consistent with "fair use" as prescribed in the U.S. Copyright

Law. Any other reproduction for any purposes or by any means shall not be allowed

without my written permission.

Signature_______________________________

Date___________________________________

University of Washington

Abstract

Measuring Physical Properties of Organic Materials Using Dielectric Spectroscopy

by Xiaobei Li

Chairperson of the Supervisory Committee

Assistant Professor Alexander Mamishev

Department of Electrical Engineering

Real-time non-invasive sensing techniques are needed for online process control

in manufacturing industries. Impedance spectroscopy is a powerful sensing tool that can

be used for real-time non-invasive process parameter control. Applications currently

under investigation in this thesis involve moisture content sensing in food and

pharmaceutical products, and hardness and coating thickness evaluation for

pharmaceutical samples. A self-containing data acquisition and sensor control system is

designed for these applications. The system is able to perform real-time capacitance and

conductance measurements, and process the data to obtain parameters of interest. The

system can be calibrated according to the requirement of each application and can be

integrated into the feedback loop of the corresponding process control system. System

calibration involves establishing a one-to-one mapping between the parameters of interest

and the measured material impedance. The effect of other parameters needs to be

eliminated or accounted for. Experimental data demonstrates good sensitivity to

parameter variation. After compensation for disturbance factors, such as moisture

diffusion, a nearly linear dependency is observed between cookie dough moisture content

and the measured sample impedance. The investigation in pharmaceutical applications is

still at a preliminary stage. Experimental results indicate the feasibility for a broad

application of this technique in the pharmaceutical industry. A large amount of

experiments need to be conducted for a comprehensive calibration process.

i

Table of Contents

List of Figures .................................................................................................................... iv

List of Tables .................................................................................................................... vii

Chapter 1. Introduction................................................................................................. 1

1.1. Background..................................................................................................... 1

1.2. Motivation....................................................................................................... 1

1.3. State of the art ................................................................................................. 2 1.3.1 Techniques for measuring material properties........................................................................2 1.3.2 Dielectrometry sensing............................................................................................................2

1.4. Outline of this thesis ....................................................................................... 3

Chapter 2. Basics of dielectrometry sensing ................................................................ 5

2.1. Introduction to the theory of dielectrics.......................................................... 5

2.2. Principle of dielectric spectroscopy sensing ................................................... 6 2.2.1 Sensing possibilities.................................................................................................................6 2.2.2 Impedance spectroscopy and dielectric spectroscopy .............................................................7 2.2.3 Calibration based sensing .......................................................................................................8 2.2.4 Differential sensing .................................................................................................................8 2.2.5 Imaging – electrical impedance tomography (EIT).................................................................8

2.3. From parallel plate sensors to fringing field sensors .................................... 11

2.4. Penetration depth .......................................................................................... 12

2.5. Disturbance factors ....................................................................................... 13 2.5.1 Surface contact quality ..........................................................................................................13 2.5.2 Stray capacitances.................................................................................................................14 2.5.3 Deviation from ideal finite element analysis model...............................................................15 2.5.4 Interfacial double layer effect................................................................................................16

Chapter 3. Interdigital Fringing Field Dielectrometry................................................ 18

3.1. Overview of the measurement system .......................................................... 18

3.2. Fringing electric field sensor design ............................................................. 20 3.2.1 Figures of merit .....................................................................................................................20 3.2.2 Major design concerns ..........................................................................................................23 3.2.3 Example of multi-channel fringing field sensor designs........................................................27

ii

3.3. Sensor interface circuit ................................................................................. 35

Chapter 4. Moisture dynamics in cookies .................................................................. 38

4.1. Definition of the problem.............................................................................. 38

4.2. Methodology................................................................................................. 39

4.3. Experimental setup........................................................................................ 39 4.3.1 The Concentric Sensor Head.................................................................................................39 4.3.2 A Voltage Divider Circuit......................................................................................................41

4.4. Experimental procedure ................................................................................ 42

4.5. Experimental result and data analysis........................................................... 42 4.5.1 Compensation for Moisture Diffusion ...................................................................................43 4.5.2 Linear Regression..................................................................................................................44 4.5.3 Compensation for Non-Uniform Air Gap ..............................................................................45 4.5.4 Moisture Content Distribution...............................................................................................45 4.5.5 Evaluation of the Calibration Model.....................................................................................46

4.6. The effect of temperature variation............................................................... 47 4.6.1 The double layer effect ..........................................................................................................49 4.6.2 Lumped circuit simulation .....................................................................................................49

4.7. Simulating the manufacturing process – the rotating table........................... 52

4.8. The chemometric challenge – temperature and moisture control chamber .. 52

4.9. Conclusions................................................................................................... 53

Chapter 5. Measuring Physical Properties of Pharmaceutical Samples ..................... 54

5.1. Problem statement......................................................................................... 54

5.2. Motivation..................................................................................................... 54

5.3. Measuring tablet hardness and coating thickness ......................................... 55 5.3.1 Information on sample physical properties ...........................................................................55 5.3.2 Experimental setup ................................................................................................................56 5.3.3 Experimental results ..............................................................................................................57

5.4. Measuring tablet coating thickness............................................................... 58 5.4.1 The experimental setup..........................................................................................................58 5.4.2 The experimental results – parallel plate ..............................................................................59 5.4.3 The experimental results – fringing field...............................................................................62

iii

5.5. Acquiring drug signature using a FEF sensor............................................... 65

5.6. Choice of sensors – FEF vs. parallel plate.................................................... 67

5.7. Measuring API concentration for powder samples....................................... 69

5.8. Conclusion and future work.......................................................................... 71

Chapter 6. Conclusions and future work .................................................................... 73

6.1. Conclusions................................................................................................... 73

6.2. Directions of future work.............................................................................. 73 6.2.1 Information decoupling for multivariable experiments .........................................................73 6.2.2 More sophisticated parameter estimation algorithms ...........................................................73 6.2.3 Statistical evaluation of experimental results........................................................................74

End notes........................................................................................................................... 75

References......................................................................................................................... 80

Appendix........................................................................................................................... 85

1. DiSPEC hardware installation guide ............................................................ 85

2. DiSPEC software guide ................................................................................ 91

iv

List of Figures

Figure Number Page

2.1. Flow diagram of the dielectrometry system................................................................. 7

2.2. Transition from parallel plate geometry to in-plane fringing field geometry............ 11

2.3. A guard ring parallel plate capacitor.......................................................................... 11

2.4. Interdigital fringing electric field sensor with spatial wavelength λ.......................... 12

2.5. Cross-section view of a fringing electric field sensor................................................ 13

2.6. Maxwell simulation layout of a sample positioned above an interdigital sensor. ..... 15

2.7. Illustration of the double layer effect......................................................................... 16

3.1. Diagram of the measurement system......................................................................... 18

3.2. Labview device control and data acquisition interface.............................................. 19

3.3. Three-channel sensor interface circuit. ...................................................................... 20

3.4. Sensor interface circuit schematics. .......................................................................... 20

3.5. Maxwell simulation result of an interdigital fringing field sensor. ........................... 22

3.6. Transparent sensor ..................................................................................................... 25

3.7. Maxwell simulation results of a concentric-ring fringing field setup........................ 27

3.8. Three-wavelength fringing electric field sensor. ....................................................... 28

3.9. Three-wavelength fringing electric field sensor. ....................................................... 29

3.10. Top down view of a concentric fringing field sensor head...................................... 30

3.11. Top down view of a shielded concentric fringing field sensor head. ...................... 30

3.12. Maxwell simulation layout of a sample above the concentric FEF sensor.............. 31

3.13. Normalized capacitance measurement from the inner sensing channel .................. 32

3.14. Normalized capacitance measurement from the outer sensing channel .................. 32

3.15. The effect of the addition of shielding electrodes.................................................... 33

3.16. Maxwell simulation results of the two concentric fringing field sensor.................. 34

3.17. Normalized capacitance data from the simulation results of Maxwell.................... 34

3.18. A couple of novel designs........................................................................................ 35

3.19. Floating voltage with ground. .................................................................................. 36

3.20. Floating voltage with guard. .................................................................................... 36

v

4.1. Top and bottom view of the concentric sensor head.................................................. 39

4.2. Side view of the sensor in a voltage divider setup..................................................... 40

4.3. Detailed circuit model considering double layer effect. ............................................ 41

4.4. Sensor geometry and experimental setup. ................................................................. 42

4.5. Capacitances measured at different moisture content levels. .................................... 43

4.6. Phase measurements at different moisture content levels.......................................... 43

4.7. Capacitance measurements against the mass of added water at 10 kHz.................... 44

4.8. Moisture content distribution across the radius of the sample .................................. 46

4.9. Moisture loss dynamics of the cookie dough sample. ............................................... 48

4.10. Capacitance measurements of cookie dough sample............................................... 48

4.11. Phase measurements of cookie dough sample against sample surface temperature.49

4.12. Lumped circuit model for the double layer effect. .................................................. 50

4.13. Frequency dependency of the lumped circuit model. .............................................. 51

4.14. Frequency dependency of the lumped circuit model. .............................................. 51

4.15. The rotating table setup............................................................................................ 52

4.16. Moisture and temperature control chamber. ............................................................ 53

5.1. Photo of the pharmaceutical samples used in the experiments.................................. 55

5.2. Tablet sample weight and thickness against sample pressure. .................................. 56

5.3. Capacitance measurements of 180 mg tablet samples against sample hardness. ...... 58

5.4. Phase measurements of 180 mg tablet samples against sample hardness. ................ 58

5.5. Fringing electric field sensor setup for measuring tablet coating thickness. ............. 59

5.6. Absolute capacitance measurements from a parallel plate sensor. ............................ 60

5.7. Capacitance variation measurement from the parallel plate setup............................. 60

5.8. Phase variation measurement from the parallel plate setup....................................... 61

5.9. Capacitance variation against sample weight using a parallel plate sensor............... 61

5.10. Absolute capacitance measurements from the FEF setup........................................ 63

5.11. Capacitance variation from a FEF sensor with spatial wavelengh of 500 µm......... 63

5.12. Phase variation from a FEF sensor with spatial wavelengh of 500 µm................... 64

5.13. Capacitance variation against sample weight. ......................................................... 64

5.14. Fringing electric field sensor setup for acquiring drug signature. ........................... 65

5.15. Capacitance measurements from a FEF sensor. ...................................................... 66

vi

5.16. Phase measurements from a FEF sensor with spatial wavelength of 500 µm. ........ 66

5.17. Normalized capacitance from the parallel plate setup. ............................................ 68

5.18. Normalized capacitance from the frining field setup............................................... 69

5.19. Capacitance measurements of powder samples of various drying time. ................. 70

5.20. Phase measurements of powder samples of various drying time. ........................... 71

5.21. Capacitance measurements of powder samples against sample drying time........... 71

vii

List of Tables

Table Number Page

4.1 Comparison between actual moisure and added moisture.......................................... 47

5.1. Tablet sample physical properties: hardness, weight, thickness................................ 56

viii

Acknowledgments

The previous one and half years of research experience at SEAL has been very

pleasant, thanks to my advisor, Prof. Alexander Mamishev, who is very understanding

and fun to work with. The thesis would not have been possible without his consistent

guidance and support.

I am also very grateful towards my lab mates, Michael for proofreading the thesis,

Kishore for providing some of the experimental data, Alexei for the many interesting

discussions and ideas we shared, Sam for the contribution in sensor design, Henry and Yu

cheung for help with software development, and Kelly and Chika for running the

Maxwell simulations.

I would also like to acknowledge the financial support from Kraft foods for the

moisture sensing project. I am especially grateful to Dr. Carol Zrybko and Dr. Robert

Magaletta, for their help and advice on the project.

1

Chapter 1. Introduction

1.1. Background Dielectrometry is widely used for determination of material physical properties due to its

non-invasiveness and wide spectrum of sensing possibilities. Applications include

agricultural products [1], soil [2], paper [3], transformer board [4], biological sensing

[5,6] and hydrophilic polymers [7].

Capacitive sensors are often used for dielectric spectroscopy. They have the

advantage of high measurement accuracy and non-invasiveness. The simplest examples

of capacitive sensors are a guard-ring parallel-plate capacitor and a coaxial cylindrical

capacitor. More complicated examples include fringing electric field sensors, which can

assume various geometries [8,9]. The penetration depth of fringing electric field sensors

is proportional to the distance between coplanar electrodes. By applying different voltage

patterns to the sensor, variable penetration depths can be achieved, thus providing FEF

sensors access to different layers of the material. This characteristic, combined with their

one-sided access capability, makes FEF sensors more flexible in use than their parallel-

plate counterparts.

1.2. Motivation Lack of control in industrial processes limits the productivity of the manufacturers. Real-

time, non-invasive sensing systems are needed for feedback control of the parameters of

interest, such as moisture content, texture, hardness, and viscosity. This need has driven

many advances in the field of dielectrometry. This thesis discusses an impedance

spectroscopy technique, where functional dependencies between material properties of

interest and electrical impedance measurements are determined empirically and used to

calibrate the sensing system. The major challenge of this technique is to achieve

selectivity for the sensing system. Material properties other than those of interest are

usually considered to be disturbance factors, the effects of which have to be accounted

for in the parameter estimation algorithm. Other challenges include optimization of

sensor geometry for the particular application and inverse problem solving.

2

1.3. State of the art

1.3.1 Techniques for measuring material properties

1.3.1.1 NMR and MRI There has been extensive application of NMR and MRI techniques in bio-sensing and

medical imaging. Compared with other imaging techniques such as X-ray, NMR and

MRI have the advantage of non-invasiveness. Applications of NMR and MRI to food

product sensing have also been developed [10,11], especially in the field of food imaging

[12-16]. Although these techniques offer high measurement accuracy, so far they are not

fit for real time control industrial applications due to their high cost.

1.3.1.2 Ultrasound sensing Another popular technique in the field of bio-sensing and medical imaging is ultrasound.

Extensive study has been done in this direction. Other applications of ultrasound

technology include monitoring the curing process of resin [17].

1.3.2 Dielectrometry sensing Most industrial applications do not have high accuracy requirements, while production

cost needs to be kept as low as possible. Compared with the sensing technologies

mentioned above, dielectrometry sensing does not require special, high-caliber

measurement devices. This offers dielectrometry sensing techniques great flexibility to be

integrated into the manufacturing control processes.

1.3.2.1 Microwave and RF sensing Microwave and RF spectroscopy techniques are available for non-invasive sensing of

materials properties. They have been used for sensing the property of agricultural

products [1] as well as food products [18]. Near infrared spectroscopy is widely

employed for a number of qualitative studies as well as quantitative analysis of material

properties. The spectral region investigated by NIR covers the wavelength range from

700 nm to about 2500 nm. Applications of NIR spectroscopy include moisture content

determination for impregnated paper [19].

3

1.3.2.2 Dielectrometry sensing of food products Despite the advantage of low cost, applications of dielectrometry sensing to food

products are relatively scarce than those of NMR and MRI. Applications already under

investigation include evaluation of dielectric property of the biscuit dough [20] and

imaging of the cooking process of bread samples [16].

1.3.2.3 Dielectrometry sensing of pharmaceutical products Although the technique of dielectrometry sensing has existed for a long time, its

application to the pharmaceutical industry is very recent. Applications already under

investigation include bioahesive gels [21], measurement of solids, detection of inter-batch

variation, measurement of emulsions and lipsome suspensions, the characterization of

proteins and biomolecules [22], the evaluation of thermal aging effect in pharmaceutical

systems [23].

1.3.2.4 Microdielectrometry Microdielectrometry was first proposed by Senturia [24]. Since then great advances have

taken place in the field MEMS, the prospect of a low-cost, power-efficient, and

disposable mini-sensor is no longer just a distant dream. However, the field of

microdielectrometry sensing has been relatively stagnant. There are some recent efforts

of applying microdielectrometry to bio-sensing. Dielectrometry sensors at MEMS scale

allow us to study the physical properties of individual cells. Another potential application

is in the study of sensitive skin, where each dielectrometry sensor cell simulates a neuron

of the human body [25].

1.4. Outline of this thesis Chapter 1 provides the state of art for dielectrometry sensing and its applications. Chapter

2 gives an overview on the basics of the dielectrometry theory. A brief description and

comparison between impedance spectroscopy, dielectric spectroscopy, and electric

impedance tomography is provided. Fundamentals of fringing electric field sensing are

also provided. Chapter 3 talks about the various aspects of experimental setup design,

which include sensor design, interface circuit design, and the circuit calibration

4

algorithm. A special focus is placed on multi-channel fringing electric field sensor

design, yet some of the issues addressed in this chapter apply to all types of sensor

geometries. Chapter 4 focuses on parameter estimation algorithms. The forward and the

inverse problem are defined in this chapter. Algorithms for disturbance factor

compensation are also discussed. Chapter 5 and 6 deal with the experimental results and

data analysis from the cookie dough and pharmaceutical applications respectively.

5

Chapter 2. Basics of dielectrometry sensing

2.1. Introduction to the theory of dielectrics

Materials are usually divided into the categories of conductors, insulators, and dielectrics.

Dielectric materials cover the whole spectrum of anything between conductors and

insulators. Dielectrics consist of polar molecules, or non-polar molecules, or very often

both. Due to the asymmetric configuration of polar molecules, material consisting of

these molecules has built-in dipole moments. Under an external electric field, the

polarized dipoles reorient in the electric field and neutralize some of the charges on the

electrodes. The most often used measure of material dielectric properties is the complex

dielectric permittivity. It is a measure of the ability of the dielectric material to reorient

and neutralize charges on the electrodes. This usually depends on how polarized the

material is and the inertial force it has to overcome to reorient. Sometimes, relative

complex dielectric permittivity is used to describe material dielectric properties. It is

defined as the ratio between the dielectric permittivity of the material and that of free

space. The dielectric permittivity of free space is 128.85 10 /F m−× .

The dielectric permittivity of most dielectric materials is frequency-dependent. In the

presence of an alternating electric field, the dipole moments inside the material oscillates

with the direction of the electric field. The higher the frequency the harder it is for the

dipole moments to catch up with the change of field direction. This results in a decreasing

ability of the material to neutralize charges on the electrodes at high frequencies. In

general, the total complex dielectric permittivity ε*(ω) is written as:

*( ) '( ) ''( )iε ω ε ω ε ω= −

(2.1)

where 'ε and ''ε are, respectively, the real permittivity and the dielectric loss factor of

the material.

Jonscher of the Chelsea Dielectric Group has been studying the problem of a

universal relaxation law [26]. But till now, no one has proved the existence of a general

model to describe the dielectric relaxation process. One of the most widely used models

for fitting dielectric relaxation data is the Havriliak-Negami (HN) function, as shown in

6

(2.2), where 0ε is the dielectric permittivity at dc and ε∞ is its asymptotic value at

infinite high frequency. The term 0ε ε∞− is the total dielectric relaxation strength and 0τ

is the relaxation time of the material. For 1β = , the Cole-Cole model emerges; whereas

for 1α = the Davison-Cole model emerges.

( )

0

0

*( )1 i

βα

ε εε ω εωτ

∞∞

−= +

+ (2.2)

''σ ωε=

(2.3)

2.2. Principle of dielectric spectroscopy sensing

2.2.1 Sensing possibilities

Dielectrometry is one of the most versatile sensing techniques. Figure 2.1 shows the flow

diagram of a dielectrometry system. Material dielectric permittivity is dependent on

various material physical properties such as geometry, texture, temperature, degree of

cure, moisture content and aging status. Changes in these physical properties will be

reflected as changes in such dielectric property variables as *ε ,σ , and tanδ , where σ is

defined in (2.3) and tanδ is defined as the ratio between the real and imaginary part of

the complex impedance. These parameters variations, in turn, lead to changes in the

impedance measurements from the sensor. The fact that dielectric measurements are

sensitive to changes of a wide range of material physical properties makes dielectrometry

sensing technique a potential candidate for a broad spectrum of sensing applications.

7

Measurecapacitancesand conductances

Calibration-basedsensing

Differentialsensing

Computedistributionof dielectricproperties

Computedistributionof physicalproperties

Imaging

Faster Slower

ε σ, δ, Μ*, tan ∗

thicknesssurface texturetemperaturedegree of curemoistureporositydensityconcentrationpercolationstructural integrityaging statuscontamination .......

Figure 2.1. Flow diagram of the dielectrometry system.

2.2.2 Impedance spectroscopy and dielectric spectroscopy

2.2.2.1 Impedance spectroscopy – lumped circuit representation of dielectrics

Rather than focus on details of what happens inside dielectric materials, electrical

engineers often analyze dielectrics from a macroscopic perspective. Impedance

spectroscopy is one such macroscopic approach. It models dielectrics as lumped circuit

elements and uses the terminal electric impedance measurement to represent the physical,

chemical and biological processes happening inside the material.

A proper choice of circuit model is crucial to obtaining good modeling results. The

dielectric spectra that can be represented by combinations of RC circuits are called

relaxation spectra, whereas those that can be represented by combinations of RL circuits

are called resonance spectra. To determine the choice of circuit models, the frequency

dependency of material dielectric constant should be examined. For relaxation spectra,

the dielectric constant only stays constant or falls with increasing frequency. In these

cases, RC circuit models should be used. If otherwise, RL or RLC circuit model should

8

be used [27].

In addition to lumped circuit models, distributed circuit models are sometimes used

to model dielectric materials as a distributed dielectric medium in bounded or unbounded

space [28,29].

2.2.2.2 Dielectric spectroscopy Dielectric spectroscopy relates material dielectric properties with corresponding physical

properties and investigates the fundamental theoretical link between them. For industrial

applications, where in-depth theoretical knowledge of material nature is unnecessary,

impedance spectroscopy is sufficient. Dielectric spectroscopy is often used for research

efforts investigating material dielectric properties.

2.2.3 Calibration based sensing

Calibration based sensing works by establishing a quantitative relationship between

material physical property of interest and the resulting impedance measurements. This

functional dependence is usually empirically determined. Very often a linear dependency

is assumed. The algorithms for such calibration-based approaches are usually quite

straightforward, yet these approaches are not sufficient to gain insight about the physical

nature of the material and the calibration parameters are always subject to changes when

a different setup is adopted.

2.2.4 Differential sensing

Differential sensing works by scanning over the test specimen and analyzing the

measurement variation at different locations. This sensing technique provides fast

relative information about the material under test. It is suitable for industrial monitoring

applications, where speed is important and high measurement accuracy is not required.

2.2.5 Imaging – electrical impedance tomography (EIT)

For inhomogeneous materials, interest exits to look at the distribution of the dielectric

and physical properties of the material. EIT is one of the most widely used technologies

9

for such applications. Applications of EIT include bio-sensing, medical imaging,

geophysics sensing, and industrial process control.

Compared with other tomography and imaging techniques, EIT is relatively

inexpensive, which makes it a popular choice for industrial applications. However, unlike

X-ray or laser imaging, EIT is the “soft field” technique in which the field lines that

penetrate through the material do not stay in a straight path. This makes the parameter

estimation algorithms for EIT much more challenging than those for other techniques.

Compared with differential sensing and calibration based sensing, EIT and other

such imaging techniques are usually slower, due to the high computation complexity

involved in the parameter estimation algorithms.

2.2.5.1 Principle of EIT The goal of EIT is to estimate the resistivity distribution of the interior of the material by

measuring the voltages or current between the electrodes positioned at the surface of the

material.

The maximum degrees of freedom achievable by an EIT system is determined by the

number of electrodes according to the relationship in (1), where α represents the degrees

of freedom and n represents the number of electrodes.

( 1)2

n nα × −= (1)

2.2.5.2 Major disturbance factors in clinical applications

1.1.1.1.1 The ill-conditioning problem

A matrix is defined as ill-conditioned or ill-posed if the ratio of the maximal and the

minimal eigenvalues is very large. Ill-conditioning causes matrix inversion to be very

inaccurate and sensitive to measurement error. In the case of EIT systems, depending on

the resistivity distribution and data collection methods, the matrix could be ill-posed. If

current does not travel through some region, the resistivity change does not yield much

voltage change at the boundary and this results in ill-conditioning.

2.2.5.2.1 The skin effect

10

The skin effect is a major disturbance factor for clinical applications of EIT. The skin is

composed of layers of dead cells. At low frequency, the impedance of unabraded skin can

be as high as 1 MΩ/cm2. Its impedance decreases if the dead skin on the surface is

scraped off. The existence of this huge shunt impedance makes it very difficult to

measure the internal distribution of resitivity accurately. Phantoms are usually used to

calibrate and evaluate imaging systems like EIT. However, it is very difficult to find a

physical analog to model human skin in the phantom, namely a thin layer of material with

high impedance. Polyimide comes closest to the requirement, but is still not good enough.

One solution to this problem is to move from the two-electrode setup to the four-

electrode setup, where current is injected through one pair of electrodes and voltage

measured from another pair of electrodes, thus reducing the effect of skin impedance.

The four-electrode setup is obviously more complicated than the two-electrode one. For

applications where only differential sensing is of interest and absolute measurement

accuracy is not required, such as those applications in geophysics, the two-electrode setup

is often used.

Another solution is to inject current from multiple pairs of electrodes. This is called

multi-reference approach. The skin effect is less pronounced in this case due to the

relatively large supply current.

2.2.5.2.2 Electrode position

One major source of uncertainty for EIT systems is the electrode position. Information on

the exact position of the electrodes is necessary for the estimation of interior resitivity

distribution. This is difficult to achieve due to the irregularity of the human body.

Electrodes can be embedded in a rigid mold before being applied to patients, yet this is

not feasible for clinical applications because of the discomfort induced for patients.

Information on the exact location of electrodes is not necessary in such dynamic imaging

applications where only changes in tissue resitivity are of interest. In differential imaging

applications like this, all collected data is referenced against an initial data set, which

already take the electrode positions into account.

11

2.3. From parallel plate sensors to fringing field sensors A fringing electric field sensor can be formed by unfolding the electrodes of a parallel

plate sensor as shown in Figure 2.2. Figure 2.2 (a) shows an ideal parallel plate sensor

where the fringing field effect at the edge is ignored. The three-electrode guarded parallel

plate setup, as shown in Figure 2.3, is often used to avoid fringing field effects.

Figure 2.2. Transition from (a) parallel plate geometry to (c) in-plane fringing field geometry by unfolding the electrodes.

Figure 2.3. A guard ring parallel plate capacitor. The guard-ring is adopted to remove fringing field effect

0 r ACd

ε ε=

(2.4)

0 AGd

ε σ=

(2.5)

For guard-ring parallel plate sensors, there exist straight forward analytical solutions,

as shown in (2.4) and (2.5), where A is the area of the parallel plate electrode, d is the

distance between the two plates, ε0 is the dielectric permittivity of free space, εr is the

relative dielectric permittivity of the material, and σ is the conductivity of the material.

Such a general analytical solution is lacking for fringing field sensors. Unlike what is

drawn in Figure 2.2 (c), the field lines of fringing field sensor are not evenly distributed,

as is the case for parallel plate sensors. Field energy tends to focus around sharp edges

12

and places closer to the electrodes. Therefore sensor measurement sensitivity to material

properties of the specimen is different at different positions. This greatly increases the

challenge of the inverse problem of solving for distribution of material properties from

the electrical measurements of FEF sensors.

Figure 2.4 shows an interdigital fringing field electric field sensor with spatial

periodicity λ, where spatial periodicity is defined as the distance between coplanar

electrodes. When an ac voltage signal is applied to the driving electrodes, the sensor

generates a fringing electric field. The field strength and distribution pattern depend both

on the input voltage signal and the sensor geometry. The concept of a very important

parameter, penetration depth, which is related to both field strength and distribution, is

explained in detail in the next section.

Figure 2.4. Interdigital fringing electric field sensor with spatial wavelength λ. Wavelength (periodicity) is defined here as the distance between coplanar electrodes.

2.4. Penetration depth For interdigital fringing electric field sensors, penetration depth is often defined as one

third, one fourth, or one fifth of the periodicity of the sensor. One way to evaluate sensor

effective penetration depth is to measure γ3%, which is defined as the position at which the

13

difference between the asymptotic value and the measured value of sensor terminal

impedance is 3%. This technique is often used to compare the performance of several

sensor designs. Both experimental results and finite element analysis results can be used

to estimate the penetration depth of a FEF sensor.

The penetration depth of interdigital fringing field sensors is proportional to the

wavelength (periodicity) of the sensor. Figure 2.5 shows the side view of a fringing

electric field sensor. By applying different voltage pattern, as shown in the figure, various

penetration depths can be achieved, thus provide the sensor access to different layers of

the material.

Figure 2.5. Cross-section view of a fringing electric field sensor with multiple penetration depths. The penetration depth of a fringing field sensor is proportional to the distance between coplanar electrodes. “D” represents the drive electrode, “S” represents the sensing electrode and “G” represents the ground electrode.

2.5. Disturbance factors

2.5.1 Surface contact quality

In a parallel plate setup, where a sample is placed between two parallel electrodes, the

terminal impedance can often be modeled as a Maxwell capacitor with different

dielectrics in series. For a Maxwell capacitor, terminal impedance measurements are not

sensitive to vertical displacements of the material under test [30]. This property offers

parallel plate sensors robustness against surface contact disturbance.

14

The scenario is quite different for FEF sensors. Since most of the field energy for

FEF sensors concentrates around electrodes, especially at the edges, the electrical

measurements from FEF sensors are very sensitive to uncertainties caused by surface

contact quality. This fact has to be considered when designing an experimental setup

involving FEF sensors. When the material under investigation is of liquid form, this issue

is not a problem. For solid samples, there are several ways to reduce this effect.

1. Since FEF sensors are capable of non-contact measurement, the sample can

be placed away from the sensor surface to a certain distance. Using air as an

intermediate dielectric, the effect of the surface irregularities of the electrode

and material surface will be attenuated. However, this method also reduces

the signal strength of the measurements.

2. Use flexible substrate and apply pressure from the top to ensure maximum

contact. Experimental results show that this method does reduce surface

disturbance to some extent, but not completely.

Due to the inherent strong non-linearity, it is difficult to find parameter estimation

algorithm that can compensate for the effect very well.

2.5.2 Stray capacitances

For the capacitive sensors used in this study, the magnitude of the electrical impedance

measurements is roughly proportional to the surface area of sensor electrodes. The

surface sensing area of fringing field sensors is usually less than that of their parallel

plate counter-parts. This results in a much weaker signal strength, which makes FEF

sensor measurements much more susceptible to noise.

Stray capacitance is a major source of disturbance. The capacitances between the

sensing electrodes and their respective back planes affect measurement accuracy if not

eliminated or compensated for.

Another source of stray capacitance is the sensor leads and the coaxial wires used to

connect the sensor with the interface circuit. These capacitances can be minimized by

15

proper shielding and reducing the length of the leads and the coaxial wires.

Operational amplifiers are usually used in the sensor interface circuit. The input

capacitance of op-amps is usually around 4 or 5 pF. The FEF sensor experimental

capacitance measurements involved in this study are usually at the scale of 0.01 to 0.1 pF.

This shows that, if not properly compensated for, the input impedance of the op-amps can

cause inaccuracy in the resulting measurements. Compensation is carried out here in this

thesis during circuit calibration. The exact procedures will be introduced in later chapters.

2.5.3 Deviation from ideal finite element analysis model

Finite element analysis simulations are often used to evaluate the performance of a sensor

and to the test the validity of experimental results. In the study of this thesis, the Ansoft

FEA software Maxell is used. Figure 2.6 shows the Maxwell simulation layout of the

interdigital sensor shown in Figure 2.4. For the simulation, several assumptions are made:

1. All fingers are of infinite length along the y direction.

2. The finger patterns are periodic along the x axis.

Deviations from this ideal model may lead to discrepancies between experimental and

Maxwell simulation results.

Figure 2.6. Maxwell simulation layout of a cookie sample positioned above an interdigital sensor. The wavelength of the sensor is 16 mm. It is assumed that the

16

fingers have infinite length and that the finger pattern is periodic along the horizontal axis.

2.5.4 Interfacial double layer effect

Figure 2.7 shows the circuit model for a test sample with double layer, where Cs and Rs

are the effective capacitance and resistance of the material under test while Cdl is the

double layer capacitance.

sZ

sCdlCdlC

sR

Figure 2.7. Illustration of the double layer effect. The double layer is formed at the metal-electrolyte boundaries [31]. Ions carrying

opposite charges are diffused into the other side of the boundary until equilibrium is

reached. The depth at which the diffusion process stops is called the Debye length. Debye

length is usually very small. This thin layer of opposite charges forms a great capacitance

Cdl.

The interfacial effect is more pronounced in case of liquid dielectrics, due to the

much higher tendency for diffusion. For example, a much stronger double layer effect is

observed for cookie dough samples than with ready-made cookies. The moisture and oil

in the cookie dough precipitate the diffusion process. One effective way of minimizing

the layer effect is to place a layer of polyimide film between the metal electrodes and the

sample under investigation. The polyamide film here serves as a liquid barrier. The water

absorption rate for polyamide is 3% of its dry weight. However, due to the high dielectric

17

permittivity of water, the dielectric property of the polyimide films changes a lot even

with minimal amount of moisture intake [7,32,33].

The DuPont Kapton is chosen as the barrier in this study. There are three common

varieties of Kapton films, Kapton HN, Kapton VN and Kapton FN, and they come in

different thicknesses ranging from 25 µm to 125 µm. Among all three, Kapton FN is most

hydrophobic, being coated on either or both sides with Teflon as an additional moisture

barrier.

The double layer capacitance causes inaccuracies in the estimates of the material

impedance Cs and Rs. Since the double layer capacitance is Cdl is usually much larger

than the effective material capacitance Cs, it dominates at low frequency. If the

operational frequency range is high enough, the interfacial effect can be ignored. The

measurements involved in this study are conducted at the frequency range of 50 Hz to

100 kHz, which is not high enough for the effect to be ignored.

It is easy to detect the double layer effect but difficult to estimate the exact value of

the Debye layer capacitance. One way to test the existence of the layer effect is by using

the Cole-Cole plot. The Cole-Cole plot of a capacitor and resistor in parallel is a semi-

circle. Deviation from a perfect semi-circle indicates the existence of the double layer.

The Debye length of the double layer can be estimated from the magnitude of the

deviation, yet to accurately quantify this effect is a non-trivial task. Therefore, the general

approach is to reduce, if not avoid, the double layer effect.

18

Chapter 3. Interdigital Fringing Field Dielectrometry

3.1. Overview of the measurement system Figure 3.1 shows the diagram of the measurement system used in the study of this thesis.

The central part of the system is the sensor array. The sensors designs have to be

optimized toward the particular application. An AC voltage signal comes from the signal

generator to the driving electrodes. The output signal from the sensing electrodes is

measured by the sensor interface circuit board. The electrical measurements then are sent

to the computer through an embedded data acquisition board. Labview software controls

the function generator through a GPIB board and also performs on-line data acquisition

and signal processing. The system works at the frequency range between 0.1 Hz to 30

kHz.

The work of the author focuses on the area of sensor design, sensor interface circuit

design, and Labview software development.

GPIB BOARD

COMPUTER

DAQ BOARD

SIGNAL GENERATOR

SENSORARRAY

SENSOR INTERFACE

POWER SUPPLY

Figure 3.1. Diagram of the measurement system.

Figure 3.2 shows the front panel of the Labview software user interface. The software allows real-time monitoring of impedance data in both the frequency and the time domain. For the particular application of cookie moisture sensing, the software can perform real-time moisture profile imaging of cookie dough samples.

19

Figure 3.2. Labview device control and data acquisition interface

Figure 3.3 shows the three channel interface circuit. The circuit is based on a voltage

divider scheme where the ratio of the magnitude of the output voltage signal and the

input voltage signal, and the phase delay between the input and the output are evaluated.

Unit gain buffers are used in the circuit to isolate the measured signal from the later

stages of the circuit. The circuit schematic is shown in Figure 3.4.

20

Figure 3.3. Three-channel sensor interface circuit.

Figure 3.4. Sensor interface circuit schematics.

3.2. Fringing electric field sensor design Parallel plate sensor designs are usually straightforward. The chapter focuses on the more

challenging task of designing FEF sensors, especially multi-channel FEF sensors.

However, some of the issues addressed here are common to all sensor designs.

3.2.1 Figures of merit

Sensor design is an optimization process, where several figures of merit have to be

considered simultaneously and necessary trade-offs are made. The major design goals are

explained in detail in this section.

21

3.2.1.1 Penetration depth The penetration depth of fringing field sensors is roughly proportional to the periodicity

of the sensor. The farther the electrodes are positioned away from each other, the higher

the penetration depth, which means the electrical field lines penetrate deeper into the test

specimen. For thick specimens, it is desirable to use sensors with large periodicity so that

sufficient penetration depth is accommodated.

3.2.1.2 Signal strength The design parameters that mostly determine the signal strength of a sensor are the total

sensor surface area and the relative metallization ratio. The greater these two parameters,

the stronger the signal strength. Metallization ratio is defined as the ratio of surface area

of the active electrodes (guard electrodes not included) against the total sensing area.

Most of the designs developed in this study use a 50% metallization ratio. Another

parameter that affects the signal strength is the magnitude of the input driving signal.

However, increasing the input also raises the noise floor. Thus, raising the input voltage

usually has little effect on the SNR, the major parameter of interest. A more efficient way

to improve SNR is to magnify the signal using instrumental amplifiers. The earlier the

amplification, but better the overall SNR performance.

3.2.1.3 Sensitivity Sensitivity is defined as the slope of the measurement curve, namely the ratio between

the measurement variation and the variation of the measured physical property. Due to

the uneven field distribution of FEF sensors, its measurement sensitivity varies at

different locations. As indicated by Figure 3.5, sensitivity decreases exponentially with

increasing distance between the specimen and the sensor. Thus, to maximize sensitivity,

it is desirable to position the specimen as close to the sensor as possible. This is

especially important when considering that signal strength of FEF sensors is usually

weak.

22

Figure 3.5. Maxwell simulation result of an interdigital fringing field sensor. Measured capacitance value decays exponentially as the sample is moved away from the sensor. Measurement sensitivity is defined as the slope of the measurement curve. It drops even faster with increasing distance between the specimen and the sensor surface. In most cases, the limiting factor for measurement sensitivity is not the sensor itself,

but rather, the resolution of the interface circuit. For example, the fringing field

concentric sensor used in the study of this paper is able to detect environmental variations

(say, a moving hand) 4 or 5 spatial periodicities away from the sensor. However, only a

subtle change in capacitance measurements (at the scale of fF for the particular setup

used in this study) results from such a variation, and it can be easily buried by the noise.

To increase circuit resolution, the noise floor has to be reduced. Shielding and signal

magnification are the common ways to reduce noise.

3.2.1.4 Dynamic Range The dynamic range is the range of the physical parameter of interest that the sensor can

measure, namely the range between the smallest and biggest signal the sensor can

measure. The challenge of the lower end lies mainly in the measurement resolution of the

electrical device; the higher end is determined by factors like the common mode range of

the operational amplifiers on the measurement board. The inherent material property

affects both.

23

3.2.1.5 Number of channels Sometimes, instead of using sensor arrays, it is desirable to fit multiple channels on the

same sensor. The more channels the sensor has, the more information available about the

material.

3.2.1.6 Noise tolerance Guard planes are usually introduced to improve sensor noise performance. This includes

both the guard ring on the top plane of the sensor surface and the guard plane at the back

of the sensor substrate. Proper positioning of these guard electrodes is essential for

optimal results. In addition, it is desirable to have all the driving electrodes on both sides

of all sensing electrodes. However, this might be difficult to accommodate in some cases

due to size limitation.

3.2.2 Major design concerns

3.2.2.1 Choice of sensor substrate

The choice of sensor substrate should be optimized toward the particular application.

Major points to consider are:

1. Should the substrate flexible or rigid?

2. What are the desired electrical, mechanical, and chemical properties for the sensor

substrate of this particular application? For example, does it matter whether the

material is hydrophilic or hydrophobic? This is of particular importance due to the

double layer effect that often occurs in dielectrometry measurements.

Proper choices should be made based on the answer to these questions. For certain

application, no substrate is necessary. The electrode can be applied directly to the test

specimen.

3.2.2.2 Choice of electrode material

3.2.2.2.1 Surface contact quality

For applications with solid dielectrics, surface contact quality is one of the major sources

of measurement uncertainties. Any air gap between the electrodes and the test specimen

24

acts as a capacitance in series with the RC component of the specimen and results in a

decreased measurement in both the material capacitance and conductance. For accurate

measurements, the test specimen should be measured by thin metallic electrodes before

put between rigid electrodes. Such products as silver paints and low-melting metals that

can be applied with a spray-gun are commercially available. They conform readily to the

surface of rough specimens and can greatly improve surface contact quality. The

disadvantages of these electrodes are that they are usually difficult to pattern and remove

[34].

In clinical applications of electrical impedance tomography, saline gels are utilized

to improve electrode and skin contact. The most widely used is NaCl gel. The

concentration of NaCl has to be carefully controlled to avoid irritating the skin. Aside

from improved contact quality, the electrolyte also helps to reduce the contact impedance

by hydrating the skin [35].

3.2.2.2.2 Novel electrode materials

Most electrodes are made from metal. Novel materials are sometimes needed for special

applications. One type of such novel materials is called transparent conductive polymer.

They are transparent polymer films coated with a thin layer of conductive material. These

films can be either cut and pasted on a substrate, or patterned through etching and act as a

flexible sensor by itself. The material is useful for application where additional optical or

visual information is necessary. Figure 3.6 shows a sensing electrode made from such

material.

25

Figure 3.6. Transparent sensor fabricated by sputtering Indium Tin Oxide onto a thin polyester sheet.

3.2.2.3 Size limitation If the sensor is embedded on a flexible substrate, the sensor can be wrapped around the

test specimen to guarantee that all parts of the material are sensed. For sensor with rigid

substrate, however, the size of the fringing field sensor is usually limited by the size of

the material under investigation. Usually, the sensor should be of the same size or smaller

than the material so that all the surface area of the sensor is covered by the material. For

small and thick specimens, it is usually hard to design a sensor that has sufficient

penetration depth due to the electrode surface area limit posed by the size of the material.

3.2.2.4 Trade off between penetration depth, signal strength and the number of channels

Due to the sensor electrode surface area limitation mentioned above, there is a trade-off

between the penetration depth of the sensor and the number of channels that can be fit on

the sensor. To fit more channels on the sensor, the sensor electrodes have to be squeezed

closer together, the penetration depth, in turn, is affected. Equally harmed is the signal

strength of the sensor due to the reduced electrode surface area for each channel.

3.2.2.5 Cross talk between channels The closer the channels are positioned together, the stronger the cross talk between the

channels. It is desirable to position the channels a far apart as possible. In this way, cross

26

talk reduction is obtained by sacrificing sensor surface area and this comes with the price

of loss in signal strength and penetration depth. One solution to this problem is by

grounding the sensing electrodes of the channels not being used. However, this method

increases the complexity of the interface circuit, and only works when simultaneous

measurements are not required.

3.2.2.6 The positioning and the geometry of the back plane The back plane and the top sensor electrodes are separated by the substrate. Most sensor

substrates are very thin compared with the periodicity of the sensor. Due to the close

proximity of the sensor back plane to the top electrodes, the distribution of the sensor

field patterns is very sensitive to the positioning and the geometry of the back plane. The

field lines and field energy tend to be drawn away from the material being sensed by the

back plane, therefore, affecting the penetration depth and signal strength. Proper

positioning of the sensor back plane is thus essential to optimize sensor performance.

This is achieved mostly from design experience and with the help of Maxwell

simulations. Aside from these factors, attention should also be taken about the voltage

potential of the back planes. Whether they are grounded or set to the same voltage as the

top sensing electrodes also makes a difference in the resulting field energy distribution.

3.2.2.7 The thickness of the substrate As mentioned above, the thickness of FEF sensor substrate determines the distance

between the back planes to the top electrodes, which in turn affects the distribution of

field energy. To further illustrate this idea, Maxwell simulations are carried out, where a

concentric-ring fringing field sensor setup is simulated and the thickness of the substrate

varied from 97% to 25% of its original value. A test sample is moved away from the

sensor surface in this setup. The simulation results are shown in Figure 3.7. Capacitance

measurements are shown to decrease with decreasing substrate thickness. The thinner the

sensor substrate, the closer the back plane to the top electrodes and the more energy is

drawn away for the test specimen.

27

0 2 4 6 8 100

0.2

0.4

0.6

0.8

1

1.2

1.4

1.6

Distance (mm)

Cap

acita

nce

(pF)

97%50%25%

Figure 3.7. Maxwell simulation results of a concentric-ring fringing field setup. The substrate thickness is varied from 97% to 25% of its original value.

3.2.2.8 Conclusion The major obstacle in fringing field sensor design is the size limitation posed by the

geometry of the test specimen. The design goal is to fit as many channels as possible on

the sensor while still accommodating sufficient penetration depth and signal strength.

Novel sensor design and excitation patterns are being investigated to achieve this goal.

Other factors such as substrate thickness and back plane geometry should also be

considered.

3.2.3 Example of multi-channel fringing field sensor designs

3.2.3.1 Interdigital FEF sensors Figure 3.8 and Figure 3.9 show two different designs of multi-channel sensors. The one

shown in Figure 3.8 has three channels, each providing a different penetration depth.

Thus, the sensor has access to different layers of the test specimen and can provide

information on the specimen at these different layers. This is, however, only true for

homogeneous materials. In the case of an inhomogeneous specimen, it is difficult to

conduct comparative analysis on the measurements from the three different channels.

This is because the spatial information along the depth of the material and the horizontal

28

surface of the material are coupled together and extracting the depth information is a non-

trivial task.

Figure 3.8. Three-wavelength fringing electric field sensor.

Figure 3.9 shows an improved design. It still offers three measurement channels. In

addition, all three channels measure the specimen at the same horizontal location. In this

way, information on the distribution of material property at different layers can be

obtained by comparing the measurements from different channels.

29

Figure 3.9. Three-wavelength fringing electric field sensor.

3.2.3.2 Concentric FEF sensors

It is often necessary to optimize sensor geometry towards the geometry of the test

specimen. Figure 3.10 shows a concentric sensor head designed for measuring moisture

content in cookie dough samples, which usually assume a near-symmetric rounds shape.

The sensor can act as a two-channel FEF sensor, by using the center electrode as the

driving electrode and the other two electrodes as the sensing electrodes. The two sensing

channels of this sensor have roughly the same penetration depth due to the similarity in

wavelength/periodicity. The sensor measures the material at two different radial

locations. Vertical information about the test specimen at different layers can be obtained

by varying the distance between the sensor surface and the specimen.

30

Figure 3.10. Top down view of a concentric fringing field sensor head. The sensor has one driving electrode and two sensing electrodes and is capable of conducting simultaneous measurement from both sensing channels. The total sensing area of the sensor has to be smaller than a normal-sized cookie.

Thus the design was confined to an area of approximately 10 cm2. The periodicity of this

design is 8 mm. For a design with 50% metallization ratio like this, the penetration depth

is roughly one third of the periodicity and it should, in this case, be roughly 2 mm. Most

cookies are, however, thicker than 2 mm. Therefore, the penetration depth needs to be

improved.

Figure 3.11. Top down view of a concentric fringing field sensor head with additional shielding electrodes. The sensor has one driving electrode and two sensing electrodes and is capable of conducting simultaneous measurement from both sensing channels. Between driving and the sensing electrodes, shielding electrodes are added for improved penetration depth. Figure 3.11 shows an improved design where shielding electrodes are added between

the driving and the sensing electrodes for improved penetration depth. However, the

31

surface area of the active electrodes is sacrificed and the overall signal strength is

reduced.

3.2.3.3 Evaluation of sensor penetration depth Maxwell simulation allows us to evaluate the performance of a particular design before

the sensor is fabricated. Penetration depth is one of the major parameters of interest.

Figure 3.12 shows the layout of the Maxwell simulation for the concentric fringing field

sensor without shielding electrodes. In this simulation, the cookie sample is assumed to

have a real dielectric permittivity of 5.0. Penetration depth γ3% is defined here as the

distance at which measured capacitance drops to 3% of its asymptotic value (at infinite

distance).

Figure 3.12. Maxwell simulation layout of a cookie sample positioned above the concentric fringing field sensor. The simulation is carried out using the RZ radial coordinates. The capacitance data is normalized to a range between 0 and 100%. Penetration

depth can be estimated from the distance corresponding to the intersection point between

the normalized measurement curve and the 3% line. Figure 3.13 and Figure 3.14 show

respectively the normalized capacitance data from the inner and outer channel of the first

concentric sensor design.

It is worthy to note that is that instead of dropping monotonically with increasing

distance, the capacitance measurement from the inner channel actually goes up a bit at

32

the distance from 6 to 10 mm. To explain this phenomenon, the field line distribution

dynamics has to be considered. When the specimen is within certain distance to the

sensor surface, the relative high dielectric permittivity of the specimen draws the electric

field energy towards itself, thus alters the field line distribution. This change in field

distribution is the major reason for the rise in the capacitance measurement.

Figure 3.13. Normalized capacitance measurement from the inner sensing channel of the first concentric sensor design. The measured capacitance doesn’t change monotonically with distance.

Figure 3.14. Normalized capacitance measurement from the outer sensing channel of the first concentric sensor design. A monotonic dependence exists between capacitance measurement and the distance.

3.2.3.4 Comparative performance analysis of the two concentric designs The motivation behind adding the shielding electrodes in the second concentric fringing

sensor design is to improve sensor penetration depth, as illustrated in Figure 3.15.

33

Figure 3.15. The addition of shielding electrodes in the fringing electric sensor increases penetration depth.

Maxwell simulations are conducted to evaluate the effectiveness of this approach. Figure

3.16 and Figure 3.17 show respectively the absolute and normalized capacitance

measurement from the inner and outer channels on both sensors. Sensor 1 refers to the

design without shielding electrodes and sensor 2 refers to the one with shielding

electrodes. For both designs, the outer channel offer better signal strength and penetration

depth. The difference in signal strength is caused by the relative larger sensing area of the

outer channel. When comparing the performance between the two sensor designs, the

second design provides greater penetration depth at the price of decreased signal strength.

34

Figure 3.16. Maxwell simulation results of the two concentric fringing field sensor. Comparison of the absolute capacitance value shows that signal strength is weakened by introducing the shielding electrodes.

Figure 3.17. Normalized capacitance data from the simulation results of Maxwell. It is proved that the addition of the shielding electrodes increases sensor penetration depth. The outer sensing channel on either sensor has a higher penetration depth than that of the inner channel.

3.2.3.5 Other novel designs In addition to multiple programmable penetration depth, fringing field sensors can

provide a vast variety of field line distribution patterns. The flexibility of fringing electric

field sensors offers a lot of space for creativity in the design process. Novel designs like

the ones shown in Figure 3.18 have field lines that penetrate samples both along the

35

radial axis and tangential to the radial axis. The field patterns can be optimized toward a

particular application through proper arrangement of the electrodes. However, these two

designs are not as space efficient as the concentric design. Thus they are not used in the

current application.

Figure 3.18. A couple of novel designs. These designs are not suitable for the current application with cookie samples due to size limitation of this application.

3.3. Sensor interface circuit The dielectric spectroscopy sensing system developed in this study uses the 3-chaneel

sensor interface board shown in Figure 3.3. The interface board allows simultaneous

sensing of three channels. The circuits for all three channels are identical to each other,

all of which based on a voltage divider scheme. Figure 3.4 shows the schematics for the

sensor interface circuit.

A simplified circuit model for the voltage divider scheme is shown in Figure 3.19,

where G12 and C12 and the terminal impedances of between the sensor’s drive and sensing

electrodes. G11 and C11 are the conductance and capacitance between the sensing

electrodes and their respective backplane. Gdg and Cdg represent the conductance and the

capacitance of between the driving electrode and the ground plane. Note that these terms

are only considered in the fringing field set up, for the parallel plate setup, these values

are equal to zero. Cs represents the stray capacitance in the circuit, most probably

introduced by the op-Amp from the next stage, and CL represents the reference

capacitance in the circuit.

36

Since Gdg and Cdg are connected directly to the voltage input, they have no effect on

the output voltage measurement. Therefore, they can be ignored in the circuit analysis.

G11 and C11, however, affect the output voltage Vs and they are unknown and difficult to

estimate.

To eliminate the effects of G11 and C11, the circuit was modified to the one shown in

Figure 3.20. In this connection scheme, the back plane for each electrode is set to the

same voltage potential as their respective sensing electrode through a unity gain buffer

operational amplifier, which removes the effects of G11 and C11.

Figure 3.19. Floating voltage with ground.

Figure 3.20. Floating voltage with guard.

37

38

Chapter 4. Moisture dynamics in cookies

4.1. Definition of the problem Lack of process control has been thwarting food manufacturer’s efforts to reduce

production cost. Moisture content is one of the major control parameters of concern.

Accurate control of moisture content is critical to achieving the right taste and texture for

food products. By incorporating a moisture sensor in the feedback loop of the

manufacturing process, the moisture content can be accurately controlled in real time.

The integration of advanced sensing technologies in the cookie manufacturing process

allows automatic production of complex cookies that has been possible so far only in

bakeries. In this investigation, moisture content is determined from the impedance

measurements of the material of interest. Moisture concentration is defined here as

follows:

1%

1 2100%MM

M M= ×

+ (4.1)

where M1 is the mass of the moisture contained in the unit volume, and M2 is the mass of

the dry portion of the material in the same unit volume.

Material impedance is a function of many variables, as shown in (4.2), where M% is

the moisture concentration in the material, T is the ambient temperature, D is the sample

density, and ω is the input signal frequency.

%( , , , )s ZZ f M T D ω= (4.2)

System calibration involves solving the inverse problem of determining the

following function:

% ( , , , )M sM f Z T D ω= (4.3)

or

% ( )M sM f Z= (4.4)

where the functional dependence between moisture concentration and the impedance is to

be determined. The effects from variables other than moisture content are either

eliminated or accounted for.

39

4.2. Methodology The material contents of food products are usually complex and varying, which renders

direct determination of sample dielectric permittivity difficult and impractical. Under

these circumstances, an indirect parameter estimation approach based on quantitative

mapping between electrical measurements and the physical variable of interest can be

used. The major challenge for such an approach lies in minimizing the effect of variables

other than moisture concentration, such as ambient temperature and sample density,

which are considered here as disturbance factors. The effects of these factors should

either be eliminated or accounted for in the calibration algorithm [36].

4.3. Experimental setup

4.3.1 The Concentric Sensor Head

Figure 4.1 shows a concentric sensor head, designed for localized measurements. It has

three electrically separated sensing electrodes, each shielded by a guard plane on the back

of the substrate.

Channel 1 BottomTop

Channel 3Channel 2

Figure 4.1. Top and bottom view of the concentric sensor head. The center plate is 10 mm in diameter. The outer two rings are 5 mm wide. The spacing between adjacent sensing plates is also 5 mm. The guard planes on the back are slightly wider than respective sensing electrodes.

The sensor head can be used as a fringing field sensor by applying an AC sinusoidal

voltage to the middle ring electrode and measuring the voltage at the two neighboring

40

electrodes. A non-linear model is needed to describe such a setup. The solution to the

Laplace equation of the non-linear model is:

0 1 2( , ) ( )( )z zr z J r c e c eβ βφ β − += + (4.5)

where φ iss electric potential, r is to the radial coordinate on the horizontal plane, z is the

vertical coordinate, J0 is the zero order Bessel function of the first kind and β is a scaling

constant such that βr is one of the zeros of J0 [9].

The fringing field setup provides one-sided access but has limited signal strength. It

is also susceptible to disturbances from the contact quality between the samples and the

surface of the sensors. The parallel plate setup, on the other hand, is a complement to the

fringing field setup. It lacks the one-sided access but offers greater signal strength and is

comparatively less sensitive to surface contact qualities.

Figure 4.2. Side view of the sensor in a voltage divider setup. A cookie is placed between the sensing and driving plates.

This chapter presents experimental data obtained with the parallel plate arrangement

of Figure 4.2. A barrier made of 300 µm thick Kapton is used to avoid the Debye layer

effect [37]. The parallel plate capacitor can be modeled as a Maxwell capacitor with three

different dielectrics in series: air, polyimide (Kapton), and the material under test. For a

Maxwell capacitor like this, terminal impedance measurements are not sensitive to

vertical displacements of the polyimide and the material under test [30]. This property

makes parallel-plate sensors more robust to surface contact disturbances.

41

4.3.2 A Voltage Divider Circuit

Figure 4.2 shows a voltage divider circuit, where Vi is the input voltage signal, Vs is the

sensing voltage signal, Zr is the reference impedance, and Zs is the sensing impedance.

The effective impedance of the parallel-plate capacitor is calculated from the voltage

divider relationship 4.6. To maximize circuit sensitivity, Zr is chosen to be close to Zs. In

this investigation, Zr = 8 pF.

s r

i r s

V ZV Z Z

=+

(4.6)

Various connection schemes are available for this voltage divider setup [38]. When

there is an electric potential difference between the sensing electrodes and their

respective guard electrodes, stray capacitances are introduced into the circuit. To prevent

these stray capacitances from affecting measurement accuracy, the guard planes are set to

the same voltage as their respective sensing electrodes by using a unity-gain voltage

buffer.

Figure 4.3. Detailed circuit model considering double layer effect.

42

Figure 4.4. Sensor geometry and experimental setup.

4.4. Experimental procedure

To calibrate the moisture sensing system, a quantitative relationship between sample

moisture content and the corresponding impedance measurements needs to be

established. The following experiment was conducted to evaluate this relationship.

1. A test specimen is placed between the sensor plates so that the center of the specimen

is aligned with channel 1 of the sensing plate.

2. A 6 volt, 10 Hz to 10 kHz frequency sweep signal is applied to the circuit in Figure

4.2 and Vs is measured.

3. The moisture content of the sample is increased by adding increments of 0.2 grams of

water to the center point.

4. Measurements are taken at each moisture content level.

4.5. Experimental result and data analysis Figure 4.5 and Figure 4.6 show respectively the capacitance and phase variations due to

moisture content increase as measured by the center sensing electrode. Change in

moisture content leads to an increase in the capacitance and phase maxima and a shift of

the curves toward higher frequencies.

43

102

103

1040.14

0.16

0.18

0.2

0.22

0.24

0.26

Cap

acita

nce

(pF)

Frequency (Hz)

0 g

1.0 g

0.8 g

0.6 g

0.4 g

0.2 g

Figure 4.5. Capacitances measured at different moisture content levels.

102

103

104-12

-10

-8

-6

-4

-2

0

2

4

Frequency (Hz)

Pha

se (d

eg)

0 g

1.0 g0.8 g0.6 g0.4 g0.2 g

Figure 4.6. Phase measurements at different moisture content levels.

For capacitance measurements, the higher the signal frequency, the greater the

measurement sensitivity to moisture content. To achieve maximum sensitivity,

capacitance data at 10 kHz is used to calibrate the system, which here involves

establishing a quantitative mapping between capacitance values and moisture content.

4.5.1 Compensation for Moisture Diffusion

The triangles in Figure 4.7 show the channel 1 capacitance data at 10 kHz averaged

across different frequency sweeps. At higher moisture content level, moisture diffusion to

the outer channels reduces the capacitance increase between neighboring samples.

44

The higher the moisture content gradient between the center channel and the outer

rings, the more intensive the moisture diffusion process. This is reflected in the

increasing discrepancy between the uncompensated and compensated capacitance data as

water is being added to the center of the sample.

0.2 0.4 0.6 0.8 1 1.20

0.02

0.04

0.06

0.08

0.1

0.12

0.14

0.16

0.18

0.2

Moisture (g)

Cap

acita

nce

(pF)

Before compensation

After compensation

*16.4390 0.0028M C∆ = × ∆ +

Figure 4.7. Capacitance measurements against the mass of added water at 10 kHz for channel 1. Saturation occurs at high water level.

To compensate for the effect of diffusion, the capacitance increase from channels 2

and 3 is measured and mapped to effective increase in channel 1. This increase, added to

the original capacitance change from channel 1, gives the new channel 1 capacitance data

after compensation, as shown in (4.7), where ∆C is the capacitance increase for each

channel, Cor is the capacitance measurement of the original sample for each channel, and

∆C1* is the channel 1 capacitance increase after compensation.

* 1 11 1 2 3

2 3

or or

or or

C CC C C CC C

∆ = ∆ + ∆ + ∆ (4.7)

As indicated by the solid line in Figure 4.7, a much better linear approximation is

achieved after compensation.

4.5.2 Linear Regression

Assuming a linear functional dependence, the following calibration equation is

determined for channel 1 by performing linear regression on the compensated data:

45

31 16.44 2.8 10M C −∆ = × ∆ + × (4.8)

In this configuration, all sensor pixels are parallel-plate capacitors with different area.

Ideally, the functional dependence for channels 2 and 3 can be obtained by scaling (4.8)

with the respective area ratio. However, the existence of a non-uniform air gap has to be

taken into account.

4.5.3 Compensation for Non-Uniform Air Gap

The air gap between the material and the top plate of the sensor is non-uniform due to an

uneven shape of the cookie samples. To compensate for this, uniform water distribution

in the original sample is assumed and the ratios of capacitance measurements from

channels 2 and channel 3 with respect to channel 1 are measured. This difference in

capacitance measurements from the three channels is caused partly by the difference in

sensing plate area and partly by the non-uniformity in air gap thickness. Taking the ratios

obtained above and using them as scaling factors, the functional dependence of

capacitance measurements on water content from channels 2 and 3 can be obtained from

(4.8).

4

2 21.43 6.22 10M C −∆ = × ∆ + × (4.9)

4

3 30.91 3.95 10M C −∆ = × ∆ + × (4.10)

4.5.4 Moisture Content Distribution

Based on the calibration equations (4.8), (4.9), and (4.10), the absolute mass of moisture

contained in the portion of the sample above each ring is calculated from the capacitance

measurements. The mass of the dry portion of the sample above each ring is determined

from the ratio of the respective sensing electrode area to the area of the whole sample.

Given the absolute mass of moisture and the dry portion of the sample, moisture content

levels for all three channels can be calculated according to (4.1), enabling real-time

imaging of moisture content distribution.

Figure 4.8 shows the moisture content distribution profile of a sample at various

moisture content levels, which is obtained from fitting the moisture content data from the

46

three channels to scaled Gaussian curves.

2

% 2( )

2

xAM x e σ

πσ

− = (4.11)

where x is the distance to the center of the sensing plate, σ is a measure of the width of

the curves and A is a scaling factor, which is determined by the moisture diffusion

coefficient of the diffusion process.

Figure 4.8. Moisture content distribution across the radius of the sample when different amount of water is added to the center.

4.5.5 Evaluation of the Calibration Model

The calibration approach discussed above involves several approximations. To evaluate

the effectiveness of the model obtained through system calibration, the absolute masses

of the moisture measured from all three channels are summed up and compared with the

mass of the moisture added to the sample. As indicated in Table 4.1, measurement error

decreases with increasing moisture content. Further processing of experimental data is

needed to reduce the error at low moisture content levels.

47

Table 4.1 Comparison between the actual mass of the moisture added to the sample and the mass of the moisture measured by the sensor.

Moisture Added (g) Moisture Measured (g) Error

0.2 0.130 35%

0.4 0.312 22%

0.6 0.570 5%

0.8 0.776 3%

1.0 0.980 2%

4.6. The effect of temperature variation In the calibration process above, constant ambient temperature is maintained. Under

temperature-varying conditions, the effect of temperature has to be taken into account.

Additional experiments are carried out to evaluate the temperature effects. In these

experiments, a piece of cookie dough sample is heated by an oven. Impedance, weight

and surface temperature of the sample are measured during the heating process.

Assuming that sample weight variation is completely caused by moisture loss, a three-

way relationship between the impedance, moisture and temperature measurements of the

sample can be established.

Figure 4.9 shows the moisture loss dynamics of the heating process. A significant

increase in the rate of moisture is witnessed when the sample surface temperature reaches

80 °C.

Figure 4.10 and Figure 4.11 show, respectively, the sample capacitance and phase

measurements against sample surface temperature.

48

Figure 4.9. Moisture loss dynamics of the cookie dough sample against sample surface temperature.

Figure 4.10. Capacitance measurements of cookie dough sample against sample surface temperature.

49

Figure 4.11. Phase measurements of cookie dough sample against sample surface temperature.

4.6.1 The double layer effect

Double layer is formed at the interface between metal and electrolytes. At the metal-

dielectrics boundary, the ions carrying opposite charges diffused into the other side of the

boundary. This thin layer of charges forms a very big capacitance in series with the

capacitance and conductance of the dielectric material under consideration.

The cookie dough has much higher moisture content than the cookies used in the

previous constant temperature experiments. Therefore, the double layer effect which is

almost negligible in the previous case is now much more pronounced. The adoption of a

layer of polyimide between the material and sensor electrode surface still can not

completely eliminate the effect.

4.6.2 Lumped circuit simulation

To confirm the existence of the double layer effect, simulation of a lumped circuit model

as shown in Figure 4.12 was carried out, where C1 and R1 represent the capacitance and

resistance of the test sample and C2 represents the capacitance due to double layer effect.

The exact value of C2 is difficult to estimate. The existence of the unknown capacitance

makes it much more challenging to measure sample impedance accurately. Therefore,

double layer effect is undesirable. Kapton was used in the previous constant temperature

50

experiment to reduce the interfacial effect. However, in this case, the effect is much more

pronounced and it can not be cancelled by the adoption of a layer of Kapton.

Figure 4.12. Lumped circuit model for the double layer effect.

Compare the simuation result of the lumped circuit model and the cookie dough

experimental data shown in Figure 4.13 and Figure 4.14, a similarity in frequency

depedency can be seen, which confirms the exitstence of double layer effect. Now instead

of a RC parallel circuit model, the double layer effect model shown in Figure 4.12

should be adopted for the estimation of sensor terminal impedances. The value of C1, R1,

and C2 should be varied until good fit is obtained between the simulation and

experimental results. Note, however, this is a underdetermined problem, where we have

three unknows and only two equations. Prior information of material properties can be

used to provide additional information for the estimation.

51

102 103 104 105100

150

200

250

300

Frequency (Hz)

Cap

acta

nce

(pF)

102 103 104 1050

0.5

1

1.5 x 10-5

Frequency (Hz)

Con

duct

ance

(S)

102 103 104 1050

0.2

0.4

0.6

0.8

1 x 10-4

Frequency (Hz)

Cur

rent

(Am

p)

102 103 104 105-90

-80

-70

-60

-50

-40

Frequency (Hz)

Pha

se (d

eg)

33 deg78 deg87 deg

Figure 4.13. Frequency dependency of the lumped circuit model.

102 1040

1

2

3

4

5

Frequency (Hz)

Cap

acita

nce

(pF)

102 10410-12

10-10

10-8

10-6

Frequency (Hz)

Con

duct

ance

(S)

102 10410-9

10-8

10-7

10-6

Frequency (Hz)

Cur

rent

(Am

p)

102 104-80

-60

-40

-20

0

Frequency (Hz)

Phas

e (d

eg)

Figure 4.14. Frequency dependency of the lumped circuit model.

52

4.7. Simulating the manufacturing process – the rotating table To simulate the cookie manufacturing process, the rotating table setup, as shown in

Figure 4.15, is designed. Samples can be positioned on the plate with computer

controllable rotational motion. A fringing field sensor is attached to a platform, whose

vertical motion is controllable. Efforts are under way to design a automatic scanner,

where the sesnor can move according to a programmed trace and collect data at different

positions.

Figure 4.15. The rotating table setup.

4.8. The chemometric challenge – temperature and moisture control chamber

Ambient temperature is the major disturbance factor for this moisture sensing application.

Information on the moisture content is coupled with effect of temperature effect and it is

usually difficult to estimate accurately the change in moisture content in the presence of

temperature variations.

Figure 4.16 shows a moisture and temperature control chamber where the ambient

moisture concention and temperature can be independently controlled. This chamber is

ideal for running single variable experiments.

53

Figure 4.16. Moisture and temperature control chamber.

4.9. Conclusions Due to the strong correlation between material moisture concentration and its dielectric

properties, dielectrometry measurements were used for sensing cookie moisture content.

The optimized concentric sensor head enables measurement at different locations of the

sample. Impedance data shows a nearly linear dependence on moisture content. The

sensor is calibrated based on a linear model and real-time moisture content imaging is

achieved.

In this investigation, moisture content is determined from the impedance

measurements of the material of interest, without calculating the distribution of the

complex dielectric permittivity ε*. Impedance spectroscopy is shown to be adequate for a

controlled experiment. Future work is likely to involve implementation of inverse

problem solution methods to determine the spatial distribution of ε*. Efforts are

underway to integrate fringing sensors into the current setup, which involves new sensor

design and modeling. The quantitative effect of temperature variation is being

investigated so that it could be incorporated into the current calibration algorithm.

54

Chapter 5. Measuring Physical Properties of Pharmaceutical Samples

5.1. Problem statement If process control is important for the food industry, it is much more so for the

pharmaceutical industry. The Food and Drug Administration (FDA) has very stringent

regulations for pharmaceutical products in such aspects as active ingredient

concentration, tablet hardness, and coating thickness. To keep their reputations intact,

pharmaceutical manufacturers always consider quality control as their top priority.

The major obstacle for quality control of pharmaceutical products is the absence of

an accurate, efficient, and non-invasive sensing technique. The physical properties of

pharmaceutical samples are often measured by destructive tests, if they are measurable at

all. In cases where the manufacturers are not sure their products meet FDA standards,

they would rather sacrifice those products, than risking jeopardizing their reputation.

5.2. Motivation It is important to control when, where, and how much of pharmaceutical ingredients

dissolve in the body of a patient. Tablet hardness, coating thickness, and coating

roughness are among the physical parameters that determine these factors. For example,

tablet hardness and surface roughness are directly related to how fast the coating

dissolves in the body. Accurate sensing and control of these physical parameters are

crucial to obtaining the desired process of drug uptake.

This chapter investigates the feasibility of applying the non-invasive dielectrometry

sensing technique to quality control of such physical properties as tablet hardness,

coating thickness, and API content of powder samples. Tests of drug signatures are also

carried out to differentiate between unpolished, polished, and placebo tablet samples. The

results show good measurement sensitivity to parameters of interest. More extensive

experiments have to be conducted to quantify the dependencies between these physical

properties and the electrical measurements and compensate for disturbance factors.

55

Proper choice of a sensor is very crucial for achieving optimal measurement results.

Both FEF sensors and parallel plate sensors are used in the experiments. Comparative

analysis of the experimental results from these two types of sensors is provided at the end

of the chapter.

5.3. Measuring tablet hardness and coating thickness

5.3.1 Information on sample physical properties Tablet samples of known hardness are used in this feasibility study. Other information,

such as average tablet thickness and weight, are also available. Sample pressure

(hardness) affects both weight and thickness. Figure 5.2 shows the dependence between

these physical parameters. Increase in pressure leads to an increase in tablet density, and,

therefore, an increase in weight; at the same time, the pressure increase results in a

decrease in tablet thickness. Notice, however, the trend for weight variation is not strictly

monotonic. Table 5.1. shows the average values of hardness, weight and thickness for the

different groups of tablet samples.

Figure 5.1. Photo of the pharmaceutical samples used in the experiments.

56

Figure 5.2. Tablet sample weight and thickness against sample pressure.

Table 5.1. Tablet sample physical properties: hardness, weight, thickness.

Sample Hardness Weight Thickness

Sample #1 25.6 (kp) 0.592 (mg) 5.56 (mm)

Sample #2 31.5 (kp) 0.607 (mg) 5.53 (mm)

Sample #3 34.3 (kp) 0.610 (mg) 5.49 (mm)

Sample #4 41.8 (kp) 0.609 (mg) 5.27 (mm)

Sample #5 45.1 (kp) 0.615 (mg) 5.26 (mm)

5.3.2 Experimental setup The experimental results presented here are from a parallel plate setup. Tablet samples of

the same hardness are arranged side by side with the same orientation between the two

electrodes of the sensor. The sensor is driven by a 1 V AC voltage signal from a Fluke

RCL meter. The meter measures the loop AC current and sensor terminal impedance. The

AC signal sweeps from 50 Hz to 100 Hz.

57

5.3.3 Experimental results Equations (2.4) and (2.5), show, respectively, the capacitance and conductance for a

parallel plate setup, where A is the area of the parallel plate electrode, d is the distance

between the two plates, ε0 is the dielectric permittivity of free space, εr is the relative

dielectric permittivity of the material and σ is the conductivity of the material.

As mentioned previously, the major variables that affect electrical measurements are

sample density and sample thickness. Density affects material dielectric permittivity εr

and conductivity σ while thickness affects d. Therefore, capacitance measurements are

sensitive to changes in both sample density and thickness. Phase measurements, on the

other hand, are determined by the relative ratio between real and imaginary part of the

impedance. Change in sample geometry affects capacitance and conductance

measurement in the same fashion and leaves their relative ratio constant. Therefore, phase

measurements are only dependent here on density variations.

Figure 5.3 and Figure 5.4 show respectively the capacitance and phase measurement

of the tablet samples against hardness at various frequencies. Conductance and current

measurements are omitted because no additional information is offered.

According to Figure 5.2, increase in hardness results in a rise in sample density and a

drop in sample thickness, which affects the capacitance measurements adversely. The

resulting measurement shown in Figure 5.3 is a trade-off between these two effects,

which explains why the trend is not monotonic. Phase measurement displays a monotonic

dependence on hardness and bears information only about samples density. Using the

combined information from capacitance and phase measurements, samples density and

thickness can be uniquely determined. The experiments carried out so far proved the

feasibility of the technique. More extensive experiments are needed to fully calibrate the

measurement system.

58

Figure 5.3. Capacitance measurements of 180 mg tablet samples against sample hardness.

Figure 5.4. Phase measurements of 180 mg tablet samples against sample hardness.

5.4. Measuring tablet coating thickness

5.4.1 The experimental setup Experiments are carried out using both a parallel plate sensor and fringing field sensor.

Figure 5.5 shows the FEF setup. For the parallel plate sensor setup, 10 samples are

arranged side by side with the same orientation between the two electrodes of the sensor.

59

Figure 5.5. Fringing electric field sensor setup for measuring tablet coating thickness. The spatial wavelength of the sensor is 500 µm.

5.4.2 The experimental results – parallel plate Figure 5.6 shows the capacitance measurements of the tablet samples. To focus on the

detailed measurement variation between samples of different coating thickness, the

capacitance and phase measurements of the original uncoated sample are used as

references and subtracted from the measurements of all other samples. The results are

shown in Figure 5.7 and Figure 5.8. A clear dependency exists between the capacitance

variation data and sample coating thickness. Note that exact information on coating

thickness is not provided with the test samples used in these experiments. Here, weight

information, which is directly related to coating thickness, is used instead. Figure 5.13

shows the capacitance measurements acquired at 1 kHz plotted against sample weight. A

near-linear dependency is witnessed.

60

Figure 5.6. Absolute capacitance measurements of tablet samples with different coating thickess using a parallel plate sensor.

Figure 5.7. Capacitance variation between samples with different coating thickenss using a parallel plate sensor. The absolute capacitance measurements of the original uncoated sample are used as references.

61

102 103 104 105-0.5

0

0.5

1

1.5

2

2.5

3

3.5

Frequency (Hz)

Pha

se v

aria

tion

(deg

)

27 mg30 mg33 mg36 mgfinal

Figure 5.8. Phase variation for tablet samples with different coating thickness using a parallel plate sensor. The absolute phase measurements of the original uncoated sample are used as references.

Figure 5.9. Capacitance variation against sample weight using a parallel plate sensor.

62

5.4.3 The experimental results – fringing field Figure 5.10 shows the capacitance measurements of the samples from the fringing field

setup. A much greater difference is witnessed in this case between the measurements of

the original tablets and those of the coated ones than in the case of the parallel plate

setup. This is easily explained by the higher sensitivity of FEF sensors to the layer of

samples in direct contact with the electrodes. Again, using the measurements from the

original tablets as references, the capacitance and phase variations of the coated tablets

are calculated. The results are shown in Figure 5.11 and Figure 5.12. Figure 5.13 shows

the capacitance variation data at 1 kHz plotted against sample weight. Compared with the

parallel plate result shown in Figure 5.9, the dependency between capacitance and sample

weight is non-linear in the fringing field setup. This is, however, within expectation

considering the non-uniform field distribution of a FEF sensor. As coating thickness

increases, the electrical measurement sensitivity to thickness variation decreases. To

attain an optimal sensitivity curve, a wavelength of the FEF sensor has to be carefully

chosen. The sensor used here provided 3 different channels with various wavelengths.

The spatial wavelength of the channel used in the experiments is 500 µm, which

corresponds to a penetration depth of around 160 µm.

63

102 103 104 1057.2

7.4

7.6

7.8

8

8.2

8.4

8.6

8.8

9

Frequency (Hz)

Cap

acita

nce

(pF)

Original27 mg30 mg33 mg36 mgfinal

Figure 5.10. Absolute capacitance measurements for tablet samples with different coating thickness using a fringing electric field sensor with spatial wavelengh of 500 µm.

102 103 104 1050.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

1.1

1.2

Frequency (Hz)

Cap

acita

nce

varia

tion

(pF) 27 mg

30 mg33 mg36 mgfinal

Figure 5.11. Capacitance variation for tablet samples with different coating thickness using a fringing electric field sensor with spatial wavelengh of 500 µm. The absolute capacitance measurements of the original uncoated sample are used as references.

64

102 103 104 1050

0.5

1

1.5

2

2.5

3

3.5

Frequency (Hz)

Pha

se v

aria

tion

(deg

) 27 mg30 mg33 mg36 mgfinal

Figure 5.12. Phase variation for tablet samples with different coating thickness using a fringing electric field sensor with spatial wavelengh of 500 µm. The absolute phase measurements of the original uncoated sample are used as references.

26 28 30 32 34 36 38 40 42

0.44

0.46

0.48

0.5

0.52

Weight (mg)

Cap

acita

nce

var

iatio

n (p

F)

Figure 5.13. Capacitance variation against sample weight for the fringing electric field setup. The absolute capacitance measurements of the original uncoated sample are used as references.

65

5.5. Acquiring drug signature using a FEF sensor There exist the need for a non-invasive sensing technique that can differentiate different

types of drugs. One of the solutions is to look at the spectroscopy meassurements of the

drug samples in frequeny domain. Information on both the absolute value and the trend of

the frequency dependencies of the electrical measurements can be used to different types

of drugs from each other.

This section investigates the feasibility of acquiring drug signatures using the

fringing field dielectrometry sensing technique. Figure 5.14 shows the experimental

setup, where eight tablet samples are positioned on a FEF sensor with spatial wavelength

of 500 µm. The goal of the experiment is to differentiate between three groups of tablet

samples: the original, the polished, and the placebos.

Figure 5.14. Fringing electric field sensor setup for acquiring drug signature. The spatial wavelength of the sensor is 500 µm.

Figure 5.15 and Figure 5.16 show respectively the capacitance and phase

measurements of the three types of tablets. A great difference is witnessed between the

measurements of the original tablets and those of the other two types of tablets, which

makes the original tablets easily distinguishable. The measurement results for the

placebos are close to those of the polished samples, the difference in capacitance

measurements being in the range of 0.01 pF. High measurement resolution is necessary

to differentiate these two groups of samples.

66

102 103 104 1056.5

6.6

6.7

6.8

6.9

7

7.1

7.2

7.3

Frequency (Hz)

Cap

acita

nce

(pF)

OriginalPolishedPlacebo

Figure 5.15. Capacitance measurements of three different types of tablet samples using a fringing electric field sensor with spatial wavelength of 500 µm.

102 103 104 105-90.5

-90

-89.5

-89

-88.5

-88

-87.5

Frequency (Hz)

Phas

e (d

eg)

OriginalPolishedPlacebo

Figure 5.16. Phase measurements of three different types of tablet samples using a fringing electric field sensor with spatial wavelength of 500 µm.

67

5.6. Choice of sensors – FEF vs. parallel plate One of the major differences between FEF and parallel plate sensors is their different

levels of sensitivity to surface contact disturbances. Parallel plate sensors are generally

much more robust to surface contact disturbances than FEF sensors, and they should be

used when the effect of the parameters of interest is very subtle and the surface contact

uncertainties need to be reduced.

To compare the measurement performance of the FEF and parallel plate setup, the

capacitance measurements from section 5.4 are normalized. Normalization is carried out

by dividing all tablet measurements with the sensor capacitance measurements in air. The

resulting measurements are shown in Figure 5.17 and Figure 5.18. The parallel plate

setup provides greater overall measurement sensitivity to coating thickness variations.

The parallel plate setup also has better linearity, as illustrated by Figure 5.7 and Figure

5.9. Therefore for this application, the parallel plate setup is preferable to the fringing

field setup.

68

102 103 104 1052.1

2.15

2.2

2.25

2.3

2.35

2.4

2.45

2.5

2.55

2.6

Frequency (Hz)

Nor

mal

ized

cap

acita

nce

Original27 mg30 mg33 mg36 mgFinal

Figure 5.17. Normalized capacitance of tablet samples with different coating thickness using the parallel plate setup. Normalization is carried out by dividing measurements for the tablets by sensor measurements in air.

69

102 103 104 1051.15

1.2

1.25

1.3

1.35

1.4

Frequency (Hz)

Nor

mal

ized

cap

acita

nce

Original27 mg30 mg33 mg36 mgFinal

Figure 5.18. Normalized capacitance of tablet samples with different coating thickness using the fringing field setup. Normalization is carried out by dividing tablet measurements by sensor measurements in air.

Sometimes, however, the high sensitivity of FEF sensors to surface contact quality

can also be used to our advantage. For example, in section 5.5, experiments are

conducted to distinguish between unpolished and polished drug samples. Due to FEF

sensors higher sensitivity to surface contact qualities, a much more pronounced

difference is witnessed in the measurements from FEF sensors.

5.7. Measuring API concentration for powder samples This section investigates the feasibility of non-invasive monitoring of the drying process

for pharmaceutical powder samples. The electrical impedance of powder samples that

have been subject to different period of drying time are measured using the impedance

spectroscopy sensing system.

Figure 5.19 and Figure 5.20 show the capacitance and phase measurement of the

various powder samples. It can be inferred from the big difference between the

measurements of ‘0 hour’ samples and ‘2 hours’ samples that most API is removed in the

first two hours of the process. Figure 5.21 shows the capacitance measurements of the

powder samples plotted against drying time at two different frequencies. The

70

exponentially decaying profile resembles that of a diffusion process, which matches with

what is expected of the drying process. This result indicates the feasibility of the

technique. More extensive experiments have to be conducted to quantify the functional

dependence between the electrical measurements and API content.

Figure 5.19. Capacitance measurements of powder samples of various drying time.

71

Figure 5.20. Phase measurements of powder samples of various drying time.

Figure 5.21. Capacitance measurements of powder samples against sample drying time.

5.8. Conclusion and future work The investigation in the application of dielectrometry sensing to pharmaceutical products

included here in the thesis is still at a preliminary stage. Measurements show good

sensitivity to the parameters of interest. Sensor designs should be optimized toward the

72

particular applications to ensure the best performance. This involves such issues as

whether to use parallel plate or fringing field setup, and what geometry to use for the

electrodes.

The results presented here are obtained from carefully arranged experimental setups,

which are hard to come by if real time sensing on the production line is desired. More

practical experimental setup needs to be designed to simulate the manufacturing process,

and the effects of disturbance factors need to be investigated.

Extensive experiments need to be conducted to evaluate the repeatability of the

results. Based on these results, statistical analysis can be carried out to accurately

quantify the functional dependencies between the electrical measurements and the

parameters of interest. The functional dependencies, in turn, can be used to calibrate the

measurement system.

73

Chapter 6. Conclusions and future work

6.1. Conclusions This thesis deals with the topic of measuring material physical properties using

impedance spectroscopy. The scope of this work involves sensor design, circuit design,

software development, experiments, and data analysis. Two applications (1) moisture

sensing for cookie dough and (2) physical property sensing for pharmaceutical products

are included in the thesis. For the cookie dough application, measurement results proved

the feasibility of integration into the cookie manufacturing control process. A prototype

system is now being tested at the manufacturing site of Kraft foods. The investigation in

the pharmaceutical application is at a preliminary stage. Good measurement sensitivity

against parameters of interest is achieved. Further study needs to be carried out to study

the effect of disturbance factors, such as surface contact quality.

6.2. Directions of future work

6.2.1 Information decoupling for multivariable experiments In the case of many applications, for example, the heating process of cookie dough, more

than one material physical parameter is varying. It is necessary to measure the physical

properties other than the ones of interest and compensate for their effects. The current

experimental setup allows for simultaneous sensing of sensor arrays formed by different

types of sensor. Temperature, mass, as well as electrical impedance data can be collected

at the same time. Complex multi-variable experiments need to be carried out to help

decouple the information about different physical variables.

6.2.2 More sophisticated parameter estimation algorithms A linear dependence was assumed for the applications involved in this study. While this

model might be sufficient for the accuracy requirement of these particular applications, a

more sophisticated algorithm needs to be investigated for the generalized case of

dielectromoetry sensing applications. The major challenge lies in finding an efficient and

effective solution to the inverse problem.

74

6.2.3 Statistical evaluation of experimental results More extensive experiments need to be conducted to collect a database for the

measurement system. Statistical analysis should be carried out to test the repeatability of

the experimental results and to fully and accurately calibrate the measurement system for

each particular application.

75

End notes [1] S. O. Nelson, S. Trabelsi, and A. W. Kraszewski, "RF Sensing of Grain and Seed

Moisture Content," IEEE Sensors Journal, vol. 1, no. 2, pp. 119-126, Aug. 2001.

[2] P. M. Johnson, D. V. Thiel, and D. A. James, "Contributions to the Measured

Capacitance by the Dielectric Properties of Water in Insulated Electrode Soil

Moisture Sensors," Sensors Proceedings of IEEE, vol. 1, 2002, pp. 495-498.

[3] S. Simula, S. Ikalailen, K. Niskanen, T. Varpula, H. Seppa, and A. Paukku,

"Measurement of the Dielectric Properties of Paper," Imaging Science and

Technology, vol. 43, no. 5, pp. 472-477, 1999.

[4] A. V. Mamishev, Y. Du, B. C. Lesieutre, and M. Zahn, "Measurement of Moisture

Spatial Profiles in Transformer Pressboard," IEEE Conference on Electrical

Insulation and Dielectric Phenomena, 1998, pp. 323-326.

[5] K. Asami, T. Yonezawa, H. Wakamatsu, and N. Koyanagi, "Dielectric

Spectroscopy of Biological Cells," Bioelectrochemistry and Bioenergetics, vol. 40,

no. 2, pp. 141-145, Aug. 1996.

[6] H. Beving and G. Eriksson, "Dielectric-Spectroscopy of Human Blood," European

Journal of Surgery, pp. 87-89, 1994.

[7] D. D. Denton, J. B. Camou, and S. D. Senturia, "Effects of Moisture Uptake on the

Dielectric Permittivity of Polyimide Films," Proceedings of the 1985 International

Symposium on Moisture and Humidity, 1985, pp. 505-513.

[8] A. V. Mamishev, Y. Du, B. C. Lesieutre, and M. Zahn, "Development and

Applications of Fringing Electric Field Sensors and Parameter Estimation

Algorithms," Journal of Electrostatics, vol. 46, pp. 109-123, 1999.

76

[9] I. C. Shay and M. Zahn, "Cylindrincal Geometry Electroquasistatic Dielectrometry

Sensors," IEEE Conference on Electrical Insulation and Dielectric Phenomena,

2002, pp. 126-129.

[10] W. L. Kerr, R. J. Kauten, M. Ozilgen, M. J. McCarthy, and D. S. Reid, "NMR

Imaging, Calorimetric, and Mathematical Modeling Studies of Food Freezing,"

Journal of Food Process Engineering, vol. 19, no. 4, pp. 363-384, Nov. 1996.

[11] B. P. Hills, J. Godward, and K. M. Wright, "Fast Radial NMR Microimaging

Studies of Pasta Drying," Journal of Food Engineering, vol. 33, no. 3-4, pp. 321-

335, Aug. 1997.

[12] C. Simoneau, M. J. McCarthy, and J. B. German, "Magnetic-Resonance-Imaging

and Spectroscopy for Food Systems," Food Research International, vol. 26, no. 5,

pp. 387-398, 1993.

[13] M. Kalab, P. Allanwojtas, and S. S. Miller, "Microscopy and Other Imaging

Techniques in Food Structure-Analysis," Trends in Food Science & Technology,

vol. 6, no. 6, pp. 177-186, June 1995.

[14] M. J. McCarthy and K. L. McCarthy, "Applications of Magnetic Resonance

Imaging to Food Research," Magnetic Resonance Imaging, vol. 14, no. 7-8, pp.

799-802, 1996.

[15] S. J. Schmidt, X. Z. Sun, and J. B. Litchfield, "Applications of Magnetic Resonance

Imaging in Food Science," Critical Reviews in Food Science and Nutrition, vol. 36,

no. 4, pp. 357-385, 1996.

[16] M. Riva and S. Liviero, "Image Analysis Approach to Characterise the Bread

Cooking Kinetic," Industrie Alimentari, vol. 39, no. 395, pp. 953-960, Sept. 2000.

[17] D. D. Shepard and K. R. Smith, "Ultrasonic Cure Monitoring of Advanced

Composites," Sensor Review, vol. 19, no. 3, pp. 187-191, 1999.

77

[18] L. C. Haynes and J. P. Locke, "Microwave Permittivities Of Cracker Dough, Starch

and Gluten," Journal of Microwave Power and Electromagnetic Energy, vol. 30,

no. 2, pp. 124-131, 1995.

[19] R. Neimanis, H. Lennholm, and R. Eriksson, "Determination Of Moisture Content

in Impregnated Paper Using Near Infrared Spectroscopy," Electrical Insulation And

Dielectric Phenomena, 1999, pp. 162-165.

[20] Y. R. Kim, M. T. Morgan, M. R. Okos, and R. L. Stroshine, "Measurement And

Prediction Of Dielectric Properties of Biscuit Dough At 27MHz," Journal of

Microwave Power and Electromagnetic Energy, vol. 33, no. 3, pp. 184-194, 1998.

[21] M. Craig and S. Tamburic, "Dielectric Analysis of Bioadhesive Gel Systems,"

European Journal of pharmaceutics and biopharmaceutics, vol. 44, no. 1, pp. 61-

70, 1997.

[22] M. Craig, "Dielectric Spectroscopy As a Novel Analytical Technique Within the

Pharmaceutical Sciences," STP-Pharma-Pratiques, vol. 5, no. 6, pp. 421-42, 1995.

[23] G. W. Bak, K. H. Bodek, B. Hilczer, and T. Pawlowski, "Thermal Aging

Phenomena in Chitosan-Related Pharmaceutical Systems," IEEE Transactions on

Dielectrics and Electrical Insulation, vol. 8, no. 3, pp. 555-558, 2001.

[24] N. F. Sheppard, D. R. Day, H. L. Lee, and S. D. Senturia, "Microdielectrometry,"

Sensors and Actuators, vol. 2, no. 3, pp. 263-274, July 1982.

[25] V. J. Lumelsky, M. S. Shur, and S. Wagner, "Sensitive Skin," IEEE Sensors

Journal, vol. 1, pp. 41-51, Mar. 2001.

[26] A. K. Jonscher, Universal Relaxation Law, Chelsea Dielectrics Press London, 1996.

[27] A. R. von Hippel, Dielectrics and Waves, John Wiley & Sons, 1954.

78

[28] A. K. Jonscher, Dielectric Relaxation in Solids, Chelsea Dielectrics Press London,

1983.

[29] w. R. Westphal, "Permittivity, Distributed Circuits," in von Hippel, A. (ed.)

Dielectric materials and applications Cambridge: The M.I.T. press, 1961, pp. 63-

122.

[30] A. V. Mamishev, A. R. Takahashi, Y. Du, B. C. Lesieutre, and M. Zahn,

"Parameter Estimation in Dielectrometry Measurements," Journal of Electrostatics,

vol. 56, pp. 465-492, 2002.

[31] H. Schlichtling, "Boundary-Layer Theory," 1974,

[32] F. Bellucci, I. Khamis, S. D. Senturia, and R. M. Latanision, "Moisture Effects on

the Electrical Conductivity of Kapton Polyimide," Journal of the Electrochemical

Society, vol. 137, no. 6, pp. 1778-1784, 1990.

[33] J. Melcher, Y. Daben, and G. Arlt, "Dielectric Effects of Moisture in Polyimide,"

IEEE Transactions on Electrical Insulation, vol. 24, no. 1, pp. 31-38, Feb. 1989.

[34] A. von Hippel, Dielectric Materials and Applications, Artech House, 1995.

[35] J. G. Webster, "Electrodes," in Webster, J. G. (ed.) Electrical Impedance

Tomography Bristol and New York: Adam Hilger, 1990, pp. 21-28.

[36] B. S. Mohamed, R. Z. Morawski, A. W. Kraszewski, A. Brawicz, and S. O. Nelson,

"Calibration of a Microwave System for Measuring Grain Moisture Content," IEEE

Transactions on Instrumentation and Measurements, vol. 48, no. 3, pp. 778-783,

June 1999.

[37] A. K. Vijh, "Electrochemical Nature of Metal-Insulator Interfaces," IEEE

International Symposium on Electrical Insulation, 1996, pp. 870-873.

79

[38] A. V. Mamishev, B. C. Lesieutre, and M. Zahn, "Optimization of Multi-

Wavelength Interdigital Dielectrometry Instrumentation and Algorithms," IEEE

Transactions on Dielectrics and Electrical Insulation, pp. 408-420, 1998.

80

References [1] S. O. Nelson, S. Trabelsi, and A. W. Kraszewski, "RF Sensing of Grain and Seed

Moisture Content," IEEE Sensors Journal, vol. 1, no. 2, pp. 119-126, Aug. 2001.

[2] P. M. Johnson, D. V. Thiel, and D. A. James, "Contributions to the Measured

Capacitance by the Dielectric Properties of Water in Insulated Electrode Soil

Moisture Sensors," Sensors Proceedings of IEEE, vol. 1, 2002, pp. 495-498.

[3] S. Simula, S. Ikalailen, K. Niskanen, T. Varpula, H. Seppa, and A. Paukku,

"Measurement of the Dielectric Properties of Paper," Imaging Science and

Technology, vol. 43, no. 5, pp. 472-477, 1999.

[4] A. V. Mamishev, Y. Du, B. C. Lesieutre, and M. Zahn, "Measurement of Moisture

Spatial Profiles in Transformer Pressboard," IEEE Conference on Electrical

Insulation and Dielectric Phenomena, 1998, pp. 323-326.

[5] K. Asami, T. Yonezawa, H. Wakamatsu, and N. Koyanagi, "Dielectric

Spectroscopy of Biological Cells," Bioelectrochemistry and Bioenergetics, vol. 40,

no. 2, pp. 141-145, Aug. 1996.

[6] H. Beving and G. Eriksson, "Dielectric-Spectroscopy of Human Blood," European

Journal of Surgery, pp. 87-89, 1994.

[7] D. D. Denton, J. B. Camou, and S. D. Senturia, "Effects of Moisture Uptake on the

Dielectric Permittivity of Polyimide Films," Proceedings of the 1985 International

Symposium on Moisture and Humidity, 1985, pp. 505-513.

[8] A. V. Mamishev, Y. Du, B. C. Lesieutre, and M. Zahn, "Development and

Applications of Fringing Electric Field Sensors and Parameter Estimation

Algorithms," Journal of Electrostatics, vol. 46, pp. 109-123, 1999.

81

[9] I. C. Shay and M. Zahn, "Cylindrincal Geometry Electroquasistatic Dielectrometry

Sensors," IEEE Conference on Electrical Insulation and Dielectric Phenomena,

2002, pp. 126-129.

[10] W. L. Kerr, R. J. Kauten, M. Ozilgen, M. J. McCarthy, and D. S. Reid, "NMR

Imaging, Calorimetric, and Mathematical Modeling Studies of Food Freezing,"

Journal of Food Process Engineering, vol. 19, no. 4, pp. 363-384, Nov. 1996.

[11] B. P. Hills, J. Godward, and K. M. Wright, "Fast Radial NMR Microimaging

Studies of Pasta Drying," Journal of Food Engineering, vol. 33, no. 3-4, pp. 321-

335, Aug. 1997.

[12] C. Simoneau, M. J. McCarthy, and J. B. German, "Magnetic-Resonance-Imaging

and Spectroscopy for Food Systems," Food Research International, vol. 26, no. 5,

pp. 387-398, 1993.

[13] M. Kalab, P. Allanwojtas, and S. S. Miller, "Microscopy and Other Imaging

Techniques in Food Structure-Analysis," Trends in Food Science & Technology,

vol. 6, no. 6, pp. 177-186, June 1995.

[14] M. J. McCarthy and K. L. McCarthy, "Applications of Magnetic Resonance

Imaging to Food Research," Magnetic Resonance Imaging, vol. 14, no. 7-8, pp.

799-802, 1996.

[15] S. J. Schmidt, X. Z. Sun, and J. B. Litchfield, "Applications of Magnetic Resonance

Imaging in Food Science," Critical Reviews in Food Science and Nutrition, vol. 36,

no. 4, pp. 357-385, 1996.

[16] M. Riva and S. Liviero, "Image Analysis Approach to Characterise the Bread

Cooking Kinetic," Industrie Alimentari, vol. 39, no. 395, pp. 953-960, Sept. 2000.

[17] D. D. Shepard and K. R. Smith, "Ultrasonic Cure Monitoring of Advanced

Composites," Sensor Review, vol. 19, no. 3, pp. 187-191, 1999.

82

[18] L. C. Haynes and J. P. Locke, "Microwave Permittivities Of Cracker Dough, Starch

and Gluten," Journal of Microwave Power and Electromagnetic Energy, vol. 30,

no. 2, pp. 124-131, 1995.

[19] R. Neimanis, H. Lennholm, and R. Eriksson, "Determination Of Moisture Content

in Impregnated Paper Using Near Infrared Spectroscopy," Electrical Insulation And

Dielectric Phenomena, 1999, pp. 162-165.

[20] Y. R. Kim, M. T. Morgan, M. R. Okos, and R. L. Stroshine, "Measurement And

Prediction Of Dielectric Properties of Biscuit Dough At 27MHz," Journal of

Microwave Power and Electromagnetic Energy, vol. 33, no. 3, pp. 184-194, 1998.

[21] M. Craig and S. Tamburic, "Dielectric Analysis of Bioadhesive Gel Systems,"

European Journal of pharmaceutics and biopharmaceutics, vol. 44, no. 1, pp. 61-

70, 1997.

[22] M. Craig, "Dielectric Spectroscopy As a Novel Analytical Technique Within the

Pharmaceutical Sciences," STP-Pharma-Pratiques, vol. 5, no. 6, pp. 421-42, 1995.

[23] G. W. Bak, K. H. Bodek, B. Hilczer, and T. Pawlowski, "Thermal Aging

Phenomena in Chitosan-Related Pharmaceutical Systems," IEEE Transactions on

Dielectrics and Electrical Insulation, vol. 8, no. 3, pp. 555-558, 2001.

[24] N. F. Sheppard, D. R. Day, H. L. Lee, and S. D. Senturia, "Microdielectrometry,"

Sensors and Actuators, vol. 2, no. 3, pp. 263-274, July 1982.

[25] V. J. Lumelsky, M. S. Shur, and S. Wagner, "Sensitive Skin," IEEE Sensors

Journal, vol. 1, pp. 41-51, Mar. 2001.

[26] A. K. Jonscher, Universal Relaxation Law, Chelsea Dielectrics Press London, 1996.

[27] A. R. von Hippel, Dielectrics and Waves, John Wiley & Sons, 1954.

83

[28] A. K. Jonscher, Dielectric Relaxation in Solids, Chelsea Dielectrics Press London,

1983.

[29] w. R. Westphal, "Permittivity, Distributed Circuits," in von Hippel, A. (ed.)

Dielectric materials and applications Cambridge: The M.I.T. press, 1961, pp. 63-

122.

[30] A. V. Mamishev, A. R. Takahashi, Y. Du, B. C. Lesieutre, and M. Zahn,

"Parameter Estimation in Dielectrometry Measurements," Journal of Electrostatics,

vol. 56, pp. 465-492, 2002.

[31] H. Schlichtling, "Boundary-Layer Theory," 1974,

[32] F. Bellucci, I. Khamis, S. D. Senturia, and R. M. Latanision, "Moisture Effects on

the Electrical Conductivity of Kapton Polyimide," Journal of the Electrochemical

Society, vol. 137, no. 6, pp. 1778-1784, 1990.

[33] J. Melcher, Y. Daben, and G. Arlt, "Dielectric Effects of Moisture in Polyimide,"

IEEE Transactions on Electrical Insulation, vol. 24, no. 1, pp. 31-38, Feb. 1989.

[34] A. von Hippel, Dielectric Materials and Applications, Artech House, 1995.

[35] J. G. Webster, "Electrodes," in Webster, J. G. (ed.) Electrical Impedance

Tomography Bristol and New York: Adam Hilger, 1990, pp. 21-28.

[36] B. S. Mohamed, R. Z. Morawski, A. W. Kraszewski, A. Brawicz, and S. O. Nelson,

"Calibration of a Microwave System for Measuring Grain Moisture Content," IEEE

Transactions on Instrumentation and Measurements, vol. 48, no. 3, pp. 778-783,

June 1999.

[37] A. K. Vijh, "Electrochemical Nature of Metal-Insulator Interfaces," IEEE

International Symposium on Electrical Insulation, 1996, pp. 870-873.

84

[38] A. V. Mamishev, B. C. Lesieutre, and M. Zahn, "Optimization of Multi-

Wavelength Interdigital Dielectrometry Instrumentation and Algorithms," IEEE

Transactions on Dielectrics and Electrical Insulation, pp. 408-420, 1998.

85

Appendix

1. DiSPEC hardware installation guide List of Instruments

The instruments listed below are necessary for the dielectric spectroscopy system:

• National Instruments PCI-GPIB, NI-488.2 with Cable • Tektronix AFG310 Arbitrary Function Generator • National Instruments PCI-6035E/PCI-6036E DAQ board • National Instruments BNC-2120 terminal block • National Instruments SH-68-68-EP shielded cable • Tektronix PS280 Triple Output Power Supply

The following is a complete list of the items provided by SEAL:

• 3-channel sensor-interface circuit • Two 2-channel fringing field electric sensors with connectors attached • A K-type thermocouple • Five SMA male – BNC male cables • 5-pin male – 3 Banana plug power cable for sensor-interface circuit • 5-pin male – single wire for relay-control signal connection • Two short stripped wires for power supply lead-to-lead connection • Kapton covers for sensor heads • DiSPEC custom software CD • A user’s manual

Installing the Hardware Step 1: Install the GPIB and the DAQ Cards

86

Plug the NI GPIB and the NI-DAQ boards into the experimental computer and install

all recommended drivers.

Step 2: Connect BNC-2120 to the DAQ Board

Connect BNC-2120 breakout box to the DAQ card in the computer using the shielded

SH-68-68-EP cable.

Step 3: Connect Thermocouple to BNC-2120

Plug thermocouple into the thermocouple socket of the BNC-2120 breakout box.

Check that the row switches for ACH0 and ACH1 are set to the right, the “Temp.Ref.” and “Thermocouple” positions.

87

Step 4: Connect the Sensor Box to BNC-2120 Attach the BNC-ends of four BNC-SMA cables to BNC-2120’s channels ACH2-

ACH5. The switches under all ACH connectors should be in GS position.

Switch ACH3 selector (above the thermocouple connector) to BNC position.

Using the BNC-SMA cables to connect the ADC0 to ADC3 terminals on the sensor interface box to the ACH2 to ACH5 of BNC-2120.

88

Place the sensor interface box and the sensor as far away as possible from the

computer, the monitor and other electronic devices to minimize noise interference.

Step 5: Connect the Relay Control Lead to BNC-2120 Connect the relay-control wire from the “Cntrl.” socket of sensor interface box to

DIO0 of BNC-2120 using a 5-pin to single wire connector.

89

To connect the wire to BNC-2120 DIO0 port, untighten a screw (on the right), let the wire end into the hole and re-tighten the screw.

Step 6: Set up the Power Supply

To properly set the polarity of the power supply, two short wires should be

connecting the outputs as shown in the picture below:

Set the power supply to independent mode by adjusting the two buttons in the middle

of the front panel. Set the display-mode switches to Voltage position. The current dial should be set to the twelve-o’clock position (halfway). Use voltage dials to set both outputs to 10 volts. DO NOT CHANGE THE VOLTAGE SETTINGS WHILE THE SENSOR INTERFACE IS CONNECTED TO THE POWER SUPPLY! Once the voltage is set, connect the interface box to the power supply using 5-pin to 3 banana plug cable.

The plug labeled GND can be connected to either of the two ground sockets available on the power supply. Connect the “+” plug to the “+” terminal of the right 0-20V supply and “-“ plug to the “-“ terminal of the left 0-20V supply.

90

Step 7: Connect the Function Generator Using the fifth BNC-SMA cable, connect the function generator to the sensor

interface box.

Use the GPIB cable to connect the function generator to the GPIB card in the computer. Note that you don’t need to change the settings of the function generator manually. The program DiSPEC will tell it what to do. AlWAYS TURN ON THE FUNCTION GENERATOR BEFORE OPENING the DiSPEC PROGRAM.

Step 8: Connect the Sensor to the Interface Box

91

The sensor is connected to the interface as follows: The cable labeled as “drive” of

the sensor is connected to any of the Drive terminals of the board. The ones labeled as “inner sense” and “outer sense” are connected to the S. 1 and S. 2 terminals on the sensor interface box respectively. 2. DiSPEC software guide Installing the Software This LABVIEW application is for viewing data in real time as well as recording data to a file. Start by opening the DiSPEC folder on the CD. Go to folder “Installer” and double click on the “setup” icon. Follow the setup guide step by step and the software will be installed on your computer. After installation, the program will show up in your computer’s start up menu as DiSPEC. Using the Software After all the hardware is installed, go to the start up menu of your computer and open the program DiSPEC. The following steps need to be followed to run the software.

1. Connect all the instruments according to the instructions in the hardware guide. 2. Open the NI software “Measure & Automation Explorer” (MAX). (The software

should be provided the DAQ board). Find out the device number for the function generator and the DAQ board. Specific instructions are available below under Device Number of the Function Generator and the DAQ Board.

3. Go to the system settings tab in the front panel and enter the device numbers found from step 1. The same values need to be manually entered each time the program is opened if they are found to be different from the default values.

4. Find out the time delay for each channel of the circuit board. For detailed instructions, go to Time Delay.

92

5. Enter the time delay values obtained from step 3 in the system settings tab. Note that if found to be different from the default values, these time delay constants have to be reentered each time the program is opened.

6. Go to the Controls tab, and configure the entries in the tab. For detailed instructions, go to Controls.

7. Turn on the function generator and the power supply. 8. Press the “Run” and the “Acquire” button. If an arrow appears on the top left

corner of the front panel, click on the arrow and the program will start running. If the arrow doesn’t appear, the code should already be running after the “Run” and the “Acquire” button is pressed down.

9. Go to the different tabs on the front panel to look at the data and the graphs. 10. To monitor change in sample capacitance more visually, go to the Real-time

Imaging tab. 11. To stop the program, press the “Acquire” button again.

Configure the System In the front panel, go to the “System Settings” tab, a window like the following will show up.

93

Back

Upon starting LABVIEW, the 'System settings' has all its entry values set to their defaults for the 2 channel fringing field sensor. The values should be the same as those shown in Error! Reference source not found.. These default values should be used unless specified otherwise below. Description of the entries can be obtained by right-clicking on an entry and selecting “Descriptions and Tips” on the menu.

Device Number of the Function Generator and the DAQ Board The device numbers of the function generator and the DAQ board are dependent on the particular setup, therefore values different from those specified as default may need to be used. To get the device numbers, open the NI software “Measurement & Automation Explorer” that is provided with the DAQ board. Double click on “Devices and Interfaces”, the DAQ board and the NI GPIB card will show up. The number listed for the DAQ board is the device number for the DAQ and the number listed for the GPIB card is the device number for the function generator.

Back

Channel Numbers on the DAQ board

94

This entry specifies the channel numbers on the DAQ board that the input and the three output signals of the circuit are connected to. The default values are “2,3,4,5”. Channels 0 and 1 on the DAQ board are saved for the thermocouple. The first channel number is the input to the DAQ board, and the following numbers are output channels, displayed in front channel as channels 1, 2, and 3 respectively. All numbers must be separated by commas. (Channel 3 is most often used in the parallel plate setup rather than the fringing field system)

Reference Capacitance The “Reference Capacitance” entries specify the reference capacitance used for each channel of the sensor circuit board. Note that these values are not the same as the values of the reference capacitors on the board. The effect of the stray capacitances (e.g. that introduced by the Op-Amp has to be accounted for.) The default values are obtained through careful calibration of the system. These values should be used unless some circuit elements are changed. Recalibration of the system is necessary if changes are made to the circuit.

Time Delay Multiplexing of the DAQ board introduces a time delay between the data stream from its different channels. The time delay causes significant phase distortion, therefore its effect has to be eliminated. Unfortunately, the time delay values are device dependent. A different computer and DAQ board will cause a change in these values, which means that these values have to be fine-tuned for each particular setup. The following procedure can be used to find the time delay constants:

1. Connect the input and all the outputs of the circuit board to the function generator. Note that since all channels are connected to the same source, ideally there should be no phase delay between the channels.

2. Run the program at the highest frequency (30 kHz). Ideally, the gain should be 1 and the phase should be zero for all channels. Adjust the time delay for each channel of the sensor in the “systems and setting” tab until the phase delays for all channels are 0.

3. The new time delay constants must be typed in manually each time the program is started.

Back

The Settings for the Thermocouple

95

Channel 0 and 1 of the DAQ board should always be used for the thermocouple as is specified in the default setting. The sampling rate for the thermocouple could be increased if an improvement in the speed of the program is desired. Otherwise, use the default value. Averaging is used here to remove noise. The number of samples for temperature averaging can be changed for different application. Controls The controls tab should look like the following window.

Back

Mass

96

Real-time monitoring of sample mass is not necessary at this point. This function is included for possible future applications. By pressing the “mass” button down, the program will acquire data from a scale that is connected to the computer through a serial port.

Temperature A K type thermocouple is connected to the computer. By pressing the “Temperature” button down, the system starts acquiring data from the thermocouple and saving the temperature data to the output file if “Saving to file” is also enabled.

Sweep When the “Sweep” button is not pressed, the system runs at a single frequency specified by the “Start Frequency” entry. The system performs frequency sweeps when the “Sweep” button is enabled. The range of the frequency sweep is defined by the start and the stop frequency. The minimum and maximum frequencies allowed by the current version of the program are 1 Hz and 30 kHz respectively. Measurements at frequencies lower than 1 Hz are comparably noisy, thus we limit the frequency to above 1 Hz. The frequency range can be easily extended for future applications.

Save to File If you wish to record the data to a file, the “Save to File” button should be pressed down. You can enter the file name and the saving directory in the “File Name” entry. If the entry is left blank, a file saving window will automatically pop up when the program starts running. The file can be saved as an Excel spreadsheet or a ‘.txt’ file.

Back

Real-time Imaging The following is a picture of the real-time imaging tab.

97

The tab provides a profile of the sample by displaying the capacitance of all channels of the sensor. Note that in a 2-channel fringing field setup, the information shown for channel 3 should be ignored. Channel 3 is included here mainly for the consideration that a 3-channel sensor might be used in the future. The capacitance values displayed in the vertical bars are all scaled to be within 0 to 1. The “Maximum Capacitance” knobs on the right refer to the actual capacitance value in “pF” when the bar displays a value of “1”. The knobs could be adjusted for the best visual effect.

Back