influência da técnica e material restaurador no comportamento biomecânico...
TRANSCRIPT
-
Universidade Estadual de Campinas
Faculdade de Odontologia de Piracicaba
RODRIGO BARROS ESTEVES LINS
Influência da técnica e material restaurador no comportamento
biomecânico em Restaurações Classe II
Influence of the restorative technique and material on the biomechanical
behavior of Class II restorations
Piracicaba
2017
-
RODRIGO BARROS ESTEVES LINS
Influência da técnica e material restaurador no comportamento
biomecânico em Restaurações Classe II
Influence of the restorative technique and material on the biomechanical
behavior of Class II restorations
Dissertação apresentada à Faculdade de
Odontologia de Piracicaba da Universidade
Estadual de Campinas como parte dos requisitos
exigidos para a obtenção do título de Mestre em
Clínica Odontológica, Área de Dentística.
Dissertation presented to the Piracicaba Dental
School of the University of Campinas in partial
fulfillment of the requirements for the degree of
Master in Odontological Clinic, in Restorative
Dentistry area.
Orientador: Professor Doutor Luis Roberto Marcondes Martins
ESTE EXEMPLAR CORRESPONDE À VERSÃO FINAL
DA DISSERTAÇÃO DEFENDIDA PELO ALUNO
RODRIGO BARROS ESTEVES LINS E ORIENTADA
PELO PROF. DR. LUIS ROBERTO MARCONDES
MARTINS.
Piracicaba
2017
-
Dedicatória
À Deus e à Nossa Senhora,
Além de dedicar a minha dissertação à Deus e à Nossa Senhora, dedico todo o meu
mestrado, meus estudos, minha vida profissional e pessoal.
Aos meus pais,
Dedico este trabalho de obtenção do título de Mestre aos meus maiores Mestres: ao
meu pai (José Lins) e à minha mãe (Mariângela Lins).
Por terem me apoiado em todas as minhas decisões, incluindo sair de casa para
correr atrás de um futuro profissional que ainda almejo em conquistar. Vocês são meus maiores
exemplos para seguir em busca dos meus ideais, segundo à vontade de Deus.
Muito obrigado por todo apoio e carinho. Amo vocês!
-
Agradecimentos
Aos meus familiares,
Agradeço a todos os meus familiares pelo apoio e incentivo dados a mim nesta nova
fase de estudos e aprendizados, principalmente às minhas avós.
Em especial aos meus irmãos José Renato Lins e Rafael Lins. A saudade é imensa,
mas sei que vocês estão e estarão sempre torcendo por mim.
A minha querida sobrinha Isabelle Lins, que sinto tanta falta. Estar longe e não
poder acompanhar pessoalmente seu crescimento é extremamente penoso, mas ter o seu carinho
em todas as viagens de retorno à João Pessoa é mais do que reconfortante.
Aos meus amigos Pessoenses,
Aos tantos amigos que fiz durante a vida, incluindo amigos da época de escola
(Augusto Ygor, Paulo Victor e Suzane Souza) e universidade (Alana Dantas, Amanda Solano,
Evllon Sá, Larissa Fernandes, Maíra Ramalho, Mariana Figueiredo, Raissa Marçal, Priscilla
Leite e tantos outros) os quais mantemos contato por redes sociais, agradeço por fazerem parte
da minha vida.
Ao meu orientador Prof. Dr. Luís Roberto Marcondes Martins,
Obrigado por ter me escolhido na seleção de mestrado e por ter acreditado no meu
potencial para desenvolver trabalhos científicos juntos. Nestes dois anos foram tempos de
muitos aprendizados que servirão para a minha vida. Tê-lo como orientador é mais do que um
exemplo.
Aos Professores da Dentística,
A todos os professores da Dentística-FOP, por todos os ensinamentos e atenção
prestados nestes dois anos de convivência diária. Grandes mestres que nos dão muito orgulho
de suas competências profissionais.
-
Aos funcionários da FOP-UNICAMP
Não poderia deixar de agradecer aos funcionários da FOP-UNICAMP que nos
prestam tantos auxílios nas mais diversas funções, desde aos funcionários da secretaria às
senhoras da limpeza.
Aos amigos Fopianos/Pessoenses,
Aos amigos Pessoenses que estudam na FOP-UNICAMP (Bruno Mariz, Emerson
Tavares, Jaiza Araújo, Jossaria Sousa, Laíse Lima, Marina Moreno, Mayara Abreu, Renally
Wanderley) que tornam a FOP um pouco nordestina e que fazem os nossos dias de estudo mais
divertidos.
Aos amigos da FOP-UNICAMP
Aos grandes amigos que conquistei nestes dois anos de mestrado (Amanda
Bandeira, Camilla Fraga, Carolina Rangel, Filipe Martins, Heloisa Pantaroto, Jairo Cordeiro,
Manoelito Silva, Mariana Barbosa, Olívia Figueiredo, Rafaela Videira, Rahyza Freire, Renato
Machado, Thiago Bessa). Agradeço por todo o companheirismo e apoio nos momentos de
felicidade e principalmente naqueles momentos de preocupação ou estresse inevitáveis à esta
fase.
Aos amigos da Dentística/FOP-UNICAMP
Aos amigos da Dentística que dividimos diariamente conhecimentos científicos
(Renata Pereira, Bruna Guerra, Caroline Mathias, Mayara Zaghi e Marina Faria) e em especial
a Josué Pierote, Mariana Flor e Maicon Sebold pela grande amizade.
À Cristiane Yanikian
Um agradecimento especial à Cristiane Yanikian. Ela que me pôs para trabalhar no
laboratório no meu primeiro dia de mestrado, mas que soube como ninguém me ensinar e
orientar por todos os obstáculos enfrentados durante o mestrado. É sem dúvida, um exemplo
que pretendo seguir.
-
À Elis Lira e Louise Dornelas
À minha família Piracicabana; às minhas colegas de graduação e pós-graduação; às
minhas grandes amigas: Elis Lira e Louise Dornelas. Mais de 7 anos de amizade. Não sei
expressar o quanto vocês foram e são importantes para mim. Ter ingressado na pós-graduação
com vocês foi, certamente, o ‘combustível’ necessário para suprir a saudade de casa. Tenho
muito orgulho pelas profissionais que são e pelas conquistas que não cessam em acontecer.
Ao Professor Doutor João Carlos Ramos e à Professora Doutora Alexandra Vinagre,
Agradeço aos Professores da Faculdade de Medicina Dentária da Universidade de
Coimbra-Portugal, que me acolheram durante o meu intercâmbio na reta final do meu mestrado.
Exemplo de dedicação à odontologia, orientadores e mestres incríveis. Profissionais que
durante 4 meses me deram suporte e me concederam grandes conhecimentos. Obrigado por
todo o tempo e atenção depositados à mim.
À Universidade Estadual de Campinas
À Faculdade de Odontologia de Piracicaba
Ao Programa de Clínica Odontológica
À CAPES
-
RESUMO
O advento de materiais restauradores de baixa contração vem ganhando espaço no mercado e
despertando interesse de cirurgiões-dentistas devido as suas características técnicas, economia
de tempo clínico e por ser um material mais estável dentro da cavidade oral. O objetivo deste
estudo foi avaliar o comportamento biomecânico de restaurações extensas classe II mésio-
ocluso-distal (MOD) com técnica direta (compósitos convencional e de baixa contração)
avaliando a tensão de contração de polimerização, deflexão de cúspide, resistência à fratura e o
padrão de fratura. Sessenta terceiros molares humanos hígidos foram selecionados e
randomizados em quatro grupos: resina composta Z100 (Z100); Tetric N-Ceram Bulk Fill
(TNC); Filtek Bulk Fill (FBF); Aura Ultra Universal Restorative Material (ABF). Cavidades
Classe II MOD foram confeccionadas com dimensões padronizadas por meio de uma máquina
de preparo cavitário. As resinas compostas bulk fill foram inseridas na cavidade em um único
incremento de 4 mm e a resina composta convencional em três incrementos de forma oblíqua e
fotopolimerizadas por um LED de alta potência. Os sensores FBG foram inseridos na interface
adesiva a fim de avaliar a tensão de contração do material resinoso durante a fotopolimerização
(n=5). Além disso, extensômetros foram fixados nas bases das cúspides para medirem a
deformação em três momentos (n=10): durante o procedimento restaurador, sob carregamento
de compressão axial em máquina de ensaio universal até 100N e até gerar a fratura do conjunto
dente-restauração. O tipo de fratura foi classificado conforme o seu padrão analisado em
Microscopia Eletrônica de Varredura (MEV): I-fratura em resina composta; II-fratura em resina
composta e estrutura dental coronal; III-fratura em resina composta e estrutura dental cervical
com possibilidade de reparo periodontal; IV-fratura radicular sem reparo. Foi realizado teste de
homogeneidade dos dados (Shapiro-Wilk, p>0.05) para todos os grupos, seguido do teste
ANOVA um fator com teste post-hoc de Tukey para a avaliação dos sensores FBG, deformação
de cúspide e resistência à fratura. O padrão de fratura foi analisado pelo teste qui-quadrado. A
avaliação da tensão de contração pelos sensores de Bragg mostrou maiores médias para o grupo
Z100 diferindo estatisticamente de todos os outros grupos (p
-
fratura do TNC foi maior que os demais, porém, estatisticamente significante apenas com o
grupo Z100 (p
-
ABSTRACT
The appearance of low contraction restorative materials has been introduced and increasing
attention from dental surgeons due to technical features, such as chair time economy and stable
within the oral cavity. The purpose of this study was to evaluate the biomechanical behavior of
extensive Class II mesial-occlusal-distal (MOD) restorations with direct technique
(conventional and low-shrinkage resin composites) by using fiber Bragg grating sensors (FBG)
and extensometry assessing the polymerization contraction stress, cuspal deformation, fracture
resistance and fracture pattern. Sixty human caries-free third molars were selected and
distributed randomly into six groups: Z100 restorative material (Z100); Tetric N-Ceram Bulk
Fill composite (TNC); Filtek Bulk Fill composite (FBF); Aura Ultra Universal Restorative
Material (ABF). Class II cavities MOD was standardized in an abrasion standardizing
equipment. The bulk fill materials were inserted in bulk increment of 4 mm and the
conventional resin composite in three oblique ones by poly-wave LED light-curing unit. The
optical FBG sensors were fixed at adhesive interface due to evaluated the shrinkage stress of
resinous material during the fotopolymerization (n=5). Moreover, strain gauges were fixed on
cuspal base to measure the deformation at three times (n=10): during restorative procedure,
subjected to axial compressive in universal testing machine up to 100N and to occur the sample
fracture. The failure mode was rated according to their standard analyzed in Scanning Electron
Microscopy (SEM): I-fracture at resin composite; II-fracture at resin composite and coronal
tooth structure; III-fracture at resin composite and cervical tooth structure with possible
periodontal repair; IV-root fracture beyond repair. The statistical analysis was performed for
homogeneity distribution (Shapiro-Wilk, p>0.05), followed by parametric statistical tests and
one-way ANOVA with post-hoc Tukey’s test. Data on fracture mode were submitted to Chi-
square test. The optical FBG sensors evaluation showed that Z100 presented the highest means
of stress shrinkage (p
-
Keywords: Composite Resins. Fiber Optical Technology. Polymerization.
-
SUMÁRIO
1 INTRODUÇÃO 14
2 ARTIGO: Biomechanical Behavior of Class II Restoration Using Fiber
Bragg Grating Sensors and Extensometry
17
3 CONCLUSÃO 36
REFERÊNCIAS 37
APÊNDICE 1: Metodologia Ilustrada 42
ANEXO 1: Certificado - Comitê de Ética em Pesquisa 46
ANEXO 2: Comprovante de submissão do artigo 47
-
14
1 INTRODUÇÃO
A exigência por parte da população em adquirir estética do sorriso e saúde bucal
satisfatória tem influenciado no advento de materiais cada vez mais estéticos na clínica
odontológica. Somado a isso, o desenvolvimento tecnológico impulsiona a geração de novos
materiais restauradores e técnicas adesivas mais eficazes e duradoras (Pashley et al., 2011) com
propriedades físicas e ópticas semelhantes aos dos dentes naturais (Nahsan et al., 2012). A partir
desta constante inovação, podemos caracterizar o estado da arte dos materiais odontológicos
(Ferracane, 2011).
A tendência atual de realizar cada vez mais procedimentos preventivos na saúde
pública para impulsionar a diminuição na prevalência da doença cárie não garante ou não
impede a necessidade de substituição de restaurações insatisfatórias, que a partir deste ciclo
restaurador acarreta em perda da estrutura dental. Com isso, na clínica odontológica são
frequentemente realizadas restaurações em cavidades classe II MOD amplas (Deliperi e
Bardwell, 2002; Bonecker et al., 2013).
A resina composta, ao longo dos anos, ganhou grande destaque e aplicabilidade na
clínica odontológica para diversos procedimentos (Ferracane, 2011). Esta é comumente
utilizada em restaurações para dentes anteriores e posteriores, com excelente desempenho
clínico, comprovado em estudos longitudinais (Demarco et al., 2012; Cetin et al., 2013; Ozakar-
Ilday et al., 2013; Demarco et al., 2015; Van Dijken e Pallesen, 2016). Este sucesso clínico,
principalmente em restaurações diretas com compósitos microhíbridos (Da Rosa Rodolpho et
al., 2011; Demarco et al., 2015) estão relacionados às propriedades mecânicas favoráveis, como
adequada resistência à fratura (Ferracane, 2011). Entretanto, algumas desvantagens também são
observadas, como: a necessidade de inserção do material de forma incremental, a qual demanda
maior tempo clínico e incorporação de bolhas no corpo da restauração; alta contração de
polimerização e, consequente, geração de tensão, provocando deslocamento do compósito
resinoso a partir da interface adesiva, criando um ‘gap’ marginal, e portanto, sensibilidade pós-
operatória, microinfiltração, cárie secundária, pigmentação marginal e falha na restauração
(Casselli et al., 2013; Campos et al. 2014; Dominguez et al., 2014; Benetti et al., 2015; Fronza
et al., 2015).
A contração de polimerização ocorre ao longo do processo de conversão dos
monômeros em polímeros, através de forças covalentes fortes, promovendo diminuição da
distância molecular. Além disso, a tensão ocorre na transição da fase gel para um estado amorfo
-
15
isotrópico quando fotoativado (Lee et al., 2013), contribuindo para a formação de tensões ao
longo da interface adesiva das restaurações (Loguercio et al., 2004).
Algumas alterações nas resinas compostas têm sido propostas com o objetivo de
diminuir os efeitos indesejados da forte tensão gerada na interface dente/restauração, como a
composição química ou estrutural do monômero da matriz resinosa; a quantidade, forma e/ou
tratamento superficial da partícula de carga, além de modificação no sistema iniciador (El-
Damanhoury e Platt, 2014; Kwon et al., 2015; Orlowski et al., 2015; Atalay et al., 2016).
Recentemente, um grupo de resinas compostas de baixa contração foram
desenvolvidas e denominadas “bulk-fill”. Estas permitem a inserção de único incremento de até
4-5 mm de material (Czasch e Ilie, 2013); possuem a capacidade de gerar menor tensão de
contração de polimerização; reduzem a deflexão de cúspide (Moorthy et al., 2012); apresentam
menor tempo e alta reatividade à fotopolimerização, devido ao aumento da translucidez e,
portanto, permitem uma maior penetração de luz e profundidade de polimerização (Roggendorf
et al, 2011; Leprince et al, 2014; Orlowski et al., 2015); além disso, apresentam a vantagem em
diminuir o tempo de execução clínico (Campos et al, 2014).
Na literatura científica encontram-se diversos testes científicos destinados a
avaliarem o comportamento de materiais resinosos, como por exemplo: análise por elementos
finitos (Anatavara; Sitthiseripratip; Senawongse, 2016), tensômetro conectado a um sistema de
cantilever com sensores (Kalliecharan et al., 2016), máquinas de ensaio universal (Witzel et al.,
2005), sistema ‘ring slitting’ (Park e Ferracane, 2005; Park e Ferracane, 2006), sensores de rede
de Bragg em fibra óptica (Rajan et al., 2016; Vinagre et al., 2016; Umesh et al., 2016),
extensometria (Bicalho et al., 2014; Pereira et al., 2015), dentre outros.
O sensor de rede de Bragg em fibra óptica possui a característica de ser
fotossensível e a capacidade de verificar variações, como: temperatura, contração, expansão e
pressão a partir da alteração do comprimento de onda (Yeh et al., 2011; Umesh et al., 2016).
Esta tecnologia tem-se tornado uma das técnicas mais usadas em sistemas de avaliação
biomecânica e reabilitação em diversas áreas científicas, como: automotiva, aeronáutica,
medicina, odontologia e especificamente a biomecânica dental, incluindo estudos in vitro e in
vivo (Al-Fakih; Abu Osman; Mahamd Adikan, 2012; Umesh et al., 2016).
Um método adicional para compreender o comportamento biomecânico de um
material resinoso a partir da alteração da deflexão de cúspides é por meio da extensometria, a
qual é um teste mecânico não destrutivo que avalia diretamente as características do processo
-
16
restaurador, ou seja, detecta alterações do comportamento da estrutura dental resultantes das
forças atuantes na camada adesiva, que podem ser influenciadas pelo material restaurador,
sistema adesivo, técnica restauradora a ser adotada e o tamanho da cavidade (Bicalho et al.,
2014; Pereira et al., 2016).
Encontram-se na literatura científica estudos sobre resinas compostas ‘bulk fill’,
contudo, o conhecimento sobre o comportamento biomecânico na interface adesiva das mesmas
ainda se apresenta escarça. Pensando nisso, o objetivo do presente estudo foi avaliar o
comportamento biomecânico de restaurações extensas classe II mésio-ocluso-distal (MOD)
com técnica direta (compósitos convencional e de baixa contração) avaliando a tensão de
contração de polimerização, deflexão de cúspide, resistência à fratura e o padrão de fratura.
-
17
2 ARTIGO:
Biomechanical Behavior of Class II Restoration Using Fiber Bragg Grating Sensors and
Extensometry
Rodrigo Barros Esteves Linsa,*, Cristiane Rumi Fujiwara Yanikiana, Thiago Henrique
Scarabello Stapea, Aline Arêdes Bicalhob, Carlos José Soaresb, Luís Roberto Marcondes
Martinsa
a Department of Restorative Dentistry, Piracicaba Dental School, State University of Campinas,
Av. Limeira, 901, Piracicaba 13414-903, SP, Brazil
b Department of Operative Dentistry and Dental Materials, Dental School, UFU – Federal
University of Uberlândia, Av. Pará, 1720, Campus Umuarama, Uberlândia 38400-902, MG,
Brazil
*Corresponding author. Tel.: +55 83 988527948
E-mail addresses: [email protected] (R.B.E.Lins), [email protected]
(R.R.F.Yanikian), [email protected] (T.H.S.Stape), [email protected]
(A.A.Bicalho), [email protected] (C.J.Soares), [email protected] (L.R.M.Martins)
Funding sources
The CAPES Foundation of Department of Education, Brazil, supported this work.
ABSTRACT
Objective. The purpose of this study was to assess the biomechanical behavior of extensive
Class II mesial-occlusal-distal (MOD) restorations performed direct technique (conventional
and low-shrinkage resin composites) by using fiber Bragg grating sensors (FBG) and research
the polymerization contraction stress as well as evaluate the low-shrinkage resin composite
regarding cuspal deformation and fracture resistance and pattern.
Methods. Sixty human caries-free third molars were selected and distributed randomly into four
groups: Z100 restorative material (Z100); Tetric N-Ceram Bulk Fill composite (TNC); Filtek
Bulk Fill composite (FBF) and Aura Ultra Universal restorative material (ABF). In all samples
were standardized class II MOD cavities. The bulk fill materials were inserted in bulk increment
-
18
and the conventional resin composite in three ones. Shrinkage stress was evaluated with optical
FBG sensors (n=5). The cuspal deformation with extensometer was measured during
restoration, compressive axial force and until the fracture on universal machine (n=10). The
fracture pattern was analyzed with Scanning Electron Microscopy (SEM). The statistical
analysis was performed for homogeneity distribution (Shapiro-Wilk, p>0.05), followed by
parametric statistical tests and one-way ANOVA with post-hoc Tukey’s test. Data on fracture
mode were submitted to Chi-square test.
Results. The optical FBG sensors evaluation showed that Z100 presented the highest means of
stress shrinkage (p
-
19
structure. This results in extensive Class II MOD cavity restorations, which are regularly
performed in clinical odontology [7,8].
The clinical success and longevity of direct restorations in posterior teeth by using
micro-hybrid composite resins are related to favorable mechanical properties [2,9], such as
adequate fracture resistance [10]. However, some drawbacks are found: incremental insertion
technique, which demands longer clinic procedures; contraction due to polymerization and
subsequent stress, with risk of dislodging of the resinous composite from the adhesive interface,
causing marginal gap and post-procedure sensitivity; micro-infiltration; secondary caries;
marginal pigmentation; and restoration failure [11-13]. This contraction might occur throughout
the process of polymerization as a result of strong covalent forces, causing abridgement of
molecular distances and tension in the passage from gel to isotropic amorphous state [14], thus
contributing to the formation of stress at the adhesive interface of the restorations [15].
In order to diminish the undesired effects of the composites, such as the tension
created on the tooth/restoration interface, some chemical and structural changes in the
composite resin composition have been proposed. This include changes in the resinous matrix,
quantity, and shape and/or surface treatment of the inorganic particle [1]. Another group of low-
shrinkage composite resins have been recently developed, the bulk-fill resins. They allow the
insertion of a single increment with 4-5-mm [16] due to their capacity of generating less
contraction stress and high reactivity to photopolymerization as result of increased
translucency, improving light penetration and depth of polymerization [17-19]. The incremental
restoration technique, which uses micro-hybrid conventional composite resin inserted into the
cavity in small increments [20], is still the most adequate photopolymerization to modify the
contraction pattern in direct restorations [21]. However, this technique has disadvantages such
as lodging of air bubbles during incremental insertion and longer chair time [22], whereas bulk-
fill composite resins have the advantages of controlling polymerization shrinkage stress [23],
shortening chair time [22] and reducing cuspal deflection [24].
The scientific literature is comprehensive about the study of bulk fill resin
composite, however the biomechanical behavior knowledge in adhesive interface is scarce,
wherefore, the purpose of this study was to evaluate the biomechanical behavior of extensive
Class II mesial-occlusal-distal (MOD) restorations with direct technique (conventional and
low-shrinkage resin composites) by using fiber Bragg grating sensors (FBG) and extensometry
forma the polymerization contraction stress, cuspal deformation, fracture resistance and
fracture pattern. The hypothesis tested here was: bulk-fill composite resins generate less
-
20
polymerization contraction stress and reduce cuspal deformation compared to conventional
composite resins.
2 MATERIALS & METHODS
Four commercial resin-based composites were investigated: one conventional
incrementally layered composite (Z100 restorative material) (positive control); three bulk-fill
composites (Tetric N-Ceram Bulk Fill, Filtek Bulk Fill and Aura Ultra Universal restorative
materials). Product specifications are listed in Table 1.
All the restorative materials were light-cured by using a poly-wave LED light-
curing unit (VALO, Ultradent Products Inc., South Jordan, UT, USA) operating with
wavelength of 350 and 550 nm and irradiance at 995 ± 2mW/cm2 according to the
manufacturer’s recommendations as follows: 40 seconds for each layer of conventional
composite (three oblique increments), 20 seconds for one bulk layer of bulk-fill composites (10
seconds for mesio-occlusal and 10 for distal-occlusal).
2.1 Specimen Preparation
Sixty human caries-free third molars were extracted and stored in buffered aqueous
solution of 0.2% sodium azide at 4 ⁰C for up to 6 months. The research project was previously
approved by the Research Ethical Committee of Piracicaba Dental School, State University of
Campinas, Piracicaba, Brazil (protocol 127/2014).
The teeth were cleaned by using periodontal curettes to remove organic and
inorganic remnants, including pumice paste, water and Robson brush at low rotation before
storage in distilled water at 4°C. The dimensions of the teeth were determined by using a digital
calliper (Mitutoyo, Tokyo, Japan) to measure the mesio-distal (MD) and lingual-buccal (LB)
dimensions of the occlusal surface. The occlusal surface area was determined by considering
the molar as a rectangle with the sides determined by the MD and LB dimensions. Teeth
selected presented occlusal surface area varying no more than 10% of the sample average.
2.1.1 Inclusion and Simulation of Periodontal Ligament
-
21
The teeth were inserted into polystyrene resin to simulate the periodontal ligament.
A 0.5 mm graphite pencil was used to mark the insertion site at 2-mm below the amelocemental
junction, thus delineating a radicular portion to be coated with #7 molten paraffin wax. The
teeth were fixed with sticky wax from the crown to the stem of a prosthetic liner. A radiographic
film with central perforation compatible with the dimension of each tooth was positioned at the
same level of the sticky wax. The mobile table of the prosthetic liner was adjusted
perpendicularly to the long axis and on the top of the teeth. Next, a PVC cylinder with a 10-
mm wide central perforation was attached to the radiographic film with sticky wax. The set was
taken to a wooden plate with individual perforations for each tooth. The self-curing polystyrene
resin was prepared and poured into the interior of the PVC cylinder. After polymerization, the
set was removed from the rack.
The teeth were removed from the artificial alveoli and cleaned with jets of baking
soda and water, and the resin cylinders were polished by using a universal sanding machine
(AROPOL-E Arotec, Cotia/São Paulo, Brazil) to eliminate the excesses. The polyether-based
casting material (Polyether Impregum Soft 3M ESPE, St. Paul, USA) was prepared and inserted
into the spaces corresponding to the periodontal ligament [25]. The teeth were inserted by finger
pressure until the 2-mm mark of the amelocemental limit was aligned to the surface of the
cylinder. After polymerization, the excesses were removed by using a scalpel blade and the
specimens were stored in distilled water at 4°C.
The teeth, with their respective artificial alveoli, were randomly divided into four
groups (n = 5 to FBG sensors and n = 10 to cuspal deformation) according to material used for
restoration.
2.1.2 Cavity Preparation
Cavity preparation was performed by using abrasion equipment with three
coordinate axes millimetrically controlled (Mitutoyo Am. Corp., Ontario, Canada). Diamond
drills 2131 and 3131 (KD Sorensen, São Paulo, Brazil) were used to prepare the cavities by
positioning them perpendicularly to the long axis of the tooth, determining 6 degrees of
expulsion in the surrounding and axial walls and inner round angles.
Class II MOD cavities were standardized as follows: lingual-buccal isthmus width
of 4 mm and occlusal depth of 3 mm, occlusal depth of 4 mm in the proximal box and 1-2 mm
above the cemento-enamel junction of the proximal box, ending mesially and distally [26].
-
22
2.1.3 Insertion of Resin Composite
The resin composite insertion of Z100 group was incremental way from oblique
technique and thickness up until 2 mm. Three incremental were inserted, the first one was to
buccal wall, the second to opposite wall and the last one was to reproduce the teeth occlusal
anatomic.
Already for TNC, FBF and ABF groups were restored with one single increment of
4 mm thickness to fill all bulk MOD cavity.
2.2 Fiber Bragg Grating Sensors (FBG)
Optical FBG sensors were used to assess the shrinkage stress of resinous material
during the photopolymerization with values expressed in micro-strain (µɛ) (n = 5). Fiber Bragg
gratings consist of longitudinal modulations in the refractive index of optical fiber’s core region
with a periodicity in the order of hundreds of nanometers (Smart Fibers Ltd. UK). The refractive
index modulation allows the fundamental core mode to be coupled to a counter propagating
core mode. Coupling occurs at a specific wavelength, λB, given by Eq. (1) – where neff is the
effective refractive index of the fundamental core mode and Λ is the grating pitch.
𝜆𝐵 = 2 𝑛𝑒𝑓𝑓 Λ (1)
As the core mode is coupled to a counter propagating one by the Bragg grating, the
optical response of a FBG can be measured in reflection. Thus, if light from a broadband light
source is launched in a fiber in which a FBG was inscribed, a peak at λB (Bragg peak) can be
measured in the reflection spectrum. Figure 1a shows a typical spectrum from a Bragg grating
used in the experiments reported herein (Λ = 535.6 nm). The sensors were monitored with an
optical spectrum interrogation analyzer (W3-2050 FBG, Smart Fibers Ltd., UK), which can
measure tension up to 42,000 µɛ at maximum measurement frequency of 50Hz [27].
The application of strain to a FBG causes its spectral peak to shift. This shifting is
due to the fact that strain induces changes in the effective refractive index of the fundamental
core mode (via strain-optic effect), neff, and variations in the grating period, Λ. By observing
Eq. (1), it is possible to conclude that variations in neff and Λ imply in a different value for λB.
In order to characterize the Bragg wavelength shift as a function of the applied
strain, one have fixed the optical fiber on motorized translation stages and applied strain
-
23
increments to the same. Figure 1b shows the evolution of a FBG spectrum as strain increments
are applied to the fiber. Figure 1c presents the wavelength shift as a function of the strain level.
Measured data allowed calculating a strain sensitivity of (1.124 ± 0.005) pm/µε.
The evaluation of resin shrinking by the employment of fiber Bragg gratings was
carried out as follows: initially, the fiber was glued on supports and, in sequence, it was
tensioned. As a second step, the tooth under test was elevated up to fiber level. At this point,
the fiber could be attached on the side buccal wall of the MOD cavity at the level of the dentine
enamel junction (Adper Single Bond 2 adhesive was employed for fixation) [28]. The
restorative technique to be probed was then performed inside the cavity and the Bragg peak was
followed during the curing procedure. Temperatures effects of the light and curing exotherm
were not eliminated.
2.3 Cuspal Deformation
Strain gauges (PA-06-060CC-350-LEN, Excel Sensores, SP, Brazil) were used to
measure cuspal deformation (µɛ) and they were attached parallel to the long axis of the teeth at
the gingival wall level of the proximal box and placed on the external surface of the buccal and
lingual cusps (n = 10). The strain gauges had an internal electrical resistance of 350Ω and gauge
factor of 2.17. Moreover, two strain gauges were positioned to another unimpaired tooth to
compensate for dimensional alteration due to temperature effects on half bridge scheme. The
fixation of strain gauges were performed with 37% acid etching for 30 seconds, cyanoacrylate-
based adhesive (Super Bonder, Loctite, Itapeví, Brazil) and digital pressure for 60 seconds and
the wires were connected to a data acquisition device (ADS0500IP, Lynx, SP, Brazil). The
cuspal deformation values were recorded at 4 Hz during the restorative technique and continued
for 10 minutes after curing the last increment until stabilization [29].
2.4 Fracture Resistance Test
After measurement of cuspal deformation during restoration, the teeth were placed
on universal testing machine (EMIC, 500DL, São José dos Pinhais, Brazil) and submitted to
axial compressive loading with a metal sphere of 8 mm in diameter at speed of 0.5 mm/min up
to 100N in order to cause fracture and assess the cuspal deformation in these moments. The
load required (N) to cause catastrophic fracture of the specimens was recorded by a 500-N load
-
24
cell hardwired to a computer with control and data acquisition software (TESC 3.04, EMIC)
[30].
2.5 Scanning Electron Microscopy (SEM)
The failure mode of each specimen was analyzed under the SEM (JEOL-JSM
5600LV, Tokyo, Japan). Each sample was gold-sputter coated, submitted to analysis at 15kV
to observe surface characteristics and classified into one of the four categories as follows: (I)
fracture at resin composite; (II) fracture at resin composite and coronal tooth structure; (III)
fracture at resin composite and cervical tooth structure with possible periodontal repair; and
(IV) root fracture beyond repair.
2.6 Statistical Analysis
Statistical analysis was performed by using SPSS 21.0 (SPSS Inc., Chicago, IL,
USA). The evaluation of optical FBG sensors, cuspal deformation and fracture resistance were
tested for normal distribution (Shapiro-Wilk, p > 0.05), followed by parametric statistical tests
and one-way ANOVA with post-hoc Tukey’s test. Data on fracture mode were submitted to
Chi-square test.
3 RESULTS
3.1 Optical FBG Sensors
Average and standard deviation values of FBG evaluation are presented in Table 2
and Figure 1. One-way ANOVA revealed statistically significant differences among the
restorative materials (p < 0.0001). The statistics indicate that the Z100 group presented the
highest values of stress shrinkage as measured by the optical FBG sensors (p < 0.05). The bulk-
fill resin groups presented low polymerization stress values, especially the TNC group which
presented statistically significant difference than FBF group (p < 0.05).
3.2 Cuspal Deformation
Average and standard deviation values of cuspal deformation and fracture
resistance are listed in Table 3. One-way ANOVA analysis demonstrated statistical differences
-
25
between the groups (p < 0.03). Z100 group presented higher values of cuspal deformation at
restoration time and axial compressive compared of others groups (p < 0.01). The cuspal
deformation until the fracture presented only difference between TNC and FBF groups (p <
0.05). The fracture resistence of TNC group was higher of all groups, but only statistically
different of Z100 group (p < 0.01).
Failure mode distribution is shown in Figure 2 and 3. The Chi-square test (116.000)
indicates that restorative technique influenced the fracture pattern distribution in the different
groups (p < 0,0001), with Z100 and FBF groups presenting predominance of type-II fracture
and TNC and ABF groups presenting type-IV fracture.
4 DISCUSSION
The hypothesis predicting that the bulk-fill composite resins generate less
polymerization contraction stress and reduce cuspal deformation compared to conventional
composite resins was accepted. The results translate the behavior of the material inside the
cavity during photopolymerization (Table 2 and 3).
In dentistry, studies have been introducing optical FBG sensors for assessment of
dental materials, such as Rajan et al. (2016), who are pioneers in evaluating physical properties
of different composite resins by using optical fibers, thus providing conclusions on the
characteristics of these materials during and after photopolymerization. Also, Vinagre et al.
(2016) evaluated the cuspal deformation by using optical FBG sensors on the tips of cusps
during photopolymerization of bulk-fill composite resins. This has helped to understand the
basic characteristics of restorative materials and the influence of polymerization shrinkage
stress regarding the position of the cusps. However, the present study addressed the
biomechanical behavior of these materials inside a large MOD cavity, allowing the evaluation
of the shrinkage stress on the cavity walls and the influenced of the restoration techniques used
through the optical FBG.
Restorative materials inside the dental cavity shrink in volume when submitted to
polymerization, hence producing a tension by contraction, which interacts with their physical
properties and determine their behavior in the tooth cavity walls [33,34]. The results found
show that Z100 composite resin has a high shrinkage stress, significantly different of the other
groups as was reported by Ilie, Kunzelmann and Hickel (2006). The restoration with Z100 was
performed by using the incremental filling technique in order to minimize the tension created
-
26
on the cavity walls. However, this tension was significantly high in the first increment and
varied strongly in the subsequent ones depending on the position of the increment, also
demanding a longer chair time for restoration.
The force distribution found between restorative material and cavity walls
influenced heavily the prognosis of the restoration procedure, since poorly adapted restorations
are one of the most frequent causes of failure, thus requiring new restoration. Orlowski,
Tarczydlo and Chalas (2015) evaluated the marginal adaptation of different bulk-fill resins,
especially the Filtek and Tetric EvoCream Bulk Fill, and they conclued that the single increment
technique using this material weakly influence the marginal adjustment and marginal gap.
The evaluation of cuspal deformation using extensometry, which is a kind of non
destructive mechanical test, is directly related to the characteristics of the restoration process,
including restorative and adhesive materials and techniques used, and the size of the cavity
influences the behavior of the dental structure under forces acting on the adhesive layer [30,34].
All the specimens were standardized for evaluation of cuspal deformation by using strain
gauges.
Bulk-fill resins have physical characteristics such as high translucence to allow light
passage, different viscosities facilitating their handling, changes in their structure (e.g. volume
filler) and modifications in the photoinitiator to avoid shrinkage stress [19,23,35]. The resin
composite TNC have an initiator system fusion, like a camphorquinone conventional and an
acyl phosphine oxide with Ivocerim® patented system, which it’s more sensitive from
photopolymerization process requiring lower wavelength comparable of conventional [36]. In
addition, it’s presented high molecular weight monomers with long chains due to compensate
the less distance between the formed polymers and particles with low modulus of elasticity [37-
39].
In the literature, there are conflicting studies on the behavior of bulk-fill resins
[12,19]. Our results allow us to state that bulk-fill resins have less polymerization stress than
the conventional composite ones. The forces generated by cuspal deformation were similar.
Hence, extensometry and optical FBG sensor tests complement each other, helping confirm the
behaviour of different materials used for restoring Class II MOD cavity as was observed in the
study of Lee et al. (2007), which evaluated the cusp deflection of premolars from cavity
proportion and selected restorative method and, from this, testify a correlation between the cusp
deflection and stress shrinkage polymerization tests.
-
27
In this study was preconized an only kind of adhesive used (Adper Single Bond 2)
due to standard the adhesive technique, making possible a specific evaluation of material and
restorative technique to be used, similar of Kim et al. (2015) study.
The fracture resistance of restorations of Class II MOD cavities was described by
Eakle (1986), and it can be augmented by composite resin restorations. The results on fracture
resistance in the present study corroborate the finding reported by Atalay et al. (2016), who
found similarity between bulk-fill and conventional resins. However, the Z100 resin presented
low fracture resistance, with predominance of type-II fractures, possibly influenced by its
behavior during photopolymerization inducing stress accumulation. The higher and significant
fracture resistance of TNC group is related to its filler content and higher flexural strength
consequently [38,43].
5 CONCLUSIONS
Despite the limitations of this in vitro study, the following conclusions may be
drawn:
The bulk fill resin composites inserted in bulk increment of 4 mm depth at short
photo-activation time generate less shrinkage stress and less cuspal deformation than the
conventional resin composite.
The TNC resin composite present low shrinkage stress, cuspal deformation, more
resistant to fractures however reproduce catastrophic fractures.
Acknowledgements
This study was supported by the Gleb Wataghin Institute of Physics (State
University of Campinas) and Dental Research Center Biomechanics, Biomaterials and Cell
Biology (Dental School of Federal University of Uberlândia).
References
[1] Kwon HJ, Oh YJ, Jang JH, Park JE, Hwang KS, Park YD. The effect of polymerization
conditions on the amounts of unreacted monomer and bisphenol A in dental composite resins.
Dent Mater J. 2015;34(3):327-35. doi: 10.4012/dmj.2014-230.
-
28
[2] Demarco FF, Collares K, Coelho-de-Souza FH, Correa MB, Cenci MS, Moraes RR,Opdam
NJ. Anterior composite restorations: A systematic review on long-term survival and reasons for
failure. Dent Mater. 2015 Oct;31(10):1214-24. doi: 10.1016/j.dental.2015.07.005.
[3] Demarco FF, Corrêa MB, Cenci MS, Moraes RR, Opdam NJ. Longevity of posterior
composite restorations: not only a matter of materials. Dent Mater. 2012 Jan;28(1):87-101. doi:
10.1016/j.dental.2011.09.003.
[4] Cetin AR, Unlu N, Cobanoglu N. A five-year clinical evaluation of direct nanofilled and
indirect composite resin restorations in posterior teeth. Oper Dent. 2013 Mar-Apr;38(2):E1-11.
doi: 10.2341/12-160-C.
[5] Pashley DH, Tay FR, Breschi L, Tjäderhane L, Carvalho RM, Carrilho M, Tezvergil-
Mutluay A. State of the art etch-and-rinse adhesives. Dent Mater. 2011 Jan;27(1):1-16. doi:
10.1016/j.dental.2010.10.016.
[6] Nahsan FP, Mondelli RF, Franco EB, Naufel FS, Ueda JK, Schmitt VL, Baseggio W.
Clinical strategies for esthetic excellence in anterior tooth restorations: understanding color and
composite resin selection. J Appl Oral Sci. 2012 Mar-Apr;20(2):151-6.
[7] Deliperi S, Bardwell DN. An alternative method to reduce polymerization shrinkage in
direct posterior composite restorations. J Am Dent Assoc. 2002 Oct;133(10):1387-98. Review.
Erratum in: J Am Dent Assoc. 2002 Dec;133(12):1614.
[8] Bönecker M, Tenuta LM, Pucca Junior GA, Costa PB, Pitts N. A social movement to reduce
caries prevalence in the world. Braz Oral Res. 2013 Jan-Feb;27(1):5-6.
[9] Da Rosa Rodolpho PA, Donassollo TA, Cenci MS, Loguércio AD, Moraes RR, Bronkhorst
EM, Opdam NJ, Demarco FF. 22-Year clinical evaluation of the performance of two posterior
composites with different filler characteristics. Dent Mater. 2011 Oct;27(10):955-63. doi:
10.1016/j.dental.2011.06.001.
[10] Ferracane JL. Resin composite--state of the art. Dent Mater. 2011 Jan;27(1):29-38. doi:
10.1016/j.dental.2010.10.020.
[11] Casselli DS, Faria-e-Silva AL, Casselli H, Martins LR. Marginal adaptation of class V
composite restorations submitted to thermal and mechanical cycling. J Appl Oral Sci. 2013 Jan-
Feb;21(1):68-73.
-
29
[12] Benetti AR, Havndrup-Pedersen C, Honoré D, Pedersen MK, Pallesen U. Bulk-fill resin
composites: polymerization contraction, depth of cure, and gap formation. Oper Dent. 2015
Mar-Apr;40(2):190-200. doi: 10.2341/13-324-L.
[13] Fronza BM, Rueggeberg FA, Braga RR, Mogilevych B, Soares LE, Martin AA,
Ambrosano G, Giannini M. Monomer conversion, microhardness, internal marginal adaptation,
and shrinkage stress of bulk-fill resin composites. Dent Mater. 2015 Dec;31(12):1542-51. doi:
10.1016/j.dental.2015.10.001.
[14] Lee SK, Kim TW, Son SA, Park JK, Kim JH, Kim HI, Kwon YH. Influence of light-curing
units on the polymerization of low-shrinkage composite resins. Dent Mater J. 2013;32(5):688-
94.
[15] Loguercio AD, Reis A, Schroeder M, Balducci I, Versluis A, Ballester RY. Polymerization
shrinkage: effects of boundary conditions and filling technique of resin composite restorations.
J Dent. 2004 Aug;32(6):459-70.
[16] Czasch P, Ilie N. In vitro comparison of mechanical properties and degree of cure of bulk
fill composites. Clin Oral Investig. 2013 Jan;17(1):227-35. doi: 10.1007/s00784-012-0702-8.
[17] Roggendorf MJ, Krämer N, Appelt A, Naumann M, Frankenberger R. Marginal quality of
flowable 4-mm base vs. conventionally layered resin composite. J Dent. 2011 Oct;39(10):643-
7. doi: 10.1016/j.jdent.2011.07.004.
[18] Leprince JG, Palin WM, Vanacker J, Sabbagh J, Devaux J, Leloup G. Physico-mechanical
characteristics of commercially available bulk-fill composites. J Dent. 2014 Aug;42(8):993-
1000. doi: 10.1016/j.jdent.2014.05.009.
[19] Orłowski M, Tarczydło B, Chałas R. Evaluation of marginal integrity of four bulk-fill
dental composite materials: in vitro study. Scientific World Journal. 2015; 2015:701262. doi:
10.1155/2015/701262.
[20] Kwon Y, Ferracane J, Lee IB. Effect of layering methods, composite type, and flowable
liner on the polymerization shrinkage stress of light cured composites. Dent Mater. 2012
Jul;28(7):801-9. doi: 10.1016/j.dental.2012.04.028.
[21] Khosravi K, Mousavinasab SM, Samani MS. Comparison of microleakage in Class II
cavities restored with silorane-based and methacrylate-based composite resins using different
restorative techniques over time. Dent Res J (Isfahan). 2015 Mar-Apr;12(2):150-6.
[22] Campos EA, Ardu S, Lefever D, Jassé FF, Bortolotto T, Krejci I. Marginal adaptation of
class II cavities restored with bulk-fill composites. J Dent. 2014 May;42(5):575-81. doi:
10.1016/j.jdent.2014.02.007.
-
30
[23] El-Damanhoury H, Platt J. Polymerization shrinkage stress kinetics and related properties
of bulk-fill resin composites. Oper Dent. 2014 Jul-Aug;39(4):374-82. doi: 10.2341/13-017-L.
[24] Moorthy A, Hogg CH, Dowling AH, Grufferty BF, Benetti AR, Fleming GJ. Cuspal
deflection and microleakage in premolar teeth restored with bulk-fill flowable resin-based
composite base materials. J Dent. 2012 Jun;40(6):500-5. doi: 10.1016/j.jdent.2012.02.015.
[25] Soares CJ, Pizi EC, Fonseca RB, Martins LR. Influence of root embedment material and
periodontal ligament simulation on fracture resistance tests. Braz Oral Res. 2005 Jan-
Mar;19(1):11-6.
[26] Frankenberger R, Krämer N, Appelt A, Lohbauer U, Naumann M, Roggendorf MJ.
Chairside vs. labside ceramic inlays: effect of temporary restoration and adhesive luting on
enamel cracks and marginal integrity. Dent Mater. 2011 Sep;27(9):892-8. doi:
10.1016/j.dental.2011.05.007.
[27] Anttila EJ, Krintilä OH, Laurila TK, Lassila LV, Vallittu PK, Hernberg RG. Evaluation of
polymerization shrinkage and hydroscopic expansion of fiber-reinforced biocomposites using
optical fiber Bragg grating sensors. Dent Mater. 2008 Dec;24(12):1720-7. doi:
10.1016/j.dental.2008.07.006.
[28] Urabe I, Nakajima S, Sano H, Tagami J. Physical properties of the dentin-enamel junction
region. Am J Dent. 2000 Jun;13(3):129-35.
[29] Bicalho AA, Pereira RD, Zanatta RF, Franco SD, Tantbirojn D, Versluis A, Soares CJ.
Incremental filling technique and composite material--part I: cuspal deformation, bond
strength, and physical properties. Oper Dent. 2014 Mar-Apr;39(2):E71-82. doi: 10.2341/12-
441-L.
[30] Pereira R, Bicalho AA, Franco SD, Tantbirojn D, Versluis A, Soares CJ. Effect of
Restorative Protocol on Cuspal Strain and Residual Stress in Endodontically Treated Molars.
Oper Dent. 2016 Jan-Feb;41(1):23-33. doi: 10.2341/14-178-L.
[31] Rajan G, Shouha P, Ellakwa A, Bhowmik K, Xi J, Prusty G. Evaluation of the physical
properties of dental resin composites using optical fiber sensing technology. Dent Mater. 2016
Sep;32(9):1113-23. doi: 10.1016/j.dental.2016.06.015.
[32] Vinagre A, Ramos J, Alves S, Messias A, Alberto N, Nogueira R. Cuspal Displacement
Induced by Bulk Fill Resin Composite Polymerization: Biomechanical Evaluation Using Fiber
Bragg Grating Sensors. Int J Biomater. 2016;2016:7134283. doi: 10.1155/2016/7134283.
[33] Ilie N, Kunzelmann KH, Hickel R. Evaluation of micro-tensile bond strengths of composite
materials in comparison to their polymerization shrinkage. Dent Mater. 2006 Nov;22(7):593-
601.
-
31
[34] Bicalho AA, Valdívia AD, Barreto BC, Tantbirojn D, Versluis A, Soares CJ. Incremental
filling technique and composite material--part II: shrinkage and shrinkage stresses. Oper Dent.
2014 Mar-Apr;39(2):E83-92. doi: 10.2341/12-442-L.
[35] Atalay C, Yazici AR, Horuztepe A, Nagas E, Ertan A, Ozgunaltay G. Fracture Resistance
of Endodontically Treated Teeth Restored With Bulk Fill, Bulk Fill Flowable, Fiber-reinforced,
and Conventional Resin Composite. Oper Dent. 2016 Jun 28.
[36] Zorzin J, Maier E, Harre S, Fey T, Belli R, Lohbauer U, Petschelt A, Taschner M. Bulk-
fill resin composites: polymerization properties and extended light curing. Dent Mater. 2015
Mar;31(3):293-301. doi: 10.1016/j.dental.2014.12.010.
[37] Miletic V, Peric D, Milosevic M, Manojlovic D, Mitrovic N. Local deformation fields and
marginal integrity of sculptable bulk-fill, low-shrinkage and conventional composites. Dent
Mater. 2016 Nov;32(11):1441-1451. doi: 10.1016/j.dental.2016.09.011.
[38] Omran TA, Garoushi S, Abdulmajeed AA, Lassila LV, Vallittu PK. Influence of increment
thickness on dentin bond strength and light transmission of composite base materials. Clin Oral
Investig. 2016 Sep 10.
[39] Rauber GB, Bernardon JK, Vieira LC, Maia HP, Horn F, Roesler CR. In Vitro Fatigue
Resistance of Teeth Restored With Bulk Fill versus Conventional Composite Resin. Braz Dent
J. 2016 Jul-Aug;27(4):452-7. doi: 10.1590/0103-6440201600836.
[40] Lee MR, Cho BH, Son HH, Um CM, Lee IB. Influence of cavity dimension and restoration
methods on the cusp deflection of premolars in composite restoration. Dent Mater. 2007
Mar;23(3):288-95.
[41] Kim RJ, Kim YJ, Choi NS, Lee IB. Polymerization shrinkage, modulus, and shrinkage
stress related to tooth-restoration interfacial debonding in bulk-fill composites. J Dent. 2015
Apr;43(4):430-9. doi: 10.1016/j.jdent.2015.02.002.
[42] Eakle WS. Fracture resistance of teeth restored with class II bonded composite resin. J
Dent Res. 1986 Feb;65(2):149-53.
[43] Zhang H, Zhang ML, Qiu LH, Yu JT, Zhan FL. [Comparison of wear resistance and
flexural strength of three kinds of bulk-fill composite resins]. Shanghai Kou Qiang Yi Xue.
2016 Jun;25(3):292-5. Chinese.
-
32
Tables
-
33
Table 2 – Results of FBG evaluation ( µε)
Z100 TNC FBF ABF
Average (Standard
Deviation) 949.1 ± 236.9 A 105.3 ± 205.3 C 525.8 ± 71.2 B 434.8 ± 306.0 BC
Average followed by the same letter are not statistically different (p > 0.05). n = 5 specimens /
group.
Table 3 – Results of cuspal deformation ( µε) and fracture resistance (N)
Cuspal Deformation
Restoration Axial Compressive (100N) Fracture Fracture Resistence
Z100 136.2 ± 41.4 A 89.8 ± 43.1 A 999.1 ± 228.4 AB 1624.2 ± 561.2 B
TNC 82.6 ± 25.0 B 50.3 ± 31.7 B 1074.5 ± 381.1 A 2925.8 ± 906.6 A
FBF 87.3 ± 25.9 B 34.1 ± 12.8 B 751.6 ± 243.6 B 1989.9 ± 846.1 AB
ABF 70.9 ± 11.8 B 34.9 ± 10.4 B 838.1 ± 119.8 AB 2166.1 ± 867.6 AB
Average followed by the same letter are not statistically different (p > 0.05). n = 10 specimens / group.
-
34
Figures
Figure 1. (a) Typical FBG reflection spectrum. (b) FBG spectra as a function of the applied
strain. (c) Wavelength shift as a function of the applied strain.
Figure 2. Percentage of specimens (%) according to fracture pattern classification.
0%
10%
20%
30%
40%
50%
60%
70%
80%
90%
100%
Z100 TNC FBF ABF
Failu
re m
od
e
IV (root fracture beyondrepair)
III (resin composite andcervical tooth structure withpossible periodontal repair)
II (resin composite andcoronal structure)
I (resin composite)
-
35
Figure 3. SEM representative images of fracture pattern classification. A) Type-1 fracture at
composite resin (RC); B) Type-2 fracture at composite resin and dentin (D); C) Type-3
fracture at composite resin and cervical tooth structure, enamel (E) and dentin (D); D) Type-4
fracture with pulp canal (PC) exposure.
-
36
3 CONCLUSÃO
De acordo com os resultados obtidos e considerando as limitações das metodologias
utilizadas, pôde-se concluir que:
1. Resinas compostas bulk fill apresentam menor tensão de contração e deformação
de cúspide durante a fotopolimerização de um único incremento com 4 mm de espessura em
relação à resina composta convencional.
2. A resina composta TNC apresenta menor tensão de contração de polimerização,
menor deformação de cúspide, além de ser mais resistente à fraturas, contudo, apresenta maior
frequência de fraturas catastróficas.
3. O padrão de fratura das resinas compostas quando induzidas por forças
compressivas é diretamente dependente do material resinoso utilizado.
-
37
REFERÊNCIAS
Al-Fakih E, Abu Osman NA, Mahamd Adikan FR. The use of fiber Bragg grating sensors in
biomechanics and rehabilitation applications: the state-of-the-art and ongoing research topics.
Sensors (Basel). 2012 Sep 25;12(10):12890-926. doi: 10.3390/s121012890.
Anatavara S, Sitthiseripratip K, Senawongse P. Stress relieving behaviour of flowable
composite liners: A finite element analysis. Dent Mater J. 2016;35(3):369-78. doi:
10.4012/dmj.2015-204
Atalay C, Yazici AR, Horuztepe A, Nagas E, Ertan A, Ozgunaltay G. Fracture Resistance of
Endodontically Treated Teeth Restored With Bulk Fill, Bulk Fill Flowable, Fiber-reinforced,
and Conventional Resin Composite. Oper Dent. 2016 Jun 28.
Benetti AR, Havndrup-Pedersen C, Honoré D, Pedersen MK, Pallesen U. Bulk-fill resin
composites: polymerization contraction, depth of cure, and gap formation. Oper Dent. 2015
Mar-Apr;40(2):190-200. doi: 10.2341/13-324-L.
Bicalho AA, Pereira RD, Zanatta RF, Franco SD, Tantbirojn D, Versluis A, Soares CJ.
Incremental filling technique and composite material--part I: cuspal deformation, bond
strength, and physical properties. Oper Dent. 2014 Mar-Apr;39(2):E71-82. doi: 10.2341/12-
441-L.
Bönecker M, Tenuta LM, Pucca Junior GA, Costa PB, Pitts N. A social movement to reduce
caries prevalence in the world. Braz Oral Res. 2013 Jan-Feb;27(1):5-6.
Campos EA, Ardu S, Lefever D, Jassé FF, Bortolotto T, Krejci I. Marginal adaptation of class
II cavities restored with bulk-fill composites. J Dent. 2014 May;42(5):575-81. doi:
10.1016/j.jdent.2014.02.007.
De acordo com as normas da UNICAMP/FOP, baseadas na padronização do International Committee of Medical
Journal Editors – Vancouver Group. Abreviatura dos periódicos em conformidade com o PubMed.
-
38
Casselli DS, Faria-e-Silva AL, Casselli H, Martins LR. Marginal adaptation of class V
composite restorations submitted to thermal and mechanical cycling. J Appl Oral Sci. 2013
Jan-Feb;21(1):68-73.
Cetin AR, Unlu N, Cobanoglu N. A five-year clinical evaluation of direct nanofilled and
indirect composite resin restorations in posterior teeth. Oper Dent. 2013 Mar-Apr;38(2):E1-
11. doi: 10.2341/12-160-C.
Czasch P, Ilie N. In vitro comparison of mechanical properties and degree of cure of bulk fill
composites. Clin Oral Investig. 2013 Jan;17(1):227-35. doi: 10.1007/s00784-012-0702-8.
Da Rosa Rodolpho PA, Donassollo TA, Cenci MS, Loguércio AD, Moraes RR, Bronkhorst
EM, et al. 22-Year clinical evaluation of the performance of two posterior composites with
different filler characteristics. Dent Mater. 2011 Oct;27(10):955-63. doi:
10.1016/j.dental.2011.06.001.
Deliperi S, Bardwell DN. An alternative method to reduce polymerization shrinkage in direct
posterior composite restorations. The Journal of the American Dental Association. 2002;
133(10): 1387-1398.
Demarco FF, Collares K, Coelho-de-Souza FH, Correa MB, Cenci MS, Moraes RR, et al.
Anterior composite restorations: A systematic review on long-term survival and reasons for
failure. Dent Mater. 2015 Oct;31(10):1214-24. doi: 10.1016/j.dental.2015.07.005.
Demarco FF, Corrêa MB, Cenci MS, Moraes RR, Opdam NJ. Longevity of posterior
composite restorations: not only a matter of materials. Dent Mater. 2012 Jan;28(1):87-101.
doi: 10.1016/j.dental.2011.09.003.
Dominguez JA, Bittencourt BF, Farago PV, Pinheiro LA, Gomes OMM. Shrinkage stress of
resin composites: effect of material composition - sistematic review. Braz Dent Sci 2014
Jul/Sep;17(3).
El-Damanhoury H, Platt J. Polymerization shrinkage stress kinetics and related properties of
bulk-fill resin composites. Oper Dent. 2014 Jul-Aug;39(4):374-82. doi: 10.2341/13-017-L.
-
39
Ferracane JL. Resin composite--state of the art. Dent Mater. 2011 Jan;27(1):29-38. doi:
10.1016/j.dental.2010.10.020.
Fronza BM, Rueggeberg FA, Braga RR, Mogilevych B, Soares LE, Martin AA, et al.
Monomer conversion, microhardness, internal marginal adaptation, and shrinkage stress of
bulk-fill resin composites. Dent Mater. 2015 Dec;31(12):1542-51. doi:
10.1016/j.dental.2015.10.001.
Kalliecharan D, Germscheid W, Price RB, Stansbury J, Labrie D. Shrinkage stress kinetics of
Bulk Fill resin-based composites at tooth temperature and long time. Dent Mater. 2016
Nov;32(11):1322-1331. doi: 10.1016/j.dental.2016.07.015.
Kwon HJ, Oh YJ, Jang JH, Park JE, Hwang KS, Park YD. The effect of polymerization
conditions on the amounts of unreacted monomer and bisphenol A in dental composite resins.
Dent Mater J. 2015;34(3):327-35. doi: 10.4012/dmj.2014-230.
Lee SK, Kim TW, Son SA, Park JK, Kim JH, Kim HI, et al. Influence of light-curing units on
the polymerization of low-shrinkage composite resins. Dent Mater J. 2013;32(5):688-94.
Leprince JG, Palin WM, Vanacker J, Sabbagh J, Devaux J, Leloup G. Physico-mechanical
characteristics of commercially available bulk-fill composites. J Dent. 2014 Aug;42(8):993-
1000. doi: 10.1016/j.jdent.2014.05.009.
Loguercio AD, Reis A, Schroeder M, Balducci I, Versluis A, Ballester RY. Polymerization
shrinkage: effects of boundary conditions and filling technique of resin composite
restorations. J Dent. 2004 Aug;32(6):459-70.
Moorthy A, Hogg CH, Dowling AH, Grufferty BF, Benetti AR, Fleming GJ. Cuspal
deflection and microleakage in premolar teeth restored with bulk-fill flowable resin-based
composite base materials. J Dent. 2012 Jun;40(6):500-5. doi: 10.1016/j.jdent.2012.02.015.
-
40
Nahsan FP, Mondelli RF, Franco EB, Naufel FS, Ueda JK, Schmitt VL, et al. Clinical
strategies for esthetic excellence in anterior tooth restorations: understanding color and
composite resin selection. J Appl Oral Sci. 2012 Mar-Apr;20(2):151-6.
Orłowski M, Tarczydło B, Chałas R. Evaluation of marginal integrity of four bulk-fill dental
composite materials: in vitro study. Scientific World Journal. 2015; 2015:701262. doi:
10.1155/2015/701262.
Ozakar-Ilday N, Zorba YO, Yildiz M, Erdem V, Seven N, Demirbuga S. Three-year clinical
performance of two indirect composite inlays compared to direct composite restorations. Med
Oral Patol Oral Cir Bucal. 2013 May 1;18(3):e521-8.
Park JW, Ferracane JL. Measuring the residual stress in dental composites using a ring slitting
method. Dent Mater. 2005 Sep;21(9):882-9.
Park JW, Ferracane JL. Residual stress in composites with the thin-ring-slitting approach. J
Dent Res. 2006 Oct;85(10):945-9.
Pashley DH, Tay FR, Breschi L, Tjäderhane L, Carvalho RM, Carrilho M, et al. State of the
art etch-and-rinse adhesives. Dent Mater. 2011 Jan;27(1):1-16. doi:
10.1016/j.dental.2010.10.016.
Pereira R, Bicalho AA, Franco SD, Tantbirojn D, Versluis A, Soares CJ. Effect of Restorative
Protocol on Cuspal Strain and Residual Stress in Endodontically Treated Molars. Oper Dent.
2016 Jan-Feb;41(1):23-33. doi: 10.2341/14-178-L.
Rajan G, Shouha P, Ellakwa A, Bhowmik K, Xi J, Prusty G. Evaluation of the physical
properties of dental resin composites using optical fiber sensing technology. Dent Mater. 2016
Sep;32(9):1113-23. doi: 10.1016/j.dental.2016.06.015.
Roggendorf MJ, Krämer N, Appelt A, Naumann M, Frankenberger R. Marginal quality of
flowable 4-mm base vs. conventionally layered resin composite. J Dent. 2011
Oct;39(10):643-7. doi: 10.1016/j.jdent.2011.07.004.
-
41
Umesh S, Padma S, Asokan S, Srinivas T. Fiber Bragg Grating based bite force measurement.
J Biomech. 2016 Jul 1. pii: S0021-9290(16)30732-1. doi:10.1016/j.jbiomech.2016.06.036.
Van Dijken JW, Pallesen U. Posterior bulk-filled resin composite restorations: A 5-year
randomized controlled clinical study. J Dent. 2016 Aug;51:29-35.
doi:10.1016/j.jdent.2016.05.008.
Vinagre A, Ramos J, Alves S, Messias A, Alberto N, Nogueira R. Cuspal Displacement
Induced by Bulk Fill Resin Composite Polymerization: Biomechanical Evaluation Using
Fiber Bragg Grating Sensors. Int J Biomater. 2016;2016:7134283. doi:
10.1155/2016/7134283.
Witzel MF, Braga RR, Ballester RY, Lima RG. Influence of specimen dimensions on nominal
polymerization contraction stress of a dental composite. J. of the Braz. Soc. of Mech. Sci. &
Eng. 2005 Jul-Sept 27(3):283-287.
Yeh CH, Chow CW, Wu PC, Tseng FG. A simple fiber Bragg grating-based sensor network
architecture with self-protecting and monitoring functions. Sensors (Basel). 2011;11(2):1375-
82. doi: 10.3390/s110201375.
-
42
APÊNDICE 1
Metodologia Ilustrada
Figura 1. Preparo das amostras. A) Seleção de molares hígidos; B) Simulação do
ligamento periodontal 2 mm abaixo da junção amelocementária com cera 7; C) Posicionamento
do dente à haste de um delineador protético; D) Centralização de uma película radiográfica com
perfuração central ao dente; E) Posicionamento de um tubo de PVC para inclusão da porção
radicular em resina de poliestireno; F) Inclusão do dente na resina de poliestireno; G) Remoção
dos dentes nos alvéolos artificiais e eliminação da cera 7 na raiz do dente; H) Realização da
cavidade MOD em máquina de preparo; I) Cavidade MOD confeccionada; J) Inclusão do
poliéter Impregum.
-
43
Figura 2. Avaliação da tensão de contração de polimerização dos materiais
restauradores pelos sensores de Bragg. A) Condicionamento com ácido fosfórico a 37% no
esmalte por 15 segundos; B) Condicionamento com ácido fosfórico sobre a dentina por 15
segundos e no esmalte por mais 15 segundos; C) Lavagem abundante com água por 30
segundos; D) Controle da umidade mantendo a dentina levemente úmida; E) Fibra óptica
esticada com a presença dos sensores de Bragg na região azul; F) Posicionamento dos sensores
de Bragg na região amelodentinária justaposto à parede vestibular da cavidade MOD; G)
Aplicação do adesivo Adper Single Bond 2 seguido da volatilização do solvente por 5 segundos;
H) Fotopolimerização do adesivo por 40 segundos; I) Fibra óptica posicionada na parede
vestibular da cavidade MOD com o sensor de Bragg aderido à parede com adesivo fotoativado.
-
44
Figura 3. Teste de extensometria para avaliação da deformação de cúspide. A)
Aplicação de ácido fosfórico por 30 segundos na região determinada para instalação do
extensômetro; B) Posicionamento do extensômetro na base da cúspide vestibular e
lingual/palatina ao nível da parede gengival das caixas proximais da cavidade MOD com cola
a base de cianoacrilato; C) Fotopolimerização das restaurações conforme o protocolo dos
grupos de tratamento e avaliação da deformação de cúspide pelos extensômetros; D) Ligação
dos extensômetros à máquina analisadora por meio de pontos de solda; E) Avaliação da
-
45
deformação de cúspide durante carregamento de 100N e até a fratura em máquina de ensaio
universal.
Figura 4. Protocolo restaurador dos grupos com resina convencional e resina bulk
fill. A) Inserção de forma incremental do G Z100; B C e D) Inserção de um único incremento
de 4mm dos grupos TNC, FBF e ABF, respectivamente.
-
46
ANEXO 1
-
47
ANEXO 2: Dental Materials