ionization chambers proportional counters geiger …sleoni/teaching/nuc-phys-det/pdf/lezione...gas...

32
Gas Detectors [the oldest detectors] Ionization chambers Proportional counters Geiger Muller counters Same basic principle, different intensity of applied electric field Basic principle: - gas ionization from radiation interaction; - electric signal originated by ion-electron pairs collected through an electric field.

Upload: dangthuy

Post on 04-Jul-2018

227 views

Category:

Documents


0 download

TRANSCRIPT

Gas Detectors [the oldest detectors]

§  Ionization chambers §  Proportional counters §  Geiger Muller counters

Same basic principle, different intensity of applied electric field

Basic principle: - gas ionization from radiation interaction; - electric signal originated by ion-electron pairs collected through an electric field.

Ionization Process

basic signal Pair formation: - direct interaction - secondary process (very energetic electrons can ionize …)

w ∼ 30 eV average energy per pair production: weak dependence on gas and particle type

< 1 ! !

# of pairs (e-,ion) depends on particle energy

Example: Ep ∼ 1 MeV # pairs ∼ 30000

Ions/electrons suffer of multiple collisions with gas molecules and lose energy ⇒ thermal equilibrium with gas, ricombination, … loss of signal

Thermal motion produces diffused charge transport

from high energy regions

Electric Field is needed to efficiently collect the charges from interaction place Net effect: superposition of thermal velocity (random) and drift velocity (ordered)

v

v

v

diffusion coefficient

v v

= a τ = F/m τ

Drift velocity

pEvdrift µ∝

Typical values: vdrift = 1 m/s Transit time (1 cm distance) = 1 ms !!!

v

v

def

Gas admixture

90%Ar+10% methane

10%Ar+90% methane vdrift v- v+

[reduced mass compared to ions]

Region of operation of gas detectors

# collected ions

1 2

3

4

5 Signal proportional to primary ionization

Signal proportional to primary+secondary ionization

Same Signal independent of energy

Loss of original Signal

Nr. Collected Ions vs. Voltage 1

2

3

4 5

Region of operation depends on ⇒

General constituents §  Container filled with gas §  Two isolated electrodes §  High Voltage

Ionization Chamber: the simplest gas detector

10-100 V E = constant

E ∼ 1/r Choice of gas: - air (γ detection) - Ar (more dense) to increase ionzation probability (p ∼ 1 atm) Mode of operation: - current mode [average rate info] - pulse mode [info on individual event]

Measured current is true indication of rate of formation of charges

Signal type and size: α particle Eα = 3 MeV w = 30 eV ⇒ n =Eα/w = 105 e-ions

SMALL signals, amplification is needed !

(low density and Z !!!)

Force acting to move charges to electrodes

Energy absorbed by e- and e+

v v

v

Time dependent signals observed in the detector

τ = RC >>tc signal rise time depends on

charge collection time tc

v v

v v-

v-( Δ

The pulse is chopped at a convenient time by a shaping circuit

Ionization Chamber pulse mode operation with Frisch grid The fine mashing grid removes the pulse-amplitude dependence on position of interaction: •  The grid is maintained to intermediate potential and is transparent to electrons •  Incident radiation directed only into active volume Chathode-Grid

d = 1.6 cm

d = 16 cm

in PRISMA Ionization Chamber @LNL Frisch

NO resistor ⇒ NO measurable signal

The signal derives from electrons drift only ⇒ NO dependence on slow ion motion !!!

Shield

Application 1 IC for Z identification of reaction products

22

mZEEmZE

dxdE

=×∝×

Bhete-Bloch

Application 2

Ionization chamber

1 µCu (∼ 0.3 µg) of 241Am τ = 432 y - α passing through ionization air-chamber produce costant current - smoke absorbes α’s ⇒ reduced ionization, lower signal, allarm .. - α’a have low penetrability: they are stopped by the plastic of detector

NEUTRON detectors Detection of secondary events produced in - nuclear reactions: (n.p), (n,α), (n,γ), (n,fission) - nuclear scattering from light charged particles, than detected For slow and thermal n: (n,p) and (n,α) mostly … Best reaction: 10B + n → 7Li* + α

γ = 0.48 MeV

•  Exotermic reaction: it occurs also for very low n energy For En = 0.025 eV, σ = 3770 b !! ⇒ ideal for n detection ⇒ useful for measuring the flux of thermal n

∼ 3000 b

∼1/v

•  Natural abundance of 10B is high: 20% ⇒ material enriched with 10B increase the efficiency

Application 3

Ionization chamber/proportional counters filled with BF3 gas

Paraffine as

moderator

Ideal response

Real Response

Wall effect

Due to partial energy deposition of lost ion

Lost 7Li Lost

7Li Lost α

Lost α

[typical sized tubes: e.g., 2 - 5 cm diameter]

Only determination of interaction, NOT energy

Wall effect prevents precise energy measurements

Proportional Counters Basic Principle: signal moltiplication to amplify charge produced by original (e-, ion) pair

⇒ Avalanche Formation: by increasing E electrons acquire kinetic energy to ionize … Ethreshold ∼ 106 V/m, p ∼ 1 atm

xenxn

dxndn

α

α

)0()( =

= fractional increase of electrons per unit path length dx

density of electrons increases exponentially in the direction of motion of the avalanche

Townsend coefficient

α = 0 if E < Eth

α ∝ E if E > Eth

Townsend coefficients

Ar-CO2 mixture

Best Geometry: Cylindrical Polarity is very important: electrons must by attracted to the center

a = anode radius b = cathod radius

Example: V = 2000 V a = 0.008 cm ⇒ E(a) = 106 V/m b = 1 cm N.B. in planar geometry equivalent field requires V ∼ 105 Volt

Electrons are more mobile than ions ⇒  Avalanche looks like a liquid drop with electrons at the head

Uniform multiplication is obtained only in a confined region

d ∼ 5 a

MonteCarlo simulation

a = anode radius b = cathod radius

Pulse signal from a cylindrical proportional counter

τ  = RC

The pulse is usually cut by an RC differentiating circuit

Total Charge produced

(E) Q = n0 e M

Gas multiplication Factor

M = 102 - 104

Application Multi Wire Proportional Counter [invented by G. Charpak, CERN 1968, Nobel Price in 1992]

Detector sensitive to position of radiation interaction: - 2 planes of wires (anodes and coathods) - Δx ∼ 2-3 mm - size ∼ 1 m2

Electric field lines and constant potential energy surface

Wire read-out may be very complicated ⇒ read-out of more than one wire at just one end

Position reconstructed by charge division

Read-Out methods 1. Separate wire read-out complicated and costly (too many wires) 1. Analog methods: - center of gravity - charge division

Center of gravity Chatode in strips: Signals from each strip is stored and center of gravity is calculated More layers: x, y, diagonal …

correction factor for noise effects

Charge Division Charge collected at the two ends of the resistive anode is divided in proportion to the length of the wire from the point at which the charge is injected

BA

A

QQQ

Ly

+=

MWPPAC: the Focal plane detector of the PRISMA spectrometer at LNL Laboratory (Padova)

- 3 electrode structure: central cathod

2 anodic wire planes (X and Y)

- cathode: 3300 wires of 20µm

0.3 mm spacing negative high voltage: 500-600 V

- X plane: 10 x 100 wires each

1mm spacing

- - Y plane: common to all cathode, -  130 wires, 1 m long, 1mm steps

- spatial resolution: ∆X ~ 1mm, ∆Y ~ 2mm

1 m

13 c

m

Focal Plane Detectors of PRISMA: in-beam tests

X position (channels)

∆t ~ 300 ps ∆X = 1 mm ∆Y = 2 mm

Y position (channels)

MWPPAC (ε ~ 100%)

different shapes due to

PRISMA optics

dispersion in X

(DIPOLE)

focusing in Y

(QUADRUPOLE)

IC 195 MeV 36S + 208Pb, Θlab = 80o

E (a.u.)

ΔE

(a.u

.)

Z=16

Z=28

240 MeV 56Fe+124Sn, Θlab = 70°

Z=26

ΔE

(a.u

.)

∆E/E < 2% ΔZ/Z ~ 60

Application Proportional Scintillator Counter Hybrid detector !!!

UV High field

With noble gases 75-97 % of energy gained by electrons from electric field is converted into light

Relative gain per collected e-

charge

secondary light

light

Energy resolution for low radiation is 2 times better than with ordinary proportional counters

Low field

X 10

Geiger-Muller Counters

for High Electric Fields: Secondary avalanches are produced by photons emitted by excited atoms M ∼ 1010

Entire volume gas is involved in the process ⇒  Loss of information on space ⇒  Loss of information on energy (NO spectroscopy) ⇒  identical output

400-1200 V depending on the gas

Noble gases To avoid formation of negative ions being based on multiplication process

portable instruments for radiation monitoring

Dead Time The Geiger discharge stops when an high ions (+) concentration reduces the field E below the moltiplication threshold

50-100 µs

Second full pulse appears when ALL ions (+) are collected

to the cathode

⇒  G-M tubes are limited to application with low counting rates Method to improve effective counting range 1. Time-to-first-count-method: The high voltage is switched between two values:

-  Normal operating voltage -  lower value below avalanche threshold

1-2 ms

2. Self-Quencing: addition of quench gas (5-10%) with complex molecule absorbing UV photons (ethyl alcohol …)

Quenching gas is consumed (molecule disassociate) à Limit 109 counts