kott applied spectrocopy 2014_68_1108

9
Official Publication of the Society for Applied Spectroscopy Official Publication of the Society for Applied Spectroscopy 62/9 SEPTEMBER 2008 ISSN: 0003-7028 Determination of a Low-Level Percent Enantiomer of a Compound with No Ultraviolet Chromophore Using Vibrational Circular Dichroism (VCD): Enantiomeric Purity by VCD of a Compound with Three Chiral Centers Laila Kott, a* Jelena Petrovic, a Dean Phelps, a Robert Roginski, b Jared Schubert c a Takeda Pharmaceuticals International Co., Analytical Development Small Molecules, 40 Landsdowne St., Cambridge, MA 02139 USA b Eigenvector Research Inc., 3905 West Eaglerock Dr., Wenatchee, WA 98801 USA c Boston University, School of Law, 765 Commonwealth Ave., Boston, MA 02215 USA

Upload: laila-kott

Post on 23-Feb-2017

38 views

Category:

Science


5 download

TRANSCRIPT

Page 1: Kott applied spectrocopy 2014_68_1108

Official Publication of the Society for Applied SpectroscopyOfficial Publication of the Society for Applied Spectroscopy

62/9SEPTEMBER 2008ISSN: 0003-7028

Determination of a Low-Level Percent Enantiomer of aCompound with No Ultraviolet Chromophore UsingVibrational Circular Dichroism (VCD): Enantiomeric Purityby VCD of a Compound with Three Chiral Centers

Laila Kott,a* Jelena Petrovic,a Dean Phelps,a Robert Roginski,b Jared Schubertc

a Takeda Pharmaceuticals International Co., Analytical Development Small Molecules, 40 Landsdowne St., Cambridge, MA 02139

USAb Eigenvector Research Inc., 3905 West Eaglerock Dr., Wenatchee, WA 98801 USAc Boston University, School of Law, 765 Commonwealth Ave., Boston, MA 02215 USA

Page 2: Kott applied spectrocopy 2014_68_1108

Determination of a Low-Level Percent Enantiomer of aCompound with No Ultraviolet Chromophore UsingVibrational Circular Dichroism (VCD): Enantiomeric Purityby VCD of a Compound with Three Chiral Centers

Laila Kott,a* Jelena Petrovic,a Dean Phelps,a Robert Roginski,b Jared Schubertc

a Takeda Pharmaceuticals International Co., Analytical Development Small Molecules, 40 Landsdowne St., Cambridge, MA 02139

USAb Eigenvector Research Inc., 3905 West Eaglerock Dr., Wenatchee, WA 98801 USAc Boston University, School of Law, 765 Commonwealth Ave., Boston, MA 02215 USA

The chiral configuration of three of the four chiral centers in the

investigational drug MLN4924 is locked by an intermediate

(1S,2S,4R)-4-amino-2-(hydroxymethyl)cyclopentanol (designated

as INT1a). The intermediate INT1a is a key component to the

molecule, but its multiple chiral centers and lack of chromophore

make it challenging to analyze for chiral purity of the desired

enantiomer when it is contaminated with a small amount of its

undesired enantiomer. Vibrational circular dichroism (VCD) is a

technique that uses the infrared (IR) regions of the electromagnetic

spectrum and as INT1a contains IR active groups, we considered

using VCD to determine the chiral purity of INT1a. Since the VCD

spectra of enantiomers are of equal intensity and opposite in sign,

it was possible to construct calibration curves to detect the

presence of low levels of this compound in the presence of its

enantiomer. By normalizing the observed intensities of the VCD

signals with the observed IR spectra, a partial least squares model

was constructed having a root mean squared error of cross

validation of 0.46% absolute over a range of 97 to 99.9% pure

enantiomer (or 97–99.8% enantiomeric excess). This work demon-

strates that VCD can be used for the low-level detection of a

compound in the presence of its enantiomer and thus eliminates

the need for an ultraviolet chromophore and chromatographic

separation of the two enantiomers.

Index Headings: Multiple chiral centers; Mid-infrared; Quantitation;

Vibrational circular dichroism; VCD; Enantiomer; a-pinene; Partial

least squares; PLS.

INTRODUCTION

The investigational oncology drug MLN4924, whose

mechanism of action is the inhibition of the Nedd8

activating enzyme (NAE), is a complex molecule with

four chiral centers. Three of these centers are locked by

the intermediate INT1a. As seen in Fig. 1, INT1a has no

ultraviolet (UV) chromophore, which is the standard

detection used in liquid chromatographic (HPLC) sepa-

ration techniques for chiral purity assays. The molecule

is small enough and possibly volatile enough for chiral

gas chromatographic (GC) analysis, but amine functional

groups make choosing a column for GC analysis difficult

because these compounds tend to tail and interact with

GC columns in such a way as to make integration and

quantitation difficult.1 There are also fewer choices of

chiral columns for analysis by GC.

Analysis of this intermediate is possible by more

traditional approaches, such as derivatization with

Marfey’s Reagent2 (1-fluoro-2,4-dinitrophenyl-5-L-ala-

nine amide, FDAA), followed by HPLC analysis. How-

ever, as often happens with derivatization reactions, the

validation of such methods is challenging due to

complicated sample preparation and proving complete

derivatization.3,4 Since this compound is nonchromo-

phoric, chiral, and is a nitrogen-containing compound,

HPLC analysis using a polarimeter, a circular dichroism

(CD) detector, and a chemiluminescence nitrogen

detector (HPLC-CLND) were tried. Chromatography was

possible with the CLND, but no peaks were observed

using the two chiral detectors which covered a wave-

length range of 230 to 700 nm. In an effort to push to

lower limits, an alternative technique was considered.

A search for other possible methods for determining

the chiral purity of INT1a led to a lesser-known

spectroscopic technique based on CD in the infrared

region. This technique is known as vibrational circular

dichroism (VCD) and most commonly makes use of the

mid-IR region from 2000 to 800 cm�1 (5000 to 12 500

nm).5,6 Note: The entire IR region could theoretically be

used and spans 14 000 to 800 cm�1 (�715 to 12 500 nm).

To employ VCD, several experimental parameters had

to be considered. A suitable solvent that is transparent in

the region of interest is required, and to ensure that the

VCD peaks are detectable above the noise, the IR bands

of the solution to be analyzed should be at a concentra-

tion that will result in absorbance between 0.2 to 0.8.7

Similar to other spectroscopic techniques, an increase in

collection time yields higher signal-to-noise ratios and a

more robust dataset.8 Determination of the optimal

spectral region and collection time that is required for

each compound to obtain the desired spectral quality

depends on the functional groups and the flexibility of the

molecule. In general, the less flexible the molecule, the

shorter the collection time that is required for a

satisfactory signal-to-noise ratio. Therefore, due to its

ring structure and limited conformational freedom, INT1a

was a good candidate for VCD analysis.

Primer on Vibrational Circular Dichroism (VCD).Vibrational circular dichroism, like the CD measure-

ments in the UV (also known as electronic CD or ECD), is

Received 18 April 2013; accepted 4 April 2014.

* Author to whom correspondence should be sent. E-mail: Laila.Kott@

takeda.com.

DOI: 10.1366/13-07112

1108 Volume 68, Number 10, 2014 APPLIED SPECTROSCOPY0003-7028/14/6810-1108/0

Q 2014 Society for Applied Spectroscopy

Page 3: Kott applied spectrocopy 2014_68_1108

a spectroscopic technique which detects differences in

absorption of left and right circularly polarized IR and

near-IR light, where vibrational transitions in molecules

are observed.9 Only chiral molecules display CD/VCD

activity.9

DA ¼ AL � AR ð1Þ

Equation 1 shows how VCD is a measure of the

difference in absorption (DA) of left (AL) and right (AR)

circularly polarized IR light. The observed signals are

typically very small, roughly four orders of magnitude

smaller than the parent IR intensities.8

The current principal use of VCD in the pharmaceutical

industry is for the determination of absolute configura-

tion of chiral active pharmaceutical ingredients, which is

accomplished by comparison of a VCD measurement to

an ab initio calculation of a selected enantiomer.8

However, VCD is an absorption measurement, and its

intensity is linearly proportional to concentration and

path-length in accordance to Beer’s law, but unlike its

parent IR, the VCD intensity is also linearly proportional

to enantiomeric purity of the sample. This feature of VCD

allows direct measurement of the percentage of desired

enantiomer without a need for separation. Simulations of

real-time reaction monitoring have shown that it is

possible to follow and quantitate the conversion of one

enantiomer to the other.10

Since VCD is fundamentally an IR measurement, with

its abundance of bands, it lends itself to a chemometrics

analysis. Chemometrics allows for the use of a defined

wavelength range of interest, takes into consideration

many more data points, and incorporates these points

along with the variability (error) in the data into the

analysis. Chemometric analysis is especially useful

during manufacturing (for in-process control measure-

ments) as it can account for cell variability, analyst

variability, temperature fluctuations, and instrument

variability. Chemometrics use, therefore, leads to a

more robust and sensitive approach to analysis of

complex VCD datasets.

MATERIALS AND METHODS

Infrared (IR) and Vibrational Circular Dichroism(VCD) Measurements for a-Pinene Study. The IR and

VCD spectra of a-pinene were measured on a BioTools

ChiralIR-2X spectrometer (Jupiter, FL) equipped with

DualPEM (two photoelastic modulators) and a Syncrocell

rotating sample cell holder. The liquid R- and S-pinene

samples (Sigma-Aldrich, St. Louis, MO) used were of

several different assay and enantiomeric purities and

were mixed volumetrically at different R : S ratios, up to

12.5% R. These solutions were placed in a BaF2 cell with

a path length of 50 lm and run both holding the sample

stage stationary and rotating it counterclockwise at

3 RPM. Data were acquired at a resolution of 4 cm�1,for 20 min. No baseline corrections were performed, and

chemometric analysis was carried out using SOLO

software (Eigenvector Research Inc., Wenatchee, WA).

Different calibration curves prepared on different days

by separate analysts were collected and all the data

combined so as to have a sufficiently large calibration

set and to account for day-to-day and person-to-person

variability within the chemometric model.

Ab Initio Calculations for Theoretical Spectra ofINT1a, its Enantiomer, and Diastereomers. For all of

the stereoisomers, an initial search of conformational

space was performed with Molecular Operating Environ-

ment (Molecular Operating Environment (MOE) 2012.10,

Chemical Computing Group) software using molecular

mechanics (MM), stochastic search, and a 7 kcal/mol

FIG. 1. Structure of (a) MLN4924, (b) the intermediate INT1a where the chirality is locked, and (c) R- and S-a-pinene.

APPLIED SPECTROSCOPY 1109

Page 4: Kott applied spectrocopy 2014_68_1108

energy window. All the conformers output from MM were

submitted for density functional theory (DFT) minimiza-

tion. After DFT minimization, Boltzmann distribution, and

redundancy check, four conformers were within 2 kcal/

mol from the lowest energy conformer for INT1a, while

two conformers were within 2 kcal/mol from the lowest

energy conformer for INT1c, INT1e, and INT1g.

The output from MOE was inputted into Gaussian09 to

perform spectral calculations for each conformer using

B3LYP/6-31G basis set. The conformers’ spectra were

then combined into a final spectrum using Boltzmann

distributions to generate a theoretical VCD and IR

spectrum for each enantiomer and diastereomer. Com-

pare VOA version 1.1 by BioTools was used to generate

the final spectra. The calculations were performed in

vacuo.

Infrared (IR) and Vibrational Circular Dichroism(VCD) Measurements for INT1a Study. The IR and

VCD spectra were measured on a BioTools ChiralIR-2X

spectrometers equipped with DualPEM. The samples

were placed in a BaF2 cell with a path length of 100 lm.

For the first calibration set, data were acquired at a

resolution of 8 cm�1, for 10 h. In order to determine the

limit of detection for the undesired enantiomer of INT1a,

a calibration curve was constructed at 100 mg/mL total

compound, ranging from 97.0:3.0 to 99.9:0.1 (desired :

undesired) enantiomer ratio of INT1a in methanol-d4(Sigma-Aldrich, St. Louis, MO). Chemometric analysis

was carried out using PLS_Toolbox (Eigenvector Re-

search Inc.) running on Matlab. Two sets of samples (ee1

and ee2) were made on two separate days to account for

variability in the technique.

A second curve was constructed at 125 mg/mL total

compound, ranging from 80:20 to 100:0 (desired : unde-

sired) enantiomer ratio of INT1a in methanol-d4 using a

different instrument to show reproducibility. Data were

acquired at a resolution of 4 cm�1, for 24 h. Data were

normalized against their molar concentrations, and all

chemometric analyses were carried out using SOLO

software (Eigenvector Research Inc.).

RESULTS AND DISCUSSION

Considering that CD spectra have been used for the

quantitation of enantiomers in the UV,11 it was thought

that the same could be applied to VCD for the

determination of the amount of an enantiomeric impurity.

The VCD spectra of enantiomers are opposite in sign;

therefore, when added together, the mixture of enantio-

mers result in a spectrum that has a lower intensity than

the spectrum of a single pure enantiomer. In the case of

a racemic mixture, the equal and opposite-signed

enantiomer spectra cancel each other out, resulting in

no VCD spectrum.12 Providing the total concentration of

all the compounds remains the same; however, the

resulting IR spectrum should show constant intensities

for the mixture of enantiomers.

Percent Enantiomer Study with a-Pinene. Since

investigational compounds such as INT1a and its

enantiomer are not readily available in large quantities,

a-pinene (see Fig. 1) was chosen for the first study, as

both the R and the S are readily available at known

purities. As this compound was a liquid, samples were

prepared by making volumetric dilutions of the R

enantiomer with the S enantiomer. Each concentration

was analyzed both while the sample was stationary and

while rotating. Three sets of samples were made fresh

daily and analyzed on different days (using both the

stationary and rotating cell configurations) by different

analysts, to ensure that there was sufficient statistical

variability in the data for a successful chemometric

analysis. All concentrations were corrected for overall

and entiomeric purity, for both the R- and the S-pinene.

To construct the chemometric model, six a-pinenedatasets were combined (three measured with a

stationary cell, three with a rotating cell). The IR peaks

of a-pinene at 1265 and 1285 cm�1 showed a change in

intensity with increasing R enantiomer. This is indicative

of a non-chiral impurity since it is apparent in the IR

spectrum, and as such, the corresponding peak in the

VCD spectrum was removed prior to chemometric

analysis. The data range used for chemometric analysis

was from 1345 to 940 cm�1, with the peak from 1285 to

1235 cm�1 excluded (see Fig. 2). For a-pinene, mean

centering preprocessing was used. No further process-

ing was required as the pinene samples were neat

liquids with the same physical characteristics, and since

there was no interference from water vapor in the IR

spectra, there was no need to correct for it during the

data analysis of these experiments.

A partial least squares (PLS) model was built with a

total of two latent variables. Venetian blind (with seven

data splits) cross validation was performed (48 samples

in the dataset). The root mean squared error of

calibration (RMSEC) was 0.62%, while the root mean

squared error of cross validation (RMSECV) was 0.69%.

This indicates that the enantiomeric purity of S-a-pinenecould be determined to 60.69%. Figure 3 shows the

results of the PLS model for predicted-to-measured

(RMSEC) and for cross-validation predicted-to-measured

(RMSECV) enantiomeric purity.

The success of this model can be shown by using it to

predict values of unknowns. We collected two more sets

of data and used them as a prediction set. We applied

FIG. 2. The VCD spectral range used for modeling the a-pinene data

(1345 to 940 cm�1)

1110 Volume 68, Number 10, 2014

Page 5: Kott applied spectrocopy 2014_68_1108

FIG. 3. Results of the PLS model to predict a-pinene enantiomeric purity. The top plot (a) shows predicted versus measured, and the bottom (b)cross validated prediction versus measured.

FIG. 4. A plot showing the predicted values of a test dataset as determined from the chemometric model for a-pinene.

APPLIED SPECTROSCOPY 1111

Page 6: Kott applied spectrocopy 2014_68_1108

our model to them, and Fig. 4 shows that the test values

(i.e. the unknowns) can be predicted using the model

developed above.

These results indicated that quantitation to low

enantiomeric levels by VCD is possible, and the work

shifted from our model compound to 1NT1a.

Theoretical Spectra of INT1a, its Enantiomer, andDiastereomers. Prior to starting the enantiomeric purity

study, we had to verify that we could differentiate

between the enantiomers and diastereomers; therefore,

the theoretical infrared (IR) and VCD spectra were

generated. A comparison of the VCD spectra of the

diastereomers is shown in Fig. 5. In theory, we could

choose bands or a small spectral range, specific to the

enantiomeric pair in question to determine chiral purity.

We could also quantify any other diastereomers off of

their unique bands. However, in this case, since we were

dealing with known standards, we simply verified that

the measured VCD spectrum matched the theoretical to

ensure that we had the proper components.

The success of developing a model for a-pinene and

confirmation that the enantiomers and diastereomers

are spectroscopically unique led to the investigation to

determine the lower limit to enantiomeric detectability

for INT1a by VCD. A chemometric analysis approach was

also applied to the INT1a data. If the levels of detection

were found to be low enough, theoretically this non-

chromatographic method could be used for either

FIG. 5. Theoretical IR and VCD spectra of INT1a, INT1c (RRR), INT1e (RSR), and INT1g (SRR). The IR spectra show that they are all the same;

however, each displays a unique VCD spectra.

1112 Volume 68, Number 10, 2014

Page 7: Kott applied spectrocopy 2014_68_1108

(offline) in-process testing or release testing of the INT1a

intermediate without any derivatization.

The compounds in this series were salts and were,

therefore, insoluble in most of the typical VCD solvents

like CCl4 and CDCl3 and had only a 0.2 mg/mL solubility

in methylene chloride. To determine a low level of

quantitation for INT1a, we chose methanol as our solvent

for two reasons. Firstly, it was the solvent that our

compound was most soluble in, and it was available in

deuterated form at high purity. Secondly, it removed the

exchangeable protons from the spectrum, thereby

simplifying the final spectrum. Since all solutions were

made up in methanol-d4, any possible solvent effects

would be present in all solutions, normalizing any

effects. As the purpose of these experiments was to

push the level of quantitation as low as possible, not to

characterize the molecule, the criteria for choice of

solvent were (1) high concentration and (2) having an

appropriate section of the VCD spectrum that would be

amenable to quantitation by chemometrics.

In the first round, two sets of calibration curves (ee1

and ee2) were made and analyzed on two different days

to ensure that there was sufficient statistical variability in

the data for a successful chemometric analysis. The

process of analysis involved the following steps:

(1) Finding the best way to compensate for atmospheric

water vapor in the IR spectra.

(2) Learning and accounting for the baseline fluctua-

tions in VCD spectra.

(3) Evaluating the best spectral regions for analysis.

(4) Normalizing the IR and VCD spectra for molar

concentration and path length fluctuations. Vibra-

tional circular dichroism is a product of molar

concentration multiplied by the percentage of

enantiomer and path length, so correction is a

critical step.13

A unique feature of this VCD instrument was the

collection of IR spectra simultaneously with VCD spectra.

This allows for the VCD spectra to be corrected for minor

changes in concentration.

Based on the spectra collected, the best region for IR

normalization was 1234–1585 cm�1 and for VCD analysis

was 1234–1474 cm�1. This region corresponds to IR

bands associated with some C–H, C–O, and C–N modes.

Since peaks in this region show up in the VCD spectra as

well, these modes are considered to be representative of

the bonds connected to the chiral centers of interest,

thereby making this region appropriate for analyses. The

FIG. 6. (a) IR spectra (1430–1630 cm�1) and (b) VCD spectra (1240–1590

cm�1) for datasets ee1 and ee2. The vertical dashed line in both plots is

coincident with the artifact observed in some of the VCD spectra at 1512

cm�1. Note that this feature is absent in the IR spectra. The sharp

features in the IR spectra between 1480–1580 cm�1 are attributed to

water vapor; negative features in the ee2 set indicate that the

background measurement contained more water vapor than the

subsequent spectral measurements.

FIG. 7. Results of the PLS model to predict INT1a enantiomeric purity

for the first calibration set. The top plot (a) shows predicted versus

measured, and the bottom (b) cross validated prediction versus

measured.

APPLIED SPECTROSCOPY 1113

Page 8: Kott applied spectrocopy 2014_68_1108

region for VCD is smaller due to the fact that the

instruments used exhibited an artifact in the noise

spectra (even when no cell is present), which then

manifested itself as small peak in the VCD spectra.

Figure 6 shows that the artifact peak in the VCD spectrum

has no equivalent in the IR spectrum. If it was a true peak

due to the sample, it would be visible in both.14

After removing water vapor absorption lines in the IR

spectrum and applying baseline correction in the VCD

spectra using Savitzky–Golay derivatives, a partial least

squares (PLS) model was built with a total of two latent

variables. The root mean squared error of calibration

(RMSEC) was 0.24%, while the root mean squared error

of cross validation (RMSECV) was 0.46%. Leave-one-out

cross-validation was performed, which is appropriate

given the small size of the dataset (10 samples in the

dataset). Figure 7 shows the results of the PLS model to

predict enantiomeric purity for both the RMSEC and

RMSECV.

The higher value of RMSECV compared to RMSEC

directly indicates that the latter is biased by the presence

of all samples in the calculation of the error. Given that

this analysis is multivariate in nature, and there are no

well-established standard methodologies to extract

traditional univariate figures of merit such as limit of

detection (LoD) and limit of quantitation (LoQ) from this

method, we will focus our attention on the best available

metric for future performance: RMSECV. In this scenario,

the value of RMSECV can be thought of as a degree of

uncertainty around the predicted analytical value, a type

of one standard deviation (1r).15

Using the root mean squared error of cross validation

approach, we have an uncertainty of 0.46% on any future

values determined from this calibration set. Since

mixtures with low concentrations of the undesired

enantiomer were used to develop this model (from 3%down to 0.1%), we could theoretically expect that we

could detect down to 0.1% undesired enantiomer, with an

error of 60.46%, or more accurately a purity of 99.9%with an error of 60.46%. A level of detection with that

confidence is appropriate for in-process testing.

Applying what was learned from the first INT1a

calibration set, a second round of data was collected.

To improve on the signal-to-noise in the VCD spectra, the

data was collected at slightly higher concentrations at a

resolution of 4 cm�1 and for a total of 24 h. Based on the

spectra collected, the best region for VCD analysis was

1275–1500 cm�1.Since the cell compartment is open to the environ-

ment, instead of using the IR spectra for normalization

and correcting for atmospheric vapor, normalization was

done against the molar concentrations, as these solu-

tions were prepared in the testing laboratory and the

concentrations were known. This was done though

preprocessing in SOLO. Once corrected, a partial least

squares (PLS) model was built with a total of two latent

variables. The RMSEC was 1.1% while the RMSECV was

2.0%. Leave-one-out cross-validation was performed,

which is appropriate given the small size of the dataset

(eight samples in the dataset).

Due to the availability of material in the second round

of experiments with INT1a, only a few datasets were

collected, and as such, the model was not as well

defined or as rugged as the first set. The trend, however,

is the same that, as we generate more data, the model

can be pushed to be more reliable at lower concentra-

tions of enantiomer.

CONCLUSIONS

This study was an initial attempt to estimate the

sensitivity of VCD for determination of the percentage of

undesired enantiomer. For a-pinene, our model showed

that we could differentiate and predict levels of enantio-

meric impurities down to about 0.69%. When considering

that some of the S-pinene used in this study had 2.7% R

enantiomer in it, a level of uncertainty of 0.69% appears

sufficient for this compound. It should be noted that one

of the biggest challenges during these experiments was

obtaining pure pinene and accurate chiral purity data.

This became a serious consideration while running the

experiments and also highlighted another possible use

for this type of analysis—raw material testing.

For the first experiments with INT1a, the RMSECV

obtained was 0.46% for the PLS model (two latent

variables), and for the second set of experiments the

RMSECV was 2.0% for the PLS model (two latent

variables); however, our dataset was smaller than the

first set of experiments, and as such, the model

generated could not be as robust. These values are

comparable to an adequate sensitivity level for an in-

process test (target limit was 0.5 to 1.0%); however, to

replace the chromatographic testing, it would be

desirable to get closer to a level of 0.05 to 0.1%. If we

can get down to these lower levels, this technique could

also be used as a release test, replacing chromatogra-

phy completely. Given that the scope of this study did not

include much in the way of parameter optimization, it is

not unreasonable to expect an improvement in perfor-

mance that would make this measurement more com-

parable to the gold standard.

The data from all the INT1a experiments could theoret-

ically be combined, but as we were interested in the

reproducibility from site to site, we did not combine them

at this time. However, for this technique to work outside of

the research lab, calibration models must be transferable,

and the differences in instrumentation have to be negated.

This is a topic we will be researching further.

A final note on instrumentation: The current VCD

optical configuration is open to the environment, and the

nitrogen sweep is not sufficient to remove any atmo-

spheric effects (i.e., water vapor). It would be our

recommendation that, if instrument manufacturers are

interested in their equipment being used for quantitation,

as presented here, that an option for a closed purged cell

be available.

ACKNOWLEDGMENTS

The authors would like to thank Ashley McCarron and Ian Armitage

for synthesizing the compounds and Fred Hicks and Beth Piro for

sharing their experiences with these molecules. The authors would

also like to thank Larry Nafie, Rina Dukor, and the team at BioTools for

their help in executing the experiments and introducing us to the

technique.

1. L. Kott, H.M. Chen. ‘‘Experimental Considerations in Headspace

Gas Chromatography’’. Pharm. Tech. 2010. 34(5): 76-79.

1114 Volume 68, Number 10, 2014

Page 9: Kott applied spectrocopy 2014_68_1108

2. R. Bhushan, H. Bruckner. ‘‘Marfey’s Reagent for Chiral Amino Acid

Analysis: A Review’’. Amino Acids. 2004. 27(3-4): 231-247.

3. G. Lunn, L.C. Hellwig. Handbook of Derivatization Reactions for

HPLC. New York: John Wiley and Sons, 1998. Pp. 639-644.

4. N.D. Danielson, P.A. Gallagher, J.J. Bao. ‘‘Chemical Reagents and

Derivatization Procedures in Drug Analysis’’. In: R.A. Meyers,

editor. Encyclopedia of Analytical Chemistry. Chichester: John

Wiley and Sons, 2000.

5. L.A. Nafie. Vibrational Optical Activity: Principles and Applications.

Chichester: John Wiley and Sons, 2011. Pp. 234-236.

6. S. Abbate, G. Longhi, E. Castiglioni. ‘‘Near-Infrared Vibrational

Circular Dichroism: NIR-VCD’’. In: N. Berova, P.L. Polavarapu, K.Nakanishi, R.W. Woody, editors. Comprehensive Chiroptical Spec-

troscopy: Instrumentation, Methodologies, and Theoretical Simula-

tions, Volume 1. New York: John Wiley and Sons, 2012. Pp. 247-273.

7. Y. He, B. Wang, R.K. Dukor, L.A. Nafie. ‘‘Determination of Absolute

Configuration of Chiral Molecules Using Vibrational Optical

Activity: A Review’’. Appl. Spectrosc. 2011. 65(7): 699-723.8. R.K. Dukor, L.A. Nafie. ‘‘Vibrational Optical Activity of Pharmaceu-

ticals and Biomolecules’’. In: Encyclopedia of Analytical Chemistry,

Online. John Wiley and Sons, 2006.

9. P.J. Stephens, F.J. Devlin, J.R. Cheeseman. VCD Spectroscopy for

Organic Chemists. Boca Raton: CRC Press, 2012. Pp. 20-24.

10. C. Guo, R.D. Shah, R.K. Dukor, X. Cao, T.B. Freedman, L. Nafie.

‘‘Determination of Enantiomeric Excess in Samples of Chiral

Molecules Using Fourier Transform Vibrational Circular Dichroism

Spectroscopy: Simulation of Real-Time Reaction Monitoring’’. Anal.

Chem. 2004. 76(23): 6956-6966.

11. H.G. Brittain. ‘‘Applications of Chiroptical Spectroscopy in the

Characterization of Compounds Having Pharmaceutical Impor-

tance’’. In: N. Berova, K. Nakanishi, R.W. Woody, editors. Circular

Dichroism: Principles and Applications. New York: John Wiley and

Sons;, 2000. Pp. 819–844.

12. G.K. Webster, L. Kott. ‘‘Method Development for Pharmaceutical

Chiral Chromatography’’. In: S. Ahuja, S. Scypiniski, editors.

Handbook of Modern Pharmaceutical Analysis. Burlington, MA:

Academic Press, 2011. P. 276.

13. R.K. Dukor, C. Guo, L.A. Nafie. ‘‘Reaction Monitoring of Chiral

Molecules Using Fourier Transform Infrared Vibrational Circular

Dichroism Spectroscopy’’. US Patent: US 7,378,283 B2. Filed 2003.

Issued 2008.

14. G. Yang, Y. Xu. ‘‘Vibrational Circular Dichroism Spectroscopy of

Chiral Molecules’’. Top. Curr. Chem. 2011. 298: 189-236.

15. K. Kjeldahl, R. Bro. ‘‘Some Common Misunderstandings in Chemo-

metrics’’. J. Chemom. 2010. 24(7–8): 558-564.

APPLIED SPECTROSCOPY 1115