manual and documentation - freiberg instruments
TRANSCRIPT
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
1 / 78
Manual and documentation
lexsyg
2012-10-12
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
2 / 78
We work constantly at the advancement of all measuring systems.
Copying or reprinting, even in extracts, is only allowed with a written approval from the
Freiberg Instruments GmbH. All rights according to copyright subject to Freiberg Instruments
GmbH.
Subject to alterations.
Freiberg Instruments GmbH
Am St. Niclas Schacht 13
09599 Freiberg
Germany
Phone: +49 3731 419 54 0
Fax: +49 3731 419 54 14
E-Mail: [email protected]
Internet: http://www.freiberginstruments.com
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
3 / 78
Table of Contents
Table of Contents .................................................................................................................. 3
Table of Figures .................................................................................................................... 6
1. Introduction .................................................................................................................... 8
1.1. Important Information ............................................................................................... 8
1.1.1. Declaration of Conformity ................................................................................. 8 1.1.2. Target Audience ............................................................................................... 8 1.1.3. Intended Application ......................................................................................... 8 1.1.4. Not intended Application .................................................................................. 9 1.1.5. Safety Installations ........................................................................................... 9 1.1.6. Impact of the Warning Notices ......................................................................... 9
1.2. Fundamental Advices .............................................................................................10
1.2.1. Preface ...........................................................................................................10 1.2.2. Security Warnings ...........................................................................................10 1.2.3. Commissioning and Operating Method ...........................................................10 1.2.4. Control of the Operating Status .......................................................................11 1.2.5. Decommissioning ............................................................................................11 1.2.6. Storage ...........................................................................................................11 1.2.7. Electricity .........................................................................................................12 1.2.8. Mechanics .......................................................................................................12 1.2.9. Hazardous Materials .......................................................................................12 1.2.10. High Temperatures .........................................................................................12
2. Equipment ....................................................................................................................13
2.1. Descriptions ............................................................................................................14
2.1.1. Module 1.1/1.2/1.3: The Measurement chamber .............................................17 2.1.2. Module 2.1. Beta Irradiation ............................................................................18 2.1.3. Module 2.2.: Alpha irradiation ..........................................................................18 2.1.4. Module 3: Heating unit ....................................................................................19 2.1.5. Module 4.1/4.2/4.3: Green/IR/blue excitation ...................................................19 2.1.6. Module 5.1.1/5.1.2/5.1.3/5.1.4: PMT Modules .................................................20 2.1.7. Module 5.2.1: EMCCD Unit .............................................................................22 2.1.8. Modules 6 .......................................................................................................23 2.1.9. Module 6.1.1: PMT optics adaption & filter changing unit ................................24 2.1.10. Module 6.1.2: Extended PMT optics adaption & filter changing unit ................24 2.1.11. Module 6.1.3: Reduced PMT optics unit ..........................................................25 2.1.12. Module 6.2.1: Imaging optics adaption & filter changing unit for TL/OSL
measurement position .....................................................................................25 2.1.13. Module 6.2.2: Imaging optics adaption & filter changing unit for RF
measurement position .....................................................................................26 2.1.14. Module 7.1: LED-based monochrome light irradiation unit ..............................27 2.1.15. Module 7.2: LED-based solar simulation unit ..................................................27
2.2. Installation ..............................................................................................................28
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
4 / 78
2.3. Operation ................................................................................................................28
2.4. Construction Drawing ..............................................................................................29
2.4.1. Electrical wiring ...............................................................................................30 2.4.2. Specification ....................................................................................................30
3. Software .......................................................................................................................31
3.1. How to get start .......................................................................................................31
3.2. Overview .................................................................................................................31
3.2.1. Application status ............................................................................................32 3.2.2. Menu bar .........................................................................................................32 3.2.3. Sample list ......................................................................................................32 3.2.4. Measurement flow ...........................................................................................33 3.2.5. First aid ...........................................................................................................33 3.2.6. Main window ...................................................................................................33 3.2.7. Hardware module status .................................................................................33
3.3. Connection and Initialization ...................................................................................34
3.4. Typical workflow – making a measurement .............................................................34
3.4.1. Load and unload a sample ..............................................................................34 3.4.2. Creating a sequence .......................................................................................36 3.4.3. Assigning a sequence to a sample slot ...........................................................37 3.4.4. Starting a measurement ..................................................................................37 3.4.5. Handle measurement data ..............................................................................38
3.5. Special hardware characteristics .............................................................................39
3.5.1. EMCCD-camera ..............................................................................................39 3.5.2. Spectrometer ..................................................................................................40
3.6. Handling particular cases ........................................................................................40
3.6.1. Change the IP-address ...................................................................................40 3.6.2. Editing cooling power and evacuation time ......................................................41 3.6.3. Breaking a measurement ................................................................................41 3.6.4. Initialization after an Incident – First aid menu .................................................42
4. Service and Repair .......................................................................................................44
4.1. Service ....................................................................................................................44
4.1.1. Checking the system pressures ......................................................................44 4.1.2. Lift up the measurement chamber ...................................................................45 4.1.3. Cleaning the measurement chamber ...............................................................46 4.1.4. Cleaning the chamber windows .......................................................................47 4.1.5. Exchanging detection filter ..............................................................................50 4.1.6. Access to internal sample loading/unloading unit ............................................51 4.1.7. Heater exchange .............................................................................................51
4.2. Spare parts .............................................................................................................52
5. Physical Background ....................................................................................................53
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
5 / 78
5.1. Literature ................................................................................................................53
Attachment: Radiation source documentation ......................................................................54
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
6 / 78
Table of Figures
Figure 1: lexsyg modular luminescence reader ....................................................................13
Figure 2: technical overview of lexsyg ..................................................................................14
Figure 3: sample handling ....................................................................................................15
Figure 4: sample loading lid ..................................................................................................16
Figure 5: automated detector changing unit, with up to four detectors ..................................16
Figure 3: overview on the different module configuration options .........................................18
Figure 7: OSL irradiation unit ................................................................................................20
Figure 8: overview on the most widely used detection windows [1].......................................23
Figure 9: figure technical drawings: see enlarged version at the end of the document .........29
Figure 10: Electrical wiring ...................................................................................................30
Figure 11: Lexsyg Studio link................................................................................................31
Figure 12: start screen of the application ..............................................................................31
Figure 13: application status .................................................................................................32
Figure 14: menu bar .............................................................................................................32
Figure 15: sequence library ..................................................................................................36
Figure 16: sequence editor ...................................................................................................37
Figure 17: measurement status ............................................................................................37
Figure 18: data view .............................................................................................................38
Figure 19: EMCCD control ...................................................................................................39
Figure 20: spectrometer control ............................................................................................40
Figure 21: options menu .......................................................................................................41
Figure 22: first aid menu .......................................................................................................42
Figure 23: system pressures overview .................................................................................44
Figure 24: checking system pressures .................................................................................45
Figure 25: measurement chamber screws ............................................................................45
Figure 26: lift the measurement chamber .............................................................................45
Figure 27: manual moving of the sample arm .......................................................................46
Figure 28: plugs of the OSL-unit ...........................................................................................47
Figure 29: screws of the OSL-unit ........................................................................................47
Figure 30: moving the OSL-unit ............................................................................................47
figure 32: overview of the OSL-unit.......................................................................................48
figure 33: removing the plugs of the extra TL position ..........................................................48
figure 34: fastener of the extra Tl position .............................................................................48
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
7 / 78
figure 35: opening the extra TL position ................................................................................49
figure 36: schematic of the filterwheel ...................................................................................50
figure 37: opening for changing filters ...................................................................................50
figure 38: location of the loading/unloading unit ....................................................................51
figure 39: removing the loading/unloading unit .....................................................................51
figure 40: maintenance panel for changing the heater unit ...................................................52
figure 41: removedmaintenance panelwith visibleheater unit ................................................52
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
8 / 78
1. Introduction
1.1. Important Information
1.1.1. Declaration of Conformity
All lexsyg equipment from Freiberg Instruments is conform to the European legal provision
about the electromagnetic compatibility (89/336/EWG), the directive of the modification of CE
marking (93/68/EWG), the EG machine directive (2006/42/EG) and the EG low tension
directive (73/23/EWG).
The CE sign is arranged on the specification plate. Please attend on the obligatory national
and local regulations!
Through a non-coordinated modification of the equipment this declaration loses its validity.
1.1.2. Target Audience
The manual is written for the operating personal of the lexsyg(luminexcence excitation
system for geological and archaeological applications).
The operators should have the essential professional competence for all work with the
equipment.
The personal is obliged to read and completely understand the manual.
The documentation ought to be placed in reach of the equipment.
1.1.3. Intended Application
• The handler is in charge to only use the lexsyg equipment in the descript manner.
• The handling of the measuring equipment is only allowed subjected to the conditions
named:
– in this manual
– on the specification plate
– in the technical specification corresponding to the respective brief.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
9 / 78
• The lexsyg equipment is used for luminescence measurements either by thermal (TL),
optical (OSL) or radiation (RF) stimulation. It is configured for the measurement of certain
materials.
1.1.4. Not intended Application
All applications notwithstanding the technical data on the specification plate or the terms
named in the contract of delivery and the usage with missing or damaged safety installations
is prohibited.
1.1.5. Safety Installations
The following arrangements serve the security of the operating personal:
• electrical access with grounding conductor and safety plug
• laser protection service
• mechanical safety doors with safety key
• electronic safety precautions
• software safety routines
• for radiation safety precautions see attachment one
Without any of these installations a usage of the equipment is forbidden.
1.1.6. Impact of the Warning Notices
Observe the warning notices! They are distinguished:
Warning notice:
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
10 / 78
1.2. Fundamental Advices
1.2.1. Preface
It’s necessary to consider the warning notices. The defiance can lead to damage of healthy
and property.
Service and repairing of the equipment is only allowed to skilled personal.
The removal of the whole equipment or single parts should be carried out by consideration of
the local legal regulations.
1.2.2. Security Warnings
1. Warning notice: Usage of LASER radiation! The radiation source is placed in the
lower department of the system. Don’t expose any body parts, e.g. eyes, arms, to the
radiation! This will lead to irreversible damages. All works at this partition of the
equipment should only be carried out by members of Freiberg Instruments.
2. Warning notice: Usage of alpha and beta radiation source! All legal precautions
have to be followed and special training as well as constant observation of operator
depending on the local regulations is required. For details see attachment one.
1.2.3. Commissioning and Operating Method
The equipment is constructed for luminescence measurements of mineral grains.
Transport:
Lexsyg systems are well prepared for transport. Open the cases carefully.
Warning: secure the system from tilting. It has a high balance point.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
11 / 78
Commissioning occurs by:
1. Positioning the equipment on a horizontal area.
2. Connecting with the main supply, as shown on the “lexsyg supply” document.
Depending on configuration and intended use: nitrogen (helium), air pressure,
vacuum, electrical power, LAN
3. Switching-on the master controller
4. Starting the connected personal computer
5. turn on the tool
Keep in mind to place the sample materials on the intended position.
Warning notice: Defiance can lead to mechanical damages of samples and equipment!
1.2.4. Control of the Operating Status
Keep an eye on the operating status! It’s displayed on the left upper corner of the monitor
and on the display.
1.2.5. Decommissioning
Decommissioning occurs by:
1. Shutting down the computer
2. turn the tool off
3. Disconnecting from the main supply
1.2.6. Storage
Only store the equipment in closed dust-free rooms.
Temperature should lie between 5 and 40 °C, relative humidity must not be higher than 80%.
Warning notice: Disconnect from the power supply when store the equipment for a longer
period of time.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
12 / 78
1.2.7. Electricity
The lexsyg equipment runs with 110V/230 V AC, 50/60 Hz. Internal low tension in the range
of +/-5 to +/-48 V DC is used.
For the connection with the main supply a grounding conductor is necessary (DIN VDE 0100-
410 (IEC 60364-4-41)).
Observe the local regulations! Electrical circuit points should be periodically checked (DIN
EN 0105, DIN EN 0702, BGV A2).
1.2.8. Mechanics
Inappropriate operation can cause damage of property and injuries.
Use the equipment only for compulsory measuring!
External tensions and vibrations should not be applied to the equipment!
For a better cooling of the equipment don’t place any other hardware at intervals of 20 cm.
1.2.9. Hazardous Materials
An assignment of hazardous materials is not intended.
1.2.10. High Temperatures
The equipment can be warmed up by electrical lost heat. It’s not allowed to excess a
temperature of 40 °C!
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
13 / 78
2. Equipment
LEXSYG is a luminescence reader designed and constructed by Freiberg Instruments in co-
operation with luminescence research and dating laboratories [ 1].
It is a highly flexible modular system for measuring different kinds of luminescence using
thermal, optical or ionizing radiation stimulation (i.e. TL, OSL/IRSL and RL) especially in
luminescence dating (geo-/archaeochronometry), mineralogical research, solid state
dosimetry, radiation control and material sciences.
A basic measurement chamber can be equipped with a variety of modules for irradiation,
luminescence stimulation and detection. Thus a versatile luminescence research instrument
can be built up. However, each modular unit incl. its unique features can be used in a more
conventional arrangement, e.g. a basic TL or TL/OSL reader.
Figure 1: lexsyg modular luminescence reader
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
14 / 78
2.1. Descriptions
Figure 2: technical overview of lexsyg
Sample handling
Sample storage and measurement is separated in order to prevent crosstalk. Samples are
transported by a pneumatic handling system from the storage position to the measurement
chamber. On top of the measurement chamber several excitation devices and detectors are
arranged in a modular way to allow a flexible tool operation. For detailed options see the
“module” descriptions. After 10 million handling steps the positioning tolerance rises up to
0.02 mm according to supplier specifications, equals a constant operation of above 30 years.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
15 / 78
Figure 3: sample handling
Loading
Several samples can be loaded at the same time by using the loading lid (Figure 4). Samples
can be exchanged at almost any time during the measurement procedure. Measurement
sequence selection has to be done by operation software. Loading position assignment can
be done by operation software as well as using the buttons at the front side of the system.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
16 / 78
Figure 4: sample loading lid
Pneumatics
Pneumatics is used for sample handling from storage wheel to the measurement position
and for opening the radiation source shutters. All components are chosen from standard
automation companies with mean times between failures of several hundred thousand
cycles. Valves can be found at the left side of the tool in the installation section.
Electronics
Two electronic departments are used. One for power conversion from AC to DC in the back
of the tool in the installation section. Main programmable electronic for tool operation can be
found at the front side of the tool in the installation section.
Figure 5: automated detector changing unit, with up to four detectors
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
17 / 78
2.1.1. Module 1.1/1.2/1.3: The Measurement chamber
Each lexsyg instrument consists of a measurement chamber prepared for modular
extensions for measuring different kind of luminescence.
prepared for vacuum (≤ 10 mbar) and inert-gas purging
80 samples storage with automatically sample changer
independent operation of chamber and sample-disc reservoir
adapts beta- and alpha irradiation units, different units for luminescence light detection and for simulating solar light
low radiation emission
negligible radiation exposure of stored samples and irradiation source leakage dose effects
easy handling and safe service
display for load/unload, sequence and status information
1x TL/OSL measurements
1x Beta-irradiation + RF measurements
1x additional measurement position** (module 1.2/1.3)
1x additional irradiation position (Alpha/Beta) (module 1.2/1.3)
1x “solar light” irradiation (LED or fiber-coupled simulator)***
1x Detector Wheel [changes up to 4 detection units (CCD(s) / PMT(s)
/ fiber-coupled detectors) between TL/OSL and RF (Beta
Irradiation) position] (module 1.3)
* Up to 2 detector positions can hold an automatic filter changer ** The additional measurement position is not suitable for a second OSL stimulation unit and has only limited filter-change capacity. It is useful for TL, ITL or RP measurements, with an additional detector (e.g. special PMT, light guide coupled spectrometer) in fixed configuration (not automatically changeable). *** If not used for a light source an additional detection unit can be adapted at fixed position
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
18 / 78
Figure 6: overview on the different module configuration options
2.1.2. Module 2.1. Beta Irradiation
lexsyg instruments can be equipped with 2 irradiation units. The Beta-irradiation unit is
special made for RF measurements and delivers its highly uniform irradiation field also to
samples for other luminescence measurements (TL, OSL, RP, ...).The Irradiation units are
delivered separately inclusive inbuilt radioactive sources.
to irradiate samples for TL, OSL etc. and for RF measurements
Radioisotope: Sr-90; activity: ≤ 1.8 GBq; dose rate ca. 0.08 – 0.1 Gy/s
high homogeneity of dose rate distribution at irradiation area:
≤ (± 2.5 %)) @ 8 mm diameter
≤ (± 3.5 %)) @ 10 mm diameter
includes shutter unit (minimum open/close cycle time 0.5 s)
easy and safe service for regular wipe-test
permission for handling irradiator must be sent to manufacturer at time of ordering
2.1.3. Module 2.2.: Alpha irradiation
The Alpha-unit is just designed for irradiation. The Irradiation units are delivered separately
inclusive inbuilt radioactive sources.
to irradiate samples (max. 10 mm diameter) for TL, OSL etc.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
19 / 78
recommended for dating techniques where the alpha-component has to be
determined (e.g. fine-grain, flint)
Radioisotope: Am-241
vacuum irradiation, fully software controlled
includes shutter unit (minimum open/close cycle time 0.5 s)
easy and safe service for regular wipe-test
permission for handling irradiator must be sent to manufacturer at time of ordering
2.1.4. Module 3: Heating unit
Thermoluminescence (TL), preheat (PH) of materials or elevated temperature measurements
(e.g. standard Quartz SAR-protocol) need a heater. LEXSYG luminescence readers can
also use this unit for elevated temperature irradiation (at ionizing irradiation and “solar-light”
position) and measurements (elevated temperature RL). Measurements are possible during
any part of a heating/cooling cycle, which may simply be a heating ramp but also consist of a
series of different steps. An extended software version (see software options section 12.)
includes further possibilities, e.g. programmable cooling and non-linear heating/cooling
functions.
temperature: room temperature (RT) to 700 deg. Celsius (higher temp. on request)
heating rate: 0.1 – 10.0 K/s (@ Tmax= 500 °C; 0.1-5.0 K/s (@ Tmax= 700 °C higher
heating rates on request
highly flexible step-wise programming of heating cooling cycles within system
parameters (i.e. temperature and heating/cooling rates)
Software options for further temperature control (e.g. non-linear functions, see 12.)
2.1.5. Module 4.1/4.2/4.3: Green/IR/blue excitation
lexsyg instruments can be equipped with an fiber-optics based OSL stimulation unit which
allows the independent, parallel use of up to 3 different IR, VIS or UV lasers for highly
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
20 / 78
uniform excitation by light. The power of each laser light-source at the sample can be
software adjusted constant or can be ramped (e.g. LM-OSL). Software options (see section
12.) allow further flexible use (e.g. non-linear CW functions, special pulsing modes, time-
resolved measurements).
unit with up to 3 laser-light sources
850 nm (max. 150 mW), 532 nm (max. 75 mW), 478 nm (max. 40 mW) laser
stimulation
uniformity of light irradiation < (± 5%) at 8 mm sample diameter
individual independent use of each laser also together with heater
light detectors are software-protected if automatically filter-changer are in use
Figure 7: OSL irradiation unit
Each laser is separated into eight fibres. Seven for homogeneous sample irradiation and one
for optical laser power stabilization. Detection takes place through the central main axis.
2.1.6. Module 5.1.1/5.1.2/5.1.3/5.1.4: PMT Modules
Each lexsyg instrument can hold at least 2 luminescence detection units (can be extended by
up to 4). PMT units with different spectral characteristics, EMCCD-camera based spatially
resolved (e.g for single-grain measurement) or wavelength resolved (luminescence
spectrometry) as well as special units (e.g. for potassium concentration determination of
feldspar samples) are available.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
21 / 78
Module 5.1.1: UV-VIS PMT Module
for standard applications in luminescence dating and dosimetry
detection unit with standard bialkaline cathode PMT (HAMAMATSU)
sensitivity 280-650 nm
peak sensitivity 420 nm ( 3.1 E5 counts s-1 pW-1)
dark counts typ. 100 cps
photon counter
counting linearity (random pulses, 10% loss) 6.0 E6 cps
electromechanical shutter, UV-VIS optics
Module 5.1.2: Extended UV-Vis PMT Module
for standard applications in luminescence dating and dosimetry
detection unit with standard bialkaline cathode PMT (Electron Tubes 9235Q)
sensitivity 160-630 nm
peak sensitivity: 30% quantum efficiency at 200 nm and 350 nm
dark counts typ. 300 cps
photon counter
counting linearity (random pulses, 10% loss): must be specified
electromechanical shutter, UV-VIS optics
Module 5.1.3: Red enhanced UV-Vis PMT Module
necessary/useful if standard UV-VIS PMT-modules yellow to red sensitivity is not
sufficient
useful e.g. for quartz(flint)/calcite orange-red or feldspar red luminescence
measurements
thermoelectric/air cooled GaAsP photocathode PMT (HAMAMATSU)
sensitivity: 300 – 720 nm,
peak-sensitivity: 40% quantum efficiency at 580 nm
cooling max. 35 deg. below RT
cooling time approx. 5 min
typ. 100 cps dark signal at 0°C
photon counting system
counting linearity (random pulses, 10% count loss) 1.5 E6 cps
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
22 / 78
electromagnetic shutter, UV-VIS optics
Module 5.1.4: Vis-NIR PMT Module
necessary for RF dose determination using the 865 nm potassium feldspar emission
useful/necessary for any other measurements where red and NIR sensitivity is
needed
thermoelectric/air cooled GaAs photocathode PMT (HAMAMATSU)
sensitivity: 380 – 890 nm,
peak-sensitivity: 12% quantum efficiency at 800 nm
cooling max. 35 deg. below RT
cooling time approx. 5 min
typ. 60 cps dark signal at -10°C (125 cps at 0°C)
photon counting system
counting linearity (random pulses, 10% count loss) 1.5 E6 cps
electromagnetic shutter, UV-NIR optics
2.1.7. Module 5.2.1: EMCCD Unit
The lexsyg can adapt light detectors based on charged coupled devices (CCD). EM (electron
multiplying technology) CCD cameras have a very high sensitivity which allows even the
detection and counting of single photons.
UV to NIR (200 – 1050 nm) image detection for spatially resolved / single-grain
measurements
works at LEXSYG TL/OSL and RF measurement position (automatically change
recommended 1.3)
512 x 512 pixel back-thinned, UV-coated EMCCD chip (electron multiplying – EM -
technology), 16 bit resolution, frame – transfer format
works also in traditional CCD mode (low noise)
max. frame rate: > 33 frames/second (full frame)
Gigabit Ethernet data interface
high quantum efficiency from UV to NIR :
ca. 75% @ UV Quartz OSL emission peak
ca. 85-90 % @ all VIS Quartz/Feldspar emission peaks
ca. 60% @ potassium feldspar NIR RF peak (865 nm)
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
23 / 78
8 mm diameter sample is imaged on the camera chip (ca. 16µm / pixel) using an
optics adaption and filter changing unit (6.2)
incl. electromechanical shutter
TE cooling min. -95 °C (-70 °C guaranteed for life), thermostatic precision +/- 0.05°C
Important: requires 6.2.1,
lexsyg measurement chamber 1.3 recommended
2.1.8. Modules 6
There is a wide variety of detection windows for different minerals. Depending on the
flexibility needs of the lexsyg and the different detection options a variety of optical systems
and automated filter changing combinations can be selected.
Figure 8: overview on the most widely used detection windows [ 11]
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
24 / 78
2.1.9. Module 6.1.1: PMT optics adaption & filter changing
unit
These optical modules are designed for use with PMT units only. They are best performed to
collect the luminescence light from the sample, suppress stray-light (OSL, PL) or thermal
noise (TL), filter it and transfer the light efficiently to the photo-cathode of the PMT.
recommended as minimum configuration to perform optimized TL/OSL/RF analyses
in dating and other applications
optics designed for high-efficiency PMT luminescence measurements
6-position filter-wheel (6.5 mm max. filter thickness)
7measurement functions are included (more information on request)
switches 6 luminescence detection windows software controlled
filter change (min. 300 ms ... max. 3 s) also within a measurement sequence
malfunctions / PMT-damage are software protected
no stray-light transmission outside the optical path
optimized filter-sets for all main quartz and feldspar minerals luminescence emissions
and such for other materials/applications are available
filter sets (interference, colour-glass optical filters) optimised for highest transmission
individual wave-bands and best blocking stimulation light (OSL) and/or thermal noise
(TL, black-body radiation)
2.1.10. Module 6.1.2: Extended PMT optics adaption & filter
changing unit
These optical modules are designed for use with PMT units only. They are best performed to
collect the luminescence light from the sample, suppress stray-light (OSL, PL) or thermal
noise (TL), filter it and transfer the light efficiently to the photo-cathode of the PMT.
recommended to perform highly optimized TL/OSL analyses in dating and other
applications
optics designed for high-efficiency PMT luminescence measurements
2 filter-wheels with each 6 positions and each 6.5 mm filter-space
15 measurement functions are included (more information on request)
switches 6 luminescence detection windows software controlled
filter change (min. 300 ms ... max. 3 s) also within a measurement sequence
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
25 / 78
no stray-light transmission outside the optical path
optimized filter-sets for all main quartz and feldspar minerals luminescence emission
and other materials/applications available
filter sets (interference, colour-glass optical filters) optimised for highest transmission
individual wave-bands and best blocking stimulation light (OSL) and/or thermal noise
(TL, black-body radiation)
modified detection windows can be ordered (includes software-adaption)
malfunction / PMT-damage software protected
2.1.11. Module 6.1.3: Reduced PMT optics unit
These optical modules are designed for use with PMT units only. They are best performed to
collect the luminescence light from the sample, suppress stray-light (OSL, PL) or thermal
noise (TL), filter it and transfer the light efficiently to the photo-cathode of the PMT.
configuration to perform optimized TL/OSL analyses in dating and other applications
optics designed for high-efficiency PMT luminescence measurements
1measurement function of 6.1.2 is included (more information on request)
malfunctions / PMT-damage are software protected
no stray-light transmission outside the optical path
optimized filter-sets for all main quartz and feldspar minerals luminescence emissions
and such for other materials/applications are available
2.1.12. Module 6.2.1: Imaging optics adaption & filter
changing unit for TL/OSL measurement position
The lexsyg can adapt imaging light detectors based on charged coupled devices; in
particular an EMCCD based module (5.2.1) is available for spatially resolved measurements,
incl. single-grain measurements. Such detection units need advanced optical design for best
imaging quality within a chosen wavelength-band to be measured which is set by the optical
filters detection- window. Imaging units also work with PMT detectors and thus assure best
performance if a LEXSYG measurement chamber with automatically detector change is in
use.
necessary for spatially resolved / single-grain TL/OSL measurements
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
26 / 78
adapts automatically detected wave-band (optical window filter set) for best image
quality
works with equivalent efficiency also with PMT detectors
2 filter-wheels with each 6 positions and each 6.5 mm filter-space
15 measurement functions are included
switches 6 luminescence detection windows software controlled
filter change (min. 300 ms ... max. 3 s) also within a measurement sequence
no stray-light transmission outside the optical path
optimized filter-sets for all main quartz and feldspar minerals luminescence emissions
and such for other materials/applications available
filter sets (interference, colour-glass optical filters) optimised for highest transmission
individual wave-bands and best blocking stimulation light (OSL) and/or thermal noise
(TL, black-body radiation)
modified detection windows can be ordered (includes software-adaption)
malfunction / detector-damage software protected
2.1.13. Module 6.2.2: Imaging optics adaption & filter
changing unit for RF measurement position
The lexsyg can adapt imaging light detectors based on charged coupled devices, in
particular an EMCCD based module (5.2.1) is available for spatially resolved measurements,
incl. single-grain measurements. Such detection units need advanced optical design for best
imaging quality within a chosen wavelength-band to be measured which is set by the optical
filters detection- window. Imaging units also work with PMT detectors and thus assure best
performance if a LEXSYG measurement chamber with automatically detector change is in
use.
necessary for spatially resolved / single-grain RF analyses
adapts automatically detected wave-band (optical window filter set) for best image
quality
works also efficiently with PMT detectors
6-position filter-wheel (6.5 mm max. filter thickness)
switches 6 luminescence detection windows software controlled
filter change (min. 300 ms ... max. 3 s) also within a measurement sequence
no stray-light transmission outside the optical path
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
27 / 78
optimized filter-sets for all main quartz and feldspar minerals luminescence emissions
and such for other materials/applications are available
filter sets (interference, colour-glass optical filters) optimised for highest transmission
individual wave-bands
6 user-defined detection window filter-sets are included
2.1.14. Module 7.1: LED-based monochrome light
irradiation unit
power LED-array based light irradiator, cooled less fan
optical power at the sample: dependent on wavelength
loop power regulated
standard colours
fully software controlled (power, irradiation time)
setting can be changed within a measurement sequence any time
special optics for highly uniform irradiation power at the sample
other colours on request
additional power-LED array heads (7.1-x+) can be ordered and changed manually
2.1.15. Module 7.2: LED-based solar simulation unit
UV, blue, green, amber, red, IR power-LED array based light irradiator, fan-cooled
light (365 nm, 462 nm, 523 nm, 590 nm, 625 nm, 850 nm) can be used for
monochromatic sample irradiation or individually mixed together for simulating e.g.
natural bleaching conditions
typical maximum optical power at the sample:
- UV 365 nm - >3 W ex. Diode, 65 mW
- RGB A (462 nm - 140 mW, 523 nm - 35 mW, 590 nm - 50 mW, 625 nm - 85 mW)
- IR 850 nm - >2 W ex. Diode, 170 mW
fully software controlled (power, irradiation time, mixing ratios)
setting can be changed within a measurement sequence any time
mixed light spectrum display
special optics for highly uniform irradiation power at the sample
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
28 / 78
2.2. Installation
Equipment is packed for shipment. Detectors are disassembled. A detail description how to
pack and unpack the equipment can be found in a separate report.
1. Open the cover and place the tool on a suitable lab table
2. Install the detectors
3. Connect the main supply needs, pressure, electricity and vacuum (optional)
2.3. Operation
Make sure the power cord and pressure is connected. See construction drawings for a
detailed overview where to find certain components.
1. Turn the system on.
2. Connect to the device from a PC with operation software
3. See software description on how to start a measurement.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
29 / 78
2.4. Construction Drawing
Figure 9: figure technical drawings: see enlarged version at the end of the document
Dimension: 71.6 x 103.1 x 85.1 cm
Weight: 90 - 130 kg (depending on options)
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
30 / 78
2.4.1. Electrical wiring
Figure 10: Electrical wiring
2.4.2. Specification
Laser depending on option: see data sheet for detailed laser safety specifications
Warning: Never open laser safety installations. Whenever opening any mechanical door in
the installation or excitation/detection section disconnect from main power supply or wear
safety goggles!
Wavelength [nm] Power [mW] Class DIN
855 <1000 3B IEC 60825-1:93 A2:2001
532 <300 3B IEC 60825-1:93 A2:2001
480 <100 3B IEC 60825-1:93 A2:2001
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
31 / 78
3. Software
3.1. How to get start
Figure 11: Lexsyg Studio link
The lexsyg measurement system is controlled by the Lexsyg Studio application. It is already
installed on the provided PC and all properties are set. The application can be started by
clicking the lexsyg Studio link (see Figure 11) on the desktop screen of the PC.
3.2. Overview
Figure 12: start screen of the application
After clicking the Lexsyg Studio link the application will start up, showing the screen in Figure
12. The start screen is divided in several different categories (1: Application status, 2: Menu
bar, 3: Sample list, 4: Measurement flow, 5: First Aid, 6: Main window, 7: Hardware module
status). Every Element is described in the following chapters.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
32 / 78
3.2.1. Application status
The connection button (Figure 13, red button)
indicates if the Lexsyg Studio application is
connected to the lexsyg measurement system
or not. By pressing the button, the connection
to the system can be created or interrupted. While not connected, it is not possible for the
system to process any operations. Never disconnect during an active measurement to
prevent any unwanted behavior.
The measurement button (Figure 13, blue button) indicates the current measurement status
of the lexsyg measurement system. It indicates if a measurement is currently running or the
system is in standby mode.
Further information on these elements is given in section 3.3 and 3.4.
3.2.2. Menu bar
The menu bar allows access to different software
functions. In the Analysis-menu completed
measurements are accessible (see section 3.4.5).
In the Controls-menu are control elements for
particularly hardware modules accessible. See section 3.5 for further information. The menus
Options and Log contain elements for particular cases and are handled in section 3.4.5.
3.2.3. Sample list
The sample list provides a set of functions for the user:
providing an overview of the samples loaded in the lexsyg measurement system
which samples are waiting for processing or are already processed
linking a sequence to a sample slot
load and unload sample discs
These functions are descripted in section 3.4.1and 3.4.3.
Figure 13: application status
Figure 14: menu bar
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
33 / 78
3.2.4. Measurement flow
This element provides an overview about the actual running sequence. The status frame
contains the name and number of the actual running step. Temperature and counts are
showing the actual temperature of the heating unit and the counts of an PMT during a
running measurement (see Figure 17).
3.2.5. First aid
The First aid button provides access to a number of functions not needed in normal work
flow. They are descripted in section 3.6.
3.2.6. Main window
The main window is the area where all open windows are positioned.
3.2.7. Hardware module status
This bar shows a symbol for every currently active hardware module. The possible hardware
modules are:
heating unit
alpha irradiation source
beta irradiation source
solar light simulator
red laser
green laser
blue laser
UV laser diodes
UV - Vis PMT
NIR PMT
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
34 / 78
EMCCD camera
spectrometer
When hover with the mouse above an icon a description of the according symbol appears.
3.3. Connection and Initialization
Before any operation with the lexsyg measurement system can be started, it is necessary to
connect and initialize the system. The First step is to power the measurement system and
press the on/off button (see Figure 2). Now click the connection button (see section 3.2).
The caption of the button will change to “connected” and the color will switch to green. If not,
proceed at section 3.6.1.
After connection, the measurement system must be initialized. Click the first aid button and
in the appearing window the button Reset Lexsyg and initialize new (Figure 22 – 1). The
button changes its color to red and in the Application status button the caption switches to
“resetting and initializing” After the initialization, the button changes its color back to grey and
the connection button is labeled “start measurements”.
The system is now ready for processing tasks.
3.4. Typical workflow – making a measurement
The work with the lexsyg measurement system consists of a number of standard tasks
repeated in constant order. The following section describes a typical workflow which will be
aligned by a customer but show a good first workflow to get acquainted with the
measurement system.
3.4.1. Load and unload a sample
An important task concerning the work with the lexsyg measurement system is to load or
unload sample discs. It is the only task in which the user must interact with the hardware unit.
All other tasks will only be executed with the Lexsyg Studio software.
Loading and Unloading sample discs is only possible while the system is initialized and idle,
which means no measurement is actual running.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
35 / 78
The lexsyg measurement system provides space for up to 80 sample discs. The actual
allocation can be determined at the sample list. It provides a list with an entry for every
sample slot containing the name of the containing sample, the slot- number, the associated
sequence and the status of the sample slot.
Warning: For mechanical reasons never use a magnetic sample disc at position 21!
Otherwise it is possible to lose this disc during the sample transport.
To load (or unload) a sample disc, the sample wheel must be rotated until the required slot is
in front of the sample storage. There a two ways to achieve this. First, it is possible to select
the relevant slot in the numeric up/down move position to Load/Unload and press the
button move. The wheel will now move the sample slot in front of the sample storage. The
second way is to go to the front side of the sample storage and press the sample wheel
forward/reverse buttons (see Figure 2) until the required slot is in front. Now it is possible to
add or remove the sample disc to or from the slot.
After this, the new status of the slots must be announced to the software. Do this by selecting
the slot in the sample list and change the status of the marked positions to one of the
three possibilities:
empty – the slot contains no sample disc and so will not be processed.
loaded – the slot contains a sample disc and will be processed when a sequence is
associated and the measurement system is running.
measured - the slot contains a sample disc but will not be processed by the
measurement system. After a measurement the system will change the status of a
slot to this state.
Be sure to always select the correct new state of a sample slot after loading/ unloading a
sample. Otherwise malfunctions of the system are possible (e.g. unloading in an already
occupied slot).
When the sample discs are loaded, a measurement sequence can be created and linked to a
sample slot.
You can select more than one slot at a time for an operation by using the strg and/ or shift
key while clicking in the sample list.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
36 / 78
3.4.2. Creating a sequence
Every measurement done with an aliquot is defined by a sequence or list of elementary
process steps done one after another. These sequences are created in the sequence editor
and are stored in a library for later assignment to a concrete sample disc or aliquot.
The sequence editor is opened by clicking on a slot in the sample list. The sequence library
(Figure 15) consists of a list of all sequences created in the past. With the three buttons to
the right a new sequence can be created, an existing edited or deleted.
Figure 15: sequence library
Clicking the new sequence button opens the sequence editor (see Figure 16). All
elementary steps available for the specific measurement system are listed in the right list.
The list on the left side contains all steps added to the sequence in the order of processing.
Clicking on a sequence step in the add sequence step list opens an edit window, which will
be added to the actual sequence. In this window the single options of a step can be edited.
With the buttons at the lower border the steps could be rearranged, duplicated or deleted.
Double click on a step reopens the edit window for the according step.
With the save sequence button a sequence will be added to the sequence library ready to
be allocated to a sample slot.
The first and last step of a sequence normally should be a sample charging and sample
discharging step. These steps are responsible for the transport of the assigned sample disc
from the sample wheel in the measurement chamber and back. If not added, the steps in the
sequence will be processed without the sample disc on the sample arm.
Warning: Never use a measurement sequence without sample charging and discharging!
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
37 / 78
Figure 16: sequence editor
3.4.3. Assigning a sequence to a sample slot
After creating and saving a sequence using the sequence library, the sequence can be
allocated to a sample slot. Therefore select the required sample slots and double click on the
sequence entry in the library. The assigned sequence name shows up in the sample list.
Editing a linked sequence also changes the flow of all linked samples as long as they are not
currently processed. So editing a linked sequence should be done with care.
3.4.4. Starting a measurement
When the measurement system is initialized and a sequence is allocated to a loaded sample
slot a measurement can be started. By clicking the start measurement button the system
will start processing the tasks. The button changes its caption to stop measurement. The
measurement flow box (Figure 17) shows the actual step and the number of remaining
steps of this sample slot.
Figure 17: measurement status
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
38 / 78
After the measurement sequence of a slot is completed, its status in the sample list changes
to measured. If a measurement contains detection via photomultiplier, at the beginning of
the data recording the data view window (Figure 18) shows up. This window provides a live
view of the current measurement. It can be hidden by closing the window and reopened by
clicking Analysis -> Open measurement live view in the menu bar.
At the beginning of every new measurement the window will be cleared in prepare for new
data to be shown. When all assigned measurements are finished, the system will go back in
standby- mode indicated by the caption of the measurement button which changes back to
start measurement. With selecting the stop after current position checkbox the
measurement flow could be stopped after the current sample has been processed.
Figure 18: data view
3.4.5. Handle measurement data
Every measurement with a standard photomultiplier is saved in two separate ways.
The measurement curve is saved in a separate file with the .csv file format and can be
accessed by clicking Analysis -> Open measurement folder. This folder contains a
subfolder for every sequence. The files in this folder are named after the sample slot name in
the sample list. The files can be imported in other applications for further processing.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
39 / 78
The second file format is the .bin file format. It can be accessed by clicking Analysis ->
Open measurement folder, as well. For every sequence a .bin file is created which
contains all measurements done with this sequence. The results are stored in the order of
measurement.
By clicking Analysis -> Open measurement data these .bin files can be opened in a
preview window.
3.5. Special hardware characteristics
The lexsyg measurement system provides a number of additional optional components
which can’t be controlled only by the sequence editor. This section describes the using of
these components.
3.5.1. EMCCD-camera
The EMCCD camera is basically controlled by its proprietary software. Only the power supply
and the focus movement are controlled by the Lexsyg Studio software. The EMCCD control
window can be accessed by clicking Controls -> EMCCD in the menu bar (Figure 19).
Before starting the control software, the camera must be powered up by selecting the
camera on radio button. The camera is being deactivated by clicking the camera off button.
The focus is completely controlled through the sequence editor. In the normal work flow no
additional action must be taken in the control window.
Figure 19: EMCCD control
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
40 / 78
3.5.2. Spectrometer
The spectrometer camera has to cool down before it can be used for measurement. Open
the spectrometer control window by clicking Controls -> spectrometer in the menu bar
(Figure 20). Activate the spectrometer by clicking the Turn on button. Enter the target
temperature (recommended: -80°C) and click start CCD and get to target temperature.
The spectrometer camera begins to cool down and informs the user when it reached the
temperature. Now it can be used. Acquired spectra are shown in the left window and are
stored automatically in the measurement folder. After measurement the spectrometer
camera must be warmed up in a controlled manner (button warm CCD to safe temperature
and turn off) before it will be deactivated. It is absolutely necessary to warm the
spectrometer up before deactivating it. Otherwise it could be damaged.
Warning: Never turn off the spectrometer camera before it has reached room temperature in
a controlled manner by the lexsyg software! Loss of Power causes uncontrolled warming and
can also damage the spectrometer camera.
Figure 20: spectrometer control
3.6. Handling particular cases
3.6.1. Change the IP-address
The application has to know the IP-address of the target lexsyg measurement system. For
example for a new software version it could be necessary to reset the address. To do this,
open the options window (Figure 21: options menu) by clicking Options -> software options
in the menu bar. Enter the IP-Address and Port, close the window and restart the
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
41 / 78
application. After clicking the connection- button lexstudio should connect to the
measurement system. If not, check the connection cables, recheck the IP address and the
network- settings of the operating system and make sure that the lexsyg measurement
system is powered and running. The IP-address of the connected PC has to be static. If you
forgot the correct IP address for your device, please call the support hotline of Freiberg
Instruments.
Figure 21: options menu
3.6.2. Editing cooling power and evacuation time
In some particular cases it is necessary to change the power of the cooling air flow or the
duration of evacuation. This can be achieved by open the options window by clicking
Options -> software options… in the menu bar (Figure 21). In the lower part of the window
the parameters can be changed.
3.6.3. Breaking a measurement
Aborting a running measurement can be achieved by clicking the measurement button while
a running measurement. The software will ask if it is correct to abort and how to handle the
actual sample disc. When not moving the sample back to the sample wheel, keep in mind
that the disc must be removed manually.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
42 / 78
3.6.4. Initialization after an Incident – First aid menu
If an uncommon event occurs during a measurement, it is necessary to reinitialize the
system. The first step is to click the Reset Lexsyg and initialize new button (Figure 22 –1)
in the first aid menu. Wait until the measurement button changes its caption to start
measurement. If no sample disc is remaining inside the measurement chamber, a new
measurement can be started.
If there is still a sample disc in the measurement chamber, this sample disc has to be
transferred back to the sample wheel before proceeding with measurements. Therefor
localize an empty sample slot in the sample wheel and open the First aid menu (Figure 22).
If there is still a sample disc is in the grabber and not on the sample arm, or vice versa,
check the radio buttons sample on arm or sample on grabber (Figure 22 – 4)
corresponding to the sample disc location. Then Click the button Send new sample
allocation (Figure 22 – 2). After that, select an empty slot which has to be filled with the
sample disc from the measurement chamber (Figure 22 – 5) and click Try to get sample
from arm or Try to get sample from grabber (Figure 22 – 3).
Figure 22: first aid menu
The sample disc is now transferred from the corresponding location to the sample wheel. If it
is unknown, if a sample disc is located in the grabber or on the sample arm after the re-
initialization, begin with transferring back from the grabber and after that with transferring
back from the sample arm to an additional free sample wheel slot.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
43 / 78
Finally check the sample slots for the transferred sample disc. If the disc remains lost, it is
necessary to open the chamber manually and find the disc before proceeding with the
measurement. Otherwise it is possible to damage the measurement system.
Warning: After each occurring incident which interrupts a running measurement it is
necessary to find and eliminate the reason. Otherwise the whole system can be damaged.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
44 / 78
4. Service and Repair
4.1. Service
Regular maintenances of the equipment are not essential, but we suggest an annual service
by the manufacturer to check all system settings and the operating ability. A vision control of
the whole system is recommended every six month.
Electrical wiring should be controlled by accreted firms during local mandatory intervals.
Avoid exposing the equipment to heat, dirt and other environmental influences. This will
elongate the lifetime of the equipment.
4.1.1. Checking the system pressures
To check the different pressures of the measurement system use the panel with all pressure
information as shown in Figure 23.
Figure 23: system pressures overview
The inert gas and air pressure can be changed by pulling out and turning the valves shown in
Figure 24.
Warning: It is not recommended to change the air pressure! For proper functionality the
system needs an air pressure of 5 bar.
Open the small panel on the left side
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
45 / 78
Figure 24: checking system pressures
4.1.2. Lift up the measurement chamber
First remove all 5 hand screws of the chamber and check the chamber pressure. The
pressure has to be almost zero. If there is lower pressure you can fill the chamber with
nitrogen by software. It is recommended to switch of the system before lifting the
measurement chamber.
Figure 25: measurement chamber screws
The measurement chamber can be lifted by the arm in the rear area inside the device, now.
Figure 26: lift the measurement chamber
Remove all 5 screws
Arm to lift the measurement
chamber
Control of chamber pressure
Control of inert gas pressure
Setting of air pressure for the pneumatic system
Setting of inert gas pressure
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
46 / 78
4.1.3. Cleaning the measurement chamber
First of all switch off the device and lift the measurement chamber (as described in chapter
4.1.2). There is a gap between the measurement chamber and the ground plate. Use this
gap and a thin brush or vacuum cleaner to clean the chamber.
Warning: Be careful and don’t damage the sample arm inside the chamber. The arm can be moved by hand using the motor wheel in the bottom area of the
measurement system (as shown in Figure 27).
Figure 27: manual moving of the sample arm
The sample arm can be moved a quarter turn to clean the other parts of the chamber.
Hand wheel to move the sample arm
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
47 / 78
4.1.4. Cleaning the chamber windows
The chamber has two measurement positions containing windows made of quartz glass.
These windows have to be cleaned regularly because heating and silicon oil can cause dust
on it.
Cleaning at the OSL position:
First remove all plugs behind the OSL position (see Figure 28).
Figure 28: plugs of the OSL-unit
Then open the hand screws in front of the OSL position (see Figure 29).
Figure 29: screws of the OSL-unit
Move the complete OSL unit backward by pulling the unit back (see Figure 30).
Figure 30: moving the OSL-unit
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
48 / 78
Now you have free access to the detection window. You can remove it by opening the tree
screws in the frame to clean or replace it.
figure 31: overview of the OSL-unit
Mounting works in reverse direction. Don’t forget to mount the two gaskets below the window
and into the window frame.
Cleaning at the extra TL position (optional):
First remove the plug and the air tube backside of the filter changing unit on this position. For
removing the air tube push on the small frame of the connector and pull the air tube out.
figure 32: removing the plugs of the extra TL position
Open the small fastener by pulling and turning. (see below)
figure 33: fastener of the extra Tl position
Now open whole unit by pulling it up. It is blocked in straight position by himself. (see below)
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
49 / 78
figure 34: opening the extra TL position
Now you have free access to the detection window. You can remove it by opening the tree
screws in the frame to clean or replace it.
Mounting works in reverse direction. Don’t forget to mount the two gaskets below the window
and into the window frame.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
50 / 78
4.1.5. Exchanging detection filter
The used filter wheels have 6 positions. Every Filterwheel has an easy reachable cap to get
access to the filters inside.
figure 35: schematic of the filterwheel
Attention: The Number that is selected means always the filter in the light beam. That means
that the changed filter has an offset of 3 positions to the shown number. The filter position
can be calculated by: |shown position – 7| = position
figure 36: opening for changing filters
Openings: one for changing filters
and one for the light beam
Number of selected filter
Opening to change filters
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
51 / 78
4.1.6. Access to internal sample loading/unloading unit
The loading/unloading unit is hidden under the big cap on the chamber.
figure 37: location of the loading/unloading unit
Open the screws and move the cap up.
figure 38: removing the loading/unloading unit
Give attention on the right position of the gasket if you close the cap.
4.1.7. Heater exchange
The heater is mounted together with the sample arm and can be changed together with the
arm.
First the arm has to be moved to the maintenance position. Than the maintenance panel has
to be opened.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
52 / 78
figure 39: maintenance panel for changing the heater unit
Now the measurement chamber has to be moved up (see point 4.1.2).
The arm can be removed by the three screws in the centre.
figure 40: removedmaintenance panelwith visibleheater unit
Then the arm can be moved up carefully and a new one can be mounted in reverse direction.
4.2. Spare parts
Spare parts:
- Chamber window glass
- Gasket for all removeable parts connected to the chamber
- Screws
- Air tubes
- Wires and plugs
- Sample arm with heater
Tools:
- Cleaning brush
- Filter sucker
- Set of Allen wrench
Requests and feedback is welcome, please contact: [email protected]
Open the panel by unmounting the
screws
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
53 / 78
5. Physical Background
5.1. Literature
[ 1 ] M. R. Krbetschek J. Götze, A. Dietrich, T. Trautmann, Radiation Measurement,Vol.
27, No. 5/6,pp. 695-748,1997, Spectral information from minerals relevant for luminescence
dating
[ 2 ] G. Erfurt*
and M.R. Krbetschek,Ancient TL Vol. 21 No. 1 2003, IRSAR – A single-
aliquot regenerative-dose dating protocol applied to the infrared radiofluorescence (IR-RF) of
coarse-grain K-feldspar
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
54 / 78
Attachment: Radiation source
documentation
Radiation Related Aspects of the lexsyg Luminescence Measuring
System
Document: RP-061
Revision: 0
Date: 02-Aug-2011
Revision History:
Date Rev. Description by
02-Aug-2011 0 Initial release RP
1 Radiation Source and Radiation Source Module
1.1 The lexsyg measuring system
The measuring system lexsyg(see fig. 1) has been designed for absolute dating of
geological and archaeological objects by means of radio fluorescence (RF), thermo
luminescence (TL) and optically stimulated luminescence (OSL). Using a beta radiation
source a predefined absorbed dose can be administered to the sample. The luminescence
signal emitted is coupled out by means of an adapted optics and detected by either a photo-
multiplier tube or CCD camera. The lexsyg standard is equipped with a beta radiation
source, and it can optionally be provided with both a beta and an alpha radiation source (see
lexsyg data sheet).
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
55 / 78
Fig. 1 Basic concept of the measuring system lexsyg
Photo-multiplier
tube
CCD camera
Radiation source
modules
Control panel
Port for sample
loading
Fig. 2 Measuring system lexsyg (housing cover open)
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
56 / 78
Fig. 3 Sample handling system of the lexsyg
Samples are placed on a circular sample holder accommodating up to 80 samples. During
the automatic measuring procedure the individual sample to be analyzed is moved to the
appropriate position (irradiation, luminescence detection, etc.) in the measuring chamber by
means of a transfer system. The sample holder is accessible to the operator through an
opening at the front of the measuring system (see figures 2 and 3).
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
57 / 78
1.2 Radiation Source Module
The radiation source module contains the radiation source (see figures 4 – 6) and provides
necessary operating and safety features.
adapter for pressurized air (shutter)
quartz glass window (fluorescence
light exit)
connectors for shutter position
sensors
Transport safety plate (beta shield
during transportation, storage and
handling)
Fig. 4 Radiation Source Module (top view) mounted on transport safety plate
Fig. 5
Radiation source module (sketch)
Dimensions: 90 mm x 85 mm x 25 mm
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
58 / 78
Position sensors Retracting spring radiation source source holder shutter
Fig. 6 Radiation source module (bottom view, transport safety plate removed)
Left: shutter closed; Right: shutter open
The radiation source is safely contained in the radiation source module. The radiation source
is mechanically fixed in a holder ensuring accurate positioning over the sample during
irradiation. The only movable part of the radiation source module is the shutter, so there is no
mechanical load on the source.
With the shutter being in its “closed” position, the downwards-facing radiation exit window is
completely covered by a 1 mm stainless steel shutter (fig. 6). For sample irradiation (typically
between one and ten minutes of duration) the shutter will be opened by a pneumatic cylinder.
The actual shutter position will be detected by two non-contact sensors. In the case of failure
of electrical power or pressurized air supply the shutter will automatically close due to a
retracting spring. Due to its bifurcated design the shutter doesn’t only cover the radiation exit
window of the source but also protects the quartz glass windows from scattered beta
radiation.
The radiation source module is firmly attached to the base plate of the lexsyg by special
screws. A 5 mm lead lined housing surrounding the source module (not shown in fig. 2, see
also fig. 12) reduces dose rate and occupational exposure. The housing of the lexsyg is fully
closed and locked during routine operation, assuring that the radiation source isn’t accessible
to the standard operator or any other person.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
59 / 78
1.3 Beta radiation source
The current design of the beta radiation source features a circular arrangement of 17 single
miniaturized, sealed beta sources (see table 1). These single sources are inserted into a
circular groove (fig. 8). A stainless steel foil of 25 µm thickness, that is micro-laser welded to
the body, fixes the sources in the groove. The two main components – the source carrier and
the source body with cover foil – are affixed to each other by means of three special screws
(see figures 7 and 8). All of the components are made of stainless steel.
The radiation source provides an absorbed dose rate of 0.1 Gy·s-1 to the sample. The
radiation field inhomogeneity is less than ± 3 % over a sample diameter of 10 mm.
Fig. 7 Radiation source (schematic): red – circular arrangement of 17 miniaturized
sources, blue – radiation exit window, 25 µm stainless steel, micro-laser welded
to the body; outer diameter 25 mm, height 9.5 mm, material: stainless steel
source carrier body with cover foil
Fig. 8 Left: Components of the radiation source (individual miniaturized sources not
shown), right: assembled source with laser engraving
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
60 / 78
The radiation source is precisely positioned and safely fixed in the source holder by three
special screws (see fig. 6, right) ensuring that the radiation exit window cannot be touched or
even damaged by the shutter.
The following information is laser engraved on the circumference of the source body (see
fig. 8):
- Radioactive trefoil symbol
- Radioisotope: Sr-90
- Maximum activity: 2 GBq
- Serial number: RFQ-xxx (xxx may range from 001 to 999)
Further, the radiation source module is uniquely identified by type and serial number (see fig.
4). The labelling of the source module also provides information on the radiation source
installed. The housing bears the radiation trefoil symbol.
Table 1 Technical specifications of the single miniaturized sources the circular source
consists of (refer to attachment 1)
Source type strontium-90 seed type Sr0.S03-1
manufacturer Eckert & Ziegler BEBIG GmbH
radioisotope Sr-90
maximum activity 118 MBq
radioactive material
physical solid
chemical SrTiO3 (strontium titanate)
form sealed
source canister stainless steel 1.4301/1.4541, 40 µm thick, laser welded
dimensions length: 2.5 mm, diameter: 0.64 mm
type of radiation beta
maximum energy 2.27 MeV
ISO 2919 classification C.63211
special tests ANSI N44.1-1977 tests: puncture, impact, temperature
leakage test immersion test per ISO 9978
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
61 / 78
In case the radiation source module is not attached to the base plate of the lexsyg the
source module’s bottom side shall be completely covered by a 4 mm thick aluminium
transport safety plate. Hence, the radiation source module can be used for safely storing and
shipping of the built-in radiation source.
2 Occupational dose assessment
2.1 Dose rates relevant for exposure assessment
Ambient dose equivalent rate H*(10) was determined at various distances from the radiation
source module. A prototype source having an activity of 1.5 GBq was utilized, and the results
were extrapolated according to a maximum source activity of 2.0 GBq.
Conditions were as follows:
- Radiation source (Sr-90)
- Radiation source module with 5 mm lead-lined housing installed in lexsyg
(prototype unit)
- Radiation source shutter open (sample irradiation)
Table 2 Ambient dose equivalent dose rate at various distances to the surface of the
radiation source module (source activity 2.0 GBq)
distance to source
cm
position equivalent dose
rate H*(10) / µSv/h
5 Position of operator’s hand during installation of the
source module in device 270
10 Position of operator’s hand during adjustment and
calibration activities 160
20 Position of operator’s hand during cleaning of
sample holder 27
30 Position of operator’s eyes during adjustment and
calibration activities 40
50 Position of operator’s hand during loading the
sample holder 2,7
70 Position of operator’s body during loading the 2,0
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
62 / 78
sample holder
120 1 m distance to housing surface of lexsyg 0,8
In the present case, the radiation source module was completely sealed against beta
radiation. Hence, the equivalent dose rate measured was solely caused by photon
(bremsstrahlung) radiation created within the source body, the shutter and the source
module housing. It made no difference whether the source shutter was opened or closed.
Note that the source shutter doesn’t completely shield all of the beta radiation emitted by the
radiations source. It is anticipated that actual dose rates may exceed those given in table 2
by more than 10 times under certain circumstances such as:
- Wipe test conducted at source shutter
- Cleaning of the measuring chamber or the sample carrier of the lexsyg
- Handling of the radiation source module during installation, wipe test and handling
emergency situations
2.2 Occupational exposure during use of the lexsyg
Occupational exposures of the user associated with typical activities during operating and
maintaining the lexsyg are determined below:
2.2.1 User activities at the control panel of the lexsyg
Process steps Loading of sample carousel, keyboard input (measuring
sequences, test runs, etc.)
duration 15 min
Typical distance to source 70 cm (eyes, body)
30 cm (hands)
Dose per handling 0.5 µSv (eyes, body)
0.7 µSv (hands)
Number of annual activities 100
notes Source shutter typically closed (may be opened for a
short period of time), minimized occupancy time of the
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
63 / 78
operator near the device
2.2.2 Exchange of optical detector/adjustment of optical components
Process steps 1. Opening of the lexsyg housing
2. Disassembling of the optical system mounted on
top of the radiation source module
3. Assembling and adjusting the optical
components
4. Test run
5. Closing the lexsyg housing
duration 20 min
Typical distance to source 30 cm (eyes, body)
10 cm (hands)
Dose per handling 13 µSv (eyes, body)
55 µSv (hands)
Number of annual activities 20
notes Shutter typically closed / can be open for a short
period of time
Beta radiation completely shielded by radiation
source module
Housing of the lexsyg opened
If applicable, use appropriate means of radiation
protection (tools)
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
64 / 78
2.2.3 Exchange of quartz glass window (top of the radiation source module)
Process steps 1. Opening of the lexsyg housing
2. Disassembling of the external optical system
mounted on top of the radiation source module
3. Removing the lead shield of the radiation source
module
4. Use of a temporary radiation shield (10 mm
acrylic glass wall)
5. Removing the window fixing ring
6. Exchanging the window (use tweezers or similar
tools)
7. Assembling the window fixing ring
8. Assembling and adjusting the optical
components
9. Test run
10. Closing the lexsyg housing
duration 20 min
Typical distance to source 30 cm (eye, body)
10 cm (hands)
Dose per handling 13 µSv (eye, body)
55 µSv (hands)
Number of annual activities 1
notes Shutter typically closed / can be open for a short
period of time
Until window is removed beta radiation is completely
shielded by radiation source module
Upon removing the quartz window beta radiation
may escape from the radiation source module
For step 6 use tweezers or other appropriate tool;
wear radiation protection goggles
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
65 / 78
2. 2.4 Cleaning of the measuring chamber/sample carousel
Process steps 1. Opening of the lexsyg housing
2. Opening of the measuring chamber
3. Cleaning of the measuring chamber by means of
special tools (e.g. long forceps)
4. Closing the lexsyg housing
5. Test run
duration 5 min
Typical distance to source 50 cm (eyes, body)
20 cm (hands)
Dose per handling 0.3 µSv (eyes, body)
3 µSv (hands)
Number of annual activities 40
notes Shutter must be closed
Upon opening the measuring chamber small amount
of beta radiation may escape from the radiation
source module
The measuring chamber can just be lifted by 15 mm
in order to avoid access to the radiation source
module with bare hands
Minimize hand exposure by using appropriate tools
(step 3)
Use mobile radiation shield (10 mm acrylic glass)
and radiation protection goggles
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
66 / 78
2.2.5 Installation of the radiation source module
Process steps 1. Opening of the lexsyg housing
2. Taking out of the radiation source module off the
shielding shipping/storage container
3. Installation of a radiation protection shield
(10 mm acrylic glass)
4. Removal of the transport safety plate from the
radiation source module
5. Visual inspection and functionality check of the
radiation source module
6. Assembling of the radiation source module and
lead shield into the lexsyg
7. Assembling and adjusting of the optical
components
8. Test run
duration 10 min
Typical distance to source 30 cm (eyes, body)
5 cm (hands)
Dose per handling 7 µSv (eyes, body)
50 µSv (hands)
Number of annual activities 5
notes Ensure shutter is closed. Upon removing the
transport safety plate of the radiation source module
beta radiation escapes from the radiation source
module.
Use mobile radiation shield (10 mm acrylic glass)
and wear radiation protection goggles. Use
appropriate tools.
Removing the radiation source module from lexsyg
research occurs in reverse order and takes about
5 min (no adjustment of optical components
necessary)
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
67 / 78
Note: The measuring system lexsyg research and the pre-calibrated radiation source
module are shipped separately. Installation of the radiation source module into the
measuring system can easily and safely be performed by the user per instructions provided
by the manufacturer. There is no need to perform any source-related adjustment and
calibration activities.
2.2.6 Radiation source leakage test (performed by approved expert)
Process steps 1. Removal of the radiation source module from the
lexsyg according to the procedure described in
2.2.5
2. Opening of shutter
3. Performing wipe test (refer to 5)
4. Visual inspection of the source and source
holder
5. Reassembling of the system in reverse order
duration 15 min
Typical distance to source 30 cm (eyes, body)
10 cm (hands)
Dose per handling 11 µSv (eyes, body)
70 µSv (hands)
Number of annual activities 1
notes Use appropriate tools and shielding
Wear protective goggles, if necessary
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
68 / 78
2.2.7 Lab user (without operating the lexsyg)
Process steps Laboratory work
Minimum distance to source 200 cm
Dose rate < 1 µSv/h
Annual dose < 1 mSv
notes Normal laboratory work without operating the
lexsyg
Mean annual occupancy time at given distance is
estimated not to exceed 200 hrs
2.2.8 Emergency situation: Shutter wouldn’t close correctly and needs to be checked
and manually adjusted
Process steps 1. Opening of the lexsyg housing
2. Disassembling of the external optical system
mounted on top of the radiation source module
3. Removing the lead shield of the radiation source
module
4. Installation of a radiation protection shield
(10 mm acrylic glass)
5. Visual inspection of the radiation source module
and the shutter
6. Repair and adjustment of components
7. Reassemble system in reverse order
8. Test run
duration 15 min
Typical distance to source 30 cm (eyes, body)
5 cm (hands)
Dose per handling 15 µSv (eyes, body)
140 µSv (hands)
Number of annual activities 1
notes Shutter assumed to not fully closed
Upon removing the radiation source module from
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
69 / 78
the lexsyg beta radiation escapes from the radiation
source module
Window of radiation source shall be covered
appropriately to minimize exposure.
Use mobile radiation shield (10 mm acrylic glass)
and wear radiation protection goggles. Use
appropriate tools.
2.3 Classification of radiation workers
Tab. 3 Estimated occupational exposure during work with the lexsyg
Person Annual Dose / mSv *
Effective Dose Lens of the Eye Extremities (Hand)
Scientist / Operator 0,4 0,4 1,6
Laboratory
Technician
0,3 0,3 0,3
Approved Expert 0,02 0,02 0,07
Annual limit for
members of the
general public
1 15 50
* includes exposure obtained during an emergency situation
The above conservative occupational dose assessment reveals that annual dose limits
according to European Directive 96/29 EURATOM, even for members of the general public,
will not be exceeded by far (table 3).
3 System safety aspects
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
70 / 78
The access to the radiation source, if installed in the lexsyg, is limited to authorized
personnel only: To open the housing of the measuring system a key is required. The
radiation module is affixed to the base transport safety plate of the lexsyg by four special
screws requiring special tooling to be unfastened.
The radiation source is safely contained within the radiation source module providing
appropriate radiation shielding. The whole system has been designed such that exposure to
operators and lab workers is limited (see section 2). However, radiation surveys are
recommended to be conducted to verify radiation safety at any time.
The source shutter mechanism (see fig. 6) is pneumatically operated and has two stable
positions, i.e., the shutter is either fully closed or fully open. The actual shutter position is
monitored by sensors. In case of complete failure of electrical power and/or pressurized air
supply the shutter will automatically close due to a retracting spring.
Prior to opening the radiation source module or any activities involving the radiation source
module it is mandatory to verify proper shutter position and to conduct a radiation survey.
The measuring chamber can just be lifted by 15 mm from the lexsyg base plate. Thus,
inadvertent access to the lower side of the radiation source module by bare hands will be
avoided.
The sample heater is designed to heat up the sample to 700 °C. Under normal operating
conditions the sample will not be heated when being placed under the radiation source. The
source is being designed to withstand a temperature of 800 °C for one hour (classification
C.63211 per ISO 2919). However, there are software and hardware based safety
mechanisms in place to avoid any thermal load on the source.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
71 / 78
4 Maintenance and service
The radiation source is exclusively used under laboratory conditions. It is not subjected to
any excessive mechanical, thermal or chemical loads other than specified in the appropriate
ISO 2919 testing protocol.
The sealed beta sources of type Sr0.S03-1 are embedded in a stainless steel holder and
protected by a cover foil (25 µm stainless steel), see 8. The special design of the source
holder as well as regularly conducted visual inspections and leakage tests (refer to section 5)
ensure radiation source integrity at any time.
The radiation source module, in particular the shutter mechanism, has undergone extensive
testing and functional performance verification. Anticipating the shutter to be activated less
than 1000 (one thousand) times annually, the manufacturer recommends an annual
inspection and functionality verification of the radiation source module. The inspection is to
be carried out by trained personnel only per instructions provided by the manufacturer.
The radiation source has a recommended working life of 10 (ten) years.
Radiation surveys are recommended to be conducted on a regular base in accordance with
local policy.
In case of any device malfunctions related to the radiation source the manufacturer is to be
informed immediately.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
72 / 78
5 Source leakage test
For source leakage testing the manufacturer recommends a wipe test per ISO 9978. The
source manufacturer recommends the wipe to be taken directly on the radiation window of
the source (see fig. 9) as well as on the accessible surfaces of the shutter. Note that the
source must not be removed from the radiation source module.
It is recommended to perform the leakage annually. Note that local policy may require
another leakage test interval.
Important: It’s the user’s responsibility to have the leakage test conducted.
The leakage test is to be conducted as follows:
- Verify shutter is closed. Perform radiation survey.
- Remove radiation source module from lexsyg base plate.
- Note that the shutter doesn’t shield beta radiation completely. Use appropriate
radiation safety measures (acrylic glass shield, 4 mm aluminium transport safety plate
for intermediate storage of the source module).
- Open shutter by hydraulic (or mechanical) means. Verify the shutter opens and
closes smoothly.
- Inspect radiation source window for mechanical damage. Verify correct source
position in source holder.
- Perform wipe test on surfaces as shown in fig. 9. Ensure the radiation window is not
damaged. The radiation source MUST NOT be removed from the source holder.
- Close shutter.
- Reinstall radiation source module on the base plate of the lexsyg.
Radiation hazard!
Dose rate at 20 cm distance from unshielded source is about 200 mSv/h, i.e. 50 µSv/sec.
Use appropriate means of radiation protection.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
73 / 78
Fixing screws for source.
Do not unfasten!
Source window (direct wipe
test)
Shutter (indirect wipe test)
Fig. 9 Surfaces for the wipe test (Shutter in “open” position providing access to the
source)
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
74 / 78
6 Transportation and installation of radiation source module
The radiation source will be assembled into the radiation source module at the
manufacturer’s site and will be supplied as calibrated and tested assembly. Separate from
the lexsyg main unit, the radiation source module is being shipped in a certified type A
shipping container (fig. 10, attachment 2).
Fig. 10 Type A shipping container (left) and source container with radiation source
module (right)
The radiation source module can be easily installed into the lexsyg per instructions provided
below and depicted by figures 10 – 12.
- Upon receipt of radioactive source shipping perform activities as required by local
radiation policy involving radiation survey, wipe test on the shipping container,
document check etc.
- Open shipping drum and remove upper part of the Styrofoam insert.
- Remove source container from Styrofoam insert.
- Open source container.
- Remove radiation source module from source container. Ensure transport safety
plate is properly affixed to the radiation source module.
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
75 / 78
- Perform radiation survey. The dose rate at 30 cm from the source module is expected
to be less than 50 µSv/h. Note that the actual reading may vary depending on the
radiation meter used.
- Verify correct shutter position (shutter can be visualized through quartz window).
- Unfasten 4 special screws and remove the transport safety plate.
- Use appropriate radiation protection measures (acrylic glass shield, if available).
Wear film badge and finger ring dosimeter. To minimize exposure always hold the
source module such that the open side directs away from your body.
- Install radiation source module in lexsyg main unit. Use 4 screws to fasten it.
- Attach electrical sensor connectors and pressurized air connector.
- Perform a test run to verify shutter functionality.
- Install radiation shield cover (5 mm lead housing).
Fig. 11 Installing the radiation source module into the lexsyg
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
76 / 78
Fig. 12 5 mm lead housing (on right) to be placed over the radiation source module after
installing in the lexsyg main unit
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
77 / 78
Attachment 1
Data sheet of beta source Sr0.S03-1 (Eckert & Ziegler BEBIG GmbH)
Freiberg Instruments GmbH ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
78 / 78
Attachment 2
Type A shipping container drawing PZ-0103-001 (Eckert & Ziegler nuclitec GmbH)