UNITEST®

Echometer 3000

Instruction Manual Cat. No. 2006 Mode d´emploi Réf. 2006 Manual de instrucciones No de ref. 2006

Page 2: Medidor Longitude de Cabos por Reflexao de Impulso e Medidor Resistencia - Manual Sonigate

Table of Contents..................................................................................Page1.0 Introduction / Scope of Supply........................................................172.0 Transport and Storage ....................................................................183.0 Safety Measures..............................................................................184.0 General Information regarding UNITEST Echometer 3000 ..............195.0 Control Elements / Display Elements ..............................................206.0 Performance of Measurements ......................................................216.1 Performance of Cable Length Measurements..................................216.2 Performance of Cable Length Measurements within

the Reference Range ......................................................................246.2.1 Calibration of Customer Specific Cable Types..................................246.2.2 Deletion of Reference Ranges ........................................................246.3 Trouble Shooting ............................................................................256.4 Performance of Resistance Measurements /

Continuity Tests ..............................................................................257.0 Internal Data Memory / Cumulative Data Memory ..........................267.1 Saving Measurement Results ..........................................................267.1.1 Deletion of all Saved Measurement Results ....................................267.1.2 Deletion of the Last Measurement Result ........................................267.2 Cumulate and Save Measurement Results ......................................268.0 Transfer of Saved Measurement Values to a PC ..............................279.0 Maintenance....................................................................................279.1 Cleaning ..........................................................................................279.2 Battery Replacement ......................................................................2710.0 Calibration Interval ..........................................................................2711.0 Technical Data ................................................................................28

16

Table of Contents

Page 3: Medidor Longitude de Cabos por Reflexao de Impulso e Medidor Resistencia - Manual Sonigate

Engl

ish

17

Introduction / Scope of Supply

References marked on instrument or in instruction manual:

Warning of a potential danger, comply with in-struction manual.

Reference. Please use utmost attention.

Caution! Dangerous voltage. Danger of electri-cal shock.

Continuous double or reinforced insulationcomplies with category II.

Conformity symbol, the instrument complieswith the valid directives. It complies with theEMC Directive (89/336/EEC), Standards EN50081-1and EN 50082-1 are fulfilled. It alsocomplies with the Low Voltage Directive(73/23/EEC), Standard EN 61010-1 is fulfilled.

The instruction manual contains informationand references, necessary for safe opera-tion and maintenance of the instrument.Prior to using the instrument (commissioning/ assembly) the user is kindly requested tothoroughly read the instruction manual andcomply with it in all sections.

Failure to read the instruction manual or tocomply with the warnings and referencescontained herein can result in serious bodi-ly injury or instrument damage.

1.0 Introduction / Scope of SupplyYou have acquired a high standard measurement in-strument supplied by the company Ch. BEHA GmbHwhich is able to perform reproducible measure-ments over a very long time period. The companyCh. BEHA GmbH is a member of the world-wide op-erating BEHA group. The head office of the BEHAgroup is located in Glottertal/Schwarzwald, togeth-er with the technological centre. The BEHA group isa leading enterprise for test and measurement in-struments.

The UNITEST Echometers are universally usablecable length measurement instruments. They havebeen built in compliance with the most recent pre-scriptions and gurantee safe and reliable measure-ments. The UNITEST Echometers are a valuablesupport for trade and industry for tally measure-ments, inventories, and residual length measure-ments.

The instruments have the following characteris-tics:• Echometer for easy and time efficient measure-

ment of cable lengths in trade, industry, utilitycompanies, and specialist electronic stores.

• Only one cable end is required. Thus, conductorsalready installed or cable drums can be measured,thus saving time.

• 58 fixed, pre-programmed measurement rangesrepresenting the most common cable cross sec-tions

• 87 variable measurement ranges, individuallyprogrammable by the user to measure customerspecific conductor and cable types

• Detecting single or multi-wire conductor inter-ruptions or short-circuits for trouble shooting

• Cable and conductor length measurement up toapprox. 2000 m (depending on the type of con-ductor)

• Internal data memory to save 500 measurementsfor inventories, tally measurements, cable returnmeasurements, etc.

• RS232 interface for transfer and further PC pro-cessing of measurement data

• Cumulative memory to measure and cumulateseveral cable rings of the same type of conductor,saving time and money

• Additional resistance measurement up to 2000Ohm and acoustic continuity test

• Display light for better visibility while working inbadly illuminated environments, such as storageareas

• Automatic power off

Page 4: Medidor Longitude de Cabos por Reflexao de Impulso e Medidor Resistencia - Manual Sonigate

The scope of supply comprises of:1 UNITEST Echometer 30002 Test leads red/black2 Alligator clamps1 Carrying holster1 Battery 9 V, IEC 6LR613 Concise reference guides1 Instruction manual

Accessories:Windows Software UNITEST ”Report Studio”Order No: 1207

2.0 Transport and StoragePlease keep the original packaging for later trans-port, e.g. for calibration. Any transport damage dueto faulty packaging will be excluded from warrantyclaims.

In order to avoid instrument damage, it is ad-vised to remove accumulators when not usingthe instrument over a certain time period. How-ever, should the instrument be contaminatedby leaking battery cells, you are kindly re-quested to return it to the factory for cleaningand inspection.

Instruments must be stored in dry and closedareas. In the case of an instrument being trans-ported in extreme temperatures, a recoverytime of minimum 2 hours is required prior toinstrument operation.

3.0 Safety MeasuresThe UNITEST Cable Length Meters have been builtand tested in compliance with the valid Regulationsand have left the company in safe and perfect con-dition. To maintain this condition, the user mustcomply with the safety references contained in thisinstruction manual.

Never apply voltage to any of the instrumentmeasurement ranges. Always verify that thecircuits are not live using a dual-pole volt-age tester (e.g. UNITEST 2000 alpha)!

Prior to any operation, ensure that connectingleads and instrument are in perfect condition.

The respective accident prevention regulationsestablished by the professional association forelectrical systems and equipment must bestrictly met at all times.

The instrument may only be used within the op-erating ranges as specified in the technical datasection.

Prior to opening, the instrument has to beswitched off and disconnected from any cir-cuit.

Avoid any heating up of the instrument by di-rect sunlight to ensure perfect functioning andlong instrument life.

Appropriate Usage

The instrument may only be used under thoseconditions and for those purposes for which itwas built. For this reason, in particular the safe-ty references,as well as the technical data in-cluding environmental conditions and theusage in dry environments must be followed atall times.

The operational safety of the instrument is noilonger guaranteedafter unauthorised modifi-cations or changes.

The instrument may only be opened by an au-thorised service technician, e.g. for fuse re-placement.

18

Transport and Storage / Safety Measures

Page 5: Medidor Longitude de Cabos por Reflexao de Impulso e Medidor Resistencia - Manual Sonigate

Engl

ish

19

General Information

4.0 General Information regarding UNITEST Echometers 3000

The UNITEST Echometers are measurement instru-ments for swift, easy and precise determination ofthe length of a cable or a conductor as well as forresistance measurements.

The cable length determination is carried out inaccordance with the pulse reflection procedure.

Hereby, pulses are send into a cable which "travel”at a certain velocity along the cable. If these pulseshit the cable end they are reflected and return to thestarting point.

The UNITEST Echometer 3000 calculates the cablelength on the basis of time difference between emis-sion and reception of the pulses.

If a short-circuit or an interruption is present in theconductor to be measured the digital display indi-cates the length of the conductor up to the fault lo-cation. This allows an easy and fast fault localiza-tion.

The travel time required only depends on the dis-persion velocity of the travelling pulse. In turn, thedispersion velocity is determined by the make andstructure of a cable. This means, that every cablehas a certain characteristic dispersion velocity and,thus, different travel times.

The cable characteristics saved in the Echometer(TAB range) represent a cable mean value estab-lished by the various manufacturers and ensure ap-proximate measurements. To carry out precisemeasurements, the user may calibrate the cablesupplied by a certain manufacturer and permanent-ly store the cable characteristics in the instrumentmemory for subsequent measurements (REFrange).

The UNITEST Echometer 3000 enables the mea-surement of the most conventional cable and con-ductor cross sections. Additionally, 87 freely ad-justable cable and conductor cross sections areavailable for programming and measurement.

Simply connect the instrument alligator clamps tothe ends of a cable or a conductor and set the re-spective measurement range. The conductor lengthis clearly indicated on the digital instrument display.

The built-in memory is used to save measurementresults and to transfer the data via interface to a PC.

Page 6: Medidor Longitude de Cabos por Reflexao de Impulso e Medidor Resistencia - Manual Sonigate

5.0 Control Elements / Display ElementsControl Elements:1. RS232 interface2. LCD3. Function keys4. Push-button on/off/start5. Selection dial for the measurement ranges6. Measurement input sockets7. Display light

Display Elements:1. Emergency reflective triangle2. Error display3. Battery replacement required4. Measurement value5. Measurement unit6. Memory display (M)7. Table / reference range display8. Cumulative value memory display (Res)9. Short-circuit display (COMP)

Description of the Menu Keys:

ShiftThe second function of the double function keys isonly activated when pressing the keys together withthe "Shift” key.

TAB/REFThe ''TAB/REF'' key is used to change into differentlevels: Table I, Table II, Reference range I, Referencerange II, and Reference range III.

m/ft Switching between the measurement unitsmeter/feet.

SendThe ''Send'' key is used to start the transfer of thedata memorised in the UNITEST Echometer 3000 tothe PC.

+ Clear all''Clear all'' deletes the total internal memory. Thecumulative data memory is not cleared.The ''Clear all'' key is activated when the ''Shift'' keyis pressed at the same time.

Res+The ''Res+'' key is used to activate the cumulativedata memory. The measurement data within thesame measurement range is added, i.e. severalrings of the same cable type can be added.

+ [Light]The ''Light key is used to switch on the display light.To reduce the battery consumption, the display lightis switched off after approx. 30s. after the last keyoption. The ''Light” is only activated when the "Shift” key ispressed at the same time.

[Arrow down]When the key and the key arepressed at the same time, the reference cablelengths within the reference range can be set.

Shift

Clear all

SendShift

Clear all

Send

m/ftm/ft

Shift

20

Control Elements / Display Elements / Description of the Menu Keys

1

2

3

4

5

6

1 2 3

4 5

67

89

7

Page 7: Medidor Longitude de Cabos por Reflexao de Impulso e Medidor Resistencia - Manual Sonigate

Engl

ish

21

Performance of Measurements / Performance of Cable Lengths Measurements

+ Store REFThe ''Store REF'' key is used to save the previouslyset reference cable length.The ''Store REF'' is only activated when the "Shift”key is pressed at the same time.

[Arrow up]The [arrow up] key together with the [arrow down]key are used to set the reference cable lengths with-in the reference range.

StoreThe ''Store'' key is used to save a measurement re-sult. If several measurement values have been cu-mulated within the cumulative data memory, thesum can be imported into the data memory bypressing the ''Store'' key.

+ Clear one''Clear one'' is pressed to clear the last measure-ment value saved. The cumulative data memory isnot affected. The ''Clear one '' key is only activatedwhen the "Shift” key is pressed at the same time.

6.0 Performance of Measurements6.1 Performance of Cable Lengths

Measurements• The leads to be measured must be positioned next

to each other. It is not possible to measure indi-vidual wires.

• For coaxial cable measurements, connect the in-ternal conductors to the read sockets and thesheath to the black socket.

• Only connect one lead per measurement connec-tion.

• Additional measurement errors up to 10 % read-ing may occur due to the differing cable structureand material, depending on the manufacturer.

" Switch on the UNITEST Echometer pressing thekey.

" Connect the red test lead to the red socket andthe black test lead to the black socket of the mrange.

" Connect two parallel leads of the cable to bemeasured using the alligator clamps.

" To set the desired measurement unit (m/ft),press the und key. ''m'' or''ft'' is displayed.

" Set the measurement range selection dial in ac-cordance with the conductor types in Table I orII. The key is used to switch to the mea-surement ranges of table II. "TAB II” appears onthe display.

If the message ''- - -'' is displayed, the UNITESTEchometer is ready to perform measurements.

" Start measurement using the key.

A single signal sound is audible and the mea-surement value of the conductor length in thepre-selected unit (m/ft) is displayed.

It is preferable to carry out the measurementwith open cable ends. If the cable end has ashort-circuit the message COMP is displayed.Short-circuited cable ends cause an increasingmeasurement error of the UNITEST Echometer3000.

Start/Off

m/ftm/ftShift

Start/Off

Shift

Shift

m

Ω mEchometer 6

29

Page 8: Medidor Longitude de Cabos por Reflexao de Impulso e Medidor Resistencia - Manual Sonigate

Switch position assignment

Table I:Switch Position Table I (TAB I) Maximum Length1 NYM-J 3x1,5 mm2 1200 m2 NYM-J 4x1,5 mm2 1200 m3 NYM-J 5x1,5 mm2 1200 m4 NYM-J 7x1,5 mm2 1200 m5 NYM-J 3x2,5 mm2 1200 m6 NYM-J 4x2,5 mm2 1200 m7 NYM-J 5x2,5 mm2 1200 m8 NYM-J 7x2,5 mm2 1200 m9 NYM-J 5x4 mm2 1200 m10 NYM-J 5x6 mm2 1200 m11 NYM-J 4x10 mm2 1200 m12 NYM-J 5x10 mm2 1200 m13 NYM-J 4x16 mm2 1200 m14 NYM-J 5x16 mm2 1200 m15 NYY-J 3x1,5 mm2 1200 m16 NYY-J 4x1,5 mm2 1200 m17 NYY-J 5x1,5 mm2 1200 m18 NYY-J 3x2,5 mm2 1200 m19 NYY-J 5x2,5 mm2 1200 m20 NYY-J 4x10 mm2 1200 m21 NYY-J 5x10 mm2 1200 m22 NYY-J 4x16 mm2 1200 m23 NYY-J 4x25 mm2 1200 m24 NYY-J 4x35 mm2 1200 m25 JYSTY 2x2x0,6 mm 500 m26 JYSTY 4x2x0,6 mm 500 m27 JYSTY 6x2x0,6 mm 500 m28 JYSTY 2x2x0,8 mm 500 m29 JYSTY 4x2x0,8 mm 500 m

22

Page 9: Medidor Longitude de Cabos por Reflexao de Impulso e Medidor Resistencia - Manual Sonigate

Engl

ish

23

Table II:Switch Position Table II (TAB II) Maximum Length1 JYSTY 6x2x0,8 mm 500 m2 SAT:LCD 58 1500 m3 SAT:LCD 61 1500 m4 SAT: LCD 79 1900 m5 SAT:LCD 90 1900 m6 SAT:LCD 95 2000 m7 SAT:LCD 99 2000 m8 SAT:LCM 13 1500 m9 SAT:LCM 14 1900 m10 SAT:0,8/3,5 ALG; 1,1/5,0 ALG; 1900 m

1,65/7,2 ALG; 2,2/9,6 ALG; SAT-MINI 1 (*)11 SAT:SAT 3000 (*) 1900 m12 SAT:2-SAT 3000 midi; 2-SAT 3000 mini, 1900 m

2-SAT 3000 mini round 4-SAT 3000 (*)13 Cable TV: 0,6L/3,0 ALG; 0,63/3,0 ALG; 0,4/2,45 ALG; 1500 m

0,8/3,7 ALG; 0,7/4,4 ALG; 0,7/4,5 ALG-H; 0,7/4,8 C ALG; 0,7/4,8 CW ALG; 1,1/7,3 ALG; 1,1/7,3 ALG-T; 1,8/11,5 FG; A-2YK2YliKx 1,1/7,3 (*)

14 Cable TV: 0,4/1,75 ALG (*) 1800 m15 Cable TV: 0,4/2,0 ALG (*) 1900 m16 Cable TV: A-2YOK2YlnKx 2; 2/8,8 A-2YOK2YlqKx 3,3/13,5(*) 2000 m17 TV:0632-75vz (*) 1500 m18 TV: 0938-60vz 1600 m

0637-75bl; 0637-75vz; 0637-75vs; 0632-75bl (*)19 TV: 1046-75bl (*) 1900 m20 RG6/U; RG8/U; RG11/U; RG58/U; RG058/U; 1500 m

RG59/U; RG174/U; RG213/U; RG214/U; RG215/U; RG216/U; RG217/U; RG218/U; RG223/U; RG316/U

21 RG178/U; RG179/U; RG180/U; RG187/U 1600 m22 RG62/U; RG71/U 1600 m23 DATA:10BaseT (RG58/U) 1500 m24 DATA:YELLOW-CABLE 1900 m25 DATA:LAN300 500 m

J-02YS(C)Y 4x2xAWG22 26 Phone:JYSTY 2x2x0,6 mm 500 m27 HIFI:NYFAZ 2x0,5 mm (**) 500 m28 HIFI:NYFAZ 2x0,75 mm (**) 500 m29 HIFI:NYFAZ 2x1,5 mm (**) 500 m

* Helu Cable ** loudspeaker cable-The values and reference ranges refer to measurements with open cable ends. Measurements with short-cir-cuited ends lend to results with increased measurement errors.

Page 10: Medidor Longitude de Cabos por Reflexao de Impulso e Medidor Resistencia - Manual Sonigate

6.2 Performance of Cable Length Measurements within the Reference Range

To allow measurement of customer specific cabletypes, reference ranges have been integrated in theUNITEST Echometer 3000. The user can calibrateand save the reference ranges for the specific cabletypes.

6.2.1 Calibration of Customer Specific Cable Types

" Switch on UNITEST Echometer using thekey.

" Change to the reference range pressingthe key, e.g. reference level I. REF I isindicated on the display.

" Connect the red test lead to the red socket andthe black test lead to the black socket of the m-range.

" Connect two leads in parallel of the cable to bemeasured using the alligator clamps.

" To set the measurement unit (m/ft) press thekey. ''m'' or ''ft'' is indicated on the dis-

play. " Set measurement range selection dial to a free

memory location, e.g. switch position 1. 100.0 m is indicated on the display.

" Set the known length of the cable to be calibrat-ed using the or keys.

" Save the length of the reference cable pressingand keys. A simple signal

sound is audible.

Note cable type, switch position and reference level

Now the length measurement of the cable type cal-ibrated can be performed as described in section6.1.

The reference range accuracy values dependon the calibration accuracy of the customerspecific cable type. To comply with the accu-racy values in accordance with the specifiedtechnical data, the user must perform the cali-bration of the cable type as described in the in-struction manual.

6.2.2 Deletion of Reference RangesIf the user requires to assign the reference rangesto other cable types, the reference range must firstbe deleted.Free memory locations of the reference ranges areindicated by the display ''100.0''. Already assignedmemory locations are indicated by the display ''- --''.

" Set UNITEST Echometer 3000 to the referencerange to be deleted.''- - -'' is indicated on the display.

" Press the key once. The message''100.0'' is displayed.

" If the alligator clamps are open press and keys. An error measure-ment is generated and indicated by the message''Error''.

The reference range is deleted.

Shift

ShiftShift

m/ftm/ft

Start/Off

24

Performance of Measurements

m

Ω mEchometer 6

29

Page 11: Medidor Longitude de Cabos por Reflexao de Impulso e Medidor Resistencia - Manual Sonigate

Engl

ish

25

Performance of Measurements

6.3 Trouble-ShootingThe UNITEST Echometer 3000 enables to locatefaults in cables or conductors.

The user must check that the circuits are notlive !

" Switch on the UNITEST Echometer 3000 press-ing the key.

" Change to the reference range pressing thekey, e.g. Reference level I. REF I is in-

dicated on the display." Connect the red test lead to the red socket and

the black test lead to the black socket of the mrange.

" Connect the faulty leads of the conductor to bemeasured using the alligator clamps.

" Set the measurement unit (m/ft) pressing thekey. ''m'' or ''ft'' is displayed.

" Set the measurement range selection dial in ac-cordance with the conductor types in Tables I orII. The key is used to change to themeasurement ranges of Table II. ''TAB II'' is in-dicated on the display.

If the message ''- - -'' is displayed, the UNITESTEchometer is ready to perform measurements.

" Start the measurement pressing the key.

A single signal sound is audible and the mea-surement value of the conductor length in thepre-selected unit (m/ft) up to the error locationis displayed. If the message "COMP is dis-played, the leads have a short-circuit.

Short-circuited cable ends cause an increasingmeasurement error of the UNITEST Echometer3000.

6.4 Performance of Resistance MeasurementsThe UNITEST Echometer 3000 can easily and pre-cisely determine electrical resistances.

The user must check that the circuits are notlive !

" Switch on the UNITEST Echometer 3000 press-ing the key.

" Set the measurement range selection to the po-sition "Ohm”.

" The message ''>1999,9 Ohm'' is displayed. " Connect the red test lead to the red socket and

the black test lead to the black socket of the Ohmrange.

" Connect both ends of the resistance to be mea-sured using the alligator clamps.

The resistance measured is indicated on the display.If the resistance is below approx. 10 Ohm anacoustic signal is audible.

Start/Off

Start/Off

m/ftm/ft

Start/Off

R < 10Ω

Ω mEchometer 6

Ω

29

Start/off

20m

Unterbrechungm

KurzschlussΩ m

Echometer 6

29

m

COMP

20m

Page 12: Medidor Longitude de Cabos por Reflexao de Impulso e Medidor Resistencia - Manual Sonigate

7.0 Internal Measurement Data Memory / Cumulative Data memory

7.1 Saving Measurement Results" Perform the measurement as described in 6.2." The measurement result is indicated on the dis-

play. " Press the key. The result measured is

stored in the internal UNITEST Echometer 3000memory. A single acoustic signal is audible andthe memory location, e.g. 1 is briefly displayed.Furthermore, the symbol ''M'' is displayed andindicates that the measurement data has beensaved in the internal memory.

When the key is pressed again a dou-ble acoustic signal is audible and indicates thatfurther saving of this measurement is not pos-sible. After a new measurement, new resultscan be stored again.

7.1.1 Deletion of all Saved Measurement Results

Press the and keys.A brief acoustic signal is audible and ''0'' is brieflydisplayed to indicate that the memory has beencleared. The symbol "M” disappears from the dis-play.

The cumulative data memory is not affected by theclearing procedure.

7.1.2 Deletion of the Last Measurement Result

Press the and keys..An acoustic signal is audible and the last assignedmemory location is briefly displayed. E.g. Afterdeleting measurement no. 5 the number 4 is brieflydisplayed.

The cumulative data memory is not affected by theclearing procedure.

7.2 Adding and Saving Measurement ResultsThe UNITEST Echometer is equipped with a cumu-lative data memory enabling the measurement ofseveral cable rings of the same type , and to performcumulation of the values.

" Perform the measurements as described in sec-tion 6.2.

" The measurement result is displayed. " Press the key. The measurement re-

sult is added to the previously stored value in thecumulative data memory (for the first cumula-tion, the result is added to 0). The sum of themeasurement values is displayed. Additionally,the symbol "RES” is displayed and indicates thatthe measurement data has been stored in the cu-mulative data memory.

" Save cumulative data memory by pressing thekey. The memory location of the mea-

surement stored is displayed.

The cumulative data memory is cleared whenthe measurement range, the table/referencelevels, or the cable material are changed.

The previously saved sum in the data memory willnot be deleted.

To obtain correct cumulative results, it is im-portant to only add the same cable types.

Shift

Clear all

SendShift

26

Performance of Measurements

Page 13: Medidor Longitude de Cabos por Reflexao de Impulso e Medidor Resistencia - Manual Sonigate

Engl

ish

27

Maintenance

8.0 Transfer of Saved Data to a PCThe measurement data which has been saved in theUNITEST Echometer 3000 can be downloaded to aPC and further processed using the softwareUNITEST Report-Studio, available as an accessory.

After pressing the software function ''Daten ausMessgerät lesen'' ((download data from instru-ment)), the command is displayed to pushthe key on the UNITEST Echometer 3000.After pressing the key the UNITESTEchometer 3000 display briefly indicates the lastmemory location assigned.

The stored measurement data has been down-loaded to the PC and can now be used for furtherprocessing.

The instruction manual of the softwareUNITEST ”Report-Studio” must be compliedwith.

Option:Windows Software UNITEST ”Report-Studio”Order No.: 1207

9.0 MaintenanceWhen using the instrument in compliance with theinstruction manual, no special maintenance is re-quired.

9.1 CleaningIf the instrument is dirty after daily usage, it is ad-vised to clean it by using a humid cloth and a mildhousehold detergent.

Prior to cleaning, ensure that instrument isswitched off and disconnected from external volt-age supply and any other instruments connected(such as UUT, control instruments, etc.).

Never use acid detergents or dissolvants for clean-ing.

9.2 Battery Replacement

Prior to battery replacement, disconnect the in-strument from all test leads connected.

Only use the battery in compliance with thespecifications of the technical data section!

Please consider your environment when youdispose of your one-way batteries or accumu-lators. They belong in a rubbish dump for haz-ardous waste. In most cases, the batteries canbe returned to their point of sale.

Please, comply with the respective valid regu-lation regarding the return, recycling and dis-posal of used batteries and accumulators.

If an instrument is not used over an extendedtime period, the accumulators or batteriesmust be removed. Should the instrument becontaminated by leaking battery cells, the in-strument has to be returned for cleaning andinspection to the factory.

" Loosen the battery case screw on the bottom ofthe instrument.

" Carefully lift the battery case." Remove discharged battery." Insert new battery and respect correct polarity." Fix the battery case with the bottom of the cas-

ing.

Battery: 9V, IEC 6LR61Order No.: EZBATT000002

10.0 Calibration IntervalThe instrument has to be periodically calibrated byour service department in order to ensure the spec-ified accuracy of measurement results. We recom-mend a calibration interval of one year.

Send

Send

Page 14: Medidor Longitude de Cabos por Reflexao de Impulso e Medidor Resistencia - Manual Sonigate

11.0 Technical DataDisplay . . . . . . . . . . . . . . . 4 1/2 digit LCD, . . . . . . . . . . . . . . . . . . . . . . 19999 digits Measurement ranges . . . . 0...2000 m or . . . . . . . . . . . . . . . . . . . . . . attenuation limitResolution . . . . . . . . . . . . . 0,1 mBasic tolerance . . . . . . . . . ± (2% rdg. + 3 m)

Values and measurement ranges relate to opencable ends. Measurements with short-circuit-ed cable ends lead to measurement results withincreased measurement errors.

Resistance . . . . . . . . . . . . . 0...2000 OhmResolution . . . . . . . . . . . . . 0.1 OhmTolerance . . . . . . . . . . . . . . ± (2% rdg. + 5 digits)Continuity . . . . . . . . . . . . . < 10 OhmTest current . . . . . . . . . . . . approx. 1 mAAmbient Conditions:Operation temperature . . . 0°...40° C (0...70% . . . . . . . . . . . . . . . . . . . . . . rel. humidity)Storage temperature . . . . . -20...60° C (0...80% . . . . . . . . . . . . . . . . . . . . . . rel. humidity)Height above sea level . . . up to 2000 mMemory:Interface. . . . . . . . . . . . . . . RS 232Memory capacity . . . . . . . . 500 data strings . . . . . . . . . . . . . . . . . . . . . . (Length determination)Battery status display . . . . Battery symbol is dis-

playedAuto power off . . . . . . . . . . after approx. 5 min.Power supply. . . . . . . . . . . 1 x 9 V, IEC 6LR61Current consumption . . . . 25 mA/ max. 75 mA. . . . . . . . . . . . . . . . . . . . . . Display lightDimensions . . . . . . . . . . . . 193 x 91 x 50 mmWeight . . . . . . . . . . . . . . . . approx. 510 g

12 month WarrantyUNITEST instruments are subject to strict qualitycontrol. However, should the instrument functionimproperly during daily use, your are protected byour 12 months warranty (valid only with invoice).

We will repair free of charge any defects in work-manship or material, provided the instrument is re-turned unopened and untampered with, i.e. with un-damaged warranty label.

Any damage due to dropping or incorrect handlingare not covered by the warranty.

If the instrument shows failure following expirationof warranty, our service department can offer you aquick and economical repair.

28

Technical Data