Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 1

Morehouse Calibration Software Instruction Manual For use with DSC-USB, 4215, 4215-HS, 4215 Plus, and HADI indicators

Page 2: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 2

Table of Contents Setup ............................................................................................................................................................. 3

Install Morehouse Software Calibration ............................................................................................... 3

Windows Defender SmartScreen .......................................................................................................... 4

SSCE Runtime ........................................................................................................................................ 5

Indicator Drivers .................................................................................................................................... 5

Operating Morehouse Calibration Software ................................................................................................ 6

Launch Morehouse Calibration Software ............................................................................................. 6

Managing Load Cells ............................................................................................................................. 7

Finding Load Cell Coefficients ............................................................................................................... 9

Managing Indicators ........................................................................................................................... 11

Calibrating A Load Cell ................................................................................................................................ 12

Calibration Controls ............................................................................................................................ 13

Calibrate Load Cell .............................................................................................................................. 14

Data Collection .................................................................................................................................... 15

FAQ .............................................................................................................................................................. 16

SQL Server Compact Exception ........................................................................................................... 16

Removing Force Units ......................................................................................................................... 17

Force Summation ................................................................................................................................ 18

Missing Indicator ................................................................................................................................. 19

Missing HADI Indicator ........................................................................................................................ 20

Resetting COM Ports ........................................................................................................................... 21

Page 3: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 3

Setup The latest version of the Morehouse Calibration Software can be found at https://mhforce.com/wp-

content/uploads/2021/04/Morehouse_Calibration_Software.zip.

If downloading from our site, it will need to be unzipped prior to continuing with the installation.

Install Morehouse Software Calibration Navigate to the USB drive, CD, or the unzipped files and double click on the Setup file.

Page 4: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 4

Windows Defender SmartScreen The installer may be blocked because it is unrecognized. Simply click More info and then Run anyway.

Follow all on screen instructions. Once successfully installed, the following screen should appear.

Page 5: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 5

SSCE Runtime The Morehouse Calibration Software requires the SQL Server Compact Edition runtime to be installed.

The runtime is named SSCERuntime_x86.exe or SSCERuntime_x64.exe. Select the one that corresponds

to your CPU’s architecture (typically x64).

Follow the instructions to complete installation. A similar screen should appear after a successful

installation.

Indicator Drivers Install the necessary drivers for the indicators to be used (DSC, HADI, and/or 4215). The driver installers

are in the Drivers/<Indicator Name> directory.

Page 6: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 6

Operating Morehouse Calibration Software Launch Morehouse Calibration Software By default, Morehouse Calibration Software is installed at C:\Program Files (x86)\Morehouse

Calibration Software\Morehouse Calibration Software. Double click the Morehouse Calibration

Software file to run the program.

Alternatively, a shortcut to the program should have been installed on the desktop.

Page 7: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 7

Managing Load Cells Click on Enter Calibration Coefficients to add, edit, and/or remove load cells.

Click Add to add a new load cell.

Page 8: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 8

Select the Coefficient Units from the drop down. And then press OK.

A new load cell is created with an empty name.

1. Enter a name where <Name Here> is.

2. Select Compression (or Tension) Units

3. Enter at least 2 coefficients

4. Optionally import from Cert via clipboard

5. Repeat for the other load type if necessary

Select any existing load cell to modify the coefficients as necessary or click the Delete button to remove

the load cell from the list.

Page 9: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 9

Finding Load Cell Coefficients Coefficients can be found on the load cell’s certificate. Verify the load direction at the top of the page.

The coefficients needed are listed as Bn and should be entered in Bn. Refer to images for an example.

Select Proceed to finish managing the load cells.

Page 10: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 10

Optional: Import from Certificate

To help reduce typos, the From Cert button was added to allow effective copy/pasting of the

coefficients. If the PC has a copy of the digital certificate on it, you can select the text as shown below

and copy (CTRL + V or Right Click->Copy) the text to your clipboard. From there, just press the From Cert

button to fill all the coefficients in.

Note: The only requirements are that the copied text includes the “B₀” text and all coefficients are

selected. There are no issues with additional text being accidentally selected.

Page 11: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 11

Managing Indicators Click on Search Indicators to detect and modify connected indicators.

1. Select an indicator from the list

2. Click the drop-down arrow to select the load cell

3. Select the currently connected load cell.

Click Proceed to save any changes.

Page 12: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 12

Calibrating A Load Cell Select the indicator connected to a load cell to calibrate. The load cell’s name is listed below the

indicator.

The indicator and load cell combo will be added to the calibration display. Please verify it is correct

before proceeding.

Page 13: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 13

Calibration Controls 1. Force/Mass Display – Direct read of force or mass

2. Unit of Force/Mass selection – Select the unit of measurement

3. Display Precision – Adjusts the number of digits shown on the Display

4. Reading Interval – How frequent the display is updated

5. Indicator Communication Toggle – Turn on to communicate with indicator

6. Tare All Button – Tare all active indicators simultaneously

7. Force/Mass Toggle – Turn on to calculate mass using local gravity value

8. Local Gravity – Used for mass calculations

9. Capture Data – Logs the current reading in the selected cell

10. Indicator Force / mV/V Readout – Display the indicator’s current reading

11. Indicator Force / mV/V Toggle – Toggle to view a raw or force reading of the indicator

12. Compression/Tension Toggle – Toggle based on load direction

13. Add to Force Sum – Add this indicator’s readings to the total force measurement

14. Current Indicator – Visual cue of selected indicator

15. Current Load Cell – Coefficients being used

16. Tare – Zero the readings

17. Remove Indicator – The indicator and load cell return to the indicator list

18. Indicator List – List of indicators that are currently connected with the name of the assigned load cell

19. Data Logging – Enter forces to be applied; select cell to capture data to

20. Export to Excel – Create a spreadsheet of the data logged

21. Clear Data – Clear all data from Data Logging

Page 14: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 14

Calibrate Load Cell 1. Enter the various forces to be applied. A new row will be added once upon selecting the empty

cell

2. Select the corresponding cell to record data in.

3. Activate this toggle to enable data to be read from the indicator

4. Press Capture Data to log the indicator’s reading. The next cell down (2) will be highlighted

automatically. The program will record both the force reading and the mV/V reading of the

indicator. Select the drop down to choose between the two.

5. Adjust the force and repeat step 4, as necessary. Once all forces have been recorded for a run,

rotate the load cell, as necessary. Repeat step 2 for all 3 runs.

Page 15: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 15

Data Collection Once data desired has been collected, press Export to Excel to generate a spreadsheet with all the

logged data. Exporting to Excel requires Excel to be installed on the same computer.

The following is an example of the outputted data in force units.

Page 16: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 16

FAQ SQL Server Compact Exception Question: I am receiving an unhandled exception mentioning SQL Server Compact. What causes it and

how do I fix it?

Answer: The exception is due to the program being unable to load the required database components because they are not installed, or the incorrect version is installed. The fix is to run the installer that is included with the program. Refer to the SSCE Runtime section of this document for instructions.

Page 17: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 17

Removing Force Units Question: Our company does more than just Force and Mass calibration/measurements. Is there a way

to remove the unit label to help remove confusion for our technicians?

Answer: Yes. Enabling the “Ignore Calibration Units” checkbox on the Start/Setup tab, all unit labels are

removed from the program. Due to a technical limitation though, the Managing Load Cells window still

requires a unit to be selected.

Please note that when this feature is enabled, the unit of Force/Mass dropdown is disabled and simply

displays “As Calibrated.” Additionally, the Mass toggle and Gravity text box are also disabled. Finally,

any points captured while this feature is enabled will not record any unit of measurement data.

Page 18: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 18

Force Summation Question: Can I use multiple indicators and load cells to calculate force?

Answer: Yes. Force summation is supported and up to 4 indicators can be used at a time.

1. Ensure all load cell coefficients and indicators are set up. See Managing Load Cells and Managing

Indicators for instructions. 2. Select the indicators to be used in the calculation. See Calibrating a Load Cell for instructions.

a. Each selected indicator/load cell will have its own region for control and display

purposes 3. An indicator can be ignored for temporarily by unchecking the Add to Force Sum checkbox for

the indicator. 4. Turn on the Force/Mass toggle. 5. Capture data and export as before. 6. The exported data will contain both devices and how each of them contributed to the force

measurement.

Page 19: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 19

Missing Indicator Question: Why isn’t my indicator showing up?

Answer:

1. If using a Morehouse 4215 indicator, ensure that the “Add Line Feeds” option is enabled via the

front panel controls.

2. The indicator may not be properly connected. Please ensure the connection to the computer is

good.

3. The indicator’s driver may not have been installed. Please refer to Indicator Drivers in this document

for further instructions.

4. The indicator may be in use by another program. Please stop any other programs that may be using

the indicator or restart the PC.

5. As a last attempt, please see Resetting COM Ports.

Page 20: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 20

Missing HADI Indicator Question: Why isn’t my HADI indicator being detected?

Answer:

1. Complete the steps in the Missing Indicator first

2. Open the computer’s Device Manager via Control Panel or right-clicking on the start menu and

selecting Device Manager

3. Find the Universal Serial Bus controllers device category and expand to view all devices

4. Locate the USB Serial Converter device and right click on it and select Properties.

5. Navigate to the Advanced tab and enable Load VCP

Page 21: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 21

Resetting COM Ports Question: I am having unexplained indicator issues, how do I reset my COM ports?

Answer:

WARNING: These steps will also remove any other COM devices. Use with caution!

1. Open the computer’s Device Manager via Control Panel or right-clicking on the start menu and

selecting Device Manager

2. Find the Ports (COM & LPT) device category and expand to view all devices

a. USB Serial Port (HADI), USB COM Port (4215), or USB DSC Port (DSC) may be listed

Page 22: Morehouse Calibration Software Instruction Manual

Morehouse Calibration Software

Instruction Manual PM-4114 4/2022 Page 22

3. Select each listed device from the Ports (COM & LPT) category and right click and select Uninstall

Device

4. Disconnect the indicator’s USB cable from the PC and then reconnect it. If it is not automatically

detected, click on the Scan for hardware changes button at the top of the device manager