REPUBLIC OF SLOVENIA MINISTRY FOR THE ENVIRONMENT AND SPATIAL PLANNING SLOVENIAN ENVIRONMENT AGENCY Vojkova 1b, 1000 Ljubljana p.p. 2608, tel.: +386(0)1 478 40 00 fax.: +386(0)1 478 40 52 21 st – 25 th November 2011 Training Workshop on Metrology, Melbourne, Australia M.Sc. Drago Groselj Head of Calibration Laboratory Service Uncertainty of relative humidity calibration WMO CIMO Training Workshop on Metrology for the English- speaking countries of Region V (South-West Pacific) Melbourne, 21st - 25th November 2011

Upload: khoi

Post on 07-Jan-2016

58 views

Category:

Documents


0 download

DESCRIPTION

WMO CIMO Training Workshop on Metrology for the English-speaking countries of Region V (South-West Pacific) Melbourne, 21st - 25th November 2011. Uncertainty of r elative humidity calibration. M.Sc. Drago Groselj Head of Calibration Laboratory Service. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: M.Sc. Drago Groselj Head of Calibration Laboratory Service

REPUBLIC OF SLOVENIA

MINISTRY FOR THE ENVIRONMENT AND SPATIAL PLANNINGSLOVENIAN ENVIRONMENT AGENCY

Vojkova 1b, 1000 Ljubljana p.p. 2608, tel.: +386(0)1 478 40 00 fax.: +386(0)1 478 40 52

21st – 25th November 2011 Training Workshop on Metrology, Melbourne, Australia

M.Sc. Drago Groselj

Head of Calibration Laboratory Service

Uncertainty of relative humidity calibration

WMO CIMO Training Workshop on Metrology for the English-speaking countries of Region V (South-West Pacific)

Melbourne, 21st - 25th November 2011

Page 2: M.Sc. Drago Groselj Head of Calibration Laboratory Service

21st – 25th November 2011 Training Workshop on Metrology, Melbourne, Australia

Humidity “generator” and reference standard

Its role is producing stable and homogeneous humidity environment Actual value of the humidity is measured with a reference dew-point

meter Types of humidity generators (two pressure, two temperature, mixed

flow, climatic chamber, ...)

Reference standard

Reference standard for humidity measurement is usually dew point meter calibrated externaly (in other laboratory)

It can be also RH hygrometer and psychrometer, but the uncertainty is then typically larger

Page 3: M.Sc. Drago Groselj Head of Calibration Laboratory Service

21st – 25th November 2011 Training Workshop on Metrology, Melbourne, Australia

Measurement methods

Depending on the DUC (RH/dew point) Parallel sampling Serial sampling Measurement of air temperature Potential pressure drops Leakage

Page 4: M.Sc. Drago Groselj Head of Calibration Laboratory Service

21st – 25th November 2011 Training Workshop on Metrology, Melbourne, Australia

Calibration with reference standard

Set-up for dew-point calibration(Device under calibration – DUC - is dew-point meter)

generator

DUC

3,45°C

Reference dew-point sensor

Page 5: M.Sc. Drago Groselj Head of Calibration Laboratory Service

21st – 25th November 2011 Training Workshop on Metrology, Melbourne, Australia

Calibration with reference standard

generator

DUC (RH)

Set-up for RH calibration(DUC is RH hygrometer)

thermometer

3,45°C

Reference dew-point sensor

Page 6: M.Sc. Drago Groselj Head of Calibration Laboratory Service

21st – 25th November 2011 Training Workshop on Metrology, Melbourne, Australia

In calibration uncertainty budget calculation following uncertainty components should be considered:

generatorDUC

• due to generation of humidity• due to uncertainty of reference standard

• due to DUC

• due to calibration method and ambient conditions

Referencestandard

t90

Calibration with reference standard

Page 7: M.Sc. Drago Groselj Head of Calibration Laboratory Service

21st – 25th November 2011 Training Workshop on Metrology, Melbourne, Australia

Uncertainty of humidity generation

time instability of generated humidity:instability can be measured during the calibration with stable reference humidity and/or temperature standard. Standard uncertainty due to instability is taken as a combination of standard deviation of both dew-point and temperature

inhomogeneity of generated humidity

The reference value (RV) is measured by calibrated dew-point sensor, so only the deviations from the measured RV need to be characterised.

The major uncertainty contributions are therefore:

Page 8: M.Sc. Drago Groselj Head of Calibration Laboratory Service

21st – 25th November 2011 Training Workshop on Metrology, Melbourne, Australia

Inhomogenity of generated humidity - DP

almost insignificant in dew-point calibration (if the inlet of the sampling tube of standard and DUC are in parallel and close to each other) problems, if reference standard and DUC as dew-point meter are sampling in serial (pressure drop, contamination,...)

DUC

Referencestandard generator

Page 9: M.Sc. Drago Groselj Head of Calibration Laboratory Service

21st – 25th November 2011 Training Workshop on Metrology, Melbourne, Australia

in RH calibration – large influence of temperature on relative humidity Therefore temperature inhomogeneity needs to be quantified –

can be very significant:

1. Temperature gradients can be evaluated previous to calibration – less expensive possibility...

2. ... or each time during the calibration (the set of thermometers has to be used all the time)

the second one is better, because the influence of DUC can also be determined, however more expensive

To evaluate inhomogeneity, thermometers are typically placed in corners of calibration area and one in the geometrical center.

Inhomogenity of generated humidity - RH

Page 10: M.Sc. Drago Groselj Head of Calibration Laboratory Service

21st – 25th November 2011 Training Workshop on Metrology, Melbourne, Australia

• Typical uncertainties of RH due to temperature inhomogeneity – from 0,1%r.h. to several %r.h.

• The inhomogeneity can be reduced if smaller calibration space is used or appointed

34 times smaller or more

Inhomogenity of generated humidity - RH

Page 11: M.Sc. Drago Groselj Head of Calibration Laboratory Service

21st – 25th November 2011 Training Workshop on Metrology, Melbourne, Australia

Uncertainty of reference standard

In general the following uncertainty contributions are to be determined:

• uncertainty, taken from the calibration certificate (dew-point thermometer & thermometer for air temperature)

• uncertainty of electrical instrument (bridge, multimeter,...)

• uncertainty of calculation of relative humidity from dewpoint,temperature (water sat.pressure eq.,enh.f.eq.)

• standard’s drift

• resolution (if not included in certificate)

Page 12: M.Sc. Drago Groselj Head of Calibration Laboratory Service

21st – 25th November 2011 Training Workshop on Metrology, Melbourne, Australia

Uncertainty of reference standard

In RH calibrations, if the reference is dew-point sensor, then the temperature and dew-point contributions have to be considered separately.

22

,

dpa tdp

ta

refrh ut

RHu

t

RHu

Page 13: M.Sc. Drago Groselj Head of Calibration Laboratory Service

21st – 25th November 2011 Training Workshop on Metrology, Melbourne, Australia

Sources of uncertainty of DUC

• instability of DUC’s readings

• resolution

• hysteresis

• drift (to be added by the end user)

• uncertainty of electrical devices (bridge, multimeter)

• drift of electrical devices

• the influence of DUC to calibration environment (self-heat, humidification – psychrometer)

Page 14: M.Sc. Drago Groselj Head of Calibration Laboratory Service

21st – 25th November 2011 Training Workshop on Metrology, Melbourne, Australia

Other sources of uncertainty (errors)

• leakage (stem of the sensor, cables)

• contamination (salts, oil, dirt) • difference between calibration measurement range and actual range (temperature, pressure,...)• reproducibility

Page 15: M.Sc. Drago Groselj Head of Calibration Laboratory Service

21st – 25th November 2011 Training Workshop on Metrology, Melbourne, Australia

Uncertainty sourceUncertainty contribution

U()

Probabilty law

Coefficient Standard

uncertainty u()

A-tip reference dew point 0.2 [%] Normal 0.5 0.1 [%]

A-tip DUC 0.2 [%] Normal 0.5 0.1 [%]

Time instability of humidity generator

0.91 [%] Normal 0.5 0.46 [%]

Spatial unhomogenity of humidity generator

0.84 [%] Normal 0.5 0.42 [%]

Uncertainty of reference dew point

0.017*RH+0.2 Normal 0.5 0.53 [%] (0.0085*RH+0.1)

DUT 0.1 [%] Rectangular 1 0.1 [%]

Multimeter 64 [V] Rectangular ×100 0.0074 [%]

Expanded uncertainty 1.71 [%]

* Uncertainty calculation at Ta=20°C and RH=95%

Example of uncertainty budget calculation

Page 16: M.Sc. Drago Groselj Head of Calibration Laboratory Service

21st – 25th November 2011 Training Workshop on Metrology, Melbourne, Australia

Reference

• ARSO, Technical instruction, Calibration of electronic hygrometers in humidity generator, TN.3.00.1, 2009

• Jovan Bojkovski, Domen Hudoklin, Uncertainty in calibrations with secondary standard, , Laboratory of Metrology and Quality (LMK), Faculty of Electrical Engineering (FE), University of Ljubljana