Routine Testing SOP for Periodic SmartCal Test of Moisture Analyzers Title: SOP for Periodic SmartCal test with Halogen Moisture Analyzers Document No.: Author: METTLER TOLEDO Laboratory & Weighing Technologies Number of Pages: 4 Controls Replacement Document: N/A Reason for Revision: N/A Release Date: 11. Oct. 2011 Release for Routine Operation Reviewed by: Date: Signature: Approved by: Date: Signature:

Upload: others

Post on 28-May-2020

6 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: of Moisture Analyzers - METTLER TOLEDO2 of 4 SOP for Periodic SmartCal Test METTLER TOLEDO Routine Testing of Moisture Analyzers General The SmartCal test is a rapid test which checks

Rout

ine

Test

ing

SOP for Periodic SmartCal™ Test of Moisture Analyzers

Title: SOP for Periodic SmartCal test with Halogen Moisture Analyzers

Document No.:

Author: METTLER TOLEDO Laboratory & Weighing Technologies

Number of Pages: 4

Controls

Replacement Document: N/A

Reason for Revision: N/A

Release Date: 11. Oct. 2011

Release for Routine Operation

Reviewed by:

Date:

Signature:

Approved by:

Date:

Signature:

Page 2: of Moisture Analyzers - METTLER TOLEDO2 of 4 SOP for Periodic SmartCal Test METTLER TOLEDO Routine Testing of Moisture Analyzers General The SmartCal test is a rapid test which checks

2 of 4 SOP for Periodic SmartCal TestMETTLER TOLEDO

2 of 4 SOP for Periodic SmartCal TestMETTLER TOLEDO

Rout

ine

Test

ing

of M

oist

ure

Anal

yzer

s General

The SmartCal test is a rapid test which checks the moisture analyzer’s overall functionality. SmartCal is a granular substance which releases a defined amount of moisture depending on the selected drying temperature. Intended use is the verification of METTLER TOLEDO Halogen Moisture Analyzers. However, it can also be used for other Moisture Analyzers, but the control limits are only valid for METTLER TOLEDO Halogen Moisture Analyzers.

Basic Rules for Handling Moisture Analyzers

•Beforeusingamoistureanalyzer,makesuretheinstrumentwasleftonpowerforasufficientperiodoftime(mentioned in the user manual).

•Makesurethemoistureanalyzerisleveled.•Minimizeenvironmentalinfluences,e.g.draftordirectsunlight.•Distributesamplesgentlyandevenlyonthesamplepan.

Basic Rules for Handling SmartCal

Important to know•Thegranulartestsubstanceisdouble-packedinastickinsideablisterpack.•Thestickshouldonlyberemovedfromtheblisterpackimmediatelybeforeuse.•Donotuseanydamagedorexpiredsticks.•AftertheSmartCaltest,theusedtestsubstancecanbedisposedofasnormalwaste.

How to Store SmartCal•SmartCalshouldbestoredinthesameroomasthemoistureanalyzertheyareusedwith,sincetemperature

differences between the test substance and their surroundings can lead to measurement errors.

How to Handle SmartCal•TheSmartCaltestresultrequiresanormalizationtoambientconditions(%RHandroomtemperature).•WhenperformingaSmartCaltest,usethesamplepanhandler.•Usethecompletecontentsofthestick(approx.weight8.5g).

Page 3: of Moisture Analyzers - METTLER TOLEDO2 of 4 SOP for Periodic SmartCal Test METTLER TOLEDO Routine Testing of Moisture Analyzers General The SmartCal test is a rapid test which checks

3 of 4SOP for Periodic SmartCal Test METTLER TOLEDO

3 of 4SOP for Periodic SmartCal Test METTLER TOLEDO

1) – Values within the control limit: no action is necessary. –Valuesbeyondthecontrollimitshowthatthemoistureanalyzerisnolongerundercontrolandimmediateactionisthereforerequired.

SmartCal Routine Test

Preparation

•Beforethetestisperformed,SmartCaltestsubstanceandthethermohygrometermustbeacclimatizedtotheambient temperature of the moisture analyzer.

•Whenthetestisperformed,theheatingunitofthemoistureanalyzershouldbeatroomtemperature.•Ifasensitivitytestneedstobedoneaswell,pleaseperformthesensitivitytestfirst(toavoidwaitingtimefor

cooling down).

Test Procedure

•UseinstrumentpredefinedtestproceduresorsetinstrumenttotheSmartCaltestparameters:– Switch-offtime:10minutes– Drying program: standard drying– Display:%MC– Drying temperature: 70, 100, 130 or 160 °C (Select the temperature closest to the drying temperature normally used)

•Placethesamplepanhandlerwiththealuminumsampledishintheinstrumentandtare.•RemoveaSmartCalstickfromtheblisterpack,openitanddistributetheentirecontentsevenlyoverthe

sample pan (whole pan covered with granules).•Startthetestmeasurementimmediatelyafterdistributingthesample.•Readthedisplayedresult(%MC)fromtheinstrumentandtheambientconditionsfromthethermohygrometer.

Evaluation

•Normalizethedisplayedresulttostandardconditions(20°C/50%RH)eithermanuallywiththenormalizationtableorautomaticallywiththeExcelmeasurementreport.

•Evaluatewhetherthenormalizedresultexceedsthedefinedcontrollimits.1)

Deviation

Control Limit 1) • Ifthecontrollimitisexceeded,reduceerrorsources(seeSmartCalusermanual)andrepeatthetest(after

cooling down of the moisture analyzer).• Ifthecontrollimitisexceededagain,reportthatthecontrollimitwasnotmettothelaboratorysupervisoror

the person responsible for the moisture analyzer.•Letthemoistureanalyzercooldown,performweightandthentemperatureadjustment,letthemoisture

analyzer cool down and repeat the SmartCal test.• Ifthecontrollimitisstillexceeded,reporttheproblemtothelaboratorysupervisororthepersonresponsible

of the moisture analyzer. Optionally, contact METTLER TOLEDO’s service organization for advice.•Markthemoistureanalyzeras“outofcontrollimits”.

Page 4: of Moisture Analyzers - METTLER TOLEDO2 of 4 SOP for Periodic SmartCal Test METTLER TOLEDO Routine Testing of Moisture Analyzers General The SmartCal test is a rapid test which checks

4 of 4 SOP for Periodic SmartCal TestMETTLER TOLEDO

Recommended Control Limits for SmartCal Test

Control limits are valid for METTLER TOLEDO Halogen Moisture Analyzers.

Drying Temperature cSmartCal SmartCal

70 °C 3.3-4.3%MC 3.2-4.4%MC

100 °C 5.3-6.3%MC 5.2-6.4%MC

130 °C 7.5-8.7%MC 7.4-8.8%MC

160 °C 10.0-11.6%MC 9.9-11.7%MC

Mettler-Toledo AGLaboratory & Weighing Technologies CH-8606GreifenseeTel. +41 44 944 22 11Fax.+41449443170

Subjecttotechnicalchanges©11/2011Mettler-ToledoAGPrintedinSwitzerland30031480GlobalMarComSwitzerland

For more informationwww.mt.com/moisture