Routine Testing SOP for Periodic Sensitivity Tests of Moisture Analyzers Title: SOP for Periodic Sensitivity Tests of Moisture Analyzers Document No.: Author: METTLER TOLEDO Laboratory & Weighing Technologies Number of Pages: 4 Controls Replacement Document: N/A Reason for Revision: N/A Release Date: 11. Oct. 2011 Release for Routine Operation Reviewed by: Date: Signature: Approved by: Date: Signature:

Upload: others

Post on 15-Aug-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: of Moisture Analyzers - Mettler Toledo€¦ · SOP for Periodic Sensitivity Test 3 of 4 METTLER TOLEDO 1) – Value within the warning limit: no action is necessary. – Value between

Rout

ine

Test

ing

SOP for Periodic Sensitivity Tests of Moisture Analyzers

Title: SOP for Periodic Sensitivity Tests of Moisture Analyzers

Document No.:

Author: METTLER TOLEDO Laboratory & Weighing Technologies

Number of Pages: 4

Controls

Replacement Document: N/A

Reason for Revision: N/A

Release Date: 11. Oct. 2011

Release for Routine Operation

Reviewed by:

Date:

Signature:

Approved by:

Date:

Signature:

Page 2: of Moisture Analyzers - Mettler Toledo€¦ · SOP for Periodic Sensitivity Test 3 of 4 METTLER TOLEDO 1) – Value within the warning limit: no action is necessary. – Value between

2 of 4 SOP for Periodic Sensitivity TestMETTLER TOLEDO

Rout

ine

Test

ing

of M

oist

ure

Anal

yzer

s General

To test the sensitivity of the weighing unit of a moisture analyzer, a large test weight (between 50% and 100% of capacity) is recommended.

Basic Rules for Handling Moisture Analyzers

•Beforeusingamoistureanalyzer,makesuretheinstrumentwasleftonpowerforasufficientperiodoftime(mentioned in the user manual).

•Makesurethemoistureanalyzerisleveled.•Minimizeenvironmentalinfluences,e.g.airdraft,vibrationsordirectsunlight.

Basic Rules for Handling Weights

Important to know•Onlyanexternalweightwithcalibrationcertificatecanmaketheweighingunitofamoistureanalyzeratrace-

able piece of equipment.•Weightsshouldalwaysbeplacedgentlyonthesamplepanholderandputbackimmediatelyintheirstorageplaceafteruse.Theymaybeparkedonacleanpieceofpaper.

•Weights(sincetheyarealsopartofmeasuringequipment)needtobere-calibratedatspecifiedintervals (ISO 9001).

•Anydamage,whichmighthaveaffectedthevalueoftheweight,shouldtriggeranimmediatere-calibration.METTLER TOLEDO’s calibration services will give advice on this.

How to Store Weights•Weightsshouldbestoredintheiroriginalbox.•Weightsshouldbestoredatroomtemperature,sincetemperaturedifferencesbetweenweightsandtheir

surrounding could lead to measurement errors.•Weightsthathavenotbeenstoredatthesametemperatureneedacclimatizationbeforeuse.

How to Move Weights•Weightsshouldonlybehandledwithappropriatetoolssuchastweezers,forks,handlesorgloves(see

METTLER TOLEDO’s accessories for weights).•Thesetoolsshouldbeexclusivelyusedfortransferringweights,duetopossiblecontamination.

Page 3: of Moisture Analyzers - Mettler Toledo€¦ · SOP for Periodic Sensitivity Test 3 of 4 METTLER TOLEDO 1) – Value within the warning limit: no action is necessary. – Value between

3 of 4SOP for Periodic Sensitivity Test METTLER TOLEDO

1) – Value within the warning limit: no action is necessary. –Valuebetweenthewarningandcontrollimitarewithinthetolerancebutmustbekeptundersurveillance.

2) – Value within the control limit, see 1)

– Value beyond the control limit show that weighing process is no longer under control and immediate action is therefore required.

Sensitivity Test

Preparation

•Beforethetestisperformed,thetestweightsmustbeacclimatizedtotheambienttemperatureofthemoistureanalyzer.

•Whenthetestisperformed,theheatingunitofthemoistureanalyzershouldbeatroomtemperature.

Test Procedure HX and HS Models

•Starttestmodeandfollowtheinstructions.

Test Procedure other Moisture Analyzers

•Removethesamplepan.•Tarethebalance.•Placethetestweight(s)onthesamplepanholder.•Readthestablevaluefromthedisplayandnoteit.

Evaluation

•Evaluatewhetherthenotedvalueexceedsthedefined“WarningLimit”.1)

•Evaluatewhetherthenotedvalueexceedsthedefined“ControlLimit”.2)

Deviation

Warning Limit 1) (where defined) • Ifthewarninglimitisexceeded,repeatthetest.• Ifthewarninglimitisexceededagain,reportthatthewarninglimitwasnotmettothelaboratorysupervisoror

the person responsible for the moisture analyzer.•Levelthemoistureanalyzer,performadjustmentwithbuilt-in(ifapplicable)orexternalweightandrepeatthe

test.• Ifthewarninglimitisstillexceeded,reporttheproblemtothelaboratorysupervisororthepersonresponsible

of the moisture analyzer. Optionally, contact METTLER TOLEDO’s service organization for advice.

Control Limit 2)

• Ifthecontrollimitisexceeded,reporttheproblemtothelaboratorysupervisororthepersonresponsibleofthemoisture analyzer.

•Markthemoistureanalyzeras“outofcontrollimits”.•ContactMETTLERTOLEDOserviceorganizationforadvice.

Page 4: of Moisture Analyzers - Mettler Toledo€¦ · SOP for Periodic Sensitivity Test 3 of 4 METTLER TOLEDO 1) – Value within the warning limit: no action is necessary. – Value between

Mettler-Toledo AGLaboratory & Weighing Technologies CH-8606GreifenseeTel. +41 44 944 22 11Fax.+41449443170

Subjecttotechnicalchanges©11/2011Mettler-ToledoAGPrintedinSwitzerland30031488GlobalMarComSwitzerland

RecommendedTestWeights,WarningandControlLimits

Recommendation is based on 0.1% weighing accuracy and safety factor of 2. The absolute tolerances depend on the mass of the test weight.Various weighing parameters can contribute to the measurement uncertainty of a weighing result. With the exceptionoftherepeatabilitytest,allcontrollimitsaresetto1/2oftheweighingaccuracyinordertohave a security reserve accounting for any other influences such as eccentricity and linearity on the result (control limit = test weight * weighing accuracy / 2)

Moisture Analyzer HX/HS HR/HG HB/MJ

Test weight 100 g 50 g 20 g

Warning Limit ± 25 mg ± 12.5 mg ± 5 mg

ControlLimit ± 50 mg ± 25 mg ± 10 mg

For more informationwww.mt.com/moisture