Operating InstructionsCooling Jacket Advanced
Cooling Jacket Advanced Cooling Jacket Advanced Extended
MICRO-EPSILONMESSTECHNIKGmbH & Co. KGKönigbacher Strasse 15
94496 Ortenburg / Germany
Tel. +49 (0) 8542 / 168-0 Fax +49 (0) 8542 / 168-90e-mail [email protected]
Cooling housing for TIM series, video pyrometer and laser pyrometer at high ambient temperatures
3 Table of contents
Table of contents
1 General Notes ..................................................................................................................................... 7
1.1 Intended Use ................................................................................................................................ 7
1.2 Warranty ....................................................................................................................................... 8
1.3 Scope of Supply ............................................................................................................................ 9
1.3.1 Versions ................................................................................................................................ 9
1.4 Mounting Accessories ................................................................................................................. 10
1.4.1 Accessories for TIM NetBox ............................................................................................... 10
1.4.2 Accessories for USB Server Gigabit ................................................................................... 11
2 Technical Data .................................................................................................................................. 12
2.1 General Specifications ................................................................................................................ 12
2.1.1 Focusing Unit and Front Part .............................................................................................. 13
2.2 Accessories................................................................................................................................. 16
4
2.2.1 High Temperature Cable .................................................................................................... 16
2.2.2 Protection Window .............................................................................................................. 17
2.2.3 Mounting Flange ................................................................................................................. 19
2.2.4 Laminar Air Purge ............................................................................................................... 19
2.2.5 Additional Accessories ........................................................................................................ 22
2.3 Dimensions ................................................................................................................................. 23
2.3.1 Cooling Jacket Advanced ................................................................................................... 23
2.3.2 Cooling Jacket Advanced Extended ................................................................................... 27
2.3.3 Mounting Flange (Accessories) TM-CJAMF-TIM ............................................................... 31
2.4 Fittings ........................................................................................................................................ 32
2.4.1 Cooling Water Fitting .......................................................................................................... 32
2.4.2 Air Purge Collar................................................................................................................... 32
2.4.3 Cable Glands on the Back .................................................................................................. 34
5 Table of contents
2.5 Cooling Properties ...................................................................................................................... 35
2.5.1 Condensation ...................................................................................................................... 36
3 Installation ......................................................................................................................................... 39
3.1 Installation ................................................................................................................................... 39
3.2 Installation of TIM Camera, Infrared Video Thermometer and Infrared Thermometer ............... 40
3.2.1 Assembling of the Focusing Unit ........................................................................................ 40
3.2.2 Assembling of the Front Part .............................................................................................. 44
3.2.3 Mounting of the Protective Window at Focusing Unit ......................................................... 47
3.2.4 Assembly of the Laminar Air Purge .................................................................................... 49
3.2.5 Assembling to the Cooling Jacket Advanced Standard Version ........................................ 53
3.2.6 Assembling to the Cooling Jacket Advanced Extended ..................................................... 64
4 Example of Installation ..................................................................................................................... 77
6
7 General notes
1 General Notes
1.1 Intended Use
The cooling housing Cooling Jacket Advanced is intended to use for TIM series, video thermometers CTVideo and
CSVideo as well as CTLaser and CSLaser for application at high ambient temperatures.
► All accessories can be ordered according to the referred part numbers in
brackets ( ).
Read the manual carefully before the initial start-up. The producer reserves the right to
change the herein described specifications in case of technical advance of the product.
In case of problems or questions which may arise when you use the infrared camera, please
contact our service department.
8
1.2 Warranty
All components of the device have been checked and tested for perfect function in the factory. In the unlikely event
that errors should occur despite our thorough quality control, this should be reported immediately to MICRO-
EPSILON.
The warranty period lasts 12 months following the day of shipment. Defective parts, except wear parts, will be
repaired or replaced free of charge within this period if you return the device free of cost to MICRO-EPSILON. This
warranty does not apply to damage resulting from abuse of the equipment and devices, from forceful handling or
installation of the devices or from repair or modifications performed by third parties.
No other claims, except as warranted, are accepted. The terms of the purchasing contract apply in full. MICRO-
EPSILON will specifically not be responsible for eventual consequential damages. MICRO-EPSILON always strives
to supply the customers with the finest and most advanced equipment. Development and refinement is therefore
performed continuously and the right to design changes without prior notice is accordingly reserved.
For translations in other languages, the data and statements in the German language operation manual are to be
taken as authoritative.
9 General notes
1.3 Scope of Supply
1.3.1 Versions
Cooling Jacket Advanced
Cooling Jacket Advanced for TIM series (Part-No.: TM-CJA-TIM), consisting of housing and chassis
The focusing unit or the front part must be ordered separately.
Cooling Jacket Advanced for CSLaser, CTLaser as well as CTVideo and CSVideo (Part-No.: TM-CJA-CTL)
The front part must be ordered separately.
Installation instructions Cooling Jacket Advanced Extended
Cooling Jacket Advanced for TIM series (Part-No.: TM-CJAExxx-TIM), consisting of housing and chassis
Cooling Jacket Advanced for CSLaser, CTLaser as well as CTVideo and CSVideo (Part-No.: TM-CJAExxx-CTL), consisting of housing and chassis
The front part must be ordered separately.
incl. mounting accessories for - TIM NetBox or USB server Gigabit - Industrial PIF - Installation instructions
Operation without focusing or front attachment is not possible.
10
1.4 Mounting Accessories
1.4.1 Accessories for TIM NetBox
Figure 1: Accessories for TIM NetBox
1 Support rods for Industrial PIF (2x distance bolt SW 5,5x6 - M3x6,
2x cylinder head screw M3x10 and 2x cylinder head screw M3x5)
2 Shaft for fixing the TIM NetBox
3 Holding plate (2x cylinder head screw M3x5)
4 Fastening rail (4x cylinder head screw M4x8)
11 General notes
1.4.2 Accessories for USB Server Gigabit
Figure 2: Accessories for USB-Server Gigabit
1 DIN rail plate for fixing the USB server Gigabit
2 Distance rings
3 Screws
4 Support rods for Industrial PIF, with thread and without
12
2 Technical Data
2.1 General Specifications
Cooling Jacket Advanced Cooling Jacket Advanced Extended
Environmental rating IP 65 IP 65
Ambient temperature up to 315 °C1) up to 315 °C1)
Relative humidity 10 ... 95 %, non-condensing 10 ... 95 %, non-condensing
Material (housing) V2A V2A
Dimensions 271 mm x 166 mm x 182 mm 426 mm x 166 mm x 182 mm
Weight 5.7 kg 7.8 kg
Air purge collar G1/4“ Internal thread G3/8“ External thread
G1/4“ Internal thread G3/8“ External thread
Cooling water fittings G1/4“ Internal thread G3/8“ External thread
G1/4“ Internal thread G3/8“ External thread
Cooling water pressure 15 bar (217 psi) 15 bar (217 psi)
1) Cable available up to 250 °C ambient temperature as well as cable cooling up to 315 °C.
13 Technical Data
2.1.1 Focusing Unit and Front Part
Is needed for mounting a thermoIMAGER TIM or pyrometer.
TIM 160
Focusing unit
for 6°, 48° and 72° optics (Part-No.: TM-CJAFU6-TIM)
for 23° optics (Part-No.: TM-CJAFU23-TIM)
TIM QVGA / QVGA-HD
Focusing unit
for 18° optics (Part-No.: TM-CJAFU18-TIM)
for 29° optics (Part-No.: TM-CJAFU29-TIM)
for 53° optics (Part-No.: TM-CJAFU53-TIM)
14
TIM 640
Focusing unit
for 15° optics (Part-No.: TM-CJAFU15-TIM)
for 33° optics (Part-No.: TM-CJAFU33-TIM)
for 60° optics (Part-No.: TM-CJAFU60-TIM)
for 90° optics (Part-No.: TM-CJAFU90-TIM)
TIM M-1 / TIM M-05
Focusing unit
for 16 mm (Part-No.: TM-CJAFUO16-TIM)
for 25 mm (Part-No.: TM-CJAFUO25-TIM)
for 50 mm (Part-No.: TM-CJAFUO50-TIM)
for 75 mm (Part-No.: TM-CJAFUO75-TIM)
TIM 2xx
Front part
for TIM 2xx (Part-No.: TM-CJAFP2xx-TIM)
15 Technical Data
Front-part
for the 80° optics (TIM QVGA) the protection window (Part-No.: TM-CJAFP80-TIM)
must be ordered separately!
TIM QVGA
Front part
for CTLaser, CSLaser, CTVideo, CSVideo (Part-No.: TM-CJAFP-CTL)
CTLaser, CSLaser, CTVideo, CSVideo
16
2.2 Accessories
2.2.1 High Temperature Cable
High temperature Ethernet cable cat.6 (180 °C), 10 m, incl. 2x RJ45 connector (Part-No.: TM-CJAETC10H-TIM)
High temperature Ethernet cable cat.6 (180 °C), 20 m, incl. 2x RJ45 connector (Part-No.: TM-CJAETC20H-TIM)
High temperature Ethernet cable cat.6 (250 °C), 10 m, incl. 2x RJ45 connector (Part-No.: TM-CJAETC10H2-TIM)
High temperature Ethernet cable cat.6 (250 °C), 20 m, incl. 2x RJ45 connector (Part-No.: TM-CJAETC20H2-TIM)
High temperature USB cable (180 °C), 5 m
(Part-No.: TM-USB5PC5HCJA-TIM)
High temperature USB cable (180 °C), 10 m
(Part-No.: TM-USB5PC10HCJA-TIM)
High temperature USB cable (250 °C), 5 m
(Part-No.: TM-USB5PC5H2CJA -TIM)
High temperature USB cable (250 °C), 10 m
(Part-No.: TM-USB5PC10H2CJA -TIM)
Pyrometer cable (available separately)
Note: A pre-installed cable gland … CGx is mandatory for the usage of the CoolingJacket.
The cables listed here already contain this cable gland.
17 Technical Data
2.2.2 Protection Window
Adequate protection windows are available for all versions.
Part-No. Description Spectral range Transmission1)
TM-CJAPWL-TIM Protective window (50.8 x 3 mm/ Ge) for TIM QVGA-G7, TIM VGA-G7 7.9 µm 0.96
TM-CJAPWL-TIM Protective window (50.8 x 3 mm/ Ge) for TIM 160, TIM QVGA, TIM QVGA-HD, TIM 640
7.5 - 13 µm 0.92
TM-CJAPWL2xx-TIM Protective window (67 x 3 mm/ ZnS) for TIM 2xx (VIS + IR transmittive), TIM 160, TIM QVGA (80° optics), TIM QVGA-HD, TIM QVGA-G7, TIM 640, TIM VGA-G7 if used together with air purge laminar
7.5 - 13 µm 0.91
TM-CJAPWZNS-TIM Protective window (50.8 x 3 mm/ ZnS) for für TIM QVGA-G7, TIM VGA-G7
7.9 µm 0.93
TM-CJAPWZNS-TIM Protective window (50.8 x 3 mm/ ZnS) for TIM 160, TIM QVGA (without 80° optics), TIM QVGA-HD, TIM 640
7.5 - 13 µm 0.91
TM-CJAPWS-CTL Protective window (67 x 3 mm/ Borofloat 33) for 1M/ 2M/ 3M-models of CTlaser-, CSlaser-, CTvideo-, CSvideo series
1.0/ 1.6/ 2.3 µm
0.92
18
TM-CJAPWM05-TIM Protective window (67 x 3 mm/ Borofloat 33) for TIM M-05 450 - 650 nm 0.98
TM-CJAPWM1-TIM Protective window (67 x 3 mm/ Borofloat 33) for TIM M-1 0.85 – 1.1 µm
0.98
1) The shown values are standard values and may vary between different delivery batches.
19 Technical Data
2.2.3 Mounting Flange
Mounting flange (Part-No.: TM-CJAMF-TIM) for front mounting of the Cooling Jacket Advanced (Standard and
Extended) including mounting screws and washers
2.2.4 Laminar Air Purge
Laminar air purge for front mounting of the
CoolingJacket Advanced (Standard and Extended).
Two different versions are available: One for standard
IR camera applications [Part-No.: TM-CJAAPLS] and
the other for line scanning applications [Part-No.: TM-
CJAAPLL].
Those two versions are fitting to all focusing units with
production date ≥01/2018. A protective window (67 x 3
mm) has to be ordered separately. If you like to mount
the air purge on an older CJ, the focusing unit should
be exchanged to the current version.
Technical data
Protective rating IP65
Ambient temperature Up to 315 °C (with water cooling)
Material Aluminum
Dimensions (W x H x D) 200 x 189 x 43 mm
Weight 1,9 kg
Air purge connection NW 7,2
Cooling water connection G3/8″ external thread
Cooling water pressure Max. 8 bar
Volume flow 40 - 120 l/min
Air pressure 1,1 - 8 bar
Protective window Necessary1)
1) A protective window (67 x 3 mm) has to be
ordered separately.
20
Dimensions of laminar air purge
21 Technical Data
Dimensions of laminar air purge (line scanner version)
22
2.2.5 Additional Accessories
Industrial PIF without housing (Part-No.: TM-CJAPIF500V2-TIM), 500 VACRMS isolation voltage between TIM and
process, 25 cm connection cable.
Industrial PIF without housing
23 Technical Data
2.3 Dimensions
2.3.1 Cooling Jacket Advanced
Figure 3: Cooling Jacket Advanced - side view
24
Figure 4: Cooling Jacket Advanced - top view
25 Technical Data
Figure 5: Cooling Jacket Advanced - front view
26
Figure 6: Cooling Jacket Advanced, standard version – complete view
27 Technical Data
2.3.2 Cooling Jacket Advanced Extended
Figure 7: Cooling Jacket Advanced Extended - side view
28
Figure 8: Cooling Jacket Advanced, extended version - top view
29 Technical Data
Figure 9: Cooling Jacket Advanced, extended version - front view
30
Figure 10: Cooling Jacket Advanced, extended version – complete view
31 Technical Data
2.3.3 Mounting Flange (Accessories) TM-CJAMF-TIM
Figure 11: Mounting flange for Cooling Jacket Advanced (Standard and Extended)
32
2.4 Fittings
2.4.1 Cooling Water Fitting
2.4.2 Air Purge Collar
The lens must be kept clean at all times from dust, smoke, fumes and other contaminants in order to avoid reading
errors. These effects can be reduced by using an air purge collar.
Use oil-free, technically clean air only.
The needed amount of air (approx. 120 l/min., 5 – 8 bar) depends on the application and the
installation conditions on-site.
The air purge collar has a G1/4“-internal thread and a G3/8“-external thread.
The cooling water input and output has a G1/4“-internal thread and a G3/8“-external thread.
The maximum cooling water pressure is 15 bar (271 psi).
While connecting the hoses keep inclined the Cooling Jacket at an angle of approx. 45° to
avoid air bubbles.
33 Technical Data
The following classes according to ISO 8573-1 are recommended for the quality of compressed air:
Classes
Oil content: class 2 or better
Water content: class 4 or better
Solid content: class 2 or better
34
2.4.3 Cable Glands on the Back
Figure 12: Cable glands with seal insert
35 Technical Data
2.5 Cooling Properties
Flow Temperature
at
1 l/min
Temperature
at
2,5 l/min
Temperature
at
5 l/min
Ambient
temperature
[255 °C]
Cooling water input [°C] 19 19 19
Camera/ Thermometer
[°C] 36 34 32
Table 1: Cooling properties with a steady cooling water input temperature and various flow
Flow Temperature at 2.5 l/min of flow
Ambient temperature
[255 °C]
Cooling water input [°C] 25 30 35 40
Camera/ Thermometer [°C] 39 42 45 51
Table 2: Cooling properties with a steady flow and various cooling water input temperature
36
2.5.1 Condensation
Example (see Table 3):
Ambient temperature 80 °C
Relative humidity 25 %
Minimum device temperature 45 °C
At an ambient temperature of 80 °C and a relative humidity of 25 % the device temperature must not be below 45 °C.
Otherwise condensation occurs on the lens or the electronic.
For applications at ambient temperatures until 100 °C and a high humidity there is
danger of condensation (see Table 3).
To avoid condensation, the temperature of the cooling media and the flow rate must
ensure a minimum device temperature.
Consider the operation temperature of the applied devices.
37 Technical Data
Relative humidity [%]
Am
bie
nt
tem
pe
ratu
re [
°C]
10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
5 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 5
10 0 0 0 0 0 0 0 0 0 0 0 0 0 5 5 5 5 5 10
15 0 0 0 0 0 0 0 0 0 5 5 5 5 10 10 10 10 10 15
20 0 0 0 0 0 0 5 5 5 10 10 10 10 15 15 15 15 15 20
25 0 0 0 0 5 5 10 10 10 10 15 15 15 20 20 20 20 20 25
30 0 0 0 5 5 10 10 15 15 15 20 20 20 20 25 25 25 25 30
35 0 0 5 10 10 15 15 20 20 20 25 25 25 25 30 30 30 30 35
40 0 5 10 10 15 20 20 20 25 25 25 30 30 30 35 35 35 35 40
45 0 10 15 15 20 25 25 25 30 30 35 35 35 35 40 40 40 40 45
50 5 10 15 20 25 25 30 30 35 35 35 40 40 40 453 45 45 45 50
60 15 20 25 30 30 35 40 40 40 45 45 50 50 50 50 50 50 50 60
70 20 25 35 35 40 45 45 50 50 50 50 50 60 60 60 60 60 60 70
80 25 35 40 45 50 50 50 60 60 60 60 60 70 70 70 70 70 70 80
90 35 40 50 50 50 60 60 60 70 70 70 70 80 80 80 80 80 80
100 40 50 50 60 60 70 70 70 80 80 80 80 80
Table 3: Minimum device temperature in relation to ambient temperature and relative humidity in [°C]
38
Please respect the maximum ambient temperature of your measuring system!
39 Installation
3 Installation
3.1 Installation
The Cooling Jacket Advanced both versions can be installed in the desired position via the mounting bracket.
Figure 13: Cooling Jacket Advanced with mounting bracket (TM-JAB-TIM)
1 Mounting bracket, adjustable in two axes
40
3.2 Installation of TIM Camera, Infrared Video Thermometer and Infrared Thermometer
3.2.1 Assembling of the Focusing Unit
Figure 14: Focusing unit (external and internal part)
1 External part of the focusing unit 2 Internal part of the focusing unit
The focusing unit consists of two parts, external (1) and internal (2) part. The external part
focuses the camera. The internal part fixes the camera.
41 Installation
Figure 15: Assembling of the 6°, 9°, 23°, 26°, 29°, 33°, 39°, 41°, 53°, 72° optics
42
Figure 16: Assembling of the 13° and 15° optics
43 Installation
Note: For the installation of the
60 ° and 90 ° optics, the
camera must be fixed upside
down on a supplied bracket.
Figure 17: Assembling of the
60° and 90° optics
44
3.2.2 Assembling of the Front Part
Figure 18: Assembling of the TIM 2xx
45 Installation
Note: The subsequent focusing on the
focusing unit in the built-in cooling
housing should be done wisely. If you
use too much force, the optics can be
over-twisted and damaged.
Figure 19: Assembling of the TIM M-1 /
TIM M-05
46
Figure 20: Assembling of the CTLaser, CSLaser, CTVideo, CSVideo
47 Installation
3.2.3 Mounting of the Protective Window at Focusing Unit
Step 1: Loosen the three countersunk head screws M2 x 5
(inside) and remove them!
Figure 21: Focusing unit
Step 2: Turn six countersunk head screws M1.6 x 4 (outside) by
two turns to the left! (Do not unscrew or remove!)
Step 3: Now carefully pull the nozzle ring out of the focusing unit!
Step 4: Place the protective window in the provided platform of
the focusing unit and the O-ring 50 x 1.5 in the provided
nut of the nozzle ring!
Step 5: Plug the nozzle ring back into the focusing unit and secure it with the three countersunk head screws
M2 x 5 (inside)!
Step 6: Now tighten the six countersunk head screws M1.6 x 4 (outside) by twisting them two turns to the
right!
48
Figure 22: Individual components of the focusing unit
49 Installation
3.2.4 Assembly of the Laminar Air Purge
Figure 23: Assembly of the laminar air purge
50
Figure 24: Assembly of the laminar air purge
For the focusing unit, the standard air purge must be removed!
When mounting from the chassis to the CJA, the nut of the focusing unit must match the spring of the focus
ring (see picture Z).
51 Installation
Adjustment of air flow
Figure 25: Adjustment of air outflow
To change the air outflow direction on the air purge, turn the knurled head screws 1 and 2 simultaneously to
the left or right to move the drawer up or down.
52
Exchanging the protective window
Figure 26: Change of the protective window
Loosen the screws from the clamping ring and remove it! Now take out the O-ring and the protective
window!
53 Installation
3.2.5 Assembling to the Cooling Jacket Advanced Standard Version
Depending on the chassis either a camera of the TIM series, a laser infrared thermometer or an infrared video
thermometer can be installed.
Figure 27: Cooling Jacket Advanced
1 Housing
2 Front part (or focusing unit)
3 Chassis
54
Installation of TIM camera
1. Mount the camera to the focusing unit or the front part as described in Chapter 3.2.1 respectively Chapter
3.2.2. (exception for the camera TIM M-1/ M-05).
2. Seat the focusing unit or front part in the chassis (Figure 28). Position it as shown in Figure 31, by pushing
it to the bottom. The TIM M-1/ TIM M-05 camera is first placed into the chassis without the front part. After
that the front part is attached to the optic. Make sure that the front part is placed at a height of approx. 5 mm
above the optic in order to be able to slide it over the chassis front plate (Figure 31).
Figure 28: Inserting of the focusing unit
55 Installation
Figure 29: Long holes on the bottom of the chassis
Figure 30: Positioning the two bars in the slot for
different cameras
56
Figure 31: Focusing unit with camera
3. Fix the camera with the provided screw on the bottom of the chassis.
57 Installation
Figure 32: Mounting of the camera to the chassis
4. Then connect the TIM camera with the provided USB cable and lead it out of the cable gland.
5. Slide the chassis with the camera in the housing. Put the pins of the hinges to the slits of the chassis and
lock the hinge by pushing it forward (Figure 33).
58
Figure 33: Locking of the hinge
6. Move the locking lever to the left (symbol: ), so that the chassis is fitted close to the inner surface of the
housing (Figure 34).
59 Installation
The alternate contact of the cooling jaws with the camera/ infrared thermometer and
inner housing generates an optimal cooling effect.
1 Contact of the cooling jaws and inner housing
2 Contact of the cooling jaws and camera/ infrared thermometer
60
Figure 34: Back side of the Cooling Jacket with locking lever
1 Locking lever
61 Installation
7. Dismount in reverse order.
Figure 35: Unlocking of the hinge
62
Installation of video thermometer or infrared thermometer
1. Screw the infrared thermometer in the thread (M48x1.5) of the front part (see Figure 20) and seat the front
part in the chassis (Figure 36).
Figure 36: Front part with infrared thermometer
2. Lead the sensor cable out of the cable gland.
63 Installation
Figure 37: Infrared thermometer mounted
3. Slide the chassis with the camera in the housing. Put the pins of the hinges to the slits of the chassis and
lock the hinge by pushing in forward (Figure 33).
4. Move the locking lever to the left (symbol: ), so that the chassis is fitted close to the inner surface of the
housing (Figure 34)
5. Dismount in reverse order (Figure 35).
64
3.2.6 Assembling to the Cooling Jacket Advanced Extended
The extended version of the Cooling Jacket Advanced provides an installation of the TIM series together with the TIM
NetBox and an Industrial PIF or with the USB server Gigabit and an Industrial PIF.
Figure 38: Cooling Jacket Advanced, extended version
1 Housing
2 Focusing unit
3 Chassis
65 Installation
Installation of TIM camera
Steps 1-3, see page 54.
Figure 39: Camera implemented (extended version)
Assembling of TIM NetBox and Industrial PIF
4. Mount the holding plate (screws M3x5). Then attach the two support rods complete with the distance bolts (SW
5,5x6 - M3x6) with the provided screws (M3x10) to the bottom of the chassis. At last mount the shaft to fix the
TIM NetBox (Figure 40).
66
Figure 40 (a-c): Mounting of the accessories for TIM NetBox: a) Holding plate for TIM NetBox (top view),
b) Support rods for Industrial PIF (view from the bottom), c) Shaft to fix the TIM NetBox (view from the
bottom)
67 Installation
5. Fix the Industrial PIF with the screws (M3x5) as shown in Figure 41.
Figure 41: Mounting of the Industrial PIF (top view)
6. To mount the TIM NetBox to the chassis screw it to the fastening rail (screws M4x8).
68
Figure 42: TIM NetBox with fastening rail
At first push the fastening rail with the TIM NetBox into the left notch (1) of the shaft. Afterwards into the right
notch (1) (Figure 43) until it is engaged (Figure 44).
By locking the chassis the shaft pushes the TIM NetBox to the inner surface of the
housing. This guarantees an optimal cooling of the TIM NetBox.
69 Installation
Figure 43: Notch to fix the TIM NetBox (top view)
1 Notch
70
Figure 44: Fitting the TIM NetBox into the chassis (right hand view)
7. Then connect the TIM camera and the TIM NetBox with the provided USB cable and the Industrial PIF with
the camera. Combine the network connector and the TIM NetBox (Figure 45).
The USB angle plug is a reversible plug. This means that the plug can be plugged freely around in
the USB socket and works in both orientations.
71 Installation
Figure 45: Chassis with TIM camera, Industrial PIF and TIM NetBox
Follow as step 5, page 57.
72
Assembling of USB server Gigabit and Industrial PIF
4. Mount the DIN rail plate with the screws (M3x5) to fix the USB server Gigabit. Then attach the support rods
with the provided screws (M3x10) to one side of the chassis; use the distance ring on the left (Figure 46).
Figure 46 (a + b): Mounting of the accessories for USB server Gigabit: a) DIN rail plate for USB server
Gigabit (view from the bottom), b) Support rods for Industrial PIF (right hand view)
5. Fix the Industrial PIF with the screws (M3x5) as shown in Figure 47.
73 Installation
Figure 47: Mounting of the Industrial PIF (top view)
6. Engage the USB server Gigabit to the DIN rail plate (Figure 48).
74
Figure 48 (a + b): Fitting the USB server Gigabit into the chassis (top view and right hand view)
7. Then connect the TIM camera and the USB server Gigabit with the provided USB cable and the Industrial
PIF with the camera. Combine the network connector and the USB server Gigabit (Figure 49).
The USB angle plug is a reversible plug. This means that the plug can be plugged freely around in
the USB socket and works in both orientations.
75 Installation
Figure 49: Chassis with TIM camera, Industrial PIF and USB server Gigabit
Follow as step 5, page 57.
The USB Server can only be powered via PoE (Power of Ethernet) in the built-in CoolingJacket.
76
Figure 50: Mounting USB and network connector to USB server
Follow as step 5, page 57.
77 Example of installation
4 Example of Installation
Cooling Jacket Advanced
HT-Ethernet cable Cat. 6
Cooling Jacket Advanced
or
TIM camera
USB server
Gigabit
HT-USB cable
PIF cable
Industrial PIF
3x Analog Output
2x Analog Input
1x Digital Input
TIM camera
TIM NetBox
78
Cooling Jacket Advanced Extended
or
TIM camera
USB server
Gigabit
Cooling Jacket Advanced
HT cable
CTLaser
PIF cable
3x Analog Output
2x Analog Input
1x Digital Input
CT Box
Industrial-PIF
HT-Ethernet cable Cat. 6
TIM NetBox
79 Example of installation
Cooling Jacket Advanced
HT cable
CTVideo
Cooling Jacket Advanced
HT cable (analog [4-20 mA]/ digital)
CSLaser
sensor signals + laser
CT Box
HT cable video signal
80
*All high temperature cables are available for temperatures up to 180 °C/ 250 °C:
- IR video thermometer and IR thermometer: 3 m, 8 m, 15 m
- HT-Ethernet cable Cat.6: 10 m und 20 m
- HT-USB cable: 5 m, 10 m
HT cable current loop + alarm + laser
Cooling Jacket Advanced
CSVideo
HT-USB cable
Cooling Jacket Advanced Extended
CSVideo
USB server
Gigabit
HT-Ethernet cable Cat. 6
HT-cable analog
81 Example of installation
Figure 51: CoolingJacket (Standard) with CSvideo
Figure 52: Cooling Jacket Advanced (Extended) with
TIM Netbox and industrial PIF
Figure 53: Cooling Jacket Advanced (Extended-
Version) with USB Server and industrial PIF
MICRO-EPSILON MESSTECHNIK GmbH & Co. KG
Koenigbacher Str. 15 · 94496 Ortenburg / Germany
Tel. +49 (0) 8542 / 168-0 · Fax +49 (0) 8542 / 168-90
[email protected] · www.micro-epsilon.com
Your local contact: www.micro-epsilon.com/contact/worldwide/
X9751341-A032100HDR
*X9751341-A03*
MICRO-EPSILON MESSTECHNIK