point-based ionizing radiation dosimetry using radiochromic materials and … · 2013. 11. 8. ·...
TRANSCRIPT
-
POINT-BASED IONIZING RADIATION DOSIMETRY USING RADIOCHROMIC
MATERIALS AND A FIBREOPTIC READOUT SYSTEM
by
Alexandra Rink
A thesis submitted in conformity with the requirements for the degree of Doctor of Philosophy
Graduate Department of Medical Biophysics University of Toronto
© Copyright by Alexandra Rink (2008)
-
ii
Abstract
Point-Based Ionizing Radiation Dosimetry Using Radiochromic Materials And Fibreoptic
Readout System
Doctor of Philosophy, 2008
Alexandra Rink
Department of Medical Biophysics
University of Toronto
Real-time feedback of absorbed dose at a point within a patient can help with radiological
quality assurance and innovation. Two radiochromic materials from GafChromic MD-55 and
EBT films have been investigated for applicability in real-time in vivo dosimetry of ionizing
radiation. Both films were able to produce a real-time measurement of optical density from a
small volume, allowing positioning onto a tip of an optical fibre in the future. The increase in
optical density was linear with absorbed dose for MD-55, and non-linear for EBT. The non-
linearity of EBT is associated with its increased sensitivity to ionizing radiation compared to
MD-55, thus reaching optical saturation at a much lower dose. The radiochromic material in
EBT film was also shown to polymerize and stabilize faster, decreasing dose rate dependence in
real-time measurements in comparison to MD-55. The response of the two media was tested
over 75 kVp – 18 MV range of x-ray beams. The optical density measured for EBT was constant
within 3% throughout the entire range, while MD-55 exhibited a nearly 40% decrease at low
energies. Both materials were also shown to be temperature sensitive, with the change in optical
density generally decreasing when the temperature increased from ~22°C to ~37°C. This was
accompanied by a shift in the peak absorbance wavelength. It was illustrated that some of this
decrease can be corrected for by tracking the peak position and then multiplying the optical
density by a correction factor based on the predicted temperature. Overall, the radiochromic
-
material in GafChromic EBT film was found to be a better candidate for in vivo real-time
dosimetry than the material in GafChromic MD-55.
A novel mathematical model was proposed linking absorbance to physical parameters
and processes of the radiochromic materials. The absorbance at every wavelength in the
spectrum was represented as a sum of absorbances from multiple absorbers, where absorbance is
characterized by its absorption coefficient, initiation constant, and polymerization constant.
Preliminary fits of this model to experimental data assuming two absorbers suggested that there
is a trade-off between EBT’s greater sensitivity and its dose linearity characteristics. This was
confirmed by experimental results.
iii
-
iv
Dedicated to my dear parents.
-
Acknowledgments The assistance and wisdom of many people went into this work. I would gratefully like to
acknowledge the contributions, in whatever form they were, of the following:
• my co-supervisors, Dr. David Jaffray and Dr. Alex Vitkin, for all the paper edits, research
guidance and advice
• committee members, Dr. Christine Allen and Dr. Mike Rauth for all the support
• Yuen Wong, Brian Taylor, Jason Ellis, and Matt Filletti for machining all the phantoms
and doing the various small “rush” jobs
• Robert Rothwell and Robert Rusnov for all the assistance with electrical and optical work
• Dr. Robert Heaton, Hamideh Alasti, Duncan Galbraith, Dr. Mohammad Islam, and Dr.
Jean-Pierre Bissonnette for their experience
• Bern Norrlinger for his experience and help with any and every accelerator that ever
broke
• Tony Manfredi for all the assistance with the Elekta accelerators
• Dr. Robert Weersink and David Giewercer for assistance and wisdom with fibre optics
• Joanne Kniaz of Advanced Optical Microscopy Facility for the microscopy work
• Dr. David Lewis and Dr. Sangya Varma of International Specialty Products for their
contributions to this work, experience and guidance
• Dr. Douglas Moseley for all the help with Matlab, the carpool rides, and outrageous
conversations on the GO train
• Steve Ansell and Graham Wilson for all the computer support, psychotherapy lunches
and Chinese noodles
• Jinzi Zheng and Jeremy Hoisak for all the coffee breaks which kept me sane
• my parents, Gala and Youri Rink, for never looking back or regretting any choices in life
v
-
Table of Contents CHAPTER 1: INTRODUCTION.................................................................................................. 1
I. Ionizing Radiation in Cancer Treatment ............................................................................... 2
II. Radiation Dose ...................................................................................................................... 3
III. Radiation Dosimetry ............................................................................................................ 5
A. Basic Interactions.............................................................................................................. 5
B. Standards and Protocols for Dosimetry............................................................................ 6
C. Estimation of Dose Delivered in Therapy......................................................................... 7
IV. The Challenges of Dose Measurement in the Clinical Setting ............................................ 7
A. Clinical Applications and Ideal Dosimeter ....................................................................... 7
B. Current in vivo Dosimeters.............................................................................................. 10
C. Optical Methods .............................................................................................................. 11
V. Outline of Thesis................................................................................................................. 13
CHAPTER 2: REAL-TIME RESPONSE OF GAFCHROMIC® MD-55 FILM TO IONIZING
RADIATION ................................................................................................................................ 21
I. Introduction ......................................................................................................................... 22
Review of GafChromic® MD-55 .......................................................................................... 22
General Experience............................................................................................................... 22
Solid-state Polymerization of Diacetylenes .......................................................................... 25
II. Methods and Materials ........................................................................................................ 29
A. ΔOD of GafChromic® MD-55 at Various Doses............................................................ 38
B. Sensitivity as a Function of Layer Thickness.................................................................. 39
C. ΔOD of GafChromic® MD-55 at Various Dose Rates.................................................... 39
D. ΔOD Dependency on Temperature ................................................................................. 40
E. Continuous Versus Pulsed Irradiation ............................................................................. 40
III. Results................................................................................................................................ 41
A. ΔOD of GafChromic® MD-55 at Various Doses............................................................ 41
B. Sensitivity as a Function of Layer Thickness.................................................................. 44
C. ΔOD of GafChromic® MD-55 at Various Dose Rates.................................................... 45
D. ΔOD Dependency on Temperature ................................................................................. 47
E. Continuous Versus Pulsed Irradiation ............................................................................. 49
IV. Discussion.......................................................................................................................... 49
vi
-
A. OD of GafChromic® MD-55 at Various Doses .............................................................. 50
B. Sensitivity as a Function of Layer Thickness.................................................................. 51
C. ΔOD of GafChromic® MD-55 at Various Dose Rates.................................................... 52
D. ΔOD Dependency on Temperature ................................................................................. 53
E. Applications..................................................................................................................... 54
V. Conclusion .......................................................................................................................... 55
ACKNOWLEDGEMENTS.................................................................................................. 56
CHAPTER 3: REAL-TIME RESPONSE OF GAFCHROMIC® EBT ....................................... 61
I. Introduction ......................................................................................................................... 62
II. Method and Materials.......................................................................................................... 62
A. ΔOD of EBT Film Versus Time ..................................................................................... 66
B. Sensitivity and Stability Comparison Between EBT and MD-55 Films......................... 66
C. Dependence of Real-Time OD Measurements on Dose Rate for the EBT Film ............ 67
D. Structure of Active Crystals in MD-55 and EBT Films.................................................. 67
III. Results and Discussion ...................................................................................................... 67
A. OD of EBT Film Versus Time........................................................................................ 67
B. Sensitivity and Stability Comparison Between EBT and MD-55 Films......................... 70
C. Dependence of Real-Time ΔOD Measurements on Dose Rate for the EBT Film.......... 73
D. Structure of Active Crystals in MD-55 and EBT Films.................................................. 75
IV. Conclusion ......................................................................................................................... 77
ACKNOWLEDGEMENTS.................................................................................................. 78
CHAPTER 4: EFFECTS OF VARYING DOSE RATE ON REAL-TIME MEASUREMENTS
OF OPTICAL DENSITY OF GAFCHROMIC® EBT ................................................................. 81
I. Introduction ......................................................................................................................... 82
II. Methods and Materials ........................................................................................................ 82
III. Results and Discussion ...................................................................................................... 85
IV. Conclusion ......................................................................................................................... 88
ACKNOWLEDGEMENTS.................................................................................................. 89
CHAPTER 5: CHARACTERIZATION OF GAFCHROMIC® EBT: TEMPERATURE AND
HUMIDITY EFFECTS................................................................................................................. 91
I. Introduction ......................................................................................................................... 92
Chemical Background and General Experience ................................................................... 92
II. Methods and Materials ........................................................................................................ 94 vii
-
A. Temperature Dependence ................................................................................................ 96
B. Absorbance and Sensitivity Dependence on Water Content............................................ 97
III. Results and Discussion ...................................................................................................... 99
A. Temperature Dependence ................................................................................................ 99
B. Absorbance and Sensitivity Dependence on Water Content.......................................... 104
IV. Conclusion ....................................................................................................................... 109
ACKNOWLEDGEMENTS................................................................................................ 109
CHAPTER 6: ENERGY DEPENDENCE OF GAFCHROMIC® ............................................. 112
MD-55 AND EBT....................................................................................................................... 112
I. Introduction ....................................................................................................................... 113
II. Methods and Materials ...................................................................................................... 114
Solid Water™ Phantom ...................................................................................................... 114
Ionizing Radiation Exposures ............................................................................................. 115
Optical Measurements ........................................................................................................ 117
III. Results and Discussion .................................................................................................... 118
IV. Conclusion ....................................................................................................................... 123
ACKNOWLEDGEMENTS................................................................................................ 124
CHAPTER 7: MATHEMATICAL MODEL OF RADIOCHROMIC MEDIUM RESPONSE TO
IONIZING RADIATION ........................................................................................................... 127
I. Introduction ....................................................................................................................... 128
II. Methods and Materials ...................................................................................................... 129
III. Results and Discussion .................................................................................................... 132
V. Conclusion ........................................................................................................................ 138
ACKNOWLEDGEMENTS................................................................................................ 138
CHAPTER 8: SUMMARY AND FUTURE DIRECTIONS..................................................... 142
I. Summary............................................................................................................................. 143
II. Future Directions............................................................................................................... 146
A. Optical Probe................................................................................................................. 146
B. Organization of Monomers and Polymers .................................................................... 148
C. Importance of Chemical Composition and Structure.................................................... 148
D. New Radiochromic Materials ..................................................................................... 149
E. Polymerization Kinetics as a Function of Dose Per Pulse.......................................... 149
F. ............................................................................ 150 Model-Fitting Algorithm and Code
viii
-
List of Tables Table 1. List of criteria for in vivo point-based real-time dosimeter. ............................................ 9
Table 2. Evaluation criteria for in vivo point-based real-time dosimeter. ................................... 23
Table 3. Comparison of inferred dose and percent error using calibration plot and pre-exposure
calibration as methods of calculation............................................................................................ 44
Table 4. Coefficients of equations of best fit characterizing ΔOD/DGy as a function of dose rate.
....................................................................................................................................................... 87
Table 5. Average percent standard deviation for each dose, uncertainty, and the difference
between the two for each dose delivered. ..................................................................................... 88
Table 6. Correction factors for the temperature correction scheme, calculated for doses of 50 –
400 cGy shown for a selection of predicted temperature values.. .............................................. 103
Table 7. The x-ray and photon beams employed in the investigations...................................... 116
Table 8. Comparison of response of EBT film, normalized to response at 6 MV, as measured
approximately 24 hours after exposure to that measured immediately at the end of exposure .. 122
Table 9. Model parameters using two absorbers for MD-55 and EBT and a fit with 1 second
pulse averaging. .......................................................................................................................... 136
List of Figures Figure 1. Schematic of a typical relationship between tumour control probability (TCP) and
normal tissue complication probability (NTCP) versus dose. ........................................................ 4
Figure 2. Structures of: (a) diacetylene monomers, upon exposure to ionizing radiation,
polymerizes into (b) butatriene structure polymer; as the polymer chain grows, it rearranges via
(c) an intermediate between butatriene structure and acetylene structure, into (d) acetylene
structure polymer. ......................................................................................................................... 26
Figure 3. A model of optical density of GafChromic® MD-55 versus time before, during and
after exposure................................................................................................................................ 28
Figure 4. Schematic of experimental setup.................................................................................. 30
Figure 5. Emission spectrum of the LED as detected by the spectrophotometer. ....................... 31
Figure 6(a-c). Solid Water™ phantom (a) assembled, (b) disassembled, (c) schematic. ............ 32
Figure 7. Schematic of cross-section of film holder in Solid Water™ phantom.. ....................... 33
ix
-
Figure 8. Change in absorbance of GafChromic® MD-55 film at various wavelengths plotted
before exposure, immediately after end of exposure, and 15 and 60 minutes after the end of
exposure. ....................................................................................................................................... 34
Figure 9. Change in optical density and rate of change in optical density for GafChromic® MD-
55 film as a function of time before, during and after exposure to 381 cGy with 6 MV X-rays.. 35
Figure 10. Termination of exposure is taken as the intercept of the two fitted lines: first line
corresponding to data obtained during exposure and second line corresponding to data obtained
after end of exposure..................................................................................................................... 36
Figure 11. Change in optical density can be calculated for any exposure by subtracting initial
OD from final OD......................................................................................................................... 37
Figure 12. Schematic of setup for temperature dependency experiments. .................................. 40
Figure 13. Change in optical density for GafChromic® MD-55 exposed to 381 cGy with 6 MV
X-rays as a function of time.......................................................................................................... 41
Figure 14. Change in OD for five pieces of film, each exposed to 381 cGy with 6 MV X-rays (at
the doserate of 285 cGy/min)........................................................................................................ 42
Figure 15. Inferred dose using ΔOD measurements and calibration plot as a function of applied
dose. .............................................................................................................................................. 42
Figure 16. Change in optical density for a piece of GafChromic® MD-55 film during several
exposures applied approximately 5 minutes apart. ....................................................................... 43
Figure 17. Optical density as a function of dose for a system utilizing one, two and four pieces
of stacked film............................................................................................................................... 45
Figure 18. Change in optical density as a function of dose for doses delivered at 95 cGy/min,
286 cGy/min and 671 cGy/min..................................................................................................... 46
Figure 19. Rate of change in optical density as given by the linear fit of data obtained during
exposure as a function of applied dose, for doses delivered at 95 cGy/min, 286 cGy/min and 571
cGy/min......................................................................................................................................... 47
Figure 20. Position of wavelength of maximum absorbance for GafChromic® MD-55 as a
function of irradiation/measurement temperature......................................................................... 48
Figure 21. Change in OD for a given dose as a function of applied/measured temperature using
both a constant spectral averaging window and a shifting spectral averaging window. .............. 48
Figure 22. Schematic of experimental setup................................................................................ 63
x
-
Figure 23. Emission of the light emitting diode used in experimental setup, as measured by
spectrometer.................................................................................................................................. 63
Figure 24. Schematic of layers in EBT film. ............................................................................... 64
Figure 25. Change in absorbance of EBT film over a range of wavelengths before, immediately
after, and at two time points post-exposure. ................................................................................. 65
Figure 26. Optical density versus time for a single piece of EBT film; optical density versus time
for five pieces of EBT film shown on a reduced time scale (inset). ............................................. 68
Figure 27. Wavelength of maximum absorbance for EBT film versus time during and after
exposure to 9.52 Gy at 2.86 Gy/min with 6 MV X-rays............................................................... 69
Figure 28. Optical denisty of EBT film versus time for various spectral averaging windows.... 70
Figure 29. Optical density for EBT and MD-55 films during and after exposure....................... 71
Figure 30. Percent increase in OD for EBT and MD-55 films after exposure, calculated with
respect to OD at the end of exposure. ........................................................................................... 72
Figure 31. Percent increase in OD for EBT and MD-55 films within one hour after exposure. . 73
Figure 32. Optical density for EBT film exposed to 9.52 Gy, delivered with 6 MV at 0.95
Gy/min and 5.71 Gy/min. ............................................................................................................. 74
Figure 33. Microscope images of monomer crystals within the sensitive media of MD-55 and
EBT films...................................................................................................................................... 76
Figure 34. Optical density versus time for a 50 cGy irradiation at 16 cGy/min.......................... 84
Figure 35. The average sensitivity as a function of dose rate, for various doses......................... 85
Figure 36. Chemical formula of pentacoasa-10,12-dyinoic acid (PCDA), and lithium salt of
PCDA (LiPCDA).. ........................................................................................................................ 93
Figure 37. Change in absorbance spectra for EBT film exposed to 1 Gy with 6 MV and 75 kVp
beams.. .......................................................................................................................................... 95
Figure 38. Schematic of the modified phantom, with plastic water hoses on either side of the
film and optical fibers. .................................................................................................................. 96
Figure 39. Wavelength of maximum change in absorbance of commercial EBT films irradiated
to 1 Gy as a function of measured temperature. ......................................................................... 100
Figure 40. Values of wavelength of maximum absorbance for various doses delivered to
commercial EBT films as a function of measured temperature.................................................. 100
Figure 41. Change in optical density for 1 Gy dose calculated for optical range of 630-640 nm,
and an optical range of 10 nm centered about wavelength of maximum absrobance, versus
measured temperature.. ............................................................................................................... 101
xi
-
Figure 42. Temperature calculated using the position of wavelength of maximum absorbance
versus measured temperature, shown with a line of best fit.. ..................................................... 102
Figure 43. Change in optical density for films irradiated to 1 Gy using a fixed optical integration
range of 630 to 640 nm, moving optical range of 10 nm about the peak of maximum absorbance,
and as calculated using the peak of maximum absorbance and temperature-dependent correction
factor.. ......................................................................................................................................... 102
Figure 44. Percent decrease in net OD for a 3 Gy dose, following different times in a desiccator
at 50 ºC........................................................................................................................................ 105
Figure 45. Spectral comparisons of absorbance of desiccated and normal unlaminated EBT film.
..................................................................................................................................................... 106
Figure 46. Absorbance of unlaminated EBT film after time in desiccator at 50 ºC. ................. 107
Figure 47. Spectral comparisons of absorbance of desiccated, rehydrated, and normal
unlaminated EBT film irradiated to 3 Gy. .................................................................................. 107
Figure 48. Absorbance spectra of exposed unlaminated films using “plate-like” form of
polymer, and the rehydrated form of “hair-like” polymer. ......................................................... 108
Figure 49. 30 cm × 30 cm × 4 cm phantom with the film insert ............................................... 114
Figure 50. Sample of time-dependent OD with time for a 1 Gy irradiation with a 75 kVp
Therapax DXT 300 beam at 8 cGy/min for MD-55, HS, and EBT film. ................................... 119
Figure 51. Un-normalized change in OD for 1 Gy total dose for MD-55, HS and EBT films for
irradiations delivered at various equivalent x-ray energies ........................................................ 119
Figure 52. Change in OD per Gy for MD-55, HS and EBT, as a function of equivalent x-ray
energy.......................................................................................................................................... 120
Figure 53. Increased sensitivity of HS and EBT films with respect to MD-55, for a dose of 1 Gy
..................................................................................................................................................... 123
Figure 54. Change in absorbance of a single absorber versus time for different A parameters,
keeping k and p parameters constant. ......................................................................................... 132
Figure 55. Change in absorbance of a single absorber versus time for different k parameters,
keeping A and p parameters constant.......................................................................................... 133
Figure 56. Change in absorbance of a single absorber versus time for different p parameters,
keeping A and k parameters constant. ......................................................................................... 134
Figure 57. Experimental absorbance of MD-55 film at the main absorbance peak, and the model
fit.. ............................................................................................................................................... 135
xii
-
Figure 58. Experimental absorbance of EBT film at the main absorbance peak, and the model
fit.. ............................................................................................................................................... 135
Figure 59. Schematics of single and dual fibre optical dosimeter prototypes. .......................... 147
xiii
-
xiv
List of Abbreviations and Symbols ΔA change in absorbance
ΔOD change in optical density
ΔODv change in visual density
ΔV small volume
ε(λ) extinction coefficient at wavelength λ
λ wavelength
λmax wavelength of maximum absorbance
ρ density 60Co Cobalt-60
A absorbance
A(λ) absorbance at wavelength λ
AAPM American Association of Physicists in Medicine
bi polymer initiation constant per dose
c concentration
cGy centiGray
D dose
DGy dose in Gy
EBT radiochromic film intended for External Beam Therapy
E energy
eV electron-volt
FWHM full width half maximum
Gy gray (J/kg)
HS radiochromic film of High Sensitivity
HVL half-value layer
I intensity
ID background intensity
IR reference intensity
I0s initial intensity
Is sample intensity
ICRU International Commission of Radiation Units
IMRT intensity modulated radiation therapy
-
IGRT image guided radiation therapy
ISP International Specialty Products
ki polymer initiation constant
keV kilo electron-volt
kVp peak kilovoltage
l length
LED light emitting diode
LINAC linear accelerator
LiPCDA Lithium pentacosa-10,12-diynoate (lithium salt of PCDA)
MD-55 radiochromic film intended for Medium Dose, size 5×5
MeV mega electron-volt
MOSFET metal oxide semiconductor field-effect transistor
MSDS Material Safety Data Sheet
MV megavolt
N0 initial number of monomer chains
Nm remaining number of monomer chains
Nip number of initiated polymer chains
Nfp number of fully-formed polymer chains
NTCP normal tissue complications probability
OD optical density
ODv visual density (weighted by known response of human eye)
OSL optically stimulated luminescence
pi polymerization kinetics constant
PCDA pentacosa-12,12-diynoic acid
PDD percent depth dose
PMMA poly-methyl methacralate
QTH quartz-tungsten-halogen
SAD source-to-axis distance
SSD source-to-surface distance
t time
TCP tumour control probability
TG Task Group
TLD thermoluminescent dosimeter
xv
-
TPR tissue-phantom ratio
UV ultraviolet
Z atomic number
xvi
-
1
CHAPTER 1: INTRODUCTION
-
2The work within this thesis describes a novel method for performing real-time dosimetry using
fibre-optic read-out of radiochromic materials. The radiochromic materials are investigated for
their applicability in clinical dosimetry measurements in vivo and in vitro. Their performance as
a function of dose and time is modeled, with parameters linked to physical properties of the
materials and the processes that occur during exposure to radiation. A system is thus established
for evaluating radiochromic dosimeters for clinical dosimetry, whereby their performance can be
at least in part be predicted by their physical properties.
In this chapter, clinical rationale for the proposed real-time dosimeter is established by
outlining the need for in vivo dosimetry and the inability of the dosimeters presently available on
the market to meet that need.
I. Ionizing Radiation in Cancer Treatment
Ionizing radiation is encountered under many circumstances in medicine, and specifically in
oncology. It is used to identify and locate the cancer, to target it, and to treat it, with
approximately 50% of cancer patients receiving radiation therapy for management of their
disease.* High energy photons (referred to as x-rays and gamma rays) are known to damage
tissue. Although the exact details of tissue damage are still not fully understood, it is believed
high energy photons induce ionization of important molecules within the cells, such as
deoxyribonucleic acid. Ionization refers to removal of an electron from a molecule, making it
unstable. These unstable molecules may react in a way that would prevent them from
functioning properly, eventually leading to cell death.
The source of radiation can be external or internal (known as brachytherapy), varying
greatly in energy and intensity. Energy can vary from 21 – 660 keV(1) gamma rays from
decaying radionuclides in brachytherapy seeds, to 18 MV x-ray or 20 MeV electron beams from
a linear accelerator (LINAC). The dose rate can be as low as 0.4-2 Gy/h(1) (where Gy=1J/kg) for
low dose rate brachytherapy, to as high as 6 Gy/min for LINAC treatments. The majority of the
external treatments are divided into dose fractions delivered daily, five days a week, over several
weeks, with only a small percentage of treatments delivered in a single large dose (known as
stereotactic radiosurgery) from a 60Co unit called GammaKnife or from a LINAC using a 6 MV
x-ray beam. In recent years, treatments have become more conformal to the tumour due to
implementation of Intensity Modulated Radiation Therapy (IMRT) and Image-Guided Radiation
Therapy (IGRT). With the development of these new technologies, a trend has been evolving * http://www.cancer.gov/cancertopics/factsheet/Therapy/radiation
-
3towards higher target absorbed dose values, fewer fractions, smaller treatment volumes, and
steeper dose gradients. These developments need to be validated in terms of actual dose
delivered.
II. Radiation Dose
Cell damage from ionizing radiation may result in several different outcomes, including repair of
damage by the cell, cell death, and survival with mutation.(2) Depending on the type of damage
and the tissue irradiated, the biological effect can take anywhere between a few hours (acute) to
many years (late) to manifest. During radiation therapy, the dose and its distribution are
important for the outcome of the treatment and prevention of further complications. A high
enough dose has to be delivered to the tumour and affected organs to obtain high probability of
tumour control, and a minimal dose should be delivered to healthy surrounding organs to limit
probability of acute or late effects.(2) To maintain high probability of tumour control, the
International Commission of Radiation Units (ICRU) recommends uniformity of tumour dose
within +7/-5% of the total prescribed dose.(3) The upper limit exists because the dose prescribed
is often limited by the dose delivered to the surrounding healthy organs during irradiation, which
is dependent on the type of treatment delivery. Generally, conformal treatment allows for higher
dose to be delivered to the tumour and for lower dose be delivered to the surrounding tissues,
though the total volume of tissue irradiated may increase. On the other hand, the probability of
cancer recurrence due to geometric miss may increase.
The probability of biological effect taking place (whether it be tumour cell kill or normal
tissue complications) versus dose is called a dose response curve.(2) The curves for tumour
control probability (TCP) and normal tissue complication probability (NTCP) are often plotted as
sigmoid relationships (Figure 1). That is, there is nearly no effect at first, then the probability
rises sharply, and levels off to a plateau. Thus delivering a smaller dose to the tumour than that
required for cure or control would sharply increase the probability of relapse. On the other hand,
if the patient receives a dose to the tumour that is much higher than that prescribed, then it is also
likely that the patient receives a higher dose to the surrounding normal tissue. Given the
sigmoidal relationship between dose and effect, the risk of acute or late effects increases
dramatically.
Because of the conformality of modern treatments, describing the dose distribution by a
single dose to the tumour and using the +7/-5% recommendation for guidance is an
oversimplification. As the treatments become tailored to each patient’s needs, the dose
-
41.0
0.5
Prob
abili
ty
Dose (Gy)
TCP NTCP
0.9
+5%-5%
30 40 50
Figure 1. Schematic of a typical relationship between tumour control probability (TCP) and
normal tissue complication probability (NTCP) versus dose. The NTCP curve is based on two
values (marked by X): doses at which 5% and 50% of patients develop complications when 2/3
of their liver is irradiated.(4) Dashed line represents a dose of 42.5 Gy, yielding 90% probability
of tumour control, and 37.5% probability of normal tissue complications (liver failure).
Increasing the tumour dose by 5% (dotted line) of the prescribed dose increases TCP by only
2.5%, but increases NTCP to 50%. On the other hand, delivering the same distribution with 5%
lower dose (dotted line), decreases NTCP to 25%, but also decreases TCP to 82.5%, which may
compromise treatment outcome.
distributions become more diverse. Systematic errors larger than 5% in dose delivery (measured
as entrance, exit dose, or combination of the two during treatment) to a small percent of patients
(~ 1%) have been published.(5-7) These can be due to inadequacies in dose calculation
algorithms,(8) setup errors, or a human error on behalf of the many individuals involved in the
process of patient treatment. Although doses to the patient can be calculated or inferred from a
-
5relative measurement, the American Association of Physicists in Medicine Task Group 40
recommended that clinics have access to an in vivo dosimetry system.(9) While radiation
transport algorithms are constantly being improved in order to perform dose calculations, and
phantoms become more complex to better represent human anatomy, conditions where dosimetry
and simulation of humans remain a challenge still exist. Accurate assessment of the dose
distribution in brachytherapy treatment and in regions near in homogeneities for external beam
treatments is vital to rapid innovation and development of new radiation therapy technologies
and techniques. Such an assessment can be done by performing in vivo dose measurements.
However, performing such measurements under clinical conditions is challenging.
III. Radiation Dosimetry
A. Basic Interactions
The interaction of ionizing radiation with matter results in a dose deposited within that
matter, where dose is defined as the absorbed energy per mass (J/kg, or Gy).(10) In radiotherapy,
the dose quoted is often dose to water, as most tissues within the body have similar radiological
properties as water (common exceptions are lung tissue, bone and teeth). The ionizing radiation
can be directly ionizing (charged particles) or indirectly ionizing (photons).(10)
The photons, as they pass through the medium, are attenuated by the medium and scatter
from their original path. The processes of attenuation are due to coherent scattering,
photoelectric effect, Compton effect, pair/triplet production, or photonuclear interactions. In
radiation therapy, the middle three are the interactions important for dose deposition, and the
probability of any of these events happening when a photon beam passes through the medium
depends on the energy of the photon, density, and the atomic number (Z) of the medium.
Photoelectric effect is the predominant interaction for low-energy photons (below 100 keV). The
probability of photon interacting in this manner in a medium of a given density increases with Z3
of the medium. Here all of the photon’s energy is transferred to one of the inner shell electrons
of an atom. This electron then continues through the medium with a kinetic energy equal to the
energy of the initial photon less the binding energy of the electron. In a Compton interaction, the
photon interacts with a valence shell electron, transferring part of photon’s energy to an electron
and the two then continue at a given angle from each other from the point of interaction, such
that the momentum is conserved. In this interaction, the photon does not transfer all of its energy
to the electron. The probability of this interaction occurring in a medium of given density is
-
6approximately independent of Z of the medium, and is predominant for photon energies around 1
mega-electron-volt (MeV). For high energy photons (over 1.022 MeV) pair production can
occur. In these interactions, as the photon interacts with the Coulomb field near the nucleus, the
photon is absorbed giving rise to an electron and a positron. If the photon interacts with the field
of the atomic electron, the atomic electron also acquires energy and escapes from the atom, thus
yielding triplet production. The process of triplet production requires the energy of the photon to
be greater than 2.044 MeV. The probability of pair production for a medium of given density is
roughly proportional to Z, while triplet production is independent of Z of the medium.(10)
The electrons that result from these interactions of photons with the medium in turn travel
through the medium themselves. The electron interacts with almost every atom in its path whose
electric field is detectable. Most of these interactions transfer small fractions of the electron’s
energy, and thus the electron is often thought of as losing the energy gradually in a frictionlike
process. Some of these energy losses result in a photon (Bremsstrahlung) being emitted when
the electron changes direction due to an electric field from a nearby nucleus. The energy lost in
this way does not contribute to the locally deposited dose. Electron interactions that do
contribute to local dose will either excite the shell electrons of a nearby atom to a higher energy
level, or ionize it. The rate of energy loss per distance decreases with increasing Z of the
medium and increasing kinetic energy of the electron.(10)
B. Standards and Protocols for Dosimetry
Ionizing radiation dose can be measured well in standard conditions following set
protocols(11-14). In most cases, using dosimetry equipment calibrated at a national standard
laboratory (such as National Research Council of Canada) or traceable to such a calibration, and
performing measurements under controlled conditions allows for accurate dosimetry in air and in
water or plastic phantoms, with uncertainties below 1%.(12,15) However, in some cases, such as
intravascular brachytherapy, absorbed dose standards may vary by as much as 10% between
measurements and different Monte Carlo calculations.(16) When using the dosimetric gold
standard, the ion chamber, some of the controlled conditions include a known temperature and
pressure, type and energy of the beam, distance from the source and depth (if phantom is
used).(12) Unfortunately, these parameters may not always be known to a high level of precision
and accuracy in clinical conditions.
-
7C. Estimation of Dose Delivered in Therapy
Doses under uncontrolled conditions, delivered during imaging and therapy, can be
obtained in several different ways. They can be computed using various algorithms as is done in
actual treatment planning,(17) with the algorithms constantly improving to include dose
calculations for brachytherapy and imaging procedures, as well as external beam irradiations.(18-
21) The dose can also be inferred from a relative dose measurement performed during the
procedure (such as skin dose measurement), or measured directly at a point of interest. Because
the points of interest on the patient may be inside the patient, such as at the tumour site,
performing a direct measurement is often trickier than the other two approaches.
IV. The Challenges of Dose Measurement in the Clinical Setting
While real-time dosimetry may be useful for in vitro measurements at several points of time
varying radiation fields, such as those in high dose rate brachytherapy or IMRT, the discussion
of clinical applications here is kept to in vivo measurements. Predicting the three-dimensional
cumulative dose distribution within a patient over the course of their treatment can be complex,
given the variations arising from minor patient positioning errors, treatment beam fluctuations,
motion during treatment, changes in anatomy as the patient loses weight or the tumour shrinks,
and possibly other sources of error. Thus measurements of dose at or within a patient are often
desired as part of a quality assurance, for investigative purposes of a new procedure, or for
implementation of new technology protocol. Performing dose measurements inside patients is
more complicated than skin dose measurements, and is not nearly as straightforward as phantom
dosimetry which is often utilized in the clinic.
A. Clinical Applications and Ideal Dosimeter
A dosimeter that can accurately measure ionizing radiation dose in vivo under various
clinical conditions may simplify and improve the current state of dosimetry. It can be used for
quality assurance of both external beam and brachytherapy treatments. If an error in positioning,
machine output, transcription or some other error occurs that results in a discrepancy of planned
dose above a set threshold, treatment can be interrupted, and the discrepancy investigated. The
dosimeter may also be used to track dose in an organ that moves in and out of the radiation field
during respiration. In a brachytherapy procedure, the real-time dose rate measurement may be
used as feedback to verify proper positioning of seeds. If the dose rate is higher or lower than
-
8anticipated, the insertion of the next set of seeds may be adjusted to get the proper dose rate and
dose at the point of measurement.
For a dosimeter to be an appropriate option for in vivo measurements in most clinical
scenarios, the overall construction and the dosimetric medium must satisfy several requirements,
listed in Table 1. The requirements can be used as a set of guidelines for evaluation of a new
dosimetric medium. The presence of the dosimeter must not perturb the tissue or the dose
distribution. It must also be sufficiently small (1) to be able to resolve a sharp dose gradient in a
radiation field. However, the size of the sensitive medium must be sufficiently large to yield
good signal statistics for high precision measurements. In part, the signal statistics can be
controlled if the dosimeter uses passive read-out method, as is discussed later in this Chapter.
The dosimeter should have near water-equivalent composition (2), such that its response
with change in energy is similar as that of water.(22) This allows for a single dosimeter to be used
across all energies encountered both in external beam and brachytherapy treatments without
performing a separate calibration. The dosimeter for quality assurance purposes should respond
in real time (3), providing dose or dose rate estimate within a few seconds from beginning of
irradiation so that treatment can be interrupted if the dose rate is outside of the value expected.
The dosimeter also has to show a dose response that is independent of the dose rate used to
deliver that dose (6). The dose estimate should have high dose resolution (5) in order to detect
small variations (~1 cGy) in the dose delivered (daily fractions are often ~200 cGy, and total
doses are 24-70 Gy, depending on the treatment and fractionation pattern), and the response
should be ideally linear with dose (4) for simple computation. Some of the other issues to be
taken into consideration are environmental conditions (7), such as the temperature dependence of
the dosimeter. The dose estimate should be independent of the temperature of the dosimeter,
which can vary from near room temperature on skin surface to as high as ~38 °C within the
patient. Finally, the dosimeter must be non-toxic (8) and biocompatible, and would preferably
be inexpensive enough to be disposable after each patient.
A dosimeter, as described above, which can be used across a wide range of energies
could be implemented in both external beam radiation therapy and brachytherapy, simplifying
the dosimetric procedures to a single device. Accurate assessment of the dose given to the
patient using a reliable dosimeter with a water-equivalent response would save time and money
over using multiple dosimeters with a significant over- or under-response to low energy X-rays
and inferring the dose via correction factors and calculations.
-
9Table 1. List of criteria for in vivo point-based real-time dosimeter.
Criterion
#
Criterion Comments
1 small size(22)
(
-
10B. Current in vivo Dosimeters
Several dosimeters currently available on the market are used for some in vivo
dosimetry measurements. These are in vivo ion chambers, diodes and MOSFETs (metal oxide
semiconductor field effect transistors). An ion chamber is classified as an absolute dosimeter: it
can be used to measure the absorbed dose to its own sensitive volume without any calibration.(10)
However, it often needs to be calibrated as knowing the exact measurement volume, and mass of
air contained, is required for absolute dosimetry. The ion chamber measures the total ionization
produced by electron interactions in air, the charge is collected by an electrode set at a high
voltage, and this value can be related to dose. The mass of air contained must yield reasonable
signal statistics, limiting in vivo ion chambers to dosimetry in bladder and rectum, too large for
use in tissue.
A silicon semiconductor diode is what is generally used as an in vivo radiation detector.
When the diode is exposed to ionizing radiation, electron-hole pairs are created throughout the
diode, and as they move through the diode, a measurable current is created. The amount of
silicon required for measurement (known as the die) is 0.01-0.1 mm3, and the current related to
deposited dose can be measured in real-time via the coaxial cable by the electrometer.(23) For in
vivo dosimetry, the die is covered with material for protection and for proper buildup. The
buildup is necessary for skin dose measurements to give the photons enough medium to interact
with in order to create the electrons that will in turn interact with the silicon.(23) This buildup
and the protective cover make the diodes rather bulky (up to 3 cm in length and ~7 mm in
diameter), making it difficult or impossible to position into all tissues of the patient because of
their size. Some diode configurations also suffer from angle dependence of their response (due
to inhomogeneity of the buildup material and coaxial cable attached), and they have been shown
to perturb the dose distribution directly behind the diode by as much as 30%, with the effect
more pronounced at lower beam energies.(24) Read-out was also shown to vary with
temperature,(24,25) complicating dosimetry further by the fact that the temperature of the diode
may not always be known during its use.
A MOSFET is a small silicon transistor.(26) A p-type silicon semiconductor sits on an
insulating oxide layer, which separates it from a conducting metal band.(27) The p-type silicon
has positive charges accumulated within, proportional to the negative voltage bias applied at the
conducting metal band. The measurement of dose is related back to the threshold voltage
(voltage required to allow current to flow through the semiconductor).(26,27) As the ionizing
radiation travels through the MOSFET, charges generated within are trapped, causing the
-
11threshold voltage to shift proportionally to deposited dose.(26,27) MOSFETs can be used as real-
time dosimeters(27) and are much smaller than diodes (some are as small as 1 mm in diameter,
known as micro-MOSFETs), decreasing beam attenuation and dose perturbation effects that are
observed for diodes.(26) Their small size also allows for dosimetry in small beams, down to a few
mm (4.4 mm) in beam diameter.(28) Having high Z components and not being water-equivalent,
MOSFETs have large differences in calibration factors,(29) and require separate calibrations to be
performed at different beam energies.(26) On the other hand, they show good agreement with the
ion chamber for a given energy down to a depth of 34 cm.(30) They have also been shown to
exhibit directional anisotropies in response because of the silicon substrate beneath the sensitive
volume,(28,31) and, like diodes, are known to exhibit temperature dependence.(26)
C. Optical Methods
Energy independence of a dosimeter can in large part be met by staying clear of metallic
components within the dosimeter. As such, there has been a considerable amount of effort over
the last few decades to find a dosimeter based on optical characteristics of a radiation sensitive
medium and fibre-optic readout. Among such radiation sensitive media are doped optical
fibres,(32-34) plastic scintillators,(35-39) thermoluminescent dosimeters (TLDs),(34,40) optically
stimulated luminescent (OSL) dosimeters,(34,41,42) and a fluorescing ruby.(43) The media can be
subdivided into two categories: light emitters and light modifiers. Light emitting dosimeters
(such as scintillators, TLDs and OSL media) produce signal that is proportional to the absorbed
dose. Thus the number of photons and the signal statistics are dependent on dose, and is out of
user’s control. On the other hand, light modifying dosimeters alter some aspect of the
interrogating light, the properties and the intensity of which is controlled by the user. This
allows for higher precision measurements, because the number of photons can be increased if the
noise is too high. The other major difference between the two types of optical dosimeters is that
the light emitting media are reusable, whereas light modifying media have to be disposed off
after a certain dose. Reusable dosimeters are often less expensive per use, but “age” and one
must be careful to not assume the signal per dose remains constant as the total dose delivered is
increased. Dosimeters that integrate dose to give a single signal at the end, such as light
modifying media, have the ability to always keep the reading, and can be measured multiple
times as the read-out is non-destructive. As they are also disposable, the need for disinfecting
between patients is eliminated, simplifying their use, as well as reducing the risk of spreading
infection.
-
12Fluorescing rubies and scintillating fibres automatically emit light when exposed to
ionizing radiation, whereas TLDs and OSL dosimeters must be stimulated by either heat or light,
respectively, to obtain a light signal. These materials work by trapping electrons in higher
energy states when they are exposed to ionizing radiation. When the electrons move down
(either automatically or due to stimulation) to their ground state, photons corresponding to the
energy difference between the two states are emitted. TLDs cannot be read out in real-time, as
they require annealing after irradiation and a lengthy read-out processes for accurate dose
estimate.(5) When one considers the high temperatures that TLDs must be heated to (at least 100
°C, depending on the type of TLD),(28,40) it is hard to imagine how this would be done safely
within a patient. Some fibre-based read-out schemes have been suggested,(40) but have not been
implemented clinically. OSL and plastic scintillator dosimeters are promising, and some have
been made to be nearly energy independent.(38) However, they continue to suffer from
interferences such as fibre scintillation and Cerenkov radiation, where removal of the latter often
requires accurate knowledge of pulse sequences and careful timing.(42,44,45) The alternative is to
use scintillators that have high emission wavelength, such that the Cerenkov radiation (which
drops off with 1/λ3) from the fibre doesn’t interfere much with the dose-related signal from the
sensor.(37) However, the signal per given dose from such scintillators is generally decreased
compare to the signal from scintillators with low emission wavelength, and thus measurements
of dose are noisy.(37) While rubies fluoresce at high enough wavelength and long enough after
the pulse such that the Cerenkov radiation is irrelevant to the measurement, they have a high Z
and are not water equivalent.(43)
Light modifying dosimeters, such as doped optical fibres and Fricke xylenol-orange
solutions or gels create light-absorbing colour centres when exposed to ionizing radiation. This
is done either via electron trapping,(46) or via formation of a complex with a dye,(47) respectively.
Doped optical fibres are generally not water equivalent due to high Z (often Pb) components
used as doping material.(33,34,46) A method for reducing Z by incorporating dopants such as Na,
Mg, and Li has been proposed.(48) However, these optical fibres have not been implemented,
likely because of reduced sensitivity compared to higher-Z counterparts. Finally, while certain
gels can be used as optical dosimeters, these are typically utilized in 3D dosimetry by making 3D
phantoms out of the gel,(49-52) and no effort to incorporate them in fibre-optic dosimeter has been
made; rather, they are used in post-exposure volumetric readout (e.g. MR).
-
13
V. Outline of Thesis
Another type of optical dosimeter makes use of what is known as a radiochromic medium. This
type of material changes colour, or gets darker, upon exposure to ionizing radiation, and is in the
category of light modifiers. Some radiochromic films are manufactured under the name of
GafChromic® (International Specialty Products, or ISP). These films contain one or two gelatin
layers with organic monomers arranged in a small crystal or micelle-like structure suspended
within them.(53-55) The monomers undergo polymerization when exposed to radiation. The
absorbance spectrum of the resulting polymer systems is then related back to the absorbed
dose.(55-57) Historically, these films are used for two-dimensional dose distribution
measurements,(58,59) and the measurements are performed 3-24 hours (depending on the
film)(54,55) after the end of irradiation to ensure stable readout. This is because the
polymerization reaction is not instantaneous, and proceeds even after the source of ionizing
radiation is removed. This, in turn, causes the absorbance to change with time, producing errors
in dose estimate. (59-61)
Despite the recommendation that these media be read out 3-24 hours after irradiation,
radiochromic media are being investigated in this work for applicability in real-time patient
dosimetry. They have some advantages, including a near water-equivalent organic
composition,(54,55) and the ability to produce signal from a sub cubic millimetre volume.(55) They
also absorb predominantly in the red region of the visible spectrum, where Cerenkov radiation
does not interfere. Because the radiochromic material is a light modifier, the signal statistics can
in part be controlled by the user, by increasing or decreasing the interrogation light. More
importantly, if the performance of these systems during and after irradiation can be characterized
and accounted for, they may provide real-time dose estimates with an acceptable error despite
the above-mentioned issues. If these systems are understood, reverse engineering may be
possible to create a radiochormic material that polymerizes faster and has appropriate sensitivity
for a given application.
In the present work, response of two films (GafChromic MD-55 and EBT) were assessed as
a function of dose, time, dose rate, temperature, and energy and results are described in Chapters
2 to 6. Chapter 2 investigates in detail the possibility of using the radiochromic material,
GafChromic MD-55, as a point-based dose measurement material in real-time. Although
throughout the experiments described in this thesis the measurements were made immediately
after the end of irradiation, the endpoint was chosen only because this is when the dose is known.
It is easy to imagine how, once the relationship between optical density and dose is established,
-
14optical density measurements can be made any time during irradiation. Thus, the dose can be
estimated during irradiation as well, making it a true real-time dosimeter.
The film was investigated for signal linearity, reproducibility, dose rate and temperature
dependence. Chapter 3 compares the performance of GafChromic MD-55 with a medium from
another film, GafChromic EBT. This chapter focuses on differences in sensitivity and linearity
of MD-55 and EBT, and discusses the fundamental chemical and structural difference between
the two monomer systems. Chapter 4 describes the investigation and quantification of the dose
rate dependence of GafChromic EBT. Chapter 5 describes temperature and humidity
investigations of GafChromic EBT (performed in collaboration with Dr. D.F. Lewis and Dr. S.
Varma, researchers of ISP). Chapter 6 compares energy dependence between two sensitive
media present in three films (MD-55, EBT and HS, where MD-55 and HS use the same
formulation). A novel mathematical model of the response to dose with time both during and for
short periods after the end of irradiation is also developed, with the preliminary results described
in Chapter 7. The parameters of the model are based on physical properties and processes
occurring during exposure to ionizing radiation and interrogation with read-out light. Ideally,
this would allow for future engineering or selection of radiochromic media that meet the in vivo
requirements, by working backwards from the desired response to radiation as predicted by the
model to physical and chemical properties. The thesis concludes with a summary of current
investigations and ideas for future work.
1. Suntharalingam N., Podgorsak E., Tolli H. Brachytherapy: Physical and Clinical Aspects.
In: Podgorsak E., ed. Radiation Oncology Physics: A Handbook for Teachers and
Students. Vienna: International Atomic Energy Agency, p. 451-484, 2005.
2. Suntharalingam N., Podgorsak E., Hendry J.H. Basic Radiobilogy. In: Podgorsak E., ed.
Radiation Oncology Physics: A Handbook for Teachers and Students. Vienna:
International Atomic Energy Agency, p. 485-504, 2005.
3. ICRU 50. Prescribing, Recording, and Reporting Photon Beam Therapy. Bethesda:
International Commission on Radiation Units and Measurements, 1993.
-
15 4. Burman C., Kutcher G., Emami B., Goitein M. Fitting of normal tissue tolerance data to
an analytic function. International Journal of Radiation Oncology Biology Physics 21:
123-35, 1991.
5. Essers M., Mijnheer B.J. In vivo dosimetry during external photon beam radiotherapy.
International Journal of Radiation Oncology Biology Physics 43: 245-59, 1999.
6. Feldman A., Edwarads F.M. The routine use of personal patient dosimeters is of little
value in deteting therapeutic misadministrations; Point/Counterpoint. Medical Physics
28: 295-7, 2001.
7. Noel A., Aletti P., Bey P., Malissard L. Detection of errors in individual patients in
radiotherapy by systematic in vivo dosimetry. Radiotherapy and Oncology 34: 144-51,
1995.
8. Papanikolaou N, Battista JJ, Boyer AL, Kappas C, Klein E, Mackie TR, Sharpe M, Van
Dyk J. Tissue Inhomogeneity Corrections for Megavoltage Photon Beams. Medical
Physics Publishing, 2004.
9. Kutcher G., Coia L., Gillin M., Hanson G.K., Leibel S., Morton R.J., Palta J.R., Purdy
J.A., Reinstein L.E., Svensson G.K., Weller M., Wingfield L. Comprehensive QA for
radiation oncology: Report of AAPM Radiation Therapy Committee Task Group 40.
Medical Physics 21: 581-618, 1994.
10. Attix F.H. Introduction to Radiological Physics and Radiation Dosimetry. Toronto: John
Wiley & Sons, 1986.
11. Schulz R.J., Almond P.R., Cunningham J.R., Holt J.G., Loevinger R., Suntharalingam N.,
Wright K.A., Nath R., Lempert G.D. A protocol for the determination of absorbed dose
from high-energy photon and electron beams. Medical Physics 10: 741-71, 1983.
12. Almond P.R., Biggs P.J., Coursey B.M., Hanson W.F., Huq M.S., Nath R., Rogers
D.W.O. AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon
and electron beams. Medical Physics 26: 1847-70, 1999.
-
16 13. Ma C.M., Coffey C.W., DeWerd L.A., Liu C., Nath R., Seltzer S.M., Seuntjens J.P.
AAPM protocol for 40-300 kV x-ray beam dosimetry in radiotherapy and radiobiology.
Medical Physics 28: 868-93, 2001.
14. Khan F.M., Doppke K., Hogstrom K.R., Kutcher G.J., Nath R., Prasad S.G., Purdy J.A.,
Rozenfeld M., Werner B.L. Clinical electron-beam dosimetry: Report of AAPM
Radiation Therapy Committee Task Group No. 25. Medical Physics 18: 73-109, 1990.
15. Shortt K.R., Ross C.K., Schneider M., Hohlfeld K., Roos M., Perroche A-M. A
comparison of absrobed dose standards for high-energy X-rays. Physics in Medicine &
Biology 38: 1937-55, 1993.
16. Chiu-Tsao S.T., Schaart D.R., Soares C.G., Nath R. Dose calculation formalisms and
consensus dosimetry parameters for intravascular brachytherapy dosimetry:
Recommendations of the AAPM Therapy Physics Committee Task Group No. 149.
Medical Physics 34: 4126-57, 2007.
17. Knöös T., Wieslander E., Cozzi L., Brink C., Fogliata A., Albers D., Nyström H., Lassen
S. Comparison of dose calculation algorithms for treatment planning in external photon
beam therapy for clinical situations. Physics in Medicine and Biology 51: 5785-807,
2006.
18. Storchi P., Woudstra E. Calculation of the absorbed dose distribution due to irregularly
shaped photon beams using pencil beam kernels derived from basic beam data. Physics in
Medicine and Biology 41: 637-56, 1996.
19. Aspradakis M.M., Redpath A.T. A technique for the fast calculation of three-dimensional
photon dose distributions using the superposition model. Physics in Medicine and
Biology 42: 1475-89, 1997.
20. Carlsson A.K., Ahnesjö A. Point kernels and superposition methods for scatter dose
calculations in brachytherapy. Physics in Medicine and Biology 45: 357-82, 2000.
21. Inanc F. A CT image based deterministic approach to dosimetry and radiography
simulations. Physics in Medicine and Biology 47: 3351-68, 2002.
-
17 22. Dutreix A., Bridier A. Dosimetry for External Beams of Photon and Electron Radiation.
In: K.R.Kase, B.E.Bjarngard, F.H.Attix, eds. The Dosimetry of Ionizing Radiation.
Toronto: Academic Press, Inc., 1985.
23. Yorke E., Alecu R., Ding L., Fontenla D., Kalend A., Kaurin D., Masterson-McGary
M.E., Marinello G., Matzen T., Saini A., Shi J., Simon W., Zhu T.C., Zhu X.R. Diode in
vivo dosimetry for patients receiving external beam radiation therapy. Medical Physics
Publishing, 2005.
24. Marre D., Marinello G. Comparison of p-type commercial electron diodes for in vivo
dosimetry. Medical Physics 31: 50-6, 2004.
25. Saini A.S., Zhu T.C. Temperature dependence of commerically available diode detectors.
Medical Physics 29: 622-30, 2002.
26. Izewska J., Rajan G. Radiation Dosimeters. In: Podgorsak E., ed. Radiation Oncology
Physics: A Handbook for Teachers and Students. Vienna: International Atomic Energy
Agency, p. 71-99, 2005.
27. Gladstone D.J., Chin L.M. Automated data collection and analysis system for MOSFET
radiation detectors. Medical Physics 18: 542-8, 1991.
28. Francescon P., Cora S., Cavedon C., Scalchi P., Reccanello S., Colombo F. Use of a new
type of radiochromic film, a new parallel-plate micro-chamber, MOSFETs, and TLD 800
microcubes in the dosimetry of small beams. Medical Physics 25: 503-11, 1998.
29. Lavallée M., Gingras L., Beaulieu L. Energy and integrated dose dependence of
MOSFET dosimeter sensitivity for irradiation energies between 30 kV and 60Co. Medical
Physics 33: 3683-9, 2006.
30. Chuang C.F., Verhey L.J., Xia P. Investigation of the use of MOSFET for clinical IMRT
dosimetric verification. Medical Physics 29: 1109-15, 2002.
31. Scalchi P., Francescon P., Rajaguru P. Characterization of a new MOSFET detector
configuration for in vivo skin dosimetry. Medical Physics 32: 1571-8, 2005.
32. Bueker H., Haesing F.W. Fiber-optic radiation sensors. SPIE 2425: 106-14, 1994.
-
18 33. Gripp S., Haesing F.W., Bueker H., Schmitt G. Clinical in vivo dosimetry using optical
fibers. Radiation Oncology Investigations 6: 142-9, 1998.
34. Huston A.L., Justus B.L., Falkenstein P.L., Miller R.W., Ning H., Altemus R. Remote
optical fiber dosimetry. Nuclear Instrumets and Methods in Physics Research Section B:
Beam Interactions with Materials and Atoms 184: 55-67, 2001.
35. Becks K.H., Drees J., Goldmann K., Gregor I.M., Heintz M. A multichannel dosimeter
based on scintillating fibers for medical applications. Nuclear Instrumets and Methods in
Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated
Equipment 454: 147-51, 2000.
36. Arnfield M.R., Gaballa H.E., Zwicker R.D, Islam Q., Schmidt-Ullrich R. Optimum
plastic scintillator and optical fiber combinatin for brachytherapy dosimetry. International
Journal of Radiation Oncology, Biology, Physics 32[Supplement 1], 296. 2004.
37. Archambault L., Arsenault J., Gingras L., Beddar A.S., Roy R., Beaulieu L. Plastic
scintillation dosimetry: Optimal selection of scintillating fibers and scintillators. Medical
Physics 32: 2271-8, 2005.
38. Beddar A.S., Briere T.M., Mourtada F.A., Vassiliev O.N., Liu H.H., Mohan R. Monte
Carlo calculations of the absrobed dose and energy dependence of plastic scintillators.
Medical Physics 32: 1265-9, 2005.
39. Petric M.P., Robar J.L., Clark B.G. Development and characterization of a tissue
equivalent plastic scintillator based dosimetry system. Medical Physics 33: 96-105, 2006.
40. Huston A.L., Justus B.L. All-optical, rapid readout, fiber-coupled thermoluminescent
dosimeter system. [Patent US 5606163]. 2004. United States.
41. Aaki F., Ishidoya T., Ikegami T., Moribe N., Yamashita Y. Application of a
radiophotoluminescent glass plate dosimeter for small field dosimetry. Medical Physics
32: 1548-54, 2005.
42. Gaza R., McKeever S.W.S., Akselrod M.S. Near-real-time radiotherapy dosimetry using
optically stimulated luminescence of Al2O3:C: Mathematical models and preliminary
results. Medical Physics 32: 1094-102, 2005.
-
19 43. Jordan K.J. Evaluation of ruby as a fluorescent sensor for optical fiber-based radiation
dosimetry. SPIE 2705: 170-8, 1996.
44. Archambault L., Beddar A.S., Gingras L., Roy R., Beaulieu L. Measurement accuracy
and Cerenkov removal for high performance, high spatial resolution scintillation
dosimetry. Medical Physics 33: 128-35, 2006.
45. Clift M.A., Johnston P.N., Webb D.V. A temporal method of avoiding the Cerenkov
radiation generated in organic scintillator dosimeters by pulsed mega-voltage electron
and photon beams. Physics in Medicine and Biology 47: 1421-33, 2002.
46. Bishay A. Radiation induced color centers inmulticomponent glasses. Journal of Non-
Crystalline Solids 3: 54-114, 1970.
47. Gupta B.L., Narayan G.R. G(Fe3+) values in the FBX gosimeter. Physics in Medicine
and Biology 30: 337-40, 1985.
48. Hasing F.W., Pfeiffer F., Buker H. Multipass cavity sensor for measuring a tissue-
equivalent radiation dose. Forschungszentrum Julich GmbH. [Patent US 6041150]. 2000.
49. Bero M.A. Development of a three-dimensional radiation dosimetry system. Medical
Physics 29(12), 2965. 2002.
50. Oldham M., Siewerdsen J.H., Kumar S., Wong J., Jaffray D.A. Optical-CT gel dosiemtry
I: Basic investigations. Medical Physics 30: 623-34, 2003.
51. Oldham M., Siewerdsen J.H., Shetty A., Jaffray D.A. High resolution gel-dosimetry by
optical-CT and MR scanning. Medical Physics 28: 1436-45, 2001.
52. Wuu C.S., Schiff P., Maryanski M.J., Liu T., Borzillary S., Weinberger J. Dosimetry
study of Re-188 liquid balloon for intravascular brachytherapy using polymer gel
dosimeters and laser-beam optical CT scanner. Medical Physics 30: 132-7, 2003.
53. Klassen N.V., van der Zwan L., Cygler J. GafChromic MD-55: Investigated as a
precision dosimeter. Medical Physics 24: 1924-34, 1997.
54. Todorovic M., Fischer M., Cremers F., Thom E., Schmidt R. Evaluation of GafChromic
EBT prototype B for external beam dose verification. Medical Physics 33: 1321-8, 2006.
-
20 55. McLaughlin W.L., Yun-Dong Chen, Soares C.G., Miller A., Van Dyk G., Lewis D.F.
Sensitometry of the response of a new radiochromic film dosimeter to gamma radiation
and electron beams. Nuclear Instruments and Methods in Physics Research Section A:
Accelerators, Spectrometers, Detectors and Associated Equipment 302: 165-76, 1991.
56. McLaughlin W.L., Puhl J.M., Al-Sheikhly M., Christou C.A., Miller A., Kovacs A.,
Wojnarovits L., Lewis D.F. Novel radiochromic films for clinical dosimetry. Radiation
Protection and Dosimetry 66: 263-8, 1996.
57. McLaughlin W.L., Al-Sheikhly M., Lewis D.F., Kovacs A., Wojnarovits L. A
radiochromic solid-state polymerization reaction. In: Irradiation of Polymers:
Fundamentals and Technological Applications. Washington: American Chemical
Society, p. 152-166, 1996.
58. Ramani R., Lightstone A.W., Mason D.L.D., O'Brien P.F. The use of radiochromic film
in treatment verification of dynamic stereotactic radiosurgery. Medical Physics 21: 389-
92, 1994.
59. Mack A., Mack G., Weltz D., Scheib S.G., Böttcher H.D., Seifert V. High precision film
dosimetry with GafChromic® films for quality assurance especially when using small
fields. Medical Physics 30: 2399-409, 2003.
60. Ali I., Costescu C., Vicic M., Dempsey J.F., Williamson J.F. Dependence of
radiochromic film optical density post-exposure kinetics on dose and dose fractionation.
Medical Physics 30: 1958-67, 2003.
61. Chu R.D.H., Van Dyk G., Lewis D.F., O'Hara K.P.J., Buckland B.W., Dinelle F.
GafChromic dosimetry media: A new high dose, thin film routine dosimeter and dose
mapping tool. Radiation Physics and Chemistry 35: 767-73, 1990.
-
21
CHAPTER 2: REAL-TIME RESPONSE OF GAFCHROMIC® MD-55 FILM TO IONIZING RADIATION
Portions of the following have been published as “Suitability of radiochromic medium for real-
time optical measurements of ionizing radiation dose” by Alexandra Rink, I. Alex Vitkin, and
David A. Jaffray in Medical Physics 32(4), p. 1140-1155 (2005)
-
22
I. Introduction
The goal of these investigations is to develop a dosimeter that is economically and logistically
acceptable (low-cost, disposable, reusable, and sterilizable). To meet these requirements, a water
equivalent dosimeter which undergoes an immediate change in optical properties upon exposure
to ionizing radiation is proposed. The difference in a particular quantitative optical property is to
be measured via optical fibers, and ionizing radiation dose is to be inferred through a calibration
model. In the initial embodiment of the device, the radiation sensitive material present in
GafChromic® MD-55 radiochromic film was investigated for suitability in the application of
external beam patient dosimetry. A list of requirements in choosing an in vivo dosimeter against
which GafChromic MD-55 film is investigated is given in Table 2.
Review of GafChromic® MD-55
To better understand the results presented in this paper, the reader is provided with a
review of literature and explanation of solid-state polymerization. An understanding of the
process that forms the basis for radiochromic dosimetry is required if the real-time dosimetry
system is to be quantified and optimized. The time course of the energy transfer and subsequent
processes that lead to changes in optical density are also important for rational design.
General Experience
Radiochromic films have been used for nearly 30 years in the field of dosimetry.(1)
Commercially available radiochromic dosimeters are manufactured by International Specialty
Products (ISP), and some are sold under the product name of GafChromic® MD-55. A broad
assessment of its characteristics suggest that it is a good candidate for the proposed point-based
dosimeter: the sensitive medium from GafChromic® MD-55 film can be packaged as a small
volume placed at the tip of an optical fiber (closed system to minimize any interference from the
tissue, such as humidity); it has response characteristics within 5% of water and striated muscle
for photons of energy in the range of 0.1-10 MeV, and electrons in range of 0.01-30 MeV.(2)
Upon exposure to heat, ultraviolet (UV) light, and high-energy photons and electrons, the
monomers polymerize to provide an absorbance spectrum with two peaks (675 and 615 nm),
creating a polymer with a blue tint.(2-4) The signal linearity requirement appears to have also
been met, since the change in absorbance is a linear function of the absorbed dose,(5) although the
dynamic range of this function depends on the wavelength at which the measurements are
-
23Table 2. Evaluation criteria for in vivo point-based real-time dosimeter.
Criterion
#
Criterion Comments
1 small size(6)
(
-
24obtained.(7-11)
The film has been reported to resolve dose to 1.5 Gy with a precision of 5% or better,
using the 671 nm absorption peak,∗ and this resolution can be further increased by increasing the
thickness of sensitive layer.(12) The requirement of real-time readout appears to be a significant
impediment to use of GafChromic® MD-55 film. Time frames of 24 – 48 hours after exposure
are recommended. Additional investigations are required to resolve this issue.
Less than 5% difference in net change in absorbance of GafChromic® MD-55 film
exposed to 10 Gy at dose rates of 0.034-3.422 Gy/min is expected.∗ However, validity of the
measurements done with this film has been questioned for low dose-rate brachytherapy. Ali and
colleagues reported in 2003 the kinetics of film darkening as function of post-exposure time
depends on the total dose, with the development being faster at the lower doses.(13) These
findings are a concern for real-time dosimetry and require further investigation. While the focus
of this study is to apply the dosimeter in the context of external beam radiotherapy, where dose-
rates are typically greater than those in brachytherapy, a range of doses and dose-rates, over
which post-exposure development from the first few fractions of the treatment does not introduce
error in the absorbance reading and final dose estimate, should be clearly defined.
McLaughlin et al. [1996] reported that propagation of the polymerization is complete
within 2 ms of a single 20 Gy 50 ns pulse.(4) It is unclear, however, if the polymerization
occurred mostly due to ionizing radiation or heat. There literature describes a continuous
increase in absorbance even after irradiation is complete,(14,15) with the absorbance being a
function of a logarithm of elapsed time.(16) Hence, it has generally been recommended to
perform the measurements 24 h (2) to 48 h later(16) by both researchers and manufacturers.*
Measurements are further complicated by the shift in wavelength of maximum absorbance (λmax)
to lower wavelengths as dose increases.(2,7,14,15)
GafChromic® MD-55 film is stable during storage or short exposures to ambient light,(17)
satisfying part of seventh requirement of in vivo dosimeter. However, the temperature
dependence of absorbance of GafChromic® MD-55 is complicated, and humidity and pressure
dependence poorly documented. Increase in temperature reported to correspond to a decrease in
absorbance and a peak shift to lower λ (λmax = 677.5 nm at 18.6°C, 673 nm at 28.0 °C for 6.9
Gy),(2,16) with the latter effect being reversible if temperature fluctuations occur during
∗ International Specialty Products (ISP) product information.
-
25measurement, not but irreversible if the temperature was varied during irradiation.(2) Others
report an increase in absorbance with an increase in temperature.(14,18) This discrepancy is likely
due to a choice of wavelength for absorbance measurements, as the λmax depends on temperature,
and also due to a range of temperatures sampled. It has been shown that a He-Ne laser operating
as low as 0.1 mW will cause an increase in absorbance of GafChromic® MD-55 in five minutes,
with this effect being stronger for films exposed to smaller doses.(19) For this reason, the
absorbance measurements should be performed using low optical powers to prevent
polymerization due to the heat produced by the light. Above 60°C, the colour of the film
changes from blue to red, as the