small field dosimetry radiochromic film - aapm...

23

Upload: hadat

Post on 13-Aug-2019

228 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael
Page 2: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

Small Field Dosimetry

&

Radiochromic film

Michael Kissick, Ph.D., Assistant Professor,

Medical Physics & Human Oncology &

Affiliated with the

Morgridge Institute for Research

North Central AAPM Chapter Symposium, 10/10/2013

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

Page 3: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

• I have no potential conflicts of interest.

Page 4: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

Outline and Thesis/Objectives:

I. Small Field Dosimetry – review

II. Radiochromic film (RCF) in general – overview

III. QC/QA for high dose hypofractionated radiotherapy – A few

concerns*, many other reasons not to be concerned.

IV. Conclusions – this film will continue to be useful

Learning Objectives:

I. Radiochromic film is fascinating. The topochemical process

involved is special in chemistry, and very useful in dosimetry

after the vendor has spent decades perfecting it for our use.

II. The QC/QA of the future will likely use this film even more than it

is used now, because of hypofractionation and the need for

anthropomorphic phantoms with high dose gradients for small

fields. Some concerns I have … mostly, RCF is an excellent

choice.

I will assume most of you know the basic positives/negatives of RCF.

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

* Composite fields / fractionation

Page 5: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

What is a small field?

According to Alfonso et al [1]

2008, page 5180, first paragraph:

“A small field is defined as a

field with a size smaller than the

lateral range of charged

particles.”

R. Alfonso et al, “A new

Formalism for Reference

Dosimetry of Small and

Nonstandard Fields,” Medical

Physics 35 (11), Novenber 2008,

5179-5186.

1νh

1νh

2νh

2νh1T

1T

2T

2T

Loss of charged particle equilibrium

for the small field � We need a

dosimeter with resolution better than

the range of the average charged

particle: FILM IS SUCH A DOSIMETER!

If charged particle equilibrium:

TKDose enc ∝∝∝ µ

If no charged particle equilibrium:

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

Then we are sensitive to unknown

aspects of the charged particle spectrum.

Page 6: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

RCF does not have the same small field issues

that ion chambers have – it does not have an air

filled cavity.

1. If one attempts to make a small ion chamber for resolution, and the media are not perfectly

matched, then delta-ray equilibrium can be a problem.b

2. Photon energy determines everything, but the proportionality of a ratio of readings to a ratio of

absorbed dose is generally robust to the photon spectrum.a

3. This is because it is well-known that ratios of stopping powers are robust to the charged

particle energy spectrum.b However, we really need charge particle equilibrium for this.

4. Because RCF is not a gas, the energy dependence resulting from a density effect mismatch is

not a concern.

5. It is the tissue equivalence of this film that is a major help for us.

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

a. F. Verhaugen et al., “Monte-Carlo dosimetry study of a 6 MV stereotactic radiosurgery unit,”

Phys. Med. Biol. 43 (1998) 2755-2768.

b. Radiological Physics and Dosimetry, Frank Herb Attix, Wiley-VCH, Germany, 2004 edition

Page 7: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

RCF has the spatial resolution needed, and it does

not itself change equilibrium.

(It is tissue equivalent also)

4 mm beamlet From:

N. Hardcastle et al. “High dose per fraction dosimetry of

small fields with Gafchromatic EBT2 film,” Medical Physics

38 (7), July 2011, 4081-4085.

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

So, we can image the penumbra from lateral electron motion.

Page 8: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

In the context of radiotherapy’s future, small

fields pose the following challenges for

dosimetry*:

1. Some innovative machines with small fields CANNOT provide the standard reference field conditions, i.e., they cannot provide the 10 cm X 10 cm uniform field at 100 cm distance. This is true of helical tomotherapy and the CyberknifeTM.

2. Also, there is no standard for composite fields, i.e., helical tomotherapy inherently provides dose with composite fields.

3. For RCF, for composite fields, an issue for me concerns repeated applications of dose – new polymerizations applied on top of previous polymerizations – Note: we have seen fractionation effects, confirmed by David Lewis.

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

* Alfonso et al Med Phys 35 (2008) 5179.

Page 9: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

How does RCF work? It is made up of

special crystals: LiPCDA.

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

From David Lewis

talk, Oct. 20, 2010

Polydiacetelyenes in hair-like crystals:

Page 10: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

The crystals are made up of stacks of

monomers that polymerize in place:

topochemical phase change, remarkable!

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

CC

CC*

R

R

*C

CCC

R

R

CC

CC

R

R

**C

CCC

*

R

R

*C

CCC

R

R

CC

CC

R

R

**

H. Gross et al., Chem. Phys. Lett. 95 (1983), p. 584

Page 11: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

The polymerization process is inherently

nonlinear: the activation energy barrier

could respond to local strain for example:

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

Polydiacetylenes, Editor: Hans-Joachim Cantow, Springer-Verlag, ‘84

These barriers could

change size as the

fraction polymerized

changes – understudied!

Page 12: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

Some complications … known for a long

time … Not perfectly topochemical –

strain develops!

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

From: R.W. Carpick et al., Langmuir 16 (2000), page 4639

Notice the

strain this

causes!

And

rotational

strain too:

Both can

cause OD

Changes!

Page 13: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

This local strain, if different from the

surrounding crystal, could cause OD

changes at low dose …

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

* N.V. Klassen et al,

Med. Phys. 24 (1997) 1924.

•R.H. Baughman, J. Chem Phys. 68 (1978) 3110.

Page 14: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

There are also temperature phase

changes due to the hydrogen bonding to

the substrates:

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

Q. Huo, et al., Langmuir 15 (1999) 3972-3980

U. G. Hofmann, et al., Langmuir 17 (2001) 1518-1524

Page 15: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

We saw even more worrisome issues:

fractionation effects! – scanner or film?

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

Fractionation Effects ...

0

0.01

0.02

0.03

0.04

0.05

0.06

0.07

0 10 20 30

total accumulated dose

Inc

rem

en

tal s

en

itiv

ity

: d

(OD

)/d

(do

se

)

1Gy fract., 5Gy pre

0.5Gy fract., 5Gy pre

2Gy fract., 5Gy pre

1Gy fract., 0Gy pre

0.5Gy fract., 0Gy pre

2Gy fract., 0Gy pre

5 Gy fract., 0 Gy pre

5 Gy fract., 5 Gy pre

E.Y. Hirata et al., “Low Dose Fraction Behavior

of High Sensitivity Radiochromic Film,” Med.

Phys. 32 (2005) 1054HS film

Biggest

issues

for small

doses atop

big doses !

Page 16: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

David Lewis made me aware of the issue

that the field flatness depends on density

in the Epson 10000XL flatbed scanner:

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

From David Lewis

talk, Oct. 20, 2010 … he claims this is due to polarization of light through film !

Note: I avoided

the issue by using

small pieces of film

here! – benefit

for small fields and

RCF !

Page 17: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

Now with EBT2,3, there are dyes:

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

Film looks yellow – red/yellow is NOT absorbed when film UNexposed,but blue is always absorbed.

Film looks green – red/yellow and blue BOTH absorbed – leaving only green to get through when exposed.

But, we have seen the scanner

Behave oddly for one channel,

But not another!

McCaw et al Med. Phys. 38 (2010) 5771-7

Page 18: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

The UW ADCL has developed (Ben Rosen)

a new scanning laser densitometer system

(LDS)* that eliminates many scanner

issues!

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

From: Ben Rosen talk at the

AAPM 55th Annual Meeting

Scientific: Therapy: Dosimetry,

Patient Safety and QA Procedures:

Calibration WE-E-141-3

8/7/2013 2:00-3:50 pm

* This will be patented by WARF

{Would be good to have

dual energy, but the laser

is adjustable in frequency}

Page 19: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

I have used RCF in anthropomorphic phantoms to explore

interplay for helical tomotherapy:

need film for this (small wiggles)!

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

The Phantoms of Medical

and Health Physics,

L. DeWerd &

Michael Kissick Ed.

Springer, 2013

M. W. Kissick et al., “A

phantom model demonstration

of tomotherapy dose painting

delivery, including managed

respiratory motion without

motion management,”

Phys. Med. Biol. 55 (2010)

2983-2995.

Page 20: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

Of great concern to me about the use of film

for DQA… can it be used to match DVH type

metrics?

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

H. Zhen et al., “Moving from gamma

Passing rates to patient DVH-based

QA metrics in pretreatment dose QA,”

Med. Phys. 38 (10) Oct. 2011, 5477-89.

•We have to be careful about

how we use film for QC/QA

on 3D volumes sampled with

2D dosimetry (film).

There is good reason to drop

the use of ‘gamma’ as a useful

metric for patient-specific QA!

Only this case showed correleations

Between DVH based and gamma based

Metrics, and it was weak.

Page 21: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

In the end, I like this film! It is improving!

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

a. high dose capable so competes with EDR film (which has lots of high Z)

b. great spatial resolution because its film

c. tissue equivalent

d. develops itself (solid state polymerization ‘in place’ (topochemical))

e. not very sensitive to visible light, but it is to UV like from sun

f. now it is mainly energy independent to a large degree

a. sensitivity varies with temperature, UV, (and low dose, fractionation issues).

b. polarized densitometer light affects OD reading (but EBT model improving this)

c. developing takes ~ days, but “saturates” at ~ weeks really:[initial phase = starts < 100 µsec, slow phase = days – weeks, ~log(time)]

d. An issue is the nonlinearity at low doses (10% of maximum): seems to be fraction dependent!

e. needs to be handled very consistently

f. orientation and position dependencies on most scanners – scanners are a big issue!

Pros Cons

1. Gafchromatic film is only made by one vendor.

2. It was invented by David Lewis.

3. It was first used for an image in 1984 for an electron beam

recorder.

4. The film activates at about 10-3 J/cm2 versus.

5. Silver halide film activates at 10-10 – 10-8 J/cm2.

6. The new yellow dye for EBT2,3 is brilliant!

Page 22: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

I am on the new TG-235, representing

the UW ADCL.

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

(I am writing the QA for Tomotherapy and Cyberknife, and MV IGRT QA parts.)

Page 23: Small Field Dosimetry Radiochromic film - AAPM Chapterchapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Kissick.pdf · Small Field Dosimetry & Radiochromic film Michael

Conclusions:

DEPARTMENT OF MEDICAL PHYSICS, UNIVERSITY OF WISCONSIN-MADISON

1. The diacytelene tomochemical phase transition process in radiochromic

film is fascinating!

2. There are still unanswered questions from: i. the film, ii. the scanner, and

iii. its use for modern QC/QA., i.e., fractionation / composite fields

issues.

3. The future appears to be very bright for small field dosimetry with

radiochromic film !!