qc analytic 1
DESCRIPTION
chemTRANSCRIPT
2007
Instrumental Analysis:Spectrophotometric Methods
•Understand interaction between light and matter
(absorbance, excitation, emission, luminescence,fluorescence, phosphorescence)
•Describe the main components of a spectrophotometer,
(sources, monochromators, detectors, interferometer, grating, ATR, ICP, )
•Make calculations using Beer’s Law(analyse mixture absorption)
•Understand the mechanism and application of UV-Vis, FTIR, Luminescence, atomic spectroscopy
By the end of this part of the course, you should be able to:
Background knowledge:
What you are expected to know before the course:
Error analysis in quantitative analysisSolve linear equationsComplementary colourExponential and logarithm
What you are recommended to know before the course:
Least square fittingBasic quantum chemistryMolecular symmetry
If you have difficulty to understand above topics, find extra reading materials!
Or discuss with me after the lecture.
If you are trying to learn above topics, please let me know.
Today’s lecture: (Instruments based on light interaction with matter)
• Properties of light• Molecular electronic structures• Interaction of photons with molecules• Spectrophotometer components
• Light sources• Single and double beam instruments• Monochrometers• Detectors
• Fluorescence spectroscopy
Next week’s lecture:
• Fourier transformed infrared spectroscopy• Interferometer
• Atomic spectroscopy• Quantitative analysis
• Beer’s law• Method validation• Dilution and spike
Light travelling speed:
in other media: c/n (n = refractive index, generally >1)
in a vacuum: c=2.998 x 108 m s-1 (n=1 exactly, in air n=1.0002926)
c/n=
Therefore:
Energy is inversely proportional to wavelength
but proportional to wavenumber
And of course, the relationship between energy and frequency:
E = h = hc/ = hc h = Planck’s constant (6.626 x 10-34 J s)
= wavenumber (most common units = cm-1)
~
~
Light is energy in the form of electromagenetic field
Review on properties of light:photon
Wavelength (): Crest-to-crest distance between waves Frequency (): Number of complete oscillations that the wave makes each second
units: number of oscillations/sec or s-1 or Hertz |(Hz)
Frequency Scanning Techniques: a few definitions
Emission method: source of light is sample
Absorption method: intensities of a source with and without the sample in place are compared
Spectrum: a plot of intensity vs. frequency/wavelength
In quantitative analysis:
common to work at 1 wavelength
running a spectrum is an important initial step (to select best conditions)
Fig. 18-2
Regions of Electromagnetic Spectrum-the “colour” of light
Electronic structures of simple molecule
S0
S1
T1
Bond length
D
Ground state
Excited stateSinglet
Excited stateTriplet
En
erg
y
Dissociated states
Vib
rati
on
sta
tes
Interaction between photon and molecule
S0
S1
T1
D
S0
S1
T1
S0 S1 transition
A F
IR
UV
-vis
P
Electronic structures
Singlet and triplet
Bond length for ground and excited states
Vibrational structures-infrared absorption/transmission (FTIR)
Internal conversion
Intersystem crossing
Photon adsorption excitation (Beer’s law, UV-vis)
Frank Condon condition and The Stokes' shift
Radionless relaxation and vibration relaxation
Luminescence-fluorescence/phosphorescence
Key concept from energy diagram
Type of optical spectroscopy
UV-vis absorption spectroscopy (UV-Vis)FT-IR absorption/transmission spectroscopy (FTIR)Atomic absorption spectroscopy (AAS)Atomic fluorescence spectroscopy (AFS)X-ray fluorescence spectroscopy (XFS)
What you will learn:
The excitation mechanism
Monochromator design
Instrument principle
Quantitative methods
Optical spectrophotometer components
Monochromators
Filters
Grating+slit
prism
Excitation sources
Deuterium Lamp
Tungsten Lamp
Laser
X-ray tube
Mercury lamp
Xenon lamp
Silicon carbide globar
Flame
Furnaces
Plasmas
Hollow-cathode lamp
Detectors
PMT
CCD/CID
Photodiode
Thermocouple
MCT
Pyroelectric detector
UV
UV-vis
X-ray, UV, vis, IR
X-ray
UV-vis
UV-vis
IR
What is the advantage and disadvantage?
Design of optical spectrophotometersSingle Beam vs. Double Beam
Fig. 13-12, pg. 315 "Instrument designs for photometers and spectrophotometers”
(a) single-beam design
(b) dual channel design with beams separated in space but simultaneous in time
(c) double-beam design in which beams alternate between two channels."
(a)
(b)
(c)
Q: what’s the advantage of double beam spectrophotometer?
Light sources
Black-body radiation for vis and IR but not UV
- a tungsten lamp is an excellent source of black-body radiation
- operates at 3000 K
- produces from 320 to 2500 nm
For UV:
- a common lamp is a deuterium arc lamp
- electric discharge causes D2 to dissociate and emit UV radiation (160 – 325 nm)
- other good sources are:
Xe (250 – 1000 nm)
Hg (280 – 1400 nm)
( How much in cm-1, J, Hz and eV?)
Lasers:
- high power
- very good for studying reactions
- narrow line width
- coherence
- can fine-tune the desired wavelength (but choice of wavelength is limited)
- £££ expensive £££
What is the important properties of a source?
BrightnessLine widthBackgroundStabilityLifetime
Sample a source containers:for UV: quartz (won’t block out the light)
for vis: glass [ 800nm (red) to 400 nm (violet)]
for IR: NaCl (to or 15384 nm or 650 cm-1)
KBr (to 22222 nm or 450 cm-1)
CsI (to 50000 nm or 200 cm-1)
Optical transmission coefficient Best material: diamond, why?
High transmissionChemically inert
Mechanically strong
Criteria
Monochromators
Early spectrophotometers used prisms
- quartz for UV
- glass for vis and IR
These are now superseded by:
Diffraction gratings:
- made by drawing lines on a glass with a diamond stylus
ca. 20 grooves mm-1 for far IR
ca. 6000 mm-1 for UV/vis
- can use plastic replicas in less expensive instruments
Think of diffraction on a CD
http://www.mrfiber.com/images/cddiffract.jpg
http://www.ii.com/images/prism.jpg
http://www.veeco.com/library/nanotheater_detail.php?type=application&id=331&app_id=34
10mx10m
Why?
Monochromators: cont’d
Polychromatic radiation enters
Second concave mirror focuses each wavelength at different point of focal plane
Orientation of the reflection grating directs only one narrow band of wavelengths to exit slit
The light is collimated the first concave mirror
Reflection grating diffracts different wavelengths at different angles
http://oco.jpl.nasa.gov/images/grating_spec-br.jpg
What is the purpose of concave mirrors?
Interference in diffraction
d
d sin()+d sin()=n
n=1, 2, 3 In-phase
n=1/2, 3/2, 5/2 out-phase
>0<0
Bragg condition
Phase relationship
Monochromators: reflection grating
Monochromators: reflection grating
Each wavelength is diffracted off the grating at a different angle
Angle of deviation of diffracted beam is wavelength dependent diffraction grating separates the incident beam into its constituent wavelengths components
Groove dimensions and spacings are on the order of the wavelength in question
In order for the emerging light to be of any use, the emerging light beams must be in phase with each other
Resolution of grating:
=nN
Angular resolution:As: d sin()+d sin()=n
So: n =d cos()
Therefore: =n/[d cos()]
What does this mean?
n: diffraction orderN: number of illuminated groves
Monochromators: slit
Bottom line:
- it is usually possible to arrange slits and mirrors so that the first order (n = 1) reflection is separated
- a waveband of ca. 0.2 nm is obtainable
However, the slit width determines the resolution and signal to noise ratio
Large slit width: more energy reaching the detector higher signal:noise
Small slit width: less energy reaching the detector BUT better resolution!
Detectors
Choice of detector depends upon what wavelength you are studying
Want the best response for the wavelength (or wavelength range) that you are studying
In a single-beam spectrophotometer, the 100% transmittance control must be adjusted each time the wavelength is changed
In a double-beam spectrophotometer, this is done for you!
: Radiation-----charger converter
Photomultiplier-single channel, but very high sensitivity
- Light falls on a photosensitive alloy (Cs3Sb, K2CsSb, Na2KSb)
- Electrons from surface are accelerated towards secondary electrodes called dynodes and gain enough energy to remove further electrons (typically 4-12, to 50 with GaP).
- For 9 stages giving 4 electrons for 1, the amplification is 49 or 2.6 x 105)
- The output is fed to an amplifier which generates a signal
- To minimise noise it is necessary to operate at the lowest possible voltage
What decide the sensitive wavelength?
Photodiode Array-multiplex, but low sensitivity
Good for quick (fraction of a second) scanning of a full spectrum
Uses semiconductor material:
Remember: n-type silicon has a conduction electron – P or As doped
p-type silicon has a ‘hole’ or electron vacancy – Al or B doped
A diode is a pn junction:
under forward bias, current flows from n-Si to p-Si
under reverse bias, no current flows
boundary is called a depletion layer or region
Photodiode Array
- Electrons excited by light partially discharge the condenser
- Current which is necessary to restore the charge can be detected
- The more radiation that strikes, the less charge remains
- Less sensitive than photomultipliers several placed on placed on single crystal
- Different wavelengths can be directed to different diodes
- Good for 500 to 1100 nm
- For some crystals (i.e. HgCdTe) the response time is about 50 ns
Could you compare photodiode with CCD detector?
Photodiode Array Spectrophotometer
- For photodiode array spectrophotometers, a white light passes through sample
- The grating polychromator disperses the light into the component wavelengths
- All wavelengths are measured simultaneously
- Resolution depends upon the distance between the diodes and amount of dispersion
No moving parts! Simple mechanical and optical design, very compact.
Photodiode Array Spectrophotometers vs Dispersive Spectrophotometers
Dispersive Spectrophotometer:
- only a narrow band of wavelengths reaches the detector at a time
- slow spectral acquisition (ca. 1 min)
- several moving parts (gratings, filters, mirrors, etc.)
- resolution: ca. 0.1 nm
- produces less stray light greater dynamic range for measuring high absorbance
- sensitive to stray light from outside sources i.e. room light
Photodiode Array Spectrophotometer:
- no moving parts rugged
- faster spectral acquisition (ca. 1 sec)
- not dramatically affect by room light
What are the components 1 to 10?From: http://www.oceanoptics.com/
Property of luminescence spectrum
Fluorescence vs phosphorescence
1. Phosphorescence is always at longer wavelength compared with fluorescence2. Phosphorescence is narrower compared with fluorescence3. Phosphorescence is weaker compared with fluorescence
Absorption vs emission
1. absorption is mirrored relative to emission2. Absorption is always on the shorter wavelength compared to emission3. Absorption vibrational progression reflects vibrational level in the electronic excited
states, while the emission vibrational progression reflects vibrational level in the electronic ground states
4. 0 transition of absorption is not overlap with the 0 of emission
Why?
Why?
Fluorescence spectroscopy
Fluorescence spectroscopy
Emission spectrum: hold the excitation wavelength steady and measure the emission at various wavelengths
Excitation spectrum: vary the excitation wavelength and vary the wavelength measured for the emitted light
Light source
Excitationmonochromator
Referencediode
8% of li
ght
Beamsplitter
sample
EmissionMonochromator
Amplifier ComputerPMT
Q: why the emission is measured at 90 relative to the excitation?
Fluorescence spectroscopy: well defined molecules
• Describe the main components of a spectrophotometer and distinguish between single double beam instruments
• Describe suitable sources for ultraviolet (UV)/visible (vis), infra red (IR) and atomic absorption (AA) instruments
• Describe and assess advantages and disadvantages of various monochromators e.g. Prism, diffraction gratings
• Explain how to asses the quality of grating
• Explain how photomultipliers and diode detectors work• Explain the advantage of multiplex detecting• Describe the luminescence spectroscopy and energy transfer process• Compare the emission and absorption spectrum
Summary of spectrophotometric techniques