stat fax 300series
TRANSCRIPT
Rev C 53
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted foruse by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
300 Series Strip ReaderService Manual
© 1993-1997 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted foruse by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair service. No partof this manual may be copied or redistributed without the express consent of Awareness Technology, Inc. There is no transfer of technology,copyright, trade name, patent, trade secret, or other proprietary right given or implied.
Rev C 11/04
Table of Contents
Introduction .................................... 1Layout ............................................. 2
Principles of Operation ................. 8Photometer ...................................... 8Display, Keypad, and Printer .......... 8System Control ................................ 8Carrier Position Detection ............... 9
Troubleshooting ...........................12Power Supply ................................ 12Photometer .................................... 12Strip Drive ..................................... 13Main PCB ..................................... 14Internal Printer ............................... 14Display .......................................... 14Keypad .......................................... 14Serial Port ...................................... 14Error Messages .............................. 17
Service Procedures .......................18Opening the Instrument ................. 18EPROM Replacement 20Lamp Replacement ....................... 22Filter Replacement ........................ 24Filter Label .................................... 26Printer Jams ................................... 28Printer Repairs ............................... 29
Print Head Replacement 29Printer Platen Replacement 31
Calibration ....................................32Software or Hardware Calibration?33Restore Calibration Data ............... 34
Diagrams .......................................35Block Diagram .............................. 35
PCB Layouts ................................36Main PCB - rev AB ...................... 36Main PCB - from rev Z ................. 37Main PCB – rev L to rev Y ........... 38Main PCB – through rev K ........... 39Printer PCB (300-200C) ............... 40Printer PCB (300-200) .................. 41
Schematics .....................................42Main PCB (rev AB) 1 of 4 ............ 42Main PCB (rev AB) 2 of 4 ............ 43Main PCB (rev AB) 3 of 4 ............ 44Main PCB (rev AB) 4 of 4 ............ 45Main PCB (from rev Z) 1 of 4 ...... 46Main PCB (from rev Z) 2 of 4 ...... 47Main PCB (from rev Z) 3 of 4 ...... 48Main PCB (from rev Z) 4 of 4 ...... 49Photometer PCB ........................... 50Printer PCB ................................... 51
Rev C 1
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
IntroductionThis manual describes service and troubleshooting procedures for 300 series instruments. This manual
is written with the service technician in mind, and contains no information for clinical chemistry
analysis, or other applications. For details on instrument operation and specifications, please refer to
the Owner's Manual specific to the instrument.
There are presently two types of 300 series instruments in field use. Both types are equipped with either
a four optical filter photometer configuration or a six optical filter photometer configuration depending
on customer requirements. One type features a single microtiter strip handling device (hereafter
referred to as “Strip Reader with Spring Clip”) and the other type can accommodate up to three
microtiter strips using a different handling device (hereafter referred to as “Strip Reader with Three-
Position Carrier”). Both types of readers will accommodate either 8-well or 12-well strips. The Strip
Reader with Three-Position Carrier is supplied with a strip carrier which will accommodate 12-well
microtiter strips. An 8-well plate carrier is available as an option. Figure 1A illustrates a Strip Reader
with Spring Clip and Figure 1B illustrates a Strip Reader with Three-Position Carrier.
All types of the Model 300 series feature an internal thermal printer which provides a printed record
of all samples and test parameters. All calibration data, test parameters, and other information is stored
in battery-backed (nonvolatile) RAM.
2 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
LayoutFigures 1A and 1B show the exterior of the single-strip and triple-strip style instruments. Figure 2
shows the rear panel. Figure 3 shows the bottom (both styles) of the instrument. Figure 4A shows the
internal chassis layout for the single-strip style instrument and Figure 4B shows the internal chassis
layout for the triple-strip style instrument.
The instrument housing is comprised of a chassis (base and rear panel combined) and a formed plastic
cover. The cover is secured to the base by five #4 self-tapping screws.
A 16 character alphanumeric LED display is mounted to the inside of the cover. A 20 column thermal
printer and the printer PCB mount to a metal printer bracket, which in turn mounts to the inside of the
cover. The display and the printer assembly both connect to the main PCB via 14 conductor ribbon
cable terminated with DIP (dual in-line pin) plugs. The keypad is a membrane-switch type sealed
beneath a graphic overlay and connects to the main PCB by the 8 conductor extension of the switch
layer, referred to as the “keytail”. The serial port connector (DB-9P) is mounted to the rear panel and
connects to the main PCB with a 4 conductor cable and SIP (single in-line pin) header.
The strip mechanism assembly (including the photometer) connects to the main PCB via a 14 conductor
ribbon cable and DIP plug as well as a 10 conductor ribbon cable and header. The 12VDC cooling
fan is mounted to the chassis base and connects to the main PCB via a 2 conductor cable and SIP header.
All AC mains components are mounted on the chassis.
The EPROM is located beneath a hatch on the underside of the instrument to facilitate easy software
upgrades. Early models which do not incorporate the EPROM hatch require the removal of the cover
for software upgrade.
CAUTION
Disconnect AC power cable before servicing.
Rev C 3
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Your particular instrument may differ from the configuration described in this manual. To describe
each instrument model in detail would be impossible due to space limitations. Omitted features may
be ignored. Differences which affect service procedures are noted where applicable.
300 series variations:
• 4 or 6 - position filter wheel
• Thermal printer omitted
• Graphics printer assembly
• Printer assembly mounted to main PCB
• Non-removable power cord installed
• Serial port omitted
• EPROM hatch omitted and EPROM mounted to component side of main PCB
• EPROM device may be one of:
27128 16K bytes
27256 32K bytes
27512 64K bytes
• RAM device may be one of:
6116 2K bytes volatile
48T08 8K bytes nonvolatile, memory-mapped clock
DS1243 8K bytes nonvolatile, serial clock
• PCB revisions–installed components, locations, and jumpers:
Revisions A through K - see “Main PCB - through rev K” under “PCB
Layouts”
Revisions L through Y - shown in illustrations, also “Main PCB - Revisions
L through Y” under “PCB Layouts”
Revisions Z and higher - see “Main PCB - from rev Z” under “PCB
Layouts”
4 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Figure 1A. Instrument Exterior
Strip Reader with Spring Clip
Figure 1B. Instrument Exterior
Strip Reader with Three-Position Carrier
Display
Printer
Paper roll cover
Strip holder
Strip track
Keypad
Paper roll cover
Keypad
Display
Printer
Strip Carrier
Strip track
6 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Figure 4A. Chassis Layout
Strip Reader with Spring Clip
Power
transformer
Strip mechanism
Strip track
Strip
holder
Lamp
bracket
Filter
wheel
motor
Yellow wires
to Strip
Mechanism1
Main PCB
1 Only on Main PCBs from rev Z
Rainbow
Ribbon
Cable to
Strip Mecha-
nism
J8
(test
points)
Gray
Ribbon
Cable to
Strip
Mechanism
EPROM
(underneath main PCB)
Fuse blockStrip drive motor
Rev C 7
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Figure 4B. Chassis Layout
Strip Reader with Three-Position Carrier
Strip mechanism
Strip track
Shuttle
Lamp
bracket
Filter
wheel
motor
Main PCB
Yellow wires
to Strip
Mechanism
Rainbow
Ribbon
Cable to
Strip Mecha-
nism
J8
(test
points)
Gray
Ribbon
Cable to
Strip
Mechanism
EPROM
(underneath main PCB)
Fuse block
Strip drive motor
Power
transformer
8 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Principles of OperationPhotometer
Light from a tungsten-halogen lamp mounted on a bracket above the strip track passes downward
though the well and the sample it contains. The sealed box beneath the sample contains a rotating filter
wheel and photodetector (GaAsP photodiode). The filter wheel contains optical band-pass filters at
selected wavelengths, and is speed controlled, via drivers U9A and U9B in parallel, to approximately
4 rps. As each filter passes in front of the photodiode, a precisely located index hole in the filter wheel
in tandem with an infrared optical switch generates a pulse (FCNT) at U13. Sample-hold capacitor C14
charges to the peak voltage produced by the photodetector and amplifiers U16 and U14B. The sampled
voltage is buffered by U15A and then fed to comparator U15B which compares the sampled voltage
to the output of an exponential capacitor decay circuit built around Q3 and C16. The positive voltage
at the comparator output enables a counter in 8254 timer U4, the value of which is ultimately read by
the microprocessor and used to compute the absorbance. The photodetector output is proportional to
the intensity of the light, whereas the width of the positive phase of the comparator output is proportional
to the absorbance. An oscilloscope can be used to view the important signals at connector J8 while the
instrument is in operation. See Figure 6, Oscilloscope Waveforms.
A DC motor under microprocessor control drives a lead screw via a pulley and neoprene belt. The lead
screw moves the strip holder along the strip track. The pulley incorporates indexing holes so that an
infrared optical switch generates pulses (CLK2) to timer U4 from which the microprocessor derives
position and speed of the strip movement. The movement is zeroed at the front extreme when the motor
stalls. Each well in the microtiter strip can thus be precisely aligned with the photometer assembly.
Display, Keypad, and PrinterA 16 character alphanumeric LED display is mounted inside the cover. The display connects to the
Z80 data and address buses on the main PCB via a 14 pin DIP cable, which also supplies +5VDC
power. A 4 x 4 membrane switch keypad is sealed behind the front panel overlay. The tail of the keypad
switch layer is fed through the cover and installed to connector K1 on the main PCB, which then
connects to 8255 PPI U5. The printer assembly, consisting of a 20 column thermal printer, the printer
PCB, and metal support, is mounted to the cover and is connected to the main PCB via a 14 pin DIP
cable, which provides +5VDC power and the control lines from U5. On early models the printer
assembly is mounted to the main PCB with threaded spacers.
System ControlThe instrument is based on the 8 bit Z80 microprocessor U1, using a 2MHz clock. The software is
permanently stored in EPROM U2. Battery-backed NVRAM (nonvolatile random access memory)
U3, incorporates a real-time clock mapped to the uppermost 8 bytes. Some early models used an
NVRAM with a serial-access real-time clock (DS1243Y). The NVRAM is used store test setups. Most
digital I/O is accomplished with 8255 PPI U5. Time intervals and pulse widths are measured using the
counter channels of 8254 timer U4.
The memory map shown below is for a 27256 EPROM and a 48T08 NVRAM. See the jumper
settings table below for other configurations
Rev C 9
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Carrier Position DetectionThe latest configuration instrument features the ability to read three microtiter strips instead of a single
strip as in earlier models. Earlier models detected the position of the first well. The three - position
carrier configuration uses indexing holes located on the carrier to distinguish between the adjacent
strips. As the carrier moves forward the instrument looks for the front edge of the carrier which blocks
light to the Photometer. A predetermined distance between the edge of the carrier (light blocked) and
the index hole (light sensed) gives the control circuitry the information it needs to determine if the carrier
is positioned to read the left or center strips. In the case of the right strip, no index hole is used and the
control circuitry is programmed so that if light is not sensed in either of the two other programmed index
hole locations (the light continues to be blocked) the carrier is positioned to read the right strip.
The position of the carrier relative to each well in a microtiter strip is tracked by counting pulses obtained
from the pulley which drives the threaded shaft which moves the shuttle (and carrier) along the guide
tracks. When the front edge of the carrier is sensed, a reference position is established.
10 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
300 Series typical memory map (see jumper settings table)0000:7FFF EPROM8000:9FF7 NVRAM9FF8:9FFF real-time clock
300 Series I/O map (fixed)00:03 8255 PPI U5
00 Port A01 Port B02 Port C03 Control Word
10:13 8254 Timer/Counter U410 Timer 011 Timer 112 Timer 213 Control Word
20:2F Alphanumeric LED Display (each I/O address maps to a character position)
300 Series Jumper Settings
PCB Rev through F:no jumpers
PCB Rev G through K:27128 27256 6116 48T08/DS1243 M E R
x x 1-2 2-3 1-2x x 1-2 2-3 2-3
x x 1-2 1-2 1-2x x 2-3 1-2 2-3
PCB Rev L through U:27128 27256 6116 48T08/DS1243 M E R
x x 1-2 1-2 1-2x x 1-2 1-2 2-3
x x 1-2 2-3 1-2x x 2-3 2-3 2-3
PCB Rev V through AA27128 27256 27512 6116 48T08 M E1 E2 Rx x 1 1-2 2-3 2-3
x x 2 1-2 1-2 1-2x x 3 2-3 1-2 1-2
Rev C 11
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Figure 5. Troubleshooting Flowchart
12 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
TroubleshootingThe flowchart in Figure 5 describes the logical steps to take if the instrument is malfunctioning. It is
only a guide to troubleshooting, and is intended to assist in determining routine problems. It is not
intended to take the place of an experienced technician, nor does it attempt to cover all possible
problems.
In the following section, each subassembly or component group and possible problems and solutions
are outlined.
Power SupplyThe power supply consists of a transformer, two diode bridges, and three regulators. Line voltage
supplied to the transformer is converted to 36 VAC, center tapped, which feeds diode bridge BR2. The
center tap is used as a reference for regulators VR4 and VR5, which provide +15 VDC and -15 VDC
for analog circuitry. Another transformer output supplies 11VAC to diode bridge BR1, which in turn
provides +12 VDC (raw, 11-13 VDC nominal) to the system, and VR3 regulates this down to +5 VDC
for the logic and other low voltage control circuits.
A “dead” instrument may be just a blown fuse. However, a blown fuse may indicate a problem on the
main PCB or transformer such as a shorted bridge rectifier or transformer winding. Also, the voltage
select switch setting must match the line voltage. If there is no input voltage present at J6, check the fuse.
To replace the fuse, disconnect the power cable and open the instrument. Locate the double clip fuse
holder mounted on the inside of the rear panel. Locate the blown fuse and remove it with a fuse puller,
or carefully pry it out with a small screwdriver. Install two (2) 0.5 A 3AG slow-blow fuse. Use only
the recommended fuse. Do not substitute any other rating!
If voltage is present at J6, check for input voltage on VR3, VR4, and VR5. If no voltage is present, the
diode bridge BR1 or BR2 may have failed.
PhotometerThe photometer assembly can be examined as several component systems:
• Lamp• Filter wheel• Photometer• Strip drive mechanism
The lamp is tungsten gas-filled, with integral lens end, rated at 6V 6W. The lamp bracket holds the lamp
in alignment with the rest of the assembly, that is, the lens end of the lamp is centered on the optical path.
The lamp is intended to be maintenance free for the life of the instrument, since the lamp is turned off
automatically after 30 minutes of instrument inactivity. Under heavy usage the lamp remains on for
longer periods and so may require replacement. Also, a physical shock to the instrument could cause
filament breakage.
Rev C 13
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
The lamp must be replaced if output becomes low or the lamp fails completely. Since the light is visiblewithout removing the instrument cover, the case of no lamp output can be observed from outside theinstrument. Sight down the strip track while the instrument is first powered on. To check for low output,select test #186. The instrument prints the wavelength, position, and the detected voltage for each filter.The voltages should be between 2 and 10 volts. If all of the voltages are below 2.0 volts, the lamp isthe likely suspect. As a final check, measure the voltage at the lamp terminals. If the lamp voltage ismuch lower than 5.5 VDC, the regulator circuitry on the main PCB (Q6 and VR1) may be the problem.Otherwise, replace the lamp. Refer to the section “Lamp Replacement”. Never adjust trim pots onthe Main PCB to achieve the proper voltages.
If only one or two filters report low detected voltages, degraded filters or an improperly adjusted lampmay be the cause. To check for an improperly adjusted lamp, refer to Figures 10 and 11, as well as theprocedure “Lamp Replacement”. The optical interference filters contained in the wheel are of metal-deposition type construction and are intended to be maintenance free for the life of the instrument. Someolder instruments incorporated other filter materials whose transmittance can be reduced dramaticallyif subjected to high humidity over very long periods. Refer to the section “Filter Replacement”.
The filter wheel turns on a shoulder screw shaft and is driven by the filter wheel motor via a neoprenebelt. The filter wheel itself requires no adjustments. Indexing holes in the filter wheel generate pulsesvia LED/phototransistor pairs, which synchronize photometric readings and provide feedback to themicroprocessor of the filter wheel speed. The photometer contains a photodiode, a high-gain op ampcircuit (U16) and phototransistors Q7 and Q8. Because the photodetector and related circuitry is solid-state, it should require no service. Avoid unnecessary handling of the photometer PCB or removal of
the optical cover.
Strip DriveThe strip drive mechanism is a 1/4-20 lead screw driven by a 12 VDC motor via a belt and pulley.Indexing holes in the pulley serve to generate tach pulses via an LED/phototransistor pair, which aremonitored by the microprocessor. To establish a reference position, the strip holder is moved forwardalong the strip track until the motor stalls. As the strip holder is moved back along the strip track, pulsesare counted to derive the position relative to the stall point. Any problem with the strip mechanism isimmediately apparent as a lack of strip movement, or noisy or erratic movement. A total lack of stripmovement could be the tach generator, motor driver U10 failure, or motor failure. Check the opticalswitch pair (LED/phototransistor) mounted next to the pulley. You should see pulses at CLK2 (U4.18)when the motor is running. Ensure that the LED and phototransistor are aligned with the holes in thepulley. Check for voltage at the motor terminals and at driver U10.
For older instruments, lubrication or adjustment of the lead screw is indicated for noisy or erraticmovement. This should only be necessary at 3 to 5 year intervals, depending on service conditions.Lubricate only the o-rings and shoulder washers at each end of the lead screw. Do not lubricate the stripholder shuttle or the lead screw threads.
If strip position errors are chronic, lead screw end play may require adjustment. Grasp the pulley andmove the lead screw along its axis. End play should be less than 1mm. To adjust end play, the pulleyposition along the drive screw must be adjusted. The drive pulley will be retained by a thumb nut orset screw depending on when it was manufactured. Loosen the nylon thumb nut or loosen the set screw(access via the pulley groove) holding the pulley. Turn the pulley on the lead screw to remove end-play. Do not tighten the pulley against the o-ring spacer. Hold the pulley and tighten the thumb nut orset screw. Recheck end play and repeat if needed.
14 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Main PCBUnder normal circumstances, there are no adjustments to be made to the main PCB. Circuit failures
are highly unlikely, but if they occur, it is recommended that the repairs be performed by factory
authorized technicians. There are ten test points on J8 which provide access to a number of vital signals.
The instrument can be observed in operation with an oscilloscope. Typical test point waveforms are
shown in Figure 6. Note that Figure 6 shows a 6 position filter wheel. The 4 position filter wheel is
very similar.
Internal PrinterThe internal printer is a 20 column thermal type device which prints ASCII characters using a 5 x 7
dot matrix. A printer PCB, incorporating a dedicated microcontroller, manages the print head motor
and the head drivers. There are no adjustments, and service is limited to replacement of the entire printer
mechanism. A printer jam can occur if the printer paper tears and becomes lodged in the printer
mechanism, thus preventing the print head from moving freely. Refer to the section “Printer Jams”.
DisplayThe display should be clearly legible with no missing or dim characters or segments. There are no
adjustments, and service is limited to replacement.
KeypadThe keypad is a sealed membrane switch layer which is serviceable only by replacement. You can test
the keypad by pressing AUX, then pressing all keys except CLEARandENTER. Note that each
key causes the instrument to beep and display a character. Press CLEARand the display clears. Press
ENTER to end the test.
Serial PortThe serial port is intended for use with an external serial printer. Early models do not support a serial
printer. By using a special cable, a serial printer or other serial device may be connected. This is a male
9 pin DB-style connector using RS-232 signals in a nonstandard pin-out. The data format is 2400 baud,
8 bits, 1 stop bit, and no parity.
Serial port pin-out
1 GND
2 TX
3 RX/DTR
4-9 NC
Rev C 15
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Figure 6. Oscilloscope Waveforms
16 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
NOTE
The mechanism for the Strip Reader with Three-Position Carrier will
look somewhat different. The motor which drives the Photometer
wheel will also be on the bottom of the mechanism.
Figure 7. Strip drive mechanism (bottom)
Photometer cover
Pulley
Strip drive
motor
Lead screw
O-ring
spacer
Cap nut
Thumb
nut
Optical
switch
Rev C 17
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Error MessagesError messages are displayed when the instrument fails to operate correctly. They are intended to help
the operator locate the problem. If error messages appear frequently, a hardware problem is usually
indicated.
The following error messages indicate possible interface or component problems.
LAMP OUTPUT IS LOW!! Low voltages were detected for all filter positions., The instrument
assumes this to be a result of low lamp output. This may be due to a
failing or misaligned lamp, degraded interference filters, or voltage
regulator VR1 failure.
LAMP FAILURE The lamp does not appear to illuminated at all. Very low voltages
were detected for all filter positions. See the section “Photometer”
under “Troubleshooting”.
PRINTER JAM The internal printer paper path is obstructed. Clear the paper path. If
the paper path is clear, Check the 14 pin DIP cable connecting the
printer PCB to the main PCB.
FILTER WHEEL STOPPED The instrument cannot correctly detect pulses from the filter wheel.
Check the photometer LED board and the phototransistors Q7 and
Q8 on the photometer PCB. See the section “Photometer” under
“Troubleshooting”.
MECHANISM ERROR The instrument was unable to correctly position the strip for reading.
This may result from a stalled drive motor, lead screw improperly
adjusted, failed driver U10, or motor failure.
POSITIONING STRIP The instrument is attempting to position the strip for reading. If this
message is displayed continuously, the mechanism may have jammed.
See “MECHANISM ERROR”.
The following error messages indicate possible loss of battery power in the NV RAM U3, which can
be checked by substitution and power cycling.
MEM ERR-TEST ENDED The checksum failed when a stored test was recalled. The test was
canceled.
FILTERS RESET Programmed filter wavelengths for key 5 and key 6 were found to be
corrupted. Run test # 248 to reprogram the filter wavelengths. Refer
to filter label and Figure 13. Filter wheel position #3 corresponds to
key 5 and filter wheel position #2 corresponds to key 6.
DO ABS SET TEST 212! The stored absorbance calibration values were corrupted or not
found. Restore the calibration data. See the section “Calibration”.
18 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Service Procedures
Opening the InstrumentThe cover must be removed to allow access to the inside of the instrument. Disconnect the power cable
and the serial cable (if connected) from the rear panel. Invert the instrument on a soft nonabrasive
padded surface such as a terrycloth towel, to prevent scratching. Refer to Figure 3, Chassis Bottom.
Locate and remove the (5) cover screws from the bottom of the instrument. Do not remove or loosen
any other screws. While holding the instrument cover in place with both hands, return the instrument
to the upright position.
On some models, the printer mechanism is mounted to the main PCB and the printer paper roll must
be removed to allow cover removal. Press inward on the sides of the paper roll cover and lift it off. Lift
out the paper roll and tear off the paper where it enters the printer mechanism.
Grasp the cover at the left side. Gently lift the cover upward and to the right, until the cabling is visible.
You can rest the cover on its right-hand side while you disconnect the cables. Refer to Figure 8, Cover
Connections. Disconnect the display cable at the left front of the main PCB assembly. Place a small
slot blade screwdriver between the 14 pin DIP plug and its socket and pry gently upwards. Do not bend
the pins. Disconnect the printer cable in the same way. Disconnect the keypad cable from its connector.
Remove the cover and set it aside.
To reinstall the cover, reverse the procedure. Position the cover on its side to the right of the chassis base.
Connect the printer, display, and keyboard cables. Carefully fit the cover to the chassis base. Install
the cover screws. Do not overtighten the cover screws.
WARNING
Hazardous line voltages are present on the rear panel.
Always disconnect the AC power cable before opening the
instrument.
Rev C 19
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Figure 8. Cover Connections
Main PCB
Red stripe
To display
To printer
To keypad
Red stripe
Remove and install
DIP plugs vertically.
Do not bend pins.
Remove and install
keypad cable verti-
cally. Do not twist,
kink or fold the key-
pad cable.
20 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
EPROM Replacement
The instrument software can be upgraded to the latest version by replacing the EPROM. On most
instruments, the EPROM is accessible through a hatch on the underside of the chassis, and can be
changed without removing the cover. On older instruments, the EPROM is accessible only by
removing the cover.
1. Refer to Figure 3. Locate the EPROM hatch on the underside of the instrument. Place the
instrument upside down on a soft surface. Remove one hatch screw and loosen the other.
Swing the hatch out of the way. If the instrument has no EPROM hatch, remove the
instrument cover.
2. Refer to Figure 9. Locate the EPROM socket and lift the small lever at the end of the socket.
This frees the EPROM for easy removal and insertion. If the EPROM socket has no such
lever, pry out the EPROM with a small screwdriver or other tool. Do not bend the EPROM
pins.
3. Gently bend the pins on the new EPROM as shown below, so that the pins are perpendicular
to the EPROM. Hold the EPROM by the body and bend the pins against a flat surface such
as a table to bend all the pins the same amount. Do not over bend the pins, since they are easy
to break if bent too far.
4. Install the new EPROM with the notch or dot (indicating pin 1) oriented toward the front of
the instrument. Press the lever down toward the PCB until it locks into place.
5. Replace the hatch screw removed in step 1, or replace the instrument cover if removed.
6. Power up the instrument and verify that the instrument operates with the new software.
Rev C 21
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Figure 9. EPROM Replacement
Lever (also on underside of PCB)
EPROM
socketed to under-
side of PCB, acces-
sible via hatch on
chassis bottom
EPROM accessible
by removing
instrument cover
Lever
22 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Lamp ReplacementThe lamp should be replaced only if it fails to light, or several filter voltages are reported as low.
Materials and equipment:• Replacement lamp• Phillips screw driver• Flat blade screw driver
Procedure:
1. Set the power switch to OFF (O). Open the instrument. Refer to Figure 4A or 4B (ChassisLayout) as appropriate for the instrument. Locate the strip mechanism and the lamp bracket.
CAUTION
Lamp is HOT. Allow the lamp to cool before handling.
3. Refer to Figure 10, Lamp Replacement. Remove the Lamp Bracket Mounting Screws (onlyif necessary to access the lamp!) and lift the Lamp Bracket away from the Guide Bracket.Loosen but do not remove the lamp terminal screws. Do not loosen or remove any otherscrews. Remove the lamp from the connector.
4. Use a pair of pliers or tweezers to handle the new lamp. Avoid handling with bare skin, sincethe oil from your skin can reduce lamp life. Insert the lamp leads into the connector until theybottom out. Re-screw the Lamp Bracket to the Guide Bracket if removal was necessary. Thelamp lens end must be centered on the aperture and the lamp body must be perpendicular tothe horizontal. While holding the lamp in alignment, tighten the lamp terminal screws.
5. Set the power switch to ON. Shield your eyes from the lamp beam and observe the projectionof the beam onto the aperture. Refer to Figure 11, Spot Alignment. The spot should be smalland centered on the aperture. The spot should be sufficiently large to just encircle the aperture.If the spot is not centered, repeat step 4.
6. Select test #186. The instrument prints the detected voltage for each filter position. Allvoltages should be between 2.00 volts and 10.00 volts. If all the voltages report low, repeatstep 4 until optimum lamp position is obtained. Never adjust trim pots on the Main PCBto achieve the proper voltages.
Rev C 23
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Figure 10. Lamp Replacement
Figure 11. Spot Alignment
Mounting
Screws
Lamp
bracket
Lamp
LAMP MOUNTING CONFIGURATIONS
Lamp
terminal
screws
Aperture
Guide
bracket
Lamp
connector
Lamp must be perpendicular to guide bracket and
projected spot must be centered on aperture.
Spot
Aperture
Projected spot must be bright,
well-focused, and centered on
aperture.
24 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Filter ReplacementMaterials and equipment:
• Replacement filter set• Phillips screw driver• Flat blade screw driver
Procedure:1. Open the instrument. Refer to Figure 4A or 4B as appropriate for the instrument. Locate the
Strip Mechanism and the ribbon cables which run from the Main PCB to the Strip Mechanism.Unplug both ribbon cables from the main PCB. On units using Main PCB rev Z and subsiquentrevisions, a third set of wires (two yellow wires connected to J17) need to be disconnected.
2. Use a pencil to mark the front and side edges of the photometer on the chassis so that thephotometer can be installed in its original position.
3. Refer to Figure 3. Locate the (2) #6 screws holding the Strip Mechanism assembly to thechassis. From the bottom of the instrument, remove these screws and washers while notingthe screw lengths and locations from which they were removed. Remove the StripMechanism.
4. Orient the Strip Mechanism so that you are looking at the bottom of the assembly. Refer toFigure 7, Strip drive mechanism, and Figure 12, Filter Replacement. Remove the four 6-32screws retaining the photometer cover. Set the photometer cover aside.
5. Remove the two 4-40 screws securing the photometer PCB. Gently move the PCB asidewithout disconnecting any wires.
6. Remove the belt from the filter wheel and pulley. Loosen the shaft (shoulder screw). Removethe shaft and filter wheel. Note the number of nylon washers and the order of installation.
WARNING
Retaining Rings (used on later production filter wheels) willpop out from filter wheel when its spring tension is released
during removal. Shield the opening with hand and wearsafety glasses.
7. Refer to Figure 13, Filter Wheel Positions. Locate the filter(s) to be removed. There arecurrently two means of filter retention in field use; silicone glue or a Retaining Ring. Referto Figure 12 and determine if a Retaining Ring is used. If a Retaining Ring is used, carefullypry two to three of the retaining ring teeth away from the filter until the ring is dislodged andremove the ring and filter. If silicone glue is present, remove the silicone and push the filterout of the wheel using the eraser end of a pencil or other soft blunt object and the glue shouldrelease. Remove any remaining filter components from that position in the wheel.
8. Locate the neutral density filters, dot screen, and transmittance filters that were included withthe replacement filter and drop them first into the cavity. Install the interference filter with themirror side down. Place the replacement retainer over the filter so that the tabs are angled awayfrom the filter and press in place with a 7/16 diameter wooden dowel or similar object.
9. Install the filter wheel taking care not to pinch washers under the shaft. Reassemble the shaftand washers as disassembled. Tighten the shaft securely. The filter wheel should spin freely.
10. Install the filter wheel belt. Be sure that the belt is centered on the filter wheel and is not twisted.
11. Install the photometer PCB taking care to center the board around the shaft.
12. Replace the photometer cover, taking care to position the cable in the slot of the cover. Donot pinch the cable under the cover. Tighten cover screws only until snug. Do not overtightenthe photometer cover screws! Doing so will bend the photometer cover and create lightleakage.
Rev C 25
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
13. Install the Strip Mechanism to the chassis by replacing the (2) #6 screws removed in step 3through the chassis base in their original locations. Line up Strip Mechanism with locatingmarks made in step 2. Connect the cables removed in step 1.
14. Attach the power cord and connect the instrument to the mains supply. Set the power switchto ON (1). Select test #186. The instrument prints the detected voltage for each filters. Allvoltages should now be between 2 and 10 volts. If not, check lamp alignment. If the lampalignment is properly set, the neutral density filters may need to be changed. Add neutralsto lower the voltage. Remove neutrals to increase the voltage. Never adjust trim pots on theMain PCB to achieve the proper voltages.
Figure 12. Filter Replacement
Remove Screws holding
photometer cover.Remove screws holding
photometer PCB.
Filter wheel Belt
Pulley
Shaft
Filter
Retainer
Remove filter wheel
belt and filter wheel. Remove adhesive or
retainer and push
filter from wheel.
Install neutral density filters,
dot screens, interference filters
and retainers.
One nylon
washer
One or two nylon washers
as disassembled.
Assemble washers and filter
wheel to filter wheel shaft.
26 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Filter Label
The filter label located on the photometer cover describes the specific filter wheel configuration for your
particular instrument. Refer to Figure 14.
Filter wheel position (POS) is the physical placement of the filter on the wheel and does not correspond
directly to the key position. The filter wheel position numbers are shown in Figure 13. Note the position
of the Index hole and the Home hole. When running test # 248, note that filter wheel position #3
corresponds to key 5 and filter wheel position #2 corresponds to key 6.
Each filter position can have several filter elements installed. Configuration (WAVELENGTH
column) describes the various screens and filters in that position . Each element is separated by a slash
(/). The first 3 digit value in this column is always the wavelength of the interference filter, in
nanometers. The other values indicate either dot screens, neutral density filters, or transmittance filters.
A “D” suffix indicates a dot screen and the first two digits are the percent blockage. A 2 digit value
followed by a percent sign signifies a dot screen. A 3 digit value (other than the first value in the line)
indicates one or more transmittance filters. The first digit is the quantity; the last two digits are the
percent transmittance. A 2 digit value separated by a decimal point indicates a neutral density filter. The
value is the absolute absorbance of the filter. “BLOCKED” indicates an opaque disk is installed in
that position.
LOT# is the interference filter manufacturers' production code lot number or date. In the new style,
the percent transmittance for the lot is included in this column.
The production date (DATE) is the date the filter wheel was assembled.
Figure 13. Filter Wheel Positions
Home
hole
Index
hole
Home
holeIndex
hole
Rotation
6
5
4
1
2
3
1
23
4
4 Position 6 Position
Rotation
Rev C 27
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Filter wheel
position
Configuration Production
date
Filter date or
lot number
Old style
POS WAVELENGTH LOT# DATE
1 405 3YGF 58.8% 5/23/94
2 545/90D/115/170 4GDB 80.6% 5/23/94
3 600/90D/115/170 3ZCJ 65% 5/23/94
4 492/215 3WEF 74.4% 5/23/94
5 630/215/15D 4FLR 67.6% 5/23/94
6 450/115/90D 4HDL 83% 5/23/94
New style
Figure 14. Filter Labels
POS WAVELENGTH LOT# DATE
1 405 2VCT 11/6/92
2 545/2.0/30% 6/30/92 11/6/92
3 600/2.0 8/10/92 11/6/92
4 492/1.0/70% 8/10/92 11/6/92
5 630/1.0/30% 8/10/92 11/6/92
6 450/80% 7/21/92 11/6/9
28 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Printer Jams
CAUTION
Do not apply force to the print head.
Do not scratch the platen or pry with sharp objects.
If the instrument displays “PRINTER JAMMED”, or if the print head does not return to the home
position, check for obstructions in the print head path.. Refer to Figure 15. Carefully remove any paper
or debris with a pair of tweezers. Turn off the power switch, wait 5 seconds, then turn on the power
switch. The print head should move to the right side and then return to home position at the left.
An alternate method of removing a particularly stubborn wad of paper is to move the print head
manually. To do this, open the instrument. Remove the two nuts holding the printer bracket to the
cover. View the front of the printer mechanism. Locate the small gear on the right side which turns the
grooved steel print head shaft. Turn the gear so that the print head moves away from paper wad and
the paper can be easily removed.
Figure 15. Clearing printer jams
Right front gear
moves print headPrint head shaft
Print head is jammed
against wad of paperHome
position
Print head
Rev C 29
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Printer Repairs
A graphics capable printer has been introduced in recent instruments. Replacement printers are
graphics capable and can be made compatible with earlier non graphics capable printer PCBs. One way
to identify which type of printer is used is to remove the ribbon cable from the keytail socket on the
printer interface PCB. Look at the connector on the ribbon cable. An earlier non graphics capable
printer will have eight conductors. The graphics capable printer will have nine. Another way to identify
which printer is used is to look at the blue inspection label on the printer interface PCB. If the number
on the label is M020445 or higher, a graphics capable printer is used. Repairs to printer mechanisms
can consist of print head replacement or printer mechanism replacement.
Print Head Replacement Replacement Print Heads are supplied for the graphics capable printer (nine conductor ribbon cable).
Determine the size of the keytail socket on the Printer Interface PCB into which the ribbon cable from
the Print Head being replaced is inserted (eight or nine slots). If there are eight slots, modify the
replacement Print Head as shown in Figure 16.
If ribbon is too wide for socket,
orient as shown and trim through
center of pad as shown.
Wide trace
Figure 16. Replacement Print Head Modification for
Non Graphics Capable Printers
Refer to Figure 17 and replace the Print Head as follows:
1. Pry out the pin (1) which holds the gear (2) in place.
2. Remove the pin (3) from the side of the print head with tweezers and pry the pin (4) from thetop of the Print Head. The Print Head should now be free to move along the Drive Screw (6).
3. Remove the snap ring (5) from the end of the drive screw (6) and withdraw the drive screwand gear assembly from the printer mechanism in the direction shown.
4. Press the end of the slider rod (7) out of the hole in the printer mechanism in the directionshown from the left side of the mechanism. It is not necessary to remove the slider rodcompletely from the mechanism.
5. Slide the Print Head off the slider rod and replace with the new Print Head. Reassemble themechanism in reverse order of the above disassembly steps.
30 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Figure 17. Printer Views for Print Head Replacement
1
2
3
4
5
6
7
7
6
LEFT SIDE
RIGHT SIDE
Rev C 31
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Printer Platen Replacement
1. Referring to Figure 18, remove the Paper Dress Cover.
2. Move the Print Head to the right-most position as shown by rotating the Screw Drive with itsdrive gear on the right side of the mechanism.
3. Lift out the left side of the platen.
4. Install the new platen so that the end which has the longer pin is oriented toward the right sideof the mechanism. Insert the right end of the platen under the print head. Use a pen point orsimilar blunt instrument to push metal spring slightly forward, then slip the left side of platenin place.
Figure 18. Printer Platen Replacement
Platen
(darkened area)
LEFT
SIDE
RIGHT
SIDE
Screw Drive
Dress Cover
Print Head
32 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
CalibrationEach instrument is calibrated during manufacturing using standards that are traceable to the National
Institute for Standards and Testing (NIST), and is tested to verify its linearity to 2A. This preset
calibration is very stable, and the instrument should never require re-calibration during normal service
life.
Absolute calibration can be verified with the use of NIST filters, or by comparison to a reference
instrument that is known to be calibrated to NIST filters.
Since most lab test results are based upon standards rather than upon absolute absorbances, the linearity
of the instrument, rather than the accuracy, is the more critical indicator of instrument performance. A
periodic verification of instrument accuracy and linearity is advised. Calibration may be confirmed
using a commercially available photometer calibration check set which can be obtained from your
distributor.
CAUTION
RE-CALIBRATION OF THE INSTRUMENT
SHOULD NOT BE REQUIRED UNDER NORMAL
CIRCUMSTANCES, AND SHOULD NOT BE
CONSIDERED UNTIL ALL POSSIBLE
INTERFERING FACTORS HAVE BEEN RULED
OUT, INCLUDING CHEMISTRY ERROR,
OPERATOR ERROR, LAMP FAILURE, AND
DEGRADED FILTERS.
Standards and blanks should be closely checked for accuracy
and results should be compared with a reference instrument.
Use of a commercially available photometer check set to
verify photometric accuracy and linearity is strongly
recommended.
Restoring the factory calibration data should always be
attempted before resorting to Re-calibration.
Rev C 33
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Software or Hardware Calibration?
Most instruments manufactured after June 1993 rely on software calibration, that is, on
calibration data stored in nonvolatile RAM. The calibration data is used by the instrument to
compensate for variations in electronic components. During factory calibration, any necessary
hardware adjustments are made using precision reference standards and specialized electronic
equipment. The calibration data is stored in the instrument memory and also recorded on a label located
on the bottom of the instrument. This requires no hardware adjustments.
The absorbance calibration data is stored as a scaling factor, which is the ratio of a known reference
absorbance to the instrument's reported absorbance. The value entered at the keypad is multiplied by
the stored value to obtain a new scaling factor. The calibration data may be printed at any time by
selecting test #213. This prints the serial number of the instrument, the date of last calibration, and the
stored absorbance scaling factor.
Calibration data is stored with a checksum that is recalculated and compared each time the instrument
is powered on, a new software mode is selected, or a stored test is recalled. Under certain circumstances,
large AC mains transients or other environmental disturbances may cause the microprocessor to
“crash” and possibly render the calibration data invalid. Failure to recover the calibration data properly
(checksum failure) will be indicated on the display and the internal printer as follows:
DO ABS SET TEST 212!
If this message is printed or displayed, it indicates that the calibration values have been lost. This
message will be printed each time the checksum failure occurs. The instrument also annunciates this
condition with multiple beeps. In this condition, the absorbance scaling factor is set to 1.000. The
maximum absorbance error is then +/-10%, provided the trimpots have not been disturbed.
The calibration data can be restored from the calibration label as described in the section “Restore
Calibration Data”.
All instruments manufactured before June 1993 utilize hardware calibration. This requires
making adjustments to the trimpots for absorbance gain and offset. If you believe your instrument
requires hardware re-calibration, contact your distributor for assistance. Never adjust trim pots on
the Main PCB to achieve the proper voltages.
To determine whether the instrument incorporates software calibration features, select test # 213. If
calibration data is printed, the instrument supports software calibration.
In any case, if the trimpots have been disturbed or components C16, R12, R16, R11, R13, RN8, R24,
U16, or the photodiode have been replaced, the instrument will probably require re-calibration. Contact
your distributor for assistance. Never adjust trim pots on the Main PCB to achieve the proper
voltages.
34 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Restore Calibration DataThis procedure restores the calibration data to the original factory settings recorded on the calibration
data label.
Procedure:
1. Select test #213 to print the calibration data. The instrument prints something similar to this(the actual value may differ):
ABS FACTOR 1.126
2. Select test #212. The display shows:
ABS FACTOR=
If the value printed in step 1 is not exactly 1.000, enter the inverse of the value printed in step1. That is, calculate
1 ABS FACTOR
and enter that value.
3. Locate the calibration data label on the underside of the instrument. Copy the data from thelabel for reference in the following steps.
4. If the date and time are incorrect, set the date and time as described in the Owner's Manual.
5. Select test #212. The display shows:
ABS FACTOR=
Enter the number from the Abs Factor line of the calibration label.
6. Select test #213 to print the calibration data. The printed data should now match the valuesshown on the calibration data label. If it does not, repeat this procedure.
Rev C 35
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
DiagramsBlock Diagram
36 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
AWARENESS
TECHNOLOGY
300-100AB(AR)300MAINPCB995001
PCB LayoutsMain PCB - rev AB
Rev C 37
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Main PCB - from rev Z
38 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Main PCB – rev L to rev Y
Rev C 39
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
TO
PHOTOMETER
TO
LAMP/MOTOR
Main PCB – through rev K
40 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Printer PCB (300-200C)
Rev C 41
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Printer PCB (300-200)
42 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
SchematicsMain PCB (rev AB) 1 of 4
Rev C 43
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Main PCB (rev AB) 2 of 4
44 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Main PCB (rev AB) 3 of 4
Rev C 45
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Main PCB (rev AB) 4 of 4
46 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Main PCB (from rev Z) 1 of 4
Rev C 47
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Main PCB (from rev Z) 2 of 4
48 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Main PCB (from rev Z) 3 of 4
Rev C 49
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Main PCB (from rev Z) 4 of 4
50 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Photometer PCB
Rev C 51
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Printer PCB
52 Rev C
300 Series Service Manual© 1995-1996 Awareness Technology, Inc. Unauthorized duplication is strictly prohibited. Information given in this manual is restricted for
use by authorized personnel and is to be used for the sole purpose of providing routine instrument maintenance and repair services.
Notes: