Rev A MN302000 1 of 38 WebCal™ Configuration Software for: EchoPod® DL10, DL14, DL24, DL34, DS14 & DX10 Series EchoSonic® LU23, LU27, LU28 & LU29 Series Flowline Inc. 10500 Humbolt Street Los Alamitos, CA 90720 Tel: (562) 5983015 Fax: (562) 4318507 www.flowline.com

Upload: unitronics

Post on 16-Jul-2015

136 views

Category:

Technology


3 download

TRANSCRIPT

  • RevA MN302000 1of38

    WebCalConfigurationSoftwarefor:

    EchoPodDL10,DL14,DL24,DL34,DS14&DX10SeriesEchoSonicLU23,LU27,LU28&LU29Series

    FlowlineInc.

    10500HumboltStreetLosAlamitos,CA90720Tel:(562)5983015Fax:(562)4318507www.flowline.com

  • 2of38 MN302000 RevA

    INTRODUCTION/TABLEOFCONTENTS StepOneWebCalPCconfigurationsoftwareisautilityprogramthatallowsuserstoeasilyconfiguretheirEchoPodorEchoSonicultrasoniclevelsensor.Developyourconfigurationusingpreprogrammedfunctionmenusasthetankgraphicandsetpointfieldsautomaticallychangetomatchyourconfiguration.TheninputyourlevelsetpointvaluesandclicktheWritetoUnitbutton.Yourconfigurationwillbeuploadedintothesensorinjustafewseconds. Last,clicktheWiringDiagrambuttontoopenawiringschematicofyourconfiguration inaPDFformat.Printthedocument,disconnectthesensorandwireitpertheschematic.Itsthatsimple!NewFeatures

    Volumetrictankshapelinearization SampleProgramforallWebCalconfigurablesensors ViewCurvetoolenablingtheviewingofnonlinearoutputs

    TableofContentsSystemRequirements:...........................................................................................................................................3SafetyPrecautions:................................................................................................................................................4GettingStarted:......................................................................................................................................................5 USBFobInterface:...................................................................................................................................6WebCal: .........................................................................................................................................................7 RelayConfigurationDL14,DL24,DL34,DS14Series:..............................................................................8 RelayConfigurationDL10,LU23,LU27,LU28&LU29Series:...............................................................15 RelayConfigurationDX10Series:..........................................................................................................17 TankShapeSelection:..............................................................................................................................20 Example:DimensionalEntryVerticalCylindrical:.....................................................................21 Example:DimensionalEntryHorizontalCylindrical:..................................................................23 TankLevels:..............................................................................................................................................25 WritetoUnit:...........................................................................................................................................26 Wiring:.......................................................................................................................................................27 WiringDiagram:...........................................................................................................................27 WiringtoEchoPodandEchoSonic:..........................................................................................27 AdvancedFeatures:.................................................................................................................................28Appendix: .......................................................................................................................................................30 SampleProgram:......................................................................................................................................30 UpdatingWebCalSoftware:..................................................................................................................31 UpdatingSensorFirmware:.....................................................................................................................32 DemoPage:..............................................................................................................................................33 StrappingTable:.......................................................................................................................................34 Linearvs.NonLinear:..............................................................................................................................35 Example#1VolumeOutput:.....................................................................................................35 Example#2CurrentOutput:.....................................................................................................36Troubleshooting:..................................................................................................................................................37Warranty,ReturnsandLimitations:.....................................................................................................................38

  • RevA MN302000 3of38

    SYSTEMREQUIREMENTS StepTwo ComputerSystemRequirementsforWebCal

    o Windows2000,XP,Vista,7,8o 32or64bitsystemo 1USB2.0porto 10mBharddrivespaceo 256mBRAMo Internetconnection

    SensorRequirementsEchoPodSeriesSensorDL10Series DL1000,DL1001,DL1010&DL1011DL14Series DL1400,DL1401,DL1410&DL1411DL24Series DL2400,DL2401,DL2410&DL2411DL34Series DL3400,DL3401,DL3410&DL3411DS14Series DS1400,DS1401,DS1410&DS1411DX10Series DX1000,DX1001,DX1010&DX1011

    EchoSonicSeriesSensorLU23Series LU2300,LU2301,LU2310,LU2311,LU2340,LU2341,LU2350&LU2351LU27Series LU2700,LU2701,LU2710,LU2711,LU2740,LU2741,LU2750&LU2751LU28Series LU2800,LU2801,LU2810,LU2811,LU2840,LU2841,LU2850,LU2851,LU2860,LU2861,LU2870&LU2871LU29Series LU2900,LU2901,LU2910,LU2911,LU2940,LU2941,LU2950,LU2951,LU2960,LU2961,LU2890&LU2971

    InterfacingRequirementso KeyFob,PartNumberLI991001

    Dependingonthesensorpartnumberordered,aKeyFobmayneedtobeorderedinadditiontothesensor.

    AKeyFob(LI991001)isrequiredtointerfacethesensorwiththecomputer.

  • 4of38 MN302000 RevA

    SAFETYPRECAUTIONS StepThree AboutthisManual:PLEASEREADTHEENTIREMANUALPRIORTOINSTALLINGASENSORORUSINGTHIS

    PRODUCT.ThismanualincludesinformationontheWebCalconfigurationsoftwarefromFLOWLINE.Pleaserefer to the sensor part number located on the sensor label to verify the exactmodel configuration andcompatibilitytotheWebCalsoftware.

    UsersResponsibilityforSafety:FLOWLINEmanufacturesabroadrangeoflevelsensingtechnologies.Whileeachofthesesensors isdesignedtooperate inawidevarietyofapplications, it istheusersresponsibilitytoselectasensormodel that isappropriate for theapplication, install itproperly,perform testsof the installedsystem,andmaintainallcomponents.Thefailuretodosocouldresultinpropertydamageorseriousinjury.

    Safety Installationshouldbedonebyproperlytrainedstaff Supplyvoltageshouldneverexceedamaximumof28VDC Makesurethesensorischemicallycompatiblewithyourapplication Designafailsafesystemthataccommodatesthepossibilityofsensorand/orpowerfailure Thissensorshouldnotbeusedinclassifiedhazardousenvironments

    MakeaFailSafeSystem: Designa failsafesystem thataccommodates thepossibilityofsensorand/orpower failure. FLOWLINErecommends theuseofredundantbackupsystemsandalarms inaddition to theprimarysystem.

    Flammable,ExplosiveorHazardousApplications:WebCalshouldnotbeusedwithinclassifiedhazardousenvironments.

  • RevA MN302000 5of38

    GETTINGSTARTED StepFourEchoPod and EchoSonic are configured throughWebCal, a PC software program. WebCal is a freedownloadfromFlowlineswebsite. YoumustdownloadandinstallWebCalpriortopluggingintheUSBFob.Pleasegotowww.flowline.com/webcal.php,andselectyourlanguageversion.

    WebCalSystemRequirementsWindows2000,XP,Vista,7,8

    32or64bitsystem1USB2.0port

    10mBharddrivespace256mBRAM

    Internetconnection

  • 6of38 MN302000 RevA

    GETTINGSTARTED(continued) StepFourUSBFobInterface:EchoPodandEchoSoniccommunicatewithWebCalthroughaUSBinterfacecalledaFob.Note:BeforepluggingyourFobintoyourcomputersUSBport,besurethatyouhaveinstalledWebCalonyourcomputer. Connectthered,green,whiteandblackwiresfromsensor intothecorrectterminalsontheFob.TightenthescrewsontheterminalsandplugyourFobintotheUSBportofyourcomputer.

    WiringisidenticalforallEchoPodandEchoSonicseriesUseonlytheRed,Black,GreenandWhitewires. Themaximumcabledistancebetweenthecomputerandsensoris15.Thisonlyapplieswhenconfiguring

    theEchoPodorEchoSonic. Once the sensor is configured and prior to installation, isolate thewhite and greenwires from active

    powertopreventashortoftheconfigurationcircuit. Note:WhenusingtheFob,donotaddVDCorVACpower.TheFob,whenconnectedtothecomputer,will

    providetherequiredpowertotheEchoPodorEchoSonic.

  • RevA MN302000 7of38

    WEBCAL StepFiveWithEchoPodorEchoSonicconnected toyourcomputer;open theWebCalsoftwarebyclickingon theWebCal icon. Followthestepsbelowtoconfigurethesensor. ClickHelp inthe lowerrighthandcornerandopen thehelpmenuofWebCal for instructionsonWebCal. IfyouneedadditionalassistanceusingWebCal,pleasecontactaFlowlineapplicationsengineerat(562)5983015.

    TabsforConfiguration,

    UpdatesandDemomode

    ModelSelector(SampleProgramOnly)

    ConfigurationSelector

    SensorDetails

    LevelConfiguration

    ActionButtons

    TankSelector

    AbovescreencaptureisfortheEchoPodDL24series.

    ConfiguringEchoPodorEchoSonicwithWebCal

    1. RelayConfigurationa. Configurestherelaysintermsofpump/valveoperationsaswellashighorlowalarms.b. Setsfailsafefortherelaysandthesensorsoutput(current,voltageorfrequency).

    2. TankShapeSelectiona. Definestheshapeofthetankaswellasthedimensionalinformationforthetankwithrespect

    tothesensorslocationonthetank.3. TankLevelConfiguration

    a. Entersthesettingsfortherelayactivationpointsaswellasconfirmstheoperationalrange.4. WritetoUnit

    a. Uploadstheconfigurationintothesensor.b. Providesacustomwiringdiagramspecifictothesignaloutputand/orrelayconfiguration.

  • 8of38 MN302000 RevA

    WEBCAL(RELAYCONFIGURATION)DL14,DL24,DL24,DL34&DS14SeriesOnly StepFiveThissectionofWebCaliswhereyouselectthelevelconfigurationsettings.Startfromthetopandworktothe bottom, choosing the selections that are applicable to your configuration. Not Applicable willautomaticallyshowwhenaselectiondoesntapplytoyourconfigurationsettings,andyoumaymoveon.AllconfigurationsettingsmustbeselectedorhaveNotApplicablebeforeyoucancontinuetothenextstep.Note:PressingtheClearScreenbuttonwillresettheconfigurationtableandallowaccesstoallfeatures.

    NumberofPumpsThisfeatureallowsyoutoselectthenumberofpumpsorvalvesusedwithEchoPod.Thissettingactivatesthecontrolcapabilitiesofoneortworelays.Controlrelaysareoftenreferredtoaslatchingrelays.

    o Switch/AlarmsOnlyThe relayswillbe standard singlepointnonlatchingrelays.Usethissettingforhighand/orlowalarms.

    o 1Pump/Valve One relaywill be configured as a control orlatchingrelay (relaywillhaveastart levelandaseparatestoplevel). Use this setting to control one pump or valve forautomaticfillingoremptyingofatank.

    o 2Pumps/Valves Two relays are configured as control orlatchingrelays. Eachrelaywillhaveauniquestart levelandacommon stop level. Use this setting tocontrol twopumpsorvalvesforautomaticfillingoremptyingofatank.

    o 420mATransmitterOnlyThissettingwilldisengagealloftherelays. Use this function if you are not using any relays andusingonlythe420mAcurrentoutput.

    Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton.

  • RevA MN302000 9of38

    WEBCAL(RELAYCONFIGURATION)DL14,DL24,DL24,DL34&DS14SeriesOnly StepFive

    Pump/ValveActionThisfeatureallowsyoutoselectwhetherthepumpsorvalveswillbeusedtoautomaticallyfilloremptythetank.For2Pump/Valvemode,bothdevicesmustbeused inthesame(automaticfillorempty)way.Youcannotsetonerelayforfillandtheotherforempty.

    o EmptiesTankSetstherelay(s)toautomaticallyemptyatank.ThestartlevelwillbeabovetheStoplevelforeachrelay.

    o Fills Tank Sets the relay(s) to automatically fill a tank. ThestartlevelwillbebelowtheStoplevelforeachrelay.

    o NotApplicable Appearswhen this function is not availablebasedonpreviousselections.

    EmptiesTank FillsTank (AutoEmpty) (AutoFill)

    Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton.

  • 10of38 MN302000 RevA

    WEBCAL(RELAYCONFIGURATION)DL14,DL24,DL24,DL34&DS14SeriesOnly StepFive

    Pump/ValveModeThisfeatureallowsyoutoselectthecontrolmodeforalatchingrelay.Pump/Valve mode is not active for Switch/Alarms Only or 420 mATransmitterOnly.

    o Simplex Allows the relay to be used for automatic fill orempty.Thisisthedefaultandonlymodewhen1Pump/Valveisselected.

    o SimplexusedtoEmptyTank

    o Lead/LagAllowstworelaystohaveuniquestart levelsanda

    commonstoplevel.Thefirstrelaywillbeidentifiedastheleadrelayandthesecondrelayasthelag.Eachtimetheleadlevelisreached,thefirstrelaywillalwaysstart.Thelagrelaywillonlystartwhen the lag level is reached. All relayswill stopat thecommonofflevel.

    o Lead/LagusedtoEmptyTank

    o Duplex Allows two relays tohave twodifferent start levels,a

    common stop levelandwillalternate the relayswhen the firststartlevelisreached.Thetworelayswillalternateeachtimetheleadlevelisreachedandtheremainingrelaywillstartwhenthelaglevelisreached.Allrelayswillstopatthecommonofflevel.

    o DuplexusedtoEmptyTank

    o NotApplicable Appearswhen this function is not available

    basedonpreviousselections.Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton.

  • RevA MN302000 11of38

    WEBCAL(RELAYCONFIGURATION)DL14,DL24,DL24,DL34&DS14SeriesOnly StepFive

    RelayFailSafeThisfeatureallowsyoutoselectthefailsafemodefortherelaysintheevent that the sensor looses echo confidence. When the sensorregains echo confidence, the output current will revert back to thecurrentlevelcondition.

    o RelaysOffTherelayswillreverttotheOFFstate.ThisappearswhenSwitch/AlarmsOnlyisselected.

    o RelaysOnTherelayswillreverttotheONstate.ThisappearswhenSwitch/AlarmsOnlyisselected.

    o HoldState The relay(s)will remain in the same stateas thelast confident echo detected. When the sensor regains echoconfidence,therelayswillreverttothecurrentlevel.

    o Pump/ValvesOffTherelayswillreverttotheOFFstate.Thisappearswhen1Pump/Valveor2Pumps/Valvesareselected.

    o Pump/ValvesOn The relayswill revert to theON state.Thisappearswhen1Pump/Valveor2Pumps/Valvesareselected.

    o NotApplicable Appearswhen this function is not availablebasedonpreviousselections.

    Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton.

  • 12of38 MN302000 RevA

    WEBCAL(RELAYCONFIGURATION)DL14,DL24,DL24,DL34&DS14SeriesOnly StepFive

    Switch/AlarmConfigurationThis featureallowsyou to select the relayoperation for the switch /alarm(usedasahighorlowalarm).Thenumberofavailablerelaysisbasedupontheprevioussettings.

    o NoAlarmTurnsOFFallremainingrelays.o HighAlarmsSets1to4HighAlarms(1High,2High,3High,

    4High).o LowAlarms Set1to4LowAlarms(1Low,2Low,3Low,4

    Low).o Combination Alarms Sets a combination of High and Low

    Alarms (1Low 1High, 1Low 2High, 2Low 1High, 2Low 2High,1Low3High,3Low1High).

    o NotApplicable Appearswhen this function is not availablebasedonpreviousselections.

    HighAlarm LowAlarms CombinationAlarms (4High) (4Low) (2Highand2Low)Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton.

  • RevA MN302000 13of38

    WEBCAL(RELAYCONFIGURATION)DL14,DL24,DL24,DL34&DS14SeriesOnly StepFive

    SwitchHysteresis/DeadbandThis feature allows you to select a hysteresis or dead band for theremaininghighand/orlowalarms.

    o Options forHysteresis/DeadbandNoHysteresis,,,1,2,cm,1cm,2cm,5cmorNotApplicable.

    o HighAlarmsRelayactivatesabove the setpoint. Relaywilldeactivatewhen the level goes below the set point plus thevalueofthehysteresis.

    o LowAlarmsRelay activatesbelow the setpoint. Relaywilldeactivatewhen the level goes above the set point plus thevalueofthehysteresis.

    HighAlarmw/Hysteresis LowAlarmw/HysteresisNote:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton

  • 14of38 MN302000 RevA

    WEBCAL(LOOPFAILSAFE)DL14,DL24,DL24,DL34&DS14SeriesOnly StepFive

    LoopFailSafe(NotAvailableonDS14series)Thisfeatureallowsyoutoselectthefailsafecurrentoutputifthesensorloosesechoconfidence.Whenthesensorregainsechoconfidence,theoutputcurrentwillrevertbacktothecurrentlevelcondition.

    o HoldLastValue Theoutputwill remain in thesamestateasthelastconfidentechodetected.Example:Iftheoutputwas6.7mA just prior to the lost signal, the sensor will continue tooutput6.7mAuntilechoconfidenceisregained.

    o Empty The output will revert to the current value for anemptycondition.When4mAatBottomisselected,thesensorwill output 4 mA during a failsafe condition. If 20 mA atBottom isselected,thesensorwilloutput20mAduringafailsafecondition.

    o Full The output will revert to the current value for a fullcondition. When4mAatBottom is selected, the sensorwilloutput20mAduringafailsafecondition. If20mAatBottomis selected, the sensor will output 4 mA during a failsafecondition.

    o Overfill(21mA)Thesensorwilloutput21mAduringafailsafecondition.

    o Overfill(22mA)Thesensorwilloutput22mAduringafailsafecondition.

    OutputatEmpty(NotAvailableonDS14series)Thisfeatureallowsyoutoselecttheorientationofthe4to20mAoutput(4to20mAor20to4mA). Choosewhichoutputsettingbestfitstheapplication. Typicalapplicationsaresetwith4mAatBottom. Factorydefaultis4mAatbottomand20mAattop.Whenconnectingyoursensortoadisplay,youmustaccountforyouroutputorientationsetting.

    o 4mA at Bottom The output currentwill be 4mAwhen thesensormeasures an empty tank and 20mAwhen the sensormeasuresafulltank.

    o 20mAatBottom Theoutputcurrentwillbe20mAwhen thesensor measures an empty tank and 4mA when the sensormeasuresafulltank.

    Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton

  • RevA MN302000 15of38

    WEBCAL(LOOPFAILSAFE)DL10,LU23,LU27,LU28&LU29SeriesOnly StepFive

    LoopFailSafeThisfeatureallowsyoutoselectthefailsafecurrentoutputifthesensorloosesechoconfidence.Whenthesensorregainsechoconfidence,theoutputcurrentwillrevertbacktothecurrentlevelcondition.

    o HoldLastValue Theoutputwill remain in thesamestateasthelastconfidentechodetected.Example:Iftheoutputwas6.7mA just prior to the lost signal, the sensor will continue tooutput6.7mAuntilechoconfidenceisregained.

    o Empty The output will revert to the current value for anemptycondition.When4mAatBottomisselected,thesensorwill output 4 mA during a failsafe condition. If 20 mA atBottom isselected,thesensorwilloutput20mAduringafailsafecondition.

    o Full The output will revert to the current value for a fullcondition. When4mAatBottom is selected, the sensorwilloutput20mAduringafailsafecondition. If20mAatBottomis selected, the sensor will output 4 mA during a failsafecondition.

    o Overfill(21mA)Thesensorwilloutput21mAduringafailsafecondition.

    o Overfill(22mA)Thesensorwilloutput22mAduringafailsafecondition.

    OutputatEmptyThisfeatureallowsyoutoselecttheorientationofthe4to20mAoutput(4to20mAor20to4mA). Choosewhichoutputsettingbestfitstheapplication. Typicalapplicationsaresetwith4mAatBottom. Factorydefault is 4mA at bottom and 20mA at top. When connecting yoursensortoadisplay,youmustaccountforyouroutputorientationsetting.

    o 4mA at Bottom The output currentwill be 4mAwhen thesensormeasures an empty tank and 20mAwhen the sensormeasuresafulltank.

    o 20mAatBottom Theoutputcurrentwillbe20mAwhen thesensor measures an empty tank and 4mA when the sensormeasuresafulltank.

    Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton

  • 16of38 MN302000 RevA

    WEBCAL(STARTUPCONDITON)DL10,LU23,LU27,LU28&LU29SeriesOnly StepFive

    StartupConditionThisfeatureallowsyoutoselectthestartupcurrentwhenpowerisfirstapplied to the sensor. The sensorwill consume the selected powerwhile it isacquiringthe liquid level. Whenthecorrect levelhasbeenidentified,theoutputwilladjusttothe leveloutput. Usethisfeaturetoavoidfalsealarmswiththecontrollerwhenpowerisfirstappliedtothesensor.

    o Empty Thecurrentoutputwillreverttothecurrentvalueforan empty condition. When 4mA at Bottom is selected, thesensorwilloutput4mAwhilethesensorpowersup. If20mAatBottom isselected, thesensorwilloutput20mAwhile thesensorpowersup.

    o Mid Tank (12mA) The sensorwilloutput 12mAwhile thesensorpowersup.

    o Full The output will revert to the current value for a fullcondition. When 4 mA at Bottom is selected, the sensor willoutput20mAwhilethesensorpowersup.If20mAatBottomisselected,thesensorwilloutput4mAwhilethesensorpowersup.

    o Overfill(22mA)Thesensorwilloutput22mAwhilethesensorpowersup.

    Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton

  • RevA MN302000 17of38

    WEBCAL(SELECTOUTPUT)DX10SeriesOnly StepFive

    SelectOutputThis featureallowsyoutoselecttheoutput fortheDX10series. Theoutputcaneitherbevoltageorfrequency.

    o VoltageSelect thisoption foravoltageoutput. Youwillbeabletoselectvoltagescaleof02.5V,03.3V,05Vor010V.

    o Frequency Select this option for a frequency/pulse output.Theoutputwillbefixedfrom976to2000Hz.

    MaximumOutputThis sets the maximum output for the DX10 series. If voltage isselected, then the maximum value will be in volts. If frequency isselected,thenthemaximumvaluewillbeinhertz.

    o VoltageChoosefrom:2.5volts,3.3volts,5voltsor10volts.o FrequencyThisvalueisfixedat2000Hz.

    MinimumOutputThis sets the minimum output for the DX10 series. If voltage isselected, then the minimum value will be in volts. If frequency isselected,thentheminimumvaluewillbeinhertz.

    o VoltageChoosefrom:0.25volts,0.5volts,0.833volts,1voltor2volts.

    o FrequencyThisvalueisfixedat976Hz.Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton

  • 18of38 MN302000 RevA

    WEBCAL(OUTPUTATEMPTY&FAILSAFEOUTPUT)DX10SeriesOnly StepFive

    OutputatEmptyThis feature allows you to select the orientation of the voltage orfrequencyoutput(ex.Empty=1voltandfull=10voltsvs.empty=10volts and full = 1 volt). Choose which output setting best fits theapplication.TypicalinstallationsaresetwithMinimumOutput.Whenconnectingyoursensortoadisplay,youmustaccountforyouroutputorientationsetting.

    o MinimumOutput The outputwill be the smallest value. Ifreadinginvoltage,thiswillbetheminimumoutputvoltage.Ifreadinginfrequency,thiswillbe976Hz.

    o MaximumOutput The outputwill be the largest value. Ifreadinginvoltage,thiswillbethemaximumoutputvoltage.Ifreadinginfrequency,thiswillbe2000Hz.

    FailSafeOutputThis feature allows you to select the failsafe output if the sensorlooses signal confidence. When the sensor regains echo confidence,theoutputwillrevertbacktothecurrentlevelcondition.

    o HoldLastValue Theoutputwill remain in thesamestateasthelastconfidentechodetected.Example:Iftheoutputwas6.7volts just prior to the lost signal, the devicewill continue tooutput6.7voltsuntilechoconfidenceisregained.

    o Empty The outputwill revert to the set value for an emptycondition. WhenMinimumOutput isselected,thesensorwilloutputthe lowestvoltageorfrequencyvalueduringafailsafecondition. If Maximum Output is selected, the sensor willoutputthehighestvoltageorfrequencyvalueduringafailsafecondition.

    o FullTheoutputwillreverttothesetvalueforafullcondition.WhenMinimumOutput isselected,thesensorwilloutputthehighestvoltageor frequencyvalueduringa failsafecondition.If Maximum Output is selected, the sensor will output thelowestvoltageorfrequencyvalueduringafailsafecondition.

    Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton

  • RevA MN302000 19of38

    WEBCAL(STARTUPCONDITION)DX10SeriesOnly StepFive

    StartupConditionThisfeatureallowsyoutoselectthestartupoutputwhenpowerisfirstappliedtothesensor.Thesensorwilloutputtheselectedsignalwhileit is acquiring the liquid level. When the correct level has beenidentified,theoutputwilladjusttothe leveloutput. Usethisfeaturetoavoidfalsealarmswiththecontrollerwhenpowerisfirstappliedtothesensor.

    o Empty The outputwill revert to the set value for an emptycondition. WhenMinimumOutput isselected,thesensorwilloutput the lowestvoltageor frequencyvalueuntil the level isacquired. If Maximum Output is selected, the sensor willoutputthehighestvoltageorfrequencyvalueuntilthe level isacquired.

    o Mid Tank The output will hold at the mid level of theMinimumandMaximumOutputuntilthelevelisacquired.

    o FullTheoutputwillreverttothesetvalueforafullcondition.WhenMinimumOutput isselected,thesensorwilloutputthehighestvoltageorfrequencyvalueuntilthelevelisacquired.IfMaximumOutputisselected,thesensorwilloutputthelowestvoltageorfrequencyvalueuntilthelevelisacquired.

    Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton

  • 20of38 MN302000 RevA

    WEBCAL(TANKSHAPESELECTION) StepFiveStep2 Tank Selection: The sensormaybe configured in volumetricunits (Gallonsor Liters)orDistance(HeightofLiquid)units(inches,cm,feetormeters).WebCalwilldefaultinDistance(HeightofLiquid)withunitsof Inches. To change theunitsor to change fromDistance toVolume,press the Select Tank Shapebuttonlocatednearthecenterofthewindow.

    ShapeSelectionWindow:ThiswindowwillshowsthedifferenttankshapeoptionsavailableinWebCal. VerticalCylindrical VerticalCylindricalwithConeBottom HorizontalCylindricalwithEndcaps HorizontalCylindricalwithSphericalEnds Spherical Rectangular StrappingTableUse this feature formanualentry

    ofmeasuredtankdistancesandvolumes.Select any of the above tank shapes and press OK toconfirm.

  • RevA MN302000 21of38

    WEBCAL(TANKSHAPESELECTION) StepFiveDimensionalEntryVerticalCylindricalExample: Choose theSensorOutputUnitsasDistanceorVolume.AfterchoosingtheSensorOutputUnits,selecttheunitsofmeasurementinthepulldowntotheleft.

    UnitsofMeasurementDistance VolumeInchesCmFeet

    Meters

    GallonsLiters

    DistanceSensorOutputUnits:Enterthedimensionsofthetank. Youmustenterdatainallfieldsshown.SensorHeight:Distancefromthebottomofthetanktothebottomofthetransducer.FillHeight:Distancefromthebottomofthetanktothemaximumliquidheight.RiserHeight:Distancethesensorisrecessedwithinariser,measuredfromthebottomofthesensortotheinsideofthetank.

    VolumeSensorOutputUnits:Enterthedimensionsofthetank. Youmustenterdatainallfieldsshown.SensorHeight:Distancefromthebottomofthetanktothebottomofthetransducer.FillHeight:Distancefromthebottomofthetanktothemaximumliquidheight.RiserHeight:Distancethesensorisrecessedwithinariser,measuredfromthebottomofthesensortotheinsideofthetank.TankHeight: Distance from the bottom ofthetanktothetopofthestraightsidewall.Diameter: Distance of the inside tankdiameter.

  • 22of38 MN302000 RevA

    WEBCAL(TANKSHAPESELECTION) StepFiveVolumeTankCapacity: Afterentering thedimensions,press theCapacitybutton toshow theCalculatedCapacityofthe tank. If theCalculatedCapacity isslightlydifferent than theexpectedcapacity,clickon theAdjustCapacityboxandentertheexpectedcapacityofthetank.IftheAdjustedCapacityismorethan10%oftheCalculatedCapacity,recheckthedimensionsinformationenteredabove.

    Whenalldimensionsareentered,presstheApplybuttontoreturntothepreviousConfigurationwindow.

    ApplyTransfersthedimensionstotheoriginalConfigurationwindow. TanksReturnstothepreviousShapeSelectionwindow. CancelReturnstotheConfigurationwindowwithoutsavinganyinformation. HelpJumpstotheHelpmenu.

  • RevA MN302000 23of38

    WEBCAL(TANKSHAPESELECTION) StepFiveDimensional Entry Horizontal Cylindricalwith End Caps Example: Choose the SensorOutputUnits asDistanceorVolume. After choosing the SensorOutputUnits, select theunitsofmeasurement in thepulldowntotheleft.

    UnitsofMeasurementDistance VolumeInchesCmFeet

    Meters

    GallonsLiters

    DistanceSensorOutputUnits:Enter thedimensionsof the tank. Youmustenterdatainallfieldsshown.SensorHeight:Distancefromthebottomofthetanktothebottomofthetransducer.FillHeight: Distancefromthebottomofthetanktothemaximumliquidheight.RiserHeight:Distancethesensorisrecessedwithinariser,measured from thebottomofthesensortotheinsideofthetank.

    VolumeSensorOutputUnits:Enter thedimensionsof the tank. Youmustenterdatainallfieldsshown.SensorHeight:Distancefromthebottomofthetanktothebottomofthetransducer.FillHeight: Distancefromthebottomofthetanktothemaximumliquidheight.RiserHeight:Distancethesensorisrecessedwithinariser,measured from thebottomofthesensortotheinsideofthetank.TankHeight:Distancefromthebottomofthetanktothetopofthestraightsidewall.CylinderLength:Distanceofthestraightlengthofthetank.EndCapLength:Distanceofoneendcap.Bothendcapswillbeusedinthevolumecalculation.

  • 24of38 MN302000 RevA

    WEBCAL(TANKSHAPESELECTION) StepFiveVolumeTankCapacity: Uponenteringthedimensions,presstheCapacitybuttontoshow theCalculatedCapacityofthe tank. If theCalculatedCapacity isslightlydifferent than theexpectedcapacity,clickon theAdjustCapacityboxandentertheexpectedcapacityofthetank.IftheAdjustedCapacityismorethan10%oftheCalculatedCapacity,recheckthedimensionsenteredabove.

    Whenalldimensionsareenterediscompleted,presstheApplybuttontoreturntothepreviousConfigurationwindow.

    ApplyTransfersthedimensionsbacktotheoriginalConfigurationwindow. TanksReturnstothepreviousShapeSelectionwindow. CancelReturnstotheConfigurationwindowwithoutsavinganyinformation. HelpJumpstotheHelpmenu.

  • RevA MN302000 25of38

    WEBCAL(TANKLEVELSELECTION) StepFiveStep3TankLevels:ThissectionofWebCaliswhereyouconfirmtheSensorHeightandFillHeightsettingswhichapplytoallsensorseriesandentertherelaysettingsifapplicable.TheDimensionalentrywindowwilltransferthesesettingstotheConfigurationwindow.IfVolumewasselected,thentheSensorHeightandFillHeightsettingswillbegrayedout.

    For theDL10,DL14,DL24 andDL34 series, the SensorHeight and FillHeight values determine theoperationalrangeforthe420mAoutput.

    FortheDX10series,theSensorHeightandFillHeightvaluesdeterminetheoperationalrangeforthevoltageorfrequencyoutput.

    SensorHeight: Distancefromthebottomofthetanktothebottomofthetransducer.FillHeight:Distancefromthebottomofthetanktothemaximumliquidheight.

    WebCalRelayValues(DL14,DL24,DL34andDS14seriesonly):ThissectionoftheConfigurationwindowiswhereyouentertheoperationalvaluesfortherelays. Youmustentervalueswithinallfieldsshown. Allvaluesmustbe in thepreviously selectedunitsofoperation and arebaseduponunitsof liquid. All relayvaluesmustbegreaterthananemptytank(0units)andlessthanafulltank(FillHeightorCapacityVolume).

    Example: Thisisa4HighLevelAlarmoperation. Asthe levelincreases from the bottomup,eachalarmwillactivateat4increments.Example: This is a 2PumpLead/Lag operation with aHigh and Low alarm. Thepumps will automaticallyempty the tank and thealarmswillactivateifthelevelgetstoohighorlow.

  • 26of38 MN302000 RevA

    WEBCAL(WRITETOUNIT) StepFiveWritetoUnitAfteryouhaveenteredconfigurations,selectedandconfiguredtheTankShapeandenteredthe TankValues, click Write toUnit and load theconfiguration into thememoryofthesensor. Whencompleted, this configurationwill remain inside thesensormemoryandwillnotchangeunlessthesensorisconnectedtoWebCalandanewconfiguration iswrittentothesensor. Lossofpowerwillnotchangeorlosetheconfigurationwithinsensormemory.

    WritetoUnitWiringdiagramSaveConfigFile

    Next,usethefilemanagementfeaturestosaveyourconfigurationbyclickingSaveConfigFileandprintyourwiringdiagrambyclickingWiringDiagram.SaveConfigFilewillsavethisconfigurationasatextfilewhichcanbeloadedbackintoWebCalbypressingthe Open Config File button. It is good practice to save the configuration file for each differentconfigurationwithauniquename foreasy identification. Ifusingmultiplesensors in identicalapplications,thenuseofasingleconfigurationfileisrecommended.WiringDiagramwilldisplayaPDF fileshowing theuniquewiring for thespecificconfigurationcreated inWebCal.ThePDFcanbeprintedorEmailed.Itisgoodpracticetosavethewiringdiagramasabackup.TechTips:

    420mAOutputOnly:Ifonlythe420mAoutputisrequiredwithaDL14,DL24orDL34(andnorelaysareused),underNumberofPumps,select420mATransmitterOnlytosimplifytheconfiguration.

    RelaySetPoints: Neversetrelayssetpointsattheextremeendoftheoperationalrange(emptyorfull).Arelayrequirestheleveltopassthesetpointbeforeitswitches.ExampleIfalowalarmissetto0gallons, itwillnevertriggerbecausethe levelmustpassbelow0gallons for ittoactivate. Bestpracticeistosetthelowalarmwithaslightbuffer.

    AlarmPrevention:Alwayssetrelayalarmsetpointswherethereistimetoreacttopreventanissue.ExampleSettinganalarmwhereatankoverflowsisnotadvisable.Bestpracticeistosetahighlevelalarmwhereoperatorshaveplentyoftimetopreventanoverflow.

  • RevA MN302000 27of38

    WEBCAL(WIRINGDIAGRAM) StepFiveWiringDiagramSample

    Diagramschangebaseduponeachuniquesensorconfiguration.

    Thediagramshownaboveisonlyasampleandshouldnotbeusedasawiringdiagram.SensorWiring: Once the sensorhasbeen installed, follow theWiringDiagramprovidedby theWebCalsoftware. A typical wiring diagram is shown above. Flowline recommends using a qualified licensedelectriciantowireEchoPodandEchoSonicwithyourapplicationscomponents.

    Configure your sensor with WebCal and use the wiring diagram button to view the appropriatediagram. Eachconfigurationwillhave itsownuniquediagram. Thediagramshownabove isonlyasampleandshouldnotbeusedasawiringdiagram.

    Alwaysusestepperrelaysbetweenthesensorandexternalloads.ForDCcircuits,alwaysuseacatchdiodesuchas1N4148,shownontheWiringdiagramabovesuppliedbyWebCal.

    Once the sensor is configured, isolate thewhiteandgreenwires fromactivepower topreventashortoftheconfigurationcircuit.

    Generalnotesforelectricalconnections,usageandsafety: Wherepersonalsafetyorsignificantpropertydamagecanoccurduetoaspill,theinstallationmust

    havearedundantbackupsafetysystem. Wiringshouldalwaysbecompletedbyalicensedelectrician. Supplyvoltageshouldneverexceed28VDC. Protectthesensorfromexcessiveelectricalspikesbyisolatingthepower,wheneverpossible. Thesensormaterialsmustbechemicallycompatiblewiththeliquidstobemeasured. Designafailsafesystemforpossiblesensorand/orpowerfailure. Neverusethesensorinenvironmentsclassifiedashazardous.

  • 28of38 MN302000 RevA

    WEBCAL(ADVANCEDFEATURES) StepFiveAdvancedFeatures: Theadvancedfeaturessettingsaredesignedtohelpsolveperformanceoroperationalissuesforspecificapplications.Changingthesesettingwillalterthefactorydefaultperformanceoroperation,ofyoursensor.PleasereadthroughthisHELPfiletoassistyouinmakingadjustmentsorifyouarestillunclearaboutaspecificissue,pleasecontactFLOWLINEapplicationsengineering.

    Note:WhentheAdvancedButtonishighlightedwithaREDborder,thisindicatesyouhaveselectedanadvancedfeature.

  • RevA MN302000 29of38

    WEBCAL(ADVANCEDFEATURES) StepFive

    IncreaseOutputFiltering:Placingacheckmarkintheboxwillincreasethefiltering(averaging)ofthe

    analogoutput.Usethisfilterifthe4to20mAoutputrequiresasmootheroutputfortheapplicationsuchasopenchannelflowmeasurement.

    DecreaseOutputFiltering:Placingacheckmarkintheboxwilleliminateallfiltering(averaging)oftheanalogoutputwhichenablesapulsebypulselevelreading.Usethisfiltertoseechangesinlevelaftereveryechopulse.Note:Nevercheckincreaseoutputfilteringanddecreaseoutputfilteringatthesametime.

    StabilizeOutputinDeadBand:PlacingacheckmarkintheboxwillactivateafiltertoholdtheoutputatFullifthelevelentersthedeadbandoftheEchoPodandEchoSonic.Thisfilterrequirestheleveltoleavethedeadbandatasmoothandsteadyrate.

    InvertRelayStates:Placingacheckmarkinanyofthefourboxeswillreversethestateofthatrelay.Forexample,ifrelay4isahighalarmthatenergizesabove50.0ofliquid,checkingtheinvertboxwillreverseitsstatesotherelaywillenergizewhenthelevelisbelow50ofliquid.Invertingtherelaywillalsoinvertthefailsafeoftherelay.IftherelayisfailsafeOn,invertingtherelaywillmakeitfailsafeOff.

    FailSafewillalso invertunlessyoucheckbelow: Placingacheckmark intheboxwillnot invertthefailsafewhenarelayisinverted.

  • 30of38 MN302000 RevA

    APPENDIX StepSixSampleProgram: AsampleversionofWebCal isavailableanytimeasensor isnotattached toWebCal.TheSampleProgramshowsallthefeatures intheConfigurationTabofWebCal. Anyconfigurationcanbeopened (OpenConfigFile),Saved (SaveConfigFile)orPrinted (PrintConfigFile)with theSampleProgram.SampleProgramcannotbeviewedifasensorisattachedtothecomputerviatheLI991001Fob.ToviewtheSampleProgram,startWebCalwhenasensorisnotattachedtothecomputer.Attheopeningscreen,selectSampleProgram.

    Selectyourmodeltypeintheupperrighthandcorner.Note:Whensavingoropeningaconfiguration,makesuretheModelNumbermatchesthesensorsyouintendtouse.

  • RevA MN302000 31of38

    APPENDIX StepSixUpdatingWebCal Software: WebCal can be updated directly fromwithin the software. Click on theUpdatesTabatthetopofthewindowandpresstheDownloadbutton. MakesurethatyourcomputerhasaccesstotheInternet.Ifnot,anerrorwindowwillappear.PresstheDownloadbuttonandthesoftwarewillchecktheversionofsoftwareyouareusingwiththemostrecentsoftwareversionatFlowline.Iftheversionsarethesame,awindowwillopenindicatingthatthemostrecentversionisinstalled.Ifnot,thenawindowwillopenaskingyoutodownloadthelatestversion.Followtheinstructionsforinstallingthelatestversion.

  • 32of38 MN302000 RevA

    APPENDIX StepSixUpdatingSensorFirmware: WebCalcanalsobeusedtoupdatethefirmware insideyoursensorwiththelatest featuresandenhancements. FirstopenWebCalwitha sensor connectedand the latestversionofWebCalinstalledonyourcomputer.

    Whenupdatingthesensorfirmware,disconnectthesensorfromallotherdevicesincludingdisplays,controllers,power supplies,PLCs,pumps,valvesandalarms. Connect thedevicesbackafter thefirmwarehasbeenupdated.

    ClickontheUpdatesTabandthenclickonSelectProgramtoselectthefirmwareupdate.

    SelectthelatestversionofthefirmwarefileandclickOK.

    ConfirmthattheaddressiscorrectandthenclickonUpdateSensortobeginthefirmwareupdate.Thisstepshouldtake lessthan1minute. Youcan followtheprogresswiththestatusbartotherightoftheUpdateSensorbutton.Whencompleted,clickontheConfigurationtabtoconfigurethesensor.Remember,whenthefirmwarehasbeenupdated,theunitwillreturntoitsoriginalfactorysettings.

    Ifthereisacommunicationinterruptionduringtheupdate,theprocesswillstop.ItisthenOKtoclickonUpdateSensortostarttheprocessoveragain.

  • RevA MN302000 33of38

    APPENDIX StepSixDemoPage:TheDemoPagegivesyoutheabilitytotestthesensorsoperationagainstafixedtarget,includingtheconfirmationoftheswitchingpointsfortherelays.

    Thelargetopnumbershowstheliquidlevelvalueintheunitslistjustbelowtotheright. Thebottomsmallernumbershowsthedistancefromthesensorfacetothetargetinthesameunits.

    o AddingbothnumberswillequaltheSensorHeightvalue. TheActivitybuttonwilllighteverytimethesensorpulses. TheRelaybuttons(Relay1,2,3&4)willlightwhentheircorrespondingrelayisenergized.

    o Note:Thisfeatureisonlyasimulation.Therelaysarenotphysicallyopeningandclosing.

    Note:Inthismode,thevoltagesuppliedbythecomputerisbelowthesensorsspecifiedlimits.Therefore,thesensormaybesluggish,relaysmaynotoperateandtheunitmaynotoperatetofullrange.Thissimulationisintendedtogiveyouaquickfunctionalcheckafterupdatingthefirmware.

  • 34of38 MN302000 RevA

    APPENDIX StepSixStrappingTable: WebCal featuresastrapping table thatenablesyou toenterup to16customreferencepoints insteadofusingthestandardtankshapes. Thisfeature is idealforoddshapedtanksortankswherespecificlevelsareknownvolumesofliquid.ToaccesstheStrappingTable,clickonStrappingTableintheShapeSelectionWindowandpressOK. Enter the SensorHeight, FillHeight,

    RiserHeight and TankHeight. Thisinformation isused to configure thesensortothetank.

    The Strapping Table also has twocolumnsof16pointsforenteringtheknowntankdata.

    Selectthedimensionsand/orvolumeat the top of the two columns andenterthetankdata.

    When done, press Apply totransfer the data and return to theConfigurationwindow.

  • RevA MN302000 35of38

    APPENDIX StepSixLinearvs.NonLinear:Twooftheshapes(VerticalCylinderTankandRectangularTank)willalwaysprovidealinearoutput,regardlessofselectingDistanceorVolume. Theremainingfourshapes(VerticalCylinderTankwithConeBottom,HorizontalCylinderTankwithEndCaps,HorizontalCylinderTankwithSphericalEndCapsandSphericalTank)willhavealinearoutputwhenDistanceisselected,butwillhaveanonlinearoutputwhenvolumeisselected.

    VerticalCylindricalTank

    In theabove illustration,10of liquidwillalwaysbeequalto100gallonsofliquid(1=10gallons).

    HorizontalCylinderTankwithEndCaps

    In the above illustration, 1 of liquid does notequal 10 gallons. The 10 at the bottomrepresentsariseof62.8gallonswherethechangebetween 10 and 20 represents an increase of109.6gallons.

    Whenvolumeisselected,the420mAoutputfromthesensorwillbeproportionaltothevolumeofthetank,nottheheightofthetank.Thismeansthatthecurrentoutputwilltrackthevolumeofthetank(ingallonsorliters)withinanonlineartank (VerticalCylinderTankwithConeBottom,HorizontalCylinderTankwithEndCaps,HorizontalCylinderTankwithSphericalEndCapsorSphericalTank).Whenconnectingthesignaloutputtoadisplay,thesignalwillfollowthevolumeofthetank.Thedisplaywillalsoreflectthevolumeofthetankandnottheheightoftheliquid.Example#1(VolumeOutput):intheaboveillustrations,@20ofliquid,thedisplaywillshow200.0gallonsintheVerticalCylindricalTank. However, intheHorizontalCylinderTankwithEndCaps,thesame levelof20wouldshow172.4gallons.

  • 36of38 MN302000 RevA

    APPENDIX StepSixExample #2 (CurrentOutput): In the illustrations on the previous page, the 4mA signal is set at 0 (0.0gallons)andthe20mAsignal issetto60(600.0gallons). IntheVerticalCylindricalTank,40of liquidwilloutputacurrentsignalof14.67mA.However,intheHorizontalCylindricalTankwithEndCaps,50ofliquidwilloutputacurrent signalof15.41mA. A simple loopdisplay setwith4mA=0gallonsand20mA=600gallonswillshow twodifferentvolumesbasedupon the tankshapeconfiguration. VerticalCylindricalTankwillshow400.0gallonswhileHorizontalCylindricalTankwithEndCapswillshow428.0gallons.

    VerticalCylindricalTank

    In theabove illustration,10of liquidwillalwaysbeequalto100gallonsofliquid(1=10gallons).

    HorizontalCylinderTankwithEndCaps

    In the above illustration, 1 of liquid does notequal 10 gallons. The 10 at the bottomrepresentsariseof62.8gallonswherethechangebetween 10 and 20 represents an increase of109.6gallons.

  • RevA MN302000 37of38

    TROUBLESHOOTING StepSevenPROBLEM SOLUTION

    NoUnitDetectedinWebCal.

    WebCal cannotdetectanEchoPod orEchoSonicconnectedtothecomputer. CheckthattheFobisconnectedtotheUSBport. Checkthatallfourwires(Red,Black,WhiteandGreen)aresecurely

    attachedtotheFob. Check Device Manager and confirm that both drivers (WebCal

    Configuration&EchoFob)arepresent.Interneterror.Theservernameoraddresscouldnotberesolved.

    ThisisawarningindicatingthatthecomputerconfiguringEchoPod orEchoSonic is not connected to the Internet. ClickOK to continue.Flowline recommends being connecting to the Internet duringconfiguration. Not being connected to the Internetwill not preventEchoPodorEchoSonicfrombeingconfigured.Toturnoffthiswarning,gototheUpdatesTabandclickonthecheckbox Automatically upload configurations. Click onNO in the newwindowandthepreviouscheckboxwillbecomeunchecked.WebCalwillnolongerattempttoconnecttotheinternet.Clickingonthecheckboxwillrestorethisfeature.

    CannotaccesssomeoftheconfigurationfeaturesinWebCal.

    AschoicesaremadeinConfiguration,WebCalwillbegintoeliminatefunctions that are not applicable to a configuration. To resetConfigurationorgetaccesstoallthefeatures,clickontheClearScreenbutton.

  • 38of38 MN302000 RevA

    WARRANTY,RETURNSANDLIMITATIONS StepEightWarrantyFlowlinewarrants to the original purchaser of its products that such productswill be free from defects inmaterialandworkmanshipundernormaluseandserviceinaccordancewithinstructionsfurnishedbyFlowlinefor a period of two years from the date ofmanufacture of such products. Flowline's obligation under thiswarrantyissolelyandexclusivelylimitedtotherepairorreplacement,atFlowline'soption,oftheproductsorcomponents, which Flowline's examination determines to its satisfaction to be defective in material orworkmanshipwithinthewarrantyperiod.Flowlinemustbenotifiedpursuanttotheinstructionsbelowofanyclaimunderthiswarrantywithinthirty(30)daysofanyclaimedlackofconformityoftheproduct.Anyproductrepairedunder thiswarrantywillbewarrantedonly for the remainderof theoriginalwarrantyperiod.Anyproductprovidedasareplacementunderthiswarrantywillbewarrantedforthefulltwoyearsfromthedateofmanufacture.ReturnsProducts cannotbe returned to Flowlinewithout Flowline'spriorauthorization. To returnaproduct that isthought to be defective, go towww.flowline.com, and submit a customer return (MRA) request form andfollow the instructions therein.Allwarrantyandnonwarrantyproduct returns toFlowlinemustbe shippedprepaidandinsured.Flowlinewillnotberesponsibleforanyproductslostordamagedinshipment.LimitationsThiswarrantydoesnotapplytoproductswhich:1)arebeyondthewarrantyperiodorareproductsforwhichthe originalpurchaserdoesnot follow thewarrantyprocedures outlinedabove;2)havebeen subjected toelectrical,mechanicalorchemicaldamageduetoimproper,accidentalornegligentuse;3)havebeenmodifiedoraltered; 4)anyoneotherthanservicepersonnelauthorizedbyFlowlinehaveattemptedtorepair;5)havebeeninvolvedinaccidentsornaturaldisasters;or6)aredamagedduringreturnshipmenttoFlowline.FlowlinereservestherighttounilaterallywaivethiswarrantyanddisposeofanyproductreturnedtoFlowlinewhere:1)thereisevidenceofapotentiallyhazardousmaterialpresentwiththeproduct;or2)theproducthasremainedunclaimedatFlowlineformorethan30daysafterFlowlinehasdutifullyrequesteddisposition.Thiswarrantycontains the sole express warranty made by Flowline in connection with its products. ALL IMPLIEDWARRANTIES,INCLUDINGWITHOUTLIMITATION,THEWARRANTIESOFMERCHANTABILITYANDFITNESSFORAPARTICULARPURPOSE,AREEXPRESSLYDISCLAIMED.Theremediesofrepairorreplacementasstatedabovearetheexclusiveremediesforthebreachofthiswarranty.INNOEVENTSHALLFLOWLINEBELIABLEFORANYINCIDENTALORCONSEQUENTIALDAMAGESOFANYKINDINCLUDINGPERSONALORREALPROPERTYORFORINJURYTOANYPERSON.THISWARRANTYCONSTITUTESTHEFINAL,COMPLETEANDEXCLUSIVESTATEMENTOF WARRANTY TERMS AND NO PERSON IS AUTHORIZED TO MAKE ANY OTHER WARRANTIES ORREPRESENTATIONSONBEHALFOFFLOWLINE.Thiswarrantywillbe interpretedpursuant to the lawsof theStateofCalifornia. Ifanyportionofthiswarranty isheldtobe invalidorunenforceableforanyreason,suchfindingwillnotinvalidateanyotherprovisionofthiswarranty.

    Forcompleteproductdocumentation,videotraining,andtechnicalsupport,gotowww.flowline.com.Forphonesupport,call5625983015from8amto5pmPST,MonFri.(PleasemakesureyouhavethePartandSerialnumberavailable.)