application of two-dimensional x-ray diffraction (xrd ) throughput xrays 1.pdf · x-ray diffraction...
TRANSCRIPT
1
24.06.2008Bruker Confidential2
Bob He
Application of Two-Dimensional X-Ray Diffraction (XRD )
High-throughput Screening with XRD
24.06.2008Bruker Confidential5
Phase Identification
Quantitative Analysis
Texture
Stress
Small Angle X-ray Scattering
High-Throughput Screening
Micro Diffraction
Mapping
Forensics and Archaeology
Thin Films
XRD2: Theory, Systems and Applications
Geometry Conventions
System and Configuration
X-ray Diffraction (XRD)
2
24.06.2008Bruker Confidential17
Two-dimensional detectors revolutionized the X-ray diffraction
What can you do with XRD ?
24.06.2008Bruker Confidential18
Comparison: Area Detector vs. Point Detector
Instant Capture of 2D pattern vs. one intensity value at a time
3
24.06.2008Bruker Confidential19
XRD & Single Crystals
24.06.2008Bruker Confidential20
XRD & Micro Samples
4
24.06.2008Bruker Confidential21
XRD & Textured Materials
24.06.2008Bruker Confidential22
XRD & Powders
5
24.06.2008Bruker Confidential23
XRD & Strained Materials
24.06.2008Bruker Confidential24
Debye Cone
Sample
Incident Beam
XRD Pattern
Discover the γ-information
Open your eyes with XRD
6
24.06.2008Bruker Confidential31
XRD2: Comparison with Conventional XRD (1)
The powder diffraction pattern in 3D space (blue) and the conventional diffractometer plane.
24.06.2008Bruker Confidential32
scintillation detector
small spot measuredscan necessarylong measuring time
PSD
large 2θ range measured simultaneouslymedium measuring time
GADDS
large 2θ and chi range measured simultaneouslymeasurement of oriented samplesvery short measuring timesintensity versus 2θ by integration of the data
XRD2: Comparison with Conventional XRD (2)
7
24.06.2008Bruker Confidential33
XRD2: Geometry Convention (1) - Diffraction Space
Diffraction rings (blue) in the laboratory axes (red).⎥
⎥⎥⎥
⎦
⎤
⎢⎢⎢⎢
⎣
⎡
−
−
−
=
⎥⎥⎥⎥
⎦
⎤
⎢⎢⎢⎢
⎣
⎡
=
γθ
γθ
θ
coscos
sincos
sin
L
z
y
x
h
h
h
h
24.06.2008Bruker Confidential34
XRD2: Ideal Detector for Diffraction Pattern in 3D Space
An ideal detector to measure 3D space diffraction pattern is a sphere with the sample in the center.The direction of a
diffracted beam is defined by γ (longitude) and 2θ (latitude).The incident X-ray
beam points from 2θ=πto 2θ=0.
8
24.06.2008Bruker Confidential35
XRD2: Diffraction Pattern on 2D Detector
A 2D detector can be treated as a detecting surface intersecting the diffraction cone.The detecting surface
can be a plane or a curved surface, such as sphere or cylinder.The conic section of a
plane may be a circle, ellipse, parabola, or hyperbola depending on the swing angle α.
24.06.2008Bruker Confidential36
XRD2: Geometry Convention (2) - Detector Space
Detector position in the laboratory coordinates is determined by the detector distance D and swing angle α.
9
24.06.2008Bruker Confidential37
XRD2: Geometry Convention (2) - Detector Space
Conversion of pixel intensity into 2θ and γintensity based on the detector position in the laboratory coordinates: D and α.
)20(,cossincos2222
1 πθααθ <<++
+= −
yxDDx
)(,)sincos(
cossincossincos
22
1 πγπαααα
ααγ ≤<−−+
−−−
= −
Dxyy
DxDx
24.06.2008Bruker Confidential38
XRD2: Geometry Convention (3)- Sample Space
Rotation axes ω, φ, χgand the laboratory axes XLYLZL (red).
Rotation axes ω, φ, χg(ψ) and translation axes XYZ (blue).
10
24.06.2008Bruker Confidential39
XRD2: Geometry Convention - Transformation Matrix
⎥⎥⎥⎥⎥
⎦
⎤
⎢⎢⎢⎢⎢
⎣
⎡
−−−
−
−−−−
ψψωψω
φψφωφψω
φωφψω
φψφωφψω
φωφψω
sincoscoscossin
coscossinsincossincos
sincoscossinsin
sincoscossinsinsincos
coscossinsinsin
The transformation matrix from the laboratory coordinates XLYLZL to the sample coordinates S1S2S3 in Eulerian geometry
24.06.2008Bruker Confidential40
XRD2: Diffraction Vector in Sample Space
The unit vector hS of the diffraction vector Hhkl in the sample coordinates S1S2S3 is given by:In matrix form:
or:
Ls Ahh =
⎥⎥⎥⎥⎥
⎦
⎤
⎢⎢⎢⎢⎢
⎣
⎡
−
−
−
⎥⎥⎥⎥⎥
⎦
⎤
⎢⎢⎢⎢⎢
⎣
⎡
−−−
−
−−−−
=
⎥⎥⎥⎥
⎦
⎤
⎢⎢⎢⎢
⎣
⎡
⎥⎥⎥⎥
⎦
⎤
⎢⎢⎢⎢
⎣
⎡
=
⎥⎥⎥⎥
⎦
⎤
⎢⎢⎢⎢
⎣
⎡
γθ
γθ
θ
ψψωψω
φψφωφψω
φωφψω
φψφωφψω
φωφψω
coscos
sincos
sin
sincoscoscossin
coscossinsincossincos
sincoscossinsin
sincoscossinsinsincos
coscossinsinsin
333231
232221
131211
3
2
1
z
y
x
h
h
h
aaa
aaa
aaa
h
h
h
)sincoscossin(sinsincoscossincoscos)coscossinsin(sinsin1
ωφωψφγθψφγθωφωψφθ
−−++=h
)sinsincossin(cossincoscoscoscoscos)cossinsinsin(cossin2
ωφωψφγθψφγθωφωψφθ
++−−−=h
ψγθωψγθωψθ sincoscoscoscossincossincossin3 −−=h
11
24.06.2008Bruker Confidential43
Key Components for XRD2
Patented Hi-Star Area Detector
designed for singleevent detectionquantum efficiency> 80%no intrinsic detectornoiseinstant read-out
Unique Sensitivity &Unique Speed
24.06.2008Bruker Confidential44
2-Dimensional MikroGap Detector –VÅNTEC-2000
12
24.06.2008Bruker Confidential45
Functional Principle of The New VÅNTEC-2000TM – MikroGapTM Technology
MikroGapTM technology with resistive anode:
shortens drift time of ionsfast electrons induce charge on readout strips
Adjusted surface resistance (105
- 107 Ω/ area): high enough to limit dischargeslow enough to support high count rates
US Patent US 6,340,819 B1
24.06.2008Bruker Confidential47
XRD2 : Choice of Detectors: Sensitivity vs. Counting Rate
MiKroGapMWPC
CCD
Image Plate
Detective Quantum Efficiency (DQE):The DQE is a parameter
defined as the square of the ratio of the output and input signal-to-nose ratios (SNR).
The DQE of a real detector is less than 100% because not every incident x-ray photon is detected, and because there is always some detector noise.
MiKroGap™ has the best overall performance.
2
)/()/(
⎟⎟⎠
⎞⎜⎜⎝
⎛=
in
out
NSNSDQE
13
24.06.2008Bruker Confidential48
XRD2:X-ray intensity. What X-ray intensity?
Four parameters are often used to describe x-ray intensity within a bandwidth ∆λ :Flux is defined as the total x-ray photons passing a plane crossing the x-ray beam
per unit time. The typical unit for flux is photons/second or pps. The flux is sometimes also referred to as the integrated intensity of the x-ray beam.
Fluence is defined as the number of x-ray photons passing a unit area of the beam crossing a plane per unit time. The typical unit is photons/second⋅mm2 or pps/mm2.An alternative terminology for fluence is flux density. Fluence is an appropriate parameter for measuring the local counting rate of area detectors.
Brightness is defined as photons passing through a surface defined by unit solid angle. The typical unit is photons/second⋅milliradian2 or pps/mrad2. Since no linear dimension is defined in brightness, it is more appropriate to be used for measuring emission strength of a point source.
Brilliance is defined as photons passing through a unit area of surface within unit solid angle. The typical unit is photons/second⋅mm2⋅milliradian2 or pps/mm2⋅mrad2. It is more appropriate to compare two sources of different focal spot size.
photons-per-second (pps) = count-per-second (cps) if measurement counts are calibrated.
24.06.2008Bruker Confidential49
XRD2: X-ray intensity and beam divergence. Liouville’s theorem?
Liouville's theorem describes the nature of the x-ray source and optics:
or
The brilliance of an x-ray source can not be increased by the optics.
The product of divergence or image size can be equal or great than the product of capture angle and source size.
A source size lager than the effective size only increases the power consumption.
The optics should be located as close as possible to the source.
α β
f1 f2
S1
A1
S2
A2
source
optics
image
x
y
ΘΦ
βα 21 SS ≤ Φ≤Θ 21 AA
14
24.06.2008Bruker Confidential51
SUPER SPEED SOLUTIONSTurbo-X-Ray SourceTM (TXS) for PPXRD
24.06.2008Bruker Confidential52
Lin
(Cps
)
0
1
2
3
4
5
6
7
8
9
2-Theta - Scale23 30 40 50
SUPER SPEED SOLUTIONSTXS vs. Sealed Tube with Corundum Plate
Gain factor > 6
15
24.06.2008Bruker Confidential53
Bright sealed tube for ultimateconvenience
Incoatec Microfocus Source – IµS
The source - tube
• High brilliance
• Low energy: 30 W
• Low maintenance
• Tube change as easy as forconventional sealed-tubes
• Air-cooled
• Spot size < 100 µm
24.06.2008Bruker Confidential54
IµS & VÅNTEC-2000 vs. Sealed TubeCorundum Comparison
Single 40mm Göbel Mirror,
45kV, 40mA,
0.3mm collimator
total counts: 78K
IµS & VÅNTEC-2000
45kV, 0.650mA,
0.3mm snout
total counts: 1235K
16
24.06.2008Bruker Confidential78
1D/2D: Point Spread Function and Resolution
Consider a very small diffraction spot (blue line-delta function)An adjacent spot – red lineRMS (root-mean-square) is another parameter for PSF:
A perfect detector - dashed blue line. A real detector - intensity in a spread distribution.Can be measured if the separation is larger than FWHM.
RMS2.3548FWHM ⋅=
24.06.2008Bruker Confidential79
2D detector resolution: Kα1-Kα2 split at 35° 2θ with NIST1976 (measured with VÅNTEC-2000)
∆λ (Kα2-Kα1) →∆2θ=0.06º→ 210 µm on the detector (20 cm)
17
24.06.2008Bruker Confidential80
XRD2 :Defocusing at low incident angle in reflection
Lower resolution when θ2 or (2θ-ω) → 90° ω
ωθθθ
sin)2sin(
sinsin
1
2 −==
bB
24.06.2008Bruker Confidential81
Cylinder detector with 5° incident angle for 5°~80° 2θ
Flat detector with several (5°,15°,25°,35°) incident angles for 5°~80° 2θ
XRD2: Defocusing effect with reflection sample depends on detector and data collection strategy
18
24.06.2008Bruker Confidential82
XRD2: Defocusing effect with reflection sample depends on detector and data collection strategy
Defocusing vs. Detectors
0
2
4
6
8
10
12
5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80Two Theta
Def
ocus
ing
Fact
or
FlatCylinderBB
Defocus effect can be minimized with data collection strategy
Cylinder detector may collect large 2θ range, but with large defocusing effect at high 2θangle
24.06.2008Bruker Confidential83
XRD2: Phase ID: γ-integration with merged frames
Software with automatic data collection strategy to minimize defocusing effect, and merge and integrate diffraction frame into diffraction profile for phase ID.
γγ
19
24.06.2008Bruker Confidential197
GADDS - all applications with ONE instrument
D8 DISCOVER with GADDS C2:Rapid Screening System for Many Applications
CatalystsChemistryFine ChemicalsBiochemicalsPetrochemicalsPharmaceuticals… ...
OpticsElectronicsSemiconductorsFuel CellsBatteriesPolymersCoatings… ...
24.06.2008Bruker Confidential203
Easy and accurate sample positioning without touching the sample surface
Video image of each material library spot can be automatically stored during data scan
XRD2: Laser/video sample alignment
20
24.06.2008Bruker Confidential204
XRD2 for Combinatorial Screening:Cross contamination at low incident angle in reflection
Cross contamination happens when θ1 or ω
sb>
1sinθ
s
24.06.2008Bruker Confidential205
Cross contamination at 4° incident angle.
XRD2 for Combinatorial Screening:Cross contamination at low incident angle in reflection
21
24.06.2008Bruker Confidential206
XRD2 for Combinatorial Screening:Reflection System (CS)
24.06.2008Bruker Confidential207
Lin
(Cps
)
0
1
2
2-Theta - Scale9 10 20 30 40
XRD2: Data Collection:Acetaminophen powder
5 second data collection 30 second data collection
22
24.06.2008Bruker Confidential208
XRD2: reflection vs. transmission
Reflection mode frame from corundum at 5°incident angle.
Transmission mode frame with perpendicular incident beam.
24.06.2008Bruker Confidential209
Lin
(Cps
)
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
1.1
1.2
1.3
1.4
1.5
2-Theta - Scale3 10 20 30
Reflection and Transmission Data Collection:Ibuprofen powder
10 second overall data collection
R
TT
23
24.06.2008Bruker Confidential210
D8 DISCOVER with GADDS CST :Combinatorial Screening Transmission (CST)
Reduce background from sample holderImprove resolution at
low incident angleEliminate cross-cell
data contaminationReduce primary
beam air scatteringUS Patent #6,859,520
24.06.2008Bruker Confidential211
Reflection Transmission convertible Geometry
US Patent #7,242,745
D8 DISCOVER with GADDS HTS :Combinatorial Screening Reflection & Transmission
24
24.06.2008Bruker Confidential212
Beam-down Transmission
D8 DISCOVER with GADDS HTS:Combinatorial Screening Reflection & Transmission
24.06.2008Bruker Confidential213
SUPER SPEED SOLUTIONS:Turbo-X-ray Source for PXRD (θ-2θ)
VÅNTEC-2000TXS
XYZ stage with ω
rotation
25
24.06.2008Bruker Confidential214
X-ray Source for XRD2:IµSTM for PPXRD θ-θ Reflection
24.06.2008Bruker Confidential215
Operations: Import1)Corundum06192007_newsource - File: Corundum06192007_newT_01.raw - Type: 2Th alone - Start: 22.000 ° - End: 55.200 ° - Step: 0.020 ° - Step time: 100. s - Temp.: Operations: Import1)corundum6282007 - File: corundum6282007_05.raw - Type: 2Th alone - Start: 22.000 ° - End: 55.200 ° - Step: 0.020 ° - Step time: 100. s - Temp.: 25 °C (Room) - Time S
Corundum06192007_newsource - Obs. Max: 35.150 ° - Max Int.: 21.1 Cps - FWHM: 0.187 °
corundum6282007 - Obs. Max: 35.144 ° - Max Int.: 1.25 Cps - FWHM: 0.189 °
Lin
(Cps
)
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
2-Theta - Scale22 30 40 50
Black: sealed tube
Red: IµS & VÅNTEC-2000
Comparison after Data Integration
Observation- (104) reflection
Black: Max Int: 1.25cps
FWHM : 0.189
Red: Max Int: 21.1cps
FWHM: 0.187
26
24.06.2008Bruker Confidential216
Comparison: IbuprofenIµS & VÅNTEC-2000 vs. Clasical set-up
IµS – XRD2 – foc
2mmX2mm on sample, and 200um spot focused on detector
small slice for integration to obtain better resolution
15 sec collection time
Sealed Tube
• 0.3 mm collimator
• Sample-Detector distance 29 cm
120 sec collection time
24.06.2008Bruker Confidential217
Combined System with XRD and RamanD8 SCREENLAB
US Patent #7269245
27
24.06.2008Bruker Confidential218
X-ray collimator
Alignment laser
Microscope w/ zoom X-ray
detector
Unilab probe
Light shield
Sample plate
Combined System with XRD and RamanD8 SCREENLAB
24.06.2008Bruker Confidential219
Carbamazepine
Lin
(Cou
nts)
0
1000
2000
3000
4000
5000
6000
7000
8000
9000
10000
11000
12000
13000
14000
15000
16000
17000
18000
2-Theta - Scale4 10 20
0
5000
10000
15000
20000
25000
30000
250 500 750 1000 1250 1500 1750 2000 2250
Carbamazepine
Cou
nts
Raman Shift
XRD&Raman Examples
XRD Image
RAMAN
28
24.06.2008Bruker Confidential220
PolySNAP for Combined Analysis:Correlation among XRD, Raman and Other Probes
Cell display
Dendrogram
3D plots
24.06.2008Bruker Confidential221
XRD2 Screening: Conclusions
Two-dimensional XRD has many advantages in combinatorial screening, including high speed and improved statistics.XRD screening can be done in either
reflection or transmission depending on the sample and plate design.The new instrument allows the XRD
screening at either reflection or transmission modes with automated conversion.