calibrations listed in certificate # 2166.01 and meets the...

16
This laboratory maintains A2LA accreditation to ISOIIEC 17025 for the specific calibrations listed in Certificate # 2166.01 and meets the relevant requirements of ISO 9001:2008 IACCREDITEDl CALlBRATION CERT #2166.01 FL.UKE @ Everett Service Center 1420 75th St. SW Everett, Washington 98203 Calibration Certificate USA e NQA ISO 9001:2008 (10100/2) Description: Manufacturer: Model: Serial Number: Customer Name: METRICA SA de CV City, State: MONTERREY N L, MX Customer Item ID: C-801 PO Number: CIA RMA Number: 4698843 Result Summary: MV<Spec = VES Received Date: 29 March 2011 CALlBRATOR FLUKE 5500A 120 6265013 Certificate Number: 945162-6265013:1301616947 Date of Calibration: 01 April 2011 Date of Certificate: 01 April2011 Recommended Due Date: Procedure Name: FLUKE 5500A:(1 YEAR) ACAL VER COMBINE Procedure Revision: 4.1 Data Type: FOUND-LEFT Temperature: 22.9 °Celsius Relative Humidity: 47 % This calibration certificate may contain data that is not covered by the A2LA Scope of Accreditation. The unaccredited material, where applicable, is indicated by an asterisk (*), or confined to clearly marked sections. PASS I FAIL tests are not accredited. If there is non accredited material, the non accredited data is compliant with ANSI/NCSLI Z540.1-1994(R2002) requirements. Measurement uncertainties at the time of test are given in the following pages, where applicable. They are calculated in accordance with the method described in NIST TN1297, for a confidence level of approximately 95% using a coverage factor of approximately 2 (K=2). In the attached measurement results, deviation may be expressed with units, Measured Value (MV) - Nominal Value (NV) or as a proportion ofthe nominal value «MV-NV)/NV), expressed without units with a scalar multiplier such as % (0.01), or as a ratio ofthe units (mAlA, INN, etc.) Descriptions such as IlAlA, IlVN, and others, where used to annotate results or column headings are the preferred replacements for what was historically labeled as "ppm" or parts-per-million and described the results in that column, unless otherwise noted by units symbols. The Data type that could be found in this certificate must be interpreted as: As-Found - Calibration data collected before the unit is adjusted and/or repaired. . As-Left - Calibration data collected after the unit is adjusted and/or repaired. Found-Left - Calibration data collected without any adjustment and/or repair performed. No statement of compliance with specifications is made or implied on this certificate. However, results are reviewed, if applicable, to establish where any measurement results exceeded the manufacturer's specifications and to communicate results by means of this certificate. Measured values areater than the Manufacturer's scecification (Scec) are indicated bv "!". Comments: ~~ Authorized Slgnature Sr. Spec;alty Tech Metrology T echnician Fluke Corporation 1420 75th 5treet 5W, Everett WA 98203 USA Telephone Facsirnile Internet Page 1 of 15 Rev 7.4, 02123/11 888.993.5853 425.446.6390 www.f1uke.com

Upload: vophuc

Post on 30-Oct-2018

213 views

Category:

Documents


0 download

TRANSCRIPT

This laboratory maintains A2LA accreditation to ISOIIEC 17025 for the specificcalibrations listed in Certificate # 2166.01 and meets the relevant requirementsof ISO 9001:2008

IACCREDITEDlCALlBRATION CERT #2166.01

FL.UKE @ Everett Service Center1420 75th St. SW

Everett, Washington 98203 Calibration CertificateUSAe

NQA ISO 9001:2008 (10100/2)

Description:Manufacturer:

Model:

Serial Number:

Customer Name:METRICA SA de CV

City, State: MONTERREY N L, MXCustomer Item ID: C-801PO Number: CIARMA Number: 4698843

Result Summary: MV<Spec = VESReceived Date: 29 March 2011

CALlBRATORFLUKE5500A 120

6265013

Certificate Number: 945162-6265013:1301616947Date of Calibration: 01 April 2011

Date of Certificate: 01 April2011Recommended Due Date:Procedure Name:

FLUKE 5500A:(1 YEAR) ACAL VER COMBINEProcedure Revision: 4.1

Data Type: FOUND-LEFTTemperature: 22.9 °Celsius

Relative Humidity: 47 %

This calibration certificate may contain data that is not covered by the A2LA Scope of Accreditation. The unaccredited material,where applicable, is indicated by an asterisk (*), or confined to clearly marked sections. PASS I FAIL tests are not accredited.If there is non accredited material, the non accredited data is compliant with ANSI/NCSLI Z540.1-1994(R2002) requirements.

Measurement uncertainties at the time of test are given in the following pages, where applicable. They are calculated in accordancewith the method described in NIST TN1297, for a confidence level of approximately 95% using a coverage factor of approximately 2(K=2).

In the attached measurement results, deviation may be expressed with units, Measured Value (MV) - Nominal Value (NV) or as aproportion ofthe nominal value «MV-NV)/NV), expressed without units with a scalar multiplier such as % (0.01), or as a ratio oftheunits (mAlA, INN, etc.) Descriptions such as IlAlA, IlVN, and others, where used to annotate results or column headings are thepreferred replacements for what was historically labeled as "ppm" or parts-per-million and described the results in that column,unless otherwise noted by units symbols.

The Data type that could be found in this certificate must be interpreted as:As-Found - Calibration data collected before the unit is adjusted and/or repaired.

. As-Left - Calibration data collected after the unit is adjusted and/or repaired.Found-Left - Calibration data collected without any adjustment and/or repair performed.

No statement of compliance with specifications is made or implied on this certificate. However, results are reviewed, if applicable, toestablish where any measurement results exceeded the manufacturer's specifications and to communicate results by means of thiscertificate. Measured values areater than the Manufacturer's scecification (Scec) are indicated bv "!".

Comments:

~~Authorized Slgnature

Sr. Spec;alty Tech

Metrology Technician

Fluke Corporation

1420 75th 5treet 5W, Everett WA 98203 USA

Telephone Facsirnile Internet Page 1 of 15

Rev 7.4, 02123/11888.993.5853 425.446.6390 www.f1uke.com

This certificate applies to only the item identified and shall not be reproduced other than in tull, without the specific written approval byFluke Corporation. Calibration certificates without signatures are not valid. This certificate shall not be used to claim product .endorsement bv A2LA.

FL.UKE@ Certificate Number:945162-6265013:1301616947

Calibration Date:1-Apr-ll

Traceability InformationFor each parameter listed below the calibration was conducted using an unbroken chain of standards to:

DCVoltageThe Voltage Reference standard group, traceable to the Fluke PrimaryElectricalStandards Laboratory,which is traceable to the U.S. representationof the volt, through the internationallyaccepted value of the Josephson constant Kj=483597.9GHzIVand a 10 VoltJosephson ArrayVoltageStandard.

Frequency and PeriodA GPS Disciplined Rubidiumoscillatorfrequency standard which is traceable to the Nationallnstitute of Standards and Technology(NIST).

ACVoltage, Resistance, DCCurrent, ACCurrent, Capacitance, Inductance, PhaseThe Fluke Primary Electrical Standards Laboratory, which is traceable to the NIST.

ACVoltage FlatnessThe Fluke Primary Electrical Standards Laboratory or Agilent Technologies Standards Laboratory which are traceable to NIST.

Humidity

The Vaisala Measurement Standards Laboratory Primary Salt calibration bath, with traceability based on the physical phenomena in which theequilibrium relative humidity values associated with certain saturated salt solutions are known.

Rise Time

The Tektronix GmbH Calibration Laboratory which is traceable to the Physikalisch-Technische Bundensanstalt (PTB), and the National PhysicalLaboratory (NPL).

Radiation TemperatureTraceableto NIST,PTB,andthe FlukePrimaryTemperatureStandardsLaboratory.

Contact TemperatureTraceable to the Fluke Primary Temperature Standards Laboratory which is traceable to the NIST.

Gas FlowThe Fluke Primary Gas Flow Laboratory which is traceable to the NIST.

Pressure

The Fluke Primary Pressure Laboratory (Phoenix), which is traceable to the Laboratoire National de Metrologie et D'Essais (LNE) and PTB.

Standards Used

Asset#10066

10242

10383

10384

13821

14386

6619

7602

7678

9220

9532

9567

9866

9911

9911S

J277

Fluke Corporation

Instrument ModelHEWLETI PACKARD 3458A DIGITAL MULTIMETER

HEWLETI PACKARD 3458A DIGITAL MULTIMETER

FLUKE Y5020 CURRENT SHUNT/SL013

FLUKE Y5020 CURRENT SHUNT/SL013

KROHN-HITE CORP 6900B DISTORTION ANAL YZER

OMEGA TJ36ICINI8UI2SMPWM J PROBE

FLUKE 5450A RESISTANCE CALIBRATOR

FLUKE 8842A MUL TIMETER

FLUKE 5450A RESISTANCE CALIBRATOR

FLUKE 8520AIPRT DIGITAL MULTIMETER W/PRT

FLUKE PM6680B TIMER/COUNTER

FLUKE 5500A CALIBRATOR

FLUKE PM6304C PROGRAMMABLE RCL METER

FLUKE 5790A AC MEASUREMENT STANDARD

FLUKE 5790A/A40 SET ACI MEASUREMENT SYSTEM

CLARKE-HESS 6000 PHASE METER

Cal Date08 March 2011

II February 2011

27 October 2010

07 December 2010

08 September 2010

20 December 2010

14 March 2011

25 September 2010

25 January 2011

19 October 2010

31 August 2010

08 October 2010

26 January 2011

21 January 2011

09 March 2011

04 August 2010

Telephone Facsimile Internet

1420 75th Street SW, Everett WA 98203 USA www.fluke.com888.993.5853 425.446.6390

CalDue08 September 2011

11 November 2011

27 October 2011

07 June 2011

08 September 2011

20 December 2011

14 July 2011

25 September 2011

25 April 2011

19 April 2011

31 August 2011

08 October 2011

26 July 2011

21 July 2011

09 April2011

04 August 2011

Page 2 of 15

Rev 7.4, 02/23/11

FLUKE@ Certificate Number: Calibration Date:945162-6265013:1301616947 l-Apr-ll

Calibration ResultsManufacturer's

Nominal Measured Measurement Specifications

Function/Range Value Value Uncertainty Lower Limit Upper Limit

1 Year Specification Limits shown

DC VOLTAGE ACCURACY TEST (NORMAL)

330 mV Range0.0000 mV 0.0000 0.00028 2.2e-007 V -0.00300 0.00300

300.000 mV 300.000 299.9994 1.2e-006 V 299.9790 300.0210

-300.000 mV -300.000 -299.9953 1.2e-006 V -300.0210 -299.9790

3.3 V Range0.000000 V 0.000000 0.0000012 8.4e-007 V -0.0000050 0.0000050

0.300000 V 0.300000 0.3000006 1.2e-006 V 0.2999800 0.3000200

-0.300000 V -0.300000 -0.2999990 1.2e-006 V -0.3000200 -0.2999800

3.000000 V 3.000000 2.9999951 9.0e-006 V 2.9998450 3.0001550

-3.000000 V -3.000000 -3.0000256 9.0e-006 V -3.0001550 -2.9998450

33 V Range0.00000 V 0.00000 0.000007 1.0e-005 V -0.000050 0.000050

30.00000 V 30.00000 29.999971 1.7e-004 V 29.998450 30.001550

-30.00000 V -30.00000 -30.000263 1.7e-004 V -30.001550 -29.998450

330 V Range30.0000 V 30.0000 30.00007 1.7e-004 V 29.99785 30.00215

-30.0000 V -30.0000 -29.99998 1.7e-004 V -30.00215 -29.99785

300.0000 V 300.0000 300.00001 1.8e-003 V 299.98300 300.01700

-300.0000 V -300.0000 -300.00304 1.8e-003 V -300.01700 -299.98300

1000 V Range100.0000 V 100.0000 100.00006 5.6e-004 V 99.99300 100.00700

-100.0000 V -100.0000 -99.99994 5.5e-004 V -100.00700 -99.99300

1000.000 V 1000.000 999.9990 1.7e-002 V 999.9435 1000.0565

-1000.000 V -1000.000 -1000.0106 1.7e-002 V -1000.0565 -999.9435

DC VOLTAGE ACCURACY TEST (AUX)

330 mV Range0.000 mV 0.000 -0.0189 3.ge-006 V -0.3500 0.3500

300.000 mV 300.000 299.9883 9.8e-006 V 299.5300 300.4700

-300.000 mV -300.000 -300.0234 9.8e-006 V -300.4700 -299.5300

3.3 V Range0.330000 V 0.330000 0.3299883 5.5e-006 V 0.3295180 0.3304820

-0.330000 V -0.330000 -0.3300170 5.5e-006 V -0.3304820 -O.3295180

3.00000 V 3.00000 3.000018 3.3e-005 V 2.998450 3.001550

-3.00000 V -3.00000 -3.000059 3.3e-005 V -3.001550 -2.998450

DC OFFSET ACCURACYTEST

50 mV Range0.000 mV 0.000 0.0021 4.1e-006 V -0.0330 0.0330

50.0000 mV 50.0000 50.01025 1.1e-006 V 49.91700 50.08300

-50.0000 mV -50.0000 -50.01090 7.8e-007 V -50.08300 -49.91700

500 mV Range

Fluke Corporation Telephone Facsimile Internet Page 3 of 15

1420 75th Street SW, Everett WA 98203 USA 888.993.5853 425.446.6390 www.fIuke.com Rev 7.4, 02/23/11

FL.UKE@ Certificate Number: Calibration Date:

945162-6265013:1301616947 l-Apr-ll

Calibration ResultsManufacturer's

Nominal Measured Measurement Specifications

Function/Range Value Value Uncertainty Lower Limit Upper Limit

0.00 mV 0.00 0.025 6.6e-005 V -0.330 0.330

500.00 mV 500.00 500.150 3.2e-005 V 499.170 500.830

-500.00 mV -500.00 -500.206 2.ge-005 V -500.830 -499.170

5 V Range0.0000 V 0.0000 0.00050 7.6e-004 V -0.00330 0.00330

5.0000 V 5.0000 5.00233 5.3e-004 V 4.99170 5.00830

-5.0000 V -5.0000 -5.00158 2.1e-004 V -5.00830 -4.99170

50 V Range0.000 V 0.000 -0.0010 1.1e-003 V -0.0330 0.0330

45.0000 V . 45.0000 45.02202 3.0e-004 V 44.92202 45.07798

-45.0000 V -45.0000 -45.01758 3.ge-004 V -45.07798 -44.92202

DC CURRENT ACCURACY TEST

3.3 mA Range0.00000 mA 0.00000 0.000006 1.8e-008 A -0.000050 0;000050

3.00000 mA 3.00000 2.999988 2.ge-008 A 2.999559 3.000441

-3.00000 mA -3.00000 -2.999951 2.ge-008 A -3.000441 -2.999559

33 mA Range0.00000 mA 0.00000 0.000041 7.ge-008 A -0.000250 0.000250

30.0000 mA 30.0000 29.99982 3.2e-007 A 29.99676 30.00324

-30.0000 mA -30.0000 -29.99950 3.2e-007 A -30.00324 -29.99676

330 mA Range0.0000 mA 0.0000 0.00049 1.6e-007 A -0.00330 0.00330

300.000 mA 300.000 299.9977 7.ge-006 A 299.9667 300.0333

-300.000 mA -300.000 -299.9924 7.ge-006 A -300.0333 -299.9667

2.2 A Range0.0000000 A 0.0000000 0.00000p21 2.6e-007 A -0.00004400 0.00004400

2.000000 A 2.000000 2.0000095 1.3e-004 A 1.9993560 2.0006440

-2.000000 A -2.000000 -2.0000249 1.3e-004 A -2.0006440 -1.9993560

11 A Range0.0000000 A 0.0000000 0.00002258 1.5e-006 A -0.00033000 0.00033000

10.0000 A 10.0000 9.99957 6.ge-004 A 9.99367 10.00633

-10.0000 A -10.0000 -10.00000 6.6e-004 A -10.00633 -9.99367

DC POWER ACCURACY TEST (NORMAL)

20 mV @ 2.9 A20.0000 mV 20.0000 20.00061 3.7e-007 V 19.99580 20.00420

20mV@11 A20.0000 mV 20.0000 20.00050 3.7e-007 V 19.99580 20.00420

DC POWER ACCURACY TEST (AUX)

0.1 mA @ 1000 V0.100000 mA 0.100000 0.1000035 2.3e-009 A 0.0999370 0.1000630

1 mA@ 1000 V1.00000 mA 1.00000 0.999993 2.0e-008 A 0.999820 1.000180

FlukeCorporation Telephone Facsimile Internet Page4 of 15

1420 75th Street SW, Everett WA 98203 USA 888.993.5853 425.446.6390 www.fluke.com Rev 7.4, 02/23/11

FL.UKE@ Certificate Number: Calibration Date:945162-6265013:1301616947 l-Apr-ll

Calibration ResultsManufacturer's

Nominal Measured Measurement Specifications

Function/Range Value Value Uncertainty Lower Limit Upper Limit

2A @ 300 V2.00000 A 2.00000 2.000009 1.3e-004 A 1.999356 2.000644

10 A @ 1000 V10.0000 A 10.0000 9.99969 6.6e-004 A 9.99367 10.00633

PHASE ACCURACY TEST (Voltage ys. Voltage)

O degrees

0.00 o @ 60 Hz 0.00 0.054 5.2e-002 o -0.150 0.150

0.00 o @ 400 Hz 0.00 -0.040 5.2e-002 o -0.900 0.900

0.00 o @ 1 kHz 0.00 -0.060 5.2e-002 o -2.000 2.000

0.00 o @ 5 kHz 0.00 -0.370 5.2e-002o -6.000 6.000

0.00 o @ 10 kHz 0.00 -0.030 5.2e-002 o -10.000 10.000

60 degrees

60.00 o @ 60 Hz 60.00 59.982 6.4e-002 o 59.850 60.150

60.00 o @ 400 Hz 60.00 59.970 5.2e-002 o 59.100 60.900

60.00 o @ 1 kHz 60.00 59.930 5.2e-002 o 58.000 62.000

60.00 o @ 5 kHz 60.00 59.618 5.2e-002 o 54.000 66.000

60.00o @ 10kHz 60.00 60.070 5.2e-002o 50.000 70.000

90 degrees

90.00 o @ 60 Hz 90.00 90.030 5.2e-002 o 89.850 90.150

90.00o @400 Hz 90.00 89.960 5.2e-002o 89.100 90.900

90.00o @ 1 kHz 90.00 89.940 5.2e-002o 88.000 92.000

90.00o @5 kHz 90.00 89.630 5.2e-002o 84.000 96.000

90.00o @ 10kHz 90.00 90.080 5.2e-002o 80.000 100.000

PHASE ACCURACY TEST (Voltage ys. Current)

0.00 degrees (33 V, 300 mA @ 65 Hz)

0.00 o @ 65 Hz 0.00 0.030 5.2e-002o -0.150 0.150

0.00 degrees (33 V, 2 A @ 65 Hz)

0.00 degrees (33 V, 2 A @ 65 Hz)

0.00 o @65 Hz 0.00 0.018 5.2e-002o -0.150 0.150

0.00 degrees (33 V, 5 A @ 65 Hz)

0.00 o @ 65 Hz 0.00 -0.004 5.3e-002o -0.150 0.150

0.00 degrees (33 V, 5 A @ 400 Hz)

0.00 o @400 Hz 0.00 -0.070 6.8e-002o -0.900 0.900

FREQUENCY ACCURACY TEST

119.00000 Hz @ 3 V 119.00000 119.000301 4.ge-005 Hz 118.996025 119.003975

120.00000 Hz @ 3 V 120.00000 119.999394 1.8e-005 Hz 119.996000 120.004000

1.00000000 kHz @ 3 V 1.00000000 1.000000484 2.0e-005 Hz 0.999974000 1.000026000

100.0000000 kHz @ 3 V 100.0000000 99.99998690 6.0e-005 Hz 99.99748500 100.00251500

DISTORTION AND NOISE TEST (NORMAL)

300 mV Range

0.000 % @ 10 Hz 0.000 0.0650 0.0000 0.1800 *

0.000 % @ 45 Hz 0.000 0.0205 0.0000 0.0650 *

Fluke Corporation Telephone Facsimile Internet Page 5 of 15

1420 75th Street SW. Everett WA98203 USA 888.993.5853 425.446.6390 www.fIuke.com Rev 7.4,02/23/11

...

FLUKE@ Certifica te Number: Calibration Date:

945162-6265013:1301616947 l-Apr-ll

Calibration ResultsManufacturer's

Nominal Measured Measurement Specifications

Function/Range Value Value Uncertainty Lower Limit Upper Limit

0.000 % @ 10 kHz 0.000 0.0204 0.0000 0.0650 *

0.000 % @ 20 kHz 0.000 0.0350 0.0000 0.0900 *0.000 % @ 50 kHz 0.000 0.0854 0.0000 0.1800 *0.000 % @ 100 kHz 0.000 0.1177 0.0000 0.2300 *

0.000 % @ 500 kHz 0.000 0.1143 0.0000 0.9300 *300 V Range0.000 % @ 45 Hz 0.000 0.0215 0.0000 0.1583 *0.000 % @ 1 kHz 0.000 0.0238 0.0000 0.0583 *0.000 % @ 10 kHz 0.000 0.0275 0.0000 0.0583 *

0.000 % @ 20 kHz 0.000 0.0514 0.0000 0.6083 *

DISTORTION AND NOISE TEST (AUX)

300 mV Range

0.000 % @ 10 Hz 0.000 0.0597 0.0000 0.2667 *

0.000 % @ 20 Hz 0.000 0.0154 0.0000 0.1267 *0.000 % @ 45 Hz 0.000 0.0182 0.0000 0.1267 *0.000 % @ 1 kHz 0.000 0.0256 0.0000 0.1467 *

0.000 % @ 5 kHz 0.000 0.1630 0.0000 0.3667 *

0.000 % @ 10 kHz 0.000 0.4313 0.0000 0.6667 *

3 V Range0.000 % @ 10 Hz 0.000 0.0649 0.0000 0.2067 *0.000 % @ 20 Hz 0.000 0.0148 0.0000 0.0667 *

0.000 % @ 45 Hz 0.000 0.0173 0.0000 0.0667 *0.000 % @ 1 kHz 0.000 0.0271 0.0000 0.0867 *0.000 % @ 5 kHz 0.000 0.0927 0.0000 0.3067 *0.000 % @ 10 kHz 0.000 0.3302 0.0000 0.6067 *

1 Year Specification Limits shown

AC VOLTAGE ACCURACY TEST (NORMAL)

33mV Range

30.00 mV @ 9.5 Hz 30.00 30.153 3.0e-005 V 28.335 31.665

30.000 mV @ 10 Hz 30.000 30.0017 6.8e-006 V 29.8750 30.1250

30.000 mV @ 45 Hz 30.000 29.9993 2.7e-006 V 29.9350 30.0650

30.000 mV @ 1 kHz 30.000 29.9991 2.7e-006 V 29.9350 30.0650

30.000 mV@ 10 kHz 30.000 29.9992 2.7e-006 V 29.9350 30.0650

30.000 mV@20 kHz 30.000 30.0012 2.7e-006 V 29.9200 30.0800

30.000 mV @ 50 kHz 30.000 30.0077 4.8e-006 V 29.9050 30.0950

30.000 mV @ 100 kHz 30.000 30.0139 8.1e-006 V 29.8620 30.1380

30.00 mV@ 500 kHz 30.00 29.988 2.3e-005 V 29.640 30.360

330mV Range300.0 mV @ 9.5 Hz 300.0 300.39 3.0e-004 V 283.35 316.65

300.00 mV @ 10 Hz 300.00 300.001 5.0e-005 V 299.200 300.800

300.000 mV @ 45 Hz 300.000 300.0014 9.1e-006 V 299.8300 300.1700

300.000 mV @ 1 kHz 300.000 300.0006 9.1e-006 V 299.8300 300.1700

Fluke Corporation Telephone Facsimile Internet Page 6 of 15

1420 75th Street SW, Everett WA98203 USA 888.993.5853 425.446.6390 www.fluke.com Rev 7.4,02/23/11

FL-UKE@ Certifica te Number: Calibration Date:945162-6265013:1301616947 l-Apr-ll

Calibration ResultsManufacturer's

Nominal Measured Measurement Specifications

Function/Range Value Value Uncertainty Lower Limit Upper Limit

300.000 mV @ 10 kHz 300.000 299.9930 9.0e-006 V 299.8300 300.1700

300.000 mV @ 20 kHz 300.000 299.9810 9.0e-006 V 299.6800 300.3200

300.00 mV @ 50 kHz 300.00 299.953 1.4e-005 V 299.480 300.520

300.00 mV @ 100 kHz 300.00 299.907 2.1 e-005 V 299.110 300.890

300.00 mV @ 500 kHz 300.00 299.372 7.8e-D05 V 297.570 302.430

3.3V Range

3.000 V @ 9.5 Hz 3.000 3.0035 3.0e-D03 V 2.8335 3.1665

3.0000 V@ 10 Hz 3.0000 2.99995 4.7e-004 V 2.99525 3.00475

3.00000 V @ 45 Hz 3.00000 2.999988 6.0e-005 V 2.999040 3.000960

3.00000 V @ 1 kHz 3.00000 3.000000 6.0e-D05V 2.999040 3.000960

3.00000 V @ 10kHz 3.00000 2.999932 6.0e-005 V 2.999040 3.000960

3.00000 V @ 20 kHz 3.00000 2.999846 6.0e-005 V 2.997540 3.002460

3.0000 V @ 50 kHz 3.0000 2.99964 1.4e-004 V 2.99550 3.00450

3.0000 V @ 100 kHz 3.0000 2.99942 2.1e-004 V 2.99110 3.00890

3.0000 V @ 500 kHz 3.0000 2.99913 9.6e-004 V 2.98170 3.01830

33V Range

30.00 V @ 9.5 Hz 30.00 30.008 3.0e-002 V 28.335 31.665

30.000 V @ 10 Hz 30.000 29.9994 4.7e-003 V 29.9525 30.0475

30.0000 V @ 45 Hz 30.0000 29.99998 7.5e-D04V 29.98740 30.01260

30.0000 V @ 1 kHz 30.0000 29.99844 7.6e-004 V 29.98740 30.01260

30.0000 V @ 10 kHz 30.0000 29.99840 7.5e-004 V 29.98740 30.01260

30.0000 V @ 20 kHz 30.0000 29.99836 7.6e-004 V 29.97340 30.02660

30.000 V @ 50 kHz 30.000 29.9971 1.4e-003 V 29.9380 30.0620

30.000 V @ 100 kHz 30.000 29.9987 2.3e-003 V 29.9110 30.0890

330V Range300.000 V @ 45 Hz 300.000 300.0094 9.7e-003 V 299.8434 300.1566

300.000 V @ 1 kHz 300.000 299.9848 9.7e-003 V 299.8434 300.1566

300.000 V @ 10 kHz 300.000 299.9776 9.7e-003 V 299.7450 300.2550

300.000 V @ 20 kHz 300.000 299.9882 9.7e-003 V 299.6970 300.3030

1000V Range

1000.00 V @ 45 Hz 1000.00 1000.050 3.2e-002 V 999.420 1000.580

1000.00 V @ 1 kHz 1000.00 1000.021 3.2e-002 V 999.420 1000.580

1000.00 V @ 5 kHz 1000.00 999.988 3.2e-002 V 997.900 1002.100

1000.00 V @ 10 kHz 1000.00 999.965 3.2e-002 V 997.500 1002.500

AC VOLTAGE ACCURACY TEST, SQUAREWAVE (NORMAL)

66mV Range60.000 mVpp @ 10Hz 60.000 60.0274 59.5200 60.4800 *60.000 mVpp @ 45 Hz 60.000 60.0036 59.6850 60.3150 *60.000 mVpp @ 1 kHz 60.000 60.0136 59.6850 60.3150 *60.000 mVpp @ 20 kHz 60.000 60.2316 59.5350 60.4650 *60.000 mVpp @ 100 kHz 60.000 61.0792 56.6700 63.3300 *660mV Range

Fluke Corporation Telephone Facsimile Internet Page 7 of 15

1420 75th Street SW, Everett WA 98203 USA 888.993.5853 425.446.6390 www.fluke.com Rev 7.4, 02/23/11

FL.UKE@ Certifica te Number: Calibration Date:945162-6265013:1301616947 l-Apr-ll

Calibration ResultsManufacturer's

Nominal Measured Measurement Specifications

Function/Range Value Value Uncertainty Lower Limit Upper Limit

600.000 mVpp @ 10 Hz 600.000 600.2320 595.2000 604.8000 *600.000 mVpp @ 45 Hz 600.000 600.0240 596.8500 603.1500 *600.000 mVpp @ 1 kHz 600.000 600.1180 596.8500 603.1500 *600.000 mVpp @ 20 kHz 600.000 602.1240 595.3500 604.6500 *600.000 mVpp @ 100 kHz 600.000 610.0640 566.7000 633.3000 *

6.6V Range

6.00000 Vpp @ 10 Hz 6.00000 6.002180 5.952000 6.048000 *6.00000 Vpp @ 45 Hz 6.00000 6.000220 5.968500 6.031500 *6.00000 Vpp @ 1 kHz 6.00000 6.001060 5.968500 6.031500 *6.00000 Vpp @ 20 kHz 6.00000 6.019520 5.953500 6.046500 *6.00000 Vpp @ 100 kHz 6.00000 6.092640 5.667000 6.333000 *

66V Range

60.0000 Vpp @ 10 Hz 60.0000 60.02220 59.52000 60.48000 *

60.0000 Vpp @ 45 Hz 60.0000 60.00220 59.68500 60.31500 *60.0000 Vpp @ 1 kHz 60.0000 60.00940 59.68500 60.31500 *60.0000 Vpp @ 20 kHz 60.0000 60.23540 59.53500 60.46500 *60.0000 Vpp @ 100 kHz 60.0000 61.29180 56.67000 63.33000 *AC VOLTAGE ACCURACY TEST, TRIANGLEWAVE (NORMAL)

93mV Range90.000 mVpp @ 10 Hz 90.000 90.0022 89.3100 90.6900 *90.000 mVpp @ 45 Hz 90.000 90.0040 89.5425 90.4575 *90.000 mVpp @ 1 kHz 90.000 90.0064 89.5425 90.4575 *90.000 mVpp @ 20 kHz 90.000 90.0123 89.3175 90.6825 *90.000 mVpp @ 100 kHz 90.000 90.0691 85.0350 94.9650 *930mV Range900.000 mVpp @ 10 Hz 900.000 899.9287 893.1000 906.9000 *900.000 mVpp @ 45 Hz 900.000 899.9945 895.4250 904.5750 *900.000 mVpp @ 1 kHz 900.000 900.0257 895.4250 904.5750 *900.000 mVpp @ 20 kHz 900.000 899.9599 893.1750 906.8250 *900.000 mVpp @ 100 kHz 900.000 899.9148 850.3500 949.6500 *9.3V Range9.00000 Vpp @ 10 Hz 9.00000 8.999079 8.931000 9.069000 *9.00000 Vpp @ 45 Hz 9.00000 8.999806 8.954250 9.045750 *9.00000 Vpp @ 1 kHz 9.00000 9.000187 8.954250 9.045750 *9.00000 Vpp @ 20 kHz 9.00000 8.999737 8.931750 9.068250 *9.00000 Vpp @ 100 kHz 9.00000 9.00167T 8.503500 9.496500 *93V Range90.0000 Vpp @ 10 Hz 90.0000 89.98940 89.31000 90.69000 *90.0000 Vpp @ 45 Hz 90.0000 89.99876 89.54250 . 90.45750 *90.0000 Vpp @ 1 kHz 90.0000 89.99772 89.54250 90.45750 *90.0000 Vpp @ 20 kHz 90.0000 90.02162 89.31750 90.68250 *90.0000 Vpp @ 100 kHz 90.0000 90.01538 85.03500 94.96500 *

Fluke Corporation Telephone Facsimile Internet Page 8 of 15

1420 75th Street SW, Everett WA 98203 USA 888.993.5853 425.446.6390 www.fluke.com Rev 7.4, 02/23/11

FLUKE@ Certifica te Number: Calibration Date:945162-6265013:1301616947 1-Apr-ll

Calibration ResultsManufacturer's

Nominal Measured Measurement SpecificationsFunction/Range Value Value Uncertainty Lower Limit Upper Limit

AC VOLTAGE ACCURACYTEST, SINEWAVE(AUX)

330mVRange10.000 mV@ 10 Hz 10.000 10.0022 5.1e-006 V 9.6100 10.3900

10.000 mV@ 20 Hz 10.000 10.0037 5.2e-006 V 9.6200 10.3800

10.000 mV@ 45 Hz 10.000 10.0047 5.3e-006 V 9.6200 10.3800

10.000 mV @ 1 kHz 10.000 10.0075 5.3e-006 V 9.6200 10.3800

10.000 mV@ 5 kHz 10.000 10.0203 5.3e-006 V 9.5300 10.4700

10.000 mV @ 10 kHz 10.000 10.0362 5.3e-006 V 9.5100 10.4900

300.0 mV @ 9.5 Hz 300.0 299.71 3.0e-004 V 283.35 316.65

300.00 mV@ 10Hz 300.00 299.945 5.1e-005 V 299.030 300.970

300.00 mV@ 20 Hz 300.00 299.990 2.1e-005 V 299.330 300.670

300.00 mV@ 45 Hz 300.00 299.995 1.2e-005 V 299.330 300.670

300.00 mV@ 1 kHz 300.00 300.036 1.2e-005 V 299.330 300.670

300.00 mV@ 5 kHz 300.00 299.973 1.2e-005 V 298.950 301.050

300.00 mV @ 10 kHz 300.00 299.927 1.4e-005 V 298.350 301.650

3.3V Range3.000 V @ 9.5 Hz 3.000 3.0042 3.0e-003 V 2.8335 3.1665

3.0000 V @ 10 Hz 3.0000 2.99947 5.1e-004 V 2.99355 3.00645

3.0000 V @ 20 Hz 3.0000 2.99994 2.1e-004 V 2.99655 3.00345

3.0000 V @ 45 Hz 3.0000 3.00000 1.2e-004 V 2.99685 3.00315

3.0000 V @ 1 kHz 3.0000 3.00023 1.2e-004 V 2.99685 3.00315

3.0000 V @ 5 kHz 3.0000 2.99994 1.2e-004 V 2.99260 3.00740

3.0000 V @ 10 kHz 3.0000 2.99991 1.2e-004 V 2.98660 3.01340

AC VOLTAGEACCURACYTEST, SQUAREWAVE(AUX)

660mV Range600.000 mVpp@ 10 Hz 600.000 600.0640 595.2000 604.8000 *

600.000 mVpp @ 45 Hz 600.000 600.0960 596.8500 603.1500 *

600.000mVpp@ 1 kHz 600.000 602.2120 596.8500 603.1500 *

600.000 mVpp @ 10 kHz 600.000 621.4640 566.7000 633.3000 *

6.6V Range6.00000 Vpp @ 10Hz 6.00000 6.000600 5.952000 6.048000 *

6.00000 Vpp @ 45 Hz 6.00000 6.001080 5.968500 6.031500 *

6.00000 Vpp @ 1 kHz 6.00000 6.024500 5.968500 6.031500*6.00000Vpp@ 10kHz 6.00000 6.236860 5.667000 6.333000*AC VOLTAGEACCURACYTEST,TRIANGLEWAVE (AUX)930mV Range

900.000 mVpp@ 10 Hz 900.000 899.8005 893.1000 906.9000 *

900.000 mVpp@ 45 Hz 900.000 899.9980 895.4250 904.5750 *

900.000mVpp@ 1 kHz 900.000 899.9564 895.4250 904.5750 *

900.000 mVpp@ 10 kHz 900.000 890.5375 850.3500 949.6500*9.3VRange9.00000Vpp@ 10Hz 9.00000 8.997797 8.931000 9.069000*

Fluke Corporation Telephone Facsirnile Internet Page 9 of 15-1420 75th Street SW, Everett WA 98203 USA 888.993.5853 425.446.6390 www.fIuke.com Rev 7.4,02/23/11

FLUKE@ Certifica te Number: Calibration Date:945162-6265013:1301616947 l-Apr-ll

Calibration ResultsManufacturer's

Nominal Measured Measurement Specifications

Function/Range Value Value Uncertainty Lower Limit Upper Limit

9.00000 Vpp @ 45 Hz 9.00000 8.999945 8.954250 9.045750 *

9.00000 Vpp @ 1 kHz 9.00000 8.998629 8.954250 9.045750 *

9.00000 Vpp @ 10kHz 9.00000 8.902430 8.503500 9.496500 *

HARMONIC AMPLlTUDE ACCURACY TEST (NORMAL)

33mV Range

30.000 mV @ 1 kHz 30.000 30.0003 29.9150 30.0850 *

30.000 mV @ 5 kHz 30.000 30.0004 29.9150 30.0850 *

30.000 mV @ 10 kHz 30.000 30.0004 29.9150 30.0850 *

330mV Range

300.000 mV @ 1 kHz 300.000 300.0030 299.8101 300.1899 *

300.000 mV @ 5 kHz 300.000 300.0000 299.8101 300.1899 *

300.000 mV @ 10 kHz 300.000 299.9920 299.8101 300.1899 *

3.3V Range

3.00000 V @ 1 kHz 3.00000 3.000010 2.998980 3.001020 *

3.00000 V @ 5 kHz 3.00000 2.999980 2.998980 3.001020 *

3.00000 V @ 10kHz 3.00000 2.999940 2.998980 3.001020 *

33V Range

30.0000 V @ 1 kHz 30.0000 29.99860 29.98680 30.01320 *

30.0000 V @ 5 kHz 30.0000 29.99850 29.98680 30.01320 *

30.0000 V @ 10 kHz 30.0000 29.99830 29.98680 30.01320 *

330V Range

300.000 V @ 1 kHz 300.000 299.9860 299.8368 300.1632 *300.000 V @ 5 kHz 300.000 299.9770 299.7300 300.2700 *300.000 V @ 10 kHz 300.000 299.9770 299.7300 300.2700 *

1000V Range

1000.00 V @ 1 kHz 1000.00 1000.022 999.340 1000.660 *

1000.00 V @ 5 kHz 1000.00 999.982 997.800 1002.200 *1000.00 V @ 10 kHz 1000.00 999.958 997.000 1003.000 *HARMONIC AMPLlTUDE ACCURACY TEST (AUX)

330mV Range300.000 mV @ 1 kHz 300.000 300.0460 298.9600 301.0400 *300.000 mV @ 5 kHz 300.000 299.9830 298.5000 301.5000 *300.000 mV @ 10kHz 300.000 299.9340 297.9000 302.1000 *3.3V Range

3.00000 V @ 1 kHz 3.00000 3.000280 2.996400 3.003600 *3.00000 V @ 5 kHz 3.00000 3.000020 2.991200 3.008800 *3.00000 V @ 10 kHz 3.00000 2.999990 2.985200 3.014800 *AC CURRENT ACCURACY TEST

330pA Range300.00 IJA@ 10 Hz 300.00 299.981 3.0e-008 A 299.100 300.900

300.00 IJA@ 20 Hz 300.00 299.993 3.0e-008 A 299.475 300.525

300.00 IJA@ 45 Hz 300.00 299.995 2.3e-008 A 299.475 300.525

FlukeCorporation Telephone Facsimile Internet Page 10 of 15

1420 75th Street SW, Everett WA 98203 USA 888.993.5853 425.446.6390 www.fluke.com Rev 7.4, 02/23/H

FL.UKE@ Certificate Number: Calibration Date:

945162-6265013:1301616947 l-Apr-ll

Calibration ResultsManufacturer's

Nominal Measured Measurement Specifications

Function/Range Value Value Uncertainty Lower Limit Upper Limit

300.00 IJA@ 1 kHz 300.00 300.032 2.3e-008 A 299.375 300.625

300.00 IJA@ 5 kHz 300.00 300.013 6.1 e-008 A 298.650 301.350

300.00 IJA@ 10 kHz 300.00 300.092 7.7e-008 A 296.100 303.900

3.3mA Range

3.0000 mA @ 10 Hz 3.0000 2.99984 3.0e-007 A 2.99370 3.00630

3.0000 mA @ 20 Hz 3.0000 2.99999 3.0e-007 A 2.99670 3.00330

3.0000 mA @ 45 Hz 3.0000 2.99999 2.0e-007 A 2.99670 3.00330

3.0000 mA @ 1 kHz 3.0000 3.00043 2.0e-007 A 2.99670 3.00330

3.0000 mA @ 5 kHz 3.0000 3.00001 2.6e-007 A 2.99370 3.00630

3.0000 mA @ 10kHz 3.0000 3.00009 3.0e-007 A 2.98170 3.01830

33mA Range30.000 mA @ 10 Hz 30.000 30.0001 4.5e-006 A 29.9370 30.0630

30.000 mA @ 20 Hz 30.000 30.0009 4.5e-006 A 29.9670 30.0330

30.000 mA @ 45 Hz 30.000 30.0013 2.ge-006 A 29.9700 30.0300

30.000 mA@ 1 kHz 30.000 30.0039 3.0e-006 A 29.9700 30.0300

30.000 mA @ 5 kHz 30.000 30.0037 , 3.0e-006 A 29.9370 30.0630

30.000 mA @ 10 kHz 30.000 30.0014 3.0e-006 A 29.8170 30.1830

330mA Range

300.00 mA @ 10 Hz 300.00 299.875 3.ge-005 A 299.370 300.630

300.00 mA @ 20 Hz 300.00 299.893 4.0e-005 A 299.670 300.330

300.00 mA @ 45 Hz 300.00 299.895 2.0e-005 A 299.700 300.300

300.00 mA @ 1 kHz 300.00 299.927 2.0e-005 A 299.700 300.300

300.00 mA @ 5 kHz 300.00 299.920 2.0e-005 A 299.370 300.630

300.00 mA @ 10 kHz 300.00 299.919 2.0e-005 A 298.170 301.830

2.2A Range

2.0000 A @ 10 Hz 2.0000 1.99994 2.6e-004 A 1.99570 2.00430

2.0000 A @ 45 Hz 2.0000 2.00006 1.6e-004 A 1.99770 2.00230

2.0000 A @ 1 kHz 2.0000 2.00005 1.6e-004 A 1.99770 2.00230

2.0000 A @ 5 kHz 2.0000 1.99989 1.6e-004 A 1.98470 2.01530

11A Range

10.000 A @ 45 Hz 10.000 10.0003 1.6e-003 A 9.9920 10.0080

10.000 A @ 65 Hz 10.000 10.0002 1.6e-003 A 9.9920 10.0080

10.000 A@ 500 Hz 10.000 10.0001 1.1e-003 A 9.9880 10.0120

10.000 A @ 1 kHz 10.000 9.9997 1.0e-003 A 9.9650 10.0350

AC POWER ACCURACY TEST (High Voltage)

3mA, 1000V (O° Phase)

3.0000 mA @ 65 Hz 3.0000 2.99997 2.0e-007 A 2.99670 3.00330

3mA, 1000V (90° Phase)

3.0000 mA @ 65 Hz 3.0000 3.00039 2.0e-007 A 2.99670 3.00330

30mA, 1000V (O° Phase)

30.000 mA @ 500 Hz: 30.000 29.9994 3.0e-006 A 29.9700 30.0300

30mA, 1000V (90° Phase)

FlukeCorporation Telephone Facsimile Internet Page 11 of 15

1420 75th Street SW, Everett WA 98203 USA 888.993.5853 425.446.6390 www.fluke.com Rev 7.4,02/23/11

FL.UKE@ Certificate Number: Calibration Date:945162-6265013:1301616947 1-Apr-ll

Calibration ResultsManufacturer's

Nominal Measured Measurement SpecificationsFunction/Range Value Value Uncertainty Lower Limit Upper Limit

30.000 mA @ 500 Hz 30.000 30.0017 3.0e-006 A 29.9700 30.0300

30mA, 1000V (00 Phase)

30mA, 1000V (00 Phase)

30.000 mA @ 1 kHz 30.000 30.0041 3.0e-006 A 29.9700 30.0300

30.000 mA @ 5 kHz 30.000 30.0016 3.0e-006 A 29.9370 30.0630

30.000 mA @ 10 kHz 30.000 30.0034 3.0e-006 A 29.8170 30.1830

AC POWER ACCURACY TEST (High Current)

33mV, 10A (00 Phase)

33.000 mV @ 65 Hz 33.000 33.0033 3.0e-006 V 32.9635 33.0365

33mV, 10A (900 Phase)

33.000 mV @ 65 Hz 33.000 33.0225 3.0e-006 V 32.9635 33.0365

33mV, 10A (00 Phase)33.000 mV @ 500 Hz 33.000 33.0117 3.0e-006 V 32.9635 33.0365

330mV, 10A (00 Phase)

0.33000 V @ 1 kHz 0.33000 0.329984 1.0e-005 V 0.329841 0.330159

3.3V, 2A (00 Phase)

3.30000 V @ 5 kHz 3.30000 3.299714 6.7e-005 V 3.298079 3.301921

3;3V, 300mA (00 Phase)

3.30000 V @ 10kHz 3.30000 3.299704 6.8e-005 V 3.298079 3.301921

AC POWER ACCURACYTEST (High Power)300V, 2A (00 Phase)

300.000 V @ 5 kHz 300.000 299.9736 9.7e-003 V 299.7450 300.2550

300V, 2A (900 Phase)300.000 V @ 5 kHz 300.000 299.9802 9.7e-003 V 299.7450 300.2550

1000V, 10A (00 Phase)

1000.00 V @ 1 kHz 1000.00 1000.012 3.2e-002 V 999.420 1000.580

1000V, 10A (900 Phase)

1000.00 V @ 1 kHz 1000.00 1000.028 3.2e-002 V 999.420 1000.580

1 Year Specification Limits shown

RESISTANCE 2W COMP TEST

2W Comp Test PASS*

RESISTANCE ACCURACY TEST

11 Ohm RangeO.OOOOOOhm 0.00000 0.000093 4.1e-0050hm -0.006000 0.006000

1.00000 Ohm 1.00000 1.000168 3.1e-005 Ohm 0.993880 1.006120

1.90000 Ohm 1.90000 1.900203 5.6e-005 Ohm 1.893772 1.906228

10.00000hm 10.0000 10.00057 1.2e-004 Ohm 9.99280 10.00720

33 Ohm Range11.90000 Ohm 11.90000 11.901017 1.4e-004 Ohm 11.888572 11.911428

19.00000 Ohm 19.00000 19.001655 2.2e-004 Ohm 18.987720 19.012280

Fluke Corporation Telephone Facsimile Internet Page 12 of 15-1420 75th Street SW, Everett WA 98203 USA 888.993.5853 425.446.6390 www.fluke.com Rev 7.4,02/23/11

FLUKE@ Certifica te Number: Calibration Date:945162-6265013:1301616947 l-Apr-ll

Calibration ResultsManufacturer's

Nominal Measured Measurement Specifications

Function/Range Value Value Uncertainty Lower Limit Upper Limit

30.0000 Ohm 30.0000 30.00237 3.5e-004 Ohm 29.98640 30.01360

110 Ohm Range33.0000 Ohm 33.0000 33.00197 3.8e-0040hm 32.98703 33.01297

100.00000hm 100.0000 100.00606 5.ge-0040hm 99.98100 100.01900

330 Ohm Range119.00000hm 119.0000 119.00427 7.0e-004 Ohm 118.97929 119.02071

190.0000hm 190.000 190.0075 1.1e-003 Ohm 189.9729 190.0271

300.000 Ohm 300.000 300.0098 2.1e-003 Ohm 299.9630 300.0370

1.1 kOhm Range0.330000 kOhm 0.330000 0.3300059 2.4e-003 Ohm 0.3299103 0.3300897

1.000000 kOhm 1.000000 1.0000191 5.8e-003 Ohm 0.9998500 1.0001500

3.3 kOhm Range1.190000 kOhm 1.190000 1.1900204 1.0e-002 Ohm 1.1898329 1.1901671

1.90000 kOhm 1.90000 1.900061 1.6e-002 Ohm 1.899769 1.900231

3.00000 kOhm 3.00000 3.000141 3.4e-002 Ohm 2.999670 3.000330

11 kOhm Range3.30000 kOhm 3.30000 3.300092 3.5e-002 Ohm 3.299103 3.300897

10.00000 kOhm 10.00000 10.000246 9.6e-002 Ohm 9.998500 10.001500

33 kOhm Range11.9000 kOhm 11.9000 11.90035 1.2e-001 Ohm 11.89833 11.90167

19.0000 kOhm 19.0000 19.00065 2.2e-001 Ohm 18.99769 19.00231

30.0000 kOhm 30.0000 30.00083 3.3e-001 Ohm 29.99670 30.00330

110 kOhm Range33.0000 kOhm 33.0000 33.00027 3.5e-001 Ohm 32.99037 33.00963

100.0000 kOhm 100.0000 100.00085 9.5e-001 Ohm 99.98300 100.01700

330 kOhm Range119.0000 kOhm 119.0000 119.00090 1.0e+000 Ohm 118.97972 119.02028

190.0000 kOhm 190.0000 190.00223 1.6e+000 Ohm 189.97120 190.02880

300.000 kOhm 300.000 300.0019 2.5e+000 Ohm 299.9580 300.0420

1.1 MOhm Range0.330000 MOhm 0.330000 0.3300014 2.8e+000 Ohm 0.3298955 0.3301045

1.000000 MOhm 1.000000 1.0000034 9.3e+000 Ohm 0.9997950 1.0002050

3.3 MOhm Range1.19000 MOhm 1.19000 1.190039 1.5e+001 Ohm 1.189767 1.190233

1.90000 MOhm 1.90000 1.900040 2.4e+001 Ohm 1.899660 1.900340

3.00000 MOhm 3.00000 3.000034 3.5e+001 Ohm 2.999495 3.000505

11 MOhm Range3.30000 MOhm 3.30000 3.300068 9.2e+001 Ohm 3.297470 3.302530

10.00000 MOhm 10.00000 10.000031 6.3e+002 Ohm 9.993450 10.006550

33 MOhm Range11.90000 MOhm 11.90000 11.899956 7.3e+0020hm 11.887550 11.912450

19.00000 MOhm 19.00000 18.999556 1.5e+003 Ohm 18.980450 19.019550

Fluke Corporation Telephone Facsimile Internet Page 13 of 15

1420 75th Street SW, EverettWA 98203 USA 888.993.5853 425.446.6390 www.fluke.com Rev 7.4, 02123/11

FLUKE@ Certifica te Number: Calibration Date:945162-6265013:1301616947 l-Apr-ll

Calibration ResultsManufacturer's

Nominal Measured Measurement Specifications

Function/Range Value Value Uncertainty Lower Limit Upper Limit

30.00000 MOhm 30.00000 29.999078 2.2e+003 Ohm 29.969450 30.030550

110MOhm Range33.00000 MOhm 33.00000 32.998928 2.8e+003 Ohm 32.829500 33.170500

100.00000 MOhm 100.00000 100.021295 2.7e+004 Ohm 99.494500 100.505500

330MOhm Range119.00000 MOhm 119.00000 118.965854 8.7e+0040hm 118.388500 119.611500

300.00000 MOhm 300.00000 300.047426 1.0e+005 Ohm 298.483500 301.516500

CAPACITANCE ACCURACY TEST

500pF Range

350 pF @ 1 kHz 350 350.2 2.ge-012 F 338.3 361.7

480 pF @ 1 kHz 480 480.6 2.2e-012 F 467.6 492.4

1.1nF Range

0.600 nF @ 1 kHz 0.600 0.6005 2.0e-012 F 0.5870 0.6130

1.000 nF @ 1 kHz 1.000 1.0008 1.8e-012 F 0.9850 1.0150

3.3nF Range1.200 nF @ 1 kHz 1.200 1.2008 1.7e-012 F 1.1840 1.2160

3.000 nF @ 1 kHz 3.000 3.0009 3.3e-012 F 2.9750 3.0250

11nF Range

3.300 nF @ 1 kHz 3.300 3.3007 4.3e-012 F 3.2735 3.3265

10.900 nF @ 1 kHz 10.900 10.9013 6.6e-012 F 10.8355 10.9645

33nF Range

12.000 nF @ 1 kHz 12.000 12.0070 7.2e-012 F 11.8700 12.1300

30.00 nF @ 1 kHz 30.00 30.007 1.8e-011 F 29.825 30.175

110nF Range

33.00 nF @ 1 kHz 33.00 33.007 2.7e-011 F 32.818 33.182

109.00 nF @ 1 kHz 109.00 109.020 6.6e-011 F 108.627 109.373

330nF Range

120.00 nF @ 1 kHz 120.00 120.017 7.3e-011 F 119.400 120.600

300.0 nF @ 1 kHz 300.0 300.03 1.8e-010 F 298.95 301.05

1.1uF Range

0.3300 F @ 100 Hz 0.3300 0.33003 2.7e-010 F 0.32818 0.33182

1.0900 F @ 100 Hz 1.0900 1.09007 6.6e-010 F 1.08627 1.09373

3.3uF Range1.2000 F @ 100 Hz 1.2000 1.20037 7.3e-010 F 1.19280 1.20720

3.000 F @ 100 Hz 3.000 3.0009 1.ge-009 F 2.9865 3.0135

11uF Range

3.300 F @ 100 Hz 3.300 3.3009 2.7e-009 F 3.2785 3.3215

10.9000 F @ 100 Hz 10.9000 10.90233 8.ge-009 F 10.85185 10.94815

33uF Range12.000 F @ 100 Hz 12.000 12.0033 9.7e-009 F 11.9220 12.0780

30.00 F @ 100 Hz 30.00 30.005 2.4e-008 F 29.850 30.150

110uF Range

Fluke Corporation Telephone Facsimile Internet Page 14 of 15

1420 75th Street SW, Everett WA 98203 USA 888.993.5853 425.446.6390 www.fluke.com Rev 7.4, 02/23/11

FLUKE@ Certifica te Number:945162-6265013:1301616947

Calibration Date:

l-Apr-ll

Calibration Results

Function/RangeNominal

ValueMeasured

Value

33.00 IJF@ 100 Hz

109.00 IJF@ 100 Hz

330uF Range120.0 IJF@ 100 Hz

300.0 IJF@ 50 Hz

1100uF Range

330.0 IJF@ 50 Hz 330.0 330.05

800.0 IJF@ 50 Hz 800.0 800.04

THERMOCOUPLE ("J" Type) MEASUREMENT ACCURACY TEST25.030 °C 25.030 24.96

THERMOCOUPLE SOURCING ACCURACY TEST (10uVlOC)0.00000 mV 0.00000 -0.000053

1.0000 mV 1.0000 0.99992

-1.0000 mV -1.0000 -1.00004

10.0000 mV 10.0000 9.99984

-10.0000 mV -10.0000 -9.99992

100.0000 mV 100.0000 99.99925

-100.0000 mV -100.0000 -99.99827

THERMOCOUPLE MEASUREMENT ACCURACY TEST (10uV/°C)0.000 mV 0.000 0.0005

100.000 mV 100.000 100.0026

-100.000 mV -100.000 -100.0003

33.00

109.00

33.001

109.007

120.0

300.0

120.13

300.00

MeasurementUncertainty

2.7e-008F1.2e-007F

1.3e-007 F

3.3e-007 F

4.3e-007 F

2.6e-006 F

4.0e-002 °C

1.Oe-007V

1.6e-006 V

1.6e-006 V

3.1e-007 V

3.0e-007 V

8.0e-007 V

8.1e-007 V

2.5e-007 V

6.8e-007 V

6.8e-007 V

Manufacturer'sSpecifications

Lower Limit Upper Limit

End of Report

Fluke Corporation Telephone

1420 75th Street SW, Everett WA 98203 USA 888.993.5853

Facsimile

425.446.6390

Internet Page 15 of 15

Rev 7.4,02/23/11www.fluke.com

32.735 33.265

108.355 109.645

118.86 121.14

297.60 302.40

326.40 333.60

791.70 808.30

24.89 25.17

-0.003000 0.003000

0.99694 1.00306

-1.00306 -0.99694

9.99640 10.00360

-10.00360 -9.99640

99.99100 100.00900

-100.00900 -99.99100

-0.0030 0.0030

99.9910 100.0090

-100.0090 -99.9910