COUPLING / DECOUPLING NETWORK (CDN) CDN S TYPE CDN S150, S200, S250, S400, S900 Used for screened cables Up to 25 lines Designed for IEC/EN 61000-4-6 Types with D-Sub sockets Performance up to 230 MHz IEC/EN 61000-4-6 specifies the design and performance of a range of coupling/decoupling networks (CDNs). Each CDN is specific to the type of cable and the intended signal carried on the cable. Teseq offers an extensive range of CDNs which fully comply with the requirements of the standard and provide a simple and reliable method of injecting RF energy into the equipment under test (EUT). Whilst the published standard only specifies the performance of CDN from 150 kHz to 80 MHz, most of Teseq CDNs are specially designed to give the same performance to a wider frequency range. This is required by some product specific standards. Each CDN is also useable for emission testing. Special types meet the requirements of CISPR 15 or can be used as impedance stabilization network (ISN) defined in CISPR 22. Technical specifications CDN S150, S200, S250, S400 and S900 CDN S900 Frequency range: 150 kHz to 230 MHz Power rating (EUT- and AE port) AC max. voltage (line to ground): 150 V (CDN S400: 34 V) DC max. voltage (line to ground): 150 V (CDN S400: 34 V) Current max : 250 mA Test voltage: 500 V DC, 2 sec Common mode impedance (EUT port) 150 kHz to 26 MHz: 150 Ω ±20 Ω 26 MHz to 230 MHz: 150 Ω +60 Ω / -45 Ω Coupling path (In/Out port/EUT) Connection: BNC 50 Ω RF voltage: <20 V Voltage division factor (RF input to EUT port) 150 kHz to 80 MHz: 9.5 dB ±1 dB 80 MHz to 230 MHz: 9.5 dB +3 dB / -2 dB Transmission bandwidth (wanted signal) EUT/AE B3 dB:* > 20 kHz sin. Decoupling of CM disturbance (RF port / AE) 150 kHz: >60 dB 1.5 MHz: >60 dB 30 MHz: >50 dB 230 MHz: >40 dB Mechanical specifications Size (W x H x D): 100 mm x 100 mm x 240 mm Weight: approx. 1.5 kg

Upload: others

Post on 01-Feb-2021

13 views

Category:

Documents


0 download

TRANSCRIPT

  • Coupling / deCoupling network (Cdn) Cdn S type

    Cdn S150, S200, S250, S400, S900

    Used for screened cables

    Up to 25 lines

    Designed for IEC/EN 61000-4-6

    Types with D-Sub sockets

    Performance up to 230 MHz

    IEC/EN 61000-4-6 specifies the design and performance of a range of coupling/decoupling networks (CDNs). Each CDN is specific to the type of cable and the intended signal carried on the cable. Teseq offers an extensive range of CDNs which fully comply with the requirements of the standard and provide a simple and reliable method of injecting RF energy into the equipment under test (EUT). Whilst the published standard only specifies the performance of CDN from 150 kHz to 80 MHz, most of Teseq CDNs are specially designed to give the same performance to a wider frequency range. This is required by some product specific standards.Each CDN is also useable for emission testing. Special types meet the requirements of CISPR 15 or can be used as impedance stabilization network (ISN) defined in CISPR 22.

    Technical specifications CDN S150, S200, S250, S400 and S900

    CDN S900

    Frequency range: 150 kHz to 230 MHz Power rating (EUT- and AE port) AC max. voltage (line to ground): 150 V (CDN S400: 34 V) DC max. voltage (line to ground): 150 V (CDN S400: 34 V) Current max : 250 mA Test voltage: 500 V DC, 2 secCommon mode impedance (EUT port) 150 kHz to 26 MHz: 150 Ω ±20 Ω 26 MHz to 230 MHz: 150 Ω +60 Ω / -45 ΩCoupling path (In/Out port/EUT) Connection: BNC 50 Ω RF voltage: 20 kHz sin.Decoupling of CM disturbance (RF port / AE) 150 kHz: >60 dB1.5 MHz: >60 dB30 MHz: >50 dB230 MHz: >40 dB

    Mechanical specifications

    Size (W x H x D): 100 mm x 100 mm x 240 mmWeight: approx. 1.5 kg

  • Coupling / deCoupling network (Cdn) Cdn S type

    Cdn S150, S200, S250, S400, S900

    Product name

    Application Connector EUT port

    Connector AE port

    CDN S200 2 wires, screened, XLR-Audio, pin 3 to ground XLR, 3 pins XLR, 3 pins

    CDN S400 4 wires, screened, DIN-Audio, 5 pins, pin 2 to ground DIN, 5 pins DIN, 5 pins

    CDN S900 9 wires, screened data line, RS232 D-Sub 9 pins D-Sub 9 pins

    CDN S150 15 wires, screened data line D-Sub 15 pins D-Sub 15 pins

    CDN S250 25 wires, screened data line D-Sub 25 pins D-Sub 25 pins

    AE-side EUT-side

    CDN S200

    100

    150/50 Ω adapterCAL S200

    RF out

    Amp in

    < +10 dBm

    Amp outch.1 < +27 dBm

    ch.2 < +20 dBm

    ch.3 < +20 dBmPower meter

    NSG 4070

    PowerUSB

    Tuning

    StSizeStSize StSize

    Local

    Back StopRun

    Hold

    0 .

    1

    4

    7 8

    5

    2 3

    kHzdBm6

    9 MHzdBµV

    HzV

    Enter

    Step1

    STO

    FRQ LVL

    RCL

    Step2

    Step3

    MOD

    2nd

    Help

    RFON/OFF

    Ground plane

    6 dBAttenuator

    No load oradapter required

    Power meter with input impedance 50 Ω

    Test generator e.g. Teseq NSG 4070with built-in RF synthesizer, brodband power amplifier and power meter

    Test set-up calibration with CDN S200

    Connector CDN S200

    AE-side EUT-side

    CDN S400

    100

    150/50 Ω adapterCAL S400

    RF out

    Amp in

    < +10 dBm

    Amp outch.1 < +27 dBm

    ch.2 < +20 dBm

    ch.3 < +20 dBmPower meter

    NSG 4070

    PowerUSB

    Tuning

    StSizeStSize StSize

    Local

    Back StopRun

    Hold

    0 .

    1

    4

    7 8

    5

    2 3

    kHzdBm6

    9 MHzdBµV

    HzV

    Enter

    Step1

    STO

    FRQ LVL

    RCL

    Step2

    Step3

    MOD

    2nd

    Help

    RFON/OFF

    Ground plane

    6 dBAttenuator

    No load oradapter required

    Power meter with input impedance 50 Ω

    Test generator e.g. Teseq NSG 4070with built-in RF synthesizer, brodband power amplifier and power meter

    Test set-up calibration with CDN S400

    Connector CDN S400

  • AE-side EUT-side

    CDN S150

    100

    150/50 Ω adapterCAL S150

    RF out

    Amp in

    < +10 dBm

    Amp outch.1 < +27 dBm

    ch.2 < +20 dBm

    ch.3 < +20 dBmPower meter

    NSG 4070

    PowerUSB

    Tuning

    StSizeStSize StSize

    Local

    Back StopRun

    Hold

    0 .

    1

    4

    7 8

    5

    2 3

    kHzdBm6

    9 MHzdBµV

    HzV

    Enter

    Step1

    STO

    FRQ LVL

    RCL

    Step2

    Step3

    MOD

    2nd

    Help

    RFON/OFF

    Ground plane

    6 dBAttenuator

    No load oradapter required

    Power meter with input impedance 50 Ω

    Test generator e.g. Teseq NSG 4070with built-in RF synthesizer, brodband power amplifier and power meter

    Test set-up calibration with CDN S150

    Coupling / deCoupling network (Cdn) Cdn S type

    Cdn S150, S200, S250, S400, S900

    Connector CDN S150

    AE-side EUT-side

    CDN S900

    100

    150/50 Ω adapterCAL S900

    RF out

    Amp in

    < +10 dBm

    Amp outch.1 < +27 dBm

    ch.2 < +20 dBm

    ch.3 < +20 dBmPower meter

    NSG 4070

    PowerUSB

    Tuning

    StSizeStSize StSize

    Local

    Back StopRun

    Hold

    0 .

    1

    4

    7 8

    5

    2 3

    kHzdBm6

    9 MHzdBµV

    HzV

    Enter

    Step1

    STO

    FRQ LVL

    RCL

    Step2

    Step3

    MOD

    2nd

    Help

    RFON/OFF

    Ground plane

    6 dBAttenuator

    No load oradapter required

    Power meter with input impedance 50 Ω

    Test generator e.g. Teseq NSG 4070with built-in RF synthesizer, brodband power amplifier and power meter

    Test set-up calibration with CDN S900

    Connector CDN S900

    AE-side EUT-side

    CDN S250

    100

    150/50 Ω adapterCAL S250

    RF out

    Amp in

    < +10 dBm

    Amp outch.1 < +27 dBm

    ch.2 < +20 dBm

    ch.3 < +20 dBmPower meter

    NSG 4070

    PowerUSB

    Tuning

    StSizeStSize StSize

    Local

    Back StopRun

    Hold

    0 .

    1

    4

    7 8

    5

    2 3

    kHzdBm6

    9 MHzdBµV

    HzV

    Enter

    Step1

    STO

    FRQ LVL

    RCL

    Step2

    Step3

    MOD

    2nd

    Help

    RFON/OFF

    Ground plane

    6 dBAttenuator

    No load oradapter required

    Power meter with input impedance 50 Ω

    Test generator e.g. Teseq NSG 4070with built-in RF synthesizer, brodband power amplifier and power meter

    Test set-up calibration with CDN S250

    Connector CDN S250

  • Teseq GmbH Landsberger Str. 255 · 12623 Berlin · Germany T + 49 30 56 59 88 35 F + 49 30 56 59 88 [email protected] www.teseq.com

    82-231008 E01 Oct. 2008

    Delivery information

    Part number Description

    231008 CDN S150 CDN S15, 230MHz (D-sub)244147 CDN S150S CDN S15, 230MHz (D-sub) with calibration adapter in suitcase231005 CDN S200 CDN S2, 230MHz (audio)244144 CDN S200S CDN S2, 230MHz (audio) with calibration adapter in suitcase231009 CDN S250 CDN S25, 230MHz (D-sub)244148 CDN S250S CDN S25, 230MHz (D-sub) with calibration adapter in suitcase231006 CDN S400 CDN S4, 230MHz (DIN)244145 CDN S400S CDN S4, 230MHz (DIN) with calibration adapter in suitcase231007 CDN S900 CDN S9, 230MHz (D-sub)244146 CDN S900S CDN S9, 230MHz (D-sub) with calibration adapter in suitcase97-231024 CDN-TC Traceable calibration (ISO17025) for IEC 61000-4-6 requirements, order only with device CDN M, AF, S or T00x type234077 CAL S150 Calibration unit for CDN S150 (150 Ω /50 Ω adapter) 234074 CAL S200 Calibration unit for CDN S200 (150 Ω /50 Ω adapter)234078 CAL S250 Calibration unit for CDN S250 (150 Ω /50 Ω adapter)234075 CAL S400 Calibration unit for CDN S400 (150 Ω /50 Ω adapter) 234076 CAL S900 Calibration unit for CDN S900 (150 Ω /50 Ω adapter)

    Coupling / deCoupling network (Cdn) Cdn S type

    Cdn S150, S200, S250, S400, S900

    Auxiliaryequipment

    RF out

    Amp in

    < +10 dBm

    Amp outch.1 < +27 dBm

    ch.2 < +20 dBm

    ch.3 < +20 dBmPower meter

    NSG 4070

    PowerUSB

    Tuning

    StSizeStSize StSize

    Local

    Back StopRun

    Hold

    0 .

    1

    4

    7 8

    5

    2 3

    kHzdBm6

    9 MHzdBµV

    HzV

    Enter

    Step1

    STO

    FRQ LVL

    RCL

    Step2

    Step3

    MOD

    2nd

    Help

    RFON/OFF

    EUTAE

    Equipmentunder test

    Insulating Ground plane Insulating

    AE-side EUT-side

    CDN

    6 dB attenuator

    Test generator e.g. Teseq NSG 4070with built-in RF synthesizer, brodband power amplifier and power meter

    EUT test set-up