ct physics – ii

41
CT PHYSICS – II Dr. Archana Koshy 4/23/22 CT PHYSICS - II 1

Upload: archana-koshy

Post on 12-Feb-2017

858 views

Category:

Healthcare


1 download

TRANSCRIPT

Page 1: Ct physics – II

Monday, May 1, 2023 1CT PHYSICS - II

CT PHYSICS – II Dr. Archana Koshy

Page 2: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 2

OVERVIEW 1. IMAGE RECONSTRUCTION

2. HOUNDSFIELD UNITS

3. IMAGE QUALITY

4. MULTIPLE SLICE COMPUTED TOMOGRAPHY

SPECIAL APPLICATIONS OF CT :

4. 3D CT IMAGING 5. CT FLUOSCOPY

6. CARDIAC CT

7.CT ANGIO

8. CT BONE DENSITOMETRY

9. ARTEFACTS

Page 3: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 3

IMAGE RECONSTRUCTION

Page 4: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 4

(i) Correction for the heterochromatic nature of the beam (ii)Weighing factor to compensate the differences between

the size and shape of the scanning beam and picture matrix .

Page 5: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 5

ALGORITHMS FOR IMAGE RECONSTRUCTION

• For the purpose of reconstruction, complex mathematical equations called algorithms are required.

• The following three methods will be discussed :

(i) Back projection

(ii)Iterative methods

(iii)Analytical methods

Page 6: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 6

BACK PROJECTION • Also known as Summation method ; one of the oldest

methods .

Page 7: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 7

ITERATIVE METHODS Depending on whether the correction sequence involves the whole matrix , one ray or a single point .

-Simultaneous reconstruction –

• All projections for the entire matrix are calculated at the beginning of the iteration.

• All corrections are made simultaneously for each iteration.

-Ray by ray correction-

• One ray sum is calculated and corrected and these are incorporated into the future ray sums .

-Point by point correction -

• Calculations and corrections are made for all rays passing through one point .

• These corrections are used in ensuing calculations m again with the process being repeated for every point .

Page 8: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 8

ANALYTICAL METHODS

• Differ from iterative methods in that exact formulas are utilized for the analytical reconstruction .

• 2 most popular methods :

A) Filtered back projection

B) Two dimensional Fourier analysis – Any function of time or space can be represented by the sum of various frequencies and various amplitudes of sine and cosine waves.

Page 9: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 9

FILTERED BACK PROJECTION

Page 10: Ct physics – II

Monday, May 1, 2023 10

CT NUMBERS • Linear transformation of the original linear

attenuation coefficient measurement of each pixel .

• CT NUMBER = K (µp- µw) µw• The original EMI scanner used a magnification

constant of 500 and the values ranged from -500 ( air ) to + 500 ( dense bone ) .

• In contemporary CT untis , these values are arranged on a scale from -1024 HU to +3071 HU. CT PHYSICS - II

Page 11: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 11

Page 12: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 12

IMAGE QUALITY

QUANTUM MOTTLE ( NOISE )

• Variation in the number of X-ray photons absorbed by the detector .

• The only way to decrease noise is to increase the number of photons absorbed by the detector ; thereby increasing X-ray dose to the patient .

• In order to avoid statistical fluctuation , iterative logarithms were employed , noise would be smoothed out and the image would appear quite homogenous .

Page 13: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 13

• The image on the left has a higher degree of image noise, measured as a standard deviation in the range of HU of 48.

• On the right, the image has a more homogeneous appearance; less noise, which is measured as 9.4 HU.

• Noise also affects spatial resolution.

Page 14: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 14

SPATIAL RESOLUTION

• Ability of the Ct scanner to display separate images of two images placed close together .

• Determined by -Scanner design -Computer reconstruction -Design • Expressed in line pairs/ cm .

Page 15: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 15

CONTRAST RESOLUTION

• Defined as the ability of an imaging system to display the image of a relatively large object , which only slightly different in density from the surroundings .

• Low contrast visibility is determined by noise .

• Contemporary CT scanners can display objects about 3 mm in diameter with density difference of 0.5 % or less .

Page 16: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 16

MULTIPLE SLICE COMPUTED TOMOGRAPHY • Conventional CT scanners use a single row of detector

elements and acquire a single slice per rotation .

• MDCT uses multiple rows and can therefore take multiple slices per rotation .

• Speed of gantry rotation is increased resulting in an overall increase in scan speed .

• Allows larger volumes to be scanned in the same time .

• Functions both in high speed and high resolution mode promises to improve performance of spiral CT dramatically .

Page 17: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 17

Page 18: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 18

DETECTOR DESIGN

• More than 30000 detector elements are placed in a 2D array .

• Two approaches to the detector design :

a) Fixed matrix detector

b) Adaptive array detector

Page 19: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 19

IMAGE RECONSTRUCTION

• In multi slice CT , the outer most rays of the X-ray beam are tilted with respect to the imaging plane by the “cone angle “.

• The artefact level depends on the ratio of the cone angle and slice collimation .

• If higher number of detector rows are activated , visible artefacts appear .

Page 20: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 20

SPECIAL APPLICATIONS OF

CT IMAGING

Page 21: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 21

• Transverse images are stacked to form a 3 D data set ,which can be rendered as an image .

• Mostly used in Multispiral CT

• The algorithms include –

(i) Maximum Intensity projection (MIP)

(ii) Minimum Intensity Projection ( MinIP)

(iii)Surface shaded display volume rendering technique.

MULTIPLANAR RECONSTRUCTION

Page 22: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 22

Page 23: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 23

Page 24: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 24

Page 25: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 25

CT FLUOROSCOPY • Computed tomographic fluoroscopy is a technical

advance resulting from slip-ring technology, x-ray tubes with improved heat capacity, high-speed array processors, and partial reconstruction algorithms .

• The images can be reconstructed at a rate of approximately 6 frames per second, allowing near real-time visualization similar to that of ultrasonography.

• Safe and effective guidance tool for percutaneous interventional procedures .

• An additional concern is the scattered radiation to the consulting doctors .

Page 26: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 26

Page 27: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 27

CARDIAC COMPUTED TOMOGRAPHY

• Synchronisation of the data acquisition to the cardiac cycle and fast speed of the data acquisition to freeze the motion of the heart .

• Achieved by ECG gating • A particular ECG signal triggers the initiation of the CT

scan • Or Data acquired is reconstructed relative to a

selected cardiac phase .

Page 28: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 28

CT ANGIO

• Utilises the principle of using narrow collimation to scan the region and reconstructing both thin and thick slices .

• Superior quality tomographic images .

• The high resolution slices allow luminal views of the vessel . (“fly through”)

Page 29: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 29

DUAL ENERGY CT • Relatively new technique which uses two different x-

ray tubes in a single CT unit.• Additional applications include tissue differentiation

and visualization of tendons and ligaments.

Page 30: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 30

Page 31: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 31

• Dual Source CT uses two rotating tubes to acquire both high and low voltage images.

• Since the images are dependent on the attenuation of the x-ray beam, which depends on the voltage applied across the tube, each image acquired is energy dependent.

Page 32: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 32

APPLICATIONS OF DUAL CT • Angiography

• Renal calculi differentiation by determining the specific properties of the calculus .

• Plaque distribution in the vessels which have calcified can be viewed to diagnose atherosclerosis.

• Differentiation of thick ligaments and tendons.

• Ventilation and perfusion images

Page 33: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 33

QUANTITATIVE COMPUTED TOMOGRAPHY

• CT numbers or x-ray attenuation of a tissue is properly referenced to a calibration standard and then used to quantify some property of the tissue.

• Measures trabecular bone and is highly sensitive to changes in skeletal density.

• Both single energy and dual energy QCT can be used.

• QCT scanning done with dual energy eliminated the effect of marrow fat ; increased accuracy .

Page 34: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 34

ARTEFACTS

Page 35: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 35

MOTION ARTEFACTS • The reconstructed image will

display the object as a streak in the drection of motion .

• Motion of objects that have densities much different from their surroundings produces more intense artefacts .

• The intensity of the streak artefact will depend on the density of the object in motion .

• Motion of metallic or gas containing structures produce striking artefacts .

Page 36: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 36

STREAK ARTEFACTS • At every position , each detector will absorb some

transmitted radiation . • If a high density material severely reduces the

transmission , the detector may not record any image .• Hence streaks appear in the image

Page 37: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 37

BEAM HARDENING ARTEFACT • As a heterogenous xray beam passes through the

patient , the low energy protons are rapidly absorbed .

• Therefore the xray beam exiting the patient contains a lower percentage of energy photons .

• Reconstruction programs anticipate and correct the variation in linear attenuation co efficients , but are not precise .

Page 38: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 38

Page 39: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 39

RING ARTEFACT

• Result of miscalibration of a detector .

• Records incorrect data in every position .

• The misinformation is reconstructed as a ring in the image .

• The radius of the ring determined by the position of the faulty detector .

Page 40: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 40

Page 41: Ct physics – II

Monday, May 1, 2023CT PHYSICS - II 41