    Einstein-IIDGSOperating Manual

    Einstein-II DGS

    CONTENTS

    1>Introducing Einstein-II DGS ... 1

    2>How does an Ultrasonic Flaw Detector work.24

    3>Parts and Controls of Einstein-II DGS .. 5

    4>Operating detail .. 9

    5>Time/Date setting, Factory Default Setting 32

    6>Interface of Printer ..33

    7>Power Supply 34

    8> Precaution ..35

    9> Specification of Einstein-II DGS .36

    10> Menu Tree ....40

  • 8/10/2019 Ein II DGS Manual

    3/24

    Einstein-II DGS

    1 Introducing Einstein-II DGSEinstein-II DGS is a portable, compact, light in weight and user friendly Digital

    Ultrasonic Flaw Detector. Einstein-II DGS has COLOR LCD display, it provides wideviewing angle and allows fast scanning speed. Fully charged battery gives continuous

    working of eight hours and charging time is just 5 to 6 hours. Trace pattern can be

    directly printed out on conventional PC printer having serial port for interface or can be

    transferred to PC via RS232 Serial port for storage of data.

    --1--

    MODSONIC

    Einstein-II DGS

    2 How does an Ultrasonic Flaw detector Work

    Einstein-II DGS is a single channel Ultrasonic-testing instrument used for the inspectionof homogeneous material for the presence of inclusions, porosity and other

    discontinuities that could affect the performance of material and components. It can also

    be used for thickness gauging of homogeneous material, requiring access from only one

    side of the test piece. High frequency sound (Ultrasonic sound) waves are introduced intothe test material/part from a transducer/probe that is usually coupled to the test piece by

    water or other suitable coupling liquid. The transducer converts electric signals to

    Ultrasound and vice versa. A short burst of Ultrasound is introduced into test material sosome or all of the energy is reflected by

    discontinuities. The reflection of the ultrasoundenergy is a function of the ratio between the

    acoustic impedance of the discontinuity and thebase material. The greater the impedance ratiothe more sound energy will be reflected. The

    principle of Ultrasonic testing is shown in fig. 1.

    It shows the ultrasonic energy generated in thetest piece and resultant instrument display.

    Thickness gauging with Einstein-II DGS

    Operates on the principle of the time-of-flight

    --2--

    MODSONIC

  • 8/10/2019 Ein II DGS Manual

    4/24

    Einstein-II DGS

    measurement. This principle utilizes the precise timing of the transit time of a short burst ofultrasound energy, through a material under test. The ultrasound waves travel to the far side of the

    test piece and reflect back to the transducer/probe and a measurement is obtained.

    Basically three typed of transducers/probes are available for different types of applications:

    1> Straight beam Probe (Normal probe)

    2> TR probe (Dual crystal probe)

    3> Angle beam Probe1> Straight beam Probe:This probe introduces ultrasound normal to the test piece surface

    utilizing longitudinal or compression waves. Normal beam probe is used mostly for flaw

    detection and thickness gauging(refer figure 1.)

    2> TR probe: This probe contains separate transmitting

    and receiving elements as showing in figure 2

    usually mounted on delay lines. This designimproves near surface resolution by separating

    the initial pulse from the received echoes. TR

    probe is suitable for the thickness gauging of pitting

    and corrosion and also for better surface resolution.

    --3--

    MODSONIC

    Einstein-II DGS

    3> Angle Beam Probe:This probe introduces ultrasound at an angle to the surface of the testpiece. In most angle beam probes,

    the wave energy is mode converted from a

    longitudinal wave to a shear wave by therefraction principle. This probe is suitable for

    inspection of the welds. The reason for this is

    its ability to position the transducer/probeaway from the weld bead yet propagate energy

    into the weld zone. Another reason to use angle

    beam testing on the welds is to position thesound beam more normal to the expected discontinuities since the flaws in welds are usually

    perpendicular to the test piece (except porosity). Figure 3 shows the principle of angled beamweld testing.

    --4--

    MODSONIC

  • 8/10/2019 Ein II DGS Manual

    5/24

    Einstein-II DGS

    3 Parts and Controls of Einstein-II DGS

    1

    2

    3

    4

    5 6

    9

    10

    11

    12

    13

    1415

    20

    1716 1918

    21 87

    --5--

    MODSONIC

    Einstein-II DGS

    1. POWER key : Einstein-II DGS can be switched ON and OFF using this key.

    2. ZOOM key : By pressing this key A-Scan spreads over to full screen. Menuparameters disappear from LCD and only A-Scan is displayed. By

    pressing any other key, A-Scan is restored to normal size and

    menu parameters appear again .

    3. FREEZE key : By pressing this key once A-Scan can freeze. By pressing second

    time selects Peak Freeze mode. In this mode it hold echo dynamicpattern which is helpful for a angle beam inspection to locate peak

    signal.

    4. LOCK key : By pressing this key all function (keys) get locked which preventsaccidental pressing of any key. To unlock, simultaneously press

    Step key (9) and LOCK key (4).

    5. LCD Screen : It displays A-Scan Trace pattern and Menu Parameters.

    6. Battery Status : When all the three cells are displayed empty, then the battery

    Indicator needs to be recharged.

    7. Gain Step : This location displays Gain Step value. This value can be changed

    by using dB STEP(10) key. It follows 0.5, 1, 2, 6, 12 and 20dB Step sequence.

    --6--

    MODSONIC

  • 8/10/2019 Ein II DGS Manual

    6/24

    Einstein-II DGS

    8. Menu keys : By using these keys the Menu can be changed.

    9. Hot key : When this key is pressed simultaneously with any other parameter

    key then the change in value takes place in larger steps.

    10.dB STEP key : Any Gain step out of 0.5, 1, 2, 6, 12 or 20 dB can be selected

    using this key.

    11.

    GAIN UP/DN : Gain can be incremented/decremented using these keys in the step

    selected by dB STEP key.

    12. UP/DN key : Respective parameters can be incremented or decremented

    using these keys.

    13. REJECT : Rejection function activation is indicated by this LED.

    LED When Rejection value is zero then LED is OFF. When rejection

    value is not zero then LED illuminates.

    14. MONITOR LED : It is visual indication of MONITOR getting triggered.(LED 1

    and LED 2 represent Monitor Gage 1 and Gate 2 respective.)

    --7--

    MODSONIC

    Einstein-II DGS

    15. Parameter : All function (parameter) values are displayed in this area of

    Values LCD.

    16. RX Connector : A probe is connected to the Receiver of Einstein-II DGS

    through this connector.(choice of BNC and LEMO )

    17. TX Connector : A probe is connected to the Transmitter of Einstein-II DGS

    through this connector.(choice of BNC and LEMO)

    18. Charge Connector: Connect the charger here.

    19. RS-232 Connector: Serial interface connector. For print-out on Serial PC

    printer/Computer connect supplied cable on this connector.

    20. Battery : Battery is fitted in this compartment.Compartment

    21. Mode Status : Mode selection SINGLE or DUAL is displayed here.Indicator

    --8--

    MODSONIC

  • 8/10/2019 Ein II DGS Manual

    7/24

    Einstein-II DGS

    4 Operation detailEinstein-II DGS is user friendly, menu driven Ultrasonic Flaw Detector. To access any

    parameter, various menus are available. Details of all menus and parameters are as per below.

    First menu is Basic menu. In this menu, Basic functions of anUltrasonic Flaw Detector (UFD) like Range, Fine Range,

    Delay and Gain functions are covered.

    ZERO : It is to calibrate Artificial zero of the probe.RANGE : The required test range can be set directly in terms of mm/

    inch. The range available for steel is from 10 to 5000 mm

    (0.50 to 200 inch) it steps through 10,25,50, 100, 125,200,

    300,500,1000,2000,3000,4000 and 5000 mm with pressing

    hot key or in step of 1mm (0.5,1,2,5, 10, 15,20,25,30,50,100150 and 200 inch) or in step of 0.2 inch).

    VELOCITY : The required velocity can be set directly in terms ofmet./Sec (inch/sec). By pressing hot key it steps through

    preset value. In fine mode 1 met./sec. (0.0001 in/us).

    DELAY : Partial Range can be set using these keys. Exact shift/delayin mm (Inch) is displayed at the bottom.

    ZERO

    2200

    RANGE

    0100 MM

    VEL M/S

    5920

    DELAY

    000.0MM

    GAIN 1

    25.0 dB

    --9--

    MODSONIC

    Einstein-II DGS

    GAIN STEP: Gain step value can be set by using GAIN STEP key, it follows dB insteps of 0.5, 1, 2, 6, 12 or 20 dB.

    GAIN : Gain can be increased/decreased as per value of step by using GAINUP/DN keys.

    Example: Einstein-II DGS is to be calibrated for 100 mm Test Range using Single

    crystal (Normal) probe.

    > Set required range to 100 mm using RANGE UP/DN keys.

    > Couple the probe on 25mm thickness of V1 reference block and set first echo at 25mmusing ZERO UP/DN keys.

    > Set 4thecho at 100 mm using VELOCITY UP/DN keys, if required.> Repeat above two steps until both the echoes are at their respective positions.

    > Amplitudes of the echoes can be varied using GAIN keys.

    --10--

    MODSONIC

  • 8/10/2019 Ein II DGS Manual

    8/24

    Einstein-II DGS

    Second Menucovers MONITOR GATE 1 functions.

    GATE 1 : Gate can be changed to PLOGIC, NLOGIC, DAC, DAC 6,

    DAC-14 trigger, EXPAND or OFF.

    PLOGIC : When PLOGIC is selected, the MONITOR is

    triggered if an echo is present within the gate and its

    height is equal to or crosses the set Threshold Level.

    NLOGIC : When NLOGIC is selected, the MONITOR is

    triggered if there is no echo within the gate or theheight is lower than the Threshold Level.

    DAC : If DAC is turned ON and DAC is selected and if

    any echo signal equals or crosses DAC curve then

    monitor gates trigger.

    DAC-6 : If DAC is made ON 6/14 and DAC-6 is selected then if

    any echo signal cross DAC-6 dB curve then monitor

    gates trigger.

    DAC-14 : If DAC is made ON 6/14 and DAC-14 is selected then if echosignal equals or crosses DAC-14 dB curve then monitor

    gates trigger.EXPAND: When EXPAND is selected then the Scale expansion is activated and

    the GATE width is expanded to Full-screen width for a magnified view.

    GATE 1

    OFF

    START

    0030.0

    END

    0100.0

    LEVEL

    040 %

    GAIN 1

    25.0 dB

    --11--

    MODSONIC

    Einstein-II DGS

    START : Gate starting position can be set using START UP/DN keys. Using thesekeys Gate start can be directly set in terms of mm/inch.

    END : Gate End position can be set using END UP/DN keys. Using these keys Gateend can be directly set in terms of mm/inch.

    LEVEL : Gate Threshold Level can be directly set in terms of the percentage of the

    Full Scale Height (FSH).

    Example: The MONITOR to be triggered when any echo crosses the 50% FSH

    between 35mm and 80mm.

    > In the Gate setting, select PLOGIC using GATE UP/DN keys.

    > Set Gate START to 35mm using START UP/DN keys.

    > Set Gate END to 80mm using END UP/DN keys.

    > Set Gate LEVEL to 50% using LEVEL UP/DN keys.

    --12--

    MODSONIC

  • 8/10/2019 Ein II DGS Manual

    9/24

    Einstein-II DGS

    Third Menucovers MONITOR GATE 2 functions.

    GATE 2 : Gate can be changed to PLOGIC, NLOGIC, or OFF.

    PLOGIC : When PLOGIC is selected, the MONITOR istriggered if an echo is present within the gate and its

    height crosses the set Threshold Level.

    NLOGIC: When NLOGIC is selected, the MONITOR is

    triggered if there is no echo within the gate or the

    height is lower than the Threshold Level.

    START : Gate starting position can be set using START

    UP/DN keys. Using these keys, Gate start can be

    directly set in terms of mm.

    END : Gate End position can be set using END UP/DN keys.

    Using these keys, Gate end can be directly set interms of mm.

    LEVEL : Gate Threshold Level can be directly set in terms of the percentage ofthe Full Scale Height (FSH).

    GATE 2

    OFF

    START

    0040.0

    END

    0100.0

    LEVEL

    050 %

    GAIN 1

    25.0 dB

    --13--

    MODSONIC

    Einstein-II DGS

    Forth Menucovers Memory storage/recall and Trace printout functions.

    MEMORY : Memory function can be selected A SCAN or SET-UP

    using MEMORY UP/DN keys.

    MEM NO : Desired memory location can be selected using MEM NO

    UP/DN keys. When changing Memory no. If Data is stored

    and valid then asterisk (*) will be displayed nextto MEM NO.

    ACTION : Required functions like SAVE, RECALL, SEND, PRINT or

    DELETE, DETAIL can be selected using ACTION UP/DN

    keys. When DETAIL is selected it shows Note data ofStored A Scan it can be scroll by pressing MEM NO UP/DN

    keys.

    ENTER :By pressing Right key of the ENTER, the ACTION selected

    Is performed. On pressing Left key Note menu will be

    displayed and Side text menu will change for Text editingas shown. If required then edit/change Note detail. By

    pressing appropriate keys Note can be edited. For exit pressExit UP/DN keys.

    MEMORY

    A SCAN

    MEM NO*

    001

    ACTION

    RECALL

    NOTE

    ENTER

    GAIN 1

    25.0 dB

    --14--

    MODSONIC

  • 8/10/2019 Ein II DGS Manual

    10/24

    Einstein-II DGS

    Example: To SAVE current Calibration and A scan in Memory no 3.

    > Get required pattern which is to be stored. Press FREEZE key.

    > Select A SCAN using Memory UP/DN keys.> Set MEM NO to 3 using MEM NO UP/DN keys.

    OPERATOR:

    ABC

    JOB DETAIL:

    NIL

    DEFECT DETAIL:

    NIL

    REMARK:

    NIL

    CURSOR

    MOVE

    CURSOR

    UP/DN

    CHAR

    SCROLL

    NOTE

    EXIT

    GAIN 1

    25.0 dB

    --15--

    MODSONIC

    Einstein-II DGS

    > Set ACTION to SAVE using ACTION UP/DN keys.

    > Then press NOTE/ENTER Left key so Note menu will be displayed. If required

    then edit it and exit from note menu.

    > Then press NOTE/ENTER Right key to store A Scan. By pressing Right key

    Current Calibration, A Scan pattern and Note DATA will be stored UP/DN key to

    store the current calibration set-up at memory no. 3.

    Note:If data is already stored at this location, then the message will be displayed on LCD

    screen as OVER WRITE DATA YES - PRESS HOT KEY,

    NO PRESS ANY KEY. So press HOT KEY to overwrite data.

    --16--

    MODSONIC

  • 8/10/2019 Ein II DGS Manual

    11/24

    Einstein-II DGS

    Fifth Menu covers Rejection of unwanted signal, Set Reference, Video setting andTesting mode selection function.

    REJECT : For the rejection of unwanted signals. Using REJECTUP/DN keys Rejection level can be set as per requirement.

    It can be set directly in terms of percentage of FSH.

    Activation of this function is indicated by REJ LED(13).

    POWER : Hi or Low Transmitter power can be selected using

    POWER UP/DN keys.

    DAMPING: Probe Damping Hi/Low can be selected using DAMPING

    UP/DN keys.

    MODE : Test Mode for the use of Single or Dual Crystal probes can

    be selected using MODE UP/DN keys.

    REJECT

    000 %

    POWER

    LOW

    DAMPING

    LOW

    MODE

    SINGLE

    GAIN 1

    25.0 dB

    --17--

    MODSONIC

    Einstein-II DGS

    Sixth menucovers the required functions for DAC curve AWS or DGS curve as per selected in

    menu 10 CONFIG parameter.

    DAC : DAC setting can be switched to DRAW, ON ,ON 6/14, TCG

    ON or OFF using DAC UP/DN keys. DRAW mode, peak points

    of echo signals are selected.

    ON mode :The selected peak points are digitally connected to form a

    DAC curve.

    ON 6/14 mode: -6dB DAC and 14dB DAC will be also displayed.

    TCG ON: Time Gain Correction activated as per DAC so it brings echo

    signal to same height irrespective of depth.

    OFF mode :DAC curve will be not displayed on LCD.

    STRT G1 :Start of the GATE 1 can be set using STRT G1 UP/DN

    keys. To select any peak of the echo signal, move the

    gate on the echo signal and then press ENTER UP/DNkey. POINT value is automatically incremented to indicate

    next point.(If GATE 1 is OFF then turn to PLOGIC).

    POINT : It indicates how many points are selected to draw DAC

    curve.

    PRESS : Press ENTER UP/DN keys to accept the peak points

    selected by the GATE 1.

    DAC

    OFF

    STRT G1

    030%

    POINT

    001

    PRESS

    ENTER

    GAIN 1

    25.0 dB

    --18--

    MODSONIC

  • 8/10/2019 Ein II DGS Manual

    12/24

    Einstein-II DGS

    PRESS/ENTER : Parameter will be changed to T-CORR when DAC is On or TCG isON. Gain of machine can be change without curve height. It can be

    used for Transfer/Scanning gain correction.

    Example:To digitally draw DAC curve to join four peaks of the echo signals using 25mm

    thick reference block and range of 100mm.

    > Set Gain so that all four peaks are visible on the LCD screen above 5% FSH.

    > Set DAC to DRAW using DAC UP/DN keys.

    > Move cursor using STRT G1 UP/DN keys to cover the first echo. If GATE 1 isOFF then switch to PLOGIC.

    > Press ENTER UP/DN key to accept it and Point No. is incremented automatically.

    > Repeat above two steps for second, third and forth echo.

    > Set DAC to ON using DAC UP/DN key. Now DAC Curve will be displayed on

    LCD.

    Note: Cursor need not touch the echo for selection and acceptance of the peak point.

    --19--

    MODSONIC

    Einstein-II DGS

    AWS Function functions will be displayed as per selected in menu 10 CONFIG parameter.

    AWS : AWS setting can be switched to SET, MEASURE or OFFusing AWS UP/DN keys. SET mode, Make GATE 1 to

    PLOGIC then Bring reference echo of reflector in between 40 to

    60% , Press ENTER parameter key so gain of reference echo isrecorded and displayed as REF GAIN. (RG)

    MEASURE mode : Display IL (Indication level) dB value to bring echo

    to reference level.

    AF : Display attenuation factor to correct for the depth of the echosignal.

    IR : Display indication rating as per AWS code. IR=IL-RG-AF.

    Example:> Select MEASURE to ON.

    > Set GATE 1 to PLOGIC> Couple probe on reference defect and set gain value so echo height within

    40 to 60% of Screen.

    > Set GATE 1 in such a way so echo signal within GATE.> Then press SET REF /ENTER key so required gain to bring echo to 50% is

    recorded and displayed as a REF Gain in GAIN box.

    > Set AWS to MEASURE so it starts to display Indicating rating as per AWS

    code.

    AWS

    OFF

    STRT G1

    030%

    GAIN 1

    25.0 dB

    MODSONIC

    --20--

  • 8/10/2019 Ein II DGS Manual

    13/24

    Einstein-II DGS

    Indication rating is based below formula.

    IR=IL-RG-AF

    IL : Indicating Level its gain value to bring echo height to reference level (50%).

    RG : Reference Gain.. Its Gain value to bring reference echo height to referencelevel(50%)

    AF : Attenuation Factor its attenuation of signal in material as per AWS standard.

    Based on below formula.AF= (Sound path (in Inch) -1) * 2

    IR : Indicating Rating Its different in dB between the indication signal and reference signal with

    attenuation correction.

    MODSONIC

    --21--

    Einstein-II DGS

    DGS functions will be displayed as per selected in menu 10 CONFIG parameter.

    DGS : DGS setting can be switched to SELECT, ON or OFF using

    DGS UP/DN keys. In SELECT mode probe type , reference

    dia., reference type, required DGS curve dia. etc. can beselected. Refer DGS selection menu. In ON mode DGS curve

    will be displayed on Screen as per what is selected in DGS

    selection menu.

    STRT G1 : To select echo peak as reference move GATE 1 using

    STRT G1 UP/DN keys. If GATE 1 is OFF then select toPLOGIC in Menu 2.

    SET-REF :When switched to ON, reference echo within gate isregistered and height of reference echo is brought to ref.curve height. When DGS is turned to ON then SET-Ref will

    automatically switched OFF.

    TL-CORR : This correction is required if test object and reference block

    have different surface qualities. By varying TL-CORR gain willchange but DGS curve height remains same.(It works when

    DGS is selected to ON).

    DGS

    OFF

    STRT G1

    030%

    DGS-REF

    OFF

    TL-CORR

    00.0dB

    GAIN 1

    25.0 dB

    --22--

    MODSONIC

  • 8/10/2019 Ein II DGS Manual

    14/24

    Einstein-II DGS

    DGS Selection Menu: When DGS is selected to SELECT DGS selection menu willbe displayed on Screen. By pressing

    Menu UP/DN keys arrow( ) shifts.

    It indicates currently selected submenu. For eg. To change SCAL NO

    move arrow to first bottom empty box

    then press corresponding UP/DN keys.

    SCAL NO : It scrolls from 0 to 18. Select 0 for custom probe which one is not listed

    between 1 to 18. For custom probe select appropriate probe frequency and

    probe effective dia. For predefined probe, 1 to 18, it is not required.

    PRBNAME : It displays Model name of predefined probe.( for scale no 1 to 18 but for

    scale 0 it will display CUSTOM). With each Memory set one Customprobe parameters will be stored.

    SCAL NO

    002

    D EFFT

    23.1 MM EXIT

    DGS

    SELECT

    PRBNAME

    M2SN

    DGS-CRV

    11.0 MM

    ATT-REF

    00.0 dB

    STRT G1

    030%

    PRBFREQ

    2.00MHZ

    REFECHO

    FBH

    ATT-OBJ

    00.0 dB

    DGS-REF

    OFFDEL-VEL

    2500M/S

    REFSIZE

    11.0 MM

    AMPCOR

    00.0 dB

    TL-CORR

    00.0 dB

    GAIN 1

    25.0 dB

    --23--

    MODSONIC

    Einstein-II DGS

    PRBFREQ : It displays probe frequency of predefined probe. For custom probe select

    appropriate probe frequency by pressing appropriate UP/DN keys.

    DEL-VEL :It displays Ultrasound velocity of probe delay line. It is fixed forpredefined probes.

    D-EFFT : It displays Effective diameter of probe crystals. It is fixed for predefined

    probe. Select appropriate value for Custom probe.

    DGS-CRV :Select appropriate value for which diameter DGS curve is required.

    REF-ECHO :Select reference type weather it is FBH or BW (Back wall).

    REF SIZE :Select appropriate value as per reference reflector size.

    ATT-REF :Select Ultrasound Attenuation factor of Reference object. (dB/meter).

    ATT-OBJ :Select Ultrasound Attenuation factor of test object. (dB/meter).

    AMPLCOR :This is gain correction. When Angle probe is used and reference echo istaken from 100 mm or 25 mm arc of V1/V2 reference block.

    Example: To plot DGS for 11 mm size equivalent reflector for M2SN probe.

    > Switch GATE 1 to PLOGIC. (NLOGIC will also work).> Select DGS in menu 10 under CONFIG parameter. It will not turn to DGS if DAC is

    selected to OFF.

    > Go to DGS menu 6.

    --24--

    MODSONIC

  • 8/10/2019 Ein II DGS Manual

    15/24

    Einstein-II DGS

    > Couple M2SN probe on 25 thick V1 block and calibrate for required range.

    > Bring first echo within screen with height between 40 to 80 % FSH preferable.

    > Set first back-wall echo signal within GATE 1 using STRT G1.

    > Change DGS to SELECT by pressing DSG UP/DN keys.

    > Move arrow below first sub menu by pressing Menu UP/DN Keys.

    > Select SCAL NO to 2 by pressing corresponding UP/DN keys.> Move arrow below second sub menu by pressing Menu UP/DN keys.

    > Select DGS-CRV to 11.0 MM by pressing corresponding UP/DN keys.

    > Select REFECHO to BW by pressing corresponding UP/DN keys.

    >ATT-REF and ATT-OBJ to be set as per appropriate value. Here it is set to 0.

    > For exit press EXIT UP/DN keys.

    >Select SET-REF to ON so reference echo height is registered and brought to reference

    curve. (make sure reference echo is within GATE 1 with 40 to 80% FSH preferable).

    > Now press DGS UP/DN key so it turns to ON and appropriate DGS curve will be

    displayed.

    --25--

    MODSONIC

    Einstein-II DGS

    Seventh menucovers the required functions for Thickness/Distance measurements.

    MEASURE : MEASURE setting can be switched to ON or OFF using

    MEASURE UP/DN keys. In ON mode it will display

    thickness/distance of echo signal which is in-side the gate.In OFF mode it will not display thickness/distance.

    X-OFF : Set X-OFF value to same as Angle probe index point to

    Front face distance so Surface Distance will be displayed

    on LCD will be distance from angle probe front face to

    defect position.

    ANGLE : Value to be set as per Angle of the probe. If angle value

    not zero and Measure is set to ON then appropriate

    Surface and Depth of defect is calculated and display onLCD.

    THICK : Thickness value should be set as per the thickness of the

    test piece, so as to enable in calculation ofdepth/distance accounting for multiple skips.

    MEASURE

    OFF

    X-OFF

    000.0

    ANGLE

    000 DEG

    THICK

    100.0MM

    GAIN 1

    25.0 dB

    --26--

    MODSONIC

  • 8/10/2019 Ein II DGS Manual

    16/24

    Einstein-II DGS

    Example: To measure Beam path, Surface and Depth distance when 45 deg. probe is

    used.

    > Go to GATE 1 menu and make Gate to PLOGIC/NLOGIC.

    > Set Gate START and Gate END value in such a way so defect signal within the GATE.

    > Set Gain or Threshold so height of the echo/defect beyond the GATE.

    > Go to measure menu and set MEASURE to ON.

    > Set angle value to 45 DEG. using angle UP/DN keys.

    > Set appropriate THICK value so it can account multiple skips for DEPTH distance

    calculation.

    > Go to menu 11 and set MEA PO1 to PD, MEA PO2 to DP and MEA PO3 to SD so

    Sound path, Depth and Surface distance will be displayed on screen.

    --27--

    MODSONIC

    Einstein-II DGS

    Eighth menucovers the required functions for Thickness/Distance measurements.

    TRIGGER: Select Measurement point Weather trigger by FLANK or

    on PEAK.

    HUD : In Head up Display selected measured value will be

    displayed if MEASURE is selected to ON in Menu 7. IfALL is selected then value of MEA PO1, MEA PO2 and

    MEA PO3 will be displayed in Head up Display. (For

    selection of measured value refer menu 11 detail)

    SET REF : To set any A-scan pattern as a reference, achievereference pattern, then set SET REF ON, so reference

    pattern sets in background for easy comparison.A-scan Pattern From Memory location can be recalled

    and set as a reference.

    UNIT : To set Measurement/Calibration Unit. It can be select Inches

    or Metric by pressing UNIT UP/DN keys.

    Note: When changing UNIT Calibration will be change.

    TRIGGER

    FLANK

    HUD

    OFF

    SET REF

    OFF

    UNIT

    INCHES

    GAIN 1

    25.0 dB

    --28--

    MODSONIC

  • 8/10/2019 Ein II DGS Manual

    17/24

    Einstein-II DGS

    Ninth Menucovers miscellaneous functions, which are described as below.

    CLOCK : Clock can be switched ON or OFF using CLOCKUP/DN keys. If CLOCK is set to ON then date and

    time will be displayed at bottom of A-Scan.

    COLOR : LCD Color combination can be selected using COLOR

    UP/DN keys.

    VIDEO : Echo signals can be selected to appear as line

    (ENVELOP) or solid (FILLED) using VIDEO UP/DN

    keys. In ENVELOP mode, echo signals appear same as in

    Analog Ultrasonic Flaw Detector. In FILLED mode, echosignals looks solid for better visibility.

    BRIGHT : Brightness of LCD can be adjusted for better visibilityusing BRIGHT UP/DN keys. Set Low brightness when

    used for indoor application for the longer battery life.

    CLOCK

    OFF

    COLOR

    005VIDEO

    ENVELOP

    BRIGHT

    050 %

    GAIN 1

    25.0 dB

    --29--

    MODSONIC

    Einstein-II DGS

    Tenth menucovers the required functions for HORN and Key Beep Setting.

    CONFIG : Evaluation mode can be selected to DAC,AWSor DGS.As per

    selection, menu 6 will turns to DAC, AWS or DGS setting.

    This selection can change only if current active

    DAC/AWS/DGS is selected to OFF.

    HORN : Horn can be switched ON or OFF using HORN UP/DN

    keys. When HORN is switched ON, then it gives out anaudible alarm each time MONITOR is triggered.

    BEEP : Key press will be acknowledge by sound if BEEP isselected to ON. Otherwise it will not give beep sound when

    set to OFF.

    PRF : Using PRF UP/DB keys HI or LO PRF can be selected.

    In HI PRF is depends on current range selection. In Low

    mode is PRF is low which is useful when ghost echo

    pattern appears when testing low attenuate material.

    CONFIG

    DAC

    HORN

    OFF

    BEEP

    ON

    PRF

    HI

    GAIN 1

    25.0 dB

    --30--

    MODSONIC

  • 8/10/2019 Ein II DGS Manual

    18/24

    Einstein-II DGS

    Eleventh menucovers selection of measured parameter to be displayed on screen.

    MEA PO1 :Measured value will be displayed on screen at Left most

    position below base line. It scrolls as per detail given below.OFF Display OFF

    SND PTH Sound path of Echo signal

    DEPTH Depth of Echo signalSURF D Surface dist. of Echo signal

    ECH HI% Percentage height of Echo signalDAC % Percentage height with respect to DAC curve.

    (When DAC is selected to ON)

    ERS VAL Equivalent Reflector Size of echo signal

    (When DGS is selected to ON)

    DGS CRV Display currently selected DGS curve size.DGS HDB dB diff between DGS curve and echo signal

    (When DGS is selected to ON)

    Signal to be within GATE 1 and MEASURE is selected to ON in menu

    7 then appropriate measured value will be displayed onscreen.

    MEA PO2 : Measured value will be displayed on screen at Middle position below

    base line. It scrolls as per detail given in MEA PO1.

    MEA PO3 : Measured value will be displayed on screen at Right side below base line.

    It scrolls as per detail given in MEA PO1.

    MEA PO1

    OFF

    MEA PO2

    OFF

    MEA PO3

    OFF

    GAIN 1

    25.0 dB

    --31--

    MODSONIC

    Einstein-II DGS

    5 Setting of timeKeep key 1 press and switch ON Einstein-II DGS. After displayof menu Parameter, release key 1 and again press key 1 so Time setting menu will be

    displayed. Now set required Time/Data by pressing

    UP/DN parameter key. Then press key 2 for to

    store/set current time to value set as per display.Press key 3 to exit without store/set current time.

    Factory Default Calibration set-upIn Einstein-II DGS last calibration set-up remain

    store So when further Einstein-II DGS Starts itwill start with last setting, But due to some reason

    if stored calibration is corrupted then it will startswith factory default calibration set-up. If any userwant to recall factory default calibration set-up

    then press key 4, keep it press and Switch on

    Einstein-II DGS. After displaying menu parameter

    release key 4.

    2

    3

    --32--

    MODSONIC

  • 8/10/2019 Ein II DGS Manual

    19/24

    Einstein-II DGS

    6 Interface of PrinterEinstein-II DGS has direct printout facility of A-Scan pattern with full calibration set-up values.

    It can be printed out on conventional PC printer having a Serial Communication connector. So

    computer or special printer is not required. Printer should be IBM-PC compatible with serial

    port. Serial port setting is as below. Serial printer EPSON FX/LQ type recommended. (To setbelow setting in serial printer refer operating manual supplied with the printer).

    Baud Rate 9600 Parity bit NoneData Bits 8 Stop Bits 1

    To print out trace pattern, follow the below given steps.

    > Connect Serial cable (supplied with Einstein-II DGS) between connector (19) of

    Einstein-II DGS and serial printer.> Switch ON Serial PC printer.

    > Switch ON Einstein-II DGS.

    > Get the required Trace pattern and freeze using Freeze key (3).> Using Menu key go to memory menu (4).

    > Using ACTION UP/DN key select PRINT.

    > Press ENTER UP/DN key to initiate printing.

    For Interface with PC Using Serial port Refer Help.txt file supplied in SoftwareCD.

    --33--

    MODSONIC

    Einstein-II DGS

    7 Power Supply

    Einstein-II DGS is powered by a Lithium-Ion battery pack ATB1, which is fitted in Einstein-II

    DGS using two screws. It electrically connects to Einstein-II DGS via two internal terminal

    contact points. With fully charged battery pack ATB 1, Einstein-II DGS can operatecontinuously for eight hours, under typical operating conditions. The battery pack ATB 1 can be

    charged when inside Einstein-II DGS by connecting Battery charger ATC 1 to the connector

    (18). Alternatively, the battery pack ATB 1 can be charged outside Einstein-II DGS whichallows uninterrupted use of Einstein-II DGS using multiple battery packs.

    Battery Charging

    The battery Charger ATC 1 supplied with Einstein-II DGS is a special quick- charge charger.

    The charging time is just 5 to 6 hours. ATC 1 battery charger is suitable for use with mains

    supply from 100 to 240 volts AC/50 Hz. The LED on charger illuminates to indicate power on.When the battery voltage is low (discharge condition) the LED glows RED. When the battery is

    almost charged the LED turns to yellow. When the battery is full charged the LED glows green.

    The output voltage from the charger is 16.4 Volts DC. The charging current is constant at 2.0Amps.

    --34--

    MODSONIC

  • 8/10/2019 Ein II DGS Manual

    20/24

    Einstein-II DGS

    8 Precaution

    > Avoid shock to the Einstein-II DGS and probes.

    > Wipe the couplant off the probe and Einstein-II DGS after use.

    > When battery pack is out-of Einstein-II DGS, keep battery contacts away from

    metal object

    > Never heat the battery pack ATB 1 nor throw into a fire.

    --35--

    MODSONIC

    Einstein-II DGS

    9 Specification of Einstein-II DGS

    Test Range : 10 mm to 5 meter (0.500 in to 200 in) (in steel). In Hot Key modeit has 13 preset values. Fine mode adjustable in step of

    1mm (0.2 in).

    Velocity : 1000 met./sec to 9999 met./sec (0.0400 in/s to 0.4000 in/s). In

    Hot Key mode it has 7 preset values. Fine Mode adjustable in

    step of 1 met/sec (0.0001 in/s).

    Delay : Variable from 0 to 3000 mm (0 to 100 in). In Hot Key mode 5mm(0.2 in) . Fine mode adjustable in step of 0.5mm (0.02 in).

    Gain : 100dB calibrated gain adjustable in 0.5, 1, 2, 6, 12 or 20 dB step.

    Rejection : 0 to 100%FSH with LED indicator.

    Rectification : Full-wave rectified with filtering.

    Frequency : Broad Band amplifier 0.5 MHz. to 10MHz.

    Test Modes : Pulse echo and Transmit/Receive.

    Transmitter :Transmission pulse Negative spike with selectable high or low

    power.

    Damping : Damping high/low is selectable.

    Connectors : BNC and LEMO (Size 1) both are provided.

    --36--

    MODSONIC

  • 8/10/2019 Ein II DGS Manual

    21/24

    Einstein-II DGS

    Freeze/Peak Freeze : A-Scan freeze and Peak Freeze available. In Peak freeze it holdsecho dynamic pattern which is useful for angle beam probing to

    locate peak signal.

    Connectors : BNC and LEMO (Size 1) both are provided.

    Monitor : Dual gate adjustable in 1% of Screen width with

    Positive/Negative Logic, Gate Expand modes.

    Gate Expand : Expands Range to width of the gate.

    A-Scan memory : 200 Trace Patterns can be stored (with Note/Detail),recalled, print

    or transferred to PC via RS-232 serial port.

    Calibration Set-up : 50 different calibration set-ups can be Stored and Recalled.

    Software : EinSoft DGS Interface software for transferring A-Scan from

    Einstein-II DGS to PC supplied.Printer Attachment : IBM compatible Printer having serial port can be directly

    attached to main unit for printing of stored A-Scan waveform

    with calibration data.

    Display : High brightness Color TFT LCD Display. Display area 320 x 240pixel (122 x 92 mm). Eleven different color scheme option.

    Full Screen : By pressing Zoom key A-Scan can be displayed in Full Screenarea.

    --37--

    MODSONIC

    Einstein-II DGS

    Reference A-Scan : Reference A-Scan pattern of standard test object can saved and

    recalled in the background for easy comparison during testing.

    DAC : Dynamic DAC curve can be Digitally plotted (Smooth parabolic

    curve) on screen with 6dB and 14dB DAC curves.. DAC curvecan be set as flaw monitor gate. Using minimum 2 to

    maximum 10 point DAC curve can be plotted.

    TCG : As per plotted DAC Time Corrected Gain activated so it bringsecho signal to same height irrespective of depth.

    AWS : Built-in Software for evaluation of defect in accordance with

    AWS standards

    DGS : Defect size evaluation 18 predefined probe settings and one

    custom probe set-up per memory location. Defect size is directlydisplayed in ERS value. (Equivalent Reflector Size).

    Digital Readout : Thickness/Depth can be displayed in digital readout when usinga normal probe and Sound path, Surface Distance and Depth are

    directly displayed when angle probe is in use. Measurement point

    can be selected to be Peak or Flank. Echo height, ERS value,dB diff of DGS curve to signal height, Echo height WRP to DAC

    can be measured.

    Measurement Unit : Metric or British unit of measurement is selectable.

    --38--

    MODSONIC

  • 8/10/2019 Ein II DGS Manual

    22/24

    Einstein-II DGS

    Video Output : Composite video signal (PAL) output for connection to a TV.

    PRF : High or low PRF is selectable.

    Update Rate : 50 Hz.

    Power : Lithium-Ion Battery pack 14.4VDC, 4AH, gives 8 hours

    continuous operation from fully charged battery.

    Charger : Input voltage 100 to 240 VAC. Charge status indicators provided.

    Temperature : 0 to 55 C.

    Size : 152 x 255 x 70 mm (HxWxD).Weight : 2.3 kg. with Battery.

    --39--

    MODSONIC

    Einstein-II DGS

    10 Menu Tree

    First Menu ZERO RANGE VEL DELAY

    Second Menu GATE 1 START END LEVEL

    OFF

    PLOGIC

    NLOGIC

    DAC

    DAC 6

    DAC 14

    EXPAND

    Third Menu GATE 2 START END LEVEL

    OFF

    PLOGIC

    NLOGIC

    Forth Menu MEMORY MEM NO ACTION NOTE

    SET-UP RECALL ENTER

    SAVE

    DELETE

    A SCAN RECALL

    DETAIL

    MODSONIC

    --40--

  • 8/10/2019 Ein II DGS Manual

    23/24

    Einstein-II DGS

    DELETE

    PRINT

    SEND

    SAVE

    Fifth Menu REJECT POWER DAMPING MODE

    LOW LOW SINGLE

    HI HI DUAL

    Sixth Menu (1) DAC STRT G1 POINT PRESSOFF ENTER

    DRAW

    ON

    ON 6/14

    TCG ON

    AWS OFF START G1

    SET START G1 SET REF

    MEASURE START G1 IL ---- AF---- RATING

    Sixth Menu (2) DGS STRT G1 DGS-REF TL-CORR

    OFF

    SELECTON

    Seventh Menu MEASURE X-OFF ANGLE THICK

    --41--

    MODSONIC

    Einstein-II DGS

    OFF

    ONEighth Menu TRIGGER HUD SET REF UNIT

    FLANK SND PTH OFF INCHES

    PEAK DEPTH ON METRIC

    SURF D

    ECH HI%

    DAC %

    ERS VAL

    DGS CRV

    DGS HDB

    ALL

    Ninth Menu CLOCK COLOR VIDEO BRIGHT

    OFF ENVELOP

    ON FILLED

    Tenth Menu CONFIG HORN BEEP PRF

    DAC OFF OFF HI

    DGS ON ON LOWEleventh Menu MEA PO1 MEA PO2 MEA PO3

    OFF OFF OFF

    SND PTH SND PTH SND PTH

    --42--

    MODSONIC

  • 8/10/2019 Ein II DGS Manual

    24/24

    Einstein-II DGS

    DEPTH DEPTH DEPTH

    SURF D SURF D SURF D

    ECH HT% ECH HT% ECH HT%

    DAC % DAC % DAC %

    ERS VAL ERS VAL ERS VAL

    DGS CRV DGS CRV DGS CRV

    DGS HDB DGS HDB DGS HDB

    --43--

    MODSONIC

    All possible care has been taken in the preparation of this operation manual. If you find

    any errors or would like to make suggestions for improvement to this operation manual,