italian product - bsi · arezzo, italy procedure in 4 simple steps. • insert the strip: the...
TRANSCRIPT
ITALIAN PRODUCT
Blood glucose monitoring is essential1 in DIABETES MELLITUS I and II, for patients that DAILY must check their GLYCEMIA2.These patients need a PERSONAL DEVICE as simple, reliable and fast as possible.
OGCARE is a blood glucose monitoring device for checking glucose concentration in whole blood from fingertip pricking.
S E L F T E S T I N G
T E L E M E D I C I N E P H A R M A C Y • P O C
L A B O R A T O R Y
1 Kirk JK, Stegner J. Self-Monitoring of Blood Glucose: Practical Aspects. Journal of Diabetes Science and Technology. 2010;4(2):435-439.
2 TunisSL, Willis WD, Foos V Self-monitoring of blood glucose (SMBG) in patients with type 2 diabetes on oral anti-diabetes drugs: cost-effectiveness in France, Germany, Italy, and Spain,Current Medical Research and Opinion, 2010, 26:1, 163-175
Personal medical devices for metabolic disorders MONITORING and PREVENTION, taking advantage of reflectometric and electrochemical technologies.
OGCAREis an economic and easy-to-use device. OGCARE is a simple and accurate system for blood glucose monitoring.OGCARE allows patients to send data to a smartphone/tablet and to share them with clinicians.*
*optional, Bluetooth connectivity available depending on product version
Arezzo, Italy
Procedure in 4 SIMPLE STEPS.• Insert the strip: the device will turn on• Prick your finger• Let the strip aspirate
one drop of blood• Results in 5 seconds
OG Care user-friendliness also makes it a very useful support in situations that require fast and reliable blood glucose measurement, as emergency and first aid units.
MEMORY 300 tests results (including time and date)
AUTO-CODINGNo need to insert a chip, switch on by strip.
REDUCED BLOOD VOLUME 0.5 µl.
READING TIME: 5 seconds
EASY TO USE
STRIPS EJECTORto avoid contaminations
WIDE DISPLAY and ergonomic shape.
CUSTOMERIZABLEwith Fast/Not Fast selection
From the accuracy tests performed on 117 patients,a linear correlation coefficient of 0.993 and the following linear regression equation: y = 0.9691x - 2.3347 were obtained.The results comply with the acceptance criteriaof the new ISO15197:2015.
600
500
400
300
200
100
100100 200 300
0
0 400 500 600
Method Amperometric
Measurement range 20-600 mg/dL (1.1-33.3 mmol/L)
Calibration Auto coding, plasma equivalent
Power source 2X3V lithium batteries, type CR2032. Batteries can be used for more than 1000 measurements approximately
Switch on Automatic switch on after strip insertion
DimensionsLenght: 88 mmHeight: 18 mmWidth: 51 mm
Weight 48 gr (including batteries)
Display LCD display (backlight is optional)
Climatic operating conditionsTemperature limits: from 10° to 40°C (50°- 104 °F)Humidity limits: from 10% to 90% RH (non-condensing)Altitude limits: 3000 meters
Memory capacity 300 results with date and time
Data transmission Connecting cable and software
Warranty 3 years
Packaging1 meter,1 lancing device,1 bag with 10 sterile lancets,1 vial with 10 test strips, 1 softcase,1 user manual,1 instruction for use for lancing device
OGCARE METER TECHNICAL SPECIFICATIONS
Measurement method Glucose Oxidase
Range of measurement 20-600 mg/dL (1.1-33.3 mmol/L)
Sample Fresh capillary blood
Reading time 5 sec
OGCARE STRIPS TECHNICAL SPECIFICATIONS
Sample volume 0.5 µL
Hematocrit dependence 29-56%
Storage condition 5-30 °C (41-86 °F)
Packaging Packs of 25, 50 pcs
BIOCHEMICAL SYSTEMS INTERNATIONAL S.P.A.
www.biosys.it - [email protected]
HeadquarterLoc. Palazzo del Pero, 23 - 52100 Arezzo (AR) - ItaliaTel. +39 0575 984164 - Fax +39 0575 984238
Instrument DivisionVia B. Buozzi, 253 - 50013 Campi Bisenzio (FI) - ItaliaTel. +39 055 8963140 - Fax +39 055 8997086