laser alignment system

17
1 Laser alignment system Laser alignment system Status report Krzysztof Oliwa Eryk Kielar, Wojciech Wierba, Leszek Zawiejski, INP PAN, Cracow, Poland Wojciech Slominski, Jagiellonian University, Cracow , Poland FCAL Collaboration Meeting , May 6-7, 2008, INP PAN - Cracow, Poland

Upload: walt

Post on 19-Jan-2016

48 views

Category:

Documents


0 download

DESCRIPTION

Laser alignment system. Status report. Krzysztof Oliwa Eryk Kielar, Wojciech Wierba, Leszek Zawiejski, INP PAN, Cracow, Poland Wojciech Slominski, Jagiellonian University, Cracow , Poland. FCAL Collaboration Meeting , May 6-7, 200 8, INP PAN - Cracow, Poland. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Laser  alignment  system

1

Laser alignment systemLaser alignment system

Status report

Krzysztof Oliwa

Eryk Kielar, Wojciech Wierba, Leszek Zawiejski, INP PAN, Cracow, Poland

Wojciech Slominski, Jagiellonian University, Cracow , Poland

FCAL Collaboration Meeting , May 6-7, 2008, INP PAN - Cracow, Poland

Page 2: Laser  alignment  system

2

LumiCal

LumiCal

min

High precision in luminosity measurement andhigh accuracy in determination of LumiCal position

Outgoingbeam

LumiCal X, Y position with respect to the incoming beam should be known with accuracy better than ~700 µm (optimal ~100 -200 µm) (LumiCal’s will be centered on outgoing beam)

Distance between two LumiCal’s should be known with accuracybetter than ~60 -100 µm (14 mrad crossing angle)

min

2

L

L

Two LumiCal’s (L,R) alignment

Single (Left / Right) LumiCal alignment

IP

Page 3: Laser  alignment  system

3

• Two laser beams, one perpendicular, second with the angle of 45˚ to the CCD/CMOS sensor surface, are used to calculate the position shift\• The CCD camera and lasers can be fixed to the LumiCal and beam pipe• Three or more sensors can be used to measure tilt of each LumiCal• Six (?) laser beams from one to another LumiCal passing

inside the ‘carbon support’ pipe can be used : to measure the relative position shift (the method

described above) the distance between two LumiCal’s (very

challenging, not solved yet)

• Measurement based on frequency scanning interferometry to measure distance between calorimeters

Laser alignment system (LAS): laser beam monitoring with CCD sensor

450 beam spot

00 beam spot

X

Z

Laser1Laser2

X

Z

Laser beam spots

on the surface of CCD camera (640 x 480 pixels)

(picture from the computer monitor)

Page 4: Laser  alignment  system

4

LAS : laser beam monitoring system with CCD camera

450 beam spot

00 beam spot

Shape of the spots

Pixels saturation -used ND filters

The picture on the face of pixel CCD silicon camera

XY

450 beam spot00 beam spot

Page 5: Laser  alignment  system

5

LAS : laboratory setup

Present setup – dual laser beam

• BW camera DX1-1394a from Kappa

company 640 x 480 with Sony ICX424AL

sensor 7.4 μm x 7.4 μm unit cell size

• Laser module LDM635/1LT

from Roithner Lasertechnik

• ThorLabs ½” travel translation stage

MT3 with micrometers (smallest div. 10 μm)

• Neutral density filters ND2

• Renishaw RG24 optical head (0,1 µm resolution) to control movement of the lasers

• New support for laser aligments

Page 6: Laser  alignment  system

6

• Lasers table was translated in steps of 50 μm. • The distances Xtrue and Ztrue was measured with Renishaw RG-24 optical head with the resolution of ±0.1 µm

Results of X & Z position measurements

x = Xcal - Xtrue displacement (m) : ± 1 m

z = Zcal - Ztrue displacement (m) : ± 1.5 m

X, Z displacement measurement relative to reference system

Xcal and Zcal positions – from improved algorithm for centre beam spot determination.

45 deg laser beam accuracy - X displacement +/- 300 um

-1,000

-0,800

-0,600

-0,400

-0,200

0,000

0,200

0,400

0,600

0,800

1,000

-400 -300 -200 -100 0 100 200 300 400

X displacement [um]

dX

[u

m]

Serie1

45 deg laser beam accuracy - Z displacement 600 um

-1,000

-0,800

-0,600

-0,400

-0,200

0,000

0,200

0,400

0,600

0,800

1,000

1,200

-400 -300 -200 -100 0 100 200 300 400

Z displacement [um]

dZ

[u

m]

Serie1

Page 7: Laser  alignment  system

7

Stability - temperature dependence (current laboratory setup)

45 degree beam Perpendicular beam

• Insulated heating box .• For each temperature point, the mean positionof the spot centers from multiple measurementswere calculated using improved algorithm

Cooling down – measurement for each 5 minutesOver the T = 5.2 0 C. Position calculated from algorithm

The temperature dependence of the beam spots position in CCD camera:heating or cooling down environment of the laser system.

The relative distance two spots

T = 34 0 C

T=28.8 0 C

Perpendicular beam45 degree beam

The observed changes on the level ~ 2 m/1 0 C

Page 8: Laser  alignment  system

8

Temperature stabilization – the small temperature changes (T ~ 0.10 C)

45 degree laser beam 0 degree laser beam

The changes in distance betwen spots: on the level ± 0.4 m

The calculated X,Y positions of both beams -the relative changes are on the level ± 0.3 m

5 minutes measurements:

Even without temperature influencesome effect coming from natureof laser spot and systematic uncertainties in used algorithmcan be important

Page 9: Laser  alignment  system

9

Temperature stability

> 8 hours measurements : temperature changes within T ~ 0. 1 degree

45 degree beam 0 degree beam

The relative distance between laser beams

The observed changes in calculated X,Y spots positions are on the level 0.5 m. Contribution from other effects ?

It is necessary to stabilize the temperature of camera It is necessary to stabilize the temperature of camera

Collimator and laser optics should be improvedCollimator and laser optics should be improved

Page 10: Laser  alignment  system

10

Temperature stability

> 24 hours measurements : temperature changes within T ~ 0. 2 degree

45 degree beam 0 degree beam

The relative distance between laser beams

The observed changes in calculated X,Y spots positions are on the level 0.5 m. Contribution from other effects ?

It is necessary to stabilize the temperature of cameraIt is necessary to stabilize the temperature of camera

Collimator and laser optics should be improvedCollimator and laser optics should be improved

Page 11: Laser  alignment  system

11

•Reflective laser distance measurement – accuracy ~1-5 µm, resolution ~0.1-0.5 µm•Mirrors glued to beam pipe•Calibration of sensors procedure – detector push-pull solution (?)•Calibration of sensors procedure after power fault (?)

Alignment (L/R) measurement based on beam pipe and BPM.s.Alignment two parts (L+R) of LumiCAl

LAS development – integration with LDC Beam pipe can be centered on detector axis or on outgoing beam:a different free space for LumiCal

Beam pipe (well measured in lab before installing, temperature and tension sensors for corrections) with installed BPM Laser beams inside ‘carbon’ pipe (need holes, but possible) – interferometric measurement

will be behind BeamCal

Page 12: Laser  alignment  system

12

LAS development – integration with LDC

Proposed method: measurement based on frequency scanning interferometry

Page 13: Laser  alignment  system

13

FSI (Frequency Scanning Interferometry)

Page 14: Laser  alignment  system

14

Changes in X,Y and Z positions of the Tungsten and Si sensors layers

Example from MC studies on the internal structure deformation

Z

Z

Z

ideal

dedormationin Z

deformationin X and Y

The changes in relative luminosity according to The changes in relative luminosity according to

changes in internal structure along Z axischanges in internal structure along Z axis

An in X,Y directionsAn in X,Y directions

Possible systematic effect on luminosity measurements is expected to be about one order Possible systematic effect on luminosity measurements is expected to be about one order smaller in comparison to possible displacement the Lumical detector as whole but still should smaller in comparison to possible displacement the Lumical detector as whole but still should be treated carefully as possible significant contribution to total error in luminosity calculationbe treated carefully as possible significant contribution to total error in luminosity calculation

Page 15: Laser  alignment  system

15

Spanned wire alignment

Active during time slots between trains

Possible interferences with FE electronics

Accuracy up to ~0,5 µm

Quite simple electronics

Need 4 coax cables for each plane

Spanned wire alignment :

Spanned wire going through the holes in sensor planes working as antenaand pickup electrodes to measure the position

LAS development – measurement of individual sensor layers

Proposed solutions for the online measurement of the LumiCal sensor planes :

Page 16: Laser  alignment  system

16

Capacitive sensors:

- Displacement -> measurnig reactance of the capacitance beetwen central electrode and conducting surface of the target

- Central electrode connected to AC source with a controlled frequency (close to 15kHz)

- Small distance (up to 1mm) to measuring

- Sensivity ~3,5mV / um

- Guard rings reduces the edge effects for the sensor electrode

Page 17: Laser  alignment  system

17

Summary

• LAS is very challenging project in respect to the requirements: precisely positioned Si sensors (inner radius accuracy < ~4 µm),

X & Y alignment with respect to the beam < ~700 µm, distance between Calorimeters < ~100 µm, tilts < ~10 mrad

• The current laboratory prototype : the accuracy in position measurements are on the level ± 1.0 μm in X,Y and ± 2 µm in Z direction thermal stability of the prototype is ~1 µm/ºC

• The final LAS design will take into account LDC geometry

• More work is ongoing on the system development : alignment of both parts of LumiCal, positions of the internal sensor layers, the more compact prototype, readout electronics for dedicated sensors

and automatic position calculations