probe c 5-2_2

Upload: rima-jailani

Post on 06-Apr-2018

216 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/2/2019 Probe C 5-2_2

    1/12

    Philips Medical Systems is part ofRoyal Philips Electronics

    Interested?

    Would you like to know more about our imaginative

    products? Please do not hesitate to contact us.

    We would be glad to hear rom you.

    On the web

    www.medical.philips.com

    Via [email protected]

    By ax

    +31 40 27 64 887

    By mail

    Philips Medical Systems

    Global Inormation Center

    P.O. Box 1168

    5602 BD Eindhoven

    The Netherlands

    By phone

    Asia

    Tel: +852 2821 5888

    Europe, Middle East, Arica

    Tel: +49 7031 463 2254

    Latin America

    Tel: +1 954 628 1000

    North America

    Tel: +800 285 5585

    2006 Koninklijke Philips Elec tronics N.V.

    All rights are reserved.

    Philips Medical Systems Nederland B.V. reserves the right to make changes in specications and/or to discontinue any product at any time without

    notice or obligation and will not be liable or any consequences re sulting rom the use o this publication.

    Printed in USA

    4522 962 15631/795 * AUG 2006

    Your nearest Philips Medical Systems representative:

  • 8/2/2019 Probe C 5-2_2

    2/12

    Philips now oers an ultrasound system with

    high denition perormance and eatures

    sized and priced or the demanding single or

    multi-specialty practice.

    HD Performance

    All-digital broadband beamormer captures

    more tissue inormation or true high

    denition perormance

    Wide dynamic range and digital ocal tuning

    provide exceptional sensitivity and detail

    Advanced Features

    Tissue Harmonic Imaging (THI) or

    better image resolution with less artiacts

    Highly sensitive PW Doppler, color

    Doppler and CPA provide dynamic

    blood fow data

    Freehand 3D surace capability or

    advanced obstetrical examinations

    Clinical applications packages covering

    Ob/Gyn, General Imaging, Vascular,

    Urology, Small Parts and Cardiology

    Mobility and Ergonomics

    Lightweight cart is easy to maneuver

    in tight spaces

    Height adjustable control panel and

    monitor help reduce work-related

    strain injuries

    Convenient on-cart storage or

    transducers, peripheral equipment

    and supplies

    Holds up to three transducers or

    comprehensive exam protocols

    Compact powerPhilips HD3 ultrasound system specications

  • 8/2/2019 Probe C 5-2_2

    3/12

    Clinical Applications

    Abdominal

    Supercial/Small Parts

    Pediatric

    Obstetrical

    Gynecological

    Urology/Prostate

    Vascular

    Cardiac

    System Specications

    Imaging Modes

    2D imaging

    Tissue Harmonic Imaging (THI)

    Pulsed wave Doppler (PW)

    M-mode

    Freehand 3D

    Optional color Doppler

    Optional Color Power Angio Imaging (CPA)

    Dual 2D/Color image mode with Cineloop review

    Duplex or simultaneous 2D and Doppler

    Triplex mode or simultaneous 2D, Doppler and

    Color/CPA

    High resolution write zoom

    Trapezoidal imaging

  • 8/2/2019 Probe C 5-2_2

    4/12

    Control Panel and User Interface

    Easy-to-learn graphical user interace

    Primary controls readily accessible and

    logically grouped Commonly used secondary controls located on

    sot keys or quick access; sot key unctions change

    dynamically based on the active mode, preset or

    system unction

    Other secondary controls accessible through

    on-screen menus

    Alphanumeric keyboard: QWERTY keyboard with

    globalization key or conversion to local language

    (English, French, German, Italian and Spanish)

    User selectable keyboard input language (Roman,

    simplied Chinese and Russian) Trackball with Select and Enter keys or easy

    system navigation

    Integrated stereo speakers

    Imaging mode keys: 2D, Color Power Angio (CPA)

    imaging, M-mode, Color Doppler, PW Doppler

    2D image controls: Depth, Dual, Freeze, THI,

    Zoom, Focus

    Image enhancement controls: Dynamic Range,

    Focus, Gain, Persistence, Post-processing Map,

    Edge Enhancement

    Tissue optimization keys: Image Optimization control,Transducer-specic presets, Tissue Harmonic Imaging

    Quantitation controls: Control, Trackball, Enter,

    Select, Next and Clear

    Doppler/Color controls: Angle/Steer, Scale, Baseline,

    Gain, Power, Volume, Duplex and Triplex

    Image Acquisition keys: Review, Print, and Acquire

    Annotation controls: Text, Clear, Pointer, Bodymarker

    Function keys: Patient Entry, System Setup

    Time Gain Compensation (TGC) slide pot controls

    Review and Report Keys

    Gray Shades

    256 (8 bits) in 2D, M-mode and Doppler

    spectral analysis

    Color Levels

    256 color levels (8 bits) in color Doppler and

    CPA modes

    256 color levels in 2D Chroma display

    Acquisition Frame Rate

    Greater than 90 rames per second (rame rates

    dependent on transducer, eld o view, depth

    and angle)

    Ergonomic, User-Centered Design High mobility cart

    Designed or easy maneuverability

    - Front handles or portability

    - Four-wheel swivel ability

    - Two-wheel ront brake

    Control panel can be raised and lowered

    Supports comortable operation rom standing

    and sitting positions

    Improves accessibility o controls and reduces

    extended reaching

  • 8/2/2019 Probe C 5-2_2

    5/12

    High-resolution LCD fat panel display

    Height adjustable with control panel Independent tilt (180 degrees) and swivel

    (180 degrees)

    High resolution LCD display (640 x 800) with

    progressive scan

    16 million color depth

    Display 29.5 cm x 31.5 cm including bezel,

    12.1 inch LCD viewing area

    Brightness control

    Integrated stereo speakers

    Cart back deck provides easy access to hardcopy/

    documentation devices Built-in ront bay or easy access to black and white

    printer or storage

    Built-in A/C line conditioner provides isolation rom

    voltage fuctuation and electrical noise intererence

    Internal high-capacity ans to optimize cooling

    eciency with minimal audible noise

    System Architecture

    All-digital broadband beamormer

    2D ocusing with Dynamic Focal Tuning

    150 dB ull time system dynamic range 256 digitally processed channels

    Display Annotation

    On-screen display o all pertinent imaging parameters

    or complete documentation, including transducer

    type and requency range, active clinical options and

    optimized presets, display depth, TGC curve, grayscale,

    color map, rame rate, compression and color

    gain, color image mode, and institution and patient

    demographic data

    Displayed data can be turned o or generating imagesused in publications and presentations

    Continuous display o Mechanical Index (MI) and

    Thermal Index (TIb, TIs)

    Body markersdisplays body-part icons appropriate

    or the active preset and indicates relative transducer

    position, selectable via sot keys and control panel

    Dual orientation mark to indicate the active buer

    System Specications

    Image Presentation

    Up/down Let/right

    Depth to 24 cm (exam and probe specic)

    Transducers

    Transducer Selection

    System supports six transducers to meet a wide range

    o clinical needs

    Two transducer connection ports

    Optional three transducer connection port

    Electronic switching o up to three transducers

    Continuous dynamic receive ocusing on all transducers

    C5-2 Broadband Curved Array

    5 to 2 MHz extended operating requency range

    Provides high-resolution images or abdominal and

    Ob/Gyn applications

    Supports 2D, 3D, THI, M-mode, color, CPA, and

    PW Doppler

    Biopsy kit available

    C7-3 Broadband Curved Array

    7 to 3 MHz extended operating requency range

    Provides high-resolution images or abdominal and

    Ob/Gyn applications

    Supports 2D, 3D, THI, M-mode, color, CPA, and

    PW Doppler

    Biopsy kit available

    C9-4 Broadband Curved Array

    9 to 4 MHz extended operating requency range

    10 mm radius o curvature, 150.3 degree eld o view

    Provides high-resolution or neonatal head

    and vascular applications

    Supports 2D, 3D, M-mode, CPA and PW Doppler

    C9-4ec Broadband Curved Array

    9 to 4 MHz extended operating requency range

    10 mm radius o curvature, 150.3 degree eld o view

    Endovaginal/Endocavitary applications

    Supports 2D, 3D, M-mode, color, CPA, and

    PW Doppler

    Biopsy kit available

  • 8/2/2019 Probe C 5-2_2

    6/12

    L9-5 Broadband Linear Array

    9 to 5 MHz extended operating requency range

    Trapezoidal imaging

    Provides high-resolution images or supercial

    applications including vascular, small parts andmusculoskeletal imaging

    Supports 2D, 3D, color, PW Doppler and Color Power

    Angio imaging

    Biopsy kit available

    C4-2 Broadband Curved Array

    4 to 2 MHz extended operating requency range

    Tightly curved radius or cardiac and abdominal

    applications

    Supports 2D, 3D, THI, M-mode, color, CPA, PW

    Doppler and Color Power Angio imaging

    Imaging Modes

    2D

    Frame rate selection

    Write/Read zoom

    Image orientation marker

    Cineloop image review (up to 256 rames)

    Persistence, adjustable in real time

    Selectable compression curves

    Sector size and steering control

    Selectable line density Up to our transmit ocal zones

    Dual imaging with Cineloop review

    Chroma imaging with 36 selectable color maps

    B/W image suppress

    M-mode

    Available with all imaging transducers

    Selectable sweeping rates

    Time markers

    Chroma colorization with 36 selectable color maps

    M-mode review or retrospective analysis oM-mode data

    Selectable display ormat (1/3-2/3, 1/2-1/2, 2/3-1/3)

    Doppler

    Display annotation including Doppler mode, scale,

    wall lter setting, gain, acoustic output status, sample

    volume size, normal/inverted, angle correction,

    grayscale curve

    Adjustable requency/velocity display ranges

    8-position zero baseline shit

    Normal/ invert display around horizontal zero line Selectable sweep speeds

    Selectable grayscale curve or optimal display

    Selectable display ormat (1/3-2/3, 1/2-1/2, 2/3-1/3)

    Doppler Review or retrospective analysis o

    Doppler data

    Pulsed Wave (PW) Doppler

    Available on all imaging transducers

    Adjustable sample volume size: .5-15 mm

    Simultaneous or duplex operation modes

    Triplex mode: displays 2D, color/CPA and PWDoppler simultaneously

    Tissue Harmonic Imaging (THI)

    System processing o second harmonic requencies

    (nonlinear energy) in tissue

    Extends high perormance imaging capabilities to

    all patient body types

    Available on the C7-3, C5-2 and C4-2 transducers

    Image display virtually ree o artiacts

  • 8/2/2019 Probe C 5-2_2

    7/12

    Color Power Angio Imaging (CPA)

    Highly sensitive mode or small vessel visualization

    Available on all transducers

    Cineloop review

    User-denable presets

    Multiple color maps

    Individual controls or gain, lters, sensitivity, scale

    and persistence

    Adjustable CPA region o interest: size and position

    User-selectable persistence

    3D Grayscale Imaging

    Freehand 3D acquisition available on C7-3, C5-2, C4-2,

    C9-4 and L9-5 transducers

    Individual controls or manipulating the on-screen 3D

    rendering and display options

    Expanded Field o View

    Trapezoidal imaging expands eld o view by 15

    degrees on linear array transducer

    Measurement Tools

    Measurement Tools and General Description

    2D distance

    2D circumerence/area by ellipse, continuous trace,

    trace by points

    2D curved-linear distance

    M-mode distance (depth, time, slope)

    Doppler distance

    Auto and manual Doppler trace

    Time/Slope measurements in Doppler and M-mode

    Doppler values containing PI, RI, S/D indices

    2D volume

    Heart rate

    Trackball-controlled electronic measurement

    calipers: 4 sets

    User-dened protocols, measurements and equations

    Fully-editable results data sheet

    System Specications

  • 8/2/2019 Probe C 5-2_2

    8/12

    Analysis Packages

    Clinical Option Analysis Packages

    Comprehensive measurements, calculations and

    application-specic reports, including expanded

    cardiac, vascular, ob/gyn and urology capabilities

    or thorough exam documentation

    Obstetrical Analysis

    Fetal biometry

    Biophysical prole

    Amniotic fuid index

    Early gestation

    Fetal long bones

    Fetal cranium

    Nuchal translucency, nuchal old

    Supports multiple gestations up to our etuses

    Other Obstetrical Measurements

    Fetal heart M-mode

    Fetal Doppler, including etal carotids, MCA,

    umbilical artery, etal aorta and ductus venosus

    Maternal uterine artery

    Placenta artery

    Fetal Echo

    User-selectable obstetric tables and equations

    User-dened obstetric tables and equations

    Obstetric trending

    Gynecological Measurements and Calculations

    Gynecology/Fertility

    Uterus

    Cervix and endometrial thickness

    Right and let ovaries

    Right and let ovarian ollicles

    Ovarian artery

  • 8/2/2019 Probe C 5-2_2

    9/12

    Other Analysis Pediatric Hip Joint

    Urology analysis

    Prostate volume

    T-Z prostate volume

    Bladder volume and residual volume

    Kidney volume

    Volume fow

    System Features

    Cineloop Review Acquisition, storage in memory and display in duplex

    modes o up to 512 rames o 2D and color images

    or retrospective review and image selection

    Single rames o Doppler data and M-mode images

    can be archived to print or electronic media

    Real-time Auto Playback o Cineloop rames

    Trackball control o rame-by-rame image selection

    Functions in 2D, M-mode, PW Doppler, Color Doppler

    and Color Power Angio imaging modes

    System Specications

    Physical Dimensions

    Depth: 31.4 in/79.8 cm (no peripherals)

    Width: 19.9 in/50.5 cm

    Ground clearance: 2.75 in/7 cm

    Height (adjustable): 40-59 in/101-149 cm

    Control panel height (adjustable): 30-37 in/78-94 cm

    Weight: 135 lbs/61 kg (no peripherals)

    Simple, easy-to-usecontrols

    Ergonomically adjustablecontrol panel andmonitor

    On-board printer port andperipheral storage

    Philips High Denition clinicalperormance

    Choice o up to three transducersor multiple scanning needs

    LCD fat-panel display or sharp colorand grayscale imaging

    Integrated CD driveand optional DICOM

    networking

    Lightweight andeasy to move

    Vascular Analysis Cerebrovascular

    Peripheral vascular

    Cardiac Analysis

    M-mode analysis

    Doppler

    Ventricle analysis

    Diastolic unction

    Cardiac output

    Valvular analysis

    Pulmonary vein analysis Ejection raction (via Teichholz or cubed method)

    Simpsons biplane and single plane

    Volume by area/length method

    Peak and mean gradients

    Pressure hal time

    Continuity equation

    Proximal Isovelocity Surace Area ( PISA)

    Qp:Qs ratio

    LV mass

    IVRT

  • 8/2/2019 Probe C 5-2_2

    10/12

    LAN connector used with DICOM networking

    Footswitch port (optional) or connecting theoptional oot switch

    USB port adapter or RS-232 port to support

    data transer

    Localization Options

    Sotware: English, French, German, Italian, Spanish,

    Training and user documentation: English, French,

    German, Italian, Spanish, Japanese, simplied Chinese,

    Greek, Danish, Dutch, Swedish, Turkish, Polish,

    Portuguese and Russian

    Power Requirements

    Power: 100-127V/200-240V; 50-60Hz

    Frequency: 47 to 63 Hz

    Voltage: 90V to 264VAC

    Power Cords

    Available or electrical standards worldwide

    Electrical Saety Standards

    CSA C22.2 No. 601.1

    IEC 60601-1 UL-60601-1

    EN60601-1

    According to CISPR 11, this is Group 1, Class A ISM

    equipment. It is suitable or use in all establishments

    other than domestic and those directly connected to

    the low-voltage power supply network that supplies

    buildings used or domestic purposes.

    Environmental

    Temperature

    System: 10-40 C at 15-95% relative humidity VCR and printers: 0-40 C at 80% relative humidity

    (non-condensing)

    Miscellaneous System Data

    Optional two pedal ootswitch: user-congurableor Dual, Freeze and Acquire controls

    Connectivity and Storage

    75 GB onboard hard drive memory storage

    Standard CDR/RW recordable

    Supports USB memory storage devices

    Direct digital storage o system conguration backup,

    including user-dened presets

    Direct digital storage o single rame color and B/W

    images to internal hard disk, CD and USB memory

    storage device Stores a minimum o 75,000 B&W still rame images

    to internal hard drive

    Ability to export images to USB fash memory or

    compact disk or PC viewing in multiple image archive

    ormats (JPG, BMP, TIF, DICOM)

    Extensive image management capability, including

    multiple ormat image review

    Patient exam directory

    Delete and replace recalled image capability

    DICOM 3.0 print and store service class user

    10BaseT Ethernet output Site congurable IP address, port and AE Title

    Study reports available as DICOM images

    Exam Documentation

    Peripherals

    Mitsubishi HS-MD3000 video cassette recorder

    (PAL and NTSC)

    Sony UP 897MD black and white thermal printer

    Sony UP21-MD color printer

    System supports a range o plain paper printers

    Input/Output Ports

    Available ports:

    Two standard USB interaces or support o qualied

    plain paper printers and USB fash memory

    Composite video: output to external monitor, VCR

    or printer

    Audio right/let

    External print trigger

  • 8/2/2019 Probe C 5-2_2

    11/12

    Transducers

    Name C5-2 C7-3 C9-4 C9-4ec L9-5 C4-2

    Extended operating requency range 5-2 MHz 7-3 MHz 9-4 MHz 9-4 MHz 9-5 MHz 4-2 MHz

    Application Exam Type

    Abdominal General

    Renal

    Vascular

    Obstetrics Early OB

    General OB

    Fetal Echo

    Gynecology Pelvis

    Fertility

    Cardiology Adult General

    Vascular Carotid

    Arterial

    Venous

    Abdominal

    Pediatric Abdomen

    Hip

    Neonatal Cephalic

    Small Parts Supercial

    Thyroid

    Testicle

    Breast

    Musculoskeletal

    Urology Prostate

    Bladder

    Renal

  • 8/2/2019 Probe C 5-2_2

    12/12

    Color Power Angio and Cineloop are trademarks o KoninklijkePhilips Electronics N.V.