920896-01 Rev T 10/2018 QSE (Q-Star) Series Electromagnetic Meter Owner’s Manual NPT Meter (1/2 in. to 2 in.) Shown with Display Mount Cover Plate and Display BSPP Meter (1/2 in. to 2 in.) Shown with Display Mount Cover Plate and Display ANSI Flanged Meter (3 in. & 4 in.) Shown with Plain Cover Plate and Pulse Out Transmitter (QSB) INTRODUCTION The QSE meter has multiple types of output electronics available. The electronics for the operation of the meter coils and flow tube are housed within the meter body casing. The cover plate is designed in two versions; a plain cover plate or a display mount cover plate. The output electronics (QSB, QSI1, QSI2 or QSI3) can be housed within either of the two cover plates. A display (Q09) is also available mounted to the “display mount” cover plate. All meters are equipped with galvanically isolated pulse-out electronics (QSB) as the default standard, regardless of style of cover plate. This manual contains overall information related only to the meter. This meter is externally powered and all external wiring connects to the electronics within the cover plate through its threaded ports. TABLE OF CONTENTS INTRODUCTION ..................................... 1 IMPORTANT NOTICE ....................... 2 PRINCIPLE OF OPERATION ............ 2 SAFETY ............................................ 2 INSTALLATION ....................................... 3 EARTH GROUND ............................. 3 Grounding Summary: ....................... 3 CONNECTIONS ............................... 6 RECOMMENDED INSTALLATION .... 8 WIRING ............................................ 9 MAINTENANCE ...................................... 9 TROUBLESHOOTING .......................... 10 SPECIFICATIONS ................................. 11 U.S. MEASUREMENT .................... 12 METRIC MEASUREMENT .............. 13 U.S. PRODUCT WEIGHT ............... 14 METRIC PRODUCT WEIGHT ......... 14 DIMENSIONS ................................. 15 TEMPERATURES ........................... 16 FLUID ELECTRICAL CONDUCTIVITY .. 17 REPLACEMENT PARTS LIST ............... 19 SERVICE............................................... 21 LIMITED WARRANTY ........................... 24

Upload: others

Post on 01-Feb-2021

3 views

Category:

Documents


0 download

TRANSCRIPT

  • 920896-01 Rev T10/2018

    QSE (Q-Star) Series Electromagnetic Meter

    Owner’s Manual

    NPT OR BSP METER

    WITH 09 ELECTRONICS

    3" POLYMER FLANGE METER WITH QB ELECTRONICS

    NPT Meter(1/2 in. to 2 in.)

    Shown with Display Mount Cover Plate and Display

    BSPP Meter(1/2 in. to 2 in.)

    Shown with Display Mount Cover Plate and Display

    ANSI Flanged Meter(3 in. & 4 in.)

    Shown with Plain Cover Plate and Pulse Out Transmitter (QSB)

    INTRODUCTION

    The QSE meter has multiple types of output electronics available. The electronics for the operation of the meter coils and flow tube are housed within the meter body casing. The cover plate is designed in two versions; a plain cover plate or a display mount cover plate. The output electronics (QSB, QSI1, QSI2 or QSI3) can be housed within either of the two cover plates. A display (Q09) is also available mounted to the “display mount” cover plate. All meters are equipped with galvanically isolated pulse-out electronics (QSB) as the default standard, regardless of style of cover plate.

    This manual contains overall information related only to the meter. This meter is externally powered and all external wiring connects to the electronics within the cover plate through its threaded ports.

    TABLE OF CONTENTSINTRODUCTION ..................................... 1

    IMPORTANT NOTICE ....................... 2PRINCIPLE OF OPERATION ............ 2SAFETY ............................................ 2

    INSTALLATION ....................................... 3

    EARTH GROUND ............................. 3Grounding Summary: ....................... 3CONNECTIONS ............................... 6RECOMMENDED INSTALLATION .... 8WIRING ............................................ 9

    MAINTENANCE ...................................... 9

    TROUBLESHOOTING .......................... 10

    SPECIFICATIONS ................................. 11

    U.S. MEASUREMENT .................... 12METRIC MEASUREMENT .............. 13U.S. PRODUCT WEIGHT ............... 14METRIC PRODUCT WEIGHT ......... 14DIMENSIONS ................................. 15TEMPERATURES ........................... 16

    FLUID ELECTRICAL CONDUCTIVITY .. 17

    REPLACEMENT PARTS LIST ............... 19

    SERVICE ............................................... 21

    LIMITED WARRANTY ........................... 24

    BSPP METER WITH

    09 ELECTRONICS

  • 2

    The magnetic coils, electrodes and other electronic components within the main meter body receive power from the electronics housed within the cover plate through a ribbon cable. See the included electronics manuals for meter wiring diagrams specific to your meter electronics.

    IMPORTANT NOTICE

    Your QSE meter is supplied ready for operation in a wide variety of applications. The meter has been factory configured to your order. It is suitable for volumetric flow measurement of non-flammable, electrically conductive liquids that have a minimum fluid conductivity of 10 µS/cm, and are compatible with the wetted components of the meter (See Specification Section).

    Fluid conductivity below 50 µS/cm may result in uncertain readings. Consult factory for use with fluids having a conductivity below 50 µS/cm.

    Use QSE series meters with water, aqueous solutions and other non-flammable, electrically conductive fluids. A fluid conductivity chart of common liquids is in the back of this manual for your reference. Do not use the meter with petroleum products (diesel fuel, unleaded gasoline, jet fuel, kerosene, etc.) or incompatible chemicals.

    QSE series meters are very sensitive to electric noise if operated within 6 inches (152mm) of some electric motors, relays, transformers or other sources of electronic noise.

    If the QSE series meters are used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.

    PRINCIPLE OF OPERATION

    Faraday’s Law of Electromagnetic Induction is the operating principle on which the QSE series meters are based. Faraday’s Law (paraphrased) states that a voltage will be induced in a conductor when it passes through a magnetic field, and the induced voltage will be directly proportional to the velocity of the conductor passing through that magnetic field. In this case, flowing liquid is the conductor and the QSE meter creates the magnetic field. The velocity of the flowing liquid, which must pass through the magnetic field, is the velocity of the conductor. A voltage is induced in the conductive liquid as it passes through the magnetic field. By placing electrodes in calculated locations on the flow tube of the meter, it is possible to accurately measure the induced voltage, thus determining the corresponding velocity and volumetric flow of the liquid.

    SAFETY

    • This product is not approved for use with petroleum products (diesel fuel, unleaded gasoline, jet fuel, kerosene, etc.), aromatic hydrocarbons, flammable fluids or other incompatible chemicals

    • This product is not approved for use in hazardous locations.

    • Be sure O-rings and seals are kept in good repair.

    • When applying power, adhere to specifications listed in appropriate electronics manual.

    • Disconnect external power before attaching or detaching input or output wires.

  • 3

    INSTALLATION

    EARTH GROUND

    When making installations, the magnetic flow meter grounding rules have to be observed. The sensors of the QSE flow meter are sensitive to any electrical noise that is always present in piping systems. The fluid in piping systems usually carries significant levels of static electricity and should be grounded for the QSE meter to be at its best performance.

    As part of the installation, it is important to understand the importance of having the QSE meter connected to “earth ground”. Earth grounding helps ensure electronic component stability and reliability by using the earth to absorb any static charge buildup or spurious signal noise that can affect the meter electronics. Earth ground in this installation is NOT the ground wire in the customer isolated power supply.

    Process Fluid Grounding

    Establishing a process fluid ground is one of the most important installation details. Proper process fluid grounding ensures that the sensor and fluid are at the same electrical potential so that only the induced flow signal is measured. The process fluid grounding provides a stable base potential reference for measuring the induced flow signal (the differential between the base potential reference and the fluid flow potential created by the sensor magnetic field). Process fluid grounding should be established on both sides (upstream and downstream) of the sensor elements.

    By connecting the QSE sensor, the fluid, and the reference used by the internal electronics to a stable, noise free reference point (earth ground), the user is ensured of getting the best performance from their magnetic flowmeter.

    Each QSE meter has a ground lug with a removable ring terminal located in the circular wall adjacent to the outlet end of the meter. This ground lug is connected internally to the electronics and sensors and is the meter connection to earth ground (the base potential reference).

    Grounding Summary:

    1. Ground sensor to meter ground lug. (Done at factory.)

    2. Ground electronics to meter ground lug. (Done at factory.)

    3. Ground process fluid to meter ground lug. (Customer responsibility.) Various process fluid grounding examples are shown on the following pages (see Figures 1,2, 3, and 4) to assist you in completing this item. The examples cover various meter installation situations.

    4. Ground meter ground lug to earth ground. (Customer responsibility)

    All of the examples on the following pages show 14-16 AWG ground wire (customer supplied) being used. A ground wire should be crimped to the ring terminal on the meter ground lug and the other end of the ground wire connected to earth ground (item 4 above).

  • 4

    Figure 2

    Figure 1

    A

    CONNECT WIRE TO RING TERMINAL.TIGHTEN RING TERMINAL WITH SCREW.(3 PLACES)

    EARTH GROUNDSIDE VIEW WITH CUT-AWAY

    DETAIL A CUT-AWAY VIEW(TYPICAL 2 PLACES)

    SCREW (HAND TIGHT)

    RING TERMINAL

    GROUNDING PROBE(APPLY THREAD TAPEOR PIPE SEALANTPRIOR TO INSTALL)

    1/8 - 27 NPT(TO BE TAPPEDINTO PORT CAVITY)

    Ø.312 (5/16) HOLE(TO BE DRILLED THRUFLOOR OF PORT CAVITY)

    14-16 AWGGROUNDING WIRE

    QSE GROUNDING - THREADED FITTINGS & ANSI FLANGEALL PROCESS PIPE MATERIALS

    USING GPI GROUNDING PROBES (Kit P/N 145630-529)

    QSE GROUNDING - THREADED FITTINGSNON-CONDUCTIVE (PLASTIC) PIPE

    USING COMMERCIALLY AVAILABLE METAL COUPLINGS

    QSE GROUNDING - ANSI FLANGENON-CONDUCTIVE (PLASTIC) PIPE

    USING COMMERCIALLY AVAILABLE GROUNDING RINGS

    EARTH GROUND

    14-16 AWGGROUNDING WIRE

    CONDUCTIVE PIPE COUPLING(STEEL, IRON, BRASS,ALUMINUM, COPPER, ETC.)(2 PLACES)

    HOSE CLAMP(2 PLACES)

    QSE GROUNDING - ANSI FLANGECONDUCTIVE METAL PIPE

    USING INTEGRATED CONTINUITY GROUND WIRE

    METER FLANGE (TYP)

    GASKET (TYP) GASKET (TYP)

    GROUNDING RING (TYP)(COMMERCIALLY AVAILABLE)

    EARTH GROUND

    CONNECT WIRE TO RING TERMINAL.TIGHTEN RING TERMINAL WITH SCREW.

    14-16 AWGGROUNDING WIRE

    INSERT WIRE UNDER HOSECLAMP AND TIGHTEN CLAMP.(2 PLACES)

    CONNECT WIRE TO RING TERMINAL.TIGHTEN RING TERMINAL WITH SCREW.

    EARTH GROUND

    GROUNDING SCREW (TYP)

    CONNECT WIRE TO RING TERMINAL.TIGHTEN RING TERMINAL WITH SCREW.

    METER FLANGE (TYP) GASKET (TYP)14-16 AWG

    GROUNDING WIRE

    14-16 AWG GROUNDING WIRE TOMAINTAIN GROUNDING CONTINUITYBETWEEN PIPE SEGMENTS.

    PLASTIC PIPE & FLANGE (TYP)

    METAL PIPE & FLANGE (TYP)

    GROUNDING SCREW (TYP)

    QSE GROUNDING LUG

    QSE GROUNDING LUG

    QSE GROUNDINGLUG

    QSE GROUNDING LUG

    A

    CONNECT WIRE TO RING TERMINAL.TIGHTEN RING TERMINAL WITH SCREW.(3 PLACES)

    EARTH GROUNDSIDE VIEW WITH CUT-AWAY

    DETAIL A CUT-AWAY VIEW(TYPICAL 2 PLACES)

    SCREW (HAND TIGHT)

    RING TERMINAL

    GROUNDING PROBE(APPLY THREAD TAPEOR PIPE SEALANTPRIOR TO INSTALL)

    1/8 - 27 NPT(TO BE TAPPEDINTO PORT CAVITY)

    Ø.312 (5/16) HOLE(TO BE DRILLED THRUFLOOR OF PORT CAVITY)

    14-16 AWGGROUNDING WIRE

    QSE GROUNDING - THREADED FITTINGS & ANSI FLANGEALL PROCESS PIPE MATERIALS

    USING GPI GROUNDING PROBES (Kit P/N 145630-529)

    QSE GROUNDING - THREADED FITTINGSNON-CONDUCTIVE (PLASTIC) PIPE

    USING COMMERCIALLY AVAILABLE METAL COUPLINGS

    QSE GROUNDING - ANSI FLANGENON-CONDUCTIVE (PLASTIC) PIPE

    USING COMMERCIALLY AVAILABLE GROUNDING RINGS

    EARTH GROUND

    14-16 AWGGROUNDING WIRE

    CONDUCTIVE PIPE COUPLING(STEEL, IRON, BRASS,ALUMINUM, COPPER, ETC.)(2 PLACES)

    HOSE CLAMP(2 PLACES)

    QSE GROUNDING - ANSI FLANGECONDUCTIVE METAL PIPE

    USING INTEGRATED CONTINUITY GROUND WIRE

    METER FLANGE (TYP)

    GASKET (TYP) GASKET (TYP)

    GROUNDING RING (TYP)(COMMERCIALLY AVAILABLE)

    EARTH GROUND

    CONNECT WIRE TO RING TERMINAL.TIGHTEN RING TERMINAL WITH SCREW.

    14-16 AWGGROUNDING WIRE

    INSERT WIRE UNDER HOSECLAMP AND TIGHTEN CLAMP.(2 PLACES)

    CONNECT WIRE TO RING TERMINAL.TIGHTEN RING TERMINAL WITH SCREW.

    EARTH GROUND

    GROUNDING SCREW (TYP)

    CONNECT WIRE TO RING TERMINAL.TIGHTEN RING TERMINAL WITH SCREW.

    METER FLANGE (TYP) GASKET (TYP)14-16 AWG

    GROUNDING WIRE

    14-16 AWG GROUNDING WIRE TOMAINTAIN GROUNDING CONTINUITYBETWEEN PIPE SEGMENTS.

    PLASTIC PIPE & FLANGE (TYP)

    METAL PIPE & FLANGE (TYP)

    GROUNDING SCREW (TYP)

    QSE GROUNDING LUG

    QSE GROUNDING LUG

    QSE GROUNDINGLUG

    QSE GROUNDING LUG

  • 5

    Figure 4

    Figure 3

    A

    CONNECT WIRE TO RING TERMINAL.TIGHTEN RING TERMINAL WITH SCREW.(3 PLACES)

    EARTH GROUNDSIDE VIEW WITH CUT-AWAY

    DETAIL A CUT-AWAY VIEW(TYPICAL 2 PLACES)

    SCREW (HAND TIGHT)

    RING TERMINAL

    GROUNDING PROBE(APPLY THREAD TAPEOR PIPE SEALANTPRIOR TO INSTALL)

    1/8 - 27 NPT(TO BE TAPPEDINTO PORT CAVITY)

    Ø.312 (5/16) HOLE(TO BE DRILLED THRUFLOOR OF PORT CAVITY)

    14-16 AWGGROUNDING WIRE

    QSE GROUNDING - THREADED FITTINGS & ANSI FLANGEALL PROCESS PIPE MATERIALS

    USING GPI GROUNDING PROBES (Kit P/N 145630-529)

    QSE GROUNDING - THREADED FITTINGSNON-CONDUCTIVE (PLASTIC) PIPE

    USING COMMERCIALLY AVAILABLE METAL COUPLINGS

    QSE GROUNDING - ANSI FLANGENON-CONDUCTIVE (PLASTIC) PIPE

    USING COMMERCIALLY AVAILABLE GROUNDING RINGS

    EARTH GROUND

    14-16 AWGGROUNDING WIRE

    CONDUCTIVE PIPE COUPLING(STEEL, IRON, BRASS,ALUMINUM, COPPER, ETC.)(2 PLACES)

    HOSE CLAMP(2 PLACES)

    QSE GROUNDING - ANSI FLANGECONDUCTIVE METAL PIPE

    USING INTEGRATED CONTINUITY GROUND WIRE

    METER FLANGE (TYP)

    GASKET (TYP) GASKET (TYP)

    GROUNDING RING (TYP)(COMMERCIALLY AVAILABLE)

    EARTH GROUND

    CONNECT WIRE TO RING TERMINAL.TIGHTEN RING TERMINAL WITH SCREW.

    14-16 AWGGROUNDING WIRE

    INSERT WIRE UNDER HOSECLAMP AND TIGHTEN CLAMP.(2 PLACES)

    CONNECT WIRE TO RING TERMINAL.TIGHTEN RING TERMINAL WITH SCREW.

    EARTH GROUND

    GROUNDING SCREW (TYP)

    CONNECT WIRE TO RING TERMINAL.TIGHTEN RING TERMINAL WITH SCREW.

    METER FLANGE (TYP) GASKET (TYP)14-16 AWG

    GROUNDING WIRE

    14-16 AWG GROUNDING WIRE TOMAINTAIN GROUNDING CONTINUITYBETWEEN PIPE SEGMENTS.

    PLASTIC PIPE & FLANGE (TYP)

    METAL PIPE & FLANGE (TYP)

    GROUNDING SCREW (TYP)

    QSE GROUNDING LUG

    QSE GROUNDING LUG

    QSE GROUNDINGLUG

    QSE GROUNDING LUG

    A

    CONNECT WIRE TO RING TERMINAL.TIGHTEN RING TERMINAL WITH SCREW.(3 PLACES)

    EARTH GROUNDSIDE VIEW WITH CUT-AWAY

    DETAIL A CUT-AWAY VIEW(TYPICAL 2 PLACES)

    SCREW (HAND TIGHT)

    RING TERMINAL

    GROUNDING PROBE(APPLY THREAD TAPEOR PIPE SEALANTPRIOR TO INSTALL)

    1/8 - 27 NPT(TO BE TAPPEDINTO PORT CAVITY)

    Ø.312 (5/16) HOLE(TO BE DRILLED THRUFLOOR OF PORT CAVITY)

    14-16 AWGGROUNDING WIRE

    QSE GROUNDING - THREADED FITTINGS & ANSI FLANGEALL PROCESS PIPE MATERIALS

    USING GPI GROUNDING PROBES (Kit P/N 145630-529)

    QSE GROUNDING - THREADED FITTINGSNON-CONDUCTIVE (PLASTIC) PIPE

    USING COMMERCIALLY AVAILABLE METAL COUPLINGS

    QSE GROUNDING - ANSI FLANGENON-CONDUCTIVE (PLASTIC) PIPE

    USING COMMERCIALLY AVAILABLE GROUNDING RINGS

    EARTH GROUND

    14-16 AWGGROUNDING WIRE

    CONDUCTIVE PIPE COUPLING(STEEL, IRON, BRASS,ALUMINUM, COPPER, ETC.)(2 PLACES)

    HOSE CLAMP(2 PLACES)

    QSE GROUNDING - ANSI FLANGECONDUCTIVE METAL PIPE

    USING INTEGRATED CONTINUITY GROUND WIRE

    METER FLANGE (TYP)

    GASKET (TYP) GASKET (TYP)

    GROUNDING RING (TYP)(COMMERCIALLY AVAILABLE)

    EARTH GROUND

    CONNECT WIRE TO RING TERMINAL.TIGHTEN RING TERMINAL WITH SCREW.

    14-16 AWGGROUNDING WIRE

    INSERT WIRE UNDER HOSECLAMP AND TIGHTEN CLAMP.(2 PLACES)

    CONNECT WIRE TO RING TERMINAL.TIGHTEN RING TERMINAL WITH SCREW.

    EARTH GROUND

    GROUNDING SCREW (TYP)

    CONNECT WIRE TO RING TERMINAL.TIGHTEN RING TERMINAL WITH SCREW.

    METER FLANGE (TYP) GASKET (TYP)14-16 AWG

    GROUNDING WIRE

    14-16 AWG GROUNDING WIRE TOMAINTAIN GROUNDING CONTINUITYBETWEEN PIPE SEGMENTS.

    PLASTIC PIPE & FLANGE (TYP)

    METAL PIPE & FLANGE (TYP)

    GROUNDING SCREW (TYP)

    QSE GROUNDING LUG

    QSE GROUNDING LUG

    QSE GROUNDINGLUG

    QSE GROUNDING LUG

  • 6

    TYPICAL INSTALLATION

    CONNECTIONS

    Install your meter in-line with either horizontal flow or vertical flow. The best meter position for horizontal flow setups is with the meter rotated slightly (about 1 o’clock or 2 o’clock) to tilt the top from the horizontal plane (see Figure 5). This prevents sediment from settling on the lower set of sensing electrodes. Install the meter with the flow arrow pointing in the direction of fluid flow.

    5 x DIA.

    3 x DIA.

    5 x DIA. 3 x DIA.

    FLOW DIRECTION

    FLO

    W D

    IRE

    CT

    ION

    20˚

    Electrode Locations

    5 x DIA.

    3 x DIA.

    5 x DIA. 3 x DIA.

    FLOW DIRECTION

    FLO

    W D

    IRE

    CT

    ION

    Figure 5

    Meter shown tilted 20 degrees in horizontal position

    Plan to install meter with minimum straight pipe lengths at inlet and outlet ends. The straight run lengths noted below represent the minimum requirements for accurate flow measurement (see Figure 5). For optimum performance, provide as much additional straight run as possible.

    • Upstream from the meter, allow a minimum straight pipe length of (5) times the pipe diameter from bends and obstructions.

    • Downstream from the turbine, allow a minimum straight pipe length of (3) times the pipe diameter from bends and obstructions.

    • Avoid downward flow that can lead to partially filled pipes.

  • 7

    For NPT Fittings: Seal all pipe threads with an appropriate non-lubricated thread sealant (such as Loctite® No More Leaks™ Plastic Pipe Thread Sealant or NSF equivalent for NSF applications). Make sure the thread sealant does not intrude into the flow path. Hand tighten the meter at the housing ends. Do not use a wrench or similar tool to tighten as this can damage the housing.

    NOTE: If connecting to new female pipe threads, burrs and curls can adversely affect accuracy. Correct the problem prior to meter installation.

    For BSPP Fittings: The BSPP meter includes a plastic bag containing (2) O-rings and (2) backup rings that must be installed onto the meter. They are used for sealing the fitting ends of the meter to process piping. Before installing the meter, install an O-ring in the O-ring gland on each housing end, then fit a backup ring over each O-ring (see Figure 6). The BSPP meter is now ready for installation. Hand tighten the meter at the housing ends. DO NOT use a wrench or similar tool to tighten as this can damage the housing.

    1. Install O-ring 2. Install backup ring

    Installed

    Flush to center O-ring with housing

    cut-away view

    Figure 6

    For ANSI and DIN Flange Fittings: The flanges supplied with your meter allow the meter to be oriented regardless of the mating flanges position. The meter flanges and their steel half-ring “keys” must be installed onto the meter by the customer before meter installation (see Figure 7).

    LOBE

    RECESS

    (TOP & BOTTOM)

    FLANGE INSTALLATION

    (POLYMER FLANGE SHOWN)

    CUT-AWAY

    Figure 7

    • Installing flanges on meter: With the mating face of the flange facing outboard, slide flange over one end of the meter. Position (2) half-rings in the retaining groove

  • 8

    and snap into place. Each end of a half-ring has a small lobe that snaps into a recess at the top and bottom of the groove for retention. Repeat at other end.

    Customer to provide:

    • Ring Gaskets or Full-Face Gaskets approved for use with type flange installed (ANSI or DIN) and the fluid being monitored (2 required).

    • 5/8 in. bolts and nuts for ANSI flanges. Four per side for 3-inch meters; eight per side for 4-inch meters.

    • 16mm bolts and nuts for DIN flanges. Four per side for 3” meters (DN80/PN10 flanges); Eight per side for 4” meters (DN100/PN10 flanges).

    • Torque bolts using a star pattern to 25 ft-lbs (33.9 N•m). For best results, always verify torque accuracy before use.

    NOTE: Do not over tighten the flange bolts. This may cause the gasket to be compressed into the flow stream and may decrease the accuracy of the meter.

    When properly installed, the flow meter will only measure flow in that portion of the piping system where the meter is installed. Choose the location with the longest straight unobstructed run of pipe, keeping in mind that in some applications it may be possible to locate the meter in either the supply or return pipe.

    NOTE: Your meter may have two 1/8 in. NPT pipe plugs installed on top of the flow tube near each end. They have been installed and sealed at the factory and are for use only with optional temperature sensing probes for energy use calculations in specialized applications.

    RECOMMENDED INSTALLATION

    WARNING: Compatibility of this product’s material and the process fluid and/or environment should be considered prior to putting into service.

    WARNING: Product should never be operated outside its published specifications for temperature or pressure. See specifications for your model.

    WARNING: Make sure flow and pressure have been eliminated from process pipe prior to installing or removing product.

    WARNING: When connecting product to process piping: For NPT fittings use non-lubricated thread sealant (such as Loctite® No More Leaks™ Plastic Pipe Thread Sealant or NSF equivalent for NSF applications). For BSPP fittings use O-ring seals. For flange fittings use flange gaskets.

    CAUTION: Installation near high electromagnetic fields and high current fields is not recommended and may result in inaccurate readings.

    CAUTION: Do not allow water to freeze in meter. Ice expansion may burst the plastic housing.

    CAUTION: Do not allow this meter to be used with steam.

    CAUTION: Using metal fittings on the plastic threaded pipe ends of the meter is not recommended.

  • 9

    WIRING

    All electronic options are associated with a matching style of meter cover plate. This cover plate has four threaded ports, compatible with PG7 threads, for gaining wiring access to the electronics inside the cover plate. The meter is shipped with the ports environmentally sealed with a threaded plug and seal. Remove one or more of these plugs as required to install the supplied port fittings below.

    Each meter is supplied with cable gland strain reliefs with O-rings and 1/2 NPT adapters with seals for use in the threaded ports of the meter cover plate. Select the port fittings that fit your process and replace the threaded plugs in the cover plate with the fittings as required. The threaded plugs installed at the factory may be left in any unused cover plate port indefinitely.

    It is recommended that a “removable” thread-locker (fluid, stick, tape, spray, etc.) be used when installing the strain reliefs or adapters into the cover plate ports.

    Strain Reliefs: The cable gland strain reliefs will accommodate a cable diameter of 0.11 - 0.26 inches (2.79 - 6.6mm) and provide an environmental seal around the cable when the dome-nut is tightened.

    NPT Adapters: The 1/2 NPT adapter fittings are used for attaching flex conduit to the meter, for those applications that require cable runs to be enclosed in conduit.

    • Cable to be provided by customer to accommodate job requirements. Cable is not included with meter.

    • This meter is externally powered. The magnetic coils, electrodes and other electronic components within the main meter body receive power from the electronics housed within the cover plate. All external wiring connects to the cover plate through threaded ports. See the included electronics manuals for meter wiring diagrams specific to this meters electronics.

    MAINTENANCE

    The meter is virtually maintenance-free. However, it is important to keep the meter clean and free of contaminants.

    CAUTION: Do not allow liquids to dry inside the meter. The electrodes may develop a film that degrades accuracy.

    Remove internal debris or deposits using soft brush or small probe.

    NOTE: Make sure the arrow on the meter is pointed in the direction of fluid flow (see Figure 2).

  • 10

    TROUBLESHOOTING

    MEASUREMENT IS NOT ACCURATE

    PROBABLE CAUSE SOLUTION

    Debris/particles in liquid Need proper filtration

    Air in liquid - No back pressureIncrease back pressure on meter to eliminate air

    Air in liquid - Plumbing installationInstall meter away from other fittings or flow obstructions. Do not allow meter to discharge to atmosphere.

    Air in liquid - Pump cavitation Install meter away from pump

    Pipe not full - No back pressureIncrease back pressure on meter to eliminate air.

    Pipe not full - Plumbing installationInstall meter where pipe is always full of liquid. Do not allow meter to discharge to atmosphere.

    Conductivity of fluid too low Consult liquid properties

    Operating outside meter limitsIncrease/decrease flow rate to proper meter specification. Review temperature limits.

    Electrical noiseInstall meter away from devices that emit EMF.

    Ground loopMeter may need to be earth grounded

    Electrodes not clean Clean electrode tips

    Interface device not properly calibrated

    Consult device instructions and specifications.

    NORMAL FLOW BUT METER DOES NOT WORK

    PROBABLE CAUSE SOLUTION

    Wiring not correctConsult wiring diagrams. Review installation.

    Operating outside meter limits Review voltage requirements

    Meter installed backwardsCheck to see if the direction of the arrows on the meter match flow direction.

  • 11

    Design Type: Electromagnetic

    INLET AND OUTLET

    NPT MODELS

    QSE05NPT 1/2 inch NPT

    QSE07NPT 3/4 inch NPT

    QSE10NPT 1 inch NPT

    QSE15NPT 1-1/2 inch NPT

    QSE20NPT 2 inch NPT

    BSPP MODELS

    QSE05BSP1/2 inch BSPP (ISO 228 - G 1/2)

    QSE07BSP3/4 inch BSPP (ISO 228 - G 3/4)

    QSE10BSP1 inch BSPP (ISO 228 - G 1)

    QSE15BSP1-1/2 inch BSPP (ISO 228 - G 1 1/2)

    QSE20BSP2 inch BSPP (ISO 228 - G 2)

    ANSI FLANGE MODELS

    QSE30FAP3 inch 150 # ANSI Flange (Polymer)

    QSE40FAP4 inch 150 # ANSI Flange (Polymer)

    QSE30FAS3 inch 150 # ANSI Flange (Steel)

    QSE40FAS4 inch 150 # ANSI Flange (Steel)

    DIN FLANGE MODELS

    QSE30FDS3 inch DN80/PN10 Flange (Steel)

    QSE40FDS4 inch DN100/PN10 Flange (Steel)

    WETTED COMPONENTS

    HousingNORYL™ GFN3 PPE+PS

    Electrodes316L Stainless Steel

    Temperature Probes

    316 Series Stainless Steel

    Pipe Plugs300 Series Stainless Steel

    O-RingsEPDM (Ethylene Propylene Diene Monomer)

    MAXIMUM WORKING PRESSURE

    De-rate maximum working pressure 0.682 PSIG per each degree °F above 70° F

    NPT150 PSIG @70° F 10.3 BAR @ 21° C

    BSPP150 PSIG @70° F 10.3 BAR @ 21° C

    ANSI FLANGE150 PSIG @70° F 10.3 BAR @ 21° C

    DIN FLANGE150 PSIG @70° F 10.3 BAR @ 21° C

    POWER SUPPLY

    Voltage Requirement

    Min. 12 VDC or 12 VAC, 50-60 Hz (Lower voltage will cause inaccurate readings)

    Max. 36 VDC or 30 VAC, 50-60 Hz (higher voltage may damage unit)

    ELECTRICAL

    Pollution Degree

    2

    Installation Category

    1

    Altitude 20000m Max.

    IP67 rated enclosure. CSA approved for indoor use only

    See electronics manuals for electronic specifications.

    SPECIFICATIONS

  • 12

    U.S. MEASUREMENT

    Unit of Measure: Gallon

    TURNDOWN 60:1

    ACCURACY

    Line SizeFlow Range

    (GPM)ACCURACY ±Uncertainty

    Typ. K-factor

    (PPG)

    1/2 inch0.16 to 10 gpm (.25 to 15 fps)

    ± 0.5% of Reading

    ± 0.023 fps

    ± .015 gpm

    4347

    3/4 inch0.33 to 20 gpm (.25 to 15 fps)

    ± 0.5% of Reading

    ± 0.012 fps

    ± .015 gpm

    1937

    1 inch0.67 to 40 gpm (.25 to 15 fps)

    ± 0.5% of Reading

    ± 0.006 fps

    ± .015 gpm

    1089

    1 1/2 inch

    1.33 to 80 gpm (.25 to 15 fps)

    ± 0.5% of Reading

    ± 0.003 fps

    ± .015 gpm

    484.1

    2 inch2.5 to 150 gpm (.25 to 15 fps)

    ± 0.5% of Reading

    ± 0.003 fps

    ± .030 gpm

    400

    3 inch5 to 300 gpm (.25 to 15 fps)

    ± 0.5% of Reading

    ± 0.003 fps

    ± .060 gpm

    121

    4 inch10 to 600 gpm (.25 to 15 fps)

    ± 0.5% of Reading

    ± 0.003 fps

    ± .120 gpm

    68.1

    FLUID OPERATING TEMPERATURE:

    +32° F to +210° F (NPT & BSPP)

    +32° F to +180° F (Polymer and Steel Flanges)

    (Do not allow fluid to freeze inside meter.)

    STORAGE TEMPERATURE:

    –40° F to +228° F

    AMBIENT AIR OPERATING

    TEMPERATURE

    0° F to +140° F ALL METERS

    There is a correlation between ambient air temperature and maximum fluid operating temperature. As ambient air temperature increases, the maximum fluid operating temperature decreases. See Maximum Temperature Conditions graph named “TEMPERATURES” on page 16 to verify that your process is within allowable meter operating limits.

  • 13

    METRIC MEASUREMENT

    Unit of Measure: Litre

    TURNDOWN 60:1

    ACCURACY

    Line Size

    Flow Range (L/min)

    ACCURACY ±UncertaintyTyp.

    K-factor (PPL)

    1/2 inch

    0.63 to 38 L/min [.076 to 4.57 m/s]

    ± 0.5% of Reading

    ± 7.0 mm/s

    ± .057 L/min

    1148.5

    3/4 inch

    1.27 to 76 L/min [.076 to 4.57 m/s]

    ± 0.5% of Reading

    ± 3.7 mm/s

    ± .057 L/min

    511.8

    1 inch2.52 to 151 L/min [.076 to 4.57 m/s]

    ± 0.5% of Reading

    ± 2.0 mm/s

    ± .057 L/min

    287.7

    1 1/2 inch

    5.05 to 303 L/min [.076 to 4.57 m/s]

    ± 0.5% of Reading

    ± 1.0 mm/s

    ± .057 L/min

    127.9

    2 inch9.47 to 568 L/min [.076 to 4.57 m/s]

    ± 0.5% of Reading

    ± 1.0 mm/s

    ± .114 L/min

    105.7

    3 inch19 to 1140 L/min [.076 to 4.57 m/s]

    ± 0.5% of Reading

    ± 1.0 mm/s

    ± .227 L/min

    30

    4 inch38 to 2270 L/min [.076 to 4.57 m/s]

    ± 0.5% of Reading

    ± 1.0 mm/s

    ± .454 L/min

    18

    FLUID OPERATING TEMPERATURE:

    0° C to +98° C (NPT & BSPP)

    0° C to +82° C (Polymer and Steel Flanges)

    (Do not allow fluid to freeze inside meter.)

    STORAGE TEMPERATURE:

    –40° C to +108° C

    AMBIENT AIR OPERATING

    TEMPERATURE

    -18° C to +60° C ALL METERS

    There is a correlation between ambient air temperature and maximum fluid operating temperature. As ambient air temperature increases, the maximum fluid operating temperature decreases. See Maximum Temperature Conditions graph named “TEMPERATURES” on page 16 to verify that your process is within allowable meter operating limits.

  • 14

    PRODUCT WEIGHT – kg:*

    NPT / BSPPANSI Polymer

    FlangeANSI Steel

    FlangeDIN Steel

    Flange

    1/2 in. 1.1 - - -

    3/4 in. 1.2 - - -

    1 in. 1.2 - - -

    1 1/2 in. 1.8 - - -

    2 in. 2.0 - - -

    3 in. - 6.4 11.1 10.2

    4 in. - 7.4 13.4 10.8

    * Weight with display. For plain cover plates, subtract 0.09 kg.

    PRODUCT WEIGHT – lb:*

    NPT / BSPPANSI Polymer

    FlangeANSI Steel

    FlangeDIN Steel

    Flange

    1/2 in. 2.5 - - -

    3/4 in. 2.6 - - -

    1 in. 2.7 - - -

    1 1/2 in. 3.9 - - -

    2 in. 4.3 - - -

    3 in. - 14 24.4 22.5

    4 in. - 16.3 29.5 23.7

    * Weight with display. For plain cover plates, subtract 0.2 lb.

    U.S. PRODUCT WEIGHT

    METRIC PRODUCT WEIGHT

  • 15

    C A

    NPT / BSPP WITH 09 ELECTRONICS NPT / BSPP WITH QB ELECTRONICS

    B DE

    QSE METER DIMENSIONS (NPT, BSP, FLANGE)

    METER SIZE & FITTING A B C D E

    1/2 in. NPT & BSP 5.20 [132.1] 5.83 [148.1] 10.50 [266.7] 5.13 [130.3] 1.83 [46.5]

    3/4 in. NPT & BSP 5.20 [132.1] 5.83 [148.1] 10.75 [273.1] 5.13 [130.3] 1.83 [46.5]

    1 in. NPT & BSP 5.20 [132.1] 5.83 [148.1] 11.00 [279.4] 5.13 [130.3] 1.83 [46.5]

    1 1/2 in. NPT & BSP 5.22 [132.6] 6.95 [176.5] 11.00 [279.4] 6.25 [158.8] 2.37 [60.2]

    2 in. NPT & BSP 5.22 [132.6] 6.95 [176.5] 11.00 [279.4] 6.25 [158.8] 2.37 [60.2]

    3 in Flange 7.50 [190.5] 9.62 [244.3] 12.00 [304.8] 8.92 [226.6] 3.25 [82.6]

    4in. Flange 9.00 [228.6] 10.37 [263.4] 12.00 [304.8] 9.67 [245.6] 4.50 [114.3]

    FLANGE WITH 09 ELECTRONICS FLANGE WITH QB ELECTRONICS

    A

    B

    E

    D

    C

    MILLIMETER DIMENSIONS SHOWN IN BRACKETS

    6.31[160.2]

    ALL METERS

    DIMENSIONS

    QSE METER DIMENSIONS (NPT, BSPP, ANSI FLANGE, DIN FLANGE) listed in inches; millimeters listed in [ ] brackets

    METER SIZE & FITTING A B C D E

    1/2 in. NPT & BSPP 5.20 [132.1] 5.83 [148.1] 10.50 [266.7] 5.13 [130.9] 1.83 [46.5]

    3/4 in. NPT & BSPP 5.20 [132.1] 5.83 [148.1] 10.75 [w273.1] 5.13 [130.9] 1.83 [46.5]

    1 in. NPT & BSPP 5.20 [132.1] 5.83 [148.1] 11.00 [279.4] 5.13 [130.9] 1.83 [46.5]

    1-1/2 in. NPT & BSPP 5.22 [132.6] 6.95 [176.5] 11.00 [279.4] 6.25 [158.8] 2.37 [60.2]

    2 in. NPT & BSPP 5.22 [132.6] 6.95 [176.5] 11.00 [279.4] 6.25 [158.8] 2.37 [60.2]

    3 in. ANSI Flange 7.50 [190.5] 9.62 [244.3] 12.00 [304.8] 8.92 [226.6] 3.75 [95.3]

    4 in. ANSI Flange 9.00 [228.6] 10.37 [263.4] 12.00 [304.8] 9.67 [245.6]4.50

    [114.3]

    3 in. DIN Flange 7.87 [200.0] 9.80 [248.9] 12.00 [304.8] 9.10 [231.1] 3.94 [100.0]

    4 in. DIN Flange 8.66 [220.0] 10.2 [259.1] 12.00 [304.8] 9.5 [241.3] 4.33 [110.0]

  • 16

    TEMPERATURES

  • 17

    The useful unit for seawater is milimhos/cm (mS/cm); seawater is around 55 mS/cm. The useful unit for freshwater is micromhos/cm (µmhos/cm, or µS/cm); tap water ranges between 50 and 800 µS/cm (depending on the source).

    The basic unit of fluid conductivity is “mho/cm”, otherwise known as 1 Siemen. However, this unit does not really occur in water, so we typically use one thousandth (mili-) or one millionths (micro-) of it for natural waters (1000 milimhos and 1,000,000 micromhos are equal to one mho).

    FLUID ELECTRICAL CONDUCTIVITY

    Fluid Electrical Conductivity Reference:

    Electrical Conductivity of Common Fluids

    FLUIDTEMPERATURE

    (°F)CONDUCTIVITY

    (µs/cm)

    Coca Cola Syrup 68 600

    Coffee Extract 183.2 5000

    Corn Syrup 89.6 16

    Ethylene Glycol (Pure) 68 1.07

    Gin 90 Proof 77 10

    ISO-propyl Alcohol 77 3.5

    Molasses 50 300

    RC Cola Syrup 77 600

    Sugar Solution Dilute 86 585

    Urea 77 5000

    Vodka 100 Proof 77 4

    Water, Distilled - 0.04

    Water, NYC 77 72

    NOTE: This table of fluid conductivity is for reference only, to show the relative conductivity of various fluids. Since the list contains flammable fluids, it is not to be construed as a list of permissible fluids to be used with QSE series meters.

    NOTE: QSE series meters are for use with water, aqueous solutions and other non-flammable, electrically conductive fluids. Do not use the meter with petroleum products (diesel fuel, unleaded gasoline, jet fuel, kerosene, etc.) or other incompatible chemicals.

  • 18

    BSPP METERFITTING END

    920896-01REV. P

    L, N1, N2 & N3

    W

    T

    B

    A1 & A2

    U

    P1, P2 & P3

    S1, S2, S3, S4, S5 & S6(Polymer flanges shown)

    V1, V2, V3, V4 & V5

    R

    C, D, E, F, G, H & J

    K, M1, M2 & M3

    X

  • 19

    REPLACEMENT PARTS LIST

    REF. PART NUMBER DESCRIPTIONNO

    REQ’D.

    A1 145500-01 Kit, 1 in. temperature sensor probe Includes: (2) probes, (2) strain reliefs w/O-rings 1

    A2 145500-02 Kit, 2 in. temperature sensor probe Includes: (2) probes, (2) strain reliefs w/O-rings. 1

    B 145500-03 Kit, pipe plug, 1/8-27 NPT stainless steel Includes: (2) plugs. 1

    C 145501-01 Kit, Q09 computer display, 1/2 in. Includes: Computer display w/decal, seal, 10 pin connector. 1

    D 145501-02 Kit, Q09 computer display, 3/4 in. Includes: Computer display w/decal, seal, 10 pin connector. 1

    E 145501-03 Kit, Q09 computer display, 1 in. Includes: Computer display w/decal, seal, 10 pin connector. 1

    F 145501-04 Kit, Q09 computer display, 1-1/2 in. Includes: Computer display w/decal, seal, 10 pin connector. 1

    G 145501-05 Kit, Q09 computer display, 2 in. Includes: Computer display w/decal, seal, 10 pin connector. 1

    H 145501-06 Kit, Q09 computer display, 3 in. Includes: Computer display w/decal, seal, 10 pin connector. 1

    J 145501-07 Kit, Q09 computer display, 4 in. Includes: Computer display w/decal, seal, 10 pin connector. 1

    K 145500-14Kit, QSB w/display cover plate. Includes: Display cover plate, QSB electronics with ribbon cable, (6) screws, cover plate seal.

    1

    L 145500-15Kit, QSB w/plain cover plate. Includes: Plain cover plate w/decal, QSB electronics with ribbon cable, (6) screws, cover plate seal.

    1

    M1 145502-01Kit, QSI1 w/display cover plate. Includes: Display cover plate, QSI1 electronics with ribbon cable, (6) screws, cover plate seal.

    1

    M2 145502-02Kit, QSI2 w/display cover plate. Includes: Display cover plate, QSI2 electronics with ribbon cable, (6) screws, cover plate seal.

    1

    M3 145502-03Kit, QSI3 w/display cover plate. Includes: Display cover plate, QSI3 electronics with ribbon cable, (6) screws, cover plate seal.

    1

    N1 145503-01Kit, QSI1 w/plain cover plate. Includes: Plain cover plate w/decal, QSI1 electronics with ribbon cable, (6) screws, cover plate seal.

    1

    N2 145503-02Kit, QSI2 w/plain cover plate. Includes: Plain cover plate w/decal, QSI2 electronics with ribbon cable, (6) screws, cover plate seal.

    1

    N3 145503-03Kit, QSI3 w/plain cover plate. Includes: Plain cover plate w/decal, QSI3 electronics with ribbon cable, (6) screws, cover plate seal.

    1

  • 20

    P1 145500-04 Kit, O-rings, 1/2 in., 3/4 in. and 1 in. housings. Includes: (2) O-rings. 1

    P2 145500-05 Kit, O-rings, 1-1/2 in. and 2 in. housings. Includes: (2) O-rings. 1

    P3 145500-06 Kit, O-rings, 3 in. and 4 in. housings. Includes: (2) O-rings. 1

    R 145500-07 Kit, seal, cover plate. Includes: (1) Cover plate seal. 1

    S1 145500-08 Kit, ANSI flange, polymer, 3 in. Includes: (2) 3 in. polymer flanges, (4) 3 in. half-ring keys. 1

    S2 145500-09 Kit, ANSI flange, polymer, 4 in. Includes: (2) 4 in. polymer flanges, (4) 4 in. half-ring keys. 1

    S3 145500-10 Kit, ANSI flange, steel, 3 in. Includes: (2) 3 in. steel flanges, (4) 3 in. half-ring keys. 1

    S4 145500-11 Kit, ANSI flange, steel, 4 in. Includes: (2) 4 in. steel flanges, (4) 4 in. half-ring keys. 1

    S5 145500-16 Kit, DIN flange, steel, DN80, PN10. Includes: (2) 3 in. steel flanges, (4) 3 in. half-ring keys. 1

    S6 145500-17 Kit, DIN flange, steel, DN100, PN10. Includes: (2) 4 in. steel flanges, (4) 4 in. half-ring keys. 1

    T 145500-12 Kit, strain relief, cover plate. Includes: (4) PG7 thread strain reliefs, .11 - .26 cable diameter range 1

    U 145500-13 Kit, strain relief, temperature sensor. Includes: (2) PG7 thread strain reliefs (reduced fit), .08 - .20 cable diameter range 1

    V1 145500-20 Kit, O-ring & backup ring, 1/2 in. BSPP. Includes: (2) O-rings, (2) backup rings. 1

    V2 145500-21 Kit, O-ring & backup ring, 3/4 in. BSPP. Includes: (2) O-rings, (2) backup rings. 1

    V3 145500-22 Kit, O-ring & backup ring, 1 in. BSPP. Includes: (2) O-rings, (2) backup rings. 1

    V4 145500-23 Kit, O-ring & backup ring, 1-1/2 in. BSPP. Includes: (2) O-rings, (2) backup rings. 1

    V5 145500-24 Kit, O-ring & backup ring, 2 in. BSPP. Includes: (2) O-rings, (2) backup rings. 1

    W 145500-25 Kit, PG7 to 1/2 NPT adapter, cover plate. Includes: (4) PG7 to 1/2 NPT male adapters 1

    X 145177-501 Assembly, Ribbon Cable (9 INCH), Spare 1

    Not Shown 145630-529

    Kit, grounding probe Includes: (2) grounding probes, (2) screws, (4) terminal rings, instruction sheet.

    1

  • 21

    SERVICE

    For warranty consideration, contact your local distributor. If you need further assistance, contact the GPI Customer Service Department at:

    1-888-996-3837

    You will need to:

    • Provide information from the decal on your meter.

    • Receive a Return Authorization number.

    • Flush any fluid from the meter before shipping to the factory.

    CAUTION: Do not return the meter without specific authority from the GPI Customer Service Department. Due to strict regulations governing transportation, handling and disposal of hazardous or flammable liquids, GPI will not accept meters for rework unless they are completely free of liquid residue.

    The Waste Electrical and ElectronicEquipment (WEEE) direct ive(2002/96/EC) was approved by theEuropean Parliament and the Councilof the European Union in 2003. Thissymbol indicates that this productcontains electrical and electronicequipment that may include bat-teries, printed circuit boards, liquid

    crystal displays or other components that may be subjectto local disposal regulations at your location. Pleaseunderstand those regulations and dispose of this productin a responsible manner.

    RoHS Compliant (2011/65/EU)

    This product is in compliance with the RoHS Directive of the European Parliament and of the Council on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.

  • 22

    NOTES

  • 23

    NOTES

  • 920896-01 Rev T10/2018

    © 2018 Great Plains Industries, Inc., All Rights Reserved.Great Plains Industries, Inc. / 888-996-3837 / FLOMEC.net

    LIMITED WARRANTY

    Great Plains Industries, Inc. 5252 E. 36th Street North, Wichita, KS USA 67220-3205, hereby provides a limited warranty against defects in material and workmanship on all products manufactured by Great Plains Industries, Inc. This product includes a 1 year warranty.

    Manufacturer’s sole obligation under the foregoing warranties will be limited to either, at Manufacturer’s option, replacing or repairing defective Goods (subject to limitations hereinaf-ter provided) or refunding the purchase price for such Goods theretofore paid by the Buyer, and Buyer’s exclusive remedy for breach of any such warranties will be enforcement of such obligations of Manufacturer. The warranty shall extend to the purchaser of this product and

    to any person to whom such product is transferred during the warranty period.

    The warranty period shall begin on the date of manufacture or on the date of purchase with an original sales receipt. This warranty shall not apply if:

    A. the product has been altered or modified outside the warrantor’s duly appointed representative;

    B. the product has been subjected to neglect, misuse, abuse or damage or has been in-stalled or operated other than in accordance with the manufacturer’s operating instructions.

    To make a claim against this warranty, contact the GPI Customer Service Department at

    316-686-7361 or 888-996-3837. Or by mail at:Great Plains Industries, Inc.

    5252 E. 36th St. NorthWichita, KS, USA 67220-3205

    The company shall, notify the customer to either send the product, transportation prepaid, to the company at its office in Wichita, Kansas, or to a duly authorized service center. The

    company shall perform all obligations imposed on it by the terms of this warranty within 60 days of receipt of the defective product.

    GREAT PLAINS INDUSTRIES, INC., EXCLUDES LIABILITY UNDER THIS WARRANTY FOR DIRECT, INDIRECT, INCIDENTAL AND CONSEQUENTIAL DAMAGES INCURRED IN THE

    USE OR LOSS OF USE OF THE PRODUCT WARRANTED HEREUNDER.The company herewith expressly disclaims any warranty of merchantability or fitness for any

    particular purpose other than for which it was designed.This warranty gives you specific rights and you may also have

    other rights which vary from U.S. state to U.S. state.Note: In compliance with MAGNUSON MOSS CONSUMER WARRANTY ACT – Part 702

    (governs the resale availability of the warranty terms).

    IP67

  • 920896-03 Rev B03/2018

    Q09 Display

    Owner’s Manual

    Q09 Display(Shown on 1 inch NPT QSE Meter)

    TABLE OF CONTENTSINTRODUCTION ..................................... 1

    IMPORTANT NOTICE ....................... 2SAFETY ............................................ 2

    INSTALLATION ....................................... 2

    OPERATION ........................................... 4

    Display .............................................. 4Batch And Cumulative Totals ........... 4

    CALIBRATION ........................................ 4

    Factory Calibration ........................... 4Field Calibration ............................... 5

    USER CONFIGURATION ........................ 6

    Changing Configuration Settings ..... 6K-Factor Entry Field Calibration ....... 6

    MAINTENANCE ...................................... 6

    SPECIFICATIONS ................................... 7

    SERVICE ................................................. 8

    Returning Parts ................................ 8

    LIMITED WARRANTY ........................... 12

    INTRODUCTION

    The Q09 display is specifically designed to operate with the electronics housed within the cover plate. The display is mounted on a specialized cover plate containing other electronics and together they make up the total Q09 electronics package. The electronics in the cover plate (QSB [not available as adapter kit]) or (QSI1, QSI2 or QSI3) when externally powered, provides correctly adjusted power to the display via a 10 pin bridge connector. Information on the operation, capabilities and wiring of the electronics within the cover plate that is mated to your Q09 display is contained in its own dedicated manual (QSB, QSI1, QSI2, QSI3).

  • 2

    The display microprocessor electronics have extremely low power requirements and high data retention capabilities in both RAM and ROM. Information is clearly displayed on a large 6-digit LCD readout with two-point floating decimal for totals from .01 to 999,999. All operations are easily accessed with the two buttons on the front panel.

    The Q09 display is available mounted to the meter you ordered or as a kit that adapts the electronics to other GPI meters. The electronics inside the cover plate converts the voltage (or pulses) from the meter into usable information. This information, along with power, is supplied to the display that translates the information into calibrated flow units shown on the display’s readout.

    The display is a precision measuring instrument and should be handled as such. Refer to the Specifications Section to confirm required features.

    IMPORTANT NOTICE

    Q09 displays are very sensitive to electric noise if operated within 6 inches of some electric motors, relays, transformers or other sources of electronic noise.

    If the Q09 Display is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.

    SAFETY

    • This product is not approved for use in hazardous locations.

    • Be sure O-rings and seals are kept in good repair.

    • When applying power, adhere to specifications in this manual.

    • Disconnect external power before attaching or detaching input or output wires.

    WARNING: Product should never be operated outside its published specifications for temperature or pressure. See specifications for your model.

    CAUTION: Installation near high electromagnetic fields and high current fields is not recommended and may result in inaccurate readings.

    INSTALLATION

    If you ordered your Q09 display with a meter, it is installed at the factory. If you ordered your Q09 display separately from a meter (i.e. repair part), follow the instructions below to install on a QSE meter. If you ordered your Q09 display with one of our several types of meter adapter kits, follow the separate instructions enclosed with the kit to install the unit. In every case, please review and thoroughly understand all manuals and installation instructions before proceeding.

    • Make sure the seal is seated in its groove on the bottom of the display (see Figure 1).

    • Insert the long pins of the 10 pin bridge connector about half-way into the 10 pin PC board connector.

    • Then, align the 10 pin connector on the display with the short pins on the 10 pin bridge connector and engage pins using gentle force until the display is seated on the cover plate (see Figure 2).

  • 3

    • When the display is seated on the cover plate, the pins are fully engaged. Secure the display to the cover plate with the four screws at the corners of the faceplate. Make sure the seal is fully seated before tightening the screws.

    QSE meters are designed to measure flow in only one direction. If the display is upside down in your installation, remove the six screws retaining the cover plate to the meter, turn the cover plate with display as required and reinstall the screws. A ribbon cable connects the electronics within the cover plate to electronics within the meter body and allows 180 degrees of movement in either direction. Make sure the cover plate seal is fully seated before tightening the screws (see Figure 3).

    Figure 1

    SEAL

    COVER PLATE

    DISPLAY

    (4) SCREWS

    10 PIN BRIDGE

    CONNECTOR,

    SHORT PINS UP

    10 PIN PC BOARD

    CONNECTOR

    Figure 3

    COVER PLATESEAL

    Figure 2

    10 PIN BRIDGECONNECTOR

    10 PIN SOCKETCONNECTER

  • 4

    OPERATION

    DISPLAY

    All operations are reflected in the LCD readout. The large center digits indicate amounts, where smaller words or “icons” located above and below indicate specific information regarding totals, flow, calibration and units of measure.

    Activate the Meter: When power is supplied to the meter, the display automatically becomes active, is on continuously and always ready to perform. It remains active until meter power is disconnected.

    BATCH AND CUMULATIVE TOTALS

    The display maintains two totals. The Cumulative Total provides continuous measurement and cannot be manually reset. The Batch Total can be reset to measure flow during a single use. The Cumulative Total is labeled TOTAL 1, Batch Total is labeled TOTAL 2 BATCH.

    When the Cumulative Total reaches a display reading of 999,999 the display will highlight an X10 icon. This indicates to the operator that a zero must be added to the 6 digits shown. When the next rollover occurs, the display will highlight an X100 icon. This indicates to the operator that two zeros must be added to the 6 digits shown.

    Press the DISPLAY button briefly to switch between the TOTAL 1, TOTAL 2 BATCH and FLOWRATE. When TOTAL 2 BATCH is displayed, you can hold the DISPLAY button for 3 seconds to reset the Batch Total to zero.

    When fluid is flowing through the meter, a small propeller icon is highlighted.

    Flowrate Feature: To use this feature, press and release DISPLAY button until FLOWRATE icon appears. The factory set time base will be highlighted to the right of FLOWRATE (M = minutes, H = hours, D = days). When FLOWRATE is invoked, the display will be indicating rate of flow.

    CALIBRATION

    FACTORY CALIBRATION

    All Q09 Displays

    All calibration units are visible to the user as icons on the top line of the display, above the numeric digits.

    All units are configured with a “factory” calibration. Both gallons and litres are available (“GL” or “LT” will be displayed). While holding the CALIBRATE button, briefly press DISPLAY to toggle between gallons and litres. This factory calibration (indicated with FAC) is permanently programmed into the display and is not user adjustable.

    NOTE: Your display may have other units of measure programmed into it. If so, holding the CALIBRATE button and momentarily pressing the DISPLAY button will toggle through all factory set units. Other possible units are: IGL (imperial gallon), QT (quart), CF (cubic feet), CM (cubic meter), BL (42 gal. barrel), CC (cubic centimeter) or OZ (ounce).

    Switching between different units will not corrupt the Total’s contents. For example, in GL mode, the display totalizes 10.00 gallons, if the user switches to LT mode, the display will

  • 5

    read 37.85 litres (the same volume, different unit). The “field” calibration may be set by the user, and can be changed or modified at any time using the calibration procedure described in the Field Calibration Section. Totals or flowrate derived from the field calibration are invoked when the FAC icon is no longer visible on the top line of the display.

    FIELD CALIBRATION

    If the meter is suspected of being incorrect, field calibration is a method of adjusting the meter K-factor to compensate for error causing factors. The method of field calibration is dependent on the electronics that is mated with your Q09 display.

    NOTE: Q09 displays not mated to QSI electronics will use the manual entry Correction Factor Method outlined below. The manual entry method of field calibration outlined below is disabled when the Q09 display is mated with QSI electronics. The QSI electronics becomes the controlling element that manages the manual calibration entries for the display. This field calibration method is available in the QSI Android Application noted in the QSI owner’s manual.

    Q09 Displays Not Mated to QSI Electronics

    A manual entry method using the buttons on the Q09 display is used. This method adjusts the K-factor by percentages to match the display volume with actual volume.

    Calibration Using Display (Correction Factor Method) Field Calibration and Factory Calibration were outlined previously in the Calibration Section.

    Factory calibration settings are programmed into each display during manufacturing, using water at 70° F (21° C). Readings using the Factory Calibration (FAC) may not be accurate in some situations. For example: Under extreme temperature conditions, non-standard plumbing configurations or with fluids other than water.

    1. To field calibrate, press and hold the CALIBRATE and DISPLAY buttons for about 3 seconds until you see FLdCAL. Release both buttons and you will see CF - 00.0. You are now in the field calibration mode and values from -99.9% to +99.9% can be entered.

    2. The +/– position appears either as an “underscore” character for positive, or as a “hyphen” character for negative. The DISPLAY button selects the position and the CALIBRATE button toggles this character.

    3. The DISPLAY button can then be pushed to select the numeric positions. Press the CALIBRATE button to scroll from 0 to 9. Enter the percentage of change you want the display to correct. When satisfied with the value, press both CALIBRATE and DISPLAY buttons simultaneously. CALEnd will be displayed and unit will go back to normal operation, less the FAC (factory calibration) icon.

    4. All enabled units-of-measure remain visible and selectable – the entered correction will be applied to all enabled units.

    5. To return to factory calibration (FAC), press and hold both CALIBRATE and DISPLAY buttons for about 3 seconds until FAcCAL is displayed. Then release buttons. The unit should return to normal operation and FAC icon is visible.

  • 6

    Android Application noted in the QSI owner’s manual.

    Access to the configuration process is restricted for security until a “password” is entered. Contact your distributor or GPI to get the password and instructions to unlock and reset configuration settings. This information is also available at www.FLOMEC.net. Configurations are entered and stored as six-digit “codes” where each digit represents a setting for one of the configuration options. New configuration settings are stored in the display’s long-term memory and will not be lost either in OFF mode or during electrical failure.

    K-FACTOR ENTRY FIELD CALIBRATION

    Presently all displays are programmed with three different field calibration methods, only one of which is active, the “correction factor” calibration procedure described above. It is possible to activate “K-Factor entry” or “dispense/display” field calibration by changing configuration settings. Contact your distributor or GPI to get the correct password, configuration code and instructions for this calibration method. This information is also available on the GPI Web site.

    MAINTENANCE

    The display is virtually maintenance-free. However, it is important to keep the display clean and free of contaminants.

    USER CONFIGURATION

    The Q09 display has been programmed with many features, most of which can be enabled by the end user by way of a configuration process. By disabling “unnecessary” features, day-to-day flowmeter operation can be greatly simplified, making the unit easier to use. There are several features that are disabled by default when shipping standard meters. (For example, K-Factor Entry Field Calibration, described below.) For more advanced users, it may be desirable to enable ALL possible features. User configurable features include:

    • Totalizers/Modes Enabled (Cumulative Total, Batch 2 Total, Flowrate Mode).

    • Flowrate Timebase (Minute, Hour, Day).

    • Factory Calibration Curve Units Enabled – Gallons, Litres, Imperial Gallons, Quarts, Ounces, Cubic Feet, Cubic Centimeters, Cubic Meters or Barrels (42 Gal).

    • Dispense/Display or K-Factor Entry Calibration.

    CHANGING CONFIGURATION SETTINGS

    Q09 Displays Not Mated to QSI Electronics

    NOTE: Q09 displays not mated to QSI electronics can use the manual entry method outlined below. The manual entry method of changing configuration settings outlined below is disabled when the Q09 display is mated with QSI electronics. The QSI electronics becomes the controlling element that manages the manual configuration entries. This configuration setting method is available in the QSI

  • 7

    SPECIFICATIONS

    STANDARD FEATURES

    • 2 Totalizing Registers

    • 1 Factory Calibration Curve

    • 1 Field Calibration Curve

    • Rate of Flow Feature

    • Flowrate Time Base in Minutes

    INPUT PULSE RATE

    Minimum Pulse In: DC

    Minimum Coil Input:

    10 Hz

    Maximum Raw: 1,000 Hz

    K-FACTOR

    Minimum: .01 pulses/unit

    Maximum:999,999 pulses/unit

    FIELD CALIBRATION CORRECTION

    Minimum: -99.9%

    Maximum: +99.9%

    READOUT TOTALS

    Min. Display: 0.01

    Max. Display: 999,999 (x100)

    TEMPERATURES

    Operational:+0° to +140° F (-18° to +60° C)

    Storage:-40° to +158° F (-40° to +70° C)

    Ambient Air Operation Temperature

    0°F to +140°F (-18°C to +60°C)

    INTERNAL POWER SUPPLY TO DISPLAY

    Voltage Requirement:

    5 VDC provided by a 10 pin bridge connector from cover plate electronics. External power to meter is required to provide necessary internal power to display.

    ELECTRICAL

    Pollution Degree

    2

    Installation Category

    1

    Altitude 2000m Max.

    Indoor use only

  • 8

    SERVICE

    For warranty consideration, contact your local distributor. If you need further assistance, contact the GPI Customer Service Department at:

    1-888-996-3837

    To obtain prompt, efficient service, always be prepared with the following information:

    1. The model number of your display electronics.

    2. The serial number and manufacturing date code of the FLOMEC® meter the display is attached to.

    3. Specific information about part numbers and descriptions.

    For warranty work always be prepared with your original sales slip or other evidence of purchase date.

    RETURNING PARTS

    Please contact the factory before returning any parts. It may be possible to diagnose the trouble and identify needed parts in a telephone call. GPI can also inform you of any special handling requirements you will need to follow covering the transportation and handling of equipment which has been used to transfer hazardous or flammable liquids.

    CAUTION: Do not return equipment without specific authority from the GPI Customer Service Department. Due to strict regulations governing transportation, handling, and disposal of hazardous or flammable liquids, GPI will not accept equipment for rework unless it is completely free of liquid residue..

    The Waste Electrical and ElectronicEquipment (WEEE) direct ive(2002/96/EC) was approved by theEuropean Parliament and the Councilof the European Union in 2003. Thissymbol indicates that this productcontains electrical and electronicequipment that may include bat-teries, printed circuit boards, liquid

    crystal displays or other components that may be subjectto local disposal regulations at your location. Pleaseunderstand those regulations and dispose of this productin a responsible manner.

    RoHS Compliant (2011/65/EU)

    This product is in compliance with the RoHS Directive of the European Parliament and of the Council on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.

  • 9

  • 10

  • 11

  • 920896-03 Rev B03/2018

    © 2018 Great Plains Industries, Inc., All Rights Reserved.Great Plains Industries, Inc. / 888-996-3837 / FLOMEC.net

    LIMITED WARRANTY

    Great Plains Industries, Inc. 5252 E. 36th Street North, Wichita, KS USA 67220-3205, hereby provides a limited warranty against defects in material and workmanship on all products manufactured by Great Plains Industries, Inc. This product includes a 1 year warranty.

    Manufacturer’s sole obligation under the foregoing warranties will be limited to either, at Manufacturer’s option, replacing or repairing defective Goods (subject to limitations hereinaf-ter provided) or refunding the purchase price for such Goods theretofore paid by the Buyer, and Buyer’s exclusive remedy for breach of any such warranties will be enforcement of such obligations of Manufacturer. The warranty shall extend to the purchaser of this product and

    to any person to whom such product is transferred during the warranty period.

    The warranty period shall begin on the date of manufacture or on the date of purchase with an original sales receipt. This warranty shall not apply if:

    A. the product has been altered or modified outside the warrantor’s duly appointed representative;

    B. the product has been subjected to neglect, misuse, abuse or damage or has been in-stalled or operated other than in accordance with the manufacturer’s operating instructions.

    To make a claim against this warranty, contact the GPI Customer Service Department at

    316-686-7361 or 888-996-3837. Or by mail at:Great Plains Industries, Inc.

    5252 E. 36th St. NorthWichita, KS, USA 67220-3205

    The company shall, notify the customer to either send the product, transportation prepaid, to the company at its office in Wichita, Kansas, or to a duly authorized service center. The

    company shall perform all obligations imposed on it by the terms of this warranty within 60 days of receipt of the defective product.

    GREAT PLAINS INDUSTRIES, INC., EXCLUDES LIABILITY UNDER THIS WARRANTY FOR DIRECT, INDIRECT, INCIDENTAL AND CONSEQUENTIAL DAMAGES INCURRED IN THE

    USE OR LOSS OF USE OF THE PRODUCT WARRANTED HEREUNDER.The company herewith expressly disclaims any warranty of merchantability or fitness for any

    particular purpose other than for which it was designed.This warranty gives you specific rights and you may also have

    other rights which vary from U.S. state to U.S. state.Note: In compliance with MAGNUSON MOSS CONSUMER WARRANTY ACT – Part 702

    (governs the resale availability of the warranty terms).

  • 920896-02 Rev D04/2018

    QSB (Q-Star) Electronics for QSE Electromagnetic Meter

    Owner’s Manual

    Plain Cover Plate andPulse-Out Transmitter (QSB)

    Plain Cover Plate QSB Shown on 1 inch NPT Meter

    Display Mount Cover Plate (to mount a Q09 Display)

    and Pulse-Out Transmitter (QSB)

    Display Mount Cover Plate QSB (with Q09 Display mounted)Shown on 1 inch NPT Meter

    TABLE OF CONTENTSINTRODUCTION ..................................... 2

    IMPORTANT NOTICE ....................... 2GENERAL INFORMATION ................ 2SAFETY ............................................ 2

    INSTALLATION ....................................... 3

    WIRING ............................................ 4ISOLATED POWER SUPPLY WIRING ............ 4ISOLATED PULSE OUTPUT WIRING ............. 5

    MAINTENANCE ...................................... 6

    SPECIFICATIONS ................................... 6

    SERVICE ................................................. 7

    LIMITED WARRANTY ............................. 8

  • 2

    INTRODUCTION

    The QSE meter has multiple types of output electronics available. The electronics for the operation of the meter coils and flow tube is housed within the meter body casing. The cover plate is designed in two versions; a plain cover plate and a display mount cover plate. The QSB pulse-out electronics can be housed within either of the two cover plates. A display (Q09) is also available mounted to the “display mount” cover plate.

    All meters are equipped with pulse-out electronics (QSB) as the default standard regardless of style of cover plate. This manual contains information and meter wiring diagrams for the QSB pulse-out electronics

    IMPORTANT NOTICE

    QSB electronics are very sensitive to electric noise if operated within 6 inches of some electric motors, relays, transformers or other sources of electronic noise.

    If the QSB electronics are used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.

    GENERAL INFORMATION

    As a conductive liquid passes through the magnetic field created by the meter, a voltage is induced into the liquid that is directly proportional to the velocity of the flow. Electrodes in the meter flow tube transfer this induced voltage to the meter electronics.

    The QSB electronics convert the induced voltage into usable information. The pulse-out information is one function of the electronics; another function supplies the display (when installed) with the information and power it needs to operate as designed.

    The QSB electronics is a signal conditioner with industry standard pulse output. It can transmit those pulses to customer equipment up to 5,000 feet away. It is powered by customer-supplied external power.

    SAFETY

    • This product is not approved for use in hazardous locations.

    • Be sure O-rings and seals are kept in good repair.

    • When applying power, adhere to specifications in this manual.

    • Disconnect external power before attaching or detaching input or output wires.

  • 3

    INSTALLATION

    CAUTION: Installation near high electromagnetic fields and high current fields is not recommended and may result in inaccurate readings.

    If you ordered your QSB electronics with a meter, it is installed at the factory. If you ordered your QSB electronics separately from a meter, follow the instructions below. In every case, please review and thoroughly understand all manuals and installation instructions before proceeding.

    QSE meters are designed to measure flow in only one direction. To mount your QSB electronics (with or without a display):

    • Install the cover plate seal.

    • Connect the QSB module PC board connector to the AFE module PC board connector inside the meter housing with the ribbon cable. The ribbon cable connectors are polarized and cannot be incorrectly connected (see Figures 1 & 2).

    • Install the cover plate with screw holes aligned.

    • Install the (6) cover plate screws and tighten. Make sure the cover plate seal is fully seated before tightening the screws.

    A ribbon cable connects the QSB electronics within the cover plate to the AFE electronics within the meter body and allows 180 degrees of movement in either direction (see Figure 1).

    COVER PLATESEALAFE PC BOARDCONNECTOR

    COVER PLATE(DISPLAY MOUNTSHOWN)

    RIBBON CABLE

    RIBBON CABLECONNECTOR(EACH END)

    Figure 1

  • 4

    WIRING

    All electronic options are associated with a matching style of meter cover plate. This cover plate has four threaded ports, compatible with PG7 threads, for gaining wiring access to the electronics inside the cover plate. The meter is shipped with the ports environmentally sealed with a threaded plug and seal. Remove one or more of these plugs as required to install the supplied port fittings below.

    Each meter is supplied with cable gland strain reliefs with O-rings and 1/2 NPT adapters with seals for use in the threaded ports of the meter cover plate. Select the port fittings that fit your process and replace the threaded plugs in the cover plate with the fittings as required. The threaded plugs installed at the factory may be left in any unused cover plate port indefinitely.

    It is recommended that a “removable” thread-locker (fluid, stick, tape, spray, etc.) be used when installing the strain reliefs or adapters into the cover plate ports.

    Strain Reliefs: The cable gland strain reliefs will accommodate a cable diameter of 0.11 - 0.26 inches (2.79 - 6.6mm) and provide an environmental seal around the cable when the dome-nut is tightened.

    • NPT Adapters: The 1/2 NPT adapter fittings are used for attaching flex conduit to the meter, for those applications that require cable runs to be enclosed in conduit.Cable to be provided by customer to accommodate job requirements. Cable is not included with meter.

    NOTE: Some connectors installed on the PC board are not used for this QSB application. Do not

    RIBBON CABLECONNECTOR

    COVER PLATE(DISPLAY MOUNTSHOWN)

    PC BOARDCONNECTOR(RIBBON CABLE)

    Figure 2

  • 5

    QSB PC BOARDWIRING DIAGRAM

    QSB

    RIBBON CABLE TO AFE BOARD CONNECTOR

    ISOLATED PWR SUPPLY12-36VDC / 12-30VAC, 50-60 Hz

    12VDC OR 12VAC, 50-60 Hz(POLARITY DOESNT MATTER)

    SIGNAL

    GROUND

    +

    (+) (-)

    Power (-)

    Power (+)

    Pulse Out (+)

    Pulse Out (-)

    10-Pin Ribbon Cable

    -

    +

    -OPEN COLLECTOR OUTPUTWILL REQUIRE A PULL-UP RESISTOR TO SUPPLY VOLTAGE

    Input: AFE BoardOutput: Pulse Output

    ISOLATEDPWR SUPPLY12-30VAC, 50-60Hz / 12-36VDC

    Customer EquipmentOpen CollectorPulse Input

    GPI SENSOR

    Figure 3

    attempt to use these connections for other envisioned unauthorized purposes or circuit damage to the electronics will occur.

    Isolated Power Supply Wiring

    The QSE meter has an internal connection to the conductive fluid through one of the electrodes. The fluid and/or plumbing in most applications should be earth grounded to prevent any static buildup and provide a proper grounding for the system. Therefore, the power supply connection to the QSB should have an earth ground isolated output to prevent ground loop currents. Failure to use an isolated power supply could result in ground faults and could cause accuracy issues caused by stray voltages due to the ground loop currents. Proper grounding is a requirement.

    The wire gauge for the power supply cable should be between 16 to 26 AWG to properly fit in the wire termination connector. The external power supply used for supplying power to the QSB should be rated for a minimum of 0.25 Amps.

    The power supply polarity is not a requirement for either the DC or AC supply types as long as the supply is properly earth ground isolated.

    Isolated Pulse Output Wiring

    The pulse output from the QSB provides 2500V galvanic isolation from the QSB digital circuitry to the external customer equipment. The QSB pulse output is an open collector design which requires an external pull up resistor to an external power supply. The resistance must be chosen such that the QSB will not sink more than 30mA of current. The external power supply voltage must also be less than +36VDC to prevent damage to the QSB.

  • 6

    SPECIFICATIONS

    MECHANICAL

    Cover Plate Port Threads Female 1/2-20 UNF-2B (Compatible with PG7 thread)

    Port Strain Relief Hubble PG7

    Grip Range 0.11”-0.26” (2.79 - 6.6mm)

    Operation Temperature +32°F to +140°F (0°C to +60°C)Ambient Air Operation Temp 0°F to +140°F (-18°C to +60°C)

    Port Conduit Adapter GPI PG7 x 1/2 in. NPT Male

    Operation Temperature +32°F to +140°F (0°C to +60°C)Ambient Air Operation Temp 0°F to +140°F (-18°C to +60°C)

    POWER SUPPLY

    Voltage RequirementMin. 12 VDC or 12 VAC, 50-60 Hz

    Max. 36 VDC or 30 VAC, 50-60 Hz (higher voltage may damage unit)

    MAX POWER CONSUMPTION

    QSE Meter with QSB75mA 2 watts @ 24VDC 3 watts @ 36VDC

    ELECTRICAL

    Pollution Degree 2

    Installation Category 1

    Altitude 2000m Max.

    Indoor use only

    QSB PC BOARDWIRING DIAGRAM

    QSB

    RIBBON CABLE TO AFE BOARD CONNECTOR

    ISOLATED PWR SUPPLY12-36VDC / 12-30VAC, 50-60 Hz

    12VDC OR 12VAC, 50-60 Hz(POLARITY DOESNT MATTER)

    SIGNAL

    GROUND

    +

    (+) (-)

    Power (-)

    Power (+)

    Pulse Out (+)

    Pulse Out (-)

    10-Pin Ribbon Cable

    -

    +

    -OPEN COLLECTOR OUTPUTWILL REQUIRE A PULL-UP RESISTOR TO SUPPLY VOLTAGE

    Input: AFE BoardOutput: Pulse Output

    ISOLATEDPWR SUPPLY12-30VAC, 50-60Hz / 12-36VDC

    Customer EquipmentOpen CollectorPulse Input

    GPI SENSOR

    Figure 4

    The wire gauge for the pulse output cable should be between 16 to 26 AWG to properly fit in the wire termination connector.

    The pulse output ground should be properly connected to the customer equipment ground, while the pulse output signal should be properly connected to both the customer equipment pulse input and an external pull up resistor to an external supply (if the customer equipment doesn’t contain the pull up resistor).

  • 7

    The Waste Electrical and ElectronicEquipment (WEEE) direct ive(2002/96/EC) was approved by theEuropean Parliament and the Councilof the European Union in 2003. Thissymbol indicates that this productcontains electrical and electronicequipment that may include bat-teries, printed circuit boards, liquid

    crystal displays or other components that may be subjectto local disposal regulations at your location. Pleaseunderstand those regulations and dispose of this productin a responsible manner.

    RoHS Compliant (2011/65/EU)

    This product is in compliance with the RoHS Directive of the European Parliament and of the Council on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.

    MAINTENANCE

    Check cable-entry seals periodically. Tighten and/or apply sealant if needed. This is especially important in environments containing heavy concentrations of dust, oil mist, or other residue.

    Check all wiring connections occasionally for oxidation or corrosion. Clean and re-seat if such conditions are noted.

    If necessary, check and re-seat any connections that may have been subjected to strain (during rework or construction, for example).

    SERVICE

    For warranty consideration, contact your local distributor. If you need further assistance, contact the GPI Customer Service Department at:

    1-888-996-3837

    You will need to:

    • Provide information from the decal on your meter.

    • Receive a Return Authorization number.

    • Flush any fluid from the meter before shipping to the factory.

    CAUTION: Do not return the meter without specific authority from the GPI Customer Service Department. Due to strict regulations governing transportation, handling and disposal of hazardous or flammable liquids, GPI will not accept meters for rework unless they are completely free of liquid residue.

  • 920896-02 Rev D04/2018

    © 2018 Great Plains Industries, Inc., All Rights Reserved.Great Plains Industries, Inc. / 888-996-3837 / FLOMEC.net

    LIMITED WARRANTY

    Great Plains Industries, Inc. 5252 E. 36th Street North, Wichita, KS USA 67220-3205, hereby provides a limited warranty against defects in material and workmanship on all products manufactured by Great Plains Industries, Inc. This product includes a 1 year warranty.

    Manufacturer’s sole obligation under the foregoing warranties will be limited to either, at Manufacturer’s option, replacing or repairing defective Goods (subject to limitations hereinaf-ter provided) or refunding the purchase price for such Goods theretofore paid by the Buyer, and Buyer’s exclusive remedy for breach of any such warranties will be enforcement of such obligations of Manufacturer. The warranty shall extend to the purchaser of this product and

    to any person to whom such product is transferred during the warranty period.

    The warranty period shall begin on the date of manufacture or on the date of purchase with an original sales receipt. This warranty shall not apply if:

    A. the product has been altered or modified outside the warrantor’s duly appointed representative;

    B. the product has been subjected to neglect, misuse, abuse or damage or has been in-stalled or operated other than in accordance with the manufacturer’s operating instructions.

    To make a claim against this warranty, contact the GPI Customer Service Department at

    316-686-7361 or 888-996-3837. Or by mail at:Great Plains Industries, Inc.

    5252 E. 36th St. NorthWichita, KS, USA 67220-3205

    The company shall, notify the customer to either send the product, transportation prepaid, to the company at its office in Wichita, Kansas, or to a duly authorized service center. The

    company shall perform all obligations imposed on it by the terms of this warranty within 60 days of receipt of the defective product.

    GREAT PLAINS INDUSTRIES, INC., EXCLUDES LIABILITY UNDER THIS WARRANTY FOR DIRECT, INDIRECT, INCIDENTAL AND CONSEQUENTIAL DAMAGES INCURRED IN THE

    USE OR LOSS OF USE OF THE PRODUCT WARRANTED HEREUNDER.The company herewith expressly disclaims any warranty of merchantability or fitness for any

    particular purpose other than for which it was designed.This warranty gives you specific rights and you may also have

    other rights which vary from U.S. state to U.S. state.Note: In compliance with MAGNUSON MOSS CONSUMER WARRANTY ACT – Part 702

    (governs the resale availability of the warranty terms).

  • 920896-05 Rev C10/2017

    QSI (Q-Star) Electronics

    Owner’s Manual

    QSI Module(Installs in Plain Cover Plate)

    QSI2 Shown

    QSI Module(Installs in Display Mount Cover Plate)

    QSI2 Shown

    TABLE OF CONTENTSINTRODUCTION ..................................... 2

    IMPORTANT NOTICE ....................... 2GENERAL INFORMATION ................ 2SAFETY ............................................ 2

    INSTALLATION ....................................... 3

    WIRING ................................................... 4

    TERMINAL CONNECTIONS CHART 5QSI1 WIRING ................................... 6QSI2 WIRING ................................... 7QSI3 WIRING ................................... 8

    FLOMEC APP FOR ANDROID ............. 10

    TEMPERATURE PROBES .................... 10

    INSTALLATION ............................... 10WIRING .......................................... 11

    MAINTENANCE .................................... 12

    REPLACING BATTERY ................... 12DISCONNECT POWER TO METER ...................................... 12WITHOUT DISPLAY ........................ 12WITH DISPLAY ............................... 12

    SPECIFICATIONS ................................. 15

    SERVICE ............................................... 15

    RETURNING PARTS ....................... 15

    LIMITED WARRANTY ........................... 16

  • 2

    INTRODUCTION

    The QSI communications module is a multiple capability electronics package. The module is available with the plain cover plate or the display mount cover plate. The QSI module has three versions (QSI1, QSI2 or QSI3) available; each with its own mix of capabilities to better serve the customer (See Versions below).

    • QSI1 is equipped with: Bluetooth®, **coil/digital pulse input, pulse output (flow or energy & scalable), RS485 (Modbus® RTU), temperature inputs, BTU calculator.

    • QSI2 is equipped with: Bluetooth, **coil/digital pulse input, pulse output (flow or energy & scalable), data logger, temperature inputs, BTU calculator.

    • QSI3 is equipped with: Bluetooth, **coil/digital pulse input, pulse output (flow scalable), data logger, 4-20mA output.

    **NOTE The coil/digital input connections are for use with equipment other than a QSE meter. QSE meters automatically provide coil/digital pulse input via the 10-pin ribbon cable from the AFE PC board.

    This manual contains information and meter wiring diagrams for the three different QSI electronics modules.

    IMPORTANT NOTICE

    QSI module electronics are very sensitive to electric noise if operated within 6 inches of some electric motors, relays, transformers or other sources of electronic noise.

    GENERAL INFORMATION

    QSI electronics are available mounted to the meter you ordered or as a kit that adapts the electronics to other GPI meters. The QSI electronics converts the voltage (or pulses) from the meter into usable information. In addition to the multiple capabilities of the module, it also supplies the display (when installed) with the information and power it needs to function as designed.

    The QSI module electronics is powered by customer-supplied external power.

    SAFETY

    • This product is not approved for use in hazardous locations.

    • Be sure O-rings and seals are kept in good repair.

    • When applying power, adhere to specifications in this manual.

    • Disconnect external power before attaching or detaching input or output wires.

  • 3

    COVER PLATESEALAFE PC BOARDCONNECTOR

    COVER PLATE(DISPLAY MOUNTSHOWN)

    RIBBON CABLE

    RIBBON CABLECONNECTOR(EACH END)

    Figure 1

    INSTALLATION

    CAUTION: Installation near high electromagnetic fields and high current fields is not recommended and may result in inaccurate readings.

    If you ordered your QSI module electronics with a QSE meter, it is installed at the factory. If you ordered your QSI module electronics separately (i.e. repair part), follow the instructions below to install on a QSE meter. If you ordered your QSI module with one of our several types of meter adapter kits, follow the separate instructions enclosed with the kit to install the unit. In every case, please review and thoroughly understand all manuals and installation instructions before proceeding.

    To mount your QSI module (with or without a display) to a QSE meter:

    • Install the cover plate seal.

    • Connect the QSI module PC board connector to the AFE module PC board connector inside the meter housing with the ribbon cable. The ribbon cable connectors are polarized and cannot be incorrectly connected (see Figure 1).

    • Install the cover plate with screw holes aligned.

    • Install the (6) cover plate screws and tighten. Make sure the cover plate seal is fully seated before tightening the screws.

    A ribbon cable connects the QSI electronics within the cover plate to the AFE electronics within the meter body and allows 180 degrees of movement in either direction (see Figure 2).

  • 4

    WIRING

    All electronic options are associated with a matching style of meter cover plate. This cover plate has four 1/2-20 UNF threaded ports, compatible with PG7 threads, for wiring access to the electronics inside the cover plate. The meter is shipped with the ports environmentally sealed with PG7 threaded strain reliefs fitted with mushroom plugs. The mushroom plugs are inserted into the cable glands of the strain reliefs to maintain the seal until the port is used for wiring. Strain reliefs with mushroom plugs can be

    left in any unused port indefinitely. Remove the mushroom plugs to run cable into the cover plates.

    The strain reliefs will accommodate a cable diameter of 0.11 - 0.26 inches (2.79 - 6.6mm) and provides an environmental seal around the cable when tightened.

    • Cable to be provided by customer to accommodate job requirements. Cable is not included with meter

    Figure 2

    RIBBON CABLECONNECTOR

    COVER PLATE(DISPLAY MOUNTSHOWN)

    QSI PC BOARDCONNECTOR(RIBBON CABLE)

  • 5

    TEMP SENSOR

    PT100-Red Sensor VCC Output

    PT100-White Input High Side

    PT100-Blue Input Return

    RS-485 (Galvanically Isolated)

    RS485 (+) Positive Signal

    RS485 (-) Negative Signal

    RS485 GND Circuit Ground*

    PULSE INPUT

    +5V5VDC output to power external sensor

    (+)Digital Pulse Frequency Input, up to 3kHz

    (-) Digital Pulse Return

    4-20mA (Galvanically Isolated)

    (+)4 to 20mA current loop - current in to transmitter

    (-)4 to 20mA current loop - current out of transmitter*

    Coil ALow level sine wave input

    Coil BLow level sine wave input

    GPI SENSOR

    Ribbon cable connection to GPI custom sensor designs

    PULSE OUTPUT (Galvanically Isolated)

    (-) Pulse Output return current*

    (+)

    Pulse Output frequency output. This is a “current sinking open collector” output.***

    POWER INPUT

    (-) 12-36V AC or DC Input A**

    (+)12-36V AC or DC Input B**

    *NOTE: It is not recommended to connect isolated ground connections together or to board common.

    **NOTE: Polarity of voltage for AC/DC is not important. However an earth ground isolated power supply is required to prevent earth ground loop currents and ground fault conditions.

    ***NOTE: This output may require an external pull up resistor if interfacing equipment does not include one.

    PT100-Red

    Coil A

    GPI Sensor

    Coil B

    -+-+

    PT100-White Sensor #1PT100-BluePT100-RedPT100-WhitePT100-BlueRS-485 +RS-485 -RS-485 GND

    +5v+-

    +-

    Sensor #2

    RS-485Pulse Input

    4-20mA

    Pulse Output

    Power

    Figure 3

    TERMINAL CONNECTIONS CHART

  • 6

    QSI1 WIRING

    QSI VERSION 1 PC BOARD

    WIRING DIAGRAM

    QSI Version 1

    RS-485BTU

    Pulse Out

    10-PinRibbon ConnectorRTDs

    RS-485

    Power In

    Inputs: Temp Sensors, GPI Custom SensorOutputs: RS-485, Pulse Output

    10-Pin Ribbon Cable

    GPI SENSOR

    (+) (-)

    Power (-)

    Power (+)

    ISOLATEDPOWER SUPPLY12-36V AC/DC

    Pulse Out (+)

    Pulse Out (-)

    Current SourcingPulse Input

    RS485 (-)RS485 (GND)

    *R2...=120Ω

    R2 NOTE: The QSI does not contain an internal termination resistor. Refer to Customer Equipment documentation for proper installation and resistance value of the external R2 resistor.

    *Shield

    Customer Equipment

    RS485 (+)RS485 (-)RS485 (GND)

    RS485 (+)PT100 (White)PT100 (Blue)

    PT100 (Red)

    PT100 (White)PT100 (Blue)

    PT100 (Red)

    R2

    Figure 4

    QSI VERSION 1 PC BOARD

    WIRING DIAGRAM

    QSI Version 1

    RS-485BTU

    Pulse Out

    10-PinRibbon ConnectorRTDs

    RS-485

    Power In

    Inputs: Temp Sensors, GPI Custom SensorOutputs: RS-485, Pulse Output

    10-Pin Ribbon Cable

    GPI SENSOR

    (+) (-)

    Power (-)

    Power (+)

    ISOLATEDPOWER SUPPLY12-36V AC/DC

    Pulse Out (+)

    Pulse Out (-)

    Current SourcingPulse Input

    RS485 (-)RS485 (GND)

    *R2...=120Ω

    R2 NOTE: The QSI does not contain an internal termination resistor. Refer to Customer Equipment documentation for proper installation and resistance value of the external R2 resistor.

    *Shield

    Customer Equipment

    RS485 (+)RS485 (-)RS485 (GND)

    RS485 (+)PT100 (Whi