(ut) dgs diagram
TRANSCRIPT
DGS-DIAGRAM: ( Michael Berke )
Present day evaluation methods produce reliable and reproducible results providing that the
testing device meets technical requirements and that the test personnel are accordingly
qualified. The position of a flaw in the workpiece is calculated quickly and very accurately on
the basis of the measured acoustic time of flight of an echo. This means that a flaw location
takes place. The echo amplitude is used for an estimation of the flaw size. However, this is
not quite as easy as flaw location because the echo amplitude is subjected to much more
influences than the acoustic time of flight. Two methods have become generally accepted in
manual ultrasonic testing worldwide:
- Reference block method ( Distance Amplitude Blocks )
- DGS method
Though the two methods differ very much with regard to their application, they are not
different with regard to the physical principles of sound propagation and sound reflection
that they are based on. The reason is that, in both methods, the inspector determines the
size (diameter) of a reference reflector (disk shaped reflector, cylindrical reflector). The size
thus determined is not identical with the actual flaw size and it is therefore termed as the
equivalent circular disk or side drilled hole diameter. The shorter term "equivalent reflector
size" (ERS) has become generally accepted when using disk shaped reflectors. The reason for
the fact that the actual flaw size does not correspond to the equivalent reflector size is
because the sound fractions reflected from a natural flaw are additionally influenced by the
shape, orientation and surface quality of the flaw. In this respect further tests are difficult
and not very practical in manual ultrasonic testing so that most specifications and guidelines
for ultrasonic testing attach the criteria for flaw recording to a defined equivalent reflector
size. This means: the inspector determines whether a detected flaw reaches or exceeds the
equivalent reflector size indicated as a limit value (recording level) in the standard
specifications. In addition to this, the inspector must carry out other tests, e.g. regarding the
recording length, echo dynamics, etc., further details of which are not discussed here.
DGS Diagram
General DGS
Diagram
The regularities of sound
propagation in material have been
theoretically known for a long time
and were confirmed in practice by
numerous experiments. The
development of modern evaluation
methods shows two ways.
Procedures
To understand this better, let us start by explaining the sequences for both evaluation
methods at this point.
The Reference Block Method
This method requires that a reference block, corresponding to the test object and containing
one or more reference reflectors, be available for the test. The distance dependence of echo
amplitudes is determined experimentally by means of drilled holes in the reference block,
the resulting curve is then transmitted to the screen display of the test instrument (DAC
Distance Amplitude Correction). This curve automatically includes all probe (sound field) and
material effects. The test object can now be scanned with the probe. An indication recording
is made when an echo reaches the DAC curve or exceeds it.
The DGS Method
The DGS method is that the corresponding DGS diagram be available for the probe used in
the test application. The reference gain of the test instrument, with which the reference
echo is at a fixed screen height (reference level), is determined for a specific reflector, i.e.
the reference reflector. After this, the instrument gain is increased by a certain value, i.e. the
test sensitivity is adjusted. If the reference reflector is a circular arc from one of the
standardized calibration blocks, then the instrument gain should be varied in accordance
with the correction value given for the angle beam probe: the amplitude correction value is
adjusted. With different surface qualities between the test object and the calibration block
the transfer correction must be determined experimentally and likewise taken into
consideration. The gain difference with regard to the reference echo is determined for the
maximum echo from a detected indication. This is followed by a graphic determination of
the equivalent reflector size using the DGS diagram. If required, the sound attenuation
correction is additionally carried out. This makes it possible to assess whether the indication
is to be recorded or not. Nevertheless, by using the DGS scale it is possible to significantly
simplify evaluation with the DGS method .
- In this connection, the inspector uses an attachment scale for the screen of the
ultrasonic instrument. This scale contains one or several ready made recording
curves. The tiresome graphic evaluation with the DGS diagram can thus be omitted.
The inspector can directly assess flaw indications by means of the curve.
With the reference block method the characteristic curve of the sound field is always determined
before carrying out an ultrasonic test, whereas in the DGS method DGS diagrams for probes are
applied for this. A DGS diagram shows the echo amplitudes of disk shaped reflectors with different
diameters and those of large, flat reflectors (back-wall) as a function of the distance.
- A comparison of the test sequences for the reference block method and DGS method
shows the advantages and disadvantages in next table.
Advantages and Disadvantages of the DGS- and reference block method
Reference Block Method DGS-Method
Advantages
Disadvantages
Electronic DGS Evaluation
The use of microprocessor controlled ultrasonic instruments considerably simplifies both
evaluation methods, resulting in saving of time and higher test reliability. The DGS
evaluation now becomes particularly easy in an ultrasonic instrument by an optional
evaluation program. There are DGS diagrams for several standard probes stored in the
instrument. However, other probes can also be programmed on the basis of their
parameters and stored in one of the data sets. A flat bottom hole (disk shaped reflector),
side drilled hole or back-wall can be selected as reference reflectors. Owing to the
operational concept, the use of the DGS method is especially easy and reliable, operating
errors by the inspector are largely excluded due to the display of warning messages on the
screen. After the input of all parameters necessary for the flaw evaluation, the
corresponding recording curve is electronically displayed on the instrument screen
The evaluation program ensures a direct evaluation of a detected indication. All the
necessary corrections are taken into consideration in this respect: exceeding of the
recording threshold, i.e. the dB value by which the flaw indication exceeds the preset
recording curve, is directly displayed on the screen. This type of evaluation meets the
practical requirements specified in most of the testing guidelines. For example, these do not
only include e.g. the widely known international standards, but also all other specifications
requiring flat bottom holes as reference reflectors.
The DAC curve contains all
test related influences, i.e.
no time consuming
corrections are required
Easy and reliable
evaluation. No reference
blocks required.
Fabrication or
procurement of a suitable
reference block.
Recording of a DAC curve
for every test application
Measurement and
consideration of different
individual corrections.
Graphical determination
of equivalent reflector
size.