Per Fabrizio – 2/26/14
Make an exposure time read on the display of the remote control at the end of exposure : must be 17 sec. + 1.5 sec.
If the motor run slowly ( THE FINAL PART WITHOUT RADIATION ) you have to increase the motor power voltage .
Using a multimeter check the12v DC motor power voltage on the power board A1 : test point TP3 and TP4 .
On the board A1 with the trimmer R 39 speed 1- R40 speed 2 adjustment the motor power voltage until you reach 15 seconds ( speed 1 > 3V speed 2 )
If after the situation not be restored check the motor type 404603