Per Fabrizio – 2/26/14 Make an exposure time read on the display of the remote control at the end of exposure : must be 17 sec. + 1.5 sec. If the motor run slowly ( THE FINAL PART WITHOUT RADIATION ) you have to increase the motor power voltage . Using a multimeter check the12v DC motor power voltage on the power board A1 : test point TP3 and TP4 . On the board A1 with the trimmer R 39 speed 1- R40 speed 2 adjustment the motor power voltage until you reach 15 seconds ( speed 1 > 3V speed 2 ) If after the situation not be restored check the motor type 404603