potential of low dose cone beam computed...

33
Potential of low dose Cone Beam Computed Tomography and Multislice Computed Tomography for zygomaticomaxillary fracture diagnosis A blinded and randomized clinical approach Dose Reduction does not decrease the diagnostic acceptability of Cone Beam Computed Tomography and Multislice Computed Tomography for the diagnosis of zygomaticomaxillary fractures in a blinded and randomized approach.

Upload: others

Post on 06-Jun-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

Potential of low dose Cone Beam Computed Tomography and Multislice Computed

Tomography for zygomaticomaxillary fracture diagnosis – A blinded and randomized clinical

approach

Dose Reduction does not decrease the diagnostic acceptability of Cone Beam Computed

Tomography and Multislice Computed Tomography for the diagnosis of zygomaticomaxillary

fractures in a blinded and randomized approach.

Page 2: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

Potential of low dose Cone Beam Computed Tomography and Multislice

Computed Tomography for zygomaticomaxillary fracture diagnosis

A blinded and randomized approach

Master Thesis

R. (Romke) Rozema, BSc

S2808323

Groningen, April 12, 2016

Medicine - Master’s Degree Program

Faculty of Medical Sciences, University of Groningen, Groningen, The Netherlands

Supervisor - B (Baucke) van Minnen, MD PhD - Consultant Oral and Maxillofacial Surgeon

Department of Oral and Maxillofacial Surgery, University Medical Center Groningen,

University of Groningen, Groningen, The Netherlands

Page 3: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

R. Rozema, BSc1, R.N. Hartman, BSc1, M.H.J. Doff, DMD, PhD1, P.M.A. van Ooijen, MSc,

PhD, CPHIT2,3, H.E. Westerlaan, MD, PhD2,3, M.F. Boomsma, MD.4, B. van Minnen, MD.,

PhD1

Group member contribution

R. Rozema Lead organizer, study design creator, observer correspondent

R.N. Hartman Study design creator

M.H.J. Doff Statistical and data analysis support

P.M.A. van Ooijen Technical PACS and DICOM viewer support

H.E. Westerlaan Clinical support, finalizing image assessment, recruiting observers

M.F. Boomsma Clinical support, recruiting observers

B. van Minnen Study supervisor, study design creator, recruiting observers

1. Department of Oral and Maxillofacial Surgery, University Medical Center Groningen, University of

Groningen, Groningen, the Netherlands

2. Center for Medical Imaging - North East Netherlands (CMINEN), University Medical Center Groningen,

University of Groningen, Groningen, the Netherlands

3. Department of Radiology, University Medical Center Groningen, University of Groningen, Groningen, the

Netherlands

4. Department of Radiology, Isala Clinics, Zwolle, the Netherlands

Page 4: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

PREFACE

This master thesis originates from a special interest in medical imaging. A radiology

department is constantly subjected to fast changing technology that provides new

opportunities to improve the clinical care. As for diagnostic devices, questions should always

be raised regarding the legitimate and correct use. This is especially the case for devices using

ionizing radiation. The focus of this study is the utilization of diagnostic devices in the field of

oral and maxillofacial surgery. One recent change is the increased use of Cone Beam

Computed Tomography. This shares similarities with the Multislice Computed Tomography,

which has also been well established in the radiology and emergency department. Both hold

advantages and drawbacks.

I have been educated in medical radiology and worked for 3 years as a diagnostic

radiographer. During this period I was concerned with the technical aspects and practical

decision making using diagnostic modalities. With this background I got in touch with

craniomaxillofacial trauma surgery, where Cone Beam Computed Tomography is especially

appreciated. However, it occurred to me that whenever I got concerned with a

craniomaxillofacial trauma patient at the emergency department, Multislice Computed

Tomography was the preferred modality. In view of these considerations I got interested in

the features and drawbacks of these imaging modalities.

I learned that Cone Beam Computed Tomography is appreciated because of the general lower

radiation dose. However, Multislice Computed Tomography is widely available in the

emergency department and the geometry enables to study any part of the human body. The

most significant drawback for both modalities are the radiation induced risks. This study

derived from the proposition that radiation dose should be reduced as low as reasonable

achievable. In this unique collaboration between the Department of Oral and Maxillofacial

Surgery and the Center of Medical Imaging North Eastern Netherlands of the University

Medical Center Groningen we were able to conduct the present study in a multidisciplinary

approach. The result is an experimental human cadaver study where the effect of radiation

dose reduction is studied for the diagnosis of midface fractures.

Romke Rozema

April 12, 2016

Page 5: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

SAMENVATTING

DOEL: Het onderzoeken van de diagnostische waarde van dosis reduceerde Cone Beam

Computed Tomography (CBCT) en Multislice Computed Tomography (MSCT) voor de

diagnostiek van zygomaticomaxillaire fracturen.

MATERIAAL EN METHODE: Unilaterale zygomaticomaxillaire fracturen werden

aangebracht op vier van zes vers gevroren menselijke kadaver preparaten. Alle preparaten

werden gescand met twee CBCT en vier MSCT protocollen met systematisch gereduceerde

stralingsdosis. De vervaardigde beelden werden gerandomiseerd en geblindeerd beoordeeld

door 16 radiologen en 8 kaakchirurgen. De resultaten werden vergeleken met een gouden

standaard waarbij de aanwezigheid van een fractuur werd geverifieerd met een chirurgische

benadering van het zygomaticomaxillaire complex.

RESULTATEN: Bij gemiddeld 90,3 procent (n=130) werd een correcte diagnose van een

zygomaticomaxillaire fractuur gesteld. Bij 7,6 procent werd een incorrecte diagnose gesteld.

Van de verschillende anatomische locaties werd de arcus zygomaticus gemiddeld het vaakst

correct gediagnosticeerd (91,0%). Betrokkenheid van de crista zygomaticus alveolaris werd

gemiddeld het minst vaak correct gediagnosticeerd (65,3%). Dosis reductie had geen invloed

op de mogelijkheid tot het beoordelen van de dislocatie, comminutie, orbitale inhoud, volume

rendering en de weke delen. Kaakchirurgen ervaarden de dosis gereduceerde protocollen

afdoende voor het plannen van de behandeling. De schatting van de effectieve dosis reikte van

122 µSv tot 28 µSv voor CBCT en 129.9 ± 9.2 µSv tot 51.0 ± 4.3 µSv voor MSCT.

CONCLUSIE: Dosis reductie lijkt geen duidelijke invloed te hebben op de diagnostische

waarde van CBCT en MSCT voor het diagnosticeren van zygomaticomaxillaire fracturen.

Page 6: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

ABSTRACT

PURPOSE: To assess the diagnostic value of low dose Cone Beam Computed Tomography

(CBCT) and Multislice Computed Tomography (MSCT) for zygomaticomaxillary fracture

diagnosis.

MATERIAL AND METHODS: Unilateral zygomaticomaxillary fractures were inflicted on

four out of six fresh frozen human cadaver head specimen. All specimen were scanned using

two CBCT and four MSCT protocols where the radiation exposure was systematically

reduced. A blinded diagnostic routine was recreated where 16 radiologists and 8 oral and

maxillofacial surgeons performed 144 randomized image assessments. As a gold standard, the

presence of fractures of the zygomatic region was verified by an open operative approach.

RESULTS: Zygomaticomaxillary fractures were correctly diagnosed in 90.3% (n=130) of the

image assessments. The zygomatic arch was most often correctly diagnosed (91.0%). The

zygomatic alveolar crest showed the lowest degree of correct diagnosis (65.3%). No

significant decrease of correctly diagnosed fracture sites was found between the baseline and

low dose CBCT and MSCT protocols. Dose reduction did not significant decrease the ability

to assess dislocation, comminution, orbital volume, volume rendering and soft tissues. OMF

surgeons considered the low dose protocols sufficient for treatment planning. The effective

dose of MSCT (129.9 to 51.0 µSv) remained well in range of CBCT (122 to 28 µSv).

CONCLUSION: Dose reduction seems not to decrease the diagnostic value of CBCT and

MSCT protocols for the diagnosis of zygomaticomaxillary fractures.

Page 7: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

1

TABLE OF CONTENTS

INTRODUCTION ...................................................................................................................... 2

MATERIAL AND METHODS ................................................................................................. 3

Research design ...................................................................................................................... 3

Research subjects .................................................................................................................... 3

Experimental design ............................................................................................................... 3

Image acquisition .................................................................................................................... 4

Gold standard .......................................................................................................................... 6

Image assessment ................................................................................................................... 6

Objective and subjective analysis ........................................................................................... 7

Statistical analysis ................................................................................................................... 8

RESULTS ................................................................................................................................... 9

Effective dose estimations ...................................................................................................... 9

Objective assessment results ................................................................................................ 10

Subjective assessment results ............................................................................................... 12

Effect of dose reduction ........................................................................................................ 12

DISCUSSION .......................................................................................................................... 15

CONCLUSION ........................................................................................................................ 20

ACKNOWLEDGEMENTS ..................................................................................................... 21

REFERENCES ......................................................................................................................... 22

ATTACHMENTS .................................................................................................................... 26

Image assessment form example for radiologists ................................................................ 26

Image assessment form example for OMF surgeons ........................................................... 27

Page 8: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

2

INTRODUCTION

Cone-Beam Computed Tomography (CBCT) and Multislice Computed Tomography (MSCT)

are the imaging modalities of choice in midface trauma diagnostics and treatment planning

[1–4]. Both modalities are utilized to detect and characterize midfacial fractures [2, 5, 6]. The

zygomaticomaxillary buttress represents the most common site of facial fractures [7, 8].

Disruption of the zygomatic position can cause an aesthetic and functional deficit. It is

therefore mandatory that zygomatic bone injury is properly diagnosed and adequately

managed [9].

Both CBCT and MSCT offer a three dimensional dataset with sub-millimeter resolution

[2, 3]. CBCT utilizes a cone shaped x-ray beam where only one rotation is needed for data

acquisition. Current systems require acquisition either sitting or standing in a natural head

position. MSCT datasets, however, are acquired with a collimated fan beam over a multitude

of 360 degree scans. The patient is transported through the gantry in a recumbent position in

synchrony with continuous data acquisition. Therefore, MSCT offers the advantage of

multipurpose utilization that allows studying of any part of the human body.

Although CBCT is considered to produce a generally lower radiation dose, there are some

clinical considerations for the commonly use of MSCT to evaluate blunt facial trauma [1, 4,

10]. In the first place, there is limited availability of CBCT within the emergency department,

where some of the patients present initially. Second, some conditions force the patient in a

recumbent bedridden position. Acquisition in natural head position as done with CBCT is

therefore not possible. Third, one fifth of the maxillofacial traumas have concomitant injuries,

especially cerebral and cervical spine injuries [7]. Since MSCT is needed for these

indications, the head is scanned in one go from a necessary and practical point of view [11].

However, there is an increasing concern regarding the potential radiation induced risks due to

the exponentially increased use of MSCT and CBCT in medicine [1, 12]. Keeping the

radiation dose as low as reasonably achievable, remains the most important strategy for

decreasing this potential risk [13]. There is limited evidence regarding the effect of dose

reduction for craniomaxillofacial trauma. In one study, evidence for the potential of

substantial dose reduction for the diagnosis of midface fractures was successfully provided

[14]. However, no previous study conducted a randomized analysis where the observer was

blinded for the outcome of interest. The aim of this study was to assess the diagnostic value of

low dose CBCT and low dose MSCT for zygomaticomaxillary fractures using this approach.

Page 9: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

3

MATERIAL AND METHODS

Research design

A prospective experimental human cadaver study was designed to recreate a diagnostic

approach. Unilateral zygomaticomaxillary fractures were inflicted on four out of six obtained

human cadaver heads. All specimens were scanned using CBCT and MSCT. Dose reduction

was achieved by repeated acquisition with adjusted tube current parameters as an independent

variable. A group of 16 radiologist and 8 oral and maxillofacial (OMF) surgeons were asked

to assess fracture visibility and diagnostic value in a blinded objective image analysis. Six

unique CBCT and MSCT protocols were evaluated for all six human cadavers.

Research subjects

Six fresh frozen human cadaver heads were obtained and approved by the section anatomy of

the Department of Neurosciences of the University Medical Centre Groningen, Groningen,

the Netherlands. These cadavers have been donated for medical research and educational

purposes according to the legal and ethical framework of the Dutch uniform anatomical gift

act. Their age ranged from 60 to 96 years old. Thawing the cadaver specimens to room

temperature occurred 48 hours prior to experimentation. The specimens were isolated at the

craniovertebral junction keeping the intracranial contents intact.

Experimental design

An experiment was designed to systemically inflict unilateral zygomaticomaxillary fractures

on four out of six specimen. Simulating a blunt facial trauma recreates the injury mechanism

found as a major cause in midface trauma etiology [7]. Therefore, a guided fall method was

used as an impact model [15, 16]. This allows for a better control of motion of the impactor

and has the possibility of a local deformation of the midface [15]. The specimen were placed

horizontally on a solid wedge for stabilisation. The direction of impact was chosen as 40

degree angled from the mid sagittal section plane in order to obtain perpendicular impact on

the malar eminence. A cylindrical 2.0 kg impactor with flat impacting surface was used as a

free falling mass. A drop height estimation of 70 cm was calculated using biomechanical

fracture tolerance data where power of collision was measured in dynamic loads [16–18]. An

independent consultant OMF surgeon physically examined the zygomaticomaxillary complex

after impact for fracture presence. The drop height was raised with 10 cm consequently if no

fracture was found.

Page 10: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

4

Figure 1 : The specimens were placed on a solid wedge for stabilisation. Zygomaticomaxillary fractures were

inflicted by a perpendicular impact of the malar eminence using a 2.0 kg impactor.

Image acquisition

Each specimen was scanned using two CBCT and four MSCT protocols. The field of view

(FOV) was standardized from the maxillary dentition up to the superior orbit rim of the

frontal sinus for both CBCT and MSCT. Scan parameter details are listed in table 1 and 2.

CBCT imaging was performed using the ProMax 3D Mid (Planmeca Oy, Helsinki, Finland).

The specimens were positioned on a fixed support in a natural head position. Alignment lasers

were used for horizontal and vertical centration. Each specimen was scanned using two 0,4

mm isotropic acquisition protocols, where one protocol was scanned in ultra-low dose (ULD)

mode. Projection data was gathered using a fixed tube voltage of 90 kV and tube current of

8,0 mA and 5,0 mA respectively in a partial arc scan geometry of 200 degree rotation. This

data was reconstructed into a single volumetric dataset using a Feldkamp-Davis-Kress (FDK)

filtered back projection algorithm.

MSCT imaging was performed using a third generation Siemens SOMATOM Force scanner

(Siemens Healthcare AG, Erlangen, Germany). The specimens were positioned supine and

centered such that the external auditory meatus was at the centre of the gantry. Based on the

lateral pre-scan topogram, we corrected the specimen position so that the hard palate was

parallel to the axial acquisition plane. Automatic exposure control (AEC) was used for

radiation dose optimization adapting tube current (CARE Dose4D, Siemens) and tube voltage

(CARE kV, Siemens). CARE Dose4D is a combined x-y-z automatic exposure control system

where the topogram localizer is used to determine attenuation characteristics at different

scanning positions. CARE kV is an automatic tube voltage optimization algorithm to optimize

radiation dose without increasing image noise. It is a recommendation regarding the optimal

combination of tube voltage and tube current taking the patient size, body habitus and clinical

question into account.

Page 11: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

5

Quality reference mAs was used as the only variable to systemically reduce the radiation dose

for each MSCT acquisition protocol. We scanned each specimen with a quality reference mAs

of 50, 40, 30 and the lower limit of 20. The raw datasets were reconstructed into slice images

using two different types of image reconstruction algorithms also referred to as kernels. We

reconstructed a bone and soft tissue volume dataset using the Siemens Healthcare head

regular 59 (Hr59d) and head regular 32 (Hr32d) kernels. These kernels optimize high and low

contrast detectability respectively.

All CBCT and MSCT bone volumetric datasets were reconstructed into a 3D volume

rendering batch. This rendering technique conveys the axial acquisitioned two dimensional

data into a three dimensional volume of information providing the spatial relationship of the

osseous midface anatomy. The observer of this study was provided with a dynamic 360

degree rotational view round a fixed vertical axis. This volume rendering was batched using a

bone template. All images were exported in a Digital Imaging and Communications in

Medicine (DICOM) standard.

Table 1 : Planmeca Promax 3D mid CBCT acquisition parameters for

the standard and ULD protocol

Tube voltage 90 kV

Tube current 8 mA

Tube current ULD 5 mA

Exposure time 13,6 s

Exposure time ULD 4,5 s

Grayscale depth 12 bit

Scan Time 18-26 s

FOV 200 x 100

Radiation source Pulsed

Rotation degree Single 200 degree rotation

Projections per rotation 300

Detector Type FPD

Voxel size (x,y,z) 0,4 mm3

Matrix 512 x 512

CBCT Cone Beam Computed Tomography

ULD Ultra Low Dose

FOV Field of View

FPD Flat Panel Detector

Page 12: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

6

Gold standard

A gold standard was created in a consensus meeting between a head and neck radiologist and

an OMF surgeon with craniomaxillofacial trauma fellowship. Both were not included in the

image analysis procedure. The fracture site presence was recorded by the radiologist who had

access to all scanned MSCT and CBCT data. These findings were validated by the OMF

surgeon using a combined local transcutaneous and maxillary vestibular approach of the

specimen’s zygomaticomaxillary complex. Discrepancies were discussed and advocated.

Image assessment

The diagnostic quality of all six acquisition protocols was read in 114 image assessments by a

group of 24 observers. This group consisted of radiologists (n=16) and OMF surgeons (n=8)

from the University Medical Center Groningen (Groningen, NL), Isala clinics (Zwolle, NL),

Nij Smellinghe hospital (Drachten, NL), Tjongerschans hospital (Heereveen, NL) and

Sionsberg hospital (Dokkum, NL).

Table 2 : Siemens SOMATOM Force MSCT acquisition parameters

for the four different scan protocols. The quality reference mAs was

used to systemically reduce the radiation dose for four protocols.

Tube voltage CARE kV

Tube current CARE Dose4D

Quality reference mAs 50, 40, 30, 20

Average effective mAs 77, 52, 40, 29

ADMIRE Strength* 1

FOV 220.0 mm

Collimation 192 x 0,6 mm

Average scan length 118 mm

Slice thickness 0.6 mm

Position increment 0.4 mm

Grayscale depth 12 bit

Pitch 0.6

Rotation time 0.5 s

Exposure time 0.5 s

Scan time 3,4 s

Matrix 512 x 512

MSCT Multislice Computed Tomography

ADMIRE Advanced Modelled Iterative Reconstruction

FOV Field of View

Page 13: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

7

Simulating a diagnostic routine, each specimen was presented as a fictive patient case with a

suspected zygomaticomaxillary fracture. The observer was blinded for the outcome of interest

and all scan parameters. Each observer performed this image assessment for all six patient

cases. One of the six acquisition protocols was random allocated for each patient case using a

research randomizer (www.randomizer.org). The order in which the patient cases were

assessed was also randomized for each observer. The use of a medical diagnostic display was

set as a condition for radiologist. These displays all met the acceptance and constancy tests

following the America Association of Physicists in Medicine (AAPM) Task Group 18 (TG18)

guidelines [19].

The DICOM datasets were provided on a mobile TeraRecon Aquarius iNtuition DICOM

Viewer (TeraRecon, Foster City, California, United States). This viewer application allowed

the features and functionalities a medical professional would use in the daily clinical care.

MSCT patient cases were provided with two datasets; a bone and soft tissue volume. Each

CBCT patient case was provided with the single reconstruction dataset. The volume rendering

batch was available for both acquisition modalities. The observer was allowed to adjust

window width and window level from a 2000/250 bone and 400/40 soft tissue preset. The

DICOM viewer interface loaded the volume datasets into orthogonal axial, sagittal and

coronal plane sections where free angle adjustment of a multi planar reconstruction was

possible.

Objective and subjective analysis

The image assessment was divided between an objective and subjective analysis. Each patient

case was provided in advance with the relevant personal particulars, mechanism of injury,

clinical presentation and the physical examination results. The objective analysis was

performed using a predefined selection of zygomaticomaxillary fracture associated anatomical

sites. First, the observer had to determine the presence of a suspected zygomaticomaxillary

fracture (yes/no/inconclusive). Subsequently, the analysis was subdivided into the

involvement of the frontozygomatic suture, lateral orbital wall, zygomatic arch, orbital floor,

zygomatic alveolar crest, anterior maxillary sinus wall, posterior maxillary sinus wall and the

zygomatic corpus (yes/no/inconclusive). The gold standard was used to verify if the outcome

for each of the observers was correct.

Each observer also performed a subjective analysis where the diagnostic acceptability for

each allocated scan protocol was assessed as a dichotomous outcome variable (yes/no). The

diagnostic acceptability covered the assessment of fracture dislocation, fracture comminution,

Page 14: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

8

orbital volume, soft tissue and volume rendering assessment. In addition, each observer got a

profession related question. Radiologists assessed the sufficiency to diagnose and assess the

zygomaticomaxillary fracture and OMF surgeons the sufficiency to determine the treatment

plan and the potential surgical approach. The image assessment was standardized for each

patient case. Examples of these image assessment forms for both radiologists and OMF

surgeons are provided attachments.

Radiation dose assessment and comparison

Effective dose estimations were made for all CBCT and MSCT acquisition protocols.

Dosimetric comparison is a key component in balancing the clinical consideration for CBCT

and MSCT. The use of effective dose estimation allows stochastic effect quantification and

comparison between CBCT and MSCT. The effective dose estimation of the used CBCT

device was provided by a previous study where dose measurements were performed using

thermoluminescent dosimeters (TLD) in anthropomorphic Alderson Radiation Therapy

(ART) phantoms [20]. The MSCT dose estimations were calculated using volume weighted

CT dose index (CTDIvol) and Dose Length Products (DLP) provided by the scanner. These

represented the total absorbed radiation dose within the entire scan volume for each

acquisition protocol. Head conversion coefficients (mSv mGy-1 cm-1) quoted by the American

Association of Physicists in Medicine (AAPM) were used to convert DLP to effective dose

[21].

Statistical analysis

The Statistical Package for the Social Sciences (SPSS, version 20.0) was used for data

analysis. A chi-square test was performed to compare the proportions of correct diagnosed

fracture sites between the baseline CBCT and MSCT acquisition protocols. This comparison

was repeated with their corresponding low dose scan protocols for both CBCT and MSCT.

The proportions of correct diagnosed fracture sites were compared between radiologist and

OMF surgeons. The proportion of correct diagnosed fracture sites was compared between

radiologist and OMF surgeons using a chi-square test. The agreement within each profession

was calculated using Fleiss Kappa statistics. The significance level of α was set at 5%.

Page 15: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

9

RESULTS

The consensus reading confirmed that zygomaticomaxillary fractures were sustained on all

four specimens in varying degree of severity (figure 2). All specimens had fractures at

multiple buttress relationships of the zygomaticomaxillary complex. Involvement of the

zygomatic alveolar crest was seen in one case. A fracture of the zygomatic corpus itself was

not seen in any of the cases. In one case, a fracture of the lateral orbital wall was combined

with an intact frontozygomatic suture. Based on the consensus reading the currently used

clinical acquisition protocols appeared sufficient to visualize fractures at all selected

anatomical sites of the midface (figure 2).

Figure 2 : The specimen sustained fractures a multiple buttress relationships of the zygomaticomaxillary

complex. Volume rendering reconstructions (A+C) are compared with the surgical dissection of the

zygomaticomaxillary complex (A+D) in a case were a fracture of the lateral orbital wall was combined

with an intact frontozygomatic suture.

Effective dose estimations

Effective dose estimations are summarized in figure 3. The CBCT effective dose ranged from

122 µSv to 28 µSv, resulting in a 77% dose reduction when using the ULD protocol [20]. The

average measured MSCT effective dose ranged from 129.9 ± 9.2 µSv (mean ± SD) to the

lower limit of 51.0 ± 4.3 µSv (mean ± SD).

Page 16: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

10

A dose reduction of 61% for MSCT was achieved by reducing the quality reference mAs from

50 to 20. The dose of the baseline MSCT protocol remained well in range of CBCT. Dose

reduced MSCT protocols produced less effective dose than the baseline CBCT protocol. The

minimal dose was produced by the CBCT ULD protocol.

Objective assessment results

Over the total of 144 image assessments, 90.3% (n=130) of the zygomaticomaxillary fractures

were correctly diagnosed and 7.6% (n=11) were misdiagnosed. In another 2.1% (n=3) the

diagnosis was reported as inconclusive. All four specimens that had a fracture were diagnosed

correct by all 24 observers (100%). From the two specimens without a fracture, one was

diagnosed correctly by 23 observers (95.8%) and the last was diagnosed correct by 11

observers (45.8%).

The zygomatic arch fracture was most often diagnosed correctly (91.0%). In five of the six

specimens, the zygomatic arch fractures were diagnosed 100% correctly. It was noticeable

that whilst the remaining specimen did not have a fracture, 13 of the 24 observers suspected a

fracture of the zygomatic arch. In one specimen there was an exceptional low degree of

correct diagnosis of the orbital floor (8.3%), the anterior maxillary sinus wall (16.7%) and the

posterior maxillary sinus wall (8.3%). These fractures were initially missed in the consensus

reading but later confirmed by the surgical approach as minimal and non-dislocated (figure 5).

Figure 3 : Radiation Dose estimation of the two Cone Beam Computed Tomography and four

Multislice Computed Tomography (50-20) acquisition protocols.

0

25

50

75

100

125

150

MSCT - 50 MSCT - 40 MSCT - 30 MSCT - 20 CBCT CBCT ULD

Effe

ctiv

e D

ose

Sv)

Page 17: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

11

The case where a lateral orbital wall fracture was combined with an intact frontozygomatic

suture was correctly diagnosed in 87.5% of the assessments. On average, the zygomatic

alveolar crest fractures showed the lowest degree of correct diagnosis (65.3%).

Figure 4 : Comparison of the baseline and low dose CBCT and MSCT scan protocols consecutively.

This right sided zygomaticomaxillary fracture included (1) zygomatic arch dislocation, (2) continuity loss

of the posterior maxillary sinus wall and (3) non-dislocated small defect of the anterior maxillary sinus

wall.

Page 18: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

12

There was no significant difference between radiologists and OMF surgeons regarding the

proportions of correctly diagnosed fracture sites (X2=1.358, n=1152, p=0.507). The

agreement among the eight OMF surgeons was substantial, k=0.617 (95% CI, 0.594 to 0.639),

p=0.000. There was a moderate agreement among the sixteen radiologists, k=0.552 (95% CI,

0.505 to 0.599), p=0.000. Also, no significant difference was found between the baseline

CBCT and MSCT protocols (X2=1.805, n=384, p=0.406).

Subjective assessment results

On average, there was a high degree of conformity that the image quality was sufficient to

assess the fracture dislocation (95.5%), comminution (89.2%), the orbital volume (95.5%), the

provided 3D volume rendering (85.6%) and the soft tissues (91.8%). The soft tissue

reconstructions were only available on MSCT protocols. OMF surgeons considered the image

quality sufficient for treatment management in 95.7% of the image assessments on average.

There was no apparent decrease within the reducing the radiation dose. Radiologist

considered the image quality sufficient for fracture assessment in 83.3% of the image

assessment on overage. Although there was no decrease within the MSCT protocols, the

CBCT ULD protocol had a substantial reduced agreement (56.3%) when compared to the

standard CBCT protocol (82.4%). The observers mentioned that the ULD protocol contained

an unacceptable amount of beam hardening artefacts.

Effect of dose reduction

The image assessments results per scan protocol are summarized in table 3. No significant

decrease of correctly diagnosed fracture sites was found between the baseline and low dose

CBCT and MSCT protocols. No significant decrease was found for the ability to assess

Figure 5 : Initial missed non-dislocated fracture of the infraorbital rim (A), and the posterior

maxillary sinus wall (B). These were confirmed during the surgical approach of the

zygomaticomaxillary complex.

Page 19: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

13

Table 3 : Results of the image assessments for each acquisition protocol

Quality Reference mAs

MSCT protocol

CBCT protocol Sig.c

(p<0.05)

50 40 30 20 Reg. ULD

Objective analysisa

Fracture presence 87.5 95.8 87.5 91.7 87.5 91.7 NSd

Frontozygomatic suture 75.0 75.0 70.8 79.2 75.0 79.2 NSd,e

Lateral orbital wall 87.5 87.5 100 83.3 91.7 87.5 NSe

Zygomatic arch 95.8 91.7 87.5 91.7 87.5 91.7 NSe

Orbital floor 70.8 75.0 75.0 70.8 79.2 79.2 NSd

Zygomatic alveolar crest 62.5 62.5 66.7 75.0 62.5 62.5 NSd

Anterior maxillary sinus wall 79.2 79.2 79.2 79.2 87.5 87.5 NSe

Posterior maxillary sinus wall 70.8 79.2 66.7 79.2 70.8 79.2 NSd,e

Zygomatic corpus 75.0 83.3 79.2 70.8 62.5 87.5 NSd,e

Subjective analysisb

Dislocation 94.4 100 94.7 94.4 94.7 94.7 NSe

Comminution 83.3 100 73.7 94.4 89.5 94.7 NSe

Orbital Volume 88.9 100 89.5 100 94.7 100 NSe

Volume Rendering 100 94.4 78.9 94.4 68.4 78.9 NSe

Soft Tissue 94.4 94.4 84.2 94.4 - - NSe

Treatment management 85.7 100 100 100 100 100 NS

Fracture assessment 81.3 93.8 87.5 100 82.4 56.3 CBCT Sig.

a percentage correctly diagnosed fracture sites verified by the consensus reading.

b percentage sufficient considered image quality for the presented parameters.

c statistical difference between the baseline and low dose protocols for both CBCT and MSCT

d Pearson Chi-squared test

e Fisher’s exact test

CBCT Cone Beam Computed Tomography

MSCT Multislice Computed Tomography

ULD Ultra Low Dose

NS Not significant

Page 20: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

14

dislocation, comminution and the orbital volume between the baseline and low dose CBCT

and MSCT protocols. Moreover, the effect of dose reduction did not significantly decrease the

considered sufficiency of volume rendering and MSCT soft tissue reconstructions. For OMF

surgeons, dose reduction did not significantly decrease the sufficiency to determine the

treatment pathway of the patient. However, for radiologists, the sufficiency to assess the

fracture did significantly decrease between the baseline and ULD CBCT protocol. This

decrease was not significant for MSCT.

Page 21: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

15

DISCUSSION

As far as we know, this is the first study in which blinded randomized image assessments

were independently performed to assess the diagnostic value of low dose CBCT and MSCT

for the diagnosis of zygomaticomaxillary fractures. An increase of image noise affects the

interpretability of a suspected fracture. This study demonstrated that dose reduction did not

significantly decrease the objective fracture detectability of anatomical related sites of the

zygomaticomaxillary complex. These results are in agreement with a previous study where

low dose protocols were found sufficient for the diagnosis of both dislocated and non-

dislocated midface fractures [14].

Despite these findings, detectability percentages varied substantially among the assessed

anatomical sites. Zygomatic alveolar crest fractures were the least likely to be diagnosed

correctly (table 3). The zygomatic alveolar crest is located in the infrazygomatic region. In

anatomical terms this site is not clearly defined. It is clinically appreciated because the

cortical bone thickness provides anchory for the osteosynthesis screws during surgery [22].

We believe that the unclear anatomical demarcation of the crest resulted in a doubtful

determination of the involvement in the fracture. Also, a fracture of the zygomatic corpus was

found in a considerable part of the image assessments whilst none of the specimen sustained a

fracture in this region. A potential explanation is the position of the zygomaticofacial foramen

that transmits the zygomaticofacial nerve through the zygomatic corpus. The foramen could

lead to a deceptive conviction that a fracture line courses the zygomatic corpus. In one case,

the orbital floor, the anterior and posterior maxillary sinus wall had an exceptional low degree

of correct diagnosis. These fractures were minimal and non-dislocated (figure 5). These

findings suggest that the used scan protocols were not sufficient to visualize these fractures

properly. However, this was refuted during the consensus meeting. Although these finding

should not be nullified, misdiagnosis would have no clinical consequences as conservative

treatment is also preferred when diagnosing these fractures correctly. Nevertheless, the

detection variability among the anatomical sites do emphasize that careful reading of the

datasets and adequate anatomical proficiency is needed to properly assess a suspected midface

fracture.

CBCT and MSCT both hold potential for diagnosing midfacial fractures [2, 5, 6]. Although it

has been stated that the image quality of CBCT is better [23], more recent studies provide

evidence that it is comparable to MSCT even though some variability exists among different

Page 22: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

16

systems [10, 24]. The results of the present study also show that there was no significant

difference between the baseline CBCT and MSCT protocols for the proportion of correctly

diagnosed fracture sites. It is important to note that the way in which data acquisition is

performed varies between CBCT and MSCT. CBCT is dedicated to the oral and maxillofacial

region only and the patient is scanned in a natural head position [1]. When MSCT is used, the

patient is transported through the gantry in synchrony with continuous data acquisition. This

feature is especially appreciated when a craniomaxillofacial trauma patient is suspected of

concomitant injuries and forced in a recumbent position. 19.6 percent of the

craniomaxillofacial traumas are associated with cerebral and cervical spine injuries and for

these cases, the initial use of MSCT is justified from a necessary and patient safety

perspective.

Furthermore, MSCT is appreciated for the ability to detect associated soft tissue injury which,

in case of midface fractures, is utilized to assess the extraocular muscles [9, 25, 26]. In

addition, the ophthalmologic assessment may sometimes be difficult due to the severity of the

head injury, the extent of periorbital oedema, inadequate cooperation or altered consciousness

of the patient [9, 27]. However, dose reduction is not recommended for the evaluation of soft

tissues, as image noise and contrast resolution are closely related [13]. Therefore, the clinical

applicability of low dose protocols in this study could be convicted. However, in contrast to

what might be expected, the results of this study show no significant decrease for the ability to

assess soft tissues using low dose MSCT protocols. Since this was assessed subjectively,

more extensive research is needed to provide more conclusive evidence.

Radiation dose is a significant parameter within the clinical consideration of the used imaging

modality. Previous studies provided effective dose estimations for CBCT and MSCT. A meta-

analysis reported CBCT effective dose measurements ranging from 46 to 1073 µSv for large

FOVs, 9-560 µSv for medium FOVs and 5-652 µSv for small FOVs. Other studies reported

MSCT effective dose estimations of 474 to 1160 µSv [4], 534 to 860 µSv [28] and 0.1 to 3.6

mSv [14]. These results support the common believe that MSCT delivers a higher radiation

exposure than CBCT does [1, 4, 6]. However, our study provides evidence that the dose of the

used MSCT is well in range of CBCT. Moreover, the MSCT protocols of this study often

remain below the enumerated CBCT doses. Widmann et al. also successfully confirmed a

dose reduction below the levels of CBCT [29].

Page 23: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

17

Nevertheless, the remarkable wide range of these dose estimations confine the generalizability

[30]. As the delivered radiation dose is the result of interplay between different elements,

optimization should be performed using an appropriate selection of scan parameters.

Therefore we would like to address the following four aspects of parameters that should be

taken into careful consideration.

First, a topogram of the complete scan range is a requisite. A careful determination of the scan

range allows limitation of exposure to the area of interest only. Also, the AEC uses the

topogram to improve consistency of the image quality and this curtails the radiation dose for

each individual patient [13]. In this study this was performed adjusting both tube current

(CARE Dose4D) and tube voltage (CAREkV). For CBCT, a distinction is needed between

small-, medium-, and large FOVs [20]. A significant reduction in effective dose is associated

with the use of small FOV sizes [30].

Second, the performance difference among scanners is associated with the radiation dose

given. Modern CBCT and MSCT scanners feature better detector performance, improved

beam shape filtering or reduced amount of overranging that allow radiation exposure

optimization better than former scanners do [13, 30]. This should be taken into account when

medical modalities are compared.

Third, the diagnostic task and the intrinsic contrast of the area of interest should be decisive

for the used scan parameters. Some studies propose the use of a full head MSCT protocol for

maxillofacial applications [4, 14]. The MSCT head protocol requisites low contrast

detectability because of the close range of intracranial tissue densities. This is greatly

influenced by the image noise and is less apparent with a higher radiation exposure. However,

MSCT protocols for craniomaxillofacial trauma are predominantly used to visualize the

osseous anatomy. This requires a lower radiation exposure than a full head protocol because

the image noise is reduced when a broad window width is combined with a bone algorithm

reconstruction.

Fourth, the reconstruction algorithm type applied to reconstruct raw MSCT projection data

has a unique role in radiation dose adjustments and the image quality. Currently, filtered back

projection (FBP) is the well-established reconstruction type for clinical MSCT protocols. The

recent availability of iterative reconstruction (IR) algorithms have the advantage in accurately

modelling the system geometry, incorporating physical effects such as beam spectrum, noise,

beam hardening effect, scatter and incomplete data sampling [13]. IR also offers the potential

for radiation dose reduction [14, 31].

Page 24: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

18

In this study FBP was combined with a modelled-based IR (ADMIRE 1, Siemens) that

incorporates modelling, both in the raw projection data and in the image domains, such that a

different statistical weighting is applied according to the quality of the projection data [31].

With respect to constant image quality, this allowed the use of a reduced quality reference

mAs. Widmann at al. concluded that the use of IR could not improve the detection rate of

fractures due to the smoothing effects [14]. Despite the dose reduction potential of IR

algorithms, the clinical applicability of the computational time requirements should be taken

into consideration. This is especially in case of craniomaxillofacial trauma where a minimum

time to diagnosis is a prerequisite. In view of limited evidence, further research should focus

on the use of IR for craniomaxillofacial trauma.

When comparing radiologists and OMF surgeons, no significant difference was found for the

proportion of objective diagnosed fracture sites. The subjective image assessments show that,

according OMF surgeons, dose reduction did not significantly decrease the sufficiency for

treatment management. For radiologists, the subjective diagnostic sufficiency did not

significant decrease for dose reduced MSCT protocols. However, it is important to note that

this was significant for CBCT (table 3). The average agreement among OMF surgeons

(95.7%) was higher than radiologists (83.3%). These findings suggest that image

interpretation by OMF surgeons is performed from a different perspective than that of

radiologists. OMF surgeons predominantly focus on the determination of the treatment

pathway and the necessity of surgery. However, radiologists carry the responsibility for an

appropriate diagnosis and therefore assess the image quality from a more critical approach.

When the image quality is compared to MSCT, CBCT is more prone to additional artefacts,

such as aliasing, scatter and a general higher noise level [32]. This could explain why a

substantial share of the radiologists considered the CBCT ULD protocol insufficient (table 3).

Also, initial interpretation of CBCT datasets is usually not performed by a radiologist but by

the OMF surgeons itself. The OMF surgeon is therefore more proficient in CBCT imaging

interpretation. This is not the case for MSCT where the diagnostic findings of a radiologist are

used by the OMF surgeon to define the treatment for a patient.

This study has limitations. First, the post-mortem status of human cadaver specimen carry

process changes that affect the interpretability of the scans. The specimen suffer cerebral

atrophy and reduced attenuation between soft tissue structures [33]. Also, trauma patients may

have indirect predictive CT findings, such as sinus opacification and periorbital contusion,

that have discriminatory benefits for predicting midface fractures [34]. These findings are

Page 25: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

19

absent in human cadavers. Nevertheless, we believe that as a surrogate model it is the most

accurate approach of a craniomaxillofacial trauma patient and moreover this prevents the

unjustifiable high radiation dose a living patient would receive upon repeated study

acquisitions. Another limitation is the paucity of fracture variability. In this study, a total of

six cases are used as representation of a midface trauma patient population. However,

clinicians are faced with a broad range of fracture patterns. Moreover, a craniomaxillofacial

trauma patient could also suffer fractures other than to the zygomaticomaxillary buttress, such

as Le Fort or naso-orbitoethmoid fractures that are not assessed in the present study.

Therefore, it is important to be circumspect about the direct applicability of low dose

protocols in a clinical environment.

Page 26: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

20

CONCLUSION

In conclusion, in this human cadaver study we found that dose reduction did not decrease the

diagnostic value of CBCT and MSCT for the diagnosis of zygomaticomaxillary fractures.

Low dose protocols were also considered sufficient for treatment management according

OMF surgeons. This study emphasizes that CBCT and MSCT datasets of a suspected midface

fracture should be interpreted carefully and that anatomical proficiency is required for an

appropriate diagnosis. The results of this study show that the effective dose of MSCT is well

in range, or even lower than CBCT. In addition, a successful method was provided to inflict

zygomaticomaxillary fractures on human cadaver specimen in varying degree of severity.

Further research should focus on the potential of IR algorithms and the applicability of low

dose CBCT and MSCT protocols in a clinical environment.

Page 27: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

21

ACKNOWLEDGEMENTS

We would like to thank Ronald Dob (Department of Radiology) and Arjan Dieters

(Department of Orthodontics) from the University Medical Center Groningen for their advice

and technical assistance during the use of the MSCT and CBCT devices respectively. Also,

we would like to thank Klaas van Linschoten and dr. Janniko Georgiadis from the Section

Anatomy, Department of Neuroscience, (University of Groningen, Groningen, the

Netherlands) for the accompaniment and support in the provision and use of the human

cadaver specimens in this study. Finally a special thanks for Steve Oosterhoff for his

persevering assistance in scanning these specimens in the late night hours.

Page 28: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

22

REFERENCES

1. De Vos W, Casselman J, Swennen GRJ (2009) Cone-beam computerized tomography

(CBCT) imaging of the oral and maxillofacial region: A systematic review of the

literature. Int J Oral Maxillofac Surg 38:609–625

2. Sohns JM, Staab W, Sohns C, Schwarz A, Streit U, Seif Amir Hosseini A, Spiro JE,

Kertész A, Zwaka PA, Lotz J (2013) Current perspective of multidetector computed

tomography (MDCT) in patients after midface and craniofacial trauma. Clin Imaging

37:728–733

3. Miracle a C, Mukherji SK (2009) Conebeam CT of the head and neck, part 2: clinical

applications. AJNR Am J Neuroradiol 30:1285–1292

4. Loubele M, Bogaerts R, Van Dijck E, Pauwels R, Vanheusden S, Suetens P, Marchal

G, Sanderink G, Jacobs R (2009) Comparison between effective radiation dose of

CBCT and MSCT scanners for dentomaxillofacial applications. Eur J Radiol 71:461–

468

5. Choudhary AB, Motwani MB, Degwekar SS, Bhowate RR, Banode PJ, Yadav AO,

Panchbhai A (2011) Utility of digital volume tomography in maxillofacial trauma. J

Oral Maxillofac Surg 69:e135–e140

6. Shintaku WH, Venturin JS, Azevedo B, Noujeim M (2009) Applications of cone-beam

computed tomography in fractures of the maxillofacial complex. Dent Traumatol

25:358–366

7. Gassner R, Tuli T, Hächl O, Rudisch A, Ulmer H (2003) Cranio-maxillofacial trauma:

A 10 year review of 9543 cases with 21 067 injuries. J Cranio-Maxillofacial Surg

31:51–61

8. Kraft A, Abermann E, Stigler R, Zsifkovits C, Pedross F, Kloss F, Gassner R (2012)

Craniomaxillofacial trauma: synopsis of 14,654 cases with 35,129 injuries in 15 years.

Craniomaxillofac Trauma Reconstr 5:41–50

9. Lee E, Mohan K, Koshy J, Hollier L (2010) Optimizing the Surgical Management of

Zygomaticomaxillary Complex Fractures. Semin Plast Surg 24:389–397

Page 29: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

23

10. Liang X, Jacobs R, Hassan B, et al (2010) A comparative evaluation of cone beam

computed tomography (CBCT) and multi-slice CT (MSCT) Part I. On subjective image

quality. Eur J Radiol 75:265–269

11. Lee BH, Hwang YJ, Hur G, Kim SY, Lee JY, Kim YH (2011) Multidetector computed

tomography (MDCT) images with soft tissue and bone algorithm reconstruction in

head and facial trauma. J Clin Neurosci 18:899–901

12. Brenner DJ, Hall EJ (2007) Computed tomography--an increasing source of radiation

exposure. N Engl J Med 357:2277–2284

13. Yu L, Liu X, Leng S, Kofler JM, Ramirez-Giraldo JC, Qu M, Christner J, Fletcher JG,

McCollough CH (2009) Radiation dose reduction in computed tomography: techniques

and future perspective. Imaging Med 1:65–84

14. Widmann G, Dalla Torre D, Hoermann R, Schullian P, Gassner EM, Bale R, Puelacher

W (2015) Ultralow-dose computed tomography imaging for surgery of midfacial and

orbital fractures using ASIR and MBIR. Int J Oral Maxillofac Surg 44:441–446

15. Verschueren P, Delye H, Depreitere B, Van Lierde C, Haex B, Berckmans D, Verpoest

I, Goffin J, Vander Sloten J, Van der Perre G (2007) A new test set-up for skull

fracture characterisation. J Biomech 40:3389–3396

16. Hampson D (1995) Facial injury: A review of biomechanical studies and test

procedures for facial injury assessment. J Biomech 28:1–7

17. Yoganandan N, Pintar F a, Sances a, Walsh PR, Ewing CL, Thomas DJ, Snyder RG

(1995) Biomechanics of skull fracture. J Neurotrauma 12:659–668

18. Yoganandan N, Sances A, Pintar F a, Maiman DJ, Hemmy DC, Larson SJ, Haughton

VM (1991) Traumatic Facial Injuries with Steering Wheel Loading. J Trauma 31:699–

710

19. Samei E, Badano A, Chakraborty D, Compton K, Cornelius C (2005) Assessment of

Display Performance for Medical Imaging Systems. Med Phys 4095:155

20. Pauwels R, Beinsberger J, Collaert B, et al (2012) Effective dose range for dental cone

beam computed tomography scanners. Eur J Radiol 81:267–271

Page 30: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

24

21. McCollough C, Cody D, Edyvean S, Geise R (2008) The measurement, reporting, and

management of radiation dose in CT. Rep AAPM Task Gr. doi: ISBN: 978-1-888340-

73-0 ISSN: 0271-7344

22. Liou EJW, Chen PH, Wang YC, Lin JCY (2007) A computed tomographic image

study on the thickness of the infrazygomatic crest of the maxilla and its clinical

implications for miniscrew insertion. Am J Orthod Dentofac Orthop 131:352–356

23. Hashimoto K, Kawashima S, Kameoka S, Akiyama Y, Honjoya T, Ejima K, Sawada K

(2007) Comparison of image validity between cone beam computed tomography for

dental use and multidetector row helical computed tomography. Dentomaxillofacial

Radiol 36:465–471

24. Watanabe H, Honda E, Tetsumura A, Kurabayashi T (2011) A comparative study for

spatial resolution and subjective image characteristics of a multi-slice CT and a cone-

beam CT for dental use. Eur J Radiol 77:397–402

25. Winegar B, Murillo H, Tantiwongkosi B (2013) Spectrum of Critical Imaging Findings

in Complex Facial Skeletal Trauma. Radiographics 33:3–19

26. Guly CM, Guly HR, Bouamra O, Gray RH, Lecky FE (2006) Ocular injuries in

patients with major trauma. Emerg Med J 23:915–917

27. Lee H, Jilani M, Frohman L, Baker S (2004) CT of orbital trauma. Emerg Radiol

10:168–72

28. Ludlow JB, Ivanovic M (2008) Comparative dosimetry of dental CBCT devices and

64-slice CT for oral and maxillofacial radiology. Oral Surgery, Oral Med Oral Pathol

Oral Radiol Endodontology 106:930–938

29. Widmann G, Fasser M, Schullian P, Zangerl A, Puelacher W, Kral F, Riechelmann H,

Jaschke W, Bale R (2012) Substantial dose reduction in modern multi-slice spiral

computed tomography (MSCT)-guided craniofacial and skull base surgery. RoFo

Fortschritte auf dem Gebiet der Rontgenstrahlen und der Bildgeb Verfahren 184:136–

142

Page 31: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

25

30. Ludlow JB, Timothy R, Walker C, Hunter R, Benavides E, Samuelson DB, Scheske

MJ (2015) Effective dose of dental CBCT-a meta analysis of published data and

additional data for nine CBCT units. Dentomaxillofac Radiol 44:20140197

31. Solomon J, Mileto A, Ramirez Giraldo JC, Samei E (2015) Diagnostic Performance of

an Advanced Modeled Iterative Reconstruction Algorithm for Multidetector CT

Scanner : Potential for Radiation Dose Reduction in a Multireader Study 1. Radiology

275:735–745

32. Schulze R, Heil U, Groß D, Bruellmann DD, Dranischnikow E, Schwanecke U,

Schoemer E (2011) Artefacts in CBCT: A review. Dentomaxillofacial Radiol 40:265–

273

33. Chew FS, Relyea-Chew A, Ochoa ER (2006) Postmortem Computed Tomography of

Cadavers Embalmed for Use in Teaching Gross Anatomy. J Comput Assist Tomogr

30:949–954

34. Smith HL, Chrischilles E, Janus TJ, Sidwell RA, Ramirez M, Peek-Asa C, Sahr SM

(2013) Clinical indicators of midface fracture in patients with trauma. Dent Traumatol

29:313–318

Page 32: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

26

ATTACHMENTS

Image assessment form example for radiologists

Case history; 48 years old male bumped into a lamppost while walking the dog. Consulted the

general practitioner with right sided facial pain and periorbital oedema the next day.

Table 1*

Yes No Inconclusive

Is this patient diagnosed with a fracture?

* Table 2 and 3 only need to be completed when answered YES

Table 2 - Classify the fracture whether the following anatomical sites are involved

Yes No Inconclusive

Frontozygomatic suture / Lateral orbital rim

Lateral orbital wall

Zygomatic arch

Orbital floor

Zygomatic alveolar crest

Anterior maxillary sinus wall

Posterior maxillary sinus wall

Corpus zygomaticus (i.e. the body itself)

Table 3 - Is the image quality sufficient to assess the following parameters?

Yes No

Dislocation of the fracture

Comminution of the fracture

Orbital Volume

3D (volume rendering)

Soft Tissue (see soft tissue series for CT)

Table 4

Yes No

Is the image quality of this acquisition protocol sufficient to

diagnose and assess midfacial fractures according your

professional standard?

Page 33: Potential of low dose Cone Beam Computed …scripties.umcg.eldoc.ub.rug.nl/FILES/root/geneeskunde/...fracture tolerance data where power of collision was measured in dynamic loads

27

Image assessment form example for OMF surgeons

Case history; 48 years old male bumped into a lamppost while walking the dog. Consulted the

general practitioner with right sided facial pain and periorbital oedema the next day.

Table 1*

Yes No Inconclusive

Is this patient diagnosed with a fracture?

* Table 2 and 3 only need to be completed when answered YES

Table 2 - Classify the fracture whether the following anatomical sites are involved

Yes No Inconclusive

Frontozygomatic suture / Lateral orbital rim

Lateral orbital wall

Zygomatic arch

Orbital floor

Zygomatic alveolar crest

Anterior maxillary sinus wall

Posterior maxillary sinus wall

Corpus zygomaticus (i.e. the body itself)

Table 3 - Is the image quality sufficient to assess the following parameters?

Yes No

Dislocation of the fracture

Comminution of the fracture

Orbital Volume

3D (volume rendering)

Soft Tissue (see soft tissue series for CT)

Table 4

Yes No

Is the image quality of this acquisition protocol sufficient for

medical treatment planning and potential surgical management

according your professional opinion?