seog oh/ duke university/oct, 2002 trt/ atlas week at cern test of the cleaning gas mixtures
DESCRIPTION
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN Test module 2TRANSCRIPT
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
Test of the cleaning gas mixtures
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
Purpose Find out the safe operating limit of the glass wire-joint in mixtures
with CF4 in case silicon cleaning is necessary. Run parameters
Upper chamber 14 glass wire-joints – 5 were pulled out on July 25. The rest was pulled
out on Sept 27.– June/July/August : Ar-CO2-CF4 (70-14-16) : 16% mixture– September : Ar-CO2(70-30)
Lower chamber 11 glass wire-joints – 4 were pulled out on July 25. The rest was pulled
out on Sept 27– Ar-CO2-CF4 (70-22- 8) : 8% mixture
Moisture level : 700 ppm on average Gain
June/July/August - ~30,000-40,000 September ~ 70,000-80,000
Current draw Average : 10 nA/cm – 30 nA/cm Around wire-joints : 30 nA-100nA/cm
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
Test module 2
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
Test setup
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
16% Results Failure (do not hold HV) occurred when the
integrated charge was 0.2-0.3 C/cm. 0.22, 0.19, 0.26, 0.21, 0.27, 0.34 C/cm
Tension loss proceeded the failure. All glass beads exhibited heavy etching reasonably
proportional to the integrated charge. Despite the etching, there was no evidence of silicon
deposit downstream (SEM & Gain measurement). This test ended at the end of August. At the time,
there were two surviving wire-joints (#30, #31). The gas mixture was changed to Ar-CO2 (70-30) to study the residual effect of CF4 (more later).
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
16% (C/cm .vs. T change)
-20
-15
-10
-5
0
5
0 0.05 0.1 0.15 0.2 0.25 0.3
C/cm
Tens
ion
loss
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
Failed wire-joints (~0.3C/cm)16%
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
16% Wire #25. 0.27C/cm. Broken (8/25)
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
Wire-joints in 16% Note that there is no Silicon peak. Si is replaced by F.
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
Wire surface scan (16%) Left : “normal”. Right : somewhat smooth
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
16% Left : “normal”. Right : somewhat smoother
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
Wire Downstream of a wire-joint. There is no silicon deposit
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
Wire #26 (0.27C/cm). Broken 8/18
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
8% Results One failure occurred when the integrated charge was
~0.45 C/cm. Tension loss happened as early as ~0.3 C/cm. Etching on the failed joint was also evident. However
compare to the failed ones in 16%, the etching was less.
There was also no evidence of silicon deposit downstream (SEM & Gain measurement).
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
8% (C/cm .vs. T change)
-25
-20
-15
-10
-5
0
5
0 0.1 0.2 0.3 0.4 0.5
C/cm
Tens
ion
loss
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
Failed wire-joint (~0.45C/cm)8% Wire #12
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
8% (#10) Wire #10. 0.31 C/cm. Lost 22 grams of tension
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
Wire-joint (8%) There is still some silicon peak
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
Wire #12 (0.45C/cm) Broken 9/14
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
Residual effect of CF4 The 16% mixture was changed to Ar-CO2 (70-30) on
September 3. The gas gain : ~70,000 (from ~30,000) Current Draw : ~17 nA/cm (=> ~50 nA around the
wire-joints) Charge accumulation
The integrated charge at the time of the switch : ~0.15 C/cm The integrated charge since the switch : ~0.1 C/cm
Results The degree of etching was consistent with ~0.15C/cm rather
than ~0.25 C/cm. No gain change was observed after the switch
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
0.15C/cm in 16%+0.1C in Ar-CO2
Wire #30, 31
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
16% -> 0% Wire #30 and #31
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
Wire-joint #30
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
16% -> 0% 4 cm downstream of wire-joint
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
16% -> 0% 20 cm downstream of wire-joint. The chunks are gold.
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
Wire #31
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
Summary Integrated charge
From the 16% and 8% test, it seems that the failure integrated charge is roughly proportional to the CF4 fraction when the moisture level of ~700 ppm.
16% : fail at 0.2-0.3 C/cm 8% : fail at 0.4-0.5 C/cm
Including the earlier measurement (CF4 ~20%, H2O ~ 2000-3000 ppm, integrated charge at the failure ~ 0.1 C/cm), it seems that the failure also roughly proportional to the H2O level.
There is no silicon deposit or gain change. There is no gross residual effect when the gas is
switched to non CF4 mixture.
Seog Oh/ Duke University/Oct, 2002 TRT/ ATLAS week at CERN
Proposal Test 4% mixture – (may not be necessary) Test the effectiveness of the cleaning mixtures
Produce silicon deposit with Ar-CO2 (70-30) <=> gain drop of ~10%
Bubble through silicon oil? Use contaminated straws?
Switch the gas to one of the cleaning mixtures. Determine the integrated charge necessary to restore the gain.
Parameters CF4 fraction Moisture level Gain Irradiation level