jenis-jenis pre-amp

Upload: fyndi-aw-aw

Post on 08-Jan-2016

42 views

Category:

Documents


1 download

DESCRIPTION

spectroscopy

TRANSCRIPT

  • An overview of a typical data acquisition setup

    COMMON COMPONENTS

    PREAMPLIFIERS

    General

    The fundamental output of all pulse type radiation detectors is a burst of charge Q liberated by the incidentradiation. For most detectors the charge is so small that it is impractical to deal with the signal pulses without an

    intermediate amplication step. The rst element in a signal-processing chain is therefore often a preamplier as

    an interface between the detector and the pulse-processing and analysis electronics that follow.

    The preamplier is usually located as close as possible to the detector. From a signal-to-noise standpoint, it

    is always preferable to minimize the capacitive loading on the detector, and therefore long interconnecting cables

    between the detector and preamplier should be avoided if possible. One function of the preamplier is to terminate

    the capacitance quickly and therefore to maximize the signal-to-noise ratio. It also serves as an impedance matcher,

    presenting a high impedance to the detector to minimize loading, while providing a low impedance output to drive

    succeeding components.

    The preamplier conventionally provides no pulse shaping, and its output is a linear tail pulse. The rise time

    of the output pulse is kept as short as possible, consistent with the charge collection time in the detector itself.

    The decay time of the pulse is made quite large (typically 50 or 100 s ) so that full collection of the charge fromdetectors with widely diering collection times can occur before signicant decay of the pulse sets in.

    Voltage- and charge- sensitive Congurations

    Figura 1: Schematic diagram of a simplied voltage-sensitive preamplier

    conguration. R2 is the feedback resistance.

    Figura 2: Simplied diagram of a charge-sensitive preamplier congu-

    ration. If the conditions indicated are met, the output pulse amplitude

    becomes independent of the input capacitance Ci. The time constant givenby the product RfCf determines the decay rate of the tail of the outputpulse.

    Preampliers can be of either the

    voltage-sensitive or charge-sensitive

    type. Historically, the voltage-

    sensitive type is the more conventio-

    nal in many electronic applications

    and consists simply of a conguration

    that provides an output pulse who-

    se amplitude is proportional to the

    amplitude of the voltage pulse sup-

    plied to its input terminals. A sche-

    matic diagram of a voltage-sensitive

    conguration is shown in Fig. 1. If

    the time constant of the input circuit

    (the parallel combination of the input

    capacitance and resistance) is large

    compared with the charge collection

    time, then the input pulse will have

    an amplitude equal to

    V =Q

    C(1)

    where C is the input capacitance. For

    most detectors, the input capacitance

    is xed so that the output pulse pro-

    duced by a voltage-sensitive pream-

    plier is proportional to the charge Q

    liberated by the incident radiation. If

    the input capacitance were to change

    however, this desirable proportionali-

    ty would no longer hold. In semicon-

    ductor diode detectors, for example,

    the detector capacitance may change

    with operating parameters. In these

    situations, a voltage-sensitive preamplier is undesirable because the proportionality between Vmax and Q is lost.The elements of a charge-sensitive conguration that can remedy this situation is shown in Fig. 2. For this

    circuit, the output voltage is proportional to the total integrated charge in the pulse provided to the input terminals,

  • as long as the duration of the input pulse is short compared with the time constant RfCf . Changes in the inputcapacitance no longer have an appreciable eect on the output voltage. Although originally developed for use with

    semiconductor diode detectors, this charge-sensitive conguration has proved its superiority in a number of other

    applications, so that preampliers used with other detectors in which the capacitance does not necessarily change

    are also often of the charge-sensitive design.

    Noise Characteristics

    Probably the most important specication for a preamplier is its noise gure. This specication is normally

    quoted as the FWHM of the response function of the system due only to the preamplier noise. The gure is

    normally given as the equivalent energy spread in the type of detector for which the preamplier is designed. The

    noise gure is a strong function of the capacitance with which the preamp input is loaded. For example good quality

    preamplier used with silicon diode detectors may have a noise gure of 2 keV with zero input capacitance, but this

    gure may double if the input is loaded with 100 pF. The input capacitance arises from both the inherent detector

    capacitance and from the connecting cable between the detector and preamplier. It is therefore important to keep

    the interconnecting cable as short as possible and to choose a detector whose inherent capacitance is no larger than

    necessary. The rise time for charge-sensible preampliers also normally increases with input capacitance.

    For a wide assortment of applications, the noise level of commercially available preampliers is suciently low

    so that their contribution to the FWHM of the system response is small compared with the inherent contributions

    of the detector itself.

    Detector Bias Voltage

    Figura 3: two congurations for supplying detector bias through a

    preamplier. (a ) an ac-coupled arrangement, in which a coupling

    capacitor is provided between the detector and preamplier circuits.

    This allows interchanging dierent values of RL without aecting thepreamplier input. (b) A de-coupled conguration that eliminates

    the coupling capacitor and generally leads to better noise performan-

    ce for critical applications. Now the detector must be isolated from

    ground, and changing the bias resistor RL may aect the input stagecharacteristics.

    Another function normally carried out by

    the preamplier is to provide a means for

    supplying bias voltage to the detector. Ar-

    rangements can always be made to supply

    voltage to the detector independent of the

    preamplier, but it is usually convenient

    to do so through the preamplier. In Fig.

    3. two congurations are shown in whi-

    ch the bias voltage is supplied through a

    load resistance R

    L

    A single cable between

    the preamplier and detector provides bo-

    th the voltage to the detector and the si-

    gnal pulse to the input of the preamplier.

    The load resistance, together with the in-

    put capacitance, determines the time con-

    stant across which the detector current is

    collected. From the standpoint of mini-

    mum noise, R

    L

    should be as large as pos-

    sible. Practical limitations always dictate

    that its value be no more than a few thou-

    sand megohms due to the fact that any dc

    signal or leakage current drawn by the de-

    tector must also ow through this resistor.

    Leakage currents are especially troubleso-

    me for semiconductor diode detectors and,

    in those cases in which R

    L

    is large, can lead

    to a substantial dc voltage drop across R

    L

    .

    In that event, the voltage actually applied

    to the detector is less than the supply vol-

    tage, and experimenters must be aware of

    the magnitude of the leakage current so

    that they can compensate for the volta-

    ge drop across R

    L

    by raising the supply

    voltage.

    The best noise performance in present

    designs results from the use of a eld eect

    transistor (FET) as the input stage. FETs are notoriously sensitive to overvoltage transients and can easily be

    damaged by the transients that can be generated by switching a detector bias supply in coarse steps or abruptly

  • disconnecting or turning o the voltage. As a result, many commercial preampliers are provided with overvoltage

    protection circuits. For the ultimate in low noise performance, however, it is often necessary to switch out the

    protection circuits, and in such circumstances the bias supply must be changed only gradually and in a continuous

    fashion.

    Test Pulse Input

    Most preampliers are also provided with an input labeled test pulse, which is intended to receive the output

    of a pulse generator for system test purposes. Fig. 4 shows one means of applying this test pulse to the input

    stage of a charge-sensitive conguration. If the pulser is suciently stable, its output should be resolved into a

    single channel by a pulse height analyzer connected to the system output. Any broadening of this response into

    more than one channel can be attributed to the inherent electronic noise of the measurement system. It is often

    interesting to measure this inherent electronic noise width because any measured detector response can never be

    better resolved than this electronic limit. It is important to make this measurement with the detector attached to

    the input of the preamplier because its capacitance is often a signicant factor in the overall noise characteristics

    of the preamplier. The test pulse input provides a convenient means of carrying out this noise determination

    as well as simply checking the overall function of the signal-processing system prior to a measurement. In more

    sophisticated systems, a test pulse input may also be required during the course of a measurement for purposes of

    dead time determination or gain stabilization

    Figura 4: A charge sensitive preamplier that has been provided with a test pulse input. If a step voltage pulse

    of amplitude Vi is applied to this input, a charge equal to ViCi is supplied to the preamplier input stage. Rt is asmall-value termination resistance.

    DETECTOR BIAS AND HIGH-VOLTAGE SUPPLIES

    With few exceptions, virtually all radiation detectors require the application of an external high voltage for their

    proper operation. This voltage is conventionally called detector bias, and high-voltage supplies used for this purpose

    are often called detector bias supplies.

    Some characteristics of detector bias supplies which can be important in specic applications are the following:

    1. The maximum (and minimum) voltage level and its polarity.

    2. The maximum current available from the supply.

    3. The degree of regulation against long-term drifts due to changes in temperature or power line voltage.

    4. The degree of ltering provided to eliminate ripple at power line frequency or other low-frequency noise.

    The sophistication required of the bias supply varies greatly with the detector type. For detectors that draw very little

    current (such as an ion chamber) the bias supply can be as simple as a dry cell battery. On the other hand, supplies that

    must simultaneously provide high voltage and relatively high current involve a substantial amount of design engineering and

    can be among the heaviest and bulkiest of the equipment normally found in a nuclear instrumentation system.

    However, the degree of regulation and ltering is again important because any high-voltage uctuations appear superim-

    posed on the signal. Semiconductor diode detectors draw relatively little current and the voltage demands seldom exceed

    1000 V.

    The voltage level on most bias supplies is adjustable either through switching in steps 1 or by means of a continuously

    adjustable helipot. When used with preampliers with FET ; input stages, continuous adjustment avoids switching transients

    that may be potentially damaging to the FET. Some designs provide alternate protection by limiting the rate of change of

    the voltage between steps.

  • Figura 5: Elements of typical signal chain for pulse counting

    PULSE GENERATORS

    An electronic pulse generator is indispensable in the initial setup and calibration of virtually any nuclear instrumen-

    tation system. Furthermore, some methods of gain stabilization and dead time determination require the output

    of a pulse generator to be mixed with signal pulses during the course of a measurement. Pulsers are therefore a

    very common element in most radiation instrumentation systems.

    A tail pulse generator with adjustable rise and decay times is probably the most useful of all pulser types. Its

    output is conveniently fed to the test pulse input on preampliers or used directly in place of the preamplier

    output. If the output amplitude is truly constant, a measurement of the amplitude distribution recorded by the

    pulse analysis system determines the electronic noise level present in the system.

    Most pulse generators also provide a front panel adjustment for the amplitude of the pulse. In some designs,

    often called precision pulsers, this adjustment is accurately controlled by a front panel dial. Such pulsers can then

    be used to check the integral linearity of a pulse-handling system simply by recording the output amplitude for

    several dierent settings of the input pulse amplitude. The output of most precision pulsers is suciently stable

    so that, in the absence of electronic noise, all pulses of constant amplitude would be resolved into a single channel

    in a multichannel analysis system.

    For normal pulse generators, the interval between pulses is uniform and periodic. However, for determining

    some system parameters such as pileup behavior and other time-dependent phenomena, a periodic source does

    not adequately represent the random time spacing encountered from actual radiation detector pulses. Therefore,

    designs have evolved which provide a source of randomly spaced pulses of constant amplitude. In such random

    pulsers, the noise signal from an internal component is often used to trigger randomly the time at which a pulse is

    produced at the output.

    PULSE HEIGHT ANALYSIS SYSTEMS

    General Considerations

    Next to the simple counting of pulses, the most common procedure in nuclear measurements involves recording the

    amplitude distribution of pulses produced by a radiation detector. Most often the object is to deduce properties of

    the incident radiation from the position of peaks in the recorded spectrum, although other aspects of the spectrum

    may be of interest in dierent situations. The performance required of the instrument system used to record the

    pulse height spectrum is largely dependent on the inherent energy resolution of the detector. If the detector energy

    resolution is relatively poor, the requirements of the recording system are undemanding and easy to meet. On

    the other hand, detectors with good energy resolution require careful attention to the pulse processing system to

    assure that additional degradation of the resolution is minimized.

    A simple pulse height analysis system is shown in Fig. 6The key element in this signal chain is the linear

    amplier, which shapes the pulses from the preamplier and provides enough amplication to match the input span

    for which the multichannel analyzer has been designed. It is the shaping function of the linear amplier that often

    dominates the performance of the pulse-processing system. For high-resolution detectors, however, consideration

    must be given to the eect of various shaping methods on the pulse properties with regard to signal-to-noise and

    pileup.

  • Figura 6: Scheme of a rudimentary system for non critical applications .

    The strategies employed in choosing parameters for the pulse-processing system change drastically with the

    expected counting rate. If the counting rate is low (say 100 counts/s or less), the system can be optimized with

    regard to processing each individual pulse without much concern for interfering eects between pulses.

    1

    1

    A quick estimate of the importance of rate-related eects can be made by calculating the duty cycle obtained by multiplying the

    eective width of the shaped pulse by the rate. If this product is less than about 103, these eects should be minimal and can oftenbe neglected. A duty cycle of 102 is a moderate rate, whereas in high-rate situations it may approach 101. For a typical pulse widthof 5 p.s, the corresponding rates are 200, 2000, and 20,000 per second.

  • Figura 7: Amplier module